Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 879 - 886
1 Jul 2007
Howie DW Wimhurst JA McGee MA Carbone TA Badaruddin BS

We investigated 219 revisions of total hip replacement (THR) in 211 patients using a collarless double-taper cemented femoral component. The mean age of the patients was 72 years (30 to 90). The 137 long and 82 standard length stems were analysed separately. The mean follow-up was six years (2 to 18), and no patient was lost to follow-up.

Survival of the long stems to re-revision for aseptic loosening at nine years was 98% (95% confidence interval (CI) 94 to 100), and for the standard stems was 93% (95% CI 85 to 100). At five years, one long stem was definitely loose radiologically and one standard stem was probably loose. Pre-operative femoral bone deficiency did not influence the results for the long stems, and corrective femoral osteotomy was avoided, as were significant subsidence, major stress shielding and persistent thigh pain.

Because of these reliable results, cemented long collarless double-taper femoral components are recommended for routine revision THR in older patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 581 - 585
1 May 2006
Oddy MJ Jones MJ Pendegrass CJ Pilling JR Wimhurst JA

In 20 patients undergoing hybrid total hip arthroplasty, the reproducibility and accuracy of templating using digital radiographs were assessed. Digital images were manipulated using either a ten-pence coin as a marker to scale for magnification, or two digital-line methods using computer software. On-screen images were templated with standard acetate templates and compared with templating performed on hard-copy digital prints.

The digital-line methods were the least reliable and accuracy of sizing compared with the inserted prostheses varied between −1.6% and +10.2%. The hard-copy radiographs showed better reproducibility than the ten-pence coin method, but were less accurate with 3.7% undersizing. The ten-pence coin method was the most accurate, with no significant differences for offset or acetabulum, and undersizing of only 0.9%.

On-screen templating of digital radiographs with standard acetate templates is accurate and reproducible if a radiopaque marker such as a ten-pence coin is included when the original radiograph is taken.