Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1603 - 1609
1 Dec 2013
Chana R Facek M Tilley S Walter WK Zicat B Walter WL

We report the clinical and radiological outcomes of a series of contemporary cementless ceramic-on-ceramic total hip replacements (THRs) at ten years in patients aged ≤ 55 years of age. Pre- and post-operative activity levels are described. A total of 120 consecutive ceramic cementless THRs were performed at a single centre in 110 patients from 1997 to 1999. The mean age of the patients at operation was 45 years (20 to 55). At ten years, four patients had died and six were lost to follow-up, comprising ten hips. The mean post-operative Harris hip score was 94.7 (55 to 100). Radiological analysis was undertaken in 90 available THRs of the surviving 106 hips at final review: all had evidence of stable bony ingrowth, with no cases of osteolysis. Wear was undetectable. There were four revisions. The survival for both components with revision for any cause as an endpoint was 96.5% (95% confidence interval 94.5 to 98.7). The mean modified University of California, Los Angeles activity level rose from a mean of 6.4 (4 to 10) pre-operatively to 9.0 (6 to 10) at the ten-year post-operative period.

Alumina ceramic-on-ceramic bearings in cementless primary THR in this series have resulted in good clinical and radiological outcomes with undetectable rates of wear and excellent function in the demanding younger patient group at ten years.

Cite this article: Bone Joint J 2013;95-B:1603–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 281 - 286
1 Feb 2012
Chana R Esposito C Campbell PA Walter WK Walter WL

We report a case of a male patient presenting with bilateral painful but apparently well-positioned and -fixed large-diameter metal-on-metal hip replacements four years post-operatively. Multiple imaging modes revealed a thick-walled, cystic expansile mass in communication with the hip joint (a pseudotumour). Implant retrieval analysis and tissue culture eliminated high bearing wear or infection as causes for the soft-tissue reaction, but noted marked corrosion of the modular neck taper adaptor and corrosion products in the tissues. Therefore, we believe corrosion products from the taper caused by mismatch of the implant components led to pseudotumour formation requiring revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 11 | Pages 1497 - 1502
1 Nov 2011
Chana R Salmon L Waller A Pinczewski L

We evaluated the safety and efficacy of total knee replacement in patients receiving continuous warfarin therapy.

We identified 24 consecutive patients receiving long-term warfarin therapy who underwent total knee replacement between 2006 and 2008 and compared them with a group of age- and gender-matched patients not on long-term anticoagulation. Primary observations were changes in haemoglobin, transfusion rates and complications. Secondary observations were fluctuations in the international normalised ratio (INR) and post-operative range of movement.

There was no significant difference between the two groups in pre- or post-operative haemoglobin, incidence of transfusion or incidence of post-operative complications. There were no surgical delays due to a high INR level. The mean change in INR during the peri-operative phase was minimal (mean 0.4; sd 0.7). There was no significant difference in the range of movement between the two groups after day two post-operatively.

Current American College of Chest Physicians guidelines recommend bridging therapy for high-risk patients receiving oral anticoagulation and undergoing major orthopaedic procedures. We have shown that a safe alternative is to continue the steady-state warfarin peri-operatively in patients on long-term anticoagulation requiring total knee replacement.