MAGnetic Expansion Control (MAGEC) rods are used in the surgical treatment of children with early onset scoliosis. The magnetically controlled lengthening mechanism enables rod distractions without the need for repeated invasive surgery. The CE certification of these devices was suspended in March 2021 due, primarily, to performance evidence gaps in the documents provided by the manufacturer to regulators and notified bodies. MAGEC rods are therefore not permitted for use in countries requiring CE marking. This was a survey of 18 MAGEC rod surgeons in the UK about their perception of the impact of the CE suspension on the clinical management of their patients. Unsurprisingly, virtually all perceived a negative impact, reflecting the complexity of this patient group. Reassuringly, these surgeons are highly experienced in alternative treatment methods. Cite this article: Abstract
We undertook a prospective non-randomised radiological study
to evaluate the preliminary results of using magnetically-controlled
growing rods (MAGEC System, Ellipse technology) to treat children
with early-onset scoliosis. Between January 2011 and January 2015, 19 children were treated
with magnetically-controlled growing rods (MCGRs) and underwent
distraction at three-monthly intervals. The mean age of our cohort
was 9.1 years (4 to 14) and the mean follow-up 22.4 months (5.1
to 35.2). Of the 19 children, eight underwent conversion from traditional growing
rods. Whole spine radiographs were carried out pre- and post-operatively:
image intensification was used during each lengthening in the outpatient
department. The measurements evaluated were Cobb angle, thoracic kyphosis,
proximal junctional kyphosis and spinal growth from T1 to S1.Aims
Patients and Methods