Advertisement for orthosearch.org.uk
Results 1 - 20 of 71
Results per page:
The Bone & Joint Journal
Vol. 99-B, Issue 11 | Pages 1467 - 1476
1 Nov 2017
van Hamersveld KT Marang-van de Mheen PJ Tsonaka R Valstar† ER Toksvig-Larsen S

Aims

The optimal method of tibial component fixation remains uncertain in total knee arthroplasty (TKA). Hydroxyapatite coatings have been applied to improve bone ingrowth in uncemented designs, but may only coat the directly accessible surface. As peri-apatite (PA) is solution deposited, this may increase the coverage of the implant surface and thereby fixation. We assessed the tibial component fixation of uncemented PA-coated TKAs versus cemented TKAs.

Patients and Methods

Patients were randomised to PA-coated or cemented TKAs. In 60 patients (30 in each group), radiostereometric analysis of tibial component migration was evaluated as the primary outcome at baseline, three months post-operatively and at one, two and five years. A linear mixed-effects model was used to analyse the repeated measurements.


Bone & Joint Open
Vol. 3, Issue 3 | Pages 173 - 181
1 Mar 2022
Sobol KR Fram BR Strony JT Brown SA

Aims. Endoprosthetic reconstruction with a distal femoral arthroplasty (DFA) can be used to treat distal femoral bone loss from oncological and non-oncological causes. This study reports the short-term implant survivorship, complications, and risk factors for patients who underwent DFA for non-neoplastic indications. Methods. We performed a retrospective review of 75 patients from a single institution who underwent DFA for non-neoplastic indications, including aseptic loosening or mechanical failure of a previous prosthesis (n = 25), periprosthetic joint infection (PJI) (n = 23), and native or periprosthetic distal femur fracture or nonunion (n = 27). Patients with less than 24 months’ follow-up were excluded. We collected patient demographic data, complications, and reoperations. Reoperation for implant failure was used to calculate implant survivorship. Results. Overall one- and five-year implant survivorship was 87% and 76%, respectively. By indication for DFA, mechanical failure had one- and five-year implant survivorship of 92% and 68%, PJI of 91% and 72%, and distal femur fracture/nonunion of 78% and 70% (p = 0.618). A total of 37 patients (49%) experienced complications and 27 patients (36%) required one or more reoperation. PJI (n = 16, 21%), aseptic loosening (n = 9, 12%), and wound complications (n = 8, 11%) were the most common complications. Component revision (n = 10, 13.3%) and single-stage exchange for PJI (n = 9, 12.0 %) were the most common reoperations. Only younger age was significantly associated with increased complications (mean 67 years (SD 9.1)) with complication vs 71 years (SD 9.9) without complication; p = 0.048). Conclusion. DFA is a viable option for distal femoral bone loss from a range of non-oncological causes, demonstrating acceptable short-term survivorship but with high overall complication rates. Cite this article: Bone Jt Open 2022;3(3):173–181


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 1028 - 1036
1 Aug 2017
Chawla H Nwachukwu BU van der List JP Eggman AA Pearle AD Ghomrawi HM

Aims. Patellofemoral arthroplasty (PFA) has experienced significant improvements in implant survivorship with second generation designs. This has renewed interest in PFA as an alternative to total knee arthroplasty (TKA) for younger active patients with isolated patellofemoral osteoarthritis (PF OA). We analysed the cost-effectiveness of PFA versus TKA for the management of isolated PF OA in the United States-based population. Patients and Methods. We used a Markov transition state model to compare cost-effectiveness between PFA and TKA. Simulated patients were aged 60 (base case) and 50 years. Lifetime costs (2015 United States dollars), quality-adjusted life year (QALY) gains and incremental cost-effectiveness ratio (ICER) were calculated from a healthcare payer perspective. Annual rates of revision were derived from the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man. Deterministic and probabilistic sensitivity analysis was performed for all parameters against a $50 000/QALY willingness to pay. . Results. PFA was more expensive ($49 811 versus $46 632) but more effective (14.3 QALYs versus 13.3 QALYs) over a lifetime horizon. The ICER associated with the additional effectiveness of PFA was $3097. The model was mainly sensitive to utility values, with PFA remaining cost-effective when its utility exceeded that of TKA by at least 1.0%. PFA provided incremental benefits at no increased cost when annual rates of revision decreased by 24.5%. . Conclusions. Recent improvements in rates of implant of survival have made PFA an economically beneficial joint-preserving procedure in younger patients, delaying TKA until implant failure or tibiofemoral OA progression. The present study quantified the minimum required marginal benefit for PFA to be cost-effective compared with TKA and identified survivorship targets for PFA to become both less expensive and more effective. These benchmarks might be used to assess clinical outcomes of PFA from an economic standpoint within the United States healthcare system. Cite this article: Bone Joint J 2017;99-B:1028–36


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1331 - 1347
1 Nov 2019
Jameson SS Asaad A Diament M Kasim A Bigirumurame T Baker P Mason J Partington P Reed M

Aims. Antibiotic-loaded bone cements (ALBCs) may offer early protection against the formation of bacterial biofilm after joint arthroplasty. Use in hip arthroplasty is widely accepted, but there is a lack of evidence in total knee arthroplasty (TKA). The objective of this study was to evaluate the use of ALBC in a large population of TKA patients. Materials and Methods. Data from the National Joint Registry (NJR) of England and Wales were obtained for all primary cemented TKAs between March 2003 and July 2016. Patient, implant, and surgical variables were analyzed. Cox proportional hazards models were used to assess the influence of ALBC on risk of revision. Body mass index (BMI) data were available in a subset of patients. Results. Of 731 214 TKAs, 15 295 (2.1%) were implanted with plain cement and 715 919 (97.9%) with ALBC. There were 13 391 revisions; 2391 were performed for infection. After adjusting for other variables, ALBC had a significantly lower risk of revision for any cause (hazard ratio (HR) 0.85, 95% confidence interval (CI) 0.77 to 0.93; p < 0.001). ALBC was associated with a lower risk of revision for all aseptic causes (HR 0.85, 95% CI 0.77 to 0.95; p < 0.001) and revisions for infection (HR 0.84, 95% CI 0.67 to 1.01; p = 0.06). The results were similar when BMI was added into the model, and in a subanalysis where surgeons using only ALBC over the entire study period were excluded. Prosthesis survival at ten years for TKAs implanted with ALBC was 96.3% (95% CI 96.3 to 96.4) compared with 95.5% (95% CI 95.0 to 95.9) in those implanted with plain cement. On a population level, where 100 000 TKAs are performed annually, this difference represents 870 fewer revisions at ten years in the ALBC group. Conclusion. After adjusting for a range of variables, ALBC was associated with a significantly lower risk of revision in this registry-based study of an entire nation of primary cemented knee arthroplasties. Using ALBC does not appear to increase midterm implant failure rates. Cite this article: Bone Joint J 2019;101-B:1331–1347


Bone & Joint Open
Vol. 5, Issue 4 | Pages 277 - 285
8 Apr 2024
Khetan V Baxter I Hampton M Spencer A Anderson A

Aims

The mean age of patients undergoing total knee arthroplasty (TKA) has reduced with time. Younger patients have increased expectations following TKA. Aseptic loosening of the tibial component is the most common cause of failure of TKA in the UK. Interest in cementless TKA has re-emerged due to its encouraging results in the younger patient population. We review a large series of tantalum trabecular metal cementless implants in patients who are at the highest risk of revision surgery.

Methods

A total of 454 consecutive patients who underwent cementless TKA between August 2004 and December 2021 were reviewed. The mean follow-up was ten years. Plain radiographs were analyzed for radiolucent lines. Patients who underwent revision TKA were recorded, and the cause for revision was determined. Data from the National Joint Registry for England, Wales, Northern Island, the Isle of Man and the States of Guernsey (NJR) were compared with our series.


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 450 - 454
1 Apr 2018
Chalmers BP Mehrotra KG Sierra RJ Pagnano MW Taunton MJ Abdel MP

Aims. Primary (or spontaneous) and secondary osteonecrosis of the knee can lead to severe joint degeneration, for which either total or unicompartmental arthroplasty may be considered. However, there are limited studies analyzing outcomes of unicompartmental knee arthroplasties (UKAs) for osteonecrosis involving an isolated compartment of the knee. The aims of this study were to analyze outcomes of UKAs for osteonecrosis with specific focus on 1) survivorship free of any revision or reoperation, 2) risk factors for failure, 3) clinical outcomes, and 4) complications. Patients and Methods. A total of 45 patients underwent 46 UKAs for knee osteonecrosis between 2002 and 2014 at our institution (The Mayo Clinic, Rochester, Minnesota). Twenty patients (44%) were female; the mean age of the patients was 66 years, and mean body mass index (BMI) was 31 kg/m. 2. Of the 46 UKAs, 44 (96%) were medial UKAs, and 35 (76%) were fixed-bearing design. Mean mechanical axis postoperatively was 1.5° varus (0° to 5° varus); 41 UKAs (89%) were performed for primary osteonecrosis. Mean follow-up was five years (2 to 12). Results. Survivorship free of any revision in the cohort was 89% (95% CI 77 to 99) and 76% (95% CI 53 to 99) at five and ten years, respectively. In patients undergoing UKA for primary osteonecrosis survivorship free of any revision was 93% (95% CI 83 to 100)at both five and ten years. Secondary osteonecrosis was a significant risk factor for poorer survivorship free of any revision or reoperation (hazard ratio 7.7, p = 0.03). Three medial UKAs (6.5%) were converted to total knee arthroplasties (TKAs): two for lateral compartment degeneration and one for development of lateral osteonecrosis. No implants were revised for loosening, fracture, or wear. Knee Society scores improved from a mean of 60 (44 to 72) preoperatively to a mean of 94 postoperatively (82 to 100) (p < 0.001). There were no surgical complications. Conclusion. When done for primary osteonecrosis of the knee, UKA resulted in reliable clinical improvement, minimal complications, and durable estimated implant survivorship free of revision at ten years. UKA done for secondary osteonecrosis was substantially less durable at mid-term follow-up. Progression of knee degeneration, rather than implant failure or loosening, was most common indication for conversion to TKA. Cite this article: Bone Joint J 2018;100-B:450–4


Bone & Joint Open
Vol. 4, Issue 5 | Pages 393 - 398
25 May 2023
Roof MA Lygrisse K Shichman I Marwin SE Meftah M Schwarzkopf R

Aims

Revision total knee arthroplasty (rTKA) is a technically challenging and costly procedure. It is well-documented that primary TKA (pTKA) have better survivorship than rTKA; however, we were unable to identify any studies explicitly investigating previous rTKA as a risk factor for failure following rTKA. The purpose of this study is to compare the outcomes following rTKA between patients undergoing index rTKA and those who had been previously revised.

Methods

This retrospective, observational study reviewed patients who underwent unilateral, aseptic rTKA at an academic orthopaedic speciality hospital between June 2011 and April 2020 with > one-year of follow-up. Patients were dichotomized based on whether this was their first revision procedure or not. Patient demographics, surgical factors, postoperative outcomes, and re-revision rates were compared between the groups.


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 269 - 276
1 Mar 2023
Tay ML Monk AP Frampton CM Hooper GJ Young SW

Aims

Unicompartmental knee arthroplasty (UKA) has higher revision rates than total knee arthroplasty (TKA). As revision of UKA may be less technically demanding than revision TKA, UKA patients with poor functional outcomes may be more likely to be offered revision than TKA patients with similar outcomes. The aim of this study was to compare clinical thresholds for revisions between TKA and UKA using revision incidence and patient-reported outcomes, in a large, matched cohort at early, mid-, and late-term follow-up.

Methods

Analyses were performed on propensity score-matched patient cohorts of TKAs and UKAs (2:1) registered in the New Zealand Joint Registry between 1 January 1999 and 31 December 2019 with an Oxford Knee Score (OKS) response at six months (n, TKA: 16,774; UKA: 8,387), five years (TKA: 6,718; UKA: 3,359), or ten years (TKA: 3,486; UKA: 1,743). Associations between OKS and revision within two years following the score were examined. Thresholds were compared using receiver operating characteristic analysis. Reasons for aseptic revision were compared using cumulative incidence with competing risk.


The Bone & Joint Journal
Vol. 105-B, Issue 12 | Pages 1279 - 1285
1 Dec 2023
Baker JF Nadar AC Jouflas AC Smith LS Sachdeva S Yakkanti MR Malkani AL

Aims

The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design.

Methods

This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 437 - 441
1 Apr 2016
Middleton S Toms A

We explored the literature surrounding whether allergy and hypersensitivity has a clinical basis for implant selection in total knee arthroplasty (TKA). In error, the terms hypersensitivity and allergy are often used synonymously. Although a relationship is present, we could not find any evidence of implant failure due to allergy. There is however increasing basic science that suggests a link between loosening and metal ion production. This is not an allergic response but is a potential problem. With a lack of evidence logically there can be no justification to use ‘hypoallergenic’ implants in patients who have pre-existing skin sensitivity to the metals used in TKA. Cite this article: Bone Joint J 2016;98-B:437–41


Bone & Joint Open
Vol. 5, Issue 10 | Pages 937 - 943
22 Oct 2024
Gregor RH Hooper GJ Frampton C

Aims

The aim of this study was to determine whether obesity had a detrimental effect on the long-term performance and survival of medial unicompartmental knee arthroplasties (UKAs).

Methods

This study reviewed prospectively collected functional outcome scores and revision rates of all medial UKA patients with recorded BMI performed in Christchurch, New Zealand, from January 2011 to September 2021. Patient-reported outcome measures (PROMs) were the primary outcome of this study, with all-cause revision rate analyzed as a secondary outcome. PROMs were taken preoperatively, at six months, one year, five years, and ten years postoperatively. There were 873 patients who had functional scores recorded at five years and 164 patients had scores recorded at ten years. Further sub-group analysis was performed based on the patient’s BMI. Revision data were available through the New Zealand Joint Registry for 2,323 UKAs performed during this time period.


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 468 - 474
1 May 2024
d'Amato M Flevas DA Salari P Bornes TD Brenneis M Boettner F Sculco PK Baldini A

Aims

Obtaining solid implant fixation is crucial in revision total knee arthroplasty (rTKA) to avoid aseptic loosening, a major reason for re-revision. This study aims to validate a novel grading system that quantifies implant fixation across three anatomical zones (epiphysis, metaphysis, diaphysis).

Methods

Based on pre-, intra-, and postoperative assessments, the novel grading system allocates a quantitative score (0, 0.5, or 1 point) for the quality of fixation achieved in each anatomical zone. The criteria used by the algorithm to assign the score include the bone quality, the size of the bone defect, and the type of fixation used. A consecutive cohort of 245 patients undergoing rTKA from 2012 to 2018 were evaluated using the current novel scoring system and followed prospectively. In addition, 100 first-time revision cases were assessed radiologically from the original cohort and graded by three observers to evaluate the intra- and inter-rater reliability of the novel radiological grading system.


Bone & Joint Open
Vol. 4, Issue 10 | Pages 808 - 816
24 Oct 2023
Scott CEH Snowden GT Cawley W Bell KR MacDonald DJ Macpherson GJ Yapp LZ Clement ND

Aims

This prospective study reports longitudinal, within-patient, patient-reported outcome measures (PROMs) over a 15-year period following cemented single radius total knee arthroplasty (TKA). Secondary aims included reporting PROMs trajectory, 15-year implant survival, and patient attrition from follow-up.

Methods

From 2006 to 2007, 462 consecutive cemented cruciate-retaining Triathlon TKAs were implanted in 426 patients (mean age 69 years (21 to 89); 290 (62.7%) female). PROMs (12-item Short Form Survey (SF-12), Oxford Knee Score (OKS), and satisfaction) were assessed preoperatively and at one, five, ten, and 15 years. Kaplan-Meier survival and univariate analysis were performed.


The Bone & Joint Journal
Vol. 106-B, Issue 11 | Pages 1240 - 1248
1 Nov 2024
Smolle MA Keintzel M Staats K Böhler C Windhager R Koutp A Leithner A Donner S Reiner T Renkawitz T Sava M Hirschmann MT Sadoghi P

Aims

This multicentre retrospective observational study’s aims were to investigate whether there are differences in the occurrence of radiolucent lines (RLLs) following total knee arthroplasty (TKA) between the conventional Attune baseplate and its successor, the novel Attune S+, independent from other potentially influencing factors; and whether tibial baseplate design and presence of RLLs are associated with differing risk of revision.

Methods

A total of 780 patients (39% male; median age 70.7 years (IQR 62.0 to 77.2)) underwent cemented TKA using the Attune Knee System) at five centres, and with the latest radiograph available for the evaluation of RLL at between six and 36 months from surgery. Univariate and multivariate logistic regression models were performed to assess associations between patient and implant-associated factors on the presence of tibial and femoral RLLs. Differences in revision risk depending on RLLs and tibial baseplate design were investigated with the log-rank test.


The Bone & Joint Journal
Vol. 106-B, Issue 8 | Pages 808 - 816
1 Aug 2024
Hall AJ Cullinan R Alozie G Chopra S Greig L Clarke J Riches PE Walmsley P Ohly NE Holloway N

Aims

Total knee arthroplasty (TKA) with a highly congruent condylar-stabilized (CS) articulation may be advantageous due to increased stability versus cruciate-retaining (CR) designs, while mitigating the limitations of a posterior-stabilized construct. The aim was to assess ten-year implant survival and functional outcomes of a cemented single-radius TKA with a CS insert, performed without posterior cruciate ligament sacrifice.

Methods

This retrospective cohort study included consecutive patients undergoing TKA at a specialist centre in the UK between November 2010 and December 2012. Data were collected using a bespoke electronic database and cross-referenced with national arthroplasty audit data, with variables including: preoperative characteristics, intraoperative factors, complications, and mortality status. Patient-reported outcome measures (PROMs) were collected by a specialist research team at ten years post-surgery. There were 536 TKAs, of which 308/536 (57.5%) were in female patients. The mean age was 69.0 years (95% CI 45.0 to 88.0), the mean BMI was 32.2 kg/m2 (95% CI 18.9 to 50.2), and 387/536 (72.2%) survived to ten years. There were four revisions (0.7%): two deep infections (requiring debridement and implant retention), one aseptic loosening, and one haemosiderosis.


The Bone & Joint Journal
Vol. 104-B, Issue 12 | Pages 1313 - 1322
1 Dec 2022
Yapp LZ Clement ND Moran M Clarke JV Simpson AHRW Scott CEH

Aims

The aim of this study was to assess factors associated with the estimated lifetime risk of revision surgery after primary knee arthroplasty (KA).

Methods

All patients from the Scottish Arthroplasty Project dataset undergoing primary KA during the period 1 January 1998 to 31 December 2019 were included. The cumulative incidence function for revision and death was calculated up to 20 years. Adjusted analyses used cause-specific Cox regression modelling to determine the influence of patient factors. The lifetime risk was calculated as a percentage for patients aged between 45 and 99 years using multiple-decrement life table methodology.


Bone & Joint Open
Vol. 3, Issue 2 | Pages 107 - 113
1 Feb 2022
Brunt ACC Gillespie M Holland G Brenkel I Walmsley P

Aims

Periprosthetic joint infection (PJI) occurs in approximately 1% to 2% of total knee arthroplasties (TKA) presenting multiple challenges, such as difficulty in diagnosis, technical complexity, and financial costs. Two-stage exchange is the gold standard for treating PJI but emerging evidence suggests 'two-in-one' single-stage revision as an alternative, delivering comparable outcomes, reduced morbidity, and cost-effectiveness. This study investigates five-year results of modified single-stage revision for treatment of PJI following TKA with bone loss.

Methods

Patients were identified from prospective data on all TKA patients with PJI following the primary procedure. Inclusion criteria were: revision for PJI with bone loss requiring reconstruction, and a minimum five years’ follow-up. Patients were followed up for recurrent infection and assessment of function. Tools used to assess function were Oxford Knee Score (OKS) and American Knee Society Score (AKSS).


Bone & Joint Open
Vol. 3, Issue 5 | Pages 390 - 397
1 May 2022
Hiranaka T Suda Y Saitoh A Tanaka A Arimoto A Koide M Fujishiro T Okamoto K

The kinematic alignment (KA) approach to total knee arthroplasty (TKA) has recently increased in popularity. Accordingly, a number of derivatives have arisen and have caused confusion. Clarification is therefore needed for a better understanding of KA-TKA. Calipered (or true, pure) KA is performed by cutting the bone parallel to the articular surface, compensating for cartilage wear. In soft-tissue respecting KA, the tibial cutting surface is decided parallel to the femoral cutting surface (or trial component) with in-line traction. These approaches are categorized as unrestricted KA because there is no consideration of leg alignment or component orientation. Restricted KA is an approach where the periarthritic joint surface is replicated within a safe range, due to concerns about extreme alignments that have been considered ‘alignment outliers’ in the neutral mechanical alignment approach. More recently, functional alignment and inverse kinematic alignment have been advocated, where bone cuts are made following intraoperative planning, using intraoperative measurements acquired with computer assistance to fulfill good coordination of soft-tissue balance and alignment. The KA-TKA approach aims to restore the patients’ own harmony of three knee elements (morphology, soft-tissue balance, and alignment) and eventually the patients’ own kinematics. The respective approaches start from different points corresponding to one of the elements, yet each aim for the same goal, although the existing implants and techniques have not yet perfectly fulfilled that goal.


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 42 - 49
1 Jan 2018
Walker T Zahn N Bruckner T Streit MR Mohr G Aldinger PR Clarius M Gotterbarm T

Aims. The aim of this independent multicentre study was to assess the mid-term results of mobile bearing unicondylar knee arthroplasty (UKA) for isolated lateral osteoarthritis of the knee joint. Patients and Methods. We retrospectively evaluated 363 consecutive, lateral UKAs (346 patients) performed using the Oxford domed lateral prosthesis undertaken in three high-volume knee arthroplasty centres between 2006 and 2014. Mean age of the patients at surgery was 65 years (36 to 88) with a mean final follow-up of 37 months (12 to 93). Results. A total of 36 (10.5%) patients underwent revision surgery, giving a survival rate of 90.1% at three years (95% confidence intervals (CI) 86.1 to 93.1; number at risk: 155) and 85.0% at five years (95% CI 77.9 to 89.9; number at risk: 43). Dislocation of the mobile bearing occurred in 18 patients (5.6%) at three years (95% CI 1.0 to 16.4; number at risk: 154) and in 20 patients (8.5%) at five years (95% CI 1.0 to 27.0; number at risk: 42). There were no significant differences in the dislocation rate between the participating centres or the surgeons. We were not able to identify an effect of each surgeon’s learning curve on the dislocation rate of the mobile bearing. The clinical outcome in patients without revision surgery at final follow-up was good to excellent, with a mean Oxford knee score of 40.3 (95% CI 39.4 to 41.2), a mean Tegner activity score of 3.2 (95% CI 3.1 to 3.3) and a mean University of California, Los Angeles score of 5.7 (95% CI 5.5 to 5.9). Conclusion. Our data, which consists of a high number of patients treated with mobile bearing UKA in the lateral compartment, indicates a high revision rate of 15% at five years with dislocation of the mobile bearing being the main reason for implant failure. Despite the good functional and clinical results and the high patient satisfaction in our study group, we therefore discontinued using mobile-bearing lateral UKA in favour of a fixed-bearing component. Cite this article: Bone Joint J 2018;100-B:42–9


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 376 - 385
1 Mar 2022
Gramlich Y Hofmann L Kress S Ruckes C Kemmerer M Klug A Hoffmann R Kremer M

Aims

This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant.

Methods

A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life.