Advertisement for orthosearch.org.uk
Results 1 - 4 of 4
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 6 | Pages 976 - 982
1 Nov 1998
Psychoyios V Crawford RW Murray DW O’Connor JJ

Many designs of unicompartmental knee replacement show early and mid-term failure due to polyethylene wear. We studied the wear rate of congruent polyethylene meniscal bearings retrieved from failed Oxford unicompartmental knee replacements.

We examined 16 bearings, 0.8 to 12.8 years after implantation, measuring their thickness and comparing it with that of 14 unused bearings. The mean rate of penetration, which included the effects of wear at both upper and lower surfaces, was 0.036 mm per year (maximum 0.08). Bearings as thin as 3.5 mm wore no faster than thicker models, but ten with evidence of impingement had greater wear. The six bearings with no impingement showed a mean rate of penetration of 0.01 mm per year.

In unicompartmental knee replacement, careful implantation of fully congruous meniscal bearings can avoid failure due to polyethylene wear.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 5 | Pages 934 - 934
1 Sep 1998
CRAWFORD RW


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 2 | Pages 279 - 281
1 Mar 1998
Crawford RW Gie GA Ling RSM Murray DW

We investigated 42 patients who were being considered for primary total hip arthroplasty (THA), but in whom it was uncertain whether the hip was the source of their pain. They were given an injection of local anaesthetic into the joint space.

Of 33 patients who gained pain relief from their injection, 32 subsequently had successful THA. The remaining patient has not had surgery. The intra-articular injection of local anaesthetic is thus at least 96% sensitive. Of the nine patients who had no or only minimal pain relief from injection, one has had an unsuccessful THA, three have been successfully treated for other conditions and five have unresolved pain for which no organic basis has been established.

We believe that the injection of local anaesthetic into the hip is a reliable test, with low morbidity. In difficult cases it will aid in the clarification of the cause of pain which possibly arises from the hip.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 5 | Pages 796 - 800
1 Sep 1997
Crawford RW Ellis AM Gie GA Ling RSM

We investigated 15 patients with painful hip arthroplasties using intra-articular injection of bupivicaine. Fourteen had pain relief and 13 of them were subsequently found to have loosening of one or both components. The relief of pain after total hip arthroplasty by intra-articular injection of bupivicaine indicates that a satisfactory result is probable after revision surgery with refixation of the components.