Baker JF Nadar AC Jouflas AC Smith LS Sachdeva S Yakkanti MR Malkani AL

Aims

The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design.

Methods

This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.


Aims

Achievement of accurate microbiological diagnosis prior to revision is key to reducing the high rates of persistent infection after revision knee surgery. The effect of change in the microorganism between the first- and second-stage revision of total knee arthroplasty for periprosthetic joint infection (PJI) on the success of management is not clear.

Methods

A two-centre retrospective cohort study was conducted to review the outcome of patients who have undergone two-stage revision for treatment of knee arthroplasty PJI, focusing specifically on isolated micro-organisms at both the first- and second-stage procedure. Patient demographics, medical, and orthopaedic history data, including postoperative outcomes and subsequent treatment, were obtained from the electronic records and medical notes.


Bone & Joint Open
Vol. 2, Issue 12 | Pages 1075 - 1081
17 Dec 2021
Suthar A Yukata K Azuma Y Suetomi Y Yamazaki K Seki K Sakai T Fujii H

Aims

This study aimed to investigate the relationship between changes in patellar height and clinical outcomes at a mean follow-up of 7.7 years (5 to 10) after fixed-bearing posterior-stabilized total knee arthroplasty (PS-TKA).

Methods

We retrospectively evaluated knee radiographs of 165 knees, which underwent fixed-bearing PS-TKA with patella resurfacing. The incidence of patella baja and changes in patellar height over a minimum of five years of follow-up were determined using Insall-Salvati ratio (ISR) measurement. We examined whether patella baja (ISR < 0.8) at final follow-up affected clinical outcomes, knee joint range of motion (ROM), and Knee Society Score (KSS). We also assessed inter- and intrarater reliability of ISR measurements and focused on the relationship between patellar height reduction beyond measurement error and clinical outcomes.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 477 - 480
1 Apr 2007
Spencer JM Chauhan SK Sloan K Taylor A Beaver RJ

We previously compared the component alignment in total knee replacement using a computer-navigated technique with a conventional jig-based method. We randomly allocated 71 patients to undergo either computer-navigated or conventional replacement. An improved alignment was seen in the computer-navigated group. The patients were then followed up post-operatively for two years, using the Knee Society score, the Short Form-36 health survey, the Western Ontario and McMaster Universities osteoarthritis index, the Bartlett Patellar pain questionnaire and the Oxford knee score, to assess functional outcome. At two years post-operatively 60 patients were available for assessment, 30 in each group and 62 patients completed a postal survey. No patient in either group had undergone revision. All variables were analysed for differences between the groups either by Student’s t-test or the Mann-Whitney U test. Differences between the two groups did not reach significance for any of the outcome measures at any time point. At two years postoperatively, the frequency of mild to severe anterior pain was not significantly different (p = 0.818), varying between 44% (14) for the computer-navigated group, and 47% (14) for the conventionally-replaced group. The Bartlett Patellar score and the Oxford knee score were also not significantly different (t-test p = 0.161 and p = 0.607, respectively). The clinical outcome of the patients with a computer-navigated knee replacement appears to be no different to that of a more conventional jig-based technique at two years post-operatively, despite the better alignment achieved with computer-navigated surgery


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 2 | Pages 195 - 198
1 Mar 2001
Walker CRC Myles C Nutton R Rowe P

We used electrogoniometers to measure the range of movement (ROM) of the knee during various activities, comparing 50 patients with osteoarthritis of the knee (OA) with 20 healthy age- and sex-matched subjects. The minimum and maximum joint angles and the ranges of excursion of the patient and control groups were tested for significant differences, using an unrelated Student’s t-test with pooled variance. Knee flexion in patients with OA was significantly reduced during all activities (p < 0.05), but differences in knee extension were not significant except when patients negotiated stairs. We believe that this reduction in ROM is caused by inhibition due to pain when load-bearing. Static non-load-bearing measurements of the ROM poorly reflected the functional ROM, with a coefficient of determination (r. 2. ) of 0.59 in the patient group and 0.60 in the control group. Electrogoniometry of the ROM of the knee provides a reliable, accurate and objective measurement of knee function


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 59 - 65
1 Jun 2020
Kwon Y Arauz P Peng Y Klemt C

Aims

The removal of the cruciate ligaments in total knee arthroplasty (TKA) has been suggested as a potential contributing factor to patient dissatisfaction, due to alteration of the in vivo biomechanics of the knee. Bicruciate retaining (BCR) TKA allows the preservation of the cruciate ligaments, thus offering the potential to reproduce healthy kinematics. The aim of this study was to compare in vivo kinematics between the operated and contralateral knee in patients who have undergone TKA with a contemporary BCR design.

Methods

A total of 29 patients who underwent unilateral BCR TKA were evaluated during single-leg deep lunges and sit-to-stand tests using a validated computer tomography and fluoroscopic imaging system. In vivo six-degrees of freedom (6DOF) kinematics were compared between the BCR TKA and the contralateral knee.


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 565 - 572
1 May 2019
Teeter MG Marsh JD Howard JL Yuan X Vasarhelyi EM McCalden RW Naudie DDR

Aims

The purpose of the present study was to compare patient-specific instrumentation (PSI) and conventional surgical instrumentation (CSI) for total knee arthroplasty (TKA) in terms of early implant migration, alignment, surgical resources, patient outcomes, and costs.

Patients and Methods

The study was a prospective, randomized controlled trial of 50 patients undergoing TKA. There were 25 patients in each of the PSI and CSI groups. There were 12 male patients in the PSI group and seven male patients in the CSI group. The patients had a mean age of 69.0 years (sd 8.4) in the PSI group and 69.4 years (sd 8.4) in the CSI group. All patients received the same TKA implant. Intraoperative surgical resources and any surgical waste generated were recorded. Patients underwent radiostereometric analysis (RSA) studies to measure femoral and tibial component migration over two years. Outcome measures were recorded pre- and postoperatively. Overall costs were calculated for each group.


Aims

Patient-specific instrumentation of total knee arthroplasty (TKA) is a technique permitting the targeting of individual kinematic alignment, but deviation from a neutral mechanical axis may have implications on implant fixation and therefore survivorship. The primary objective of this randomized controlled study was to compare the fixation of tibial components implanted with patient-specific instrumentation targeting kinematic alignment (KA+PSI) versus components placed using computer-assisted surgery targeting neutral mechanical alignment (MA+CAS). Tibial component migration measured by radiostereometric analysis was the primary outcome measure (compared longitudinally between groups and to published acceptable thresholds). Secondary outcome measures were inducible displacement after one year and patient-reported outcome measures (PROMS) over two years. The secondary objective was to assess the relationship between alignment and both tibial component migration and inducible displacement.

Patients and Methods

A total of 47 patients due to undergo TKA were randomized to KA+PSI (n = 24) or MA+CAS (n = 23). In the KA+PSI group, there were 16 female and eight male patients with a mean age of 64 years (sd 8). In the MA+CAS group, there were 17 female and six male patients with a mean age of 63 years (sd 7). Surgery was performed using cemented, cruciate-retaining Triathlon total knees with patellar resurfacing, and patients were followed up for two years. The effect of alignment on tibial component migration and inducible displacement was analyzed irrespective of study group.


Bone & Joint Open
Vol. 1, Issue 2 | Pages 8 - 12
18 Feb 2020
Bhimani SJ Bhimani R Smith A Eccles C Smith L Malkani A

Aims

Robotic-assisted total knee arthroplasty (RA-TKA) has been introduced to provide accurate bone cuts and help achieve the target knee alignment, along with symmetric gap balancing. The purpose of this study was to determine if any early clinical benefits could be realized following TKA using robotic-assisted technology.

Methods

In all, 140 consecutive patients undergoing RA-TKA and 127 consecutive patients undergoing conventional TKA with minimum six-week follow-up were reviewed. Differences in visual analogue scores (VAS) for pain at rest and with activity, postoperative opiate usage, and length of stay (LOS) between the RA-TKA and conventional TKA groups were compared.


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 559 - 564
1 May 2019
Takemura S Minoda Y Sugama R Ohta Y Nakamura S Ueyama H Nakamura H

Aims

The use of vitamin E-infused highly crosslinked polyethylene (HXLPE) in total knee prostheses is controversial. In this paper we have compared the clinical and radiological results between conventional polyethylene and vitamin E-infused HXLPE inserts in total knee arthroplasty (TKA).

Patients and Methods

The study included 200 knees (175 patients) that underwent TKA using the same total knee prostheses. In all, 100 knees (77 patients) had a vitamin E-infused HXLPE insert (study group) and 100 knees (98 patients) had a conventional polyethylene insert (control group). There were no significant differences in age, sex, diagnosis, preoperative knee range of movement (ROM), and preoperative Knee Society Score (KSS) between the two groups. Clinical and radiological results were evaluated at two years postoperatively.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1081 - 1086
1 Sep 2019
Murphy WS Harris S Pahalyants V Zaki MM Lin B Cheng T Talmo C Murphy SB

Aims

The practice of alternating operating theatres has long been used to reduce surgeon idle time between cases. However, concerns have been raised as to the safety of this practice. We assessed the payments and outcomes of total knee arthroplasty (TKA) performed during overlapping and nonoverlapping days, also comparing the total number of the surgeon’s cases and the total time spent in the operating theatre per day.

Materials and Methods

A retrospective analysis was performed on the Centers for Medicare & Medicaid Services (CMS) Limited Data Set (LDS) on all primary elective TKAs performed at the New England Baptist Hospital between January 2013 and June 2016. Using theatre records, episodes were categorized into days where a surgeon performed overlapping and nonoverlapping lists. Clinical outcomes, economic outcomes, and demographic factors were calculated. A regression model controlling for the patient-specific factors was used to compare groups. Total orthopaedic cases and aggregate time spent operating (time between skin incision and closure) were also compared.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 77 - 83
1 Jul 2019
James EW Blevins JL Gausden EB Turcan S Denova TA Satalich JR Ranawat AS Warren RF Ranawat AS

Aims

Anterior cruciate ligament (ACL) and multiligament knee (MLK) injuries increase the risk of development of knee osteoarthritis and eventual need for total knee arthroplasty (TKA). There are limited data regarding implant use and outcomes in these patients. The aim of this study was to compare the use of constrained implants and outcomes among patients undergoing TKA with a history of prior knee ligament reconstruction (PKLR) versus a matched cohort of patients undergoing TKA with no history of PKLR.

Patients and Methods

Patients with a history of ACL or MLK reconstruction who underwent TKA between 2007 and 2017 were identified in a single-institution registry. There were 223 patients who met inclusion criteria (188 ACL reconstruction patients, 35 MLK reconstruction patients). A matched cohort, also of 223 patients, was identified based on patient age, body mass index (BMI), sex, and year of surgery. There were 144 male patients and 79 female patients in both cohorts. Mean age at the time of TKA was 57.2 years (31 to 88). Mean BMI was 29.7 kg/m2 (19.5 to 55.7).


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 3 - 9
1 Jul 2019
Shohat N Tarabichi M Tan TL Goswami K Kheir M Malkani AL Shah RP Schwarzkopf R Parvizi J

Aims

The best marker for assessing glycaemic control prior to total knee arthroplasty (TKA) remains unknown. The purpose of this study was to assess the utility of fructosamine compared with glycated haemoglobin (HbA1c) in predicting early complications following TKA, and to determine the threshold above which the risk of complications increased markedly.

Patients and Methods

This prospective multi-institutional study evaluated primary TKA patients from four academic institutions. Patients (both diabetics and non-diabetics) were assessed using fructosamine and HbA1c levels within 30 days of surgery. Complications were assessed for 12 weeks from surgery and included prosthetic joint infection (PJI), wound complication, re-admission, re-operation, and death. The Youden’s index was used to determine the cut-off for fructosamine and HbA1c associated with complications. Two additional cut-offs for HbA1c were examined: 7% and 7.5% and compared with fructosamine as a predictor for complications.


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 552 - 558
1 May 2019
Clark D Stevens JM Tortonese D Whitehouse MR Simpson D Eldridge J

Aims

The aim of this study was to determine and compare the congruency of the articular surface contact area of the patellofemoral joint (PFJ) during both active and passive movement of the knee with the use of an MRI mapping technique in both the stable and unstable PFJ.

Patients and Methods

A prospective case-control MRI imaging study of patients with a history of PFJ instability and a control group of volunteers without knee symptoms was performed. The PFJs were imaged with the use of an MRI scan during both passive and active movement from 0° through to 40° of flexion. The congruency through measurement of the contact surface area was mapped in 5-mm intervals on axial slices. In all, 40 patients were studied. The case group included 31 patients with symptomatic patellofemoral instability and the control group of nine asymptomatic volunteers. The ages were well matched between the case and control groups. The mean age was 25 years (16 to 42; sd 6.9) in the case group and 26 years (19 to 32; sd 5.1) in the control group. There were 19 female and 12 male patients in the case group.


The Bone & Joint Journal
Vol. 101-B, Issue 3 | Pages 331 - 339
1 Mar 2019
McEwen P Balendra G Doma K

Aims

The results of kinematic total knee arthroplasty (KTKA) have been reported in terms of limb and component alignment parameters but not in terms of gap laxities and differentials. In kinematic alignment (KA), balance should reflect the asymmetrical balance of the normal knee, not the classic rectangular flexion and extension gaps sought with gap-balanced mechanical axis total knee arthroplasty (MATKA). This paper aims to address the following questions: 1) what factors determine coronal joint congruence as measured on standing radiographs?; 2) is flexion gap asymmetry produced with KA?; 3) does lateral flexion gap laxity affect outcomes?; 4) is lateral flexion gap laxity associated with lateral extension gap laxity?; and 5) can consistent ligament balance be produced without releases?

Patients and Methods

A total of 192 KTKAs completed by a single surgeon using a computer-assisted technique were followed for a mean of 3.5 years (2 to 5). There were 116 male patients (60%) and 76 female patients (40%) with a mean age of 65 years (48 to 88). Outcome measures included intraoperative gap laxity measurements and component positions, as well as joint angles from postoperative three-foot standing radiographs. Patient-reported outcome measures (PROMs) were analyzed in terms of alignment and balance: EuroQol (EQ)-5D visual analogue scale (VAS), Knee Injury and Osteoarthritis Outcome Score (KOOS), KOOS Joint Replacement (JR), and Oxford Knee Score (OKS).


The Bone & Joint Journal
Vol. 101-B, Issue 6 | Pages 667 - 674
1 Jun 2019
Schwarzkopf R Novikov D Anoushiravani AA Feng JE Vigdorchik J Schurko B Dwyer MK Bedair HS

Aims

With an ageing population of patients who are infected with hepatitis C virus (HCV), the demand for total knee arthroplasty (TKA) in this high-risk group continues to grow. It has previously been shown that HCV infection predisposes to poor outcomes following TKA. However, there is little information about the outcome of TKA in patients with HCV who have been treated successfully. The purpose of this study was to compare the outcomes of TKA in untreated HCV patients and those with HCV who have been successfully treated and have a serologically confirmed remission.

Patients and Methods

A retrospective review of all patients diagnosed with HCV who underwent primary TKA between November 2011 and April 2018 was conducted. HCV patients were divided into two groups: 1) those whose HCV was cured (HCV-C); and 2) those in whom it was untreated (HCV-UT). All variables including demographics, HCV infection characteristics, surgical details, and postoperative medical and surgical outcomes were evaluated. There were 64 patients (70 TKAs) in the HCV-C group and 63 patients (71 TKAs) in the HCV-UT cohort. The mean age at the time of surgery was 63.0 years (sd 7.5; 44 to 79) in the HCV-C group and 61.7 years (sd 6.9; 47 to 88) in the HCV-UT group.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1087 - 1092
1 Sep 2019
Garceau S Warschawski Y Dahduli O Alshaygy I Wolfstadt J Backstein D

Aims

The aim of this study was to assess the effects of transferring patients to a specialized arthroplasty centre between the first and second stages (interstage) of prosthetic joint infection (PJI) of the knee.

Patients and Methods

A search of our institutional database was performed to identify patients having undergone two-stage revision total knee arthroplasty (TKA) for PJI. Two cohorts were created: continuous care (CC) and transferred care (TC). Baseline characteristics and outcomes were collected and compared between cohorts.


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1230 - 1237
1 Oct 2019
Kayani B Konan S Horriat S Ibrahim MS Haddad FS

Aims

The aim of this study was to assess the effect of posterior cruciate ligament (PCL) resection on flexion-extension gaps, mediolateral soft-tissue laxity, fixed flexion deformity (FFD), and limb alignment during posterior-stabilized (PS) total knee arthroplasty (TKA).

Patients and Methods

This prospective study included 110 patients with symptomatic osteoarthritis of the knee undergoing primary robot-assisted PS TKA. All operations were performed by a single surgeon using a standard medial parapatellar approach. Optical motion capture technology with fixed femoral and tibial registration pins was used to assess gaps before and after PCL resection in extension and 90° knee flexion. Measurements were made after excision of the anterior cruciate ligament and prior to bone resection. There were 54 men (49.1%) and 56 women (50.9%) with a mean age of 68 years (sd 6.2) at the time of surgery. The mean preoperative hip-knee-ankle deformity was 4.1° varus (sd 3.4).


The Bone & Joint Journal
Vol. 101-B, Issue 1 | Pages 24 - 33
1 Jan 2019
Kayani B Konan S Tahmassebi J Rowan FE Haddad FS

Aims

The objectives of this study were to compare postoperative pain, analgesia requirements, inpatient functional rehabilitation, time to hospital discharge, and complications in patients undergoing conventional jig-based unicompartmental knee arthroplasty (UKA) versus robotic-arm assisted UKA.

Patients and Methods

This prospective cohort study included 146 patients with symptomatic medial compartment knee osteoarthritis undergoing primary UKA performed by a single surgeon. This included 73 consecutive patients undergoing conventional jig-based mobile bearing UKA, followed by 73 consecutive patients receiving robotic-arm assisted fixed bearing UKA. All surgical procedures were performed using the standard medial parapatellar approach for UKA, and all patients underwent the same postoperative rehabilitation programme. Postoperative pain scores on the numerical rating scale and opiate analgesia consumption were recorded until discharge. Time to attainment of predefined functional rehabilitation outcomes, hospital discharge, and postoperative complications were recorded by independent observers.


Bone & Joint Research
Vol. 9, Issue 1 | Pages 15 - 22
1 Jan 2020
Clement ND Bell A Simpson P Macpherson G Patton JT Hamilton DF

Aims

The primary aim of the study was to compare the knee-specific functional outcome of robotic unicompartmental knee arthroplasty (rUKA) with manual total knee arthroplasty (mTKA) for the management of isolated medial compartment osteoarthritis. Secondary aims were to compare length of hospital stay, general health improvement, and satisfaction between rUKA and mTKA.

Methods

A powered (1:3 ratio) cohort study was performed. A total of 30 patients undergoing rUKA were propensity score matched to 90 patients undergoing mTKA for isolated medial compartment arthritis. Patients were matched for age, sex, body mass index (BMI), and preoperative function. The Oxford Knee Score (OKS) and EuroQol five-dimension questionnaire (EQ-5D) were collected preoperatively and six months postoperatively. The Forgotten Joint Score (FJS) and patient satisfaction were collected six months postoperatively. Length of hospital stay was also recorded.


The Bone & Joint Journal
Vol. 101-B, Issue 1 | Pages 34 - 40
1 Jan 2019
Kraus Schmitz J Lindgren V Janarv P Forssblad M Stålman A

Aims

The aim of this study was to investigate the incidence, risk factors, and outcome of venous thromboembolism (VTE) following anterior cruciate ligament (ACL) reconstruction in a nationwide cohort.

Patients and Methods

All ACL reconstructions, primary and revision, that were recorded in the Swedish Knee Ligament Register (SKLR) between 2006 and 2013 were linked with data from the Swedish National Board of Health and Welfare. The incidence of VTE was determined by entries between the day of surgery until 90 days postoperatively based on diagnosis codes and the prescription of anticoagulants. Risk factors, outcome, and the use of thromboprophylaxis were analyzed. Descriptive statistics with multivariate analysis were used to describe the findings.


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1330 - 1335
1 Oct 2018
Ponzio DY Weitzler L deMeireles A Esposito CI Wright TM Padgett DE

Aims

The aim of this study was to evaluate the surface damage, the density of crosslinking, and oxidation in retrieved antioxidant-stabilized highly crosslinked polyethylene (A-XLPE) tibial inserts from total knee arthroplasty (TKA), and to compare the results with a matched cohort of standard remelted highly crosslinked polyethylene (XLPE) inserts.

Materials and Methods

A total of 19 A-XLPE tibial inserts were retrieved during revision TKA and matched to 18 retrieved XLPE inserts according to the demographics of the patients, with a mean length of implantation of 15 months (1 to 42). The percentage areas of PE damage on the articular surfaces and the modes of damage were measured. The density of crosslinking of the PE and oxidation were measured at loaded and unloaded regions on these surfaces.


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1348 - 1355
1 Nov 2019
Gascoyne T Parashin S Teeter M Bohm E Laende E Dunbar M Turgeon T

Aims

A retrospective study was conducted to measure short-term in vivo linear and volumetric wear of polyethylene (PE) inserts in 101 total knee arthroplasty (TKA) patients using model-based radiostereometric analysis (MBRSA).

Patients and Methods

Nonweightbearing supine RSA exams were performed postoperatively and at six, 12, and 24 months. Weightbearing standing RSA exams were performed on select patients at 12 and 24 months. Wear was measured both linearly (joint space) and volumetrically (digital model overlap) at each available follow-up. Precision of both methods was assessed by comparing double RSA exams. Patient age, sex, body mass index, and Oxford Knee Scores were analyzed for any association with PE wear.


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1336 - 1344
1 Oct 2018
Powell AJ Crua E Chong BC Gordon R McAuslan A Pitto RP Clatworthy MG

Aims

This study compares the PFC total knee arthroplasty (TKA) system in a prospective randomized control trial (RCT) of the mobile-bearing rotating-platform (RP) TKA against the fixed-bearing (FB) TKA. This is the largest RCT with the longest follow-up where cruciate-retaining PFC total knee arthroplasties are compared in a non-bilateral TKA study.

Patients and Methods

A total of 167 patients (190 knees with 23 bilateral cases), were recruited prospectively and randomly assigned, with 91 knees receiving the RP and 99 knees receiving FB. The mean age was 65.5 years (48 to 82), the mean body mass index (BMI) was 29.7 kg/m2 (20 to 52) and 73 patients were female. The Knee Society Score (KSS), Knee Society Functional Score (KSFS), Oxford Knee Score (OKS), Western Ontario and McMaster Universities Arthritis Index (WOMAC), and 12-Item Short-Form Health Survey Physical and Mental Component Scores (SF-12 PCS, SF-12 MCS) were gathered and recorded preoperatively, at five-years’ follow-up, and at ten years’ follow-up. Additionally, Knee Injury and Osteoarthritis Outcome Scores (KOOS) were collected at five- and ten-year follow-ups. The prevalence of radiolucent lines (RL) on radiographs and implant survival were recorded at five- and ten-year follow-ups.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1058 - 1062
1 Sep 2019
van Kuijk KSR Reijman M Bierma-Zeinstra SMA Waarsing JH Meuffels DE

Aims

Little is known about the risk factors that predispose to a rupture of the posterior cruciate ligament (PCL). Identifying risk factors is the first step in trying to prevent a rupture of the PCL from occurring. The morphology of the knee in patients who rupture their PCL may differ from that of control patients. The purpose of this study was to identify any variations in bone morphology that are related to a PCL.

Patients and Methods

We compared the anteroposterior (AP), lateral, and Rosenberg view radiographs of 94 patients with a ruptured PCL to a control group of 168 patients matched by age, sex, and body mass index (BMI), but with an intact PCL after a knee injury. Statistical shape modelling software was used to assess the shape of the knee and determine any difference in anatomical landmarks.


The Bone & Joint Journal
Vol. 101-B, Issue 1 | Pages 41 - 46
1 Jan 2019
Clement ND Howard TA Immelman RJ MacDonald D Patton JT Lawson GM Burnett R

Aims

The primary aim of this study was to compare the knee-specific functional outcome of patellofemoral arthroplasty with total knee arthroplasty (TKA) in the management of patients with patellofemoral osteoarthritis.

Patients and Methods

A total of 54 consecutive Avon patellofemoral arthroplasties were identified and propensity-score-matched to a group of 54 patients undergoing a TKA with patellar resurfacing for patellofemoral osteoarthritis. The Oxford Knee Score (OKS), the 12-Item Short-Form Health Survey (SF-12), and patient satisfaction were collected at a mean follow up of 9.2 years (8 to 15). Survival was defined by revision or intention to revise.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 104 - 107
1 Jul 2019
Greenwell PH Shield WP Chapman DM Dalury DF

Aims

The aim of this study was to establish the results of isolated exchange of the tibial polyethylene insert in revision total knee arthroplasty (RTKA) in patients with well-fixed femoral or tibial components. We report on a series of RTKAs where only the polyethylene was replaced, and the patients were followed for a mean of 13.2 years (10.0 to 19.1).

Patients and Methods

Our study group consisted of 64 non-infected, grossly stable TKA patients revised over an eight-year period (1998 to 2006). The mean age of the patients at time of revision was 72.2 years (48 to 88). There were 36 females (56%) and 28 males (44%) in the cohort. All patients had received the same cemented, cruciate-retaining patella resurfaced primary TKA. All subsequently underwent an isolated polyethylene insert exchange. The mean time from the primary TKA to RTKA was 9.1 years (2.2 to 16.1).


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 108 - 114
1 Jul 2019
Ji G Xu R Niu Y Li N Ivashkiv L Bostrom MPG Greenblatt MB Yang X

Aims

It is increasingly appreciated that coordinated regulation of angiogenesis and osteogenesis is needed for bone formation. How this regulation is achieved during peri-implant bone healing, such as osseointegration, is largely unclear. This study examined the relationship between angiogenesis and osteogenesis in a unique model of osseointegration of a mouse tibial implant by pharmacologically blocking the vascular endothelial growth factor (VEGF) pathway.

Materials and Methods

An implant was inserted into the right tibia of 16-week-old female C57BL/6 mice (n = 38). Mice received anti-VEGF receptor-1 (VEGFR-1) antibody (25 mg/kg) and VEGF receptor-2 (VEGFR-2) antibody (25 mg/kg; n = 19) or an isotype control antibody (n = 19). Flow cytometric (n = 4/group) and immunofluorescent (n = 3/group) analyses were performed at two weeks post-implantation to detect the distribution and density of CD31hiEMCNhi endothelium. RNA sequencing analysis was performed using sorted CD31hiEMCNhi endothelial cells (n = 2/group). Osteoblast lineage cells expressing osterix (OSX) and osteopontin (OPN) were also detected with immunofluorescence. Mechanical pull-out testing (n = 12/group) was used at four weeks post-implantation to determine the strength of the bone-implant interface. After pull-out testing, the tissue attached to the implant surface was harvested. Whole mount immunofluorescent staining of OSX and OPN was performed to determine the amount of osteoblast lineage cells.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 64 - 69
1 Jul 2019
Wodowski AJ Pelt CE Erickson JA Anderson MB Gililland JM Peters CL

Aims

The Bundled Payments for Care Improvement (BPCI) initiative has identified pathways for improving the value of care. However, patient-specific modifiable and non-modifiable risk factors may increase costs beyond the target payment. We sought to identify risk factors for exceeding our institution’s target payment, the so-called ‘bundle busters’.

Patients and Methods

Using our data warehouse and Centers for Medicare and Medicaid Services (CMS) data we identified all 412 patients who underwent total joint arthroplasty and qualified for our institution’s BPCI model, between July 2015 and May 2017. Episodes where CMS payments exceeded the target payment were considered ‘busters’ (n = 123). Risk ratios (RRs) were calculated using a modified Poisson regression analysis.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 22 - 27
1 Jul 2019
Kalbian IL Tan TL Rondon AJ Bonaddio VA Klement MR Foltz C Lonner JH

Aims

Unicompartmental knee arthroplasty (UKA) provides improved early functional outcomes and less postoperative morbidity and pain compared with total knee arthroplasty (TKA). Opioid prescribing has increased in the last two decades, and recently states in the USA have developed online Prescription Drug Monitoring Programs to prevent overprescribing of controlled substances. This study evaluates differences in opioid requirements between patients undergoing TKA and UKA.

Patients and Methods

We retrospectively reviewed 676 consecutive TKAs and 241 UKAs. Opioid prescriptions in morphine milligram equivalents (MMEs), sedatives, benzodiazepines, and stimulants were collected from State Controlled Substance Monitoring websites six months before and nine months after the initial procedures. Bivariate and multivariate analysis were performed for patients who had a second prescription and continued use.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 84 - 90
1 Jul 2019
Jennings JM Loyd BJ Miner TM Yang CC Stevens-Lapsley J Dennis DA

Aims

The aim of this study was to determine whether closed suction drain (CSD) use influences recovery of quadriceps strength and to examine the effects of drain use on secondary outcomes: quadriceps activation, intra-articular effusion, bioelectrical measure of swelling, range of movement (ROM), pain, and wound healing complications.

Patients and Methods

A total of 29 patients undergoing simultaneous bilateral total knee arthroplasty (TKA) were enrolled in a prospective, randomized blinded study. Patients were randomized to receive a CSD in one limb while the contralateral limb had the use of a subcutaneous drain (SCDRN) without the use of suction (‘sham drain’). Isometric quadriceps strength was collected as the primary outcome. Secondary outcomes consisted of quadriceps activation, intra-articular effusion measured via ultrasound, lower limb swelling measured with bioelectrical impendence and limb girth, knee ROM, and pain. Outcomes were assessed preoperatively and postoperatively at day two, two and six weeks, and three months. Differences between limbs were determined using paired Student’s t-tests or Wilcoxon’s signed-rank tests.


Bone & Joint Research
Vol. 7, Issue 9 | Pages 541 - 547
1 Sep 2018
Eijkenboom JFA Waarsing JH Oei EHG Bierma-Zeinstra SMA van Middelkoop M

Objectives

It has been hypothesized that patellofemoral pain, a common knee condition in adolescents and young adults, may be a precursor of degenerative joint changes and may ultimately lead to patellofemoral osteoarthritis. Since both conditions share several mechanical disease characteristics, such as altered contact area between the femur and patella and increased joint stress, we investigated whether these conditions share similar and different shape characteristics of the patella compared with normal controls.

Methods

This cross-sectional study compared three different study populations: 32 patellofemoral pain subjects (mean age, 32 years (22 to 45); 72% female); 56 isolated radiological patellofemoral osteoarthritis subjects (mean age, 54 years (44 to 58); 89% female); and 80 healthy control subjects (mean age, 52 years (44 to 58); 74% female). Measurements included questionnaires, and lateral and skyline radiographs of the knee. Two separate 30-point 2D statistical shape models of the patella were created from the lateral and skyline radiographs. A general linear model was used to test for differences in standardized shape modes (a specific shape variant of the patella) between patellofemoral osteoarthritis, patellofemoral pain, and controls, using Bonferroni correction and adjustment for body mass index and gender.


The Bone & Joint Journal
Vol. 101-B, Issue 3 | Pages 340 - 347
1 Mar 2019
Elkassabany NM Cai LF Badiola I Kase B Liu J Hughes C Israelite CL Nelson CL

Aims

Adductor canal block (ACB) has emerged as an alternative to femoral nerve block (FNB) for analgesia after total knee arthroplasty (TKA). The optimal duration of maintenance of the ACB is still questionable. The purpose of this study was to compare the analgesic benefits and physiotherapy (PT) outcomes of single-shot ACB to two different regimens of infusion of the continuous ACB, 24-hour and 48-hour infusion.

Patients and Methods

This was a prospective, randomized, unblinded study. A total of 159 American Society of Anesthesiologists (ASA) physical status I to III patients scheduled for primary TKA were randomized to one of three study groups. Three patients did not complete the study, leaving 156 patients for final analysis. Group A (n = 53) was the single-shot group (16 female patients and 37 male patients with a mean age of 63.9 years (sd 9.6)), group B (n = 51) was the 24-hour infusion group (22 female patients and 29 male patients with a mean age of 66.5 years (sd 8.5)), and group C (n = 52) was the 48-hour infusion group (18 female patients and 34 male patients with a mean age of 62.2 years (sd 8.7)). Pain scores, opioid requirements, PT test results, and patient-reported outcome instruments were compared between the three groups.


Bone & Joint Research
Vol. 7, Issue 2 | Pages 166 - 172
1 Feb 2018
Bujnowski K Getgood A Leitch K Farr J Dunning C Burkhart TA

Aim

It has been suggested that the use of a pilot-hole may reduce the risk of fracture to the lateral cortex. Therefore the purpose of this study was to determine the effect of a pilot hole on the strains and occurrence of fractures at the lateral cortex during the opening of a high tibial osteotomy (HTO) and post-surgery loading.

Materials and Methods

A total of 14 cadaveric tibias were randomized to either a pilot hole (n = 7) or a no-hole (n = 7) condition. Lateral cortex strains were measured while the osteotomy was opened 9 mm and secured in place with a locking plate. The tibias were then subjected to an initial 800 N load that increased by 200 N every 5000 cycles, until failure or a maximum load of 2500 N.


The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1471 - 1476
1 Nov 2018
Weston JT Watts CD Mabry TM Hanssen AD Berry DJ Abdel MP

Aims

The results of irrigation and debridement with component retention (IDCR) in the treatment of acutely infected total knee arthroplasties (TKAs) have been variable. The aim of this study was to assess the outcome after IDCR when combined with chronic antibiotic suppression. We also evaluated survivorship free from subsequent infection, removal of the components, and death, as well as the risk factors for failure.

Patients and Methods

This was a single-centre retrospective review of 134 infected primary TKAs that were treated with IDCR. Infections within four weeks of the procedure were defined as acute postoperative infections, and those occurring more than four weeks after the procedure with symptoms for less than three weeks were defined as acute haematogenous infections. Patients were treated with intravenous antibiotics for four to six weeks, followed by chronic oral antibiotic suppression. Estimates of survival were made using a competing risk analysis. The mean follow-up was five years (2.1 to 13).


The Bone & Joint Journal
Vol. 100-B, Issue 2 | Pages 161 - 169
1 Feb 2018
Clement ND Bardgett M Weir D Holland J Gerrand C Deehan DJ

Aims

The primary aim of this study was to assess whether patient satisfaction one year after total knee arthroplasty (TKA) changed with longer follow-up. The secondary aims were to identify predictors of satisfaction at one year, persistence of patient dissatisfaction, and late onset dissatisfaction in patients that were originally satisfied at one year.

Patients and Methods

A retrospective cohort consisting of 1369 patients undergoing a primary TKA for osteoarthritis that had not undergone revision were identified from an established arthroplasty database. Patient demographics, comorbidities, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores, and Short Form 12 (SF-12) questionnaire scores were collected preoperatively, and one and five years postoperatively. In addition, patient satisfaction was assessed at one and five years postoperatively. Logistic regression analysis was used to identify independent predictors of satisfaction at one and five years.


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1313 - 1318
1 Oct 2017
Nakamura R Komatsu N Fujita K Kuroda K Takahashi M Omi R Katsuki Y Tsuchiya H

Aims

Open wedge high tibial osteotomy (OWHTO) for medial-compartment osteoarthritis of the knee can be complicated by intra-operative lateral hinge fracture (LHF). We aimed to establish the relationship between hinge position and fracture types, and suggest an appropriate hinge position to reduce the risk of this complication.

Patients and Methods

Consecutive patients undergoing OWHTO were evaluated on coronal multiplanar reconstruction CT images. Hinge positions were divided into five zones in our new classification, by their relationship to the proximal tibiofibular joint (PTFJ). Fractures were classified into types I, II, and III according to the Takeuchi classification.


Bone & Joint Research
Vol. 6, Issue 9 | Pages 535 - 541
1 Sep 2017
Zan P Mol MO Yao JJ Fan L Yang D Liu K Li G

Objectives

The length of the tourniquet time during total knee arthroplasty (TKA) is related to the incidence of post-operative deep vein thrombosis (DVT). Our aim in this study was to investigate the effect of the early release of the tourniquet on the incidence of DVT in patients undergoing TKA.

Methods

A total of 200 patients who underwent TKA between November 2015 and November 2016 were prospectively enrolled. The tourniquet was inflated before surgery and released immediately after the introduction of the components (early release group). This group was compared with a retrospective cohort of 200 primary TKAs, in which the tourniquet was released after the dressings had been applied (late release group). The presence of a DVT was detected using bilateral lower limb ultrasonography. Peri-operative clinical and follow-up data were collected for analysis.


The Bone & Joint Journal
Vol. 100-B, Issue 5 | Pages 579 - 583
1 May 2018
Xu S Chen JY Lo NN Chia SL Tay DKJ  Pang HN Hao Y Yeo SJ

Aims

This study investigated the influence of body mass index (BMI) on patients’ function and quality of life ten years after total knee arthroplasty (TKA).

Patients and Methods

A total of 126 patients who underwent unilateral TKA in 2006 were prospectively included in this retrospective study. They were categorized into two groups based on BMI: < 30 kg/m2 (control) and ≥ 30 kg/m2 (obese). Functional outcome was assessed using the Knee Society Function Score (KSFS), Knee Society Knee Score (KSKS), and Oxford Knee Score (OKS). Quality of life was assessed using the Physical (PCS) and Mental Component Scores (MCS) of the 36-Item Short-Form Health Survey.


The Bone & Joint Journal
Vol. 100-B, Issue 3 | Pages 285 - 293
1 Mar 2018
Nakamae A Adachi N Deie M Ishikawa M Nakasa T Ikuta Y Ochi M

Aims

To investigate the risk factors for progression of articular cartilage damage after anatomical anterior cruciate ligament (ACL) reconstruction.

Patients and Methods

A total of 174 patients who underwent second-look arthroscopic evaluation after anatomical ACL reconstruction were enrolled in this study. The graded condition of the articular cartilage at the time of ACL reconstruction was compared with that at second-look arthroscopy. Age, gender, body mass index (BMI), ACL reconstruction technique, meniscal conditions, and other variables were assessed by regression analysis as risk factors for progression of damage to the articular cartilage.


Bone & Joint Research
Vol. 6, Issue 3 | Pages 172 - 178
1 Mar 2017
Clement ND MacDonald DJ Hamilton DF Burnett R

Objectives

Preservation of posterior condylar offset (PCO) has been shown to correlate with improved functional results after primary total knee arthroplasty (TKA). Whether this is also the case for revision TKA, remains unknown. The aim of this study was to assess the independent effect of PCO on early functional outcome after revision TKA.

Methods

A total of 107 consecutive aseptic revision TKAs were performed by a single surgeon during an eight-year period. The mean age was 69.4 years (39 to 85) and there were 59 female patients and 48 male patients. The Oxford Knee Score (OKS) and Short-form (SF)-12 score were assessed pre-operatively and one year post-operatively. Patient satisfaction was also assessed at one year. Joint line and PCO were assessed radiographically at one year.


The Bone & Joint Journal
Vol. 100-B, Issue 2 | Pages 152 - 160
1 Feb 2018
Subramanyam KN Khanchandani P Tulajaprasad PV Jaipuria J Mundargi AV

Aims

The aims of this study were to compare the efficacy and safety of intra-articular and intravenous (IV) tranexamic acid (TXA) in controlling perioperative blood loss in total knee arthroplasty (TKA) using a randomized, double-blinded equivalence trial.

Patients and Methods

A total of 182 patients aged between 45 and 75 years undergoing unilateral TKA at a tertiary centre were randomized to receive TXA, either 1.5 g intra-articularly after closure of the wound (n = 91) or two doses of 10 mg/kg IV (n = 91). The primary outcome measure was the reduction in the level of haemoglobin (Hb) in the blood on the fifth postoperative day. Secondary outcome measures were the total, visible, and hidden blood losses (TBL, VBL, HBL). We assumed equivalence of the primary outcome in both routes with a margin of ± 0.35gm/dl. Block randomization using computer-generated random numbers was used. The patients and the assessor of outcome were blinded.


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1620 - 1624
1 Dec 2016
Pailhé R Cognault J Massfelder J Sharma A Rouchy R Rubens-Duval B Saragaglia D

Aims

The role of high tibial osteotomy (HTO) is being questioned by the use of unicompartmental knee arthroplasty (UKA) in the treatment of medial compartment femorotibial osteoarthritis. Our aim was to compare the outcomes of revision HTO or UKA to a total knee arthroplasty (TKA) using computer-assisted surgery in matched groups of patients.

Patients and Methods

We conducted a retrospective study to compare the clinical and radiological outcome of patients who underwent revision of a HTO to a TKA (group 1) with those who underwent revision of a medial UKA to a TKA (group 2). All revision procedures were performed using computer-assisted surgery. We extracted these groups of patients from our database. They were matched by age, gender, body mass index, follow-up and pre-operative functional score. The outcomes included the Knee Society Scores (KSS), radiological outcomes and the rate of further revision.


Bone & Joint Research
Vol. 5, Issue 7 | Pages 280 - 286
1 Jul 2016
Ozkurt B Sen T Cankaya D Kendir S Basarır K Tabak Y

Objectives

The purpose of this study was to develop an accurate, reliable and easily applicable method for determining the anatomical location of the joint line during revision knee arthroplasty.

Methods

The transepicondylar width (TEW), the perpendicular distance between the medial and lateral epicondyles and the distal articular surfaces (DMAD, DLAD) and the distance between the medial and lateral epicondyles and the posterior articular surfaces (PMAD, DLAD) were measured in 40 knees from 20 formalin-fixed adult cadavers (11 male and nine female; mean age at death 56.9 years, sd 9.4; 34 to 69). The ratios of the DMAD, PMAD, DLAD and PLAD to TEW were calculated.


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 887 - 893
1 Jul 2017
Ogawa H Matsumoto K Akiyama H

Aims

We aimed to investigate factors related to the technique of medial opening wedge high tibial osteotomy which might predispose to the development of a lateral hinge fracture.

Patients and Methods

A total of 71 patients with 82 osteotomies were included in the study. Their mean age was 62.9 years (37 to 80). The classification of the type of osteotomy was based on whether it extended beyond the fibular head. The level of the osteotomy was classified according to the height of its endpoint.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 483 - 488
1 Apr 2017
Pinsornsak P Nangnual S Boontanapibul K

Aims

Multimodal infiltration of local anaesthetic provides effective control of pain in patients undergoing total knee arthroplasty (TKA). There is little information about the added benefits of posterior capsular infiltration (PCI) using different combinations of local anaesthetic agents. Our aim was to investigate the effectiveness of the control of pain using multimodal infiltration with and without infiltration of the posterior capsule of the knee.

Patients and Methods

In a double-blind, randomised controlled trial of patients scheduled for unilateral primary TKA, 86 were assigned to be treated with multimodal infiltration with (Group I) or without (Group II) PCI. Routine associated analgesia included the use of bupivacaine, morphine, ketorolac and epinephrine. All patients had spinal anaesthesia and patient-controlled analgesia (PCA) post-operatively. A visual analogue scale (VAS) for pain and the use of morphine were recorded 24 hours post-operatively. Side effects of the infiltration, blood loss, and length of stay in hospital were recorded.


Bone & Joint Research
Vol. 6, Issue 6 | Pages 376 - 384
1 Jun 2017
Stentz-Olesen K Nielsen ET De Raedt S Jørgensen PB Sørensen OG Kaptein BL Andersen MS Stilling M

Objectives

Static radiostereometric analysis (RSA) using implanted markers is considered the most accurate system for the evaluation of prosthesis migration. By using CT bone models instead of markers, combined with a dynamic RSA system, a non-invasive measurement of joint movement is enabled. This method is more accurate than current 3D skin marker-based tracking systems. The purpose of this study was to evaluate the accuracy of the CT model method for measuring knee joint kinematics in static and dynamic RSA using the marker method as the benchmark.

Methods

Bone models were created from CT scans, and tantalum beads were implanted into the tibia and femur of eight human cadaver knees. Each specimen was secured in a fixture, static and dynamic stereoradiographs were recorded, and the bone models and marker models were fitted to the stereoradiographs.


Bone & Joint Research
Vol. 5, Issue 5 | Pages 169 - 174
1 May 2016
Wang Y Chu M Rong J Xing B Zhu L Zhao Y Zhuang X Jiang L

Objectives

Previous genome-wide association studies (GWAS) have reported significant association of the single nucleotide polymorphism (SNP) rs8044769 in the fat mass and obesity-associated gene (FTO) with osteoarthritis (OA) risk in European populations. However, these findings have not been confirmed in Chinese populations.

Methods

We systematically genotyped rs8044769 and evaluated the association between the genetic variants and OA risk in a case-controlled study including 196 OA cases and 442 controls in a northern Chinese population. Genotyping was performed using the Sequenom MassARRAY iPLEX platform.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 606 - 613
1 May 2006
Abu-Rajab RB Watson WS Walker B Roberts J Gallacher SJ Meek RMD

We compared peri-prosthetic bone mineral density between identical cemented and cementless LCS rotating platform total knee arthroplasties. Two matched cohorts had dual energy x-ray absorptiometry scans two years post-operatively using a modified validated densitometric analysis protocol, to assess peri-prosthetic bone mineral density. The knee that was not operated on was also scanned to enable the calculation of a relative bone mineral density difference. Oxford Knee and American Knee Society scores were comparable in the two cohorts.

Statistical analysis revealed no significant difference in absolute, or relative peri-prosthetic bone mineral density with respect to the method of fixation. However, the femoral peri-prosthetic bone mineral density and relative bone mineral density difference were significantly decreased, irrespective of the method of fixation, particularly in the anterior distal portion of the femur, with a mean reduction in relative bone mineral density difference of 27%.

There was no difference in clinical outcome between the cemented and cementless LCS total knee arthroplasty. However, both produce stress-shielding around the femoral implants. This leads us to question the use of more expensive cementless total knee components.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1503 - 1509
1 Nov 2014
Ritter MA Davis KE Small SR Merchun JG Farris A

The relationship between post-operative bone density and subsequent failure of total knee replacement (TKR) is not known. This retrospective study aimed to determine the relationship between bone density and failure, both overall and according to failure mechanism. All 54 aseptic failures occurring in 50 patients from 7760 consecutive primary cemented TKRs between 1983 and 2004 were matched with non-failing TKRs, and 47 failures in 44 patients involved tibial failures with the matching characteristics of age (65.1 for failed and 69.8 for non-failed), gender (70.2% female), diagnosis (93.6% OA), date of operation, bilaterality, pre-operative alignment (0.4 and 0.3 respectively), and body mass index (30.2 and 30.0 respectively). In each case, the density of bone beneath the tibial component was assessed at each follow-up interval using standardised, calibrated radiographs. Failing knees were compared with controls both overall and, as a subgroup analysis, by failure mechanism. Knees were compared with controls using univariable linear regression.

Significant and continuous elevation in tibial density was found in knees that eventually failed by medial collapse (p < 0.001) and progressive radiolucency (p < 0.001) compared with controls, particularly in the medial region of the tibia. Knees failing due to ligamentous instability demonstrated an initial decline in density (p = 0.0152) followed by a non-decreasing density over time (p = 0.034 for equivalence). Non-failing knees reported a decline in density similar to that reported previously using dual-energy x-ray absorptiometry (DEXA). Differences between failing and non-failing knees were observable as early as two months following surgery. This tool may be used to identify patients at risk of failure following TKR, but more validation work is needed.