Advertisement for orthosearch.org.uk
Results 1 - 10 of 10
Results per page:
The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 88 - 96
1 Jan 2023
Vogt B Rupp C Gosheger G Eveslage M Laufer A Toporowski G Roedl R Frommer A

Aims. Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences. Methods. A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic junction and locking bolts at four different stages. Results. Osteolysis next to the telescopic junction was observed in 31/48 segments (65%) lengthened with the STRYDE nail before implant removal compared to 1/91 segment (1%) in the PRECICE cohort. In the STRYDE cohort, osteolysis initially increased, but decreased or resolved in almost all lengthened segments (86%) after implant removal. Implant failure was observed in 9/48 STRYDE (19%) and in 8/92 PRECICE nails (9%). Breakage of the distal locking bolts was found in 5/48 STRYDE nails (10%) compared to none in the PRECICE cohort. Treatment-associated pain was generally recorded as mild and found in 30/48 patients (63%) and 39/92 (42%) in the STRYDE and PRECICE cohorts, respectively. Temporary range of motion (ROM) limitations under distraction were registered in 17/48 (35%) segments treated with the STRYDE and 35/92 segments (38%) treated with the PRECICE nail. Conclusion. Osteolysis and periosteal reaction, implant breakage, and pain during lengthening and consolidation is more likely in patients treated with the STRYDE nail compared to the PRECICE nail. Temporary ROM limitations during lengthening occurred independent of the applied device. Implant-related osseous alterations seem to remodel after implant removal. Cite this article: Bone Joint J 2023;105-B(1):88–96


The Bone & Joint Journal
Vol. 106-B, Issue 3 | Pages 293 - 302
1 Mar 2024
Vogt B Lueckingsmeier M Gosheger G Laufer A Toporowski G Antfang C Roedl R Frommer A

Aims. As an alternative to external fixators, intramedullary lengthening nails (ILNs) can be employed for distraction osteogenesis. While previous studies have demonstrated that typical complications of external devices, such as soft-tissue tethering, and pin site infection can be avoided with ILNs, there is a lack of studies that exclusively investigated tibial distraction osteogenesis with motorized ILNs inserted via an antegrade approach. Methods. A total of 58 patients (median age 17 years (interquartile range (IQR) 15 to 21)) treated by unilateral tibial distraction osteogenesis for a median leg length discrepancy of 41 mm (IQR 34 to 53), and nine patients with disproportionate short stature treated by bilateral simultaneous tibial distraction osteogenesis, with magnetically controlled motorized ILNs inserted via an antegrade approach, were retrospectively analyzed. The median follow-up was 37 months (IQR 30 to 51). Outcome measurements were accuracy, precision, reliability, bone healing, complications, and patient-reported outcome assessed by the Limb Deformity-Scoliosis Research Society Score (LD-SRS-30). Results. A median tibial distraction of 44 mm (IQR 31 to 49) was achieved with a mean distraction index of 0.5 mm/day (standard deviation 0.13) and median consolidation index of 41.2 days/cm (IQR 34 to 51). Accuracy, precision, and reliability were 91%, 92%, and 97%, respectively. New temporary range of motion limitations occurred in 51% of segments (34/67). Distraction-related equinus deformity treated by Achilles tendon lengthening was the most common major complication recorded in 16% of segments (11/67). In 95% of patients (55/58) the distraction goal was achieved with 42% unplanned additional interventions per segment (28/67). The median postoperative LD-SRS-30 score was 4.0 (IQR 3.6 to 4.3). Conclusion. Tibial distraction osteogenesis using motorized ILNs inserted via an antegrade approach appears to be a reliable and precise procedure. Temporary joint stiffness of the knee or ankle should be expected in up to every second patient. A high rate and wide range of complications of variable severity should be anticipated. Cite this article: Bone Joint J 2024;106-B(3):293–302


Bone & Joint Open
Vol. 4, Issue 8 | Pages 643 - 651
24 Aug 2023
Langit MB Tay KS Al-Omar HK Barlow G Bates J Chuo CB Muir R Sharma H

Aims

The standard of wide tumour-like resection for chronic osteomyelitis (COM) has been challenged recently by adequate debridement. This paper reviews the evolution of surgical debridement for long bone COM, and presents the outcome of adequate debridement in a tertiary bone infection unit.

Methods

We analyzed the retrospective record review from 2014 to 2020 of patients with long bone COM. All were managed by multidisciplinary infection team (MDT) protocol. Adequate debridement was employed for all cases, and no case of wide resection was included.


Bone & Joint Open
Vol. 3, Issue 5 | Pages 359 - 366
1 May 2022
Sadekar V Watts AT Moulder E Souroullas P Hadland Y Barron E Muir R Sharma HK

Aims

The timing of when to remove a circular frame is crucial; early removal results in refracture or deformity, while late removal increases the patient morbidity and delay in return to work. This study was designed to assess the effectiveness of a staged reloading protocol. We report the incidence of mechanical failure following both single-stage and two stage reloading protocols and analyze the associated risk factors.

Methods

We identified consecutive patients from our departmental database. Both trauma and elective cases were included, of all ages, frame types, and pathologies who underwent circular frame treatment. Our protocol is either a single-stage or two-stage process implemented by defunctioning the frame, in order to progressively increase the weightbearing load through the bone, and promote full loading prior to frame removal. Before progression, through the process we monitor patients for any increase in pain and assess radiographs for deformity or refracture.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 6 | Pages 1041 - 1045
1 Nov 1999
Simpson AHRW Cole AS Kenwright J

Distraction osteogenesis is widely used for leg lengthening, but often requires a long period of external fixation which carries risks of pin-track sepsis, malalignment, stiffness of the joint and late fracture of the regenerate. We present the results of 20 cases in which, in an attempt to reduce the rate of complications, a combination of external fixation and intramedullary nailing was used. The mean gain in length was 4.7 cm (2 to 8.6). The mean time of external fixation was 20 days per centimetre gain in length. All distracted segments healed spontaneously without refracture or malalignment. There were three cases of deep infection, two of which occurred in patients who had had previous open fractures of the bone which was being lengthened. All resolved with appropriate treatment. This method allows early rehabilitation, with a rapid return of knee movement. There is a lower rate of complications than occurs when external fixation is used on its own. The time of external fixation is shorter than in other methods of leg lengthening. The high risk of infection calls for caution


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1168 - 1172
1 Jun 2021
Iliadis AD Wright J Stoddart MT Goodier WD Calder P

Aims

The STRYDE nail is an evolution of the PRECICE Intramedullary Limb Lengthening System, with unique features regarding its composition. It is designed for load bearing throughout treatment in order to improve patient experience and outcomes and allow for simultaneous bilateral lower limb lengthening. The literature published to date is limited regarding outcomes and potential problems. We report on our early experience and raise awareness for the potential of adverse effects from this device.

Methods

This is a retrospective review of prospective data collected on all patients treated in our institution using this implant. We report the demographics, nail accuracy, reliability, consolidation index, and cases where concerning clinical and radiological findings were encountered. There were 14 STRYDE nails implanted in nine patients (three male and six female) between June 2019 and September 2020. Mean age at surgery was 33 years (14 to 65). Five patients underwent bilateral lengthening (two femoral and three tibial) and four patients unilateral femoral lengthening for multiple aetiologies.


Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims

This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported.

Methods

This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 128 - 133
1 Jan 2012
Kim S Agashe MV Song S Choi H Lee H Song H

Lengthening of the humerus is now an established technique. We compared the complications of humeral lengthening with those of femoral lengthening and investigated whether or not the callus formation in the humerus proceeds at a higher rate than that in the femur. A total of 24 humeral and 24 femoral lengthenings were performed on 12 patients with achondroplasia. We measured the pixel value ratio (PVR) of the lengthened area on radiographs and each radiograph was analysed for the shape, type and density of the callus. The quality of life (QOL) of the patients after humeral lengthening was compared with that prior to surgery. The complication rate per segment of humerus and femur was 0.87% and 1.37%, respectively. In the humerus the PVR was significantly higher than that of the femur. Lower limbs were associated with an increased incidence of concave, lateral and central callus shapes. Humeral lengthening had a lower complication rate than lower-limb lengthening, and QOL increased significantly after humeral lengthening. Callus formation in the humerus during the distraction period proceeded at a significantly higher rate than that in the femur.

These findings indicate that humeral lengthening has an important role in the management of patients with achondroplasia.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1394 - 1399
1 Oct 2009
Oh C Song H Kim J Choi J Min W Park B

Ten patients, who were unsuitable for limb lengthening over an intramedullary nail, underwent lengthening with a submuscular locking plate. Their mean age at operation was 18.5 years (11 to 40). After fixing a locking plate submuscularly on the proximal segment, an external fixator was applied to lengthen the bone after corticotomy. Lengthening was at 1 mm/day and on reaching the target length, three or four screws were placed in the plate in the distal segment and the external fixator was removed. All patients achieved the pre-operative target length at a mean of 4.0 cm (3.2 to 5.5). The mean duration of external fixation was 61.6 days (45 to 113) and the mean external fixation index was 15.1 days/cm (13.2 to 20.5), which was less than one-third of the mean healing index (48 days/cm (41.3 to 55). There were only minor complications.

Lengthening with a submuscular locking plate can successfully permit early removal of the fixator with fewer complications and is a useful alternative in children or when nailing is difficult.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 938 - 942
1 Jul 2006
Singh S Lahiri A Iqbal M

Limb lengthening by callus distraction and external fixation has a high rate of complications. We describe our experience using an intramedullary nail (Fitbone) which contains a motorised and programmable sliding mechanism for limb lengthening and bone transport. Between 2001 and 2004 we lengthened 13 femora and 11 tibiae in ten patients (seven men and three women) with a mean age of 32 years (21 to 47) using this nail. The indications for operation were short stature in six patients and developmental or acquired disorders in the rest.

The mean lengthening achieved was 40 mm (27 to 60). The mean length of stay in hospital was seven days (5 to 9). The mean healing index was 35 days/cm (18.8 to 70.9). There were no cases of implant-related infection or malunion.