The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 3 | Pages 354 - 357
1 Apr 2003
Potter D Claydon P Stanley D

Between 1993 and 1996, we undertook 35 Kudo 5 total elbow replacements in a consecutive series of 31 rheumatoid patients. A total of 25 patients (29 procedures) was evaluated at a mean follow-up of six years (5 to 7.5) using the Mayo Clinic performance index. In addition, all patients were assessed for loosening using standard anteroposterior and lateral radiographs. At review, 19 elbows (65%) had either no pain or mild pain, ten (35%) had moderate pain and none had severe pain. The mean arc of flexion/extension was 94° (35 to 130) and supination/pronation was 128° (30 to 165). A fracture of the medial epicondyle occurred during surgery in one patient. This was successfully treated with a single AO screw and a standard Kudo 5 implant was inserted. Postoperatively, there were no infections. One patient had a dislocation which was treated by closed reduction and five had neurapraxia of the ulnar nerve. Radiologically, there was no evidence of loosening of the humeral component, but two ulnar components had progressive radiolucent lines suggestive of loosening. Two other ulnar components had incomplete and non-progressive radiolucent lines. With definite radiological loosening as the endpoint, the probability of survival of the Kudo 5 prosthesis at five years using the Kaplan-Meier method was 89%


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 387 - 392
1 Mar 2010
Kasten P Pape G Raiss P Bruckner T Rickert M Zeifang F Loew M

We have investigated the mid-term outcome of total shoulder replacement using a keeled cemented glenoid component and a modern cementing technique with regard to the causes of failure and loosening of the components.

Between 1997 and 2003 we performed 96 total shoulder replacements on 88 patients, 24 men and 64 women with a mean age of 69.7 years (31 to 82). The minimum follow-up was five years and at the time of review 87 shoulders (77 patients) were examined at a mean follow-up of 89.1 months (60 to 127). Cumulative survival curves were generated with re-operations (accomplished and planned), survivorship of the proshesis, loosening of the glenoid (defined as tilt > 5° or subsidence > 5 mm), the presence of radiolucent lines and a Constant score of < 30 as the endpoints.

There were two re-operations not involving revision of the implants and the survival rate of the prosthesis was 100.0% for the follow-up period, with an absolute Constant score of > 30 as the endpoint the survival rate was 98%. Radiological glenoid loosening was 9% after five years, and 33% after nine years. There was an incidence of 8% of radiolucent lines in more than three of six zones in the immediate post-operative period, of 37.0% after the first year which increased to 87.0% after nine years. There was no correlation between the score of Boileau and the total Constant score at the latest follow-up, but there was correlation between glenoid loosening and pain (p = 0.001).

We found that total shoulder replacement had an excellent mid-term survivorship and clinical outcome. The surgical and cementing techniques were related to the decrease in radiolucent lines around the glenoid compared with earlier studies. One concern, however, was the fact that radiolucent lines increased over time and there was a rate of glenoid loosening of 9% after five years and 33% after nine years. This suggests that the design of the glenoid component, and the implantation and cementing techniques may need further improvement.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 343 - 348
1 Mar 2008
Prasad N Dent C

We analysed the outcome of the Coonrad-Morrey total elbow replacement used for fracture of the distal humerus in elderly patients with no evidence of inflammatory arthritis and compared the results for early versus delayed treatment. We studied a total of 32 patients with 15 in the early treatment group and 17 in the delayed treatment group. The mean follow-up was 56.1 months (18 to 88). The percentage of excellent to good results based on the Mayo elbow performance score was not significantly different, 84% in the early group and 79% in the delayed group. Subjective satisfaction was 92% in both the groups.

One patient in the early group developed chronic regional pain syndrome and another type 4 aseptic loosening. Two elbows in the early group also showed type 1 radiological loosening.

Two patients in the delayed group had an infection, two an ulnar nerve palsy, one developed heterotopic ossification and one type 4 aseptic loosening. Two elbows in this group also showed type 1 radiological loosening. The Kaplan-Meier survivorship analysis for the early and delayed treatment groups was 93% at 88 months and 76% at 84 months, respectively.

No significant difference was found between the two groups.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1460 - 1463
1 Nov 2006
Landor I Vavrik P Jahoda D Guttler K Sosna A

We assessed the long-term results of 58 Souter-Strathclyde total elbow replacements in 49 patients with rheumatoid arthritis. The mean length of follow-up was 9.5 years (0.7 to 16.7). The mean pre-operative Mayo Elbow Performance Score was 30 (15 to 80) and at final follow-up was 82 (60 to 95). A total of 13 elbows (22.4%) were revised, ten (17.2%) for aseptic loosening, one (1.7%) for instability, one (1.7%) for secondary loosening after fracture, and one elbow (1.7%) was removed because of deep infection. The Kaplan-Meier survival rate was 70% and 53% at ten and 16 years, respectively. Failure of the ulnar component was found to be the main problem in relation to the loosening. Anterior transposition of the ulnar nerve had no influence on ulnar nerve paresthaesiae in these patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 75 - 81
1 Jan 2009
Cil A Veillette CJH Sanchez-Sotelo J Sperling JW Schleck C Cofield RH

Between 1976 and 2004, 38 revision arthroplasties (35 patients) were performed for aseptic loosening of the humeral component. The mean interval from primary arthroplasty to revision was 7.1 years (0.4 to 16.6). A total of 35 shoulders (32 patients) were available for review at a mean follow-up of seven years (2 to 19.3).

Pre-operatively, 34 patients (97%) had moderate or severe pain; at final follow-up, 29 (83%) had no or only mild pain (p < 0.0001). The mean active abduction improved from 88° to 107° (p < 0.01); and the mean external rotation from 37° to 46° (p = 0.27). Excellent or satisfactory results were achieved in 25 patients (71%) according to the modified Neer rating system. Humeral components were cemented in 29, with ingrowth implants used in nine cases. There were 19 of standard length and 17 were longer (two were custom replacements and are not included). Bone grafting was required for defects in 11 humeri. Only two glenoid components were left unrevised. Intra-operative complications included cement extrusion in eight cases, fracture of the shaft of the humerus is two and of the tuberosity in four. There were four re-operations, one for recurrent humeral loosening, with 89% survival free of re-operations at ten years.

Revision surgery for aseptic loosening of the humeral component provides reliable pain relief and modest improvement of movement, although there is a substantial risk of intra-operative complications. Revision to a total shoulder replacement gives better results than to a hemiarthroplasty.