Advertisement for orthosearch.org.uk
Results 1 - 4 of 4
Results per page:
Bone & Joint Open
Vol. 3, Issue 1 | Pages 61 - 67
18 Jan 2022
van Lingen CP Ettema HB Bosker BH Verheyen CCPM

Aims

Large-diameter metal-on-metal (MoM) total hip arthroplasty (THA) has demonstrated unexpected high failure rates and pseudotumour formation. The purpose of this prospective cohort study is to report ten-year results in order to establish revision rate, prevalence of pseudotumour formation, and relation with whole blood cobalt levels.

Methods

All patients were recalled according to the guidelines of the Dutch Orthopaedic Association. They underwent clinical and radiographical assessments (radiograph and CT scan) of the hip prosthesis and whole blood cobalt ion measurements. Overall, 94 patients (95 hips) fulfilled our requirements for a minimum ten-year follow-up.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 616 - 622
1 May 2013
Horstmann WG Swierstra MJ Ohanis D Castelein RM Kollen BJ Verheyen CCPM

Autologous retransfusion and no-drainage are both blood-saving measures in total hip replacement (THR). A new combined intra- and post-operative autotransfusion filter system has been developed especially for primary THR, and we conducted a randomised controlled blinded study comparing this with no-drainage.

A total of 204 THR patients were randomised to autologous blood transfusion (ABT) (n = 102) or no-drainage (n = 102). In the ABT group, a mean of 488 ml (sd 252) of blood was retransfused. The mean lowest post-operative haemoglobin level during the hospital stay was higher in the autotransfusion group (10.6 g/dl (7.8 to 13.9) vs 10.2 g/dl (7.5 to 13.3); p = 0.01). The mean haemoglobin levels for the ABT and no-drainage groups were not significantly different on the first day (11.3 g/dl (7.8 to 13.9) vs 11.0 g/dl (8.1 to 13.4); p = 0.07), the second day (11.1 g/dl (8.2 to 13.8) vs 10.8 g/dl (7.5 to 13.3); p = 0.09) or the third day (10.8 g/dl (8.0 to 13.0) vs 10.6 g/dl (7.5 to 14.1); p = 0.15). The mean total peri-operative net blood loss was 1464 ml (sd 505) in the ABT group and 1654 ml (sd 553) in the no-drainage group (p = 0.01). Homologous blood transfusions were needed in four patients (3.9%) in the ABT group and nine (8.8%) in the no-drainage group (p = 0.15). No statistically significant difference in adverse events was found between the groups.

The use of a new intra- and post-operative autologous blood transfusion filter system results in less total blood loss and a smaller maximum decrease in haemoglobin levels than no-drainage following primary THR.

Cite this article: Bone Joint J 2013;95-B:616–22.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 755 - 761
1 Jun 2012
Bosker BH Ettema HB Boomsma MF Kollen BJ Maas M Verheyen CCPM

Peri-articular soft-tissue masses or ‘pseudotumours’ can occur after large-diameter metal-on-metal (MoM) resurfacing of the hip and conventional total hip replacement (THR). Our aim was to assess the incidence of pseudotumour formation and to identify risk factors for their formation in a prospective cohort study.

A total of 119 patients who underwent 120 MoM THRs with large-diameter femoral heads between January 2005 and November 2007 were included in the study. Outcome scores, serum metal ion levels, radiographs and CT scans were obtained. Patients with symptoms or an identified pseudotumour were offered MRI and an ultrasound-guided biopsy.

There were 108 patients (109 hips) eligible for evaluation by CT scan at a mean follow-up of 3.6 years (2.5 to 4.5); 42 patients (39%) were diagnosed with a pseudotumour. The hips of 13 patients (12%) were revised to a polyethylene acetabular component with small-diameter metal head. Patients with elevated serum metal ion levels had a four times increased risk of developing a pseudotumour.

This study shows a substantially higher incidence of pseudotumour formation and subsequent revisions in patients with MoM THRs than previously reported. Because most revision cases were identified only after an intensive screening protocol, we recommend close monitoring of patients with MoM THR.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1475 - 1479
1 Nov 2005
Goosen JHM Verheyen CCPM Tulp NJA

We investigated the rate of polyethylene wear of a cementless acetabular component at different periods of follow-up in order to test the hypothesis than an irrecoverable deformation process (creep) was followed by an initially low, but gradually increasing wear rate. We studied prospectively 93 uncemented total hip arthroplasties in 83 patients (mean age 50 years (22 to 63)) with a mean follow-up of 8.2 years (3 to 12). We measured the penetration of the femoral head from radiographs taken immediately after surgery at three, six and nine years, or at the latest follow-up.

The median wear rate was 0.17 mm per year in the first three years, a finding which we considered to be caused by creep. Thereafter, the rate of wear declined to 0.07 mm per year (four- to six-year period) and then increased to 0.17 mm per year (seven to nine years) and 0.27 mm per year (more than nine years), which we considered to be a reflection of genuine polyethylene wear. After the nine-year follow-up the wear rates were higher in patients with marked osteolysis. We found no relationship between the inclination angle of the acetabular component or femoral head orientation and the rate of wear. No acetabular component required revision.