Aims. The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the risk of revision for instability or loosening. Methods. A total of 202,511 primary total hip arthroplasties (THAs) with uncemented acetabular components were identified from the National Joint Registry (NJR) dataset between 2003 and 2017. The effect of liner geometry on the risk of revision for instability or loosening was investigated using competing risk
Aims. Navigation devices are designed to improve a surgeon’s accuracy in positioning the acetabular and femoral components in total hip arthroplasty (THA). The purpose of this study was to both evaluate the accuracy of an optical computer-assisted surgery (CAS) navigation system and determine whether preoperative spinopelvic mobility (categorized as hypermobile, normal, or stiff) increased the risk of acetabular component placement error. Methods. A total of 356 patients undergoing primary THA were prospectively enrolled from November 2016 to March 2018. Clinically relevant error using the CAS system was defined as a difference of > 5° between CAS and 3D radiological reconstruction measurements for acetabular component inclination and anteversion. Univariate and multiple logistic
Aims. Although there is increasing legalization of the use of cannabis in the USA, few well-powered studies have evaluated the association between cannabis use disorder and outcomes following primary total hip arthroplasty (THA). Thus, the aim of this study was to determine whether patients who use cannabis and undergo primary THA have higher rates of in-hospital length of stay (LOS), medical complications, implant-related complications, and costs. Methods. Using an administrative database, patients with cannabis use disorder undergoing primary THA were matched to a control group in a 1:5 ratio by age, sex, and various medical comorbidities. This yielded 23,030 patients (3,842 in the study group matched with 19,188 in the control group). The variables which were studied included LOS, 90-day medical complications, two-year implant-related complications, and 90-day costs of care. Mann-Whitney U tests were used to compare LOS and costs. Multivariate logistic
Aims. The purpose of this study was to examine whether leg-length discrepancy (LLD) following unilateral total hip arthroplasty (THA) affects the incidence of contralateral head collapse and subsequent THA in patients with bilateral osteonecrosis, and to determine factors associated with subsequent collapse. Patients and Methods. We identified 121 patients with bilateral non-traumatic osteonecrosis who underwent THA between 2003 and 2011 to treat a symptomatic hip, and who also exhibited medium-to-large lesions (necrotic area ≥ 30%) in an otherwise asymptomatic non-operated hip. Of the 121 patients, 71 were male (59%) and 50 were female (41%), with a mean age of 51 years (19 to 71) at the time of initial THA. All patients were followed for at least five years and were assessed according to the presence of a LLD (non-LLD vs LLD group), as well as the LLD type (longer non-operated side vs shorter non-operated side group). Results. Overall, 68 hips (56%) became painful and progressed to collapse at a mean of 2.6 years (0.2 to 13.8), resulting in 59 THAs (49%). The five-year collapse-free survival rate for the non-LLD group was 59% (95% confidence interval (CI) 46.8 to 71.8) compared with 45% (95% CI 32.9 to 57.5) for the LLD group (p = 0.036), and 66% (95% CI 55.2 to 77.2) for the longer non-operated side group compared with 32% (95% CI 19.1 to 44.9) for the shorter non-operated side group (p < 0.001). Multivariate
Aims. The aim of this study was to identify modifiable risk factors associated with mortality in patients requiring revision total hip arthroplasty (THA) for periprosthetic hip fracture. Methods. The electronic records of consecutive patients undergoing revision THA for periprosthetic hip fracture between December 2011 and October 2018 were reviewed. The data which were collected included age, sex, body mass index (BMI), American Society of Anesthesiologists (ASA) classification, the preoperative serum level of haemoglobin, time to surgery, operating time, blood transfusion, length of hospital stay, and postoperative surgical and medical complications. Univariate and multivariate logistic
A prospective cohort of 222 patients who underwent revision hip replacement between April 2001 and March 2004 was evaluated to determine predictors of function, pain and activity level between one and two years post-operatively, and to define quality of life outcomes using validated patient reported outcome tools. Predictive models were developed and proportional odds
Aims. Total hip arthroplasty (THA) is gaining popularity as a treatment for displaced femoral neck fractures (FNFs), especially in physiologically younger patients. While THA for osteoarthritis (OA) has demonstrated low complication rates and increased quality of life, results of THA for acute FNF are not as clear. Currently, a THA performed for FNF is included in an institutional arthroplasty bundle without adequate risk adjustment, potentially placing centres participating in fracture care at financial disadvantage. The purpose of this study is to report on perioperative complication rates after THA for FNF compared with elective THA performed for OA of the hip. Patients and Methods. The American College of Surgeons National Surgical Quality Improvement Program (ACS NSQIP) database between 2008 and 2016 was queried. Patients were identified using the THA Current Procedural Terminology (CPT) code and divided into groups by diagnosis: OA in one and FNF in another. Univariate statistics were performed. Continuous variables were compared between groups using Student’s t-test, and the chi-squared test was used to compare categorical variables. Multivariate and propensity-matched logistic
Objectives. We have encountered patients who developed large joint fluid collections with massive elevations in chromium (Cr) and cobalt (Co) concentrations following metal-on-metal (MoM) hip arthroplasties. In some cases, retrieval analysis determined that these ion concentrations could not be explained simply by the wear rates of the components. We hypothesized that these effects may be associated with aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL). Patients and Methods. We examined the influence of the ALVAL grade on synovial fluid Co and Cr concentrations following adjustment for patient and device variables, including volumetric wear rates. Initially restricting the analysis to include only patients with one MoM hip resurfacing device, we performed multiple
After implementation of a ‘fast-track’ rehabilitation
protocol in our hospital, mean length of hospital stay for primary
total hip arthroplasty decreased from 4.6 to 2.9 nights for unselected
patients. However, despite this reduction there was still a wide
range across the patients’ hospital duration. The purpose of this
study was to identify which specific patient characteristics influence
length of stay after successful implementation of a ‘fast-track’
rehabilitation protocol. A total of 477 patients (317 female and
160 male, mean age 71.0 years; 39.3 to 92.6, mean BMI 27.0 kg/m. 2. ;18.8
to 45.2) who underwent primary total hip arthroplasty between 1
February 2011 and 31 January 2013, were included in this retrospective
cohort study. A length of stay greater than the median was considered
as an increased duration. Logistic
Aims. The aims of this study were to describe the prevalence of previous
lumbar surgery in patients who undergo total hip arthroplasty (THA)
and to investigate their patient-reported outcomes (PROMs) one year
post-operatively. Patients and Methods. Data from the Swedish Hip Arthroplasty Register and the Swedish
Spine Register gathered from 2002 to 2013 were merged to identify
a group of patients who had undergone lumbar surgery before THA
(n = 997) and a carefully matched one-to-one control group. We investigated
differences in the one-year post-operative PROMs between the groups.
Linear
Custom-made partial pelvis replacements (PPRs) are increasingly used in the reconstruction of large acetabular defects and have mainly been designed using a triflange approach, requiring extensive soft-tissue dissection. The monoflange design, where primary intramedullary fixation within the ilium combined with a monoflange for rotational stability, was anticipated to overcome this obstacle. The aim of this study was to evaluate the design with regard to functional outcome, complications, and acetabular reconstruction. Between 2014 and 2023, 79 patients with a mean follow-up of 33 months (SD 22; 9 to 103) were included. Functional outcome was measured using the Harris Hip Score and EuroQol five-dimension questionnaire (EQ-5D). PPR revisions were defined as an endpoint, and subgroups were analyzed to determine risk factors.Aims
Methods
Transfusion after primary total hip arthroplasty (THA) has become rare, and identification of causative factors allows preventive measures. The aim of this study was to determine patient-specific factors that increase the risk of needing a blood transfusion. All patients who underwent elective THA were analyzed retrospectively in this single-centre study from 2020 to 2021. A total of 2,892 patients were included. Transfusion-related parameters were evaluated. A multiple logistic regression was performed to determine whether age, BMI, American Society of Anesthesiologists (ASA) grade, sex, or preoperative haemoglobin (Hb) could predict the need for transfusion within the examined patient population.Aims
Methods
Implant waste during total hip arthroplasty (THA) represents a significant cost to the USA healthcare system. While studies have explored methods to improve THA cost-effectiveness, the literature comparing the proportions of implant waste by intraoperative technology used during THA is limited. The aims of this study were to: 1) examine whether the use of enabling technologies during THA results in a smaller proportion of wasted implants compared to navigation-guided and conventional manual THA; 2) determine the proportion of wasted implants by implant type; and 3) examine the effects of surgeon experience on rates of implant waste by technology used. We identified 104,420 implants either implanted or wasted during 18,329 primary THAs performed on 16,724 patients between January 2018 and June 2022 at our institution. THAs were separated by technology used: robotic-assisted (n = 4,171), imageless navigation (n = 6,887), and manual (n = 7,721). The primary outcome of interest was the rate of implant waste during primary THA.Aims
Methods
The risk factors for abnormal spinopelvic mobility (SPM), defined as an anterior rotation of the spinopelvic tilt (∆SPT) ≥ 20° in a flexed-seated position, have been described. The implication of pelvic incidence (PI) is unclear, and the concept of lumbar lordosis (LL) based on anatomical limits may be erroneous. The distribution of LL, including a unusual shape in patients with a high lordosis, a low pelvic incidence, and an anteverted pelvis seems more relevant. The clinical data of 311 consecutive patients who underwent total hip arthroplasty was retrospectively analyzed. We analyzed the different types of lumbar shapes that can present in patients to identify their potential associations with abnormal pelvic mobility, and we analyzed the potential risk factors associated with a ∆SPT ≥ 20° in the overall population.Aims
Methods
The primary aim of this study was to compare surgical methods (sliding hip screw (SHS) vs intramedullary nailing (IMN)) for trochanteric hip fracture in relation to death within 120 days after surgery and return to independent living. The secondary aim was to assess whether the associations between surgical method and death or ability to return to independent living varied depending on fracture subtype or other patient characteristics. A total of 27,530 individuals from the Swedish Hip Fracture Register RIKSHÖFT (SHR) aged ≥ 70 years, admitted to hospital between 1 January 2014 and 31 December 2019 with trochanteric hip fracture, were included. Within this cohort, 12,041 individuals lived independently at baseline, had follow-up information in the SHR, and were thus investigated for return to independent living. Death within 120 days after surgery was analyzed using Cox regression with SHS as reference and adjusted for age and fracture type. Return to independent living was analyzed using logistic regression adjusted for age and fracture type. Analyses were repeated after stratification by fracture type, age, and sex.Aims
Methods
The aim of this study was to evaluate the suitability of the tapered cone stem in total hip arthroplasty (THA) in patients with excessive femoral anteversion and after femoral osteotomy. We included patients who underwent THA using Wagner Cone due to proximal femur anatomical abnormalities between August 2014 and January 2019 at a single institution. We investigated implant survival time using the endpoint of dislocation and revision, and compared the prevalence of prosthetic impingements between the Wagner Cone, a tapered cone stem, and the Taperloc, a tapered wedge stem, through simulation. We also collected Oxford Hip Score (OHS), visual analogue scale (VAS) satisfaction, and VAS pain by postal survey in August 2023 and explored variables associated with those scores.Aims
Methods
Although periacetabular osteotomies are widely used for the treatment of symptomatic dysplastic hips, long-term surgical outcomes and patient-reported outcome measures (PROMs) are still unclear. Accordingly, we assessed hip survival and PROMs at 20 years after transpositional osteotomy of the acetabulum (TOA). A total of 172 hips in 159 patients who underwent TOA were followed up at a mean of 21.02 years (16.6 to 24.6) postoperatively. Kaplan-Meier analysis was used to assess survivorship with an endpoint of total hip arthroplasty (THA). PROMs included the visual analogue scale (VAS) Satisfaction, VAS Pain, Oxford Hip Score (OHS), and Forgotten Joint Score-12 (FJS-12). Thresholds for favourable outcomes for OHS (≥ 42) and FJS-12 (≥ 51) were obtained using the receiver operating characteristic curve with VAS Satisfaction ≥ 50 and VAS Pain < 20 as anchors.Aims
Methods
Lateral femoral cutaneous nerve (LFCN) injury is a potential complication after the direct anterior approach for total hip arthroplasty (DAA-THA). The aim of this study was to determine how the location of the fasciotomy in DAA-THA affects LFCN injury. In this trial, 134 patients were randomized into a lateral fasciotomy (n = 67) or a conventional fasciotomy (n = 67) group. This study was a dual-centre, double-blind, prospective randomized controlled two-arm trial with parallel group design and a 1:1 allocation ratio. The primary endpoint was the presence of LFCN injury, which was determined by the presence of numbness, decreased sensation, tingling, jolt-like sensation, or pain over the lateral aspect of the thigh, excluding the surgical scar, using a patient-based questionnaire. The secondary endpoints were patient-reported outcome measures (PROMs) using the Western Ontario and McMaster Universities osteoarthritis index (WOMAC), Japanese Orthopaedic Association Hip Disease Evaluation Questionnaire (JHEQ), and the Forgotten Joint Score-12 (FJS-12). Assessments were obtained three months after surgery.Aims
Methods
It is not known whether preservation of the capsule of the hip positively affects patient-reported outcome measures (PROMs) in total hip arthroplasty using the direct anterior approach (DAA-THA). A recent randomized controlled trial found no clinically significant difference at one year postoperatively. This study aimed to determine whether preservation of the anterolateral capsule and anatomical closure improve the outcome and revision rate, when compared with resection of the anterolateral capsule, at two years postoperatively. Two consecutive groups of patients whose operations were performed by the senior author were compared. The anterolateral capsule was resected in the first group of 430 patients between January 2012 and December 2014, and preserved and anatomically closed in the second group of 450 patients between July 2015 and December 2017. There were no other technical changes between the two groups. Patient characteristics, the Charlson Comorbidity Index (CCI), and surgical data were collected from our database. PROM questionnaires, consisting of the Oxford Hip Score (OHS) and Core Outcome Measures Index (COMI-Hip), were collected two years postoperatively. Data were analyzed with generalized multiple regression analysis.Aims
Methods
To find out if there is an inverse association between estimated glomerular filtration rate (eGFR) and whole blood cobalt (Co) and chromium (Cr) levels in patients with metal-on-metal (MoM) hip arthroplasties and renal insufficiency, suggesting that renal insufficiency could cause accumulation of Co and Cr in blood. Out of 2,520 patients with 3,013 MoM hip arthroplasties, we identified 1,244 patients with whole blood Co, Cr, and creatinine measured within no more than a one-year interval. We analyzed the correlation of blood metal ion levels and eGFR to identify a potential trend of accumulating Co or Cr with decreasing eGFR.Aims
Methods
To determine if primary cemented acetabular component geometry (long posterior wall (LPW), hooded, or offset reorientating) influences the risk of revision total hip arthroplasty (THA) for instability or loosening. The National Joint Registry (NJR) dataset was analyzed for primary THAs performed between 2003 and 2017. A cohort of 224,874 cemented acetabular components were included. The effect of acetabular component geometry on the risk of revision for instability or for loosening was investigated using log-binomial regression adjusting for age, sex, American Society of Anesthesiologists grade, indication, side, institution type, operating surgeon grade, surgical approach, polyethylene crosslinking, and prosthetic head size. A competing risk survival analysis was performed with the competing risks being revision for other indications or death.Aims
Methods
This study aimed to describe preoperative waiting times for surgery in hip fracture patients in Norway, and analyze factors affecting waiting time and potential negative consequences of prolonged waiting time. Overall, 37,708 hip fractures in the Norwegian Hip Fracture Register from January 2014 to December 2018 were linked with data in the Norwegian Patient Registry. Hospitals treating hip fractures were characterized according to their hip fracture care. Waiting time (hours from admission to start of surgery), surgery within regular working hours, and surgery on the day of or on the day after admission, i.e. ‘expedited surgery’ were estimated.Aims
Methods
The main aims were to identify risk factors predictive of a radiolucent line (RLL) around the acetabular component with an interface bioactive bone cement (IBBC) technique in the first year after THA, and evaluate whether these risk factors influence the development of RLLs at five and ten years after THA. A retrospective review was undertaken of 980 primary cemented THAs in 876 patients using cemented acetabular components with the IBBC technique. The outcome variable was any RLLs that could be observed around the acetabular component at the first year after THA. Univariate analyses with univariate logistic regression and multivariate analyses with exact logistic regression were performed to identify risk factors for any RLLs based on radiological classification of hip osteoarthritis.Aims
Methods
The aim of this study was to determine whether national standards of best practice are associated with improved health-related quality of life (HRQoL) outcomes in hip fracture patients. This was a multicentre cohort study conducted in 20 acute UK NHS hospitals treating hip fracture patients. Patients aged ≥ 60 years treated operatively for a hip fracture were eligible for inclusion. Regression models were fitted to each of the “Best Practice Tariff” indicators and overall attainment. The impact of attainment on HRQoL was assessed by quantifying improvement in EuroQol five-dimension five-level questionnaire (EQ-5D-5L) from estimated regression model coefficients.Aims
Methods
Total hip arthroplasty (THA) patients undergoing or having a prior lumbar spine fusion (LSF) have an increased risk of mechanical complications. The aim of this registry-based, retrospective comparative cohort study is to assess the longer term survival of THA in patients who have undergone a LSF during a 17-year period (2000 to 2017). A registry-based population study was conducted on 679 patients who underwent both THA and LSF surgeries. Patients were identified from the regional arthroplasty data base and cross linked to patients with LSF from the regional hospital discharge database between 2000 and 2017. Demographic data, diagnosis leading to primary THA, primary implant survival, perioperative complications, number and causes of failure, and patients requiring revision arthroplasty were collated and compared. For comparison, data from 67,919 primary THAs performed during the same time time period were also retrieved and analyzed.Aims
Methods
There is evidence that prior lumbar fusion increases the risk of dislocation and revision after total hip arthroplasty (THA). The relationship between prior lumbar fusion and the effect of femoral head diameter on THA dislocation has not been investigated. We examined the relationship between prior lumbar fusion or discectomy and the risk of dislocation or revision after THA. We also examined the effect of femoral head component diameter on the risk of dislocation or revision. Data used in this study were compiled from several Finnish national health registers, including the Finnish Arthroplasty Register (FAR) which was the primary source for prosthesis-related data. Other registers used in this study included the Finnish Health Care Register (HILMO), the Social Insurance Institutions (SII) registers, and Statistics Finland. The study was conducted as a prospective retrospective cohort study. Cox proportional hazards regression and Kaplan-Meier survival analysis were used for analysis.Aims
Methods
Dual mobility (DM) bearings are an attractive treatment option to obtain hip stability during challenging primary and revision total hip arthroplasty (THA) cases. The purpose of this study was to analyze data submitted to the American Joint Replacement Registry (AJRR) to characterize utilization trends of DM bearings in the USA. All primary and revision THA procedures reported to AJRR from 2012 to 2018 were analyzed. Patients of all ages were included and subdivided into DM and traditional bearing surface cohorts. Patient demographics, geographical region, hospital size, and teaching affiliation were assessed. Associations were determined by chi-squared analysis and logistic regression was performed to assess outcome variables.Aims
Methods
Although good clinical outcomes have been reported for monolithic tapered, fluted, titanium stems (TFTS), early results showed high rates of subsidence. Advances in stem design may mitigate these concerns. This study reports on the use of a current monolithic TFTS for a variety of indications. A multi-institutional retrospective study of all consecutive total hip arthroplasty (THA) and revision total hip arthroplasty (rTHA) patients who received the monolithic TFTS was conducted. Surgery was performed by eight fellowship-trained arthroplasty surgeons at four institutions. A total of 157 hips in 153 patients at a mean follow-up of 11.6 months (SD7.8) were included. Mean patient age at the time of surgery was 67.4 years (SD 13.3) and mean body mass index (BMI) was 28.9 kg/m2 (SD 6.5). Outcomes included intraoperative complications, one-year all-cause re-revisions, and subsidence at postoperative time intervals (two weeks, six weeks, six months, nine months, and one year).Aims
Methods
The aim of this study was to assess the influence of operating time on 30-day complications following total hip arthroplasty (THA). We identified patients aged 18 years and older who underwent THA between 2006 and 2016 from the American College of Surgeons National Surgical Quality Improvement Program (NSQIP) database. We identified 131 361 patients, with a mean age of 65 years (Aims
Patients and Methods
The aim of this study was to compare patient-reported outcome measures (PROMs), radiological measurements, and total hip arthroplasty (THA)-free survival in patients who underwent periacetabular osteotomy (PAO) for mild, moderate, or severe developmental dysplasia of the hip. We performed a retrospective study involving 336 patients (420 hips) who underwent PAO by a single surgeon at an academic centre. After exclusions, 124 patients (149 hips) were included. The preoperative lateral centre-edge angle (LCEA) was used to classify the severity of dysplasia: 18° to 25° was considered mild (n = 20), 10° to 17° moderate (n = 66), and < 10° severe (n = 63). There was no difference in patient characteristics between the groups (all, p > 0.05). Pre- and postoperative radiological measurements were made. The National Institute of Health’s Patient Reported Outcomes Measurement Information System (PROMIS) outcome measures (physical function computerized adaptive test (PF CAT), Global Physical and Mental Health Scores) were collected. Failure was defined as conversion to THA or PF CAT scores < 40, and was assessed with Kaplan–Meier analysis. The mean follow-up was five years (2 to 10) ending in either failure or the latest contact with the patient.Aims
Patients and Methods
Increasingly, patients with bilateral hip arthritis wish to undergo staged total hip arthroplasty (THA). With the rise in demand for arthroplasty, perioperative risk assessment and counselling is crucial for shared decision making. However, it is unknown if complications that occur after a unilateral hip arthroplasty predict complications following surgery of the contralateral hip. We used nationwide linked discharge data from the Healthcare Cost and Utilization Project between 2005 and 2014 to analyze the incidence and recurrence of complications following the first- and second-stage operations in staged bilateral total hip arthroplasty (BTHAs). Complications included perioperative medical adverse events within 30 to 60 days, and infection and mechanical complications within one year. Conditional probabilities and odds ratios (ORs) were calculated to determine whether experiencing a complication after the first stage of surgery increased the risk of developing the same complication after the second stage.Aims
Patients and Methods
The aim of this study was to investigate the differences in 30-day outcomes between patients undergoing revision for an infected total hip arthroplasty (THA) compared with an aseptic revision THA. This was a retrospective review of prospectively collected data from the American College of Surgeons National Surgical Quality Improvement Program (ACS NSQIP) database, between 2012 and 2017, using Current Procedural Terminology (CPT) codes for patients undergoing a revision THA (27134, 27137, 27138). International Classification of Diseases Ninth Revision/Tenth Revision (ICD-9-CM, ICD-10-CM) diagnosis codes for infection of an implant or device were used to identify patients undergoing an infected revision THA. CPT-27132 coupled with ICD-9-CM/ICD-10-CM codes for infection were used to identify patients undergoing a two-stage revision. A total of 13 556 patients were included; 1606 (11.8%) underwent a revision THA due to infection and there were 11 951 (88.2%) aseptic revisions.Aims
Patients and Methods
Previous studies have suggested that metal-on-metal (MoM) Pinnacle (DePuy Synthes, Warsaw, Indiana) hip arthroplasties implanted after 2006 exhibit higher failure rates. This was attributed to the production of implants with reduced diametrical clearances between their bearing surfaces, which, it was speculated, were outside manufacturing tolerances. This study aimed to better understand the performance of Pinnacle Systems manufactured before and after this event. A total of 92 retrieved MoM Pinnacle hips were analyzed, of which 45 were implanted before 2007, and 47 from 2007 onwards. The ‘pre-2007’ group contained 45 implants retrieved from 21 male and 24 female patients, with a median age of 61.3 years (interquartile range (IQR) 57.1 to 65.5); the ‘2007 onwards’ group contained 47 implants retrieved from 19 male and 28 female patients, with a median age of 61.8 years (IQR 58.5 to 67.8). The volume of material lost from their bearing and taper surfaces was measured using coordinate and roundness measuring machines. These outcomes were then compared statistically using linear regression models, adjusting for potentially confounding factors.Objectives
Methods
Aims
Patients and Methods
There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs. We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling.Aims
Patients and Methods
Accurate placement of the acetabular component is essential in
total hip arthroplasty (THA). The purpose of this study was to determine
if the ability to achieve inclination of the acetabular component
within the ‘safe-zone’ of 30° to 50° could be improved with the
use of an inclinometer. We reviewed 167 primary THAs performed by a single surgeon over
a period of 14 months. Procedures were performed at two institutions:
an inpatient hospital, where an inclinometer was used (inclinometer
group); and an ambulatory centre, where an inclinometer was not
used as it could not be adequately sterilized (control group). We excluded
47 patients with a body mass index (BMI) of > 40 kg/m2,
age of > 68 years, or a surgical indication other than osteoarthritis
whose treatment could not be undertaken in the ambulatory centre.
There were thus 120 patients in the study, 68 in the inclinometer
group and 52 in the control group. The inclination angles of the acetabular
component were measured from de-identified plain radiographs by
two blinded investigators who were not involved in the surgery.
The effect of the use of the inclinometer on the inclination angle
was determined using multivariate regression analysis.Aims
Patients and Methods
The influence of identifiable pre-operative factors on the outcome
of eccentric rotational acetabular osteotomy (ERAO) is unknown.
We aimed to determine the factors that might influence the outcome,
in order to develop a scoring system for predicting the prognosis
for patients undergoing this procedure. We reviewed 700 consecutive ERAOs in 54 men and 646 women with
symptomatic acetabular dysplasia or early onset osteoarthritis (OA)
of the hip, which were undertaken between September 1989 and March
2013. The patients’ pre-operative background, clinical and radiological
findings were examined retrospectively. Multivariate Cox regression
analysis was performed using the time from the day of surgery to
a conversion to total hip arthroplasty (THA) as an endpoint. A risk
score was calculated to predict the prognosis for conversion to
THA, and its predictive capacity was investigated.Aims
Patients and Methods
Periprosthetic fracture (PF) after primary total hip arthroplasty
(THA) is an uncommon but potentially devastating complication. This
study aims to investigate the influence of cemented stem designs
on the risk of needing a revision for a PF. We analysed data on 257 202 primary THAs with cemented stems
and 390 linked first revisions for PF recorded in the National Joint
Registry (NJR) of England, Wales and Northern Ireland to determine
if a cemented femoral stem brand was associated with the risk of
having revision for a PF after primary THA. All cemented femoral
stem brands with more than 10 000 primary operations recorded in
the NJR were identified. The four most commonly used cemented femoral
stems were the Exeter V40 (n = 146 409), CPT (n = 24 300), C-Stem
(n = 15 113) and Charnley (n = 20 182). We compared the revision risk ratios due to PF amongst the stems
using a Poisson regression model adjusting for patient factors.
Compared with the Exeter V40, the age, gender and ASA grade adjusted
revision rate ratio was 3.89 for the cemented CPT stem (95% confidence
interval (CI) 3.07 to 4.93), 0.89 for the C-Stem (95% CI 0.57 to
1.41) and 0.41 for the Charnley stem (95% CI 0.24 to 0.70).Aims
Patients and Methods
We compared the incidence of pseudotumours after
large head metal-on-metal (MoM) total hip arthroplasty (THA) with
that after conventional metal-on-polyethylene (MoP) THA and assessed
the predisposing factors to pseudotumour formation. From a previous randomised controlled trial which compared large
head (38 mm to 60 mm) cementless MoM THA with conventional head
(28 mm) cementless MoP THA, 93 patients (96 THAs: 41 MoM (21 males,
20 females, mean age of 64 years, standard deviation ( The incidence of pseudotumours, measured using a standardised
CT protocol was 22 (53.7%) after MoM THA and 12 (21.8%) after MoP
THA. Women with a MoM THA were more likely to develop a pseudotumour
than those with a MoP THA (15 Contrary to popular belief, pseudotumours occur frequently around
MoP THAs. Women with a MoM THA and an elevated cobalt level are
at greatest risk. In this study, pseudotumours had no effect on
the functional outcome after either large head MoM or conventional
MoP THA. Cite this article:
Total hip replacement causes a short-term increase
in the risk of mortality. It is important to quantify this and to identify
modifiable risk factors so that the risk of post-operative mortality
can be minimised. We performed a systematic review and critical
evaluation of the current literature on the topic. We identified
32 studies published over the last 10 years which provide either
30-day or 90-day mortality data. We estimate the pooled incidence
of mortality during the first 30 and 90 days following hip replacement
to be 0.30% (95% CI 0.22 to 0.38) and 0.65% (95% CI 0.50 to 0.81),
respectively. We found strong evidence of a temporal trend towards
reducing mortality rates despite increasingly co-morbid patients.
The risk factors for early mortality most commonly identified are
increasing age, male gender and co-morbid conditions, particularly
cardiovascular disease. Cardiovascular complications appear to have
overtaken fatal pulmonary emboli as the leading cause of death after
hip replacement. Cite this article:
We present the development and results of a nationwide, prospective, observational follow-up programme including patient-reported outcome measures (PROMs) for the Swedish Hip Arthroplasty Register. The programme started in 2002 and has gradually expanded to include all units performing total hip replacement in Sweden. The self-administered PROMs protocol comprises the EQ-5D instrument, the Charnley class categorisation and visual analogue scales for pain and satisfaction. These current analyses include 34 960 total hip replacements with complete pre- and one-year post-operative questionnaires. Patients eligible for total hip replacement generally report low health-related quality of life and suffer from pain. One year post-operatively the mean EQ-5D index increased to above the level of an age- and gender-matched population, with a considerable reduction of pain (p <
0.001). Females, younger patients and those with Charnley category C reported a lower EQ-5D index pre-operatively than males, older patients and Charnley category A or B, respectively (all p <
0.001). In a multivariable regression analysis Charnley category C, male gender and higher age were associated with less improvement in health-related quality of life (p <
0.001). Nationwide implementation of a PROMs programme requires a structured organisation and effective data capture. Patients’ response rates to the Registry are good. The continuous collection of PROMs permits local and national improvement work and allows for further health-economic evaluation.
The Norwegian Arthroplasty Register has shown that several designs of uncemented femoral stems give good or excellent survivorship. The overall findings for uncemented total hip replacement however, have been disappointing because of poor results with the use of metal-backed acetabular components. In this study, we exclusively investigated the medium-to long-term performance of primary uncemented metal-backed acetabular components. A total of 9113 primary uncemented acetabular components were implanted in 7937 patients between 1987 and 2007. These were included in a prospective, population-based observational study. All the implants were modular and metal-backed with ultra-high-molecular-weight polyethylene liners. The femoral heads were made of stainless steel, cobalt-chrome (CoCr) alloy or alumina ceramic. In all, seven different designs of acetabular component were evaluated by the Kaplan-Meier survivorship method and Cox regression analysis. Most acetabular components performed well up to seven years. When the endpoint was revision of the acetabular component because of aseptic loosening, the survival ranged between 87% and 100% at ten years. However, when the endpoint was revision for any reason, the survival estimates were 81% to 92% for the same implants at ten years. Aseptic loosening, wear, osteolysis and dislocation were the main reasons for the relatively poor overall performance of the acetabular components. Prostheses with alumina heads performed slightly better than those with stainless steel or CoCr alloy in subgroups. Whereas most acetabular components performed well at seven years, the survivorship declined with longer follow-up. Fixation was generally good. None of the metal-backed uncemented acetabular components with ultra-high-molecular-weight polyethylene liners in our study had satisfactory long-term results because of high rates of wear, osteolysis, aseptic loosening and dislocation.
The systemic use of steroids and habitual alcohol
intake are two major causative factors in the development of idiopathic
osteonecrosis of the femoral head (ONFH). To examine any interaction
between oral corticosteroid use and alcohol intake on the risk of
ONFH, we conducted a hospital-based case-control study of 71 cases
with ONFH (mean age 45 years (20 to 79)) and 227 matched controls
(mean age 47 years (18 to 79)). Alcohol intake was positively associated
with ONFH among all subjects: the adjusted odds ratio (OR) of subjects
with ≥ 3032 drink-years was 3.93 (95% confidence interval (CI) 1.18
to 13.1) compared with never-drinkers. When stratified by steroid use,
the OR of such drinkers was 11.1 (95% CI 1.30 to 95.5) among those
who had never used steroids, but 1.10 (95% CI 0.21 to 4.79) among
those who had. When we assessed any interaction based on a two-by-two
table of alcohol and steroid use, the OR of those non-drinkers who
did use steroids was markedly elevated (OR 31.5) compared with users
of neither. However, no further increase in OR was noted for the
effect of using both (OR 31.6). We detected neither a multiplicative
nor an additive interaction (p for multiplicative interaction 0.19;
synergy index 0.95), suggesting that the added effect of alcohol
may be trivial compared with the overwhelming effect of steroids
in the development of ONFH. Cite this article:
This prospective study assessed the effect of social deprivation on the Oxford hip score at one year after total hip replacement. An analysis of 1312 patients undergoing 1359 primary total hip replacements for symptomatic osteoarthritis was performed over a 35-month period. Social deprivation was assessed using the Carstairs index. Those patients who were most deprived underwent surgery at an earlier age (p = 0.04), had more comorbidities (p = 0.02), increased severity of symptoms at presentation (p = 0.001), and were not as satisfied with their outcome (p = 0.03) compared with more affluent patients. There was a significant improvement in Oxford scores at 12 months relative to pre-operative scores for all socioeconomic categories (p <
0.001). Social deprivation was a significant independent predictor of mean improvement in Oxford scores at 12 months, after adjusting for confounding variables (p = 0.001). Deprivation was also associated with an increased risk of dislocation (odds ratio 5.3, p <
0.001) and mortality at 90 days (odds ratio 3.2, p = 0.02). Outcome, risk of dislocation and early mortality after a total hip replacement are affected by the socioeconomic status of the patient
We have evaluated the extent to which diabetes affects the revision rate following total hip replacement (THR). Through the Danish Hip Arthroplasty Registry we identified all patients undergoing a primary THR (n = 57 575) between 1 January 1996 and 31 December 2005, of whom 3278 had diabetes. The presence of diabetes among these patients was identified through the Danish National Registry of Patients and the Danish National Drug Prescription Database. We estimated the relative risk for revision and the 95% confidence intervals for patients with diabetes compared to those without, adjusting for the confounding factors. Diabetes is associated with an increased risk of revision due to deep infection (relative risk = 1.45 (95% confidence interval 1.00 to 2.09), particularly in those with type 2 diabetes (relative risk = 1.49 (95% confidence interval 1.02 to 2.18)), those with diabetes for less than five years prior to THR (relative risk = 1.69 (95% confidence interval 1.24 to 2.32)), those with complications due to diabetes (relative risk = 2.11 (95% confidence interval 1.41 to 3.17)), and those with cardiovascular comorbidities prior to surgery (relative risk = 2.35 (95% confidence interval 1.39 to 3.98)). Patients and surgeons should be aware of the relatively elevated risk of revision due to deep infection following THR in diabetes particularly in those with insufficient control of their glucose level.
Metal ions generated from joint replacements are a cause for concern. There is no consensus on the best surrogate measure of metal ion exposure. This study investigates whether serum and whole blood concentrations can be used interchangeably to report results of cobalt and chromium ion concentrations. Concentrations of serum and whole blood were analysed in 262 concurrent specimens using high resolution inductively-coupled plasma mass-spectrometry. The agreement was assessed with normalised scatterplots, mean difference and the Bland and Altman limits of agreement. The wide variability seen in the normalised scatterplots, in the Bland and Altman plots and the statistically significant mean differences between serum and whole blood concentrations suggest that they cannot be used interchangeably. A bias was demonstrated for both ions in the Bland-Altman plots. Regression analysis provided a possible conversion factor of 0.71 for cobalt and 0.48 for chromium. However, even when the correction factors were applied, the limits of agreement were greater than ±67% for cobalt and greater than ±85% for chromium, suggesting that serum and whole blood cannot be used interconvertibly. This suggests that serum metal concentrations are not useful as a surrogate measure of systemic metal ion exposure.
We have analysed from the Finnish Arthroplasty Register the long-term survivorship of the 12 most commonly-used cemented implants between 1980 and 2005 in patients aged 55 years or older with osteoarthritis. Only two designs of femoral component, the Exeter Universal and the Müller Straight femoral component had a survivorship of over 95% at ten years with revision for aseptic loosening as the endpoint. At 15 years of the femoral and acetabular component combinations, only the Exeter Universal/Exeter All-poly implant had a survival rate of over 90% with revision for aseptic loosening as the endpoint. In the subgroup of patients aged between 55 and 64 years, survivorship overall was less than 90% at ten years. The variation in the long-term rates of survival of different cemented hip implants was considerable in patients aged 55 years or older. In those aged between 55 and 64 years, none of the cemented prostheses studied yielded excellent long-term survival rates (≥ 90% at 15 years).
Primary uncemented femoral stems reported to the Norwegian arthroplasty register between 1987 and 2005 were included in this prospective observational study. There were 11 516 hips (9679 patients) and 14 different designs of stem. Kaplan-Meier survival probabilities and Cox regression were used to analyse the data. With aseptic loosening as the end-point, all currently used designs performed excellently with survival of 96% to 100% at ten years. With the end-point as stem revision for any cause, the long-term results of the different designs varied from poor to excellent, with survival at 15 years ranging between 29% and 97%. Follow-up for longer than seven years was needed to identify some of the poorly-performing designs. There were differences between the stems; the Corail, used in 5456 hips, was the most frequently used stem with a survival of 97% at 15 years. Male gender was associated with an increased risk of revision of × 1.3 (95% confidence interval 1.05 to 1.52), but age and diagnosis had no influence on the results. Overall, modern uncemented femoral stems performed well. Moderate differences in survival between well-performing stems should be interpreted with caution since the differences may be caused by factors other than the stem itself.
We evaluated the concentrations of chromium and cobalt ions in blood after metal-on-metal surface replacement arthroplasty using a wrought-forged, high carbon content chromium-cobalt alloy implant in 64 patients. At one year, mean whole blood ion levels were 1.61 μg/L (0.4 to 5.5) for chromium and 0.67 μg/L (0.23 to 2.09) for cobalt. The pre-operative ion levels, component size, female gender and the inclination of the acetabular component were inversely proportional to the values of chromium and/or cobalt ions at one year postoperatively. Other factors, such as age and level of activity, did not correlate with the levels of metal ions. We found that the levels of the ions in the serum were 1.39 and 1.37 times higher for chromium and cobalt respectively than those in the whole blood. The levels of metal ions obtained may be specific to the hip resurfacing implant and reflect its manufacturing process.