Aims. Total knee arthroplasty (TKA) is a common and safe orthopaedic procedure. Zimmer Biomet's NexGen is the second most popular brand of implant used in the UK. The primary cause of revision after the first year is aseptic loosening. We present our experience of using this implant, with significant concerns around its performance with regards early aseptic loosening of the tibial component. Methods. A retrospective, single-surgeon review was carried out of all of the NexGen Legacy Posterior Stabilized (LPS) TKAs performed in this institute. The specific model used for the index procedures was the NexGen Complete Knee System (Legacy Knee-Posterior Stabilized LPS-Flex Articular Surface, LPS-Flex Femoral Component Option, and Stemmed Nonaugmentable Tibial Component Option). Results. Between 2013 and 2016, 352 NexGen TKAs were carried out on 331 patients. A total of 62 TKAs have been revised to date, giving an all-cause revision rate of 17.6% at a minimum of five years. Three of these revisions were due to infection. Overall, 59 of the revisions were performed for aseptic loosening (16.7%) of the tibial component. The tibial component was removed intraoperatively without instrumentation due to significant tibial debonding between the implant-cement interface. Conclusion. While overall, we believe that early aseptic loosening is multi-factorial in nature, the significantly high aseptic revision rate, as seen by an experienced fellowship-trained
Aims. Our primary aim was to establish the proportion of female orthopaedic consultants who perform arthroplasty via cases submitted to the National Joint Registry (NJR), which covers England, Wales, Northern Ireland, the Isle of Man, and Guernsey. Secondary aims included comparing time since specialist registration, private practice participation, and number of hospitals worked in between male and female surgeons. Methods. Publicly available data from the NJR was extracted on the types of arthroplasty performed by each surgeon, and the number of procedures of each type undertaken. Each surgeon was cross-referenced with the General Medical Council (GMC) website, using GMC number to extract surgeon demographic data. These included sex, region of practice, and dates of full and specialist registration. Results. Of 2,895 surgeons contributing to the NJR in 2023, 102 (4%) were female. The highest proportions of female surgeons were among those who performed elbow (n = 25; 5%), shoulder (n = 24; 4%), and ankle (n = 8; 4%) arthroplasty. Hip (n = 66; 3%) and knee arthroplasty (n = 39; 2%) had the lowest female representation. Female surgeons had been practising for a median of 10.4 years since specialist registration compared to 13.7 years for males (p < 0.001). Northern Ireland was the region with the highest proportion of female
Aims. Total hip arthroplasty (THA) using the direct anterior approach (DAA) is undertaken with the patient in the supine position, creating an opportunity to replace both hips under one anaesthetic. Few studies have reported simultaneous bilateral DAA-THA. The aim of this study was to characterize a cohort of patients selected for this technique by a single, high-volume
Aims. Pelvic incidence (PI) is a position-independent spinopelvic parameter traditionally used by spinal surgeons to determine spinal alignment. Its relevance to the
Aims. Perthes’ disease (PD) often results in femoral head deformity and leg length discrepancy (LLD). Our objective was to analyze femoral morphology in PD patients at skeletal maturity to assess where the LLD originates, and evaluate the effect of contralateral epiphysiodesis for length equalization on proximal and subtrochanteric femoral lengths. Methods. All patients treated for PD in our institution between January 2013 and June 2020 were reviewed retrospectively. Patients with unilateral PD, LLD of ≥ 5 mm, and long-leg standing radiographs at skeletal maturity were included. Total leg length, femoral and tibial length, articulotrochanteric distance (ATD), and subtrochanteric femoral length were compared between PD side and the unaffected side. Furthermore, we compared leg length measurements between patients who did and who did not have a contralateral epiphysiodesis. Results. Overall, 79 patients were included, of whom 21 underwent contralateral epiphysiodesis for leg length correction. In the complete cohort, the mean LLD was 1.8 cm (95% confidence interval (CI) 1.5 to 2.0), mean ATD difference was 1.8 cm (95% CI -2.1 to -1.9), and mean subtrochanteric difference was -0.2 cm (95% CI -0.4 to 0.1). In the epiphysiodesis group, the mean LLD before epiphysiodesis was 2.7 cm (95% CI 1.3 to 3.4) and 1.3 cm (95% CI -0.5 to 3.8) at skeletal maturity. In the nonepiphysiodesis group the mean LLD was 2.0 cm (95% CI 0.5 to 5.1; p = 0.016). The subtrochanteric region on the PD side was significantly longer at skeletal maturity in the epiphysiodesis group compared to the nonepiphysiodesis group (-1.0 cm (95% CI -2.4 to 0.6) vs 0.1 cm (95% CI -1.0 to 2.1); p < 0.001). Conclusion. This study demonstrates that LLD after PD originates from the proximal segment only. In patients who had contralateral epiphysiodesis to balance leg length, this is achieved by creating a difference in subtrochanteric length.
Aims. Traditionally, acetabular component insertion during total hip arthroplasty (THA) is visually assisted in the posterior approach and fluoroscopically assisted in the anterior approach. The present study examined the accuracy of a new surgeon during anterior (NSA) and posterior (NSP) THA using robotic arm-assisted technology compared to two experienced surgeons using traditional methods. Methods. Prospectively collected data was reviewed for 120 patients at two institutions. Data were collected on the first 30 anterior approach and the first 30 posterior approach surgeries performed by a newly graduated
Obesity is an epidemic across both the developed
and developing nations that is possibly the most important current
public health factor affecting the morbidity and mortality of the global
population. Obese patients have the potential to pose several challenges
for
Aims. Although good clinical outcomes have been reported for monolithic tapered, fluted, titanium stems (TFTS), early results showed high rates of subsidence. Advances in stem design may mitigate these concerns. This study reports on the use of a current monolithic TFTS for a variety of indications. Methods. A multi-institutional retrospective study of all consecutive total hip arthroplasty (THA) and revision total hip arthroplasty (rTHA) patients who received the monolithic TFTS was conducted. Surgery was performed by eight fellowship-trained
Aims. The aim of this study was to determine the general postoperative opioid consumption and rate of appropriate disposal of excess opioid prescriptions in patients undergoing primary unilateral total knee arthroplasty (TKA). Patients and Methods. In total, 112 patients undergoing surgery with one of eight
The aims of this study were to determine the incidence and factors for developing periprosthetic joint infection (PJI) following hemiarthroplasty (HA) for hip fracture, and to evaluate treatment outcome and identify factors associated with treatment outcome. A retrospective review was performed of consecutive patients treated for HA PJI at a tertiary referral centre with a mean 4.5 years’ follow-up (1.6 weeks to 12.9 years). Surgeries performed included debridement, antibiotics, and implant retention (DAIR) and single-stage revision. The effect of different factors on developing infection and treatment outcome was determined.Aims
Methods
Blood transfusion and postoperative anaemia are complications of total knee arthroplasty (TKA) that are associated with substantial healthcare costs, morbidity, and mortality. There are few data from large datasets on the risk factors for these complications. We retrospectively reviewed the records of TKA patients from a single tertiary care institution from February 2016 to December 2020. There were a total of 14,901 patients in this cohort with a mean age of 67.9 years (SD 9.2), and 5,575 patients (37.4%) were male. Outcomes included perioperative blood transfusion and postoperative anaemia, defined a priori as haemoglobin level < 10 g/dl measured on the first day postoperatively. In order to establish a preoperative haemoglobin cutoff, we investigated a preoperative haemoglobin level that would limit transfusion likelihood to ≤ 1% (13 g/dl) and postoperative anaemia likelihood to 4.1%. Risk factors were assessed through multivariable Poisson regression modelling with robust error variance.Aims
Methods
A novel enhanced cement fixation (EF) tibial implant with deeper cement pockets and a more roughened bonding surface was released to market for an existing total knee arthroplasty (TKA) system.This randomized controlled trial assessed fixation of the both the EF (ATTUNE S+) and standard (Std; ATTUNE S) using radiostereometric analysis. Overall, 50 subjects were randomized (21 EF-TKA and 23 Std-TKA in the final analysis), and had follow-up visits at six weeks, and six, 12, and 24 months to assess migration of the tibial component. Low viscosity bone cement with tobramycin was used in a standardized fashion for all subjects. Patient-reported outcome measure data was captured at preoperative and all postoperative visits.Aims
Methods
Understanding spinopelvic mechanics is important for the success of total hip arthroplasty (THA). Despite significant advancements in appreciating spinopelvic balance, numerous challenges remain. It is crucial to recognize the individual variability and postoperative changes in spinopelvic parameters and their consequential impact on prosthetic component positioning to mitigate the risk of dislocation and enhance postoperative outcomes. This review describes the integration of advanced diagnostic approaches, enhanced technology, implant considerations, and surgical planning, all tailored to the unique anatomy and biomechanics of each patient. It underscores the importance of accurately predicting postoperative spinopelvic mechanics, selecting suitable imaging techniques, establishing a consistent nomenclature for spinopelvic stiffness, and considering implant-specific strategies. Furthermore, it highlights the potential of artificial intelligence to personalize care. Cite this article:
This study aimed to investigate patients’ attitudes towards day-case hip and knee arthroplasty and to describe patient characteristics associated with different attitudes, with the purpose of providing an insight into the information requirements for patients that surgeons should address when informing patients about day-case surgery. A total of 5,322 patients scheduled for hip or knee arthroplasty between 2016 and 2022 were included in the study. Preoperatively, patients were asked if they were interested in day-case surgery (‘Yes’, ‘Do not know’, ‘No’). Patient demographics including age, BMI, sex, and patient-reported outcome measures (PROMs) such as the EuroQol five-dimension three-level questionnaire (EQ-5D-3L) were examined within each attitude group. Additionally, changes in attitude were assessed among patients who had completed the questionnaire in association with prior hip or knee arthroplasty.Aims
Methods
The aim of this study was to compare the pattern of initial fixation and changes in periprosthetic bone mineral density (BMD) between patients who underwent total hip arthroplasty (THA) using a traditional fully hydroxyapatite (HA)-coated stem (T-HA group) and those with a newly introduced fully HA-coated stem (N-HA group). The study included 36 patients with T-HA stems and 30 with N-HA stems. Dual-energy X-ray absorptiometry was used to measure the change in periprosthetic BMD, one and two years postoperatively. The 3D contact between the stem and femoral cortical bone was evaluated using a density-mapping system, and clinical assessment, including patient-reported outcome measurements, was recorded.Aims
Methods
Precise implant positioning, tailored to individual spinopelvic biomechanics and phenotype, is paramount for stability in total hip arthroplasty (THA). Despite a few studies on instability prediction, there is a notable gap in research utilizing artificial intelligence (AI). The objective of our pilot study was to evaluate the feasibility of developing an AI algorithm tailored to individual spinopelvic mechanics and patient phenotype for predicting impingement. This international, multicentre prospective cohort study across two centres encompassed 157 adults undergoing primary robotic arm-assisted THA. Impingement during specific flexion and extension stances was identified using the virtual range of motion (ROM) tool of the robotic software. The primary AI model, the Light Gradient-Boosting Machine (LGBM), used tabular data to predict impingement presence, direction (flexion or extension), and type. A secondary model integrating tabular data with plain anteroposterior pelvis radiographs was evaluated to assess for any potential enhancement in prediction accuracy.Aims
Methods
As an increasing number of female surgeons are choosing orthopaedics, it is important to recognize the impact of pregnancy within this cohort. The aim of this review was to examine common themes and data surrounding pregnancy, parenthood, and fertility within orthopaedics. A systematic review was conducted by searching Medline, Emcare, Embase, PsycINFO, OrthoSearch, and the Cochrane Library in November 2022. The Preferred Reporting Items for Systematic Reviews and Meta Analysis were adhered to. Original research papers that focused on pregnancy and/or parenthood within orthopaedic surgery were included for review.Aims
Methods
Modular dual-mobility (DM) articulations are increasingly used during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS)-MRI abnormalities with serum metal ion levels in patients with DM articulations. A total of 45 patients (50 hips) with a modular DM articulation were included with mean follow-up of 3.7 years (SD 1.2). Enrolled patients with an asymptomatic, primary THA and DM articulation with over two years’ follow-up underwent MARS-MRI. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.Aims
Methods
The primary objective of this study was to develop a validated classification system for assessing iatrogenic bone trauma and soft-tissue injury during total hip arthroplasty (THA). The secondary objective was to compare macroscopic bone trauma and soft-tissues injury in conventional THA (CO THA) versus robotic arm-assisted THA (RO THA) using this classification system. This study included 30 CO THAs versus 30 RO THAs performed by a single surgeon. Intraoperative photographs of the osseous acetabulum and periacetabular soft-tissues were obtained prior to implantation of the acetabular component, which were used to develop the proposed classification system. Interobserver and intraobserver variabilities of the proposed classification system were assessed.Aims
Methods
Periprosthetic fractures (PPFs) following hip arthroplasty are complex injuries. This study evaluates patient demographic characteristics, management, outcomes, and risk factors associated with PPF subtypes over a decade. Using a multicentre collaborative study design, independent of registry data, we identified adults from 29 centres with PPFs around the hip between January 2010 and December 2019. Radiographs were assessed for the Unified Classification System (UCS) grade. Patient and injury characteristics, management, and outcomes were compared between UCS grades. A multinomial logistic regression was performed to estimate relative risk ratios (RRR) of variables on UCS grade.Aims
Methods