The Oxford unicompartmental knee replacement (UKR) was designed to minimise wear utilising a fully-congruent, mobile, polyethylene bearing. Wear of polyethylene is a significant cause of revision surgery in UKR in the first decade, and the incidence increases in the second decade. Our study used model-based radiostereometric analysis to measure the combined wear of the upper and lower bearing surfaces in 13 medial-compartment Oxford UKRs at a mean of 20.9 years (17.2 to 25.9) post-operatively. The mean linear penetration of the polyethylene bearing was 1.04 mm (0.307 to 2.15), with a mean annual wear rate of 0.045 mm/year (0.016 to 0.099). The annual wear rate of the phase-2 bearings (mean 0.022 mm/year) was significantly less (p = 0.01) than that of phase-1 bearings (mean 0.07 mm/year). The linear wear rate of the Oxford UKR remains very low into the
The aim of this study was to determine the polyethylene wear rate of Phase 3 Oxford Unicompartmental Knee Replacement bearings and to investigate the effects of resin type and manufacturing process. A total of 63 patients with at least ten years’ follow-up with three bearing types (1900 resin machined, 1050 resin machined, and 1050 resin moulded) were recruited. Patients underwent full weight-bearing model-based radiostereometric analysis to determine the bearing thickness. The linear wear rate was estimated from the change in thickness divided by the duration of implantation.Objectives
Methods
There have been several reports of good survivorship and excellent function at ten years with fixed-bearing unicompartmental knee replacement. However, little is known about survival beyond ten years. From the Bristol database of over 4000 knee replacements, we identified 203 St Georg Sled unicompartmental knee replacements (174 patients) which had already survived ten years. The mean age of the patients at surgery was 67.1 years (35.7 to 85) with 67 (38.5%) being under 65 years at the time of surgery. They were reviewed at a mean of 14.8 years (10 to 29.4) from surgery to determine survivorship and function. There were 99 knees followed up for 15 years, 21 for 20 years and four for 25 years. The remainder failed, were withdrawn, or the patient had died. In 58 patients (69 knees) the implant was The mean Bristol knee score of the surviving knees fell from 86 (34 to 100) to 79 (42 to 100) during the second decade. Survivorship to 20 years was 85.9% (95% CI 82.9% to 88.9%) and at 25 years was 80% (95% CI 70.2% to 89.8%). Satisfactory survival of a fixed-bearing unicompartmental knee replacement can be achieved into the second decade and beyond.