Aims. Endoprosthetic reconstruction with a distal femoral arthroplasty (DFA) can be used to treat distal femoral bone loss from oncological and non-oncological causes. This study reports the short-term implant survivorship, complications, and risk factors for patients who underwent DFA for non-neoplastic indications. Methods. We performed a retrospective review of 75 patients from a single institution who underwent DFA for non-neoplastic indications, including aseptic loosening or mechanical failure of a previous prosthesis (n = 25), periprosthetic joint infection (PJI) (n = 23), and native or periprosthetic distal femur fracture or nonunion (n = 27). Patients with less than 24 months’ follow-up were excluded. We collected patient demographic data, complications, and reoperations. Reoperation for
Aims. Patellofemoral arthroplasty (PFA) has experienced significant
improvements in implant survivorship with second generation designs.
This has renewed interest in PFA as an alternative to total knee
arthroplasty (TKA) for younger active patients with isolated patellofemoral
osteoarthritis (PF OA). We analysed the cost-effectiveness of PFA versus TKA
for the management of isolated PF OA in the United States-based
population. Patients and Methods. We used a Markov transition state model to compare cost-effectiveness
between PFA and TKA. Simulated patients were aged 60 (base case)
and 50 years. Lifetime costs (2015 United States dollars), quality-adjusted
life year (QALY) gains and incremental cost-effectiveness ratio
(ICER) were calculated from a healthcare payer perspective. Annual rates
of revision were derived from the National Joint Registry for England,
Wales, Northern Ireland and the Isle of Man. Deterministic and probabilistic
sensitivity analysis was performed for all parameters against a
$50 000/QALY willingness to pay. . Results. PFA was more expensive ($49 811 versus $46 632)
but more effective (14.3 QALYs versus 13.3 QALYs)
over a lifetime horizon. The ICER associated with the additional
effectiveness of PFA was $3097. The model was mainly sensitive to utility
values, with PFA remaining cost-effective when its utility exceeded
that of TKA by at least 1.0%. PFA provided incremental benefits
at no increased cost when annual rates of revision decreased by
24.5%. . Conclusions. Recent improvements in rates of implant of survival have made
PFA an economically beneficial joint-preserving procedure in younger
patients, delaying TKA until
Aims. Antibiotic-loaded bone cements (ALBCs) may offer early protection against the formation of bacterial biofilm after joint arthroplasty. Use in hip arthroplasty is widely accepted, but there is a lack of evidence in total knee arthroplasty (TKA). The objective of this study was to evaluate the use of ALBC in a large population of TKA patients. Materials and Methods. Data from the National Joint Registry (NJR) of England and Wales were obtained for all primary cemented TKAs between March 2003 and July 2016. Patient, implant, and surgical variables were analyzed. Cox proportional hazards models were used to assess the influence of ALBC on risk of revision. Body mass index (BMI) data were available in a subset of patients. Results. Of 731 214 TKAs, 15 295 (2.1%) were implanted with plain cement and 715 919 (97.9%) with ALBC. There were 13 391 revisions; 2391 were performed for infection. After adjusting for other variables, ALBC had a significantly lower risk of revision for any cause (hazard ratio (HR) 0.85, 95% confidence interval (CI) 0.77 to 0.93; p < 0.001). ALBC was associated with a lower risk of revision for all aseptic causes (HR 0.85, 95% CI 0.77 to 0.95; p < 0.001) and revisions for infection (HR 0.84, 95% CI 0.67 to 1.01; p = 0.06). The results were similar when BMI was added into the model, and in a subanalysis where surgeons using only ALBC over the entire study period were excluded. Prosthesis survival at ten years for TKAs implanted with ALBC was 96.3% (95% CI 96.3 to 96.4) compared with 95.5% (95% CI 95.0 to 95.9) in those implanted with plain cement. On a population level, where 100 000 TKAs are performed annually, this difference represents 870 fewer revisions at ten years in the ALBC group. Conclusion. After adjusting for a range of variables, ALBC was associated with a significantly lower risk of revision in this registry-based study of an entire nation of primary cemented knee arthroplasties. Using ALBC does not appear to increase midterm
The mean age of patients undergoing total knee arthroplasty (TKA) has reduced with time. Younger patients have increased expectations following TKA. Aseptic loosening of the tibial component is the most common cause of failure of TKA in the UK. Interest in cementless TKA has re-emerged due to its encouraging results in the younger patient population. We review a large series of tantalum trabecular metal cementless implants in patients who are at the highest risk of revision surgery. A total of 454 consecutive patients who underwent cementless TKA between August 2004 and December 2021 were reviewed. The mean follow-up was ten years. Plain radiographs were analyzed for radiolucent lines. Patients who underwent revision TKA were recorded, and the cause for revision was determined. Data from the National Joint Registry for England, Wales, Northern Island, the Isle of Man and the States of Guernsey (NJR) were compared with our series.Aims
Methods
Aims. Primary (or spontaneous) and secondary osteonecrosis of the knee
can lead to severe joint degeneration, for which either total or
unicompartmental arthroplasty may be considered. However, there
are limited studies analyzing outcomes of unicompartmental knee arthroplasties
(UKAs) for osteonecrosis involving an isolated compartment of the
knee. The aims of this study were to analyze outcomes of UKAs for
osteonecrosis with specific focus on 1) survivorship free of any
revision or reoperation, 2) risk factors for failure, 3) clinical outcomes,
and 4) complications. Patients and Methods. A total of 45 patients underwent 46 UKAs for knee osteonecrosis
between 2002 and 2014 at our institution (The Mayo Clinic, Rochester,
Minnesota). Twenty patients (44%) were female; the mean age of the
patients was 66 years, and mean body mass index (BMI) was 31 kg/m. 2. Of
the 46 UKAs, 44 (96%) were medial UKAs, and 35 (76%) were fixed-bearing
design. Mean mechanical axis postoperatively was 1.5° varus (0°
to 5° varus); 41 UKAs (89%) were performed for primary osteonecrosis.
Mean follow-up was five years (2 to 12). Results. Survivorship free of any revision in the cohort was 89% (95%
CI 77 to 99) and 76% (95% CI 53 to 99) at five and ten years, respectively.
In patients undergoing UKA for primary osteonecrosis survivorship
free of any revision was 93% (95% CI 83 to 100)at both five and ten
years. Secondary osteonecrosis was a significant risk factor for
poorer survivorship free of any revision or reoperation (hazard
ratio 7.7, p = 0.03). Three medial UKAs (6.5%) were converted to
total knee arthroplasties (TKAs): two for lateral compartment degeneration and
one for development of lateral osteonecrosis. No implants were revised
for loosening, fracture, or wear. Knee Society scores improved from
a mean of 60 (44 to 72) preoperatively to a mean of 94 postoperatively
(82 to 100) (p < 0.001). There were no surgical complications. Conclusion. When done for primary osteonecrosis of the knee, UKA resulted
in reliable clinical improvement, minimal complications, and durable
estimated implant survivorship free of revision at ten years. UKA
done for secondary osteonecrosis was substantially less durable
at mid-term follow-up. Progression of knee degeneration, rather
than
Revision total knee arthroplasty (rTKA) is a technically challenging and costly procedure. It is well-documented that primary TKA (pTKA) have better survivorship than rTKA; however, we were unable to identify any studies explicitly investigating previous rTKA as a risk factor for failure following rTKA. The purpose of this study is to compare the outcomes following rTKA between patients undergoing index rTKA and those who had been previously revised. This retrospective, observational study reviewed patients who underwent unilateral, aseptic rTKA at an academic orthopaedic speciality hospital between June 2011 and April 2020 with > one-year of follow-up. Patients were dichotomized based on whether this was their first revision procedure or not. Patient demographics, surgical factors, postoperative outcomes, and re-revision rates were compared between the groups.Aims
Methods
Unicompartmental knee arthroplasty (UKA) has higher revision rates than total knee arthroplasty (TKA). As revision of UKA may be less technically demanding than revision TKA, UKA patients with poor functional outcomes may be more likely to be offered revision than TKA patients with similar outcomes. The aim of this study was to compare clinical thresholds for revisions between TKA and UKA using revision incidence and patient-reported outcomes, in a large, matched cohort at early, mid-, and late-term follow-up. Analyses were performed on propensity score-matched patient cohorts of TKAs and UKAs (2:1) registered in the New Zealand Joint Registry between 1 January 1999 and 31 December 2019 with an Oxford Knee Score (OKS) response at six months (n, TKA: 16,774; UKA: 8,387), five years (TKA: 6,718; UKA: 3,359), or ten years (TKA: 3,486; UKA: 1,743). Associations between OKS and revision within two years following the score were examined. Thresholds were compared using receiver operating characteristic analysis. Reasons for aseptic revision were compared using cumulative incidence with competing risk.Aims
Methods
The transepicondylar axis is a well-established reference for the determination of femoral component rotation in total knee arthroplasty (TKA). However, when severe bone loss is present in the femoral condyles, rotational alignment can be more complicated. There is a lack of validated landmarks in the supracondylar region of the distal femur. Therefore, the aim of this study was to analyze the correlation between the surgical transepicondylar axis (sTEA) and the suggested dorsal cortex line (DCL) in the coronal plane and the inter- and intraobserver reliability of its CT scan measurement. A total of 75 randomly selected CT scans were measured by three experienced surgeons independently. The DCL was defined in the coronal plane as a tangent to the dorsal femoral cortex located 75 mm above the joint line in the frontal plane. The difference between sTEA and DCL was calculated. Descriptive statistics and angulation correlations were generated for the sTEA and DCL, as well as for the distribution of measurement error for intra- and inter-rater reliability.Aims
Methods
The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design. This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.Aims
Methods
We explored the literature surrounding whether
allergy and hypersensitivity has a clinical basis for implant selection
in total knee arthroplasty (TKA). In error, the terms hypersensitivity
and allergy are often used synonymously. Although a relationship
is present, we could not find any evidence of
Obtaining solid implant fixation is crucial in revision total knee arthroplasty (rTKA) to avoid aseptic loosening, a major reason for re-revision. This study aims to validate a novel grading system that quantifies implant fixation across three anatomical zones (epiphysis, metaphysis, diaphysis). Based on pre-, intra-, and postoperative assessments, the novel grading system allocates a quantitative score (0, 0.5, or 1 point) for the quality of fixation achieved in each anatomical zone. The criteria used by the algorithm to assign the score include the bone quality, the size of the bone defect, and the type of fixation used. A consecutive cohort of 245 patients undergoing rTKA from 2012 to 2018 were evaluated using the current novel scoring system and followed prospectively. In addition, 100 first-time revision cases were assessed radiologically from the original cohort and graded by three observers to evaluate the intra- and inter-rater reliability of the novel radiological grading system.Aims
Methods
The aim of this study was to determine whether obesity had a detrimental effect on the long-term performance and survival of medial unicompartmental knee arthroplasties (UKAs). This study reviewed prospectively collected functional outcome scores and revision rates of all medial UKA patients with recorded BMI performed in Christchurch, New Zealand, from January 2011 to September 2021. Patient-reported outcome measures (PROMs) were the primary outcome of this study, with all-cause revision rate analyzed as a secondary outcome. PROMs were taken preoperatively, at six months, one year, five years, and ten years postoperatively. There were 873 patients who had functional scores recorded at five years and 164 patients had scores recorded at ten years. Further sub-group analysis was performed based on the patient’s BMI. Revision data were available through the New Zealand Joint Registry for 2,323 UKAs performed during this time period.Aims
Methods
This prospective study reports longitudinal, within-patient, patient-reported outcome measures (PROMs) over a 15-year period following cemented single radius total knee arthroplasty (TKA). Secondary aims included reporting PROMs trajectory, 15-year implant survival, and patient attrition from follow-up. From 2006 to 2007, 462 consecutive cemented cruciate-retaining Triathlon TKAs were implanted in 426 patients (mean age 69 years (21 to 89); 290 (62.7%) female). PROMs (12-item Short Form Survey (SF-12), Oxford Knee Score (OKS), and satisfaction) were assessed preoperatively and at one, five, ten, and 15 years. Kaplan-Meier survival and univariate analysis were performed.Aims
Methods
This multicentre retrospective observational study’s aims were to investigate whether there are differences in the occurrence of radiolucent lines (RLLs) following total knee arthroplasty (TKA) between the conventional Attune baseplate and its successor, the novel Attune S+, independent from other potentially influencing factors; and whether tibial baseplate design and presence of RLLs are associated with differing risk of revision. A total of 780 patients (39% male; median age 70.7 years (IQR 62.0 to 77.2)) underwent cemented TKA using the Attune Knee System) at five centres, and with the latest radiograph available for the evaluation of RLL at between six and 36 months from surgery. Univariate and multivariate logistic regression models were performed to assess associations between patient and implant-associated factors on the presence of tibial and femoral RLLs. Differences in revision risk depending on RLLs and tibial baseplate design were investigated with the log-rank test.Aims
Methods
Total knee arthroplasty (TKA) with a highly congruent condylar-stabilized (CS) articulation may be advantageous due to increased stability versus cruciate-retaining (CR) designs, while mitigating the limitations of a posterior-stabilized construct. The aim was to assess ten-year implant survival and functional outcomes of a cemented single-radius TKA with a CS insert, performed without posterior cruciate ligament sacrifice. This retrospective cohort study included consecutive patients undergoing TKA at a specialist centre in the UK between November 2010 and December 2012. Data were collected using a bespoke electronic database and cross-referenced with national arthroplasty audit data, with variables including: preoperative characteristics, intraoperative factors, complications, and mortality status. Patient-reported outcome measures (PROMs) were collected by a specialist research team at ten years post-surgery. There were 536 TKAs, of which 308/536 (57.5%) were in female patients. The mean age was 69.0 years (95% CI 45.0 to 88.0), the mean BMI was 32.2 kg/m2 (95% CI 18.9 to 50.2), and 387/536 (72.2%) survived to ten years. There were four revisions (0.7%): two deep infections (requiring debridement and implant retention), one aseptic loosening, and one haemosiderosis.Aims
Methods
The aim of this study was to assess factors associated with the estimated lifetime risk of revision surgery after primary knee arthroplasty (KA). All patients from the Scottish Arthroplasty Project dataset undergoing primary KA during the period 1 January 1998 to 31 December 2019 were included. The cumulative incidence function for revision and death was calculated up to 20 years. Adjusted analyses used cause-specific Cox regression modelling to determine the influence of patient factors. The lifetime risk was calculated as a percentage for patients aged between 45 and 99 years using multiple-decrement life table methodology.Aims
Methods
Periprosthetic joint infection (PJI) occurs in approximately 1% to 2% of total knee arthroplasties (TKA) presenting multiple challenges, such as difficulty in diagnosis, technical complexity, and financial costs. Two-stage exchange is the gold standard for treating PJI but emerging evidence suggests 'two-in-one' single-stage revision as an alternative, delivering comparable outcomes, reduced morbidity, and cost-effectiveness. This study investigates five-year results of modified single-stage revision for treatment of PJI following TKA with bone loss. Patients were identified from prospective data on all TKA patients with PJI following the primary procedure. Inclusion criteria were: revision for PJI with bone loss requiring reconstruction, and a minimum five years’ follow-up. Patients were followed up for recurrent infection and assessment of function. Tools used to assess function were Oxford Knee Score (OKS) and American Knee Society Score (AKSS).Aims
Methods
The kinematic alignment (KA) approach to total knee arthroplasty (TKA) has recently increased in popularity. Accordingly, a number of derivatives have arisen and have caused confusion. Clarification is therefore needed for a better understanding of KA-TKA. Calipered (or true, pure) KA is performed by cutting the bone parallel to the articular surface, compensating for cartilage wear. In soft-tissue respecting KA
Aims. The aim of this independent multicentre study was to assess the
mid-term results of mobile bearing unicondylar knee arthroplasty
(UKA) for isolated lateral osteoarthritis of the knee joint. Patients and Methods. We retrospectively evaluated 363 consecutive, lateral UKAs (346
patients) performed using the Oxford domed lateral prosthesis undertaken
in three high-volume knee arthroplasty centres between 2006 and
2014. Mean age of the patients at surgery was 65 years (36 to 88)
with a mean final follow-up of 37 months (12 to 93). Results. A total of 36 (10.5%) patients underwent revision surgery, giving
a survival rate of 90.1% at three years (95% confidence intervals
(CI) 86.1 to 93.1; number at risk: 155) and 85.0% at five years
(95% CI 77.9 to 89.9; number at risk: 43). Dislocation of the mobile
bearing occurred in 18 patients (5.6%) at three years (95% CI 1.0
to 16.4; number at risk: 154) and in 20 patients (8.5%) at five
years (95% CI 1.0 to 27.0; number at risk: 42). There were no significant differences
in the dislocation rate between the participating centres or the
surgeons. We were not able to identify an effect of each surgeon’s
learning curve on the dislocation rate of the mobile bearing. The
clinical outcome in patients without revision surgery at final follow-up
was good to excellent, with a mean Oxford knee score of 40.3 (95%
CI 39.4 to 41.2), a mean Tegner activity score of 3.2 (95% CI 3.1
to 3.3) and a mean University of California, Los Angeles score of
5.7 (95% CI 5.5 to 5.9). Conclusion. Our data, which consists of a high number of patients treated
with mobile bearing UKA in the lateral compartment, indicates a
high revision rate of 15% at five years with dislocation of the
mobile bearing being the main reason for
This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant. A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life.Aims
Methods
Unicompartmental and total knee arthroplasty (UKA and TKA) are successful treatments for osteoarthritis, but the solid metal implants disrupt the natural distribution of stress and strain which can lead to bone loss over time. This generates problems if the implant needs to be revised. This study investigates whether titanium lattice UKA and TKA implants can maintain natural load transfer in the proximal tibia. In a cadaveric model, UKA and TKA procedures were performed on eight fresh-frozen knee specimens, using conventional (solid) and titanium lattice tibial implants. Stress at the bone-implant interfaces were measured and compared to the native knee.Aims
Methods
Aims. The Advance Medial-Pivot total knee arthroplasty (TKA) was designed
to reflect contemporary data regarding the kinematics of the knee.
We wished to examine the long-term results obtained with this prosthesis
by extending a previous evaluation. Patients and Methods. We retrospectively evaluated prospectively collected data from
225 consecutive patients (41 men and 184 women; mean age at surgery
71 years, 52 to 84) who underwent 284 TKAs with a mean follow-up
of 13.4 years (11 to 15).
It has previously been shown that higher-volume hospitals have better outcomes following revision total knee arthroplasty (rTKA). We were unable to identify any studies which investigated the effect of surgeon volume on the outcome of rTKA. We sought to investigate whether patients of high-volume (HV) rTKA surgeons have better outcomes following this procedure compared with those of low-volume (LV) surgeons. This retrospective study involved patients who underwent aseptic unilateral rTKA between January 2016 and March 2019, using the database of a large urban academic medical centre. Surgeons who performed ≥ 19 aseptic rTKAs per year during the study period were considered HV and those who performed < 19 per year were considered LV. Demographic characteristics, surgical factors, and postoperative outcomes were compared between the two groups.Aims
Methods
Limb alignment in total knee arthroplasty (TKA) influences periarticular soft-tissue tension, biomechanics through knee flexion, and implant survival. Despite this, there is no uniform consensus on the optimal alignment technique for TKA. Neutral mechanical alignment facilitates knee flexion and symmetrical component wear but forces the limb into an unnatural position that alters native knee kinematics through the arc of knee flexion. Kinematic alignment aims to restore native limb alignment, but the safe ranges with this technique remain uncertain and the effects of this alignment technique on component survivorship remain unknown. Anatomical alignment aims to restore predisease limb alignment and knee geometry, but existing studies using this technique are based on cadaveric specimens or clinical trials with limited follow-up times. Functional alignment aims to restore the native plane and obliquity of the joint by manipulating implant positioning while limiting soft tissue releases, but the results of high-quality studies with long-term outcomes are still awaited. The drawbacks of existing studies on alignment include the use of surgical techniques with limited accuracy and reproducibility of achieving the planned alignment, poor correlation of intraoperative data to long-term functional outcomes and implant survivorship, and a paucity of studies on the safe ranges of limb alignment. Further studies on alignment in TKA should use surgical adjuncts (e.g. robotic technology) to help execute the planned alignment with improved accuracy, include intraoperative assessments of knee biomechanics and periarticular soft-tissue tension, and correlate alignment to long-term functional outcomes and survivorship.
The purpose of this study was to compare the radiological outcomes of manual versus robotic-assisted medial unicompartmental knee arthroplasty (UKA). Postoperative radiological outcomes from 86 consecutive robotic-assisted UKAs (RAUKA group) from a single academic centre were retrospectively reviewed and compared to 253 manual UKAs (MUKA group) drawn from a prior study at our institution. Femoral coronal and sagittal angles (FCA, FSA), tibial coronal and sagittal angles (TCA, TSA), and implant overhang were radiologically measured to identify outliers.Aims
Methods
A limited number of investigations with conflicting results have described perivascular lymphocytic infiltration (PVLI) in the setting of total knee arthroplasty (TKA). The purpose of this study was to determine if PVLI found in TKAs at the time of aseptic revision surgery was associated with worse clinical outcomes and survivorship. A retrospective review was conducted on 617 patients who underwent aseptic TKA revision who had histological analysis for PVLI at the time of surgery. Clinical and radiological data were obtained pre- and postoperatively, six weeks postoperatively, and then every year thereafter.Aims
Methods
This study aimed to identify the tibial component and femoral component coronal angles (TCCAs and FCCAs), which concomitantly are associated with the best outcomes and survivorship in a cohort of fixed-bearing, cemented, medial unicompartmental knee arthroplasties (UKAs). We also investigated the potential two-way interactions between the TCCA and FCCA. Prospectively collected registry data involving 264 UKAs from a single institution were analyzed. The TCCAs and FCCAs were measured on postoperative radiographs and absolute angles were analyzed. Clinical assessment at six months, two years, and ten years was undertaken using the Knee Society Knee score (KSKS) and Knee Society Function score (KSFS), the Oxford Knee Score (OKS), the 36-Item Short-Form Health Survey questionnaire (SF-36), and range of motion (ROM). Fulfilment of expectations and satisfaction was also recorded. Implant survivorship was reviewed at a mean follow-up of 14 years (12 to 16). Multivariate regression models included covariates, TCCA, FCCA, and two-way interactions between them. Partial residual graphs were generated to identify angles associated with the best outcomes. Kaplan-Meier analysis was used to compare implant survivorship between groups.Aims
Methods
Recent total knee arthroplasty (TKA) designs have featured more anatomical morphologies and shorter tibial keels. However, several reports have raised concerns about the impact of these modifications on implant longevity. The aim of this study was to report the early performance of a modern, cemented TKA design. All patients who received a primary, cemented TKA between 2012 and 2017 with a minimum two-year follow-up were included. The implant investigated features an asymmetrical tibial baseplate and shortened keel. Patient demographic details, Knee Society Scores (KSS), component alignment, and the presence of radiolucent lines at final follow-up were recorded. Kaplan-Meier analyses were performed to estimate survivorship.Aims
Methods
The aim of this study was to compare ten-year longitudinal healthcare costs and revision rates for patients undergoing unicompartmental knee arthroplasty (UKA) and total knee arthroplasty (TKA). The Humana database was used to compare 2,383 patients undergoing UKA between 2007 and 2009, who were matched 1:1 from a cohort of 63,036 patients undergoing primary TKA based on age, sex, and Elixhauser Comorbidity Index. Medical and surgical complications were tracked longitudinally for one year following surgery. Rates of revision surgery and cumulative mean healthcare costs were recorded for this period of time and compared between the cohorts.Aims
Methods
Total knee arthroplasty (TKA) using functional alignment aims to implant the components with minimal compromise of the soft-tissue envelope by restoring the plane and obliquity of the non-arthritic joint. The objective of this study was to determine the effect of TKA with functional alignment on mediolateral soft-tissue balance as assessed using intraoperative sensor-guided technology. This prospective study included 30 consecutive patients undergoing robotic-assisted TKA using the Stryker PS Triathlon implant with functional alignment. Intraoperative soft-tissue balance was assessed using sensor-guided technology after definitive component implantation; soft-tissue balance was defined as intercompartmental pressure difference (ICPD) of < 15 psi. Medial and lateral compartment pressures were recorded at 10°, 45°, and 90° of knee flexion. This study included 18 females (60%) and 12 males (40%) with a mean age of 65.2 years (SD 9.3). Mean preoperative hip-knee-ankle deformity was 6.3° varus (SD 2.7°).Aims
Methods
Enhanced perioperative protocols have significantly improved patient recovery following primary total knee arthroplasty (TKA). Little has been investigated the effectiveness of these protocols for revision TKA (RTKA). We report on a matched group of aseptic revision and primary TKA patients treated with an identical pain and rehabilitation programmes. Overall, 40 aseptic full-component RTKA patients were matched (surgical date, age, sex, and body mass index (BMI)) to a group of primary cemented TKA patients. All RTKAs had new uncemented stemmed femoral and tibial components with metaphyseal sleeves. Both groups were treated with an identical postoperative pain protocol. Patients were followed for at least two years. Knee Society Scores (KSS) at six weeks and at final follow-up were recorded for both groups.Aims
Methods
The aim of this study was to compare robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) in order to determine the changes in the anatomy of the knee and alignment of the lower limb following surgery. An analysis of 38 patients who underwent TKA and 32 who underwent bi-UKA was performed as a secondary study from a prospective, single-centre, randomized controlled trial. CT imaging was used to measure coronal, sagittal, and axial alignment of the knee preoperatively and at three months postoperatively to determine changes in anatomy that had occurred as a result of the surgery. The hip-knee-ankle angle (HKAA) was also measured to identify any differences between the two groups.Aims
Methods
Although bone cement is the primary mode of fixation in total knee arthroplasty (TKA), cementless fixation is gaining interest as it has the potential of achieving lasting biological fixation. By 3D printing an implant, highly porous structures can be manufactured, promoting osseointegration into the implant to prevent aseptic loosening. This study compares the migration of cementless, 3D-printed TKA to cemented TKA of a similar design up to two years of follow-up using radiostereometric analysis (RSA) known for its ability to predict aseptic loosening. A total of 72 patients were randomized to either cementless 3D-printed or a cemented cruciate retaining TKA. RSA and clinical scores were evaluated at baseline and postoperatively at three, 12, and 24 months. A mixed model was used to analyze the repeated measurements.Aims
Methods
Total knee arthroplasty is an established treatment for knee osteoarthritis with excellent long-term results, but there remains controversy about the role of uncemented prostheses. We present the long-term results of a randomized trial comparing an uncemented tantalum metal tibial component with a conventional cemented component of the same implant design. Patients under the age of 70 years with symptomatic osteoarthritis of the knee were randomized to receive either an uncemented tantalum metal tibial monoblock component or a standard cemented modular component. The mean age at time of recruitment to the study was 63 years (50 to 70), 46 (51.1%) knees were in male patients, and the mean body mass index was 30.4 kg/m2 (21 to 36). The same cruciate retaining total knee system was used in both groups. All patients received an uncemented femoral component and no patients had their patella resurfaced. Patient outcomes were assessed preoperatively and postoperatively using the modified Oxford Knee Score, Knee Society Score, and 12-Item Short-Form Health Survey questionnaire (SF-12) score. Radiographs were analyzed using the American Knee Society Radiograph Evaluation score. Operative complications, reoperations, or revision surgery were recorded. A total of 90 knees were randomized and at last review 77 knees were assessed. In all, 11 patients had died and two were lost to follow-up.Aims
Methods
Aseptic loosening of the tibial component is a frequent cause of failure in primary total knee arthroplasty (TKA). Management options include an isolated tibial revision or full component revision. A full component revision is frequently selected by surgeons unfamiliar with the existing implant or who simply wish to “start again”. This option adds morbidity compared with an isolated tibial revision. While isolated tibial revision has a lower morbidity, it is technically more challenging due to difficulties with exposure and maintaining prosthetic stability. This study was designed to compare these two reconstructive options. Patients undergoing revision TKA for isolated aseptic tibial loosening between 2012 and 2017 were identified. Those with revision implants or revised for infection, instability, osteolysis, or femoral component loosening were excluded. A total of 164 patients were included; 88 had an isolated tibial revision and 76 had revision of both components despite only having a loose tibial component. The demographics and clinical and radiological outcomes were recorded.Aims
Methods
Varus-valgus constrained (VVC) implants are often used during revision total knee arthroplasty (TKA) to gain coronal plane stability. However, the increased mechanical torque applied to the bone-cement interface theoretically increases the risk of aseptic loosening. We assessed mid-term survivorship, complications, and clinical outcomes of a fixed-bearing VVC device in revision TKAs. A total of 416 consecutive revision TKAs (398 patients) were performed at our institution using a single fixed-bearing VVC TKA from 2007 to 2015. Mean age was 64 years (33 to 88) with 50% male (199). Index revision TKA diagnoses were: instability (n = 122, 29%), aseptic loosening (n = 105, 25%), and prosthetic joint infection (PJI) (n = 97, 23%). All devices were cemented on the epiphyseal surfaces. Femoral stems were used in 97% (n = 402) of cases, tibial stems in 95% (n = 394) of cases; all were cemented. In total, 93% (n = 389) of cases required a stemmed femoral and tibial component. Femoral cones were used in 29%, and tibial cones in 40%. Survivorship was assessed via competing risk analysis; clinical outcomes were determined using Knee Society Scores (KSSs) and range of movement (ROM). Mean follow-up was four years (2 to 10).Aims
Methods
The objectives of this study were to assess the effect of anterior cruciate ligament (ACL) resection on flexion-extension gaps, mediolateral soft tissue laxity, maximum knee extension, and limb alignment during primary total knee arthroplasty (TKA). This prospective study included 140 patients with symptomatic knee osteoarthritis undergoing primary robotic-arm assisted TKA. All operative procedures were performed by a single surgeon using a standard medial parapatellar approach. Optical motion capture technology with fixed femoral and tibial registration pins was used to assess study outcomes pre- and post-ACL resection with knee extension and 90° knee flexion. This study included 76 males (54.3%) and 64 females (45.7%) with a mean age of 64.1 years (SD 6.8) at time of surgery. Mean preoperative hip-knee-ankle deformity was 6.1° varus (SD 4.6° varus).Aims
Methods
The treatment of patients with allergies to metal in total joint arthroplasty is an ongoing debate. Possibilities include the use of hypoallergenic prostheses, as well as the use of standard cobalt-chromium (CoCr) alloy. This non-designer study was performed to evaluate the clinical outcome and survival rates of unicondylar knee arthroplasty (UKA) using a standard CoCr alloy in patients reporting signs of a hypersensitivity to metal. A consecutive series of patients suitable for UKA were screened for symptoms of metal hypersensitivity by use of a questionnaire. A total of 82 patients out of 1737 patients suitable for medial UKA reporting cutaneous metal hypersensitivity to cobalt, chromium, or nickel were included into this study and prospectively evaluated to determine the functional outcome, possible signs of hypersensitivity, and short-term survivorship at a minimum follow-up of 1.5 years.Aims
Patients and Methods
The objectives of this study were to compare postoperative pain, analgesia requirements, inpatient functional rehabilitation, time to hospital discharge, and complications in patients undergoing conventional jig-based unicompartmental knee arthroplasty (UKA) This prospective cohort study included 146 patients with symptomatic medial compartment knee osteoarthritis undergoing primary UKA performed by a single surgeon. This included 73 consecutive patients undergoing conventional jig-based mobile bearing UKA, followed by 73 consecutive patients receiving robotic-arm assisted fixed bearing UKA. All surgical procedures were performed using the standard medial parapatellar approach for UKA, and all patients underwent the same postoperative rehabilitation programme. Postoperative pain scores on the numerical rating scale and opiate analgesia consumption were recorded until discharge. Time to attainment of predefined functional rehabilitation outcomes, hospital discharge, and postoperative complications were recorded by independent observers.Aims
Patients and Methods
The aim of this study was to assess the effect of posterior cruciate ligament (PCL) resection on flexion-extension gaps, mediolateral soft-tissue laxity, fixed flexion deformity (FFD), and limb alignment during posterior-stabilized (PS) total knee arthroplasty (TKA). This prospective study included 110 patients with symptomatic osteoarthritis of the knee undergoing primary robot-assisted PS TKA. All operations were performed by a single surgeon using a standard medial parapatellar approach. Optical motion capture technology with fixed femoral and tibial registration pins was used to assess gaps before and after PCL resection in extension and 90° knee flexion. Measurements were made after excision of the anterior cruciate ligament and prior to bone resection. There were 54 men (49.1%) and 56 women (50.9%) with a mean age of 68 years (Aims
Patients and Methods
Unicompartmental knee arthroplasty (UKA) is one surgical option for treating symptomatic medial osteoarthritis. Clinical studies have shown the functional benefits of UKA; however, the optimal alignment of the tibial component is still debated. The purpose of this study was to evaluate the effects of tibial coronal and sagittal plane alignment in UKA on knee kinematics and cruciate ligament tension, using a musculoskeletal computer simulation. The tibial component was first aligned perpendicular to the mechanical axis of the tibia, with a 7° posterior slope (basic model). Subsequently, coronal and sagittal plane alignments were changed in a simulation programme. Kinematics and cruciate ligament tensions were simulated during weight-bearing deep knee bend and gait motions. Translation was defined as the distance between the most medial and the most lateral femoral positions throughout the cycle.Objectives
Methods
Between 15% and 20% of patients remain dissatisfied following total knee arthroplasty (TKA). The SAIPH knee system (MatOrtho, Surrey, United Kingdom) is a medial ball and socket TKA that has been designed to replicate native knee kinematics in order to maximize the range of movement, stability, and function. This system is being progressively introduced in a stepwise fashion, with this study reporting the mid-term clinical and radiological outcomes. A retrospective review was undertaken of the first 100 consecutive patients with five-year follow-up following SAIPH TKA performed by the senior authors. The data that were collected included the demographics of the patients, clinical findings, the rate of intraoperative ligamentous release, patient-reported outcome measures (PROMS), radiological assessment, complications, and all-cause revision. Revision data were cross-checked with a national registry.Aims
Patients and Methods
The primary aim of this study was to determine the surgical team’s
learning curve for introducing robotic-arm assisted unicompartmental
knee arthroplasty (UKA) into routine surgical practice. The secondary
objective was to compare accuracy of implant positioning in conventional
jig-based UKA versus robotic-arm assisted UKA. This prospective single-surgeon cohort study included 60 consecutive
conventional jig-based UKAs compared with 60 consecutive robotic-arm
assisted UKAs for medial compartment knee osteoarthritis. Patients
undergoing conventional UKA and robotic-arm assisted UKA were well-matched
for baseline characteristics including a mean age of 65.5 years
(Aims
Patients and Methods
The number of revision total knee arthroplasties (TKA) that are
performed is expected to increase. However, previous reports of
the causes of failure after TKA are limited in that they report
the causes at specific institutions, which are often dependent on
referral patterns. Our aim was to report the most common indications
for re-operations and revisions in a large series of posterior-stabilised
TKAs undertaken at a single institution, excluding referrals from
elsewhere, which may bias the causes of failure. A total of 5098 TKAs which were undertaken between 2000 and 2012
were included in the study. Re-operations, revisions with modular
component exchange, and revisions with non-modular component replacement
or removal were identified from the medical records. The mean follow-up
was five years (two to 12).Aims
Patients and Methods
Little biomechanical information is available about kinematically aligned (KA) total knee arthroplasty (TKA). The purpose of this study was to simulate the kinematics and kinetics after KA TKA and mechanically aligned (MA) TKA with four different limb alignments. Bone models were constructed from one volunteer (normal) and three patients with three different knee deformities (slight, moderate and severe varus). A dynamic musculoskeletal modelling system was used to analyse the kinematics and the tibiofemoral contact force. The contact stress on the tibial insert, and the stress to the resection surface and medial tibial cortex were examined by using finite element analysis.Objectives
Materials and Methods
Total knee arthroplasty (TKA) is a cost effective
and extremely successful operation. As longevity increases, the demand
for primary TKA will continue to rise. The success and survivorship
of TKAs are dependent on the demographics of the patient, surgical
technique and implant-related factors. Currently the risk of failure of a TKA requiring revision surgery
ten years post-operatively is 5%. The most common indications for revision include aseptic loosening
(29.8%), infection (14.8%), and pain (9.5%). Revision surgery poses
considerable clinical burdens on patients and financial burdens
on healthcare systems. We present a current concepts review on the epidemiology of failed
TKAs using data from worldwide National Joint Registries. Cite this article:
Since redesign of the Oxford phase III mobile-bearing unicompartmental
knee arthroplasty (UKA) femoral component to a twin-peg design,
there has not been a direct comparison to total knee arthroplasty
(TKA). Thus, we explored differences between the two cohorts. A total of 168 patients (201 knees) underwent medial UKA with
the Oxford Partial Knee Twin-Peg. These patients were compared with
a randomly selected group of 177 patients (189 knees) with primary
Vanguard TKA. Patient demographics, Knee Society (KS) scores and
range of movement (ROM) were compared between the two cohorts. Additionally,
revision, re-operation and manipulation under anaesthesia rates
were analysed.Aims
Patients and Methods
Previous studies of failure mechanisms leading
to revision total knee replacement (TKR) performed between 1986 and
2000 determined that many failed early, with a disproportionate
amount accounted for by infection and implant-associated factors
including wear, loosening and instability. Since then, efforts have
been made to improve implant performance and instruct surgeons in
best practice. Recently our centre participated in a multi-centre evaluation
of 844 revision TKRs from 2010 to 2011. The purpose was to report
a detailed analysis of failure mechanisms over time and to see if
failure modes have changed over the past 10 to 15 years. Aseptic
loosening was the predominant mechanism of failure (31.2%), followed
by instability (18.7%), infection (16.2%), polyethylene wear (10.0%),
arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to
failure was 5.9 years (ten days to 31 years), 35.3% of all revisions
occurred at less than two years, and 60.2% in the first five years.
With improvements in implant and polyethylene manufacture, polyethylene
wear is no longer a leading cause of failure. Early mechanisms of
failure are primarily technical errors. In addition to improving
implant longevity, industry and surgeons must work together to decrease
these technical errors. All reports on failure of TKR contain patients
with unexplained pain who not infrequently have unmet expectations.
Surgeons must work to achieve realistic patient expectations pre-operatively,
and therefore, improve patient satisfaction post-operatively. Cite this article:
This prospective study reports the 15-year survival and ten-year
functional outcome of a consecutive series of 1000 minimally invasive
Phase 3 Oxford medial UKAs (818 patients, 393 men, 48%, 425 women,
52%, mean age 66 years; 32 to 88). These were implanted by two surgeons
involved with the design of the prosthesis to treat anteromedial
osteoarthritis and spontaneous osteonecrosis of the knee, which
are recommended indications. Patients were prospectively identified
and followed up independently for a mean of 10.3 years (5.3 to 16.6). At ten years, the mean Oxford Knee Score was 40 (standard deviation
( This is the only large series of minimally invasive UKAs with
15-year survival data. The results support the continued use of
minimally invasive UKA for the recommended indications. Cite this article: