From 1984 to 1995, 68 ankylosed elbows and 11 which were unstable after trauma were replaced in India by Baksi sloppy hinge prostheses. The mean age of the patients was 28.6 years (17 to 70) and the mean follow-up 9.6 years (2 to 13.5). Of the 68 ankylosed elbows, 59 (87%) regained a mean arc of painless movement of 88.5° (27 to 115). The mean improvement of supination was 24° and of pronation 16.5°. There were 54 good results (80%), eight fair and three poor. There were two complete failures due to infection, and one due to a broken humeral stem. Of the 11 unstable elbows, the nine with good results had a mean arc of 125° (15 to 140) of painless stable movement, with a mean improvement in supination of 26° and of pronation of 19.5°. There was one fair result and one failure due to loosening with subsequent late infection. There were significant complications in 14 cases with infection in seven and aseptic loosening in four. Patients with loosening or late removal of the prosthesis often retained reasonably stable elbow movement because periprosthetic fibrosis had connected the approximated bone ends, and muscle balance had been restored.
We reviewed the outcome of 28 patients who had been treated using the Aequalis fracture prosthesis for an acute fracture of the proximal humerus at a mean follow-up of 39.3 months (24 to 63). The mean age of the patients at the time of the fracture was 66.3 years (38 to 80). The mean Constant score was 68.2 (37 to 84) for the operated shoulder, which represented 89.5% of the mean score for the uninjured side (p <
0.001). The quality of the reconstruction as shown on the immediate post-operative radiographs was categorised into three types, anatomical, acceptable, and unacceptable, depending on the position of the tuberosities relative to the
Replacement of the shoulder in juvenile idiopathic arthritis is not often performed and there have been no published series to date. We present nine glenohumeral hemiarthroplasties in eight patients with systemic or polyarticular juvenile idiopathic arthritis. The mean follow-up was six years (59 to 89 months). The mean age at the time of surgery was 32 years. Surgery took place at a mean of 27 years after diagnosis. The results indicated excellent relief from pain. There was restoration of useful function which deteriorated with time, in part because of progression of the systemic disease in this severely affected group. No patient has required revision to date and there has been no radiological evidence of loosening or osteolysis around the implants. We discuss the pathoanatomical challenges unique to this group. There was very little space for a
Mason type III fractures of the radial head are treated by open reduction and internal fixation, resection or
A reversed Delta III total shoulder prosthesis was retrieved post-mortem, eight months after implantation. A significant notch was evident at the inferior pole of the scapular neck which extended beyond the inferior fixation screw. This bone loss was associated with a corresponding, erosive defect of the polyethylene cup. Histological examination revealed a chronic foreign-body reaction in the joint capsule. There were, however, no histological signs of loosening of the glenoid base plate and the stability of the
Between 1978 and 1986, 59 patients received a GSB III elbow prosthesis, six of them in both elbows. Rheumatoid arthritis (RA) was the underlying cause in 51 of the patients and post-traumatic osteoarthritis (PTOA) in eight. Of these, 24 patients (28 prostheses) have since died; two, both operated on bilaterally, had had their implants for more than ten years and had already been assessed for inclusion in the long-term follow-up. Two patients, each with one elbow prosthesis, have been lost to follow-up and three males who are still living (two with PTOA, one with juvenile RA) had their prosthesis removed before ten years had elapsed. The remaining 32 patients (28 RA, 4 PTOA) with 36 GSB III elbows were examined clinically and radiologically after a mean period of 13.5 years. Pain was considerably reduced in 91.6%. Mobility was increased by 37° in those with RA and by 67° in those with PTOA. There were three cases of aseptic loosening and three of deep infection. The main complication was disassembly of the
We report the effectiveness of revision of total
elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows
in 52 patients were treated with re-cementing of a total elbow replacement
into part or all of the existing cement mantle or into the debrided
host-bone interface, without the use of structural bone augmentation
or a custom prosthesis. The original implant revision was still
We retrospectively reviewed 11 consecutive patients with an infected reverse shoulder prosthesis. Patients were assessed clinically and radiologically, and standard laboratory tests were carried out. Peroperative samples showed Propionbacterium acnes in seven, coagulase-negative Staphylococcus in five, methicillin-resistant A one-stage revision arthroplasty reduces the cost and duration of treatment. It is reliable in eradicating infection and good functional outcomes can be achieved.
Radiological changes and differences between cemented and uncemented components of Grammont reverse shoulder arthroplasties (DePuy) were analysed at a mean follow-up of 9.6 years (8 to 12). Of 122 reverse shoulder arthroplasties implanted in five shoulder centres between 1993 and 2000, a total of 68 (65 patients) were available for study. The indications for reversed shoulder arthroplasty were cuff tear arthropathy in 48 shoulders, revision of shoulder prostheses of various types in 11 and massive cuff tear in nine. The development of scapular notching, bony scapular spur formation, heterotopic ossification, glenoid and humeral radiolucencies, stem subsidence, radiological signs of stress shielding and resorption of the tuberosities were assessed on standardised true anteroposterior and axillary radiographs. A scapular notch was observed in 60 shoulders (88%) and was associated with the superolateral approach (p = 0.009). Glenoid radiolucency was present in 11 (16%), bony scapular spur and/or ossifications in 51 (75%), and subsidence of the stem and humeral radiolucency in more than three zones were present in three (8.8%) and in four (11.8%) of 34 cemented components, respectively, and in one (2.9%) and two (5.9%) of 34 uncemented components, respectively. Radiological signs of stress shielding were significantly more frequent with uncemented components (p <
0.001), as was resorption of the greater (p <
0.001) and lesser tuberosities (p = 0.009).
Between 1976 and 2004, 38 revision arthroplasties (35 patients) were performed for aseptic loosening of the humeral component. The mean interval from primary arthroplasty to revision was 7.1 years (0.4 to 16.6). A total of 35 shoulders (32 patients) were available for review at a mean follow-up of seven years (2 to 19.3). Pre-operatively, 34 patients (97%) had moderate or severe pain; at final follow-up, 29 (83%) had no or only mild pain (p <
0.0001). The mean active abduction improved from 88° to 107° (p <
0.01); and the mean external rotation from 37° to 46° (p = 0.27). Excellent or satisfactory results were achieved in 25 patients (71%) according to the modified Neer rating system. Humeral components were cemented in 29, with ingrowth implants used in nine cases. There were 19 of standard length and 17 were longer (two were custom replacements and are not included). Bone grafting was required for defects in 11 humeri. Only two glenoid components were left unrevised. Intra-operative complications included cement extrusion in eight cases, fracture of the shaft of the humerus is two and of the tuberosity in four. There were four re-operations, one for recurrent humeral loosening, with 89% survival free of re-operations at ten years. Revision surgery for aseptic loosening of the humeral component provides reliable pain relief and modest improvement of movement, although there is a substantial risk of intra-operative complications. Revision to a total shoulder replacement gives better results than to a hemiarthroplasty.
The results of proximal humeral replacement following trauma are substantially worse than for osteoarthritis or rheumatoid arthritis. The stable reattachment of the lesser and greater tuberosity fragments to the rotator cuff and the restoration of shoulder biomechanics are difficult. In 1992 we developed a prosthesis designed to improve fixation of the tuberosity fragments in comminuted fractures of the proximal humerus. The implant enables fixation of the fragments to the shaft of the prosthesis and the diaphyseal fragment using screws, washers and a special toothed plate. Between 1992 and 2003 we used this technique in 50 of 76 patients referred to our institution for shoulder reconstruction after trauma. In the remaining 26, reconstruction with a prosthesis and nonabsorbable sutures was performed, as the tuberosity fragments were too small and too severely damaged to allow the use of screws and the toothed plate. The Constant score two years post-operatively was a mean of 12 points better in the acute trauma group and 11 points better in the late post-traumatic group than in the classical suture group. We recommend this technique in patients where the tuberosity fragments are large enough to allow fixation with screws, washers and a toothed plate.
Bone loss involving articular surface is a challenging
problem faced by the orthopaedic surgeon. In the hand and wrist,
there are articular defects that are amenable to autograft reconstruction
when primary fixation is not possible. In this article, the surgical
techniques and clinical outcomes of articular reconstructions in
the hand and wrist using non-vascularised osteochondral autografts
are reviewed.
We assessed the short- to mid-term survival of
metallic press-fit radial head prostheses in patients with radial
head fractures and acute traumatic instability of the elbow. The medical records of 42 patients (16 males, 26 females) with
a mean age of 56 years (23 to 85) with acute unstable elbow injuries,
including a fracture of the radial head requiring metallic replacement
of the radial head, were reviewed retrospectively. Survival of the
prosthesis was assessed from the radiographs of 37 patients after
a mean follow-up of 50 months (12 to 107). The functional results
of 31 patients were assessed using range-of-movement, Mayo elbow
performance score (MEPS), Disabilities of the Arm, Shoulder and
Hand (DASH) score and the RAND 36-item health survey. At the most recent follow-up 25 prostheses were still well fixed,
nine had been removed because of loosening, and three remained implanted
but were loose. The mean time from implantation to loosening was
11 months (2 to 24). Radiolucent lines that developed around the
prosthesis before removal were mild in three patients, moderate
in one and severe in five. Range of movement parameters and mass
grip strength were significantly lower in the affected elbow than
in the unaffected side. The mean MEPS score was 86 (40 to 100) and
the mean DASH score was 23 (0 to 81). According to RAND-36 scores,
patients had more pain and lower physical function scores than normal population
values. Loosening of press-fit radial head prostheses is common, occurs
early, often leads to severe osteolysis of the proximal radius,
and commonly requires removal of the prosthesis.
The outcome of an anatomical shoulder replacement
depends on an intact rotator cuff. In 1981 Grammont designed a novel
large-head reverse shoulder replacement for patients with cuff deficiency.
Such has been the success of this replacement that it has led to
a rapid expansion of the indications. We performed a systematic
review of the literature to evaluate the functional outcome of each
indication for the reverse shoulder replacement. Secondary outcome
measures of range of movement, pain scores and complication rates
are also presented.
This study reviews the predisposing features, the clinical, and laboratory findings at the time of diagnosis and the results of single-stage revision of prosthetic replacement of the elbow for infection. Deep infection occurred in six of 305 (1.9%) primary total elbow replacements. The mean follow-up after revision was 6.8 years (6 months to 16 years) and the mean age at the time of revision was 62.7 years (56 to 74). All six cases with infection had rheumatoid arthritis and had received steroid therapy. The infective organism was
We compared six patients with a mean age of 70 years (49 to 80) with severe bilateral, painful glenohumeral joint destruction who underwent a single-stage bilateral total shoulder replacement, with eight patients of mean age 61 years (22 to 89) who underwent bilateral total shoulder replacement in two stages, at a mean interval of 18 months (6 to 43). The overall function, pain and strength improved significantly in both groups. The subjective shoulder value, relative Constant score, active external rotation and the strength were improved significantly more in the single-stage group. Active elevation, abduction and overall function improved, significantly more in the single-stage group. Both the total duration of hospitalisation and the time off work per shoulder were substantially shorter in the single-stage group. The overall rate of complication was lower in the single-stage group. Our findings indicated that single-stage bilateral total shoulder replacement yielded significantly better clinical results with shorter hospitalisation and rehabilitation than staged replacement, and was not associated with any increase in complications.
The combination of an irreparable tear of the rotator cuff and destructive arthritis of the shoulder joint may cause severe pain, disability and loss of independence in the aged. Standard anatomical shoulder replacements depend on a functioning rotator cuff, and hence may fail in the presence of tears in the cuff. Many designs of non-anatomical constrained or semi-constrained prostheses have been developed for cuff tear arthropathy, but have proved unsatisfactory and were abandoned. The DePuy Delta III reverse prosthesis, designed by Grammont, medialises and stabilises the centre of rotation of the shoulder joint and has shown early promise. This study evaluated the mid-term clinical and radiological results of this arthroplasty in a consecutive series of 50 shoulders in 43 patients with a painful pseudoparalysis due to an irreparable cuff tear and destructive arthritis, performed over a period of seven years by a single surgeon. A follow-up of 98% was achieved, with a mean duration of 39 months (8 to 81). The mean age of the patients at the time of surgery was 81 years (59 to 95). The female to male ratio was 5:1. During the seven years, six patients died of natural causes. The clinical outcome was assessed using the American Shoulder and Elbow score, the Oxford Shoulder Score and the Short-form 36 score. A radiological review was performed using the Sirveaux score for scapular notching. The mean American Shoulder and Elbow score was 19 (95% confidence interval (CI) 14 to 23) pre-operatively, and 65 (95% CI 48 to 82) (paired The mean maximum elevation improved from 55° pre-operatively to 105° at final follow-up. There were seven complications during the whole series, although only four patients required further surgery.
There have been few reports in the literature of total elbow arthroplasty extending beyond 10 to 15 years. We reviewed 40 patients (41 elbows) with a mean age of 56 years (19 to 83) who had undergone a Coonrad/Coonrad-Morrey elbow arthroplasty by one surgeon for various diagnoses between 1974 and 1994. Surgical selection excluded patients with previous elbow infection or who refused to accept a sedentary level of elbow activity postoperatively. Objective data were collected from charts, radiographs, clinical photographs and supplemented by the referring orthopaedic surgeons’ records and radiographs if health or distance prevented a patient from returning for final review. Subjective outcome was defined by patient satisfaction. Of the 41 elbows, 21 were functional between 10 and 14 years after operation, ten between 15 and 19 years and ten between 20 and 31 years. There were 14 complications and 13 revisions, but no cases of acute infection, or permanent removal of any implant.
We studied retrospectively the results of revision arthroplasty of the elbow using a linked Coonrad-Morrey implant in 23 patients (24 elbows) after a mean follow-up period of 55 months. According to the Mayo Elbow Performance Score, 19 elbows were satisfactory, nine were excellent and ten good. The median total score had improved from 35 points (20 to 75) before the primary arthroplasty to 85 points (40 to 100) at the latest follow-up. There was a marked relief of pain, but the range of movement showed no overall improvement. Two patients had a second revision because of infection and two for aseptic loosening. The estimated five-year survival rate of the prosthesis was 83.1% (95% confidence interval 61.1 to 93.3). Revision elbow arthroplasty using the Coonrad-Morrey implant provided satisfactory results but with complications occurring in 13 cases.
We report the long-term clinical and radiological outcomes of the Aequalis total shoulder replacement with a cemented all-polyethylene flat-back keeled glenoid component implanted for primary osteoarthritis between 1991 and 2003 in nine European centres. A total of 226 shoulders in 210 patients were retrospectively reviewed at a mean of 122.7 months (61 to 219) or at revision. Clinical outcome was assessed using the Constant score, patient satisfaction score and range of movement. Kaplan-Meier survivorship analysis was performed with glenoid revision for loosening and radiological glenoid loosening ( Younger patient age and the curettage technique for glenoid preparation correlated with loosening. The rate of glenoid revision and radiological loosening increased with duration of follow-up, but not until a follow-up of five years. Therefore, we recommend that future studies reporting radiological outcomes of new glenoid designs should report follow-up of at least five to ten years.
We have previously described the short-term outcome of the use of reverse shoulder arthroplasty in the treatment of acute complex proximal humeral fractures in the elderly. We now report the clinical and radiological outcome of 36 fractures at a mean of 6.6 years (1 to 16). Previously, at a mean follow-up of 6 years (1 to 12) the mean Constant score was 58.5; this was reduced to 53 points with the further follow-up. A total of 23 patients (63%) had radiological evidence of loosening of the glenoid component. Nevertheless, only one patient had aseptic loosening of the baseplate at 12 years’ follow-up. The reduction in the mean Constant score with longer follow-up and the further development of scapular notching is worrying. New developments in design, bearing surfaces and surgical technique, and further follow-up, will determine whether reverse shoulder arthroplasty has a place in the management of complex proximal humeral fractures in the elderly.
Between 1996 and 2008, nine patients with severe post-traumatic arthritis underwent revision of a failed interposition arthroplasty of the elbow with a further interposition procedure using an allograft of tendo Achillis at a mean of 5.6 years (0.7 to 13.1) after the initial procedure. There were eight men and one woman with a mean age of 47 years (36 to 56). The mean follow-up was 4.7 years (2 to 8). The mean Mayo Elbow Performance score improved from 49 (15 to 65) pre-operatively to 73 (55 to 95) (p = 0.04). The mean Disability of the Arm, Shoulder and Hand score was 26 (7 to 42). One patient was unavailable for clinical follow-up and one underwent total elbow replacement three months post-operatively. Of the remaining patients, one had an excellent, two had good, three fair and one a poor result. Subjectively, five of the nine patients were satisfied. Four continued manual labour. Revision interposition arthroplasty is an option for young, active patients with severe post-traumatic arthritis who require both mobility and durability of the elbow.
Between 1982 and 1997, 403 consecutive patients (522 elbows) with rheumatoid arthritis underwent Souter-Strathclyde total elbow replacement. By the end of 2007, there had been 66 revisions for aseptic loosening in 60 patients. The mean time of follow-up was 10.6 years (0 to 25) The survival rates at five-, ten, 15 and 19 years were 96% (95%, confidence interval (CI) 95 to 98), 89% (95% CI 86 to 92), 83% (95% CI 78 to 87), and 77% (95% CI 69 to 85), respectively. The small and medium-sized short-stemmed primary humeral components had a 5.6-fold and 3.6-fold risk of revision for aseptic loosening respectively, compared to the medium-sized long-stemmed component. The small and medium-sized all-polyethylene ulnar components had respectively a 28.2-fold and 8.4-fold risk of revision for aseptic loosening, compared to the metal-backed ulnar components. The use of retentive ulnar components was not associated with an increased risk of aseptic loosening compared to non-retentive implants.
Displaced fractures of the lateral end of the clavicle in young patients have a high incidence of nonunion and a poor functional outcome after conservative management. Operative treatment is therefore usually recommended. However, current techniques may be associated with complications which require removal of the fixation device. We have evaluated the functional and radiological outcomes using a novel technique of open reduction and internal fixation. A series of 16 patients under 60 years of age with displaced fractures of the lateral end were treated by open reduction and fixation using a twin coracoclavicular endobutton technique. They were followed up for the first year after their injury. At one year the mean Constant score was 87.1 and the median Disabilities of the Arm, Shoulder and Hand score was 3.3. All fractures had united, except in one patient who developed an asymptomatic fibrous union. One patient had post-traumatic stiffness of the shoulder, which resolved with physiotherapy. None required re-operation. This technique produces good functional and radiological outcomes with a low prevalence of complications and routine implant removal is not necessary.
There is little information available at present regarding the mechanisms of failure of modern metallic radial head implants. Between 1998 and 2008, 44 consecutive patients (47 elbows) underwent removal of a failed metallic radial head replacement. In 13 patients (13 elbows) the initial operation had been undertaken within one week of a fracture of the radial head, at one to six weeks in seven patients (seven elbows) and more than six weeks (mean of 2.5 years (2 to 65 months)) in 22 patients (25 elbows). In the remaining two elbows the replacement was inserted for non-traumatic reasons. The most common indication for further surgery was painful loosening (31 elbows). Revision was undertaken for stiffness in 18 elbows, instability in nine, and deep infection in two. There were signs of over-lengthening of the radius in 11 elbows. Degenerative changes were found in all but one. Only three loose implants had been fixed with cement. Instability was not identified in any of the bipolar implants.
We describe the longer term clinical and radiological findings in a prospectively followed series of 49 rheumatoid patients (58 shoulders) who had undergone Neer II total shoulder replacement. The early and intermediate results have been published previously. At a mean follow-up of 19.8 years (16.5 to 23.8) 14 shoulders survived. Proximal migration of the humeral component was associated with progressive loosening of the glenoid and humeral components, but was independent of the state of the rotator cuff at the time of operation. Despite these changes the range of movement was preserved. Most patients had little or no pain in the shoulder, could sleep undisturbed and could attend to personal hygiene and grooming.
In a prospective study between 2000 and 2005, 22 patients with primary osteoarthritis of the shoulder had a total shoulder arthroplasty with a standard five-pegged glenoid component, 12 with non-offset humeral head and ten with offset humeral head components. Over a period of 24 months the relative movement of the glenoid component with respect to the scapula was measured using radiostereometric analysis. Nine glenoids needed reaming for erosion. There was a significant increase in rotation about all three axes with time (p <
0.001), the largest occurring about the longitudinal axis (anteversion-retroversion), with mean values of 3.8° and 1.9° for the non-offset and offset humeral head eroded subgroups, respectively. There was also a significant difference in rotation about the anteversion-retroversion axis (p = 0.01) and the varus-valgus (p <
0.001) z-axis between the two groups. The offset humeral head group reached a plateau at early follow-up with rotation about the z-axis, whereas the mean of the non-offset humeral head group at 24 months was three times greater than that of the offset group accounting for the highly significant difference between them.
Mechanical loosening which begins with early-onset migration of the prosthesis is the major reason for failure of the Souter-Strathclyde elbow replacement. In a prospective study of 18 Souter-Strathclyde replacements we evaluated the patterns of migration using roentgen stereophotogrammetric analysis. We had previously reported the short-term results after a follow-up of two years which we have now extended to a mean follow-up of 8.2 years (1 to 11.3). Migration was assessed along the co-ordinal axes and overall micromovement was expressed as the maximum total point movement. The alignment of the prosthesis and the presence of radiolucent lines were examined on conventional standardised radiographs. All the humeral components showed increased and variable patterns of migration at the extended follow-up and four humeral components were revised. The maximum total point movement at two years in the revised prostheses was 1.8 mm (
Our aim in this prospective study was to evaluate the outcome of total shoulder replacement in the treatment of young and middle-aged active patients with primary glenohumeral osteoarthritis. We reviewed 21 patients (21 shoulders) with a mean age of 55 years (37 to 60). The mean follow-up was seven years (5 to 9). The same anatomical, third-generation, cemented implant had been used in all patients. All the patients were evaluated radiologically and clinically using the Constant and Murley score. No patients required revision. In one a tear of the supraspinatus tendon occurred. Overall, 20 patients (95%) were either very satisfied (n = 18) or satisfied (n = 2) with the outcome. Significant differences (p <
0.0001) were found for all categories of the Constant and Murley score pre- and post-operatively. The mean Constant and Murley score increased from 24.1 points (10 to 45) to 64.5 points (39 to 93), and the relative score from 30.4% (11% to 50%) to 83% (54% to 116%). No clinical or radiological signs of loosening of the implant were seen. For young and middle-aged patients with osteoarthritis, third-generation total shoulder replacement is a viable method of treatment with a low rate of complications and excellent results in the mid-term.
We have examined the relief from pain and the functional outcome in 18 patients who underwent resection arthroplasty of the shoulder as a salvage operation between 1988 and 2002. The indications included failed shoulder replacement in 17, with infection in 13, and chronic septic arthritis in one. The mean follow-up was 8.3 years (2.5 to 16.6). Two intra-operative fractures of the humerus occurred, both of which healed. The level of pain was significantly decreased ( Despite applying this procedure principally to failed shoulder replacements, the results were similar to those reported in the literature for patients after severe fracture-dislocation. Reduction of pain is possible in one half to two-thirds of patients. The outcome of this operation in providing relief from pain cannot be guaranteed, but the shoulder is usually comfortable at rest, albeit with profound functional limitations.
Excision is not a suitable treatment for all comminuted fractures of the radial head. In elbows where instability can be predicted, a replacement arthroplasty of the radial head is more effective. The aim of this paper was to present the medium-term results of the Judet floating radial head prosthesis. This operation was performed on 14 patients between 1992 and 2003, of whom 12 were reviewed at a mean follow-up of five years and three months (1 to 12 years). The outcome was assessed using the Mayo elbow performance score and a modified Disability of Arm Shoulder Hand (DASH) questionnaire. There were six excellent results, four good, one fair and one poor, as graded by the Mayo score. The mean DASH score was 23.9/100 (0 to 65.8/100). The only significant complication occurred in one patient who developed a severe complex regional pain syndrome. There were no patients with secondary instability of the elbow, implant loosening, cubitus valgus, osteoporosis of the capitellum, or pain in the forearm and wrist. Our experience, combined with that of other authors using this device, has encouraged us to continue using the Judet prosthesis in comminuted fractures of the elbow where instability is a potential problem.
We report the use of the reverse shoulder prosthesis in the revision of a failed shoulder hemiarthroplasty in 19 shoulders in 18 patients (7 men, 11 women) with severe pain and loss of function. The primary procedure had been undertaken for glenohumeral arthritis associated with severe rotator cuff deficiency. Statistically significant improvements were seen in pain and functional outcome. After a mean follow-up of 44 months (24 to 89), mean forward flexion improved by 26.4° and mean abduction improved by 35°. There were six prosthesis-related complications in six shoulders (32%), five of which had severe bone loss of the glenoid, proximal humerus or both. Three shoulders (16%) had non-prosthesis related complications. The use of the reverse shoulder prosthesis provides improvement in pain and function for patients with failure of a hemiarthroplasty for glenohumeral arthritis and rotator cuff deficiency. However, high rates of complications were associated with glenoid and proximal humeral bone loss.
We divided 309 patients with an inflammatory arthritis who had undergone primary elbow replacement using the Souter-Strathclyde implant into two groups according to their age. The mean follow-up in the older group (mean age 64 years) was 7.3 years while in the younger patients (mean age 42 years) it was 12 years. Survivorship for three different failure end-points (revision, revision because of aseptic loosening of the humeral component, and gross loosening of the humeral implant), was compared in both groups. Our findings showed that there was no significant difference in the incidence of loosening when young rheumatoid patients were compared with an older age group.
There are few reports in the literature of the diagnosis and treatment of the infected shoulder arthroplasty. Most deal with resection arthroplasty and two-stage exchange surgery. We present our results of one-stage exchange operation as treatment for the infected shoulder arthroplasty. Our group comprised 16 patients (ten men, six women) with 16 infected arthroplasties. By the time of follow-up, two patients had died (mean 5.8 years), two could not be located and three had already undergone revision surgery. Nine patients were thus available for clinical examination and assessment. The infections were largely caused by staphylococci, Further revision was performed in three patients. One sustained a peri-prosthetic humeral fracture, another developed an acromial pseudarthrosis after transacromial surgery and the third suffered recurrent dislocations. No patient had a recurrence of infection. A one-stage exchange procedure using antibiotic-loaded bone cement eradicated infection in all our patients and we suggest that such a procedure is at least as successful as either a resection arthroplasty or a two-stage exchange in the management of the infected shoulder arthroplasty.