Advertisement for orthosearch.org.uk
Results 1 - 40 of 40
Results per page:

Aims

The aim of this study was to compare the migration of the femoral component, five years postoperatively, between patients with a highly cross-linked polyethylene (HXLPE) insert and those with a conventional polyethylene (PE) insert in an uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary aims included clinical outcomes and patient-reported outcome measures (PROMs). We have previously reported the migration and outcome of the tibial components in these patients.

Methods

A double-blinded randomized controlled trial was conducted including 96 TKAs. The migration of the femoral component was measured with radiostereometry (RSA) at three and six months and one, two, and five years postoperatively. PROMs were collected preoperatively and at all periods of follow-up.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 74 - 81
1 May 2024
Callary SA Broekhuis D Barends J Ramasamy B Nelissen RGHH Solomon LB Kaptein BL

Aims

The aim of this study was to compare the biomechanical models of two frequently used techniques for reconstructing severe acetabular defects with pelvic discontinuity in revision total hip arthroplasty (THA) – the Trabecular Metal Acetabular Revision System (TMARS) and custom triflange acetabular components (CTACs) – using virtual modelling.

Methods

Pre- and postoperative CT scans from ten patients who underwent revision with the TMARS for a Paprosky IIIB acetabular defect with pelvic discontinuity were retrospectively collated. Computer models of a CTAC implant were designed from the preoperative CT scans of these patients. Computer models of the TMARS reconstruction were segmented from postoperative CT scans using a semi-automated method. The amount of bone removed, the implant-bone apposition that was achieved, and the restoration of the centre of rotation of the hip were compared between all the actual TMARS and the virtual CTAC implants.


Bone & Joint Open
Vol. 5, Issue 4 | Pages 260 - 268
1 Apr 2024
Broekhuis D Meurs WMH Kaptein BL Karunaratne S Carey Smith RL Sommerville S Boyle R Nelissen RGHH

Aims

Custom triflange acetabular components (CTACs) play an important role in reconstructive orthopaedic surgery, particularly in revision total hip arthroplasty (rTHA) and pelvic tumour resection procedures. Accurate CTAC positioning is essential to successful surgical outcomes. While prior studies have explored CTAC positioning in rTHA, research focusing on tumour cases and implant flange positioning precision remains limited. Additionally, the impact of intraoperative navigation on positioning accuracy warrants further investigation. This study assesses CTAC positioning accuracy in tumour resection and rTHA cases, focusing on the differences between preoperative planning and postoperative implant positions.

Methods

A multicentre observational cohort study in Australia between February 2017 and March 2021 included consecutive patients undergoing acetabular reconstruction with CTACs in rTHA (Paprosky 3A/3B defects) or tumour resection (including Enneking P2 peri-acetabular area). Of 103 eligible patients (104 hips), 34 patients (35 hips) were analyzed.


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 136 - 143
1 Feb 2024
van der Lelij TJN Marang-van de Mheen PJ Kaptein BL Koster LA Ljung P Nelissen RGHH Toksvig-Larsen S

Aims

The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface.

Methods

Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1168 - 1176
1 Nov 2023
Yüksel Y Koster LA Kaptein BL Nelissen RGHH den Hollander P

Aims

Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes.

Methods

A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively.


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 977 - 984
1 Sep 2023
Kamp T Gademan MGJ van Zon SKR Nelissen RGHH Vliet Vlieland TPM Stevens M Brouwer S

Aims

For the increasing number of working-age patients undergoing total hip or total knee arthroplasty (THA/TKA), return to work (RTW) after surgery is crucial. We investigated the association between occupational class and time to RTW after THA or TKA.

Methods

Data from the prospective multicentre Longitudinal Leiden Orthopaedics Outcomes of Osteoarthritis Study were used. Questionnaires were completed preoperatively and six and 12 months postoperatively. Time to RTW was defined as days from surgery until RTW (full or partial). Occupational class was preoperatively assessed and categorized into four categories according to the International Standard Classification of Occupations 2008 (blue-/white-collar, high-/low-skilled). Cox regression analyses were conducted separately for THA and TKA patients. Low-skilled blue-collar work was used as the reference category.


Aims

The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs).

Methods

A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with radiostereometric analysis (RSA) at three, six, 12, 24, and 60 months postoperatively. PROMS were collected preoperatively and at all follow-up timepoints.


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 148 - 157
1 Feb 2023
Koster LA Rassir R Kaptein BL Sierevelt IN Schager M Nelissen RGHH Nolte PA

Aims

The primary aim of this study was to compare the migration of the femoral and tibial components of the cementless rotating platform Attune and Low Contact Stress (LCS) total knee arthroplasty (TKA) designs, two years postoperatively, using radiostereometric analysis (RSA) in order to assess the risk of the development of aseptic loosening. A secondary aim was to compare clinical and patient-reported outcome measures (PROMs) between the designs.

Methods

A total of 61 TKAs were analyzed in this randomized clinical RSA trial. RSA examinations were performed one day and three, six, 12, and 24 months postoperatively. The maximal total point motion (MPTM), translations, and rotations of the components were analyzed. PROMs and clinical data were collected preoperatively and at six weeks and three, six, 12, and 24 months postoperatively. Linear mixed effect modelling was used for statistical analyses.


Bone & Joint Research
Vol. 12, Issue 1 | Pages 72 - 79
18 Jan 2023
Welling MM Warbroek K Khurshid C van Oosterom MN Rietbergen DDD de Boer MGJ Nelissen RGHH van Leeuwen FWB Pijls BG Buckle T

Aims

Arthroplasty surgery of the knee and hip is performed in two to three million patients annually. Periprosthetic joint infections occur in 4% of these patients. Debridement, antibiotics, and implant retention (DAIR) surgery aimed at cleaning the infected prosthesis often fails, subsequently requiring invasive revision of the complete prosthetic reconstruction. Infection-specific imaging may help to guide DAIR. In this study, we evaluated a bacteria-specific hybrid tracer (99mTc-UBI29-41-Cy5) and its ability to visualize the bacterial load on femoral implants using clinical-grade image guidance methods.

Methods

99mTc-UBI29-41-Cy5 specificity for Stapylococcus aureus was assessed in vitro using fluorescence confocal imaging. Topical administration was used to highlight the location of S. aureus cultured on femoral prostheses using fluorescence imaging and freehand single photon emission CT (fhSPECT) scans. Gamma counting and fhSPECT were used to quantify the bacterial load and monitor cleaning with chlorhexidine. Microbiological culturing helped to relate the imaging findings with the number of (remaining) bacteria.


Bone & Joint Research
Vol. 11, Issue 9 | Pages 629 - 638
1 Sep 2022
Pijls BG Sanders IMJG Kuijper EJ Nelissen RGHH

Aims

Here we used a mature seven-day biofilm model of Staphylococcus aureus, exposed to antibiotics up to an additional seven days, to establish the effectiveness of either mechanical cleaning or antibiotics or non-contact induction heating, and which combinations could eradicate S. aureus in mature biofilms.

Methods

Mature biofilms of S. aureus (ATCC 29213) were grown on titanium alloy (Ti6Al4V) coupons for seven days and were subjected to the following treatments or their combinations: antibiotics, mechanical cleaning, or heat shock by induction heating of 60°C for one minute. Experiments were repeated at least five times.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 221 - 226
1 Feb 2022
Edwards NM Varnum C Nelissen RGHH Overgaard S Pedersen AB

Aims

The aim of this study was to examine whether socioeconomic status (SES) is associated with a higher risk of infections following total hip arthroplasty (THA) at 30 and 90 days.

Methods

We obtained individual-based information on SES markers (cohabitation, education, income, and savings) on 103,901 THA patients from Danish health registries between 1 January 1995 and 31 December 2017. The primary outcome measure was any hospital-treated infection (i.e. all infections). The secondary outcomes were further specified to specific hospital-treated infections (pneumonia, urinary tract infection, and periprosthetic joint infection). The primary timepoint was within 90 days. In addition, the outcomes were further evaluated within 30 days. We calculated the cumulative incidence, and used the pseudo-observation method and generalized linear regression to estimate adjusted risk ratios (RRs) with 95% confidence intervals (CIs) for each marker.


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 855 - 863
1 May 2021
Koster LA Meinardi JE Kaptein BL Van der Linden - Van der Zwaag E Nelissen RGHH

Aims

The objective of this study was to compare the two-year migration pattern and clinical outcomes of a total knee arthroplasty (TKA) with an asymmetrical tibial design (Persona PS) and a well-proven TKA with a symmetrical tibial design (NexGen LPS).

Methods

A randomized controlled radiostereometric analysis (RSA) trial was conducted including 75 cemented posterior-stabilized TKAs. Implant migration was measured with RSA. Maximum total point motion (MTPM), translations, rotations, clinical outcomes, and patient-reported outcome measures (PROMs) were assessed at one week postoperatively and at three, six, 12, and 24 months postoperatively.


The Bone & Joint Journal
Vol. 102-B, Issue 9 | Pages 1158 - 1166
14 Sep 2020
Kaptein BL den Hollander P Thomassen B Fiocco M Nelissen RGHH

Aims

The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs).

Methods

A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively.


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1016 - 1024
1 Aug 2020
Hasan S van Hamersveld KT Marang-van de Mheen PJ Kaptein BL Nelissen RGHH Toksvig-Larsen S

Aims

Although bone cement is the primary mode of fixation in total knee arthroplasty (TKA), cementless fixation is gaining interest as it has the potential of achieving lasting biological fixation. By 3D printing an implant, highly porous structures can be manufactured, promoting osseointegration into the implant to prevent aseptic loosening. This study compares the migration of cementless, 3D-printed TKA to cemented TKA of a similar design up to two years of follow-up using radiostereometric analysis (RSA) known for its ability to predict aseptic loosening.

Methods

A total of 72 patients were randomized to either cementless 3D-printed or a cemented cruciate retaining TKA. RSA and clinical scores were evaluated at baseline and postoperatively at three, 12, and 24 months. A mixed model was used to analyze the repeated measurements.


Bone & Joint Research
Vol. 9, Issue 4 | Pages 192 - 199
1 Apr 2020
Pijls BG Sanders IMJG Kujiper EJ Nelissen RGHH

Aims

Induction heating is a noninvasive, nonantibiotic treatment modality that can potentially be used to cause thermal damage to the bacterial biofilm on the metal implant surface. The purpose of this study was to determine the effectiveness of induction heating on killing Staphylococcus epidermidis from biofilm and to determine the possible synergistic effect of induction heating and antibiotics.

Methods

S. epidermidis biofilms were grown on titanium alloy (Ti6Al4V) coupons for 24 hours (young biofilm) and seven days (mature biofilm). These coupons with biofilm were heated to temperatures of 50°C, 55°C, 60°C, 65°C, 70°C, 80°C, and 90°C for 3.5 minutes and subsequently exposed to vancomycin and rifampicin at clinically relevant concentrations.


Bone & Joint Research
Vol. 7, Issue 11 | Pages 609 - 619
1 Nov 2018
Pijls BG Sanders IMJG Kuijper EJ Nelissen RGHH

Objectives

Prosthetic joint infection (PJI) is a devastating complication following total joint arthroplasty. Non-contact induction heating of metal implants is a new and emerging treatment for PJI. However, there may be concerns for potential tissue necrosis. It is thought that segmental induction heating can be used to control the thermal dose and to limit collateral thermal injury to the bone and surrounding tissues. The purpose of this study was to determine the thermal dose, for commonly used metal implants in orthopaedic surgery, at various distances from the heating centre (HC).

Methods

Commonly used metal orthopaedic implants (hip stem, intramedullary nail, and locking compression plate (LCP)) were heated segmentally using an induction heater. The thermal dose was expressed in cumulative equivalent minutes at 43°C (CEM43) and measured with a thermal camera at several different distances from the HC. A value of 16 CEM43 was used as the threshold for thermal damage in bone.


The Bone & Joint Journal
Vol. 100-B, Issue 3 | Pages 309 - 317
1 Mar 2018
Kolk A Henseler JF Overes FJ Nagels J Nelissen RGHH

Aims

Since long-term outcome of teres major tendon transfer surgery for irreparable posterosuperior rotator cuff (RC) tears is largely unknown, the primary aim of this study was to evaluate the long-term outcome of the teres major transfer. We also aimed to report on the results of a cohort of patients with a similar indication for surgery that underwent a latissimus dorsi tendon transfer.

Patients and Methods

In this prospective cohort study, we reported on the long-term results of 20 consecutive patients with a teres major tendon transfer for irreparable massive posterosuperior RC tears. Additionally, we reported on the results of the latissimus dorsi tendon transfer (n = 19). The mean age was 60 years (47 to 77). Outcomes included the Constant score (CS), and pain at rest and during movement using the Visual Analogue Scale (VAS).


Bone & Joint Research
Vol. 6, Issue 5 | Pages 323 - 330
1 May 2017
Pijls BG Sanders IMJG Kuijper EJ Nelissen RGHH

Objectives

Infection of implants is a major problem in elective and trauma surgery. Heating is an effective way to reduce the bacterial load in food preparation, and studies on hyperthermia treatment for cancer have shown that it is possible to heat metal objects with pulsed electromagnetic fields selectively (PEMF), also known as induction heating. We therefore set out to answer the following research question: is non-contact induction heating of metallic implants effective in reducing bacterial load in vitro?

Methods

Titanium alloy cylinders (Ti6Al4V) were exposed to PEMF from an induction heater with maximum 2000 watts at 27 kHz after being contaminated with five different types of micro-organisms: Staphylococcus epidermidis; Staphylococcus aureus; Pseudomonas aeruginosa; spore-forming Bacillus cereus; and yeast Candida albicans. The cylinders were exposed to incremental target temperatures (35°C, 45°C, 50°C, 55°C, 60°C, 65°C, 70°C) for up to 3.5 minutes.


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1333 - 1341
1 Oct 2016
van der Voort P Valstar ER Kaptein BL Fiocco M van der Heide HJL Nelissen RGHH

Aims

The widely used and well-proven Palacos R (a.k.a. Refobacin Palacos R) bone cement is no longer commercially available and was superseded by Refobacin bone cement R and Palacos R + G in 2005. However, the performance of these newly introduced bone cements have not been tested in a phased evidence-based manner, including roentgen stereophotogrammetric analysis (RSA).

Patients and Methods

In this blinded, randomised, clinical RSA study, the migration of the Stanmore femoral component was compared between Refobacin bone cement R and Palacos R + G in 62 consecutive total hip arthroplasties. The primary outcome measure was femoral component migration measured using RSA and secondary outcomes were Harris hip score (HHS), Hip disability and Osteoarthritis Outcome Score (HOOS), EuroQol 5D (EQ-5D) and Short Form 36 (SF-36).


Bone & Joint Research
Vol. 5, Issue 8 | Pages 320 - 327
1 Aug 2016
van IJsseldijk EA Valstar ER Stoel BC Nelissen RGHH Baka N van’t Klooster R Kaptein BL

Objectives

An important measure for the diagnosis and monitoring of knee osteoarthritis is the minimum joint space width (mJSW). This requires accurate alignment of the x-ray beam with the tibial plateau, which may not be accomplished in practice. We investigate the feasibility of a new mJSW measurement method from stereo radiographs using 3D statistical shape models (SSM) and evaluate its sensitivity to changes in the mJSW and its robustness to variations in patient positioning and bone geometry.

Materials and Methods

A validation study was performed using five cadaver specimens. The actual mJSW was varied and images were acquired with variation in the cadaver positioning. For comparison purposes, the mJSW was also assessed from plain radiographs. To study the influence of SSM model accuracy, the 3D mJSW measurement was repeated with models from the actual bones, obtained from CT scans.


Bone & Joint Research
Vol. 5, Issue 6 | Pages 215 - 217
1 Jun 2016
Pijls BG Nelissen RGHH


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 391 - 397
1 Mar 2015
van Embden D Stollenwerck GANL Koster LA Kaptein BL Nelissen RGHH Schipper IB

The aim of this study was to quantify the stability of fracture-implant complex in fractures after fixation. A total of 15 patients with an undisplaced fracture of the femoral neck, treated with either a dynamic hip screw or three cannulated hip screws, and 16 patients with an AO31-A2 trochanteric fracture treated with a dynamic hip screw or a Gamma Nail, were included. Radiostereometric analysis was used at six weeks, four months and 12 months post-operatively to evaluate shortening and rotation.

Migration could be assessed in ten patients with a fracture of the femoral neck and seven with a trochanteric fracture. By four months post-operatively, a mean shortening of 5.4 mm (-0.04 to 16.1) had occurred in the fracture of the femoral neck group and 5.0 mm (-0.13 to 12.9) in the trochanteric fracture group. A wide range of rotation occurred in both types of fracture. Right-sided trochanteric fractures seem more rotationally stable than left-sided fractures.

This prospective study shows that migration at the fracture site occurs continuously during the first four post-operative months, after which stabilisation occurs. This information may allow the early recognition of patients at risk of failure of fixation.

Cite this article: Bone Joint J 2015;97-B:391–7.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1520 - 1524
1 Nov 2014
van der Zwaal P Pijls BG Thomassen BJW Lindenburg R Nelissen RGHH van de Sande MAJ

The purpose of this study was to evaluate the natural history of rheumatoid disease of the shoulder over an eight-year period. Our hypothesis was that progression of the disease is associated with a decrease in function with time.

A total of 22 patients (44 shoulders; 17 women, 5 men, (mean age 63)) with rheumatoid arthritis were followed for eight years. All shoulders were assessed using the Constant score, anteroposterior radiographs (Larsen score, Upward-Migration-Index (UMI)) and ultrasound (US). At final follow-up, the Short Form-36, disabilities of the arm, shoulder and hand (DASH) Score, erythrocyte sedimentation rate and use of anti-rheumatic medication were determined.

The mean Constant score was 72 points (50 to 88) at baseline and 69 points (25 to 100) at final follow-up. Radiological evaluation showed progressive destruction of the peri-articular structures with time. This progression of joint and rotator cuff destruction was significantly associated with the Constant score. However, at baseline only the extent of rotator cuff disease and the UMI could predict the Constant score at final follow-up.

A plain anteroposterior radiograph of the shoulder is sufficient to assess any progression of rheumatoid disease and to predict functional outcome in the long term by using the UMI as an indicator of rotator cuff degeneration.

Cite this article: Bone Joint J 2014;96-B:1520–4.


Bone & Joint Research
Vol. 3, Issue 10 | Pages 289 - 296
1 Oct 2014
van IJsseldijk EA Harman MK Luetzner J Valstar ER Stoel BC Nelissen RGHH Kaptein BL

Introduction

Wear of polyethylene inserts plays an important role in failure of total knee replacement and can be monitored in vivo by measuring the minimum joint space width in anteroposterior radiographs. The objective of this retrospective cross-sectional study was to compare the accuracy and precision of a new model-based method with the conventional method by analysing the difference between the minimum joint space width measurements and the actual thickness of retrieved polyethylene tibial inserts.

Method

Before revision, the minimum joint space width values and their locations on the insert were measured in 15 fully weight-bearing radiographs. These measurements were compared with the actual minimum thickness values and locations of the retrieved tibial inserts after revision.


The Bone & Joint Journal
Vol. 96-B, Issue 6 | Pages 765 - 771
1 Jun 2014
Thomassen BJW den Hollander PHC Kaptijn HH Nelissen RGHH Pilot P

We hypothesised there was no clinical value in using an autologous blood transfusion (ABT) drain in either primary total hip (THR) or total knee replacement (TKR) in terms of limiting allogeneic blood transfusions when a modern restrictive blood management regime was followed. A total of 575 patients (65.2% men), with a mean age of 68.9 years (36 to 94) were randomised in this three-arm study to no drainage (group A), or to wound drainage with an ABT drain for either six hours (group B) or 24 hours (group C). The primary outcome was the number of patients receiving allogeneic blood transfusion. Secondary outcomes were post-operative haemoglobin (Hb) levels, length of hospital stay and adverse events.

This study identified only 41 transfused patients, with no significant difference in distribution between the three groups (p = 0.857). The mean pre-operative haemoglobin (Hb) value in the transfused group was 12.8 g/dL (9.8 to 15.5) versus 14.3 g/dL (10.6 to 18.0) in the non-transfused group (p < 0.001, 95% confidence interval: 1.08 to 1.86). Post-operatively, the median of re-transfused shed blood in patients with a THR was 280 mL (Interquartile range (IQR) 150 to 400) and in TKR patients 500 mL (IQR 350 to 650) (p <  0.001). ABT drains had no effect on the proportion of transfused patients in primary THR and TKR. The secondary outcomes were also comparable between groups.

Cite this article: Bone Joint J 2014;96-B:765–71.


Bone & Joint Research
Vol. 3, Issue 2 | Pages 38 - 47
1 Feb 2014
Hogendoorn S Duijnisveld BJ van Duinen SG Stoel BC van Dijk JG Fibbe WE Nelissen RGHH

Objectives

Traumatic brachial plexus injury causes severe functional impairment of the arm. Elbow flexion is often affected. Nerve surgery or tendon transfers provide the only means to obtain improved elbow flexion. Unfortunately, the functionality of the arm often remains insufficient. Stem cell therapy could potentially improve muscle strength and avoid muscle-tendon transfer. This pilot study assesses the safety and regenerative potential of autologous bone marrow-derived mononuclear cell injection in partially denervated biceps.

Methods

Nine brachial plexus patients with insufficient elbow flexion (i.e., partial denervation) received intramuscular escalating doses of autologous bone marrow-derived mononuclear cells, combined with tendon transfers. Effect parameters included biceps biopsies, motor unit analysis on needle electromyography and computerised muscle tomography, before and after cell therapy.


Bone & Joint Research
Vol. 3, Issue 1 | Pages 7 - 13
1 Jan 2014
Keurentjes JC Van Tol FR Fiocco M So-Osman C Onstenk R Koopman-Van Gemert AWMM Pöll RG Nelissen RGHH

Objectives

To define Patient Acceptable Symptom State (PASS) thresholds for the Oxford hip score (OHS) and Oxford knee score (OKS) at mid-term follow-up.

Methods

In a prospective multicentre cohort study, OHS and OKS were collected at a mean follow-up of three years (1.5 to 6.0), combined with a numeric rating scale (NRS) for satisfaction and an external validation question assessing the patient’s willingness to undergo surgery again. A total of 550 patients underwent total hip replacement (THR) and 367 underwent total knee replacement (TKR).


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1656 - 1661
1 Dec 2013
Kraal T van der Heide HJL van Poppel BJ Fiocco M Nelissen RGHH Doets HC

Little is known about the long-term outcome of mobile-bearing total ankle replacement (TAR) in the treatment of end-stage arthritis of the ankle, and in particular for patients with inflammatory joint disease. The aim of this study was to assess the minimum ten-year outcome of TAR in this group of patients.

We prospectively followed 76 patients (93 TARs) who underwent surgery between 1988 and 1999. No patients were lost to follow-up. At latest follow-up at a mean of 14.8 years (10.7 to 22.8), 30 patients (39 TARs) had died and the original TAR remained in situ in 28 patients (31 TARs). The cumulative incidence of failure at 15 years was 20% (95% confidence interval (CI) 11 to 28). The mean American Orthopaedic Foot and Ankle Society (AOFAS) ankle–hindfoot score of the surviving patients at latest follow-up was 80.4 (95% CI 72 to 88). In total, 21 patients (23 TARs) underwent subsequent surgery: three implant exchanges, three bearing exchanges and 17 arthrodeses. Neither design of TAR described in this study, the LCS and the Buechel–Pappas, remains currently available. However, based both on this study and on other reports, we believe that TAR using current mobile-bearing designs for patients with end-stage arthritis of the ankle due to inflammatory joint disease remains justified.

Cite this article: Bone Joint J 2013;95-B:1656–61.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1626 - 1631
1 Dec 2013
van der Weegen W Brakel K Horn RJ Hoekstra HJ Sijbesma T Pilot P Nelissen RGHH

The aim of this study was to establish the natural course of unrevised asymptomatic pseudotumours after metal-on-metal (MoM) hip resurfacing during a six- to 12-month follow-up period. We used repeated metal artefact reduction sequence (MARS)-magnetic resonance imaging (MRI), serum metal ion analysis and clinical examination to study 14 unrevised hips (mean patient age 52.7 years, 46 to 68, 5 female, 7 male) with a pseudotumour and 23 hips (mean patient age 52.8 years, 38 to 69, 7 female, 16 male) without a pseudotumour. The mean post-operative time to the first MARS-MRI scan was 4.3 years (2.2 to 8.3), and mean time between the first and second MARS-MRI scan was eight months (6 to 12). At the second MRI scan, the grade of severity of the pseudotumour had not changed in 35 hips. One new pseudotumour (Anderson C2 score, moderate) was observed, and one pseudotumour was downgraded from C2 (moderate) to C1 (mild). In general, the characteristics of the pseudotumours hardly changed.

Repeated MARS-MRI scans within one year in patients with asymptomatic pseudotumours after MoM hip resurfacing showed little or no variation. In 23 patients without pseudotumour, one new asymptomatic pseudotumour was detected.

This is the first longitudinal study on the natural history of pseudotumours using MARS-MRI scans in hip resurfacing, and mirrors recent results for 28 mm diameter MoM total hip replacement.

Cite this article: Bone Joint J 2013;95-B:1626–31.


Bone & Joint Research
Vol. 2, Issue 11 | Pages 238 - 244
1 Nov 2013
Keurentjes JC Fiocco M So-Osman C Onstenk R Koopman-Van Gemert AWMM Pöll RG Nelissen RGHH

Objectives

Electronic forms of data collection have gained interest in recent years. In orthopaedics, little is known about patient preference regarding pen-and-paper or electronic questionnaires. We aimed to determine whether patients undergoing total hip (THR) or total knee replacement (TKR) prefer pen-and-paper or electronic questionnaires and to identify variables that predict preference for electronic questionnaires.

Methods

We asked patients who participated in a multi-centre cohort study investigating improvement in health-related quality of life (HRQoL) after THR and TKR using pen-and-paper questionnaires, which mode of questionnaire they preferred. Patient age, gender, highest completed level of schooling, body mass index (BMI), comorbidities, indication for joint replacement and pre-operative HRQoL were compared between the groups preferring different modes of questionnaire. We then performed logistic regression analyses to investigate which variables independently predicted preference of electronic questionnaires.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1209 - 1216
1 Sep 2013
van der Voort P Pijls BG Nouta KA Valstar ER Jacobs WCH Nelissen RGHH

Mobile-bearing (MB) total knee replacement (TKR) was introduced to reduce the risk of aseptic loosening and wear of polyethylene inserts. However, no consistent clinical advantages of mobile- over fixed-bearing (FB) TKR have been found. In this study we evaluated whether mobile bearings have an advantage over fixed bearings with regard to revision rates and clinical outcome scores. Furthermore, we determined which modifying variables affected the outcome.

A systematic search of the literature was conducted to collect clinical trials comparing MB and FB in primary TKR. The primary outcomes were revision rates for any reason, aseptic loosening and wear. Secondary outcomes included range of movement, Knee Society score (KSS), Oxford knee score (OKS), Short-Form 12 (SF-12) score and radiological parameters. Meta-regression techniques were used to explore factors modifying the observed effect.

Our search yielded 1827 publications, of which 41 studies met our inclusion criteria, comprising over 6000 TKRs. Meta-analyses showed no clinically relevant differences in terms of revision rates, clinical outcome scores or patient-reported outcome measures between MB and FB TKRs. It appears that theoretical assumptions of superiority of MB over FB TKR are not borne out in clinical practice.

Cite this article: Bone Joint J 2013;95-B:1209–16.


Bone & Joint Research
Vol. 2, Issue 6 | Pages 116 - 121
1 Jun 2013
Duijnisveld BJ Saraç Ç Malessy MJA Brachial Plexus Advisory Board TI Vliet Vlieland TPM Nelissen RGHH

Background

Symptoms of obstetric brachial plexus injury (OBPI) vary widely over the course of time and from individual to individual and can include various degrees of denervation, muscle weakness, contractures, bone deformities and functional limitations. To date, no universally accepted overall framework is available to assess the outcome of patients with OBPI. The objective of this paper is to outline the proposed process for the development of International Classification of Functioning, Disability and Health (ICF) Core Sets for patients with an OBPI.

Methods

The first step is to conduct four preparatory studies to identify ICF categories important for OBPI: a) a systematic literature review to identify outcome measures, b) a qualitative study using focus groups, c) an expert survey and d) a cross-sectional, multicentre study. A first version of ICF Core Sets will be defined at a consensus conference, which will integrate the evidence from the preparatory studies. In a second step, field-testing among patients will validate this first version of Core Sets for OBPI.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 718 - 718
1 May 2013
Pijls BG van der Heide HJL Nelissen RGHH

We welcome letters to the Editor concerning articles that have recently been published. Such letters will be subject to the usual stages of selection and editing; where appropriate the authors of the original article will be offered the opportunity to reply.


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 523 - 529
1 Apr 2013
Henseler JF Nagels J van der Zwaal P Nelissen RGHH

Surgical repair of posterosuperior rotator cuff tears has a poorer outcome and a higher rate of failure compared with repairs of supraspinatus tears. In this prospective cohort study 28 consecutive patients with an irreparable posterosuperior rotator cuff tear after failed conservative or surgical treatment underwent teres major tendon transfer. Their mean age was 60 years (48 to 71) and the mean follow-up was 25 months (12 to 80). The mean active abduction improved from 79° (0° to 150°) pre-operatively to 105° (20° to 180°) post-operatively (p = 0.011). The mean active external rotation in 90° abduction improved from 25° (0° to 70°) pre-operatively to 55° (0° to 90°) post-operatively (p < 0.001). The mean Constant score improved from 43 (18 to 78) pre-operatively to 65 (30 to 86) post-operatively (p < 0.001). The median post-operative VAS (0 to 100) for pain decreased from 63 (0 to 96) pre-operatively to 5 (0 to 56) post-operatively (p < 0.001).

In conclusion, teres major transfer effectively restores function and relieves pain in patients with irreparable posterosuperior rotator cuff tears and leads to an overall clinical improvement in a relatively young and active patient group with limited treatment options.

Cite this article: Bone Joint J 2013;95-B:523–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1366 - 1371
1 Oct 2012
Pijls BG Valstar ER Kaptein BL Nelissen RGHH

This single-blinded randomised controlled trial investigated whether one design of mobile-bearing (MB) total knee replacement (TKR) has any advantage over a fixed-bearing (FB) design on long-term fixation as measured by radiostereometry. The amount of wear underneath the mobile bearing was also evaluated. A series of 42 knees was randomised to MB or FB tibial components with appropriate polyethylene inserts and followed for between ten and 12 years, or until the death of the patient. The polyethylene in the MB group was superior in that it was gamma-irradiated in inert gas and was calcium-stearate free; the polyethylene in the FB group was gamma-irradiated in air and contained calcium stearate. In theory this should be advantageous to the wear rate of the MB group. At final follow-up the overall mean migration was 0.75 mm (sd 0.76) in the MB group and 0.66 mm (sd 0.4) in the FB group, with the FB group demonstrating more posterior tilt and the MB group more internal rotation. In the FB group there was one revision for aseptic loosening, but none in the MB group. There were no significant differences in clinical or radiological scores.

For the MB group, the mean linear wear rate on the under-surface was 0.026 mm/year (sd 0.014). This was significantly smaller than the wear rate of 0.11 mm/year (sd 0.06) in the MB between femur and polyethylene (p < 0.001). Nevertheless, even in a best-case setting the mobile bearings of this TKR design had no apparent advantage in terms of fixation over the FB knee prosthesis at ten to 12 years. The wear underneath the mobile bearing was small and is unlikely to be clinically relevant.


Bone & Joint Research
Vol. 1, Issue 10 | Pages 258 - 262
1 Oct 2012
Keurentjes JC Fiocco M Schreurs BW Pijls BG Nouta KA Nelissen RGHH

Objectives

The Kaplan-Meier estimation is widely used in orthopedics to calculate the probability of revision surgery. Using data from a long-term follow-up study, we aimed to assess the amount of bias introduced by the Kaplan-Meier estimator in a competing risk setting.

Methods

We describe both the Kaplan-Meier estimator and the competing risk model, and explain why the competing risk model is a more appropriate approach to estimate the probability of revision surgery when patients die in a hip revision surgery cohort. In our study, a total of 62 acetabular revisions were performed. After a mean of 25 years, no patients were lost to follow-up, 13 patients had undergone revision surgery and 33 patients died of causes unrelated to their hip.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 2 | Pages 235 - 241
1 Feb 2010
van der Lugt JCT Valstar ER Witvoet-Braam SW Nelissen RGHH

Mechanical loosening which begins with early-onset migration of the prosthesis is the major reason for failure of the Souter-Strathclyde elbow replacement. In a prospective study of 18 Souter-Strathclyde replacements we evaluated the patterns of migration using roentgen stereophotogrammetric analysis. We had previously reported the short-term results after a follow-up of two years which we have now extended to a mean follow-up of 8.2 years (1 to 11.3). Migration was assessed along the co-ordinal axes and overall micromovement was expressed as the maximum total point movement. The alignment of the prosthesis and the presence of radiolucent lines were examined on conventional standardised radiographs.

All the humeral components showed increased and variable patterns of migration at the extended follow-up and four humeral components were revised. The maximum total point movement at two years in the revised prostheses was 1.8 mm (sd 1.0) and in the non-revised 0.7 mm (sd 0.5, p = 0.01). Most humeral components migrated into external rotation resulting in an anterior and varus tilt. The ulnar components remained stable.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 504 - 510
1 Apr 2009
Henkus HE de Witte PB Nelissen RGHH Brand R van Arkel ERA

In a prospective randomised study we compared the results of arthroscopic subacromial bursectomy alone with debridement of the subacromial bursa followed by acromioplasty. A total of 57 patients with a mean age of 47 years (31 to 60) suffering from primary subacromial impingement without a rupture of the rotator cuff who had failed previous conservative treatment were entered into the trial. The type of acromion was classified according to Bigliani. Patients were assessed at follow-up using the Constant score, the simple shoulder test and visual analogue scores for pain and functional impairment. One patient was lost to follow-up.

At a mean follow-up of 2.5 years (1 to 5) both bursectomy and acromioplasty gave good clinical results. No statistically significant differences were found between the two treatments. The type of acromion and severity of symptoms had a greater influence on the clinical outcome than the type of treatment. As a result, we believe that primary subacromial impingement syndrome is largely an intrinsic degenerative condition rather than an extrinsic mechanical disorder.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 4 | Pages 460 - 466
1 Apr 2006
Keizer SB Kock NB Dijkstra PDS Taminiau AHM Nelissen RGHH

This retrospective study describes the long-term results of core decompression and placement of a non-vascularised bone graft in the management of avascular necrosis of the femoral head. We treated 80 hips in 65 patients, 18 by a cortical tibial autograft and 62 by a fibular allograft. The mean age of the patients was 36 years (sd 13.2). A total of 78 hips were available for evaluation of which pre-operatively six were Ficat-Arlet stage 0, three stage I, 31 stage IIA, 16 stage IIB, 13 stage III and nine stage IV.

A total of 34 hips (44%) were revised at a mean of four years (sd 3.8). Survivorship analysis using a clinical end-point showed a survival rate of 59% five years after surgery. We found a significant difference (p = 0.002) in survivorship, when using a clinical and radiological end-point, between the two grafts, in favour of the tibial autograft. We considered this difference to be the result of the better quality and increased volume of tibial bone compared with that from the trochanteric region used with the fibular allograft.

This is a relatively simple, extra-articular and reproducible procedure. In our view core decompression, removal of the necrotic tissue and packing of the cancellous grafts into the core track are vital parts of the procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 2 | Pages 368 - 369
1 Mar 1999
NELISSEN RGHH VALSTAR ER ROZING PM