Advertisement for orthosearch.org.uk
Results 1 - 20 of 27
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1296 - 1300
1 Sep 2015
Jauregui JJ Bor N Thakral R Standard SC Paley D Herzenberg JE

External fixation is widely used in orthopaedic and trauma surgery. Infections around pin or wire sites, which are usually localised, non-invasive, and are easily managed, are common. Occasionally, more serious invasive complications such as necrotising fasciitis (NF) and toxic shock syndrome (TSS) may occur.

We retrospectively reviewed all patients who underwent external fixation between 1997 and 2012 in our limb lengthening and reconstruction programme. A total of eight patients (seven female and one male) with a mean age of 20 years (5 to 45) in which pin/wire track infections became limb- or life-threatening were identified. Of these, four were due to TSS and four to NF. Their management is described. A satisfactory outcome was obtained with early diagnosis and aggressive medical and surgical treatment.

Clinicians caring for patients who have external fixation and in whom infection has developed should be aware of the possibility of these more serious complications. Early diagnosis and aggressive treatment are required in order to obtain a satisfactory outcome.

Cite this article: Bone Joint J 2015;97-B:1296–1300.


The Bone & Joint Journal
Vol. 106-B, Issue 3 | Pages 293 - 302
1 Mar 2024
Vogt B Lueckingsmeier M Gosheger G Laufer A Toporowski G Antfang C Roedl R Frommer A

Aims. As an alternative to external fixators, intramedullary lengthening nails (ILNs) can be employed for distraction osteogenesis. While previous studies have demonstrated that typical complications of external devices, such as soft-tissue tethering, and pin site infection can be avoided with ILNs, there is a lack of studies that exclusively investigated tibial distraction osteogenesis with motorized ILNs inserted via an antegrade approach. Methods. A total of 58 patients (median age 17 years (interquartile range (IQR) 15 to 21)) treated by unilateral tibial distraction osteogenesis for a median leg length discrepancy of 41 mm (IQR 34 to 53), and nine patients with disproportionate short stature treated by bilateral simultaneous tibial distraction osteogenesis, with magnetically controlled motorized ILNs inserted via an antegrade approach, were retrospectively analyzed. The median follow-up was 37 months (IQR 30 to 51). Outcome measurements were accuracy, precision, reliability, bone healing, complications, and patient-reported outcome assessed by the Limb Deformity-Scoliosis Research Society Score (LD-SRS-30). Results. A median tibial distraction of 44 mm (IQR 31 to 49) was achieved with a mean distraction index of 0.5 mm/day (standard deviation 0.13) and median consolidation index of 41.2 days/cm (IQR 34 to 51). Accuracy, precision, and reliability were 91%, 92%, and 97%, respectively. New temporary range of motion limitations occurred in 51% of segments (34/67). Distraction-related equinus deformity treated by Achilles tendon lengthening was the most common major complication recorded in 16% of segments (11/67). In 95% of patients (55/58) the distraction goal was achieved with 42% unplanned additional interventions per segment (28/67). The median postoperative LD-SRS-30 score was 4.0 (IQR 3.6 to 4.3). Conclusion. Tibial distraction osteogenesis using motorized ILNs inserted via an antegrade approach appears to be a reliable and precise procedure. Temporary joint stiffness of the knee or ankle should be expected in up to every second patient. A high rate and wide range of complications of variable severity should be anticipated. Cite this article: Bone Joint J 2024;106-B(3):293–302


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 146 - 152
1 Jan 2010
Bilen FE Kocaoglu M Eralp L Balci HI

We report the results of using a combination of fixator-assisted nailing with lengthening over an intramedullary nail in patients with tibial deformity and shortening. Between 1997 and 2007, 13 tibiae in nine patients with a mean age of 25.4 years (17 to 34) were treated with a unilateral external fixator for acute correction of deformity, followed by lengthening over an intramedullary nail with a circular external fixator applied at the same operating session. At the end of the distraction period locking screws were inserted through the intramedullary nail and the external fixator was removed. The mean amount of lengthening was 5.9 cm (2 to 8). The mean time of external fixation was 90 days (38 to 265). The mean external fixation index was 15.8 days/cm (8.9 to 33.1) and the mean bone healing index was 38 days/cm (30 to 60). One patient developed an equinus deformity which responded to stretching and bracing. Another developed a drop foot due to a compartment syndrome, which was treated by fasciotomy. It recovered in three months. Two patients required bone grafting for poor callus formation. We conclude that the combination of fixator-assisted nailing with lengthening over an intramedullary nail can reduce the overall external fixation time and prevent fractures and deformity of the regenerated bone


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1394 - 1399
1 Oct 2009
Oh C Song H Kim J Choi J Min W Park B

Ten patients, who were unsuitable for limb lengthening over an intramedullary nail, underwent lengthening with a submuscular locking plate. Their mean age at operation was 18.5 years (11 to 40). After fixing a locking plate submuscularly on the proximal segment, an external fixator was applied to lengthen the bone after corticotomy. Lengthening was at 1 mm/day and on reaching the target length, three or four screws were placed in the plate in the distal segment and the external fixator was removed. All patients achieved the pre-operative target length at a mean of 4.0 cm (3.2 to 5.5). The mean duration of external fixation was 61.6 days (45 to 113) and the mean external fixation index was 15.1 days/cm (13.2 to 20.5), which was less than one-third of the mean healing index (48 days/cm (41.3 to 55). There were only minor complications. Lengthening with a submuscular locking plate can successfully permit early removal of the fixator with fewer complications and is a useful alternative in children or when nailing is difficult


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 283 - 288
1 Feb 2017
Hughes A Heidari N Mitchell S Livingstone J Jackson M Atkins R Monsell F

Aims. Computer hexapod assisted orthopaedic surgery (CHAOS), is a method to achieve the intra-operative correction of long bone deformities using a hexapod external fixator before definitive internal fixation with minimally invasive stabilisation techniques. The aims of this study were to determine the reliability of this method in a consecutive case series of patients undergoing femoral deformity correction, with a minimum six-month follow-up, to assess the complications and to define the ideal group of patients for whom this treatment is appropriate. Patients and Methods. The medical records and radiographs of all patients who underwent CHAOS for femoral deformity at our institution between 2005 and 2011 were retrospectively reviewed. Records were available for all 55 consecutive procedures undertaken in 49 patients with a mean age of 35.6 years (10.9 to 75.3) at the time of surgery. Results. Patients were assessed at a mean interval of 44 months (6 to 90) following surgery. The indications were broad; the most common were vitamin D resistant rickets (n = 10), growth plate arrest (n = 6) and post-traumatic deformity (n = 20). Multi-planar correction was required in 33 cases. A single level osteotomy was performed in 43 cases. Locking plates were used to stabilise the osteotomy in 33 cases and intramedullary nails in the remainder. Complications included two nonunions, one death, one below-knee deep vein thrombosis, one deep infection and one revision procedure due to initial under-correction. There were no neurovascular injuries or incidence of compartment syndrome. Conclusion. This is the largest reported series of femoral deformity corrections using the CHAOS technique. This series demonstrates that precise intra-operative realignment is possible with a hexapod external fixator prior to definitive stabilisation with contemporary internal fixation. This combination allows reproducible correction of complex femoral deformity from a wide variety of diagnoses and age range with a low complication rate. Cite this article: Bone Joint J 2017;99-B:283–8


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 88 - 96
1 Jan 2023
Vogt B Rupp C Gosheger G Eveslage M Laufer A Toporowski G Roedl R Frommer A

Aims

Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences.

Methods

A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic junction and locking bolts at four different stages.


The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1095 - 1100
1 Sep 2022
McNally MA Ferguson JY Scarborough M Ramsden A Stubbs DA Atkins BL

Aims

Excision of chronic osteomyelitic bone creates a dead space which must be managed to avoid early recurrence of infection. Systemic antibiotics cannot penetrate this space in high concentrations, so local treatment has become an attractive adjunct to surgery. The aim of this study was to present the mid- to long-term results of local treatment with gentamicin in a bioabsorbable ceramic carrier.

Methods

A prospective series of 100 patients with Cierny-Mader Types III and IV chronic ostemyelitis, affecting 105 bones, were treated with a single-stage procedure including debridement, deep tissue sampling, local and systemic antibiotics, stabilization, and immediate skin closure. Chronic osteomyelitis was confirmed using strict diagnostic criteria. The mean follow-up was 6.05 years (4.2 to 8.4).


Bone & Joint Open
Vol. 4, Issue 8 | Pages 643 - 651
24 Aug 2023
Langit MB Tay KS Al-Omar HK Barlow G Bates J Chuo CB Muir R Sharma H

Aims

The standard of wide tumour-like resection for chronic osteomyelitis (COM) has been challenged recently by adequate debridement. This paper reviews the evolution of surgical debridement for long bone COM, and presents the outcome of adequate debridement in a tertiary bone infection unit.

Methods

We analyzed the retrospective record review from 2014 to 2020 of patients with long bone COM. All were managed by multidisciplinary infection team (MDT) protocol. Adequate debridement was employed for all cases, and no case of wide resection was included.


Bone & Joint Open
Vol. 3, Issue 5 | Pages 359 - 366
1 May 2022
Sadekar V Watts AT Moulder E Souroullas P Hadland Y Barron E Muir R Sharma HK

Aims

The timing of when to remove a circular frame is crucial; early removal results in refracture or deformity, while late removal increases the patient morbidity and delay in return to work. This study was designed to assess the effectiveness of a staged reloading protocol. We report the incidence of mechanical failure following both single-stage and two stage reloading protocols and analyze the associated risk factors.

Methods

We identified consecutive patients from our departmental database. Both trauma and elective cases were included, of all ages, frame types, and pathologies who underwent circular frame treatment. Our protocol is either a single-stage or two-stage process implemented by defunctioning the frame, in order to progressively increase the weightbearing load through the bone, and promote full loading prior to frame removal. Before progression, through the process we monitor patients for any increase in pain and assess radiographs for deformity or refracture.


Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims

This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported.

Methods

This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed.


Bone & Joint Open
Vol. 2, Issue 9 | Pages 705 - 709
1 Sep 2021
Wright J Timms A Fugazzotto S Goodier D Calder P

Aims

Patients undergoing limb reconstruction surgery often face a challenging and lengthy process to complete their treatment journey. The majority of existing outcome measures do not adequately capture the patient-reported outcomes relevant to this patient group in a single measure. Following a previous systematic review, the Stanmore Limb Reconstruction Score (SLRS) was designed with the intent to address this need for an effective instrument to measure patient-reported outcomes in limb reconstruction patients. We aim to assess the face validity of this score in a pilot study.

Methods

The SLRS was designed following structured interviews with several groups including patients who have undergone limb reconstruction surgery, limb reconstruction surgeons, specialist nurses, and physiotherapists. This has subsequently undergone further adjustment for language and clarity. The score was then trialled on ten patients who had undergone limb reconstruction surgery, with subsequent structured questioning to understand the perceived suitability of the score.


Bone & Joint Open
Vol. 2, Issue 8 | Pages 599 - 610
1 Aug 2021
Hothi H Bergiers S Henckel J Iliadis AD Goodier WD Wright J Skinner J Calder P Hart AJ

Aims

The aim of this study was to present the first retrieval analysis findings of PRECICE STRYDE intermedullary nails removed from patients, providing useful information in the post-market surveillance of these recently introduced devices.

Methods

We collected ten nails removed from six patients, together with patient clinical data and plain radiograph imaging. We performed macro- and microscopic analysis of all surfaces and graded the presence of corrosion using validated semiquantitative scoring methods. We determined the elemental composition of surface debris using energy dispersive x-ray spectroscopy (EDS) and used metrology analysis to characterize the surface adjacent to the extendable junctions.


Bone & Joint Open
Vol. 1, Issue 10 | Pages 645 - 652
19 Oct 2020
Sheridan GA Hughes AJ Quinlan JF Sheehan E O'Byrne JM

Aims

We aim to objectively assess the impact of COVID-19 on mean total operative cases for all indicative procedures (as outlined by the Joint Committee on Surgical Training (JCST)) experienced by orthopaedic trainees in the deanery of the Republic of Ireland. Subjective experiences were reported for each trainee using questionnaires.

Methods

During the first four weeks of the nationwide lockdown due to COVID-19, the objective impact of the pandemic on each trainee’s surgical caseload exposure was assessed using data from individual trainee logbook profiles in the deanery of the Republic of Ireland. Independent predictor variables included the trainee grade (ST 3 to 8), the individual trainee, the unit that the logbook was reported from, and the year in which the logbook was recorded. We used the analysis of variance (ANOVA) test to assess for any statistically significant predictor variables. The subjective experience of each trainee was captured using an electronic questionnaire.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1168 - 1176
1 Sep 2019
Calder PR McKay JE Timms AJ Roskrow T Fugazzotto S Edel P Goodier WD

Aims

The Precice intramedullary limb-lengthening system has demonstrated significant benefits over external fixation lengthening methods, leading to a paradigm shift in limb lengthening. This study compares outcomes following antegrade and retrograde femoral lengthening in both adolescent and adult patients.

Patients and Methods

A retrospective review of prospectively collected data was undertaken of a consecutive series of 107 femoral lengthening operations in 92 patients. In total, 73 antegrade nails and 34 retrograde nails were inserted. Outcome was assessed by the regenerate healing index (HI), hip and knee range of movement (ROM), and the presence of any complications.


Bone & Joint Open
Vol. 1, Issue 6 | Pages 229 - 235
9 Jun 2020
Lazizi M Marusza CJ Sexton SA Middleton RG

Aims

Elective surgery has been severely curtailed as a result of the COVID-19 pandemic. There is little evidence to guide surgeons in assessing what processes should be put in place to restart elective surgery safely in a time of endemic COVID-19 in the community.

Methods

We used data from a stand-alone hospital admitting and operating on 91 trauma patients. All patients were screened on admission and 100% of patients have been followed-up after discharge to assess outcome.


The Bone & Joint Journal
Vol. 101-B, Issue 2 | Pages 178 - 188
1 Feb 2019
Chaudhary MM Lakhani PH

Aims

Double-level lengthening, bone transport, and bifocal compression-distraction are commonly undertaken using Ilizarov or other fixators. We performed double-level fixator-assisted nailing, mainly for the correction of deformity and lengthening in the same segment, using a straight intramedullary nail to reduce the time in a fixator.

Patients and Methods

A total of 23 patients underwent this surgery, involving 27 segments (23 femora and four tibiae), over a period of ten years. The most common indication was polio in ten segments and rickets in eight; 20 nails were inserted retrograde and seven antegrade. A total of 15 lengthenings were performed in 11 femora and four tibiae, and 12 double-level corrections of deformity without lengthening were performed in the femur. The mean follow-up was 4.9 years (1.1 to 11.4). Four patients with polio had tibial lengthening with arthrodesis of the ankle. We compared the length of time in a fixator and the external fixation index (EFI) with a control group of 27 patients (27 segments) who had double-level procedures with external fixation. The groups were matched for the gain in length, age, and level of difficulty score.


Bone & Joint Open
Vol. 1, Issue 8 | Pages 494 - 499
18 Aug 2020
Karia M Gupta V Zahra W Dixon J Tayton E

Aims

The aim of this study is to determine the effects of the UK lockdown during the COVID-19 pandemic on the orthopaedic admissions, operations, training opportunities, and theatre efficiency in a large district general hospital.

Methods

The number of patients referred to the orthopaedic team between 1 April 2020 and 30 April 2020 were collected. Other data collected included patient demographics, number of admissions, number and type of operations performed, and seniority of primary surgeon. Theatre time was collected consisting of anaesthetic time, surgical time, time to leave theatre, and turnaround time. Data were compared to the same period in 2019.


Bone & Joint Open
Vol. 1, Issue 6 | Pages 287 - 292
19 Jun 2020
Iliadis AD Eastwood DM Bayliss L Cooper M Gibson A Hargunani R Calder P

Introduction

In response to the COVID-19 pandemic, there was a rapidly implemented restructuring of UK healthcare services. The The Royal National Orthopaedic Hospital, Stanmore, became a central hub for the provision of trauma services for North Central/East London (NCEL) while providing a musculoskeletal tumour service for the south of England, the Midlands, and Wales and an urgent spinal service for London. This study reviews our paediatric practice over this period in order to share our experience and lessons learned. Our hospital admission pathways are described and the safety of surgical and interventional radiological procedures performed under general anaesthesia (GA) with regards to COVID-19 in a paediatric population are evaluated.

Methods

All paediatric patients (≤ 16 years) treated in our institution during the six-week peak period of the pandemic were included. Prospective data for all paediatric trauma and urgent elective admissions and retrospective data for all sarcoma admissions were collected. Telephone interviews were conducted with all patients and families to assess COVID-19 related morbidity at 14 days post-discharge.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 582 - 589
1 May 2015
Brennan SA Ní Fhoghlú C Devitt BM O’Mahony FJ Brabazon D Walsh A

Implant-associated infection is a major source of morbidity in orthopaedic surgery. There has been extensive research into the development of materials that prevent biofilm formation, and hence, reduce the risk of infection. Silver nanoparticle technology is receiving much interest in the field of orthopaedics for its antimicrobial properties, and the results of studies to date are encouraging. Antimicrobial effects have been seen when silver nanoparticles are used in trauma implants, tumour prostheses, bone cement, and also when combined with hydroxyapatite coatings. Although there are promising results with in vitro and in vivo studies, the number of clinical studies remains small. Future studies will be required to explore further the possible side effects associated with silver nanoparticles, to ensure their use in an effective and biocompatible manner. Here we present a review of the current literature relating to the production of nanosilver for medical use, and its orthopaedic applications.

Cite this article: Bone Joint J 2015; 97-B:582–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 128 - 133
1 Jan 2012
Kim S Agashe MV Song S Choi H Lee H Song H

Lengthening of the humerus is now an established technique. We compared the complications of humeral lengthening with those of femoral lengthening and investigated whether or not the callus formation in the humerus proceeds at a higher rate than that in the femur. A total of 24 humeral and 24 femoral lengthenings were performed on 12 patients with achondroplasia. We measured the pixel value ratio (PVR) of the lengthened area on radiographs and each radiograph was analysed for the shape, type and density of the callus. The quality of life (QOL) of the patients after humeral lengthening was compared with that prior to surgery. The complication rate per segment of humerus and femur was 0.87% and 1.37%, respectively. In the humerus the PVR was significantly higher than that of the femur. Lower limbs were associated with an increased incidence of concave, lateral and central callus shapes. Humeral lengthening had a lower complication rate than lower-limb lengthening, and QOL increased significantly after humeral lengthening. Callus formation in the humerus during the distraction period proceeded at a significantly higher rate than that in the femur.

These findings indicate that humeral lengthening has an important role in the management of patients with achondroplasia.