Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1206 - 1214
1 Jul 2021
Tsikandylakis G Mortensen KRL Gromov K Mohaddes M Malchau H Troelsen A

Aims

We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty.

Methods

In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 451 - 457
1 Apr 2017
Gromov K Bersang A Nielsen CS Kallemose T Husted H Troelsen A

Aims

The aim of this study was to identify patient- and surgery-related risk factors for sustaining an early periprosthetic fracture following primary total hip arthroplasty (THA) performed using a double-tapered cementless femoral component (Bi-Metric femoral stem; Biomet Inc., Warsaw, Indiana).

Patients and Methods

A total of 1598 consecutive hips, in 1441 patients receiving primary THA between January 2010 and June 2015, were retrospectively identified. Level of pre-operative osteoarthritis, femoral Dorr type and cortical index were recorded. Varus/valgus placement of the stem and canal fill ratio were recorded post-operatively. Periprosthetic fractures were identified and classified according to the Vancouver classification. Regression analysis was performed to identify risk factors for early periprosthetic fracture.