Advertisement for orthosearch.org.uk
Results 1 - 20 of 31
Results per page:
Bone & Joint Research
Vol. 13, Issue 5 | Pages 226 - 236
9 May 2024
Jürgens-Lahnstein JH Petersen ET Rytter S Madsen F Søballe K Stilling M

Aims

Micromotion of the polyethylene (PE) inlay may contribute to backside PE wear in addition to articulate wear of total knee arthroplasty (TKA). Using radiostereometric analysis (RSA) with tantalum beads in the PE inlay, we evaluated PE micromotion and its relationship to PE wear.

Methods

A total of 23 patients with a mean age of 83 years (77 to 91), were available from a RSA study on cemented TKA with Maxim tibial components (Zimmer Biomet). PE inlay migration, PE wear, tibial component migration, and the anatomical knee axis were evaluated on weightbearing stereoradiographs. PE inlay wear was measured as the deepest penetration of the femoral component into the PE inlay.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1695 - 1701
1 Nov 2021
Currier JH Currier BH Abdel MP Berry DJ Titus AJ Van Citters DW

Aims

Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance.

Methods

An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1586 - 1594
1 Oct 2021
Sharma N Rehmatullah N Kuiper JH Gallacher P Barnett AJ

Aims

The Oswestry-Bristol Classification (OBC) is an MRI-specific assessment tool to grade trochlear dysplasia. The aim of this study is to validate clinically the OBC by demonstrating its use in selecting treatments that are safe and effective.

Methods

The OBC and the patellotrochlear index were used as part of the Oswestry Patellotrochlear Algorithm (OPTA) to guide the surgical treatment of patients with patellar instability. Patients were assigned to one of four treatment groups: medial patellofemoral ligament reconstruction (MPFLr); MPFLr + tibial tubercle distalization (TTD); trochleoplasty; or trochleoplasty + TTD. A prospective analysis of a longitudinal patellofemoral database was performed. Between 2012 and 2018, 202 patients (233 knees) with a mean age of 24.2 years (SD 8.1), with recurrent patellar instability were treated by two fellowship-trained consultant sports/knee surgeons at The Robert Jones and Agnes Hunt Orthopaedic Hospital. Clinical efficacy of each treatment group was assessed by Kujala, International Knee Documentation Committee (IKDC), and EuroQol five-dimension questionnaire (EQ-5D) scores at baseline, and up to 60 months postoperatively. Their safety was assessed by complication rate and requirement for further surgery. The pattern of clinical outcome over time was analyzed using mixed regression modelling.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 165 - 170
1 Jun 2021
Larson DJ Rosenberg JH Lawlor MA Garvin KL Hartman CW Lyden E Konigsberg BS

Aims

Stemmed tibial components are frequently used in revision total knee arthroplasty (TKA). The purpose of this study was to evaluate patient satisfaction, overall pain, and diaphyseal tibial pain in patients who underwent revision TKA with cemented or uncemented stemmed tibial components.

Methods

This is a retrospective cohort study involving 110 patients with revision TKA with cemented versus uncemented stemmed tibial components. Patients who underwent revision TKA with stemmed tibial components over a 15-year period at a single institution with at least two-year follow-up were assessed. Pain was evaluated through postal surveys. There were 63 patients with cemented tibial stems and 47 with uncemented stems. Radiographs and Knee Society Scores were used to evaluate for objective findings associated with pain or patient dissatisfaction. Postal surveys were analyzed using Fisher’s exact test and the independent-samples t-test. Logistic regression was used to adjust for age, sex, and preoperative bone loss.


The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 102 - 107
1 Jan 2020
Sharma N Brown A Bouras T Kuiper JH Eldridge J Barnett A

Aims

Trochlear dysplasia is a significant risk factor for patellofemoral instability. The Dejour classification is currently considered the standard for classifying trochlear dysplasia, but numerous studies have reported poor reliability on both plain radiography and MRI. The severity of trochlear dysplasia is important to establish in order to guide surgical management. We have developed an MRI-specific classification system to assess the severity of trochlear dysplasia, the Oswestry-Bristol Classification (OBC). This is a four-part classification system comprising normal, mild, moderate, and severe to represent a normal, shallow, flat, and convex trochlear, respectively. The purpose of this study was to assess the inter- and intraobserver reliability of the OBC and compare it with that of the Dejour classification.

Methods

Four observers (two senior and two junior orthopaedic surgeons) independently assessed 32 CT and axial MRI scans for trochlear dysplasia and classified each according to the OBC and the Dejour classification systems. Assessments were repeated following a four-week interval. The inter- and intraobserver agreement was determined by using Fleiss’ generalization of Cohen’s kappa statistic and S-statistic nominal and linear weights.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1058 - 1062
1 Sep 2019
van Kuijk KSR Reijman M Bierma-Zeinstra SMA Waarsing JH Meuffels DE

Aims

Little is known about the risk factors that predispose to a rupture of the posterior cruciate ligament (PCL). Identifying risk factors is the first step in trying to prevent a rupture of the PCL from occurring. The morphology of the knee in patients who rupture their PCL may differ from that of control patients. The purpose of this study was to identify any variations in bone morphology that are related to a PCL.

Patients and Methods

We compared the anteroposterior (AP), lateral, and Rosenberg view radiographs of 94 patients with a ruptured PCL to a control group of 168 patients matched by age, sex, and body mass index (BMI), but with an intact PCL after a knee injury. Statistical shape modelling software was used to assess the shape of the knee and determine any difference in anatomical landmarks.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 22 - 27
1 Jul 2019
Kalbian IL Tan TL Rondon AJ Bonaddio VA Klement MR Foltz C Lonner JH

Aims

Unicompartmental knee arthroplasty (UKA) provides improved early functional outcomes and less postoperative morbidity and pain compared with total knee arthroplasty (TKA). Opioid prescribing has increased in the last two decades, and recently states in the USA have developed online Prescription Drug Monitoring Programs to prevent overprescribing of controlled substances. This study evaluates differences in opioid requirements between patients undergoing TKA and UKA.

Patients and Methods

We retrospectively reviewed 676 consecutive TKAs and 241 UKAs. Opioid prescriptions in morphine milligram equivalents (MMEs), sedatives, benzodiazepines, and stimulants were collected from State Controlled Substance Monitoring websites six months before and nine months after the initial procedures. Bivariate and multivariate analysis were performed for patients who had a second prescription and continued use.


Bone & Joint Research
Vol. 7, Issue 9 | Pages 541 - 547
1 Sep 2018
Eijkenboom JFA Waarsing JH Oei EHG Bierma-Zeinstra SMA van Middelkoop M

Objectives

It has been hypothesized that patellofemoral pain, a common knee condition in adolescents and young adults, may be a precursor of degenerative joint changes and may ultimately lead to patellofemoral osteoarthritis. Since both conditions share several mechanical disease characteristics, such as altered contact area between the femur and patella and increased joint stress, we investigated whether these conditions share similar and different shape characteristics of the patella compared with normal controls.

Methods

This cross-sectional study compared three different study populations: 32 patellofemoral pain subjects (mean age, 32 years (22 to 45); 72% female); 56 isolated radiological patellofemoral osteoarthritis subjects (mean age, 54 years (44 to 58); 89% female); and 80 healthy control subjects (mean age, 52 years (44 to 58); 74% female). Measurements included questionnaires, and lateral and skyline radiographs of the knee. Two separate 30-point 2D statistical shape models of the patella were created from the lateral and skyline radiographs. A general linear model was used to test for differences in standardized shape modes (a specific shape variant of the patella) between patellofemoral osteoarthritis, patellofemoral pain, and controls, using Bonferroni correction and adjustment for body mass index and gender.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1485 - 1490
1 Nov 2014
Kim CW Seo SS Kim JH Roh SM Lee CR

The aim of this study was to find anatomical landmarks for rotational alignment of the tibial component in total knee replacement (TKR) in a CT-based study. Pre-operative CT scanning was performed on 94 South Korean patients (nine men, 85 women, 188 knees) with osteoarthritis of the knee joint prior to TKR. The tibial anteroposterior (AP) axis was defined as a line perpendicular to the femoral surgical transepicondylar axis and passing through the centre of the posterior cruciate ligament (PCL). The angles between the defined tibial AP axis and anatomical landmarks at various levels of the tibia were measured. The mean values of the angles between the defined tibial AP axis and the line connecting the anterior border of the proximal third of the tibia to the centre of the PCL was -0.2° (-17 to 14.1, sd 4.1). This was very close to the defined tibial axis, and remained so regardless of lower limb alignment and the degree of tibial bowing. Therefore, AP axis defined as described, is a reliable anatomical landmark for rotational alignment of tibial components.

Cite this article: Bone Joint J 2014; 96-B:1485–90.


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 914 - 922
1 Jul 2014
Lee SY Bae JH Kim JG Jang KM Shon WY Kim KW Lim HC

The aim of this study was to evaluate the risk factors for dislocation of the bearing after a mobile-bearing Oxford medial unicompartmental knee replacement (UKR) and to test the hypothesis that surgical factors, as measured from post-operative radiographs, are associated with its dislocation

From a total of 480 UKRs performed between 2001 and 2012, in 391 patients with a mean age of 66.5 years (45 to 82) (316 female, 75 male), we identified 17 UKRs where bearing dislocation occurred. The post-operative radiological measurements of the 17 UKRs and 51 matched controls were analysed using conditional logistic regression analysis. The post-operative radiological measurements included post-operative change in limb alignment, the position of the femoral and tibial components, the resection depth of the proximal tibia, and the femoral component-posterior condyle classification.

We concluded that a post-operative decrease in the posterior tibial slope relative to the pre-operative value was the only significant determinant of dislocation of the bearing after medial Oxford UKR (odds ratio 1.881; 95% confidence interval 1.272 to 2.779). A post-operative posterior tibial slope < 8.45° and a difference between the pre-operative and post-operative posterior tibial slope of > 2.19° may increase the risk of dislocation.

Cite this article: Bone Joint J 2014; 96-B:914–22.


The Bone & Joint Journal
Vol. 96-B, Issue 6 | Pages 743 - 751
1 Jun 2014
Shin YS Ro KH Jeon JH Lee DH

We used immediate post-operative in vivo three-dimensional computed tomography to compare graft bending angles and femoral tunnel lengths in 155 patients who had undergone single-bundle reconstruction of the anterior cruciate ligament using the transtibial (n = 37), anteromedial portal (n = 72) and outside-in (n = 46) techniques.

The bending angles in the sagittal and axial planes were significantly greater but the coronal-bending angle was significantly less in the transtibial group than in the anteromedial portal and outside-in groups (p < 0.001 each). The mean length of the femoral tunnel in all three planes was significantly greater in the transtibial group than the anteromedial portal and outside-in groups (p < 0.001 each), but all mean tunnel lengths in the three groups exceeded 30 mm. The only significant difference was the coronal graft- bending angle in the anteromedial portal and outside-in groups (23.5° vs 29.8°, p = 0.012).

Compared with the transtibial technique, the anteromedial portal and outside-in techniques may reduce the graft-bending stress at the opening of the femoral tunnel. Despite the femoral tunnel length being shorter in the anteromedial portal and outside-in techniques than in the transtibial technique, a femoral tunnel length of more than 30 mm in the anteromedial portal and outside-in techniques may be sufficient for the graft to heal.

Cite this article: Bone Joint J 2014;96-B:743–51.


The Bone & Joint Journal
Vol. 96-B, Issue 6 | Pages 737 - 742
1 Jun 2014
Eggerding V van Kuijk KSR van Meer BL Bierma-Zeinstra SMA van Arkel ERA Reijman M Waarsing JH Meuffels DE

We have investigated whether shape of the knee can predict the clinical outcome of patients after an anterior cruciate ligament rupture. We used statistical shape modelling to measure the shape of the knee joint of 182 prospectively followed patients on lateral and Rosenberg view radiographs of the knee after a rupture of the anterior cruciate ligament. Subsequently, we associated knee shape with the International Knee Documentation Committee subjective score at two years follow-up. The mean age of patients was 31 years (21 to 51), the majority were male (n = 121) and treated operatively (n = 135). We found two modes (shape variations) that were significantly associated with the subjective score at two years: one for the operatively treated group (p = 0.002) and one for the non-operatively treated group (p = 0.003). Operatively treated patients who had higher subjective scores had a smaller intercondylar notch and a smaller width of the intercondylar eminence. Non-operatively treated patients who scored higher on the subjective score had a more pyramidal intercondylar notch as opposed to one that was more dome-shaped. We conclude that the shape of the femoral notch and the intercondylar eminence is predictive of clinical outcome two years after a rupture of the anterior cruciate ligament.

Cite this article: Bone Joint J 2014;96-B:737–42.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1354 - 1358
1 Oct 2013
Singh G Tan JH Sng BY Awiszus F Lohmann CH Nathan SS

The optimal management of the tibial slope in achieving a high flexion angle in posterior-stabilised (PS) total knee replacement (TKR) is not well understood, and most studies evaluating the posterior tibial slope have been conducted on cruciate-retaining TKRs. We analysed pre- and post-operative tibial slope differences, pre- and post-operative coronal knee alignment and post-operative maximum flexion angle in 167 patients undergoing 209 TKRs. The mean pre-operative posterior tibial slope was 8.6° (1.3° to 17°) and post-operatively it was 8.0° (0.1° to 16.7°). Multiple linear regression analysis showed that the absolute difference between pre- and post-operative tibial slope (p < 0.001), post-operative coronal alignment (p = 0.02) and pre-operative range of movement (p < 0.001) predicted post-operative flexion. The variance of change in tibial slope became larger as the post-operative maximum flexion angle decreased. The odds ratio of having a post-operative flexion angle < 100° was 17.6 if the slope change was > 2°. Our data suggest that recreation of the anatomical tibial slope appears to improve maximum flexion after posterior-stabilised TKR, provided coronal alignment has been restored.

Cite this article: Bone Joint J 2013;95-B:1354–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 908 - 913
1 Jul 2012
Seo JG Moon YW Park SH Lee JH Kang HM Kim SM

Peri-prosthetic patellar fracture following resurfacing as part of total knee replacement (TKR) is an infrequent yet challenging complication. This case-control study was performed to identify clinical, radiological and surgical factors that increase the risk of developing a spontaneous patellar fracture after TKR. Patellar fractures were identified in 74 patients (88 knees) from a series of 7866 consecutive TKRs conducted between 1998 and 2009. After excluding those with a previous history of extensor mechanism realignment or a clear traumatic event, a metal-backed patella, any uncemented component or subsequent infection, the remaining 64 fractures were compared with a matched group of TKRs with an excellent outcome defined by the Knee Society score. The mean age of patients with a fracture was 70 years (51 to 81) at the time of TKR. Patellar fractures were detected at a mean of 13.4 months (2 to 84) after surgery. The incidence of patellar fracture was found to be strongly associated with the number of previous knee operations, greater pre-operative mechanical malalignment, smaller post-operative patellar tendon length, thinner post-resection patellar thickness, and a lower post-operative Insall-Salvati ratio.

An understanding of the risk factors associated with spontaneous patellar fracture following TKR provides a valuable insight into prevention of this challenging complication.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 914 - 918
1 Jul 2012
Jameson SS Baker PN Charman SC Deehan DJ Reed MR Gregg PJ Van der Meulen JH

We compared thromboembolic events, major haemorrhage and death after knee replacement in patients receiving either aspirin or low-molecular-weight heparin (LMWH). Data from the National Joint Registry for England and Wales were linked to an administrative database of hospital admissions in the English National Health Service. A total of 156 798 patients between April 2003 and September 2008 were included and followed for 90 days. Multivariable risk modelling was used to estimate odds ratios adjusted for baseline risk factors (AOR). An AOR < 1 indicates that risk rates are lower with LMWH than with aspirin. In all, 36 159 patients (23.1%) were prescribed aspirin and 120 639 patients (76.9%) were prescribed LMWH. We found no statistically significant differences between the aspirin and LMWH groups in the rate of pulmonary embolism (0.49% vs 0.45%, AOR 0.88 (95% confidence interval (CI) 0.74 to 1.05); p = 0.16), 90-day mortality (0.39% vs 0.45%, AOR 1.13 (95% CI 0.94 to 1.37); p = 0.19) or major haemorrhage (0.37% vs 0.39%, AOR 1.01 (95% CI 0.83 to 1.22); p = 0.94). There was a significantly greater likelihood of needing to return to theatre in the aspirin group (0.26% vs 0.19%, AOR 0.73 (95% CI 0.58 to 0.94); p = 0.01). Between patients receiving LMWH or aspirin there was only a small difference in the risk of pulmonary embolism, 90-day mortality and major haemorrhage.

These results should be considered when the existing guidelines for thromboprophylaxis after knee replacement are reviewed.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 337 - 344
1 Mar 2011
Yoo JH Chang CB Kang YG Kim SJ Seong SC Kim TK

We aimed to document the pre-operative expectations in Korean patients undergoing total knee replacement using an established survey form and to determine whether expectations were influenced by sociodemographic factors or pre-operative functional status. Expectations regarding 17 items in the Knee Replacement Expectation Survey form were investigated in 454 patients scheduled for total knee replacement. The levels and distribution patterns of summated expectation and of five expectation categories (relief from pain, baseline activity, high flexion activity, social activity and psychological well-being) constructed from the 17 items were assessed. Univariate analyses and multivariate logistic regression were performed to examine the associations of expectations with the sociodemographic factors and the functional status.

The top three expectations were relief from pain, restoration of walking ability, and psychological well-being. Of the five expectation categories, relief from pain was ranked the highest, followed by psychological well-being, restoration of baseline activity, ability to perform high flexion activities and ability to participate in social activities. An age of < 65 years, being employed, a high Western Ontario and McMaster Universities osteoarthritis index function score and a low Short-form 36 social score were found to be significantly associated with higher overall expectations.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 180 - 184
1 Feb 2009
Lim HC Bae JH Hur CR Oh JK Han SH

We retrospectively evaluated eight patients who underwent arthrodesis of the knee using cannulated screws. There were six women and two men, with a mean age of 53 years. The indications for arthrodesis were failed total knee arthroplasty, septic arthritis, tuberculosis, and recurrent persistent infection. Solid union was achieved in all patients at a mean of 6.1 months. One patient required autogenous bone graft for delayed union. One suffered skin necrosis which was treated with skin grafting. The mean limb-length discrepancy was 3.1 cm. On a visual analogue scale, the mean pain score improved from 7.9 to 3.3. According to the Knee Injury and Osteoarthritis Outcome score quality of life items, the mean score improved from 38.3 pre-operatively to 76.6 at follow-up. Cannulated screws provide a high rate of union in arthrodesis of the knee with minimal complications, patient convenience, and a simple surgical technique.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 879 - 884
1 Jul 2008
Porteous AJ Hassaballa MA Newman JH

We identified 148 patients who had undergone a revision total knee replacement using a single implant system between 1990 and 2000. Of these 18 patients had died, six had developed a peri-prosthetic fracture and ten had incomplete records or radiographs. This left 114 with prospectively-collected radiographs and Bristol knee scores available for study. The height of the joint line before and after revision total knee replacement was measured and classified as either restored to within 5 mm of the pre-operative height or elevated if it was positioned more than 5 mm above the pre-operative height. The joint line was elevated in 41 knees (36%) and restored in 73 (64%).

Revision surgery significantly improved the mean Bristol knee score from 41.1 (sd 15.9) pre-operatively to 80.5 (sd 15) post-operatively (p < 0.001). At one year post-operatively both the total Bristol knee score and its functional component were significantly better in the restored group than in the elevated group (p < 0.01). Overall, revision from a unicondylar knee replacement required less use of bone graft, fewer component augments, restored the joint line more often and gave a significantly better total Bristol knee score (p < 0.02) and functional score (p < 0.01) than revision from total knee replacement.

Our findings show that restoration of the joint line at revision total knee replacement gives a significantly better result than leaving it unrestored by more than 5 mm. We recommend the greater use of distal femoral augments to help to achieve this goal.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 31 - 36
1 Jan 2008
Yoo JH Kang YG Chang CB Seong SC Kim TK

We examined the placement of the stem in relation to the medial tibial cortex when using total knee replacements (TKRs) with medially-offset tibial stems in Korean patients. Measurements were performed on the pre- and post-operative radiographs of 246 osteoarthritic knees replaced between January 2005 and May 2006 using the Genesis II or E-motion TKR with a medially-offset stem. Pre-operatively, we measured the distance between the mechanical axis and that of the tibial shaft and post-operatively, that between the midline of the tibial stem and the axis of the shaft.

Knees were identified in which there was radiological contact between the tip of the stem and the medial tibial cortex. The mechanical axis was located medial to the axis of the shaft in 203 knees (82.5%). Post-operatively, the midline of the tibial stem was located medial to the tibial shaft axis in 196 knees (79.7%). In 16 knees (6.5%) there was radiological contact between the tibial stem or cement mantle and the medial tibial cortex.

Our study has shown that the medially-offset stem in the tibial component may not be a good option for knees undergoing replacement for advanced osteoarthritis in some Korean patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 893 - 900
1 Jul 2007
Baker PN van der Meulen JH Lewsey J Gregg PJ

A postal questionnaire was sent to 10 000 patients more than one year after their total knee replacement (TKR). They were assessed using the Oxford knee score and were asked whether they were satisfied, unsure or unsatisfied with their TKR. The response rate was 87.4% (8231 of 9417 eligible questionnaires) and a total of 81.8% (6625 of 8095) of patients were satisfied. Multivariable regression modelling showed that patients with higher scores relating to the pain and function elements of the Oxford knee score had a lower level of satisfaction (p < 0.001), and that ongoing pain was a stronger predictor of this. Female gender and a primary diagnosis of osteoarthritis were found to be predictors of lower levels of patient satisfaction. Differences in the rate of satisfaction were also observed in relation to age, the American Society of Anesthesiologists grade and the type of prosthesis.

This study has provided data on the Oxford knee score and the expected levels of satisfaction at one year after TKR. The results should act as a benchmark of practice in the United Kingdom and provide a baseline for peer comparison between institutions.