Advertisement for orthosearch.org.uk
Results 1 - 20 of 69
Results per page:
The Bone & Joint Journal
Vol. 99-B, Issue 11 | Pages 1435 - 1441
1 Nov 2017
Buttaro MA Oñativia JI Slullitel PA Andreoli M Comba F Zanotti G Piccaluga F

Aims. The Corail stem has good long-term results. After four years of using this stem, we have detected a small group of patients who have presented with symptomatic metaphyseal debonding. The aim of this study was to quantify the incidence of this complication, to delineate the characteristics of patients presenting with this complication and to compare these patients with asymptomatic controls to determine any important predisposing factors. Patients and Methods. Of 855 Corail collarless cementless stems implanted for osteoarthritis, 18 presented with symptomatic metaphyseal debonding. A control group of 74 randomly selected patients was assembled. Clinical and radiological parameters were measured and a logistic regression model was created to evaluate factors associated with metaphyseal debonding. Results. The prevalence of this complication was 2.1% in our series. In the multivariable model, the presence of a Dorr B-type proximal femur was associated with metaphyseal debonding (odds ratio (OR) 10.73, 95% confidence interval (CI) 2.31 to 49.97, p = 0.002), as was a body mass index > 25 kg/m. 2. (OR 6.85, 95% CI 1.06 to 44.28, p = 0.04). Smaller stems and the use of a polyethylene acetabular liner appeared to be protective when compared with metal and ceramic setting hard-on-hard bearings. Conclusion. We have described an uncommon but important mode of failure of the Corail stem. Surgeons should be aware of this phenomenon; overweight patients with Dorr B-type femurs and in whom hard bearings are used appear to be particularly at risk. Cite this article: Bone Joint J 2017;99-B:1435–41


Aims. The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail). Methods. Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using radiostereometric analysis (RSA), change of bone mineral density (BMD) around the stem, the development of radiolucent lines, and additional patient-reported outcome measures (PROMs). Results. There were no statistically significant differences between the groups regarding PROMs (median OHS (CFP 45 (interquartile range (IQR) 35 to 48); Corail 45 (IQR 40 to 48); p = 0.568). RSA showed stable stems in both groups, with little or no further subsidence between two and five years. Resorption of the femoral neck was evident in nine patients in the CFP group and in none of the 15 Corail stems with a collar that could be studied. Dual X-ray absorbiometry showed a significantly higher loss of BMD in the proximal Gruen zones in the CFP group (mean changes in BMD: Gruen zone 1, CFP -9.5 (95% confidence interval (CI) -14.8 to -4.2), Corail 1.0 (95% CI 3.4 to 5.4); Gruen zone 7, CFP -23.0 (95% CI -29.4 to -16.6), Corail -7.2 (95% CI -15.9 to 1.4). Two CFP stems were revised before two years’ follow-up due to loosening, and one Corail stem was revised after two years due to chronic infection. Conclusion. The CFP stem has a similar clinical outcome and subsidence pattern when compared with the Corail stem. More pronounced proximal stress-shielding was seen with the CFP stem, suggesting diaphyseal fixation, and questioning its femoral neck-sparing properties in the long term. Cite this article: Bone Joint J 2022;104-B(5):581–588


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1589 - 1596
1 Dec 2016
Magill P Blaney J Hill JC Bonnin MP Beverland DE

Aims. Our aim was to report survivorship data and lessons learned with the Corail/Pinnacle cementless total hip arthroplasty (THA) system. Patients and Methods. Between August 2005 and March 2015, a total of 4802 primary cementless Corail/Pinnacle THAs were performed in 4309 patients. In March 2016, we reviewed these hips from a prospectively maintained database. Results . A total of 80 hips (1.67%) have been revised which is equivalent to a cumulative risk of revision of 2.5% at ten years. The rate of revision was not significantly higher in patients aged ≥ 70 years (p = 0.93). The leading indications for revision were instability (n = 22, 0.46%), infection (n = 20, 0.42%), aseptic femoral loosening (n = 15, 0.31%) and femoral fracture (n = 6, 0.12%). There were changes in the surgical technique with respect to the Corail femoral component during the ten-year period involving a change to collared components and a trend towards larger size. These resulted in a decrease in the rate of iatrogenic femoral fracture and a decrease in the rate of aseptic loosening. Conclusion. The rate of revision in this series is comparable with the best performing THAs in registry data. Most revisions were not directly related to the implants. Despite extensive previous experience with cemented femoral components, the senior author noted a learning curve requiring increased focus on primary stability. The number of revisions related to the femoral component is reducing. Any new technology has a learning curve that may be independent of surgical experience. Cite this article: Bone Joint J 2016;98-B:1589–96


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 144 - 150
1 Feb 2024
Lynch Wong M Robinson M Bryce L Cassidy R Lamb JN Diamond O Beverland D

Aims

The aim of this study was to determine both the incidence of, and the reoperation rate for, postoperative periprosthetic femoral fracture (POPFF) after total hip arthroplasty (THA) with either a collared cementless (CC) femoral component or a cemented polished taper-slip (PTS) femoral component.

Methods

We performed a retrospective review of a consecutive series of 11,018 THAs over a ten-year period. All POPFFs were identified using regional radiograph archiving and electronic care systems.


The Bone & Joint Journal
Vol. 105-B, Issue 10 | Pages 1045 - 1051
1 Oct 2023
Turgeon TR Righolt CH Burnell CD Gascoyne TC Hedden DR Bohm ER

Aims. The primary aim of this trial was to compare the subsidence of two similar hydroxyapatite-coated titanium femoral components from different manufacturers. Secondary aims were to compare rotational migration (anteversion/retroversion and varus/valgus tilt) and patient-reported outcome measures between both femoral components. Methods. Patients were randomized to receive one of the two femoral components (Avenir or Corail) during their primary total hip arthroplasty between August 2018 and September 2020. Radiostereometric analysis examinations at six, 12, and 24 months were used to assess the migration of each implanted femoral component compared to a baseline assessment. Patient-reported outcome measures were also recorded for these same timepoints. Overall, 50 patients were enrolled (62% male (n = 31), with a mean age of 65.7 years (SD 7.3), and mean BMI of 30.2 kg/m. 2. (SD 5.2)). Results. The two-year subsidence was similar for Avenir (-0.018 mm (95% confidence interval (CI) -0.053 to 0.018) and Corail (0.000 mm (95% CI -0.027 to 0.026; p = 0.428). Both anteversion/retroversion (Avenir 0.139° (95% CI -0.204 to 0.481°); Corail -0.196° (95% CI -0.445 to 0.053°; p = 0.110) and varus/valgus tilt (Avenir -0.024° (95% CI -0.077 to 0.028); Corail -0.049° (95% CI -0.098 to 0.000°; p = 0.473) were not statistically significantly different. After two years, patients reported similar improvements in EuroQol five-dimension five-level health questionnaire (Avenir 0.22 (SD 0.2); Corail 0.22 (SD 0.18); p = 0.965) and other outcomes scores. Patient satisfaction on a five-point Likert scale was also similar between both groups after two years (Avenir 1.38 (SD 0.88); Corail 1.33 (SD 0.57); p = 0.846). Conclusion. The performance of both femoral components was similar in terms of stability and patient outcomes. Cite this article: Bone Joint J 2023;105-B(10):1045–1051


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 19 - 26
1 Jan 2022
Sevaldsen K Schnell Husby O Lian ØB Farran KM Schnell Husby V

Aims. Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study. Methods. In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis. Results. Results from mixed model analyses revealed a lower mean retroversion for line-to-line (0.72° (95% confidence interval (CI) 0.38° to 1.07°; p < 0.001), but no significant differences in subsidence between the techniques (-0.15 mm (95% CI -0.53 to 0.227; p = 0.429) at 24 months. Radiolucent lines measuring < 2 mm wide were found in three and five arthroplasties cemented by the standard and line-to-line method, respectively. Conclusion. The cemented Corail stem with a force-closed design seems to settle earlier and better with the line-to-line cementing method, although for subsidence the difference was not significant. However, the lower rate of migration into retroversion may reduce the wear and cement deformation, contributing to good long-term fixation and implant survival. Cite this article: Bone Joint J 2022;104-B(1):19–26


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 299 - 305
1 Mar 2014
Bell KR Clement ND Jenkins PJ Keating JF

We performed a case–control study to compare the rates of further surgery, revision and complications, operating time and survival in patients who were treated with either an uncemented hydroxyapatite-coated Corail bipolar femoral stem or a cemented Exeter stem for a displaced intracapsular fracture of the hip. The mean age of the patients in the uncemented group was 82.5 years (53 to 97) and in the cemented group was 82.7 years (51 to 99) We used propensity score matching, adjusting for age, gender and the presence or absence of dementia and comorbidities, to produce a matched cohort receiving an Exeter stem (n = 69) with which to compare the outcome of patients receiving a Corail stem (n = 69). The Corail had a significantly lower all-cause rate of further surgery (p = 0.016; odds ratio (OR) 0.18, 95% CI 0.04 to 0.84) and number of hips undergoing major further surgery (p = 0.029; OR 0.13, 95% CI 0.01 to 1.09). The mean operating time was significantly less for the Corail group than for the cemented Exeter group (59 min [12 to 136] vs 70 min [40 to 175], p = 0.001). The Corail group also had a lower risk of a peri-prosthetic fracture (p = 0.042; OR 0.19, 95% CI 0.01 to 1.42) . There was no difference in the mortality rate between the groups. There were significantly fewer complications in the uncemented group, suggesting that the use of this stem would result in a decreased rate of morbidity in these frail patients. Whether this relates to an improved functional outcome remains unknown. Cite this article: Bone Joint J 2014;96-B:299–305


Bone & Joint Open
Vol. 2, Issue 5 | Pages 293 - 300
3 May 2021
Lewis PM Khan FJ Feathers JR Lewis MH Morris KH Waddell JP

Aims. “Get It Right First Time” (GIRFT) and NHS England’s Best Practice Tariff (BPT) have published directives advising that patients over the ages of 65 (GIRFT) and 69 years (BPT) receiving total hip arthroplasty (THA) should receive cemented implants and have brought in financial penalties if this policy is not observed. Despite this, worldwide, uncemented component use has increased, a situation described as a ‘paradox’. GIRFT and BPT do, however, acknowledge more data are required to support this edict with current policies based on the National Joint Registry survivorship and implant costs. Methods. This study compares THA outcomes for over 1,000 uncemented Corail/Pinnacle constructs used in all age groups/patient frailty, under one surgeon, with identical pre- and postoperative pathways over a nine-year period with mean follow-up of five years and two months (range: nine months to nine years and nine months). Implant information, survivorship, and regular postoperative Oxford Hip Scores (OHS) were collected and two comparisons undertaken: a comparison of those aged over 65 years with those 65 and under and a second comparison of those aged 70 years and over with those aged under 70. Results. Overall revision rate was 1.3% (13/1,004). A greater number of revisions were undertaken in those aged over 65 years, but numbers were small and did not reach significance. The majority of revisions were implant-independent. Single component analysis revealed a 99.9% and 99.6% survival for the uncemented cup and femoral component, respectively. Mean patient-reported outcome measures (PROMs) improvement for all ages outperformed the national PROMs and a significantly greater proportion of those aged over 65/69 years reached and maintained a meaningful improvement in their OHS earlier than their younger counterparts (p < 0.05/0.01 respectively). Conclusion. This study confirms that this uncemented THA system can be used safely and effectively in patient groups aged over 65 years and those over 69 years, with low complication and revision rates. Cite this article: Bone Jt Open 2021;2(5):293–300


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 762 - 770
1 Jun 2015
Pennington MW Grieve R van der Meulen JH

There is little evidence on the cost effectiveness of different brands of hip prostheses. We compared lifetime cost effectiveness of frequently used brands within types of prosthesis including cemented (Exeter V40 Contemporary, Exeter V40 Duration and Exeter V40 Elite Plus Ogee), cementless (Corail Pinnacle, Accolade Trident, and Taperloc Exceed) and hybrid (Exeter V40 Trilogy, Exeter V40 Trident, and CPT Trilogy). We used data from three linked English national databases to estimate the lifetime risk of revision, quality-adjusted life years (QALYs) and cost. . For women with osteoarthritis aged 70 years, the Exeter V40 Elite Plus Ogee had the lowest risk of revision (5.9% revision risk, 9.0 QALYs) and the CPT Trilogy had the highest QALYs (10.9% revision risk, 9.3 QALYs). Compared with the Corail Pinnacle (9.3% revision risk, 9.22 QALYs), the most commonly used brand, and assuming a willingness-to-pay of £20 000 per QALY gain, the CPT Trilogy is most cost effective, with an incremental net monetary benefit of £876. Differences in cost effectiveness between the hybrid CPT Trilogy and Exeter V40 Trident and the cementless Corail Pinnacle and Taperloc Exceed were small, and a cautious interpretation is required, given the limitations of the available information. However, it is unlikely that cemented brands are among the most cost effective. Similar patterns of results were observed for men and other ages. The gain in quality of life after total hip arthroplasty, rather than the risk of revision, was the main driver of cost effectiveness. Cite this article: Bone Joint J 2015;97-B:762–70


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1559 - 1564
1 Dec 2018
McConnell JS Saunders PRJ Young SK

Aims. Cementless femoral stems must be correctly sized and well-seated to obtain satisfactory biological fixation. The change in sound that occurs during impaction of the femoral broach is said to indicate good fit, but this has not been widely studied. We set out to find whether the presence or absence of these sound changes could predict correct sizing. Patients and Methods. We recorded the sound generated during femoral broaching for 105 cementless total hip arthroplasties using the Corail stem. Four cases were excluded, leaving 101 recordings for analysis. There were 36 male patients and 65 female patients, with a mean age of 69.9 years (. sd. 12.3) and median body mass index (BMI) of 29 kg/m. 2. (interquartile range (IQR) 26 to 32). The recordings were analyzed to identify the frequencies of the sounds produced during impaction of the femoral broach. Results. The emergence of a low-frequency band of sound in the 1 kHz range, during the final femoral broaching, was a strong predictor of a well-sized implant stem. The frequency was related to femoral length, supporting our hypothesis that the sound arose from the bone itself. Conclusion. The low-frequency sound generated during femoral broaching can be monitored spectrographically, its frequency can be predicted from femoral length, and it is a good predictor of appropriate stem sizing


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 741 - 748
1 Jun 2015
Bonnin MP Neto CC Aitsiselmi T Murphy CG Bossard N Roche S

The aim of this study was to investigate the relationship between the geometry of the proximal femur and the incidence of intra-operative fracture during uncemented total hip arthroplasty (THA). We studied the pre-operative CT scans of 100 patients undergoing THA with an uncemented femoral component. We measured the anteroposterior and mediolateral dimensions at the level of division of the femoral neck to calculate the aspect ratio of the femur. Wide variations in the shape of the femur were observed, from round, to very narrow elliptic. The femurs of women were narrower than those of men (p < 0.0001) and small femurs were also narrower than large ones. Patients with an intra-operative fracture of the calcar had smaller and narrower femurs than those without a fracture (p < 0.05) and the implanted Corail stems were smaller in those with a fracture (mean size 9 vs 12, p < 0.0001). The variability of the shape of the femoral neck at the level of division contributes to the understanding of the causation of intra-operative fractures in uncemented THA. Cite this article: Bone Joint J 2015;97-B:741–8


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1574 - 1580
1 Dec 2007
Hallan G Lie SA Furnes O Engesaeter LB Vollset SE Havelin LI

Primary uncemented femoral stems reported to the Norwegian arthroplasty register between 1987 and 2005 were included in this prospective observational study. There were 11 516 hips (9679 patients) and 14 different designs of stem. Kaplan-Meier survival probabilities and Cox regression were used to analyse the data. With aseptic loosening as the end-point, all currently used designs performed excellently with survival of 96% to 100% at ten years. With the end-point as stem revision for any cause, the long-term results of the different designs varied from poor to excellent, with survival at 15 years ranging between 29% and 97%. Follow-up for longer than seven years was needed to identify some of the poorly-performing designs. There were differences between the stems; the Corail, used in 5456 hips, was the most frequently used stem with a survival of 97% at 15 years. Male gender was associated with an increased risk of revision of × 1.3 (95% confidence interval 1.05 to 1.52), but age and diagnosis had no influence on the results. Overall, modern uncemented femoral stems performed well. Moderate differences in survival between well-performing stems should be interpreted with caution since the differences may be caused by factors other than the stem itself


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 548 - 554
1 Jun 2024
Ohyama Y Minoda Y Masuda S Sugama R Ohta Y Nakamura H

Aims

The aim of this study was to compare the pattern of initial fixation and changes in periprosthetic bone mineral density (BMD) between patients who underwent total hip arthroplasty (THA) using a traditional fully hydroxyapatite (HA)-coated stem (T-HA group) and those with a newly introduced fully HA-coated stem (N-HA group).

Methods

The study included 36 patients with T-HA stems and 30 with N-HA stems. Dual-energy X-ray absorptiometry was used to measure the change in periprosthetic BMD, one and two years postoperatively. The 3D contact between the stem and femoral cortical bone was evaluated using a density-mapping system, and clinical assessment, including patient-reported outcome measurements, was recorded.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 592 - 600
1 May 2017
Matharu GS Nandra RS Berryman F Judge A Pynsent PB Dunlop DJ

Aims. To determine ten-year failure rates following 36 mm metal-on-metal (MoM) Pinnacle total hip arthroplasty (THA), and identify predictors of failure. Patients and Methods. We retrospectively assessed a single-centre cohort of 569 primary 36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012 according to Medicines and Healthcare Products Regulation Agency recommendations. All-cause failure rates (all-cause revision, and non-revised cross-sectional imaging failures) were calculated, with predictors for failure identified using multivariable Cox regression. Results. Failure occurred in 97 hips (17.0%). The ten-year cumulative failure rate was 27.1% (95% confidence interval (CI) 21.6 to 33.7). Primary implantation from 2006 onwards (hazard ratio (HR) 4.30; 95% CI 1.82 to 10.1; p = 0.001) and bilateral MoM hip arthroplasty (HR 1.59; 95% CI 1.03 to 2.46; p = 0.037) predicted failure. The effect of implantation year on failure varied over time. From four years onwards following surgery, hips implanted since 2006 had significantly higher failure rates (eight years 28.3%; 95% CI 23.1 to 34.5) compared with hips implanted before 2006 (eight years 6.3%; 95% CI 2.4 to 15.8) (HR 15.2; 95% CI 2.11 to 110.4; p = 0.007). Conclusion. We observed that 36 mm MoM Pinnacle THAs have an unacceptably high ten-year failure rate, especially if implanted from 2006 onwards or in bilateral MoM hip patients. Our findings regarding implantation year and failure support recent concerns about the device manufacturing process. We recommend all patients undergoing implantation since 2006 and those with bilateral MoM hips undergo regular investigation, regardless of symptoms. Cite this article: Bone Joint J 2017;99-B:592–600


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 826 - 832
1 Jul 2022
Stadelmann VA Rüdiger HA Nauer S Leunig M

Aims

It is not known whether preservation of the capsule of the hip positively affects patient-reported outcome measures (PROMs) in total hip arthroplasty using the direct anterior approach (DAA-THA). A recent randomized controlled trial found no clinically significant difference at one year postoperatively. This study aimed to determine whether preservation of the anterolateral capsule and anatomical closure improve the outcome and revision rate, when compared with resection of the anterolateral capsule, at two years postoperatively.

Methods

Two consecutive groups of patients whose operations were performed by the senior author were compared. The anterolateral capsule was resected in the first group of 430 patients between January 2012 and December 2014, and preserved and anatomically closed in the second group of 450 patients between July 2015 and December 2017. There were no other technical changes between the two groups. Patient characteristics, the Charlson Comorbidity Index (CCI), and surgical data were collected from our database. PROM questionnaires, consisting of the Oxford Hip Score (OHS) and Core Outcome Measures Index (COMI-Hip), were collected two years postoperatively. Data were analyzed with generalized multiple regression analysis.


Bone & Joint Open
Vol. 3, Issue 12 | Pages 960 - 968
23 Dec 2022
Hardwick-Morris M Wigmore E Twiggs J Miles B Jones CW Yates PJ

Aims

Leg length discrepancy (LLD) is a common pre- and postoperative issue in total hip arthroplasty (THA) patients. The conventional technique for measuring LLD has historically been on a non-weightbearing anteroposterior pelvic radiograph; however, this does not capture many potential sources of LLD. The aim of this study was to determine if long-limb EOS radiology can provide a more reproducible and holistic measurement of LLD.

Methods

In all, 93 patients who underwent a THA received a standardized preoperative EOS scan, anteroposterior (AP) radiograph, and clinical LLD assessment. Overall, 13 measurements were taken along both anatomical and functional axes and measured twice by an orthopaedic fellow and surgical planning engineer to calculate intraoperator reproducibility and correlations between measurements.


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 540 - 547
1 Jun 2024
Nandra RS Elnahal WA Mayne A Brash L McBryde CW Treacy RBC

Aims

The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years.

Methods

The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.


Bone & Joint Research
Vol. 10, Issue 12 | Pages 780 - 789
1 Dec 2021
Eslam Pour A Lazennec JY Patel KP Anjaria MP Beaulé PE Schwarzkopf R

Aims

In computer simulations, the shape of the range of motion (ROM) of a stem with a cylindrical neck design will be a perfect cone. However, many modern stems have rectangular/oval-shaped necks. We hypothesized that the rectangular/oval stem neck will affect the shape of the ROM and the prosthetic impingement.

Methods

Total hip arthroplasty (THA) motion while standing and sitting was simulated using a MATLAB model (one stem with a cylindrical neck and one stem with a rectangular neck). The primary predictor was the geometry of the neck (cylindrical vs rectangular) and the main outcome was the shape of ROM based on the prosthetic impingement between the neck and the liner. The secondary outcome was the difference in the ROM provided by each neck geometry and the effect of the pelvic tilt on this ROM. Multiple regression was used to analyze the data.


Bone & Joint Research
Vol. 11, Issue 3 | Pages 180 - 188
1 Mar 2022
Rajpura A Asle SG Ait Si Selmi T Board T

Aims

Hip arthroplasty aims to accurately recreate joint biomechanics. Considerable attention has been paid to vertical and horizontal offset, but femoral head centre in the anteroposterior (AP) plane has received little attention. This study investigates the accuracy of restoration of joint centre of rotation in the AP plane.

Methods

Postoperative CT scans of 40 patients who underwent unilateral uncemented total hip arthroplasty were analyzed. Anteroposterior offset (APO) and femoral anteversion were measured on both the operated and non-operated sides. Sagittal tilt of the femoral stem was also measured. APO measured on axial slices was defined as the perpendicular distance between a line drawn from the anterior most point of the proximal femur (anterior reference line) to the centre of the femoral head. The anterior reference line was made parallel to the posterior condylar axis of the knee to correct for rotation.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 193 - 199
1 Feb 2022
Wang Q Wang H A G Xiao T Kang P

Aims

This study aimed to use intraoperative free electromyography to examine how the placement of a retractor at different positions along the anterior acetabular wall may affect the femoral nerve during total hip arthroplasty (THA) when undertaken using the direct anterior approach (THA-DAA).

Methods

Intraoperative free electromyography was performed during primary THA-DAA in 82 patients (94 hips). The highest position of the anterior acetabular wall was defined as the “12 o’clock” position (middle position) when the patient was in supine position. After exposure of the acetabulum, a retractor was sequentially placed at the ten, 11, 12, one, and two o’clock positions (right hip; from superior to inferior positions). Action potentials in the femoral nerve were monitored with each placement, and the incidence of positive reactions (defined as explosive, frequent, or continuous action potentials, indicating that the nerve was being compressed) were recorded as the primary outcome. Secondary outcomes included the incidence of positive reactions caused by removing the femoral head, and by placing a retractor during femoral exposure; and the incidence of femoral nerve palsy, as detected using manual testing of the strength of the quadriceps muscle.