header advert
Results 1 - 1 of 1
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 63-B, Issue 1 | Pages 7 - 11
1 Feb 1981
Wardlaw D McLauchlan J Pratt D Bowker P

The off-loading characteristics of the cast-braces of 30 patients with fractures of the shaft of the femur have been investigated, during axial loading, using strain-gauge transducers. These were applied at the level of the fracture, where the cast was circumferentially split, and to the hinges of the brace at the knee. They measured the load transferred between the two portions of the thigh cast, and between the thigh cast as a whole and the below-knee cast; by subtraction from the total load on the limb, the skeletal force at the fracture level and at the knee could be calculated. In all patients there was an increase in the fracture load as union progressed which was thought to be due to physiological feedback mechanism from the fracture site. The load carried by the two portions of the thigh cast and by the thigh cast as a whole was proportionately high at first and stabilised at an average of 35 per cent of body weight.