The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 6 | Pages 908 - 913
1 Nov 1997
Masterson EL Masri BA Duncan CP Rosenberg A Cabanela M Gross M

An analysis of the cement mantle obtained with the Exeter impaction allografting system at one centre showed that it was either deficient or absent in almost 47% of Gruen zones. We therefore examined the mantle obtained using this system at another hospital and compared the results with those from the CPT and Harris Precoat Systems at other centres.

The surgical indications for the procedure and the patient details were broadly similar in all four hospitals. There was some variation in the frequency of use of cortical strut allografts, cerclage wires and wire mesh to supplement the impaction allograft. Analysis of the cement mantles showed that when uncertain Gruen zones were excluded, the incidence of zones with areas of absence or deficiency of the cement was 47% and 50%, respectively, for the two centres using the Exeter system, 21% for the CPT system and 18% for the Harris Precoat system.

We measured the difference in size between the proximal allograft impactors and the definitive prosthesis for each system. The Exeter system impactors are shorter than the definitive prosthesis and taper sharply so that the cavity created is inadequate, especially distally. The CPT proximal impactors are considerably longer than the definitive prosthesis and are designed to give a mantle of approximately 2 mm medially and laterally and 1.5 mm anteriorly and posteriorly. The Harris Precoat proximal impactors allow for a mantle with a circumference of 0.75 mm in the smaller sizes and 1 mm in the larger.

Many reports link the longevity of a cemented implant to the adequacy of the cement mantle. For this reason, femoral impaction systems require careful design to achieve a cement mantle which is uninterrupted in its length and adequate in its thickness. Our results suggest that some current systems require modification.


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 6 | Pages 865 - 869
1 Nov 1995
McGrory B Morrey B Cahalan T An K Cabanela M

At a minimum of one year after operation, we studied 64 patients with 86 total hip arthroplasties (THA) by standard anteroposterior hip and pelvic radiographs and measurement of range of motion and of isometric abduction strength. The femoral offset correlated positively with the range of abduction (p = 0.046). Abduction strength correlated positively with both femoral offset (p = 0.0001) and the length of the abductor lever arm (p = 0.005). Using multiple regression, abduction strength correlated with height (p = 0.017), gender (p = 0.0005), range of flexion (p = 0.047) and the abductor lever arm (p = 0.060). Our findings suggest that greater femoral offset after THA allows both an increased range of abduction and greater abductor strength.


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 4 | Pages 575 - 578
1 Jul 1994
Berry D Barnes C Scott R Cabanela M Poss R

Ten cases are described of catastrophic failure of the polyethylene liner of three different designs of uncemented acetabular component. Failure occurred as a result of either 'wearthrough' to the metal backing, liner fracture or a combination of both, at a mean of 4.6 years after implantation (2 to 7.6). At revision there was metallosis in all hips and osteolysis of the femur or the pelvis in six. Catastrophic failure was seen only in cups with a minimum polyethylene thickness of less than 5 mm.


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 2 | Pages 252 - 257
1 Mar 1994
Ferguson G Cabanela M Ilstrup D

From 1969 through 1982, 305 hips in 290 patients had total hip arthroplasty for failed femoral intertrochanteric osteotomy. Of these, we reviewed 215 hips (70.5%) with a minimum follow-up of five years. The results were good or excellent in 79%, but there were technical problems at operation in 23% and a total perioperative complication rate of 11.8%. Late complications occurred in 13.1% including seven late infections (3.2%). At a mean follow-up of ten years, 39 hips had been revised (18.1%), there was probable loosening in 19.5% of stems and 12.6% of cups and possible loosening in 11.4% of stems and 7.2% of cups. The cumulative probability of failure at ten years was 20.6%. We recommend that intertrochanteric osteotomy be advised, planned and executed bearing in mind the possible need for a subsequent total hip arthroplasty, and that screws and plates should be routinely removed soon after union of the osteotomy.


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 4 | Pages 597 - 601
1 Jul 1993
Dean M Cabanela M

We reviewed 18 hips in 17 patients at a mean of five years after performing Sugioka's transtrochanteric anterior rotational osteotomy for avascular necrosis. The results were satisfactory in only three hips (17%). Twelve hips had been revised by hip replacement, revision was pending in one and two others were unsatisfactory. Hip replacement was not compromised by the previous Sugioka osteotomy. Fifteen hips (83%) had shown further collapse of the femoral head, and we conclude from this and from isotope scans that the osteotomy may have impaired the residual blood supply of the femoral head. It seems that ethnic origin may be a factor in the outcome of this procedure; we have abandoned its use.


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 6 | Pages 977 - 981
1 Nov 1991
Beyer C Cabanela M Berquist T

We treated 36 patients with unilateral facet dislocations or fracture-dislocations of the cervical spine at the Mayo Clinic between 1975 and 1986. Adequate records were available for 34: ten patients were treated by open reduction and posterior fusion, and 24 by nonoperative management. Of these, 19 had halo traction followed by halo-thoracic immobilisation, four had a simple cervicothoracic orthosis, and one received no active treatment. Anatomical reduction was achieved more frequently in the operative group (60% compared with 25%). Nonoperative treatment was more likely to result in cervical translation on flexion/extension lateral radiographs, and in significant symptoms. Only 36% of the patients treated by halo traction achieved anatomical alignment; in 25% halo traction failed to achieve or maintain any degree of reduction. During halo-thoracic immobilisation, half of the patients lost some degree of reduction and patient satisfaction with the appliance was low. Open reduction and internal fixation of unilateral facet injuries gave better results. 6