Small SR Rogge RD Malinzak RA Reyes EM Cook PL Farley KA Ritter MA

Objectives. Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone. Methods. Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques. Results. Rotational malalignment between femoral and tibial components generated 40% less overall tibial tray micromotion in RP designs than in standard fixed bearing tibial trays. RP trays reduced micromotion by up to 172 µm in axial compression and 84 µm in rotational malalignment models. Conclusions. Reduced torque transfer at the tibiofemoral interface in RP tibial trays reduces relative component micromotion and may aid long-term stability in cases of revision TKA or poor bone quality. Cite this article: Mr S. R. Small. Micromotion at the tibial plateau in primary and revision total knee arthroplasty: fixed versus rotating platform designs. Bone Joint Res 2016;5:122–129. DOI: 10.1302/2046-3758.54.2000481


Bone & Joint Open
Vol. 2, Issue 10 | Pages 825 - 833
8 Oct 2021
Dailey HL Schwarzenberg P Webb, III EB Boran SAM Guerin S Harty JA

Aims. The study objective was to prospectively assess clinical outcomes for a pilot cohort of tibial shaft fractures treated with a new tibial nailing system that produces controlled axial interfragmentary micromotion. The hypothesis was that axial micromotion enhances fracture healing compared to static interlocking. Methods. Patients were treated in a single level I trauma centre over a 2.5-year period. Group allocation was not randomized; both the micromotion nail and standard-of-care static locking nails (control group) were commercially available and selected at the discretion of the treating surgeons. Injury risk levels were quantified using the Nonunion Risk Determination (NURD) score. Radiological healing was assessed until 24 weeks or clinical union. Low-dose CT scans were acquired at 12 weeks and virtual mechanical testing was performed to objectively assess structural bone healing. Results. A total of 37 micromotion patients and 46 control patients were evaluated. There were no significant differences between groups in terms of age, sex, the proportion of open fractures, or NURD score. There were no nonunions (0%) in the micromotion group versus five (11%) in the control group. The proportion of fractures united was significantly higher in the micromotion group compared to control at 12 weeks (54% vs 30% united; p = 0.043), 18 weeks (81% vs 59%; p = 0.034), and 24 weeks (97% vs 74%; p = 0.005). Structural bone healing scores as assessed by CT scans tended to be higher with micromotion compared to control and this difference reached significance in patients who had biological comorbidities such as smoking. Conclusion. In this pilot study, micromotion fixation was associated with improved healing compared to standard tibial nailing. Further prospective clinical studies will be needed to assess the strength and generalizability of any potential benefits of micromotion fixation. Cite this article: Bone Jt Open 2021;2(10):825–833


Bone & Joint Research
Vol. 10, Issue 11 | Pages 714 - 722
1 Nov 2021
Qi W Feng X Zhang T Wu H Fang C Leung F

Aims. To fully verify the reliability and reproducibility of an experimental method in generating standardized micromotion for the rat femur fracture model. Methods. A modularized experimental device has been developed that allows rat models to be used instead of large animal models, with the aim of reducing systematic errors and time and money constraints on grouping. The bench test was used to determine the difference between the measured and set values of the micromotion produced by this device under different simulated loading weights. The displacement of the fixator under different loading conditions was measured by compression tests, which was used to simulate the unexpected micromotion caused by the rat’s ambulation. In vivo preliminary experiments with a small sample size were used to test the feasibility and effectiveness of the whole experimental scheme and surgical scheme. Results. The bench test showed that a weight loading < 500 g did not affect the operation of experimental device. The compression test demonstrated that the stiffness of the device was sufficient to keep the uncontrollable motion between fracture ends, resulting from the rat’s daily activities, within 1% strain. In vivo results on 15 rats prove that the device works reliably, without overburdening the experimental animals, and provides standardized micromotion reproductively at the fracture site according to the set parameters. Conclusion. Our device was able to investigate the effect of micromotion parameters on fracture healing by generating standardized micromotion to small animal models. Cite this article: Bone Joint Res 2021;10(11):714–722


Bone & Joint Research
Vol. 7, Issue 2 | Pages 187 - 195
1 Feb 2018
Ziebart J Fan S Schulze C Kämmerer PW Bader R Jonitz-Heincke A

Objectives. Enhanced micromotions between the implant and surrounding bone can impair osseointegration, resulting in fibrous encapsulation and aseptic loosening of the implant. Since the effect of micromotions on human bone cells is sparsely investigated, an in vitro system, which allows application of micromotions on bone cells and subsequent investigation of bone cell activity, was developed. Methods. Micromotions ranging from 25 µm to 100 µm were applied as sine or triangle signal with 1 Hz frequency to human osteoblasts seeded on collagen scaffolds. Micromotions were applied for six hours per day over three days. During the micromotions, a static pressure of 527 Pa was exerted on the cells by Ti6Al4V cylinders. Osteoblasts loaded with Ti6Al4V cylinders and unloaded osteoblasts without micromotions served as controls. Subsequently, cell viability, expression of the osteogenic markers collagen type I, alkaline phosphatase, and osteocalcin, as well as gene expression of osteoprotegerin, receptor activator of NF-κB ligand, matrix metalloproteinase-1, and tissue inhibitor of metalloproteinase-1, were investigated. Results. Live and dead cell numbers were higher after 25 µm sine and 50 µm triangle micromotions compared with loaded controls. Collagen type I synthesis was downregulated in respective samples. The metabolic activity and osteocalcin expression level were higher in samples treated with 25 µm micromotions compared with the loaded controls. Furthermore, static loading and micromotions decreased the osteoprotegerin/receptor activator of NF-κB ligand ratio. Conclusion. Our system enables investigation of the behaviour of bone cells at the bone-implant interface under shear stress induced by micromotions. We could demonstrate that micromotions applied under static pressure conditions have a significant impact on the activity of osteoblasts seeded on collagen scaffolds. In future studies, higher mechanical stress will be applied and different implant surface structures will be considered. Cite this article: J. Ziebart, S. Fan, C. Schulze, P. W. Kämmerer, R. Bader, A. Jonitz-Heincke. Effects of interfacial micromotions on vitality and differentiation of human osteoblasts. Bone Joint Res 2018;7:187–195. DOI: 10.1302/2046-3758.72.BJR-2017-0228.R1


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 875 - 883
1 Jul 2022
Mills K Wymenga AB van Hellemondt GG Heesterbeek PJC

Aims. Both the femoral and tibial component are usually cemented at revision total knee arthroplasty (rTKA), while stems can be added with either cemented or press-fit (hybrid) fixation. The aim of this study was to compare the long-term stability of rTKA with cemented and press-fitted stems, using radiostereometric analysis (RSA). Methods. This is a follow-up of a randomized controlled trial, initially involving 32 patients, of whom 19 (nine cemented, ten hybrid) were available for follow-up ten years postoperatively, when further RSA measurements were made. Micromotion of the femoral and tibial components was assessed using model-based RSA software (RSAcore). The clinical outcome was evaluated using the Knee Society Score (KSS), the Knee injury and Osteoarthritis Outcome Score (KOOS), and visual analogue scale (pain and satisfaction). Results. The median total femoral translation and rotation at ten years were 0.39 mm (interquartile range (IQR) 0.20 to 0.54) and 0.59° (IQR 0.46° to 0.73°) for the cemented group and 0.70 mm (IQR 0.15 to 0.77) and 0.78° (IQR 0.47° to 1.43°) for the hybrid group. For the tibial components this was 0.38 mm (IQR 0.33 to 0.85) and 0.98° (IQR 0.38° to 1.34°) for the cemented group and 0.42 mm (IQR 0.30 to 0.52) and 0.72° (IQR 0.62° to 0.82°) for the hybrid group. None of these values were significantly different between the two groups and there were no significant differences between the clinical scores in the two groups at this time. There was only one re-revision, in the hybrid group, for infection and not for aseptic loosening. Conclusion. These results show good long-term fixation with no difference in micromotion and clinical outcome between fully cemented and hybrid fixation in rTKA, which builds on earlier short- to mid-term results. The patients all had type I or II osseous defects, which may in part explain the good results. Cite this article: Bone Joint J 2022;104-B(7):875–883


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 3 | Pages 484 - 489
1 May 1995
Won C Hearn T Tile M

Adult human cadaver pelves were tested to determine micromotion at the prosthesis-bone interface in cementless hemispherical acetabular components during simulated single-limb stance. The micromotion of non-press-fit components with screw fixation in response to cyclic loads to a maximum of 1500N was compressive (interface closing) at the superior iliac rim and distractive (interface opening) at the inferior ischial rim; that of press-fit components was compressive all around the acetabular rim regardless of screw fixation. Adding screws to the component decreased the micromotion at the site of the screw, but sometimes increased it at the opposite side. Two dome screws with the press-fit component decreased the micromotion at the superior iliac rim but at the inferior ischial rim there was either no change or increased movement. A press-fit cup shows less micromotion than a non-press-fit cup with screw fixation. The addition of screws to a press-fit cup does not necessarily increase the initial stability


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 1 | Pages 33 - 37
1 Jan 1991
Burke D O'Connor D Zalenski E Jasty M Harris W

We evaluated the initial stability of cemented and uncemented femoral components within the femoral canals of cadaver femurs during simulated single limb stance and stair climbing. Both types were very stable in simulated single limb stance (maximum micromotion of 42 microns for cemented and 30 microns for uncemented components). However, in simulated stair climbing, the cemented components were much more stable than the uncemented components (76 microns as against 280 microns). There was also greater variation in the stability of uncemented components in simulated stair climbing, with two of the seven components moving 200 microns or more. Future implant designs should aim to improve the initial stability of cementless femoral components under torsional loads; this should improve the chances of bony ingrowth


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 6 | Pages 912 - 917
1 Nov 1994
Karrholm J Borssen B Lowenhielm G Snorrason F

Roentgen stereophotogrammetry was used to measure the migration of the centre of the femoral head in 84 cemented Lubinus SP I hip arthroplasties (58 primary operations, 26 revisions). Four to seven years later, seven femoral components had been revised because of painful loosening. These implants showed greater subsidence, medial migration and posterior migration during the first two postoperative years than did the hips which had not been revised. Six months after operation, subsidence of more than 0.33 mm combined with a total migration of more than 0.85 mm predicted an increased risk of subsequent revision; the amount of subsidence at two years was an even better predictor. The probability of revision was greater than 50% if the subsidence at two years was 1.2 mm or more.


Bone & Joint Research
Vol. 13, Issue 10 | Pages 611 - 621
24 Oct 2024
Wan Q Han Q Liu Y Chen H Zhang A Zhao X Wang J

Aims. This study aimed to investigate the optimal sagittal positioning of the uncemented femoral component in total knee arthroplasty to minimize the risk of aseptic loosening and periprosthetic fracture. Methods. Ten different sagittal placements of the femoral component, ranging from -5 mm (causing anterior notch) to +4 mm (causing anterior gap), were analyzed using finite element analysis. Both gait and squat loading conditions were simulated, and Von Mises stress and interface micromotion were evaluated to assess fracture and loosening risk. Results. During gait, varied sagittal positioning did not lead to excessive Von Mises stress or micromotion. However, under squat conditions, posterior positioning (-4 and -5 mm) resulted in stress exceeding 150 MPa at the femoral notch, indicating potential fracture risk. Conversely, +1 mm and 0 mm sagittal positions demonstrated minimal interface micromotion. Conclusion. Slightly anterior sagittal positioning (+1 mm) or neutral positioning (0 mm) effectively reduced stress concentration at the femoral notch and minimized interface micromotion. Thus, these positions are deemed suitable to decrease the risk of aseptic loosening and periprosthetic femoral fracture


Bone & Joint Research
Vol. 13, Issue 5 | Pages 226 - 236
9 May 2024
Jürgens-Lahnstein JH Petersen ET Rytter S Madsen F Søballe K Stilling M

Aims. Micromotion of the polyethylene (PE) inlay may contribute to backside PE wear in addition to articulate wear of total knee arthroplasty (TKA). Using radiostereometric analysis (RSA) with tantalum beads in the PE inlay, we evaluated PE micromotion and its relationship to PE wear. Methods. A total of 23 patients with a mean age of 83 years (77 to 91), were available from a RSA study on cemented TKA with Maxim tibial components (Zimmer Biomet). PE inlay migration, PE wear, tibial component migration, and the anatomical knee axis were evaluated on weightbearing stereoradiographs. PE inlay wear was measured as the deepest penetration of the femoral component into the PE inlay. Results. At mean six years’ follow-up, the PE wear rate was 0.08 mm/year (95% confidence interval 0.06 to 0.09 mm/year). PE inlay external rotation was below the precision limit and did not influence PE wear. Varus knee alignment did not influence PE wear (p = 0.874), but increased tibial component total translation (p = 0.041). Conclusion. The PE inlay was well fixed and there was no relationship between PE stability and PE wear. The PE wear rate was low and similar in the medial and lateral compartments. Varus knee alignment did not influence PE wear. Cite this article: Bone Joint Res 2024;13(5):226–236


Bone & Joint Research
Vol. 10, Issue 4 | Pages 250 - 258
1 Apr 2021
Kwak D Bang S Lee S Park J Yoo J

Aims. There are concerns regarding initial stability and early periprosthetic fractures in cementless hip arthroplasty using short stems. This study aimed to investigate stress on the cortical bone around the stem and micromotions between the stem and cortical bone according to femoral stem length and positioning. Methods. In total, 12 femoral finite element models (FEMs) were constructed and tested in walking and stair-climbing. Femoral stems of three different lengths and two different positions were simulated, assuming press-fit fixation within each FEM. Stress on the cortical bone and micromotions between the stem and bone were measured in each condition. Results. Stress concentration was observed on the medial and lateral interfaces between the cortical bone and stem. With neutral stem insertion, mean stress over a region of interest was greater at the medial than lateral interface regardless of stem length, which increased as the stem shortened. Mean stress increased in the varus-inserted stems compared to the stems inserted neutrally, especially at the lateral interface in contact with the stem tip. The maximum stress was observed at the lateral interface in a varus-inserted short stem. All mean stresses were greater in stair-climbing condition than walking. Each micromotion was also greater in shorter stems and varus-inserted stems, and in stair-climbing condition. Conclusion. The stem should be inserted neutrally and stair-climbing movement should be avoided in the early postoperative period, in order to preserve early stability and reduce the possibility of thigh pain, especially when using a shorter stem. Cite this article: Bone Joint Res 2021;10(4):250–258


Bone & Joint Research
Vol. 9, Issue 4 | Pages 162 - 172
1 Apr 2020
Xie S Conlisk N Hamilton D Scott C Burnett R Pankaj P

Aims. Metaphyseal tritanium cones can be used to manage the tibial bone loss commonly encountered at revision total knee arthroplasty (rTKA). Tibial stems provide additional fixation and are generally used in combination with cones. The aim of this study was to examine the role of the stems in the overall stability of tibial implants when metaphyseal cones are used for rTKA. Methods. This computational study investigates whether stems are required to augment metaphyseal cones at rTKA. Three cemented stem scenarios (no stem, 50 mm stem, and 100 mm stem) were investigated with 10 mm-deep uncontained posterior and medial tibial defects using four loading scenarios designed to mimic activities of daily living. Results. Small micromotions (mean < 12 µm) were found to occur at the bone-implant interface for all loading cases with or without a stem. Stem inclusion was associated with lower micromotion, however these reductions were too small to have any clinical significance. Peak interface micromotion, even when the cone is used without a stem, was too small to effect osseointegration. The maximum difference occurred with stair descent loading. Stress concentrations in the bone occurred around the inferior aspect of each implant, with the largest occurring at the end of the long stem; these may lead to end-of-stem pain. Stem use is also found to result in stress shielding in the bone along the stem. Conclusion. When a metaphyseal cone is used at rTKA to manage uncontained posterior or medial defects of up to 10 mm depth, stem use may not be necessary. Cite this article:Bone Joint Res. 2020;9(4):162–172


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 35 - 46
1 Jan 2023
Mills K Wymenga AB Bénard MR Kaptein BL Defoort KC van Hellemondt GG Heesterbeek PJC

Aims. The aim of this study was to compare a bicruciate-retaining (BCR) total knee arthroplasty (TKA) with a posterior cruciate-retaining (CR) TKA design in terms of kinematics, measured using fluoroscopy and stability as micromotion using radiostereometric analysis (RSA). Methods. A total of 40 patients with end-stage osteoarthritis were included in this randomized controlled trial. All patients performed a step-up and lunge task in front of a monoplane fluoroscope one year postoperatively. Femorotibial contact point (CP) locations were determined at every flexion angle and compared between the groups. RSA images were taken at baseline, six weeks, three, six, 12, and 24 months postoperatively. Clinical and functional outcomes were compared postoperatively for two years. Results. The BCR-TKA demonstrated a kinematic pattern comparable to the natural knee’s screw-home mechanism in the step-up task. In the lunge task, the medial CP of the BCR-TKA was more anterior in the early flexion phase, while laterally the CP was more posterior during the entire movement cycle. The BCR-TKA group showed higher tibial migration. No differences were found for the clinical and functional outcomes. Conclusion. The BCR-TKA shows a different kinematic pattern in early flexion/late extension compared to the CR-TKA. The difference between both implants is mostly visible in the flexion phase in which the anterior cruciate ligament is effective; however, both designs fail to fully replicate the motion of a natural knee. The higher migration of the BCR-TKA was concerning and highlights the importance of longer follow-up. Cite this article: Bone Joint J 2023;105-B(1):35–46


The Bone & Joint Journal
Vol. 100-B, Issue 9 | Pages 1148 - 1156
1 Sep 2018
Ferguson RJ Broomfield JA Malak TT Palmer AJR Whitwell D Kendrick B Taylor A Glyn-Jones S

Aims. The aim of this study was to determine the stability of a new short femoral stem compared with a conventional femoral stem in patients undergoing cementless total hip arthroplasty (THA), in a prospective randomized controlled trial using radiostereometric analysis (RSA). Patients and Methods. A total of 53 patients were randomized to receive cementless THA with either a short femoral stem (MiniHip, 26 patients, mean age: 52 years, nine male) or a conventional length femoral stem (MetaFix, 23 patients, mean age: 53 years, 11 male). All patients received the same cementless acetabular component. Two-year follow-up was available on 38 patients. Stability was assessed through migration and dynamically inducible micromotion. Radiographs for RSA were taken postoperatively and at three, six, 12, 18, and 24 months. Results. At two years, there was significantly less subsidence (inferior migration) of the short femoral stem (head, 0.26 mm, 95% confidence interval (CI) 0.08 to 0.43, . sd. 0.38; tip, 0.11 mm, 95% CI -0.08 to 0.31, . sd. 0.42) compared with the conventional stem (head, 0.62 mm, 95% CI 0.34 to 0.90, . sd. 0.56, p = 0.02; tip, 0.43 mm, 95% CI 0.21 to 0.65, . sd. 0.44, p = 0.03). There was no significant difference in dynamically inducible micromotion, rate of complications or functional outcome. Conclusion. This study demonstrates that the short femoral stem has a stable and predictable migration. However, longer-term survival analysis still needs to be determined. Cite this article: Bone Joint J 2018;100-B:1148–56


Bone & Joint Research
Vol. 11, Issue 2 | Pages 91 - 101
1 Feb 2022
Munford MJ Stoddart JC Liddle AD Cobb JP Jeffers JRT

Aims. Unicompartmental and total knee arthroplasty (UKA and TKA) are successful treatments for osteoarthritis, but the solid metal implants disrupt the natural distribution of stress and strain which can lead to bone loss over time. This generates problems if the implant needs to be revised. This study investigates whether titanium lattice UKA and TKA implants can maintain natural load transfer in the proximal tibia. Methods. In a cadaveric model, UKA and TKA procedures were performed on eight fresh-frozen knee specimens, using conventional (solid) and titanium lattice tibial implants. Stress at the bone-implant interfaces were measured and compared to the native knee. Results. Titanium lattice implants were able to restore the mechanical environment of the native tibia for both UKA and TKA designs. Maximum stress at the bone-implant interface ranged from 1.2 MPa to 3.3 MPa compared with 1.3 MPa to 2.7 MPa for the native tibia. The conventional solid UKA and TKA implants reduced the maximum stress in the bone by a factor of 10 and caused > 70% of bone surface area to be underloaded compared to the native tibia. Conclusion. Titanium lattice implants maintained the natural mechanical loading in the proximal tibia after UKA and TKA, but conventional solid implants did not. This is an exciting first step towards implants that maintain bone health, but such implants also have to meet fatigue and micromotion criteria to be clinically viable. Cite this article: Bone Joint Res 2022;11(2):91–101


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 962 - 970
1 Jul 2007
Albert C Patil S Frei H Masri B Duncan C Oxland T Fernlund G

This study explored the relationship between the initial stability of the femoral component and penetration of cement into the graft bed following impaction allografting. Impaction allografting was carried out in human cadaveric femurs. In one group the cement was pressurised conventionally but in the other it was not pressurised. Migration and micromotion of the implant were measured under simulated walking loads. The specimens were then cross-sectioned and penetration of the cement measured. Around the distal half of the implant we found approximately 70% and 40% of contact of the cement with the endosteum in the pressure and no-pressure groups, respectively. The distal migration/micromotion, and valgus/varus migration were significantly higher in the no-pressure group than in that subjected to pressure. These motion components correlated negatively with the mean area of cement and its contact with the endosteum. The presence of cement at the endosteum appears to play an important role in the initial stability of the implant following impaction allografting


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 896 - 901
1 Jul 2014
Reiner T Jaeger S Schwarze M Klotz MC Beckmann NA Bitsch RG

Aseptic loosening of the femoral component is an important indication for revision surgery in unicompartmental knee replacement (UKR). A new design of femoral component with an additional peg was introduced for the cemented Oxford UKR to increase its stability. The purpose of this study was to compare the primary stability of the two designs of component. Medial Oxford UKR was performed in 12 pairs of human cadaver knees. In each pair, one knee received the single peg and one received the twin peg design. Three dimensional micromotion and subsidence of the component in relation to the bone was measured under cyclical loading at flexion of 40° and 70° using an optical measuring system. Wilcoxon matched pairs signed-rank test was performed to detect differences between the two groups. . There was no significant difference in the relative micromotion (p = 0.791 and 0.380, respectively) and subsidence (p = 0.301 and 0.176, respectively) of the component between the two groups at both angles of flexion. Both designs of component offered good strength of fixation in this cadaver study. Cite this article: Bone Joint J 2014;96-B:896–901


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 674 - 679
1 May 2017
Nuttall D Birch A Haines JF Watts AC Trail IA

Aims. Radiostereometric analysis (RSA) allows an extremely accurate measurement of early micromotion of components following arthroplasty. . Patients and Methods. In this study, RSA was used to measure the migration of 11 partially cemented fluted pegged glenoid components in patients with osteoarthritis who underwent total shoulder arthroplasty using an improved surgical technique (seven men, four women, mean age 68). Patients were evaluated clinically using the American Shoulder and Elbow Surgeons (ASES) and Constant-Murley scores and by CT scans two years post-operatively. . Results. There were two patterns of migration, the first showing little, if any, migration and the second showing rotation by > 6° as early as three months post-operatively. At two years, these two groups could be confirmed on CT scans, one with osseointegration around the central peg, and the second with cystic changes. Patients with osteolysis around the central peg were those with early migration and those with osseointegration had minimal early migration. Both groups,however,had similar clinical results. . Conclusion. Rapid early migration associated with focal lucency and absence of osseointegration was observed in three of 11 glenoid components, suggesting that lack of initial stability leads to early movement and failure of osseointegration. Cite this article: Bone Joint J 2017;99-B:674–9


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 879 - 883
1 Sep 2024
Kayani B Staats K Haddad FS


The Bone & Joint Journal
Vol. 105-B, Issue 10 | Pages 1045 - 1051
1 Oct 2023
Turgeon TR Righolt CH Burnell CD Gascoyne TC Hedden DR Bohm ER

Aims

The primary aim of this trial was to compare the subsidence of two similar hydroxyapatite-coated titanium femoral components from different manufacturers. Secondary aims were to compare rotational migration (anteversion/retroversion and varus/valgus tilt) and patient-reported outcome measures between both femoral components.

Methods

Patients were randomized to receive one of the two femoral components (Avenir or Corail) during their primary total hip arthroplasty between August 2018 and September 2020. Radiostereometric analysis examinations at six, 12, and 24 months were used to assess the migration of each implanted femoral component compared to a baseline assessment. Patient-reported outcome measures were also recorded for these same timepoints. Overall, 50 patients were enrolled (62% male (n = 31), with a mean age of 65.7 years (SD 7.3), and mean BMI of 30.2 kg/m2 (SD 5.2)).


Bone & Joint Open
Vol. 4, Issue 9 | Pages 682 - 688
6 Sep 2023
Hampton M Balachandar V Charalambous CP Sutton PM

Aims

Aseptic loosening is the most common cause of failure following cemented total knee arthroplasty (TKA), and has been linked to poor cementation technique. We aimed to develop a consensus on the optimal technique for component cementation in TKA.

Methods

A UK-based, three-round, online modified Delphi Expert Consensus Study was completed focusing on cementation technique in TKA. Experts were identified as having a minimum of five years’ consultant experience in the NHS and fulfilling any one of the following criteria: a ‘high volume’ knee arthroplasty practice (> 150 TKAs per annum) as identified from the National joint Registry of England, Wales, Northern Ireland and the Isle of Man; a senior author of at least five peer reviewed articles related to TKA in the previous five years; a surgeon who is named trainer for a post-certificate of comletion of training fellowship in TKA.


Aims

The tibial component of total knee arthroplasty can either be an all-polyethylene (AP) implant or a metal-backed (MB) implant. This study aims to compare the five-year functional outcomes of AP tibial components to MB components in patients aged over 70 years. Secondary aims are to compare quality of life, implant survivorship, and cost-effectiveness.

Methods

A group of 130 patients who had received an AP tibial component were matched for demographic factors of age, BMI, American Society of Anesthesiologists (ASA) grade, sex, and preoperative Knee Society Score (KSS) to create a comparison group of 130 patients who received a MB tibial component. Functional outcome was assessed prospectively by KSS, quality of life by 12-Item Short-Form Health Survey questionnaire (SF-12), and range of motion (ROM), and implant survivorships were compared. The SF six-dimension (6D) was used to calculate the incremental cost effectiveness ratio (ICER) for AP compared to MB tibial components using quality-adjusted life year methodology.


Bone & Joint Research
Vol. 12, Issue 3 | Pages 155 - 164
1 Mar 2023
McCarty CP Nazif MA Sangiorgio SN Ebramzadeh E Park S

Aims

Taper corrosion has been widely reported to be problematic for modular total hip arthroplasty implants. A simple and systematic method to evaluate taper damage with sufficient resolution is needed. We introduce a semiquantitative grading system for modular femoral tapers to characterize taper corrosion damage.

Methods

After examining a unique collection of retrieved cobalt-chromium (CoCr) taper sleeves (n = 465) using the widely-used Goldberg system, we developed an expanded six-point visual grading system intended to characterize the severity, visible material loss, and absence of direct component contact due to corrosion. Female taper sleeve damage was evaluated by three blinded observers using the Goldberg scoring system and the expanded system. A subset (n = 85) was then re-evaluated following destructive cleaning, using both scoring systems. Material loss for this subset was quantified using metrology and correlated with both scoring systems.


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1189 - 1195
1 Nov 2023
Kim JS Kim SH Kim SC Park JH Kim HG Lee SM Yoo JC

Aims

The aim of this study was to compare the clinical and radiological outcomes of reverse shoulder arthroplasty (RSA) using small and standard baseplates in Asian patients, and to investigate the impact of a mismatch in the sizes of the glenoid and the baseplate on the outcomes.

Methods

This was retrospective analysis of 50 and 33 RSAs using a standard (33.8 mm, ST group) and a small (29.5 mm, SM group) baseplate of the Equinoxe reverse shoulder system, which were undertaken between January 2017 and March 2021. Radiological evaluations included the size of the glenoid, the β-angle, the inclination of the glenoid component, inferior overhang, scapular notching, the location of the central cage in the baseplate within the vault and the mismatch in size between the glenoid and baseplate. Clinical evaluations included the range of motion (ROM) and functional scores. In subgroup analysis, comparisons were performed between those in whom the vault of the glenoid was perforated (VP group) and those in whom it was not perforated (VNP group).


Bone & Joint 360
Vol. 12, Issue 2 | Pages 19 - 24
1 Apr 2023

The April 2023 Foot & Ankle Roundup360 looks at: Outcomes following a two-stage revision total ankle arthroplasty for periprosthetic joint infection; Temporary bridge plate fixation and joint motion after an unstable Lisfranc injury; Outcomes of fusion in type II os naviculare; Total ankle arthroplasty versus arthrodesis for end-stage ankle osteoarthritis; Normal saline for plantar fasciitis: placebo or therapeutic?; Distraction arthroplasty for ankle osteoarthritis: does it work?; Let there be movement: ankle arthroplasty after previous fusion; Morbidity and mortality after diabetic Charcot foot arthropathy.


Aims

There are concerns regarding nail/medullary canal mismatch and initial stability after cephalomedullary nailing in unstable pertrochanteric fractures. This study aimed to investigate the effect of an additional anteroposterior blocking screw on fixation stability in unstable pertrochanteric fracture models with a nail/medullary canal mismatch after short cephalomedullary nail (CMN) fixation.

Methods

Eight finite element models (FEMs), comprising four different femoral diameters, with and without blocking screws, were constructed, and unstable intertrochanteric fractures fixed with short CMNs were reproduced in all FEMs. Micromotions of distal shaft fragment related to proximal fragment, and stress concentrations at the nail construct were measured.


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 148 - 157
1 Feb 2023
Koster LA Rassir R Kaptein BL Sierevelt IN Schager M Nelissen RGHH Nolte PA

Aims

The primary aim of this study was to compare the migration of the femoral and tibial components of the cementless rotating platform Attune and Low Contact Stress (LCS) total knee arthroplasty (TKA) designs, two years postoperatively, using radiostereometric analysis (RSA) in order to assess the risk of the development of aseptic loosening. A secondary aim was to compare clinical and patient-reported outcome measures (PROMs) between the designs.

Methods

A total of 61 TKAs were analyzed in this randomized clinical RSA trial. RSA examinations were performed one day and three, six, 12, and 24 months postoperatively. The maximal total point motion (MPTM), translations, and rotations of the components were analyzed. PROMs and clinical data were collected preoperatively and at six weeks and three, six, 12, and 24 months postoperatively. Linear mixed effect modelling was used for statistical analyses.


The Bone & Joint Journal
Vol. 105-B, Issue 6 | Pages 610 - 621
1 Jun 2023
Prodromidis AD Chloros GD Thivaios GC Sutton PM Pandit H Giannoudis PV Charalambous CP

Aims

Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening.

Methods

A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.


Bone & Joint Research
Vol. 11, Issue 5 | Pages 260 - 269
3 May 2022
Staats K Sosa BR Kuyl E Niu Y Suhardi V Turajane K Windhager R Greenblatt MB Ivashkiv L Bostrom MPG Yang X

Aims

To develop an early implant instability murine model and explore the use of intermittent parathyroid hormone (iPTH) treatment for initially unstable implants.

Methods

3D-printed titanium implants were inserted into an oversized drill-hole in the tibiae of C57Bl/6 mice (n = 54). After implantation, the mice were randomly divided into three treatment groups (phosphate buffered saline (PBS)-control, iPTH, and delayed iPTH). Radiological analysis, micro-CT (µCT), and biomechanical pull-out testing were performed to assess implant loosening, bone formation, and osseointegration. Peri-implant tissue formation and cellular composition were evaluated by histology.


Bone & Joint 360
Vol. 11, Issue 3 | Pages 29 - 32
1 Jun 2022


Bone & Joint Research
Vol. 11, Issue 4 | Pages 210 - 213
1 Apr 2022
Fontalis A Haddad FS


The Bone & Joint Journal
Vol. 106-B, Issue 3 | Pages 220 - 223
1 Mar 2024
Kayani B Luo TD Haddad FS


The Bone & Joint Journal
Vol. 105-B, Issue 8 | Pages 912 - 919
1 Aug 2023
Cunningham LJ Walton M Bale S Trail IA

Aims

Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA).

Methods

A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion.


Bone & Joint Open
Vol. 4, Issue 8 | Pages 612 - 620
21 Aug 2023
Martin J Johnson NA Shepherd J Dias J

Aims

There is ambiguity surrounding the degree of scaphoid union required to safely allow mobilization following scaphoid waist fracture. Premature mobilization could lead to refracture, but late mobilization may cause stiffness and delay return to normal function. This study aims to explore the risk of refracture at different stages of scaphoid waist fracture union in three common fracture patterns, using a novel finite element method.

Methods

The most common anatomical variant of the scaphoid was modelled from a CT scan of a healthy hand and wrist using 3D Slicer freeware. This model was uploaded into COMSOL Multiphysics software to enable the application of physiological enhancements. Three common waist fracture patterns were produced following the Russe classification. Each fracture had differing stages of healing, ranging from 10% to 90% partial union, with increments of 10% union assessed. A physiological force of 100 N acting on the distal pole was applied, with the risk of refracture assessed using the Von Mises stress.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 892 - 897
1 Sep 2024
Mancino F Fontalis A Kayani B Magan A Plastow R Haddad FS

Advanced 3D imaging and CT-based navigation have emerged as valuable tools to use in total knee arthroplasty (TKA), for both preoperative planning and the intraoperative execution of different philosophies of alignment. Preoperative planning using CT-based 3D imaging enables more accurate prediction of the size of components, enhancing surgical workflow and optimizing the precision of the positioning of components. Surgeons can assess alignment, osteophytes, and arthritic changes better. These scans provide improved insights into the patellofemoral joint and facilitate tibial sizing and the evaluation of implant-bone contact area in cementless TKA. Preoperative CT imaging is also required for the development of patient-specific instrumentation cutting guides, aiming to reduce intraoperative blood loss and improve the surgical technique in complex cases. Intraoperative CT-based navigation and haptic guidance facilitates precise execution of the preoperative plan, aiming for optimal positioning of the components and accurate alignment, as determined by the surgeon’s philosophy. It also helps reduce iatrogenic injury to the periarticular soft-tissue structures with subsequent reduction in the local and systemic inflammatory response, enhancing early outcomes. Despite the increased costs and radiation exposure associated with CT-based navigation, these many benefits have facilitated the adoption of imaged based robotic surgery into routine practice. Further research on ultra-low-dose CT scans and exploration of the possible translation of the use of 3D imaging into improved clinical outcomes are required to justify its broader implementation.

Cite this article: Bone Joint J 2024;106-B(9):892–897.


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 312 - 318
1 Apr 2024
Sheth NP Jones SA Sanghavi SA Manktelow A

The advent of modular porous metal augments has ushered in a new form of treatment for acetabular bone loss. The function of an augment can be seen as reducing the size of a defect or reconstituting the anterosuperior/posteroinferior columns and/or allowing supplementary fixation. Depending on the function of the augment, the surgeon can decide on the sequence of introduction of the hemispherical shell, before or after the augment. Augments should always, however, be used with cement to form a unit with the acetabular component. Given their versatility, augments also allow the use of a hemispherical shell in a position that restores the centre of rotation and biomechanics of the hip. Progressive shedding or the appearance of metal debris is a particular finding with augments and, with other radiological signs of failure, should be recognized on serial radiographs. Mid- to long-term outcomes in studies reporting the use of augments with hemispherical shells in revision total hip arthroplasty have shown rates of survival of > 90%. However, a higher risk of failure has been reported when augments have been used for patients with chronic pelvic discontinuity.

Cite this article: Bone Joint J 2024;106-B(4):312–318.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 82 - 88
1 May 2024
Villa JM Rajschmir K Hosseinzadeh S Manrique-Succar J Grieco P Higuera-Rueda CA Riesgo AM

Aims

Large bone defects resulting from osteolysis, fractures, osteomyelitis, or metastases pose significant challenges in acetabular reconstruction for total hip arthroplasty. This study aimed to evaluate the survival and radiological outcomes of an acetabular reconstruction technique in patients at high risk of reconstruction failure (i.e. periprosthetic joint infection (PJI), poor bone stock, immunosuppressed patients), referred to as Hip Reconstruction In Situ with Screws and Cement (HiRISC). This involves a polyethylene liner embedded in cement-filled bone defects reinforced with screws and/or plates for enhanced fixation.

Methods

A retrospective chart review of 59 consecutive acetabular reconstructions was performed by four surgeons in a single institution from 18 October 2018 to 5 January 2023. Cases were classified based on the Paprosky classification, excluding type 1 cases (n = 26) and including types 2 or 3 for analysis (n = 33). Radiological loosening was evaluated by an orthopaedic surgeon who was not the operating surgeon, by comparing the immediate postoperative radiographs with the ones at latest follow-up. Mean follow-up was 557 days (SD 441; 31 to 1,707).


Bone & Joint Research
Vol. 13, Issue 4 | Pages 193 - 200
23 Apr 2024
Reynolds A Doyle R Boughton O Cobb J Muirhead-Allwood S Jeffers J

Aims

Manual impaction, with a mallet and introducer, remains the standard method of installing cementless acetabular cups during total hip arthroplasty (THA). This study aims to quantify the accuracy and precision of manual impaction strikes during the seating of an acetabular component. This understanding aims to help improve impaction surgical techniques and inform the development of future technologies.

Methods

Posterior approach THAs were carried out on three cadavers by an expert orthopaedic surgeon. An instrumented mallet and introducer were used to insert cementless acetabular cups. The motion of the mallet, relative to the introducer, was analyzed for a total of 110 strikes split into low-, medium-, and high-effort strikes. Three parameters were extracted from these data: strike vector, strike offset, and mallet face alignment.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 121 - 129
1 Mar 2024
Orce Rodríguez A Smith PN Johnson P O'Sullivan M Holder C Shimmin A

Aims

In recent years, the use of a collared cementless femoral prosthesis has risen in popularity. The design intention of collared components is to transfer some load to the resected femoral calcar and prevent implant subsidence within the cancellous bone of the metaphysis. Conversely, the load transfer for a cemented femoral prosthesis depends on the cement-component and cement-bone interface interaction. The aim of our study was to compare the three most commonly used collared cementless components and the three most commonly used tapered polished cemented components in patients aged ≥ 75 years who have undergone a primary total hip arthroplasty (THA) for osteoarthritis (OA).

Methods

Data from the Australian Orthopaedic Association National Joint Replacement Registry from 1 September 1999 to 31 December 2022 were analyzed. Collared cementless femoral components and cemented components were identified, and the three most commonly used components in each group were analyzed. We identified a total of 11,278 collared cementless components and 47,835 cemented components. Hazard ratios (HRs) from Cox proportional hazards models, adjusting for age and sex, were obtained to compare the revision rates between the groups.


Bone & Joint Research
Vol. 11, Issue 2 | Pages 82 - 90
7 Feb 2022
Eckert JA Bitsch RG Sonntag R Reiner T Schwarze M Jaeger S

Aims

The cemented Oxford unicompartmental knee arthroplasty (OUKA) features two variants: single and twin peg OUKA. The purpose of this study was to assess the stability of both variants in a worst-case scenario of bone defects and suboptimal cementation.

Methods

Single and twin pegs were implanted randomly allocated in 12 pairs of human fresh-frozen femora. We generated 5° bone defects at the posterior condyle. Relative movement was simulated using a servohydraulic pulser, and analyzed at 70°/115° knee flexion. Relative movement was surveyed at seven points of measurement on implant and bone, using an optic system.


The Bone & Joint Journal
Vol. 105-B, Issue 10 | Pages 1099 - 1107
1 Oct 2023
Henry JK Shaffrey I Wishman M Palma Munita J Zhu J Cody E Ellis S Deland J Demetracopoulos C

Aims

The Vantage Total Ankle System is a fourth-generation low-profile fixed-bearing implant that has been available since 2016. We aimed to describe our early experience with this implant.

Methods

This is a single-centre retrospective review of patients who underwent primary total ankle arthroplasty (TAA) with a Vantage implant between November 2017 and February 2020, with a minimum of two years’ follow-up. Four surgeons contributed patients. The primary outcome was reoperation and revision rate of the Vantage implant at two years. Secondary outcomes included radiological alignment, peri-implant complications, and pre- and postoperative patient-reported outcomes.


Bone & Joint Open
Vol. 5, Issue 1 | Pages 20 - 27
17 Jan 2024
Turgeon TR Vasarhelyi E Howard J Teeter M Righolt CH Gascoyne T Bohm E

Aims

A novel enhanced cement fixation (EF) tibial implant with deeper cement pockets and a more roughened bonding surface was released to market for an existing total knee arthroplasty (TKA) system.This randomized controlled trial assessed fixation of the both the EF (ATTUNE S+) and standard (Std; ATTUNE S) using radiostereometric analysis.

Methods

Overall, 50 subjects were randomized (21 EF-TKA and 23 Std-TKA in the final analysis), and had follow-up visits at six weeks, and six, 12, and 24 months to assess migration of the tibial component. Low viscosity bone cement with tobramycin was used in a standardized fashion for all subjects. Patient-reported outcome measure data was captured at preoperative and all postoperative visits.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 66 - 73
1 May 2024
Chaudhry F Daud A Greenberg A Braunstein D Safir OA Gross AE Kuzyk PR

Aims

Pelvic discontinuity is a challenging acetabular defect without a consensus on surgical management. Cup-cage reconstruction is an increasingly used treatment strategy. The present study evaluated implant survival, clinical and radiological outcomes, and complications associated with the cup-cage construct.

Methods

We included 53 cup-cage construct (51 patients) implants used for hip revision procedures for pelvic discontinuity between January 2003 and January 2022 in this retrospective review. Mean age at surgery was 71.8 years (50.0 to 92.0; SD 10.3), 43/53 (81.1%) were female, and mean follow-up was 6.4 years (0.02 to 20.0; SD 4.6). Patients were implanted with a Trabecular Metal Revision Shell with either a ZCA cage (n = 12) or a TMARS cage (n = 40, all Zimmer Biomet). Pelvic discontinuity was diagnosed on preoperative radiographs and/or intraoperatively. Kaplan-Meier survival analysis was performed, with failure defined as revision of the cup-cage reconstruction.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 898 - 906
1 Sep 2024
Kayani B Wazir MUK Mancino F Plastow R Haddad FS

Aims

The primary objective of this study was to develop a validated classification system for assessing iatrogenic bone trauma and soft-tissue injury during total hip arthroplasty (THA). The secondary objective was to compare macroscopic bone trauma and soft-tissues injury in conventional THA (CO THA) versus robotic arm-assisted THA (RO THA) using this classification system.

Methods

This study included 30 CO THAs versus 30 RO THAs performed by a single surgeon. Intraoperative photographs of the osseous acetabulum and periacetabular soft-tissues were obtained prior to implantation of the acetabular component, which were used to develop the proposed classification system. Interobserver and intraobserver variabilities of the proposed classification system were assessed.


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 2 | Pages 252 - 258
1 Mar 1990
Albrektsson B Ryd L Carlsson L Freeman M Herberts P Regner L Selvik G

We studied the effect of a metal tray with an intramedullary stem on the micromotion of the tibial component in total knee arthroplasty. Of 32 uncemented Freeman-Samuelson knee arthroplasties performed in London and Gothenburg, nine had a metal backing and stem added to the tibial component. Micromotion of the tibial components, expressed as migration and inducible displacement, was analysed using roentgen stereophotogrammetric analysis up to two years follow-up. The addition of a metal back and a 110 mm stem to the standard polyethylene component significantly reduced both migration over two years and inducible displacement


Bone & Joint Research
Vol. 13, Issue 10 | Pages 559 - 572
8 Oct 2024
Wu W Zhao Z Wang Y Liu M Zhu G Li L

Aims

This study aimed to demonstrate the promoting effect of elastic fixation on fracture, and further explore its mechanism at the gene and protein expression levels.

Methods

A closed tibial fracture model was established using 12 male Japanese white rabbits, and divided into elastic and stiff fixation groups based on different fixation methods. Two weeks after the operation, a radiograph and pathological examination of callus tissue were used to evaluate fracture healing. Then, the differentially expressed proteins (DEPs) were examined in the callus using proteomics. Finally, in vitro cell experiments were conducted to investigate hub proteins involved in this process.


Bone & Joint Research
Vol. 11, Issue 8 | Pages 575 - 584
17 Aug 2022
Stoddart JC Garner A Tuncer M Cobb JP van Arkel RJ

Aims

The aim of this study was to determine the risk of tibial eminence avulsion intraoperatively for bi-unicondylar knee arthroplasty (Bi-UKA), with consideration of the effect of implant positioning, overstuffing, and sex, compared to the risk for isolated medial unicondylar knee arthroplasty (UKA-M) and bicruciate-retaining total knee arthroplasty (BCR-TKA).

Methods

Two experimentally validated finite element models of tibia were implanted with UKA-M, Bi-UKA, and BCR-TKA. Intraoperative loads were applied through the condyles, anterior cruciate ligament (ACL), medial collateral ligament (MCL), and lateral collateral ligament (LCL), and the risk of fracture (ROF) was evaluated in the spine as the ratio of the 95th percentile maximum principal elastic strains over the tensile yield strain of proximal tibial bone.


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 76 - 82
1 Jan 2022
ten Brinke B Hesseling B Eygendaal D Hoelen MA Mathijssen NMC

Aims

Stemless humeral implants have been developed to overcome stem-related complications in total shoulder arthroplasty (TSA). However, stemless implant designs may hypothetically result in less stable initial fixation, potentially affecting long-term survival. The aim of this study is to investigate early fixation and migration patterns of the stemless humeral component of the Simpliciti Shoulder System and to evaluate clinical outcomes.

Methods

In this prospective cohort study, radiostereometric analysis (RSA) radiographs were obtained in 24 patients at one day, six weeks, six months, one year, and two years postoperatively. Migration was calculated using model-based RSA. Clinical outcomes were evaluated using the visual analogue scale (VAS), the Oxford Shoulder Score (OSS), the Constant-Murley Score (CMS), and the Disabilities of the Arm, Shoulder and Hand (DASH) score.


Bone & Joint Open
Vol. 2, Issue 11 | Pages 1004 - 1016
26 Nov 2021
Wight CM Whyne CM Bogoch ER Zdero R Chapman RM van Citters DW Walsh WR Schemitsch E

Aims

This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads.

Methods

In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.


Bone & Joint Open
Vol. 3, Issue 6 | Pages 495 - 501
14 Jun 2022
Keohane D Sheridan GA Masterson E

Aims

Total knee arthroplasty (TKA) is a common and safe orthopaedic procedure. Zimmer Biomet's NexGen is the second most popular brand of implant used in the UK. The primary cause of revision after the first year is aseptic loosening. We present our experience of using this implant, with significant concerns around its performance with regards early aseptic loosening of the tibial component.

Methods

A retrospective, single-surgeon review was carried out of all of the NexGen Legacy Posterior Stabilized (LPS) TKAs performed in this institute. The specific model used for the index procedures was the NexGen Complete Knee System (Legacy Knee-Posterior Stabilized LPS-Flex Articular Surface, LPS-Flex Femoral Component Option, and Stemmed Nonaugmentable Tibial Component Option).


Bone & Joint Research
Vol. 10, Issue 8 | Pages 488 - 497
10 Aug 2021
Cleemann R Sorensen M West A Soballe K Bechtold JE Baas J

Aims

We wanted to evaluate the effects of a bone anabolic agent (bone morphogenetic protein 2 (BMP-2)) on an anti-catabolic background (systemic or local zoledronate) on fixation of allografted revision implants.

Methods

An established allografted revision protocol was implemented bilaterally into the stifle joints of 24 canines. At revision surgery, each animal received one BMP-2 (5 µg) functionalized implant, and one raw implant. One group (12 animals) received bone graft impregnated with zoledronate (0.005 mg/ml) before impaction. The other group (12 animals) received untreated bone graft and systemic zoledronate (0.1 mg/kg) ten and 20 days after revision surgery. Animals were observed for an additional four weeks before euthanasia.


Aims

The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail).

Methods

Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using radiostereometric analysis (RSA), change of bone mineral density (BMD) around the stem, the development of radiolucent lines, and additional patient-reported outcome measures (PROMs).


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1238 - 1246
1 Jul 2021
Hemmerling KJ Weitzler L Bauer TW Padgett DE Wright TM

Aims

Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues.

Methods

A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy.


Aims

The aim of this study was to evaluate medium-term outcomes and complications of the S-ROM NOILES Rotating Hinge Knee System (DePuy, USA) in revision total knee arthroplasty (rTKA) at a tertiary unit.

Methods

A retrospective consecutive study of all patients who underwent a rTKA using this implant from January 2005 to December 2018. Outcome measures included reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 150 - 157
1 Jun 2021
Anderson LA Christie M Blackburn BE Mahan C Earl C Pelt CE Peters CL Gililland J

Aims

Porous metaphyseal cones can be used for fixation in revision total knee arthroplasty (rTKA) and complex TKAs. This metaphyseal fixation has led to some surgeons using shorter cemented stems instead of diaphyseal engaging cementless stems with a potential benefit of ease of obtaining proper alignment without being beholden to the diaphysis. The purpose of this study was to evaluate short term clinical and radiographic outcomes of a series of TKA cases performed using 3D-printed metaphyseal cones.

Methods

A retrospective review of 86 rTKAs and nine complex primary TKAs, with an average age of 63.2 years (SD 8.2) and BMI of 34.0 kg/m2 (SD 8.7), in which metaphyseal cones were used for both femoral and tibial fixation were compared for their knee alignment based on the type of stem used. Overall, 22 knees had cementless stems on both sides, 52 had cemented stems on both sides, and 15 had mixed stems. Postoperative long-standing radiographs were evaluated for coronal and sagittal plane alignment. Adjusted logistic regression models were run to assess malalignment hip-knee-ankle (HKA) alignment beyond ± 3° and sagittal alignment of the tibial and femoral components ± 3° by stem type.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 135 - 144
1 Jul 2021
Kuyl E Shu F Sosa BR Lopez JD Qin D Pannellini T Ivashkiv LB Greenblatt MB Bostrom MPG Yang X

Aims

Aseptic loosening is a leading cause of uncemented arthroplasty failure, often accompanied by fibrotic tissue at the bone-implant interface. A biological target, neutrophil extracellular traps (NETs), was investigated as a crucial connection between the innate immune system’s response to injury, fibrotic tissue development, and proper bone healing. Prevalence of NETs in peri-implant fibrotic tissue from aseptic loosening patients was assessed. A murine model of osseointegration failure was used to test the hypothesis that inhibition (through Pad4-/- mice that display defects in peptidyl arginine deiminase 4 (PAD4), an essential protein required for NETs) or resolution (via DNase 1 treatment, an enzyme that degrades the cytotoxic DNA matrix) of NETs can prevent osseointegration failure and formation of peri-implant fibrotic tissue.

Methods

Patient peri-implant fibrotic tissue was analyzed for NETs biomarkers. To enhance osseointegration in loose implant conditions, an innate immune system pathway (NETs) was either inhibited (Pad4-/- mice) or resolved with a pharmacological agent (DNase 1) in a murine model of osseointegration failure.


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 68 - 75
1 Jan 2022
Harris NJ Nicholson G Pountos I

Aims

The ideal management of acute syndesmotic injuries in elite athletes is controversial. Among several treatment methods used to stabilize the syndesmosis and facilitate healing of the ligaments, the use of suture tape (InternalBrace) has previously been described. The purpose of this study was to analyze the functional outcome, including American Orthopaedic Foot & Ankle Society (AOFAS) scores, knee-to-wall measurements, and the time to return to play in days, of unstable syndesmotic injuries treated with the use of the InternalBrace in elite athletes.

Methods

Data on a consecutive group of elite athletes who underwent isolated reconstruction of the anterior inferior tibiofibular ligament using the InternalBrace were collected prospectively. Our patient group consisted of 19 elite male athletes with a mean age of 24.5 years (17 to 52). Isolated injuries were seen in 12 patients while associated injuries were found in seven patients (fibular fracture, medial malleolus fracture, anterior talofibular ligament rupture, and posterior malleolus fracture). All patients had a minimum follow-up period of 17 months (mean 27 months (17 to 35)).


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 34 - 44
1 Jan 2022
Beckers L Dandois F Ooms D Berger P Van Laere K Scheys L Vandenneucker H

Aims

Higher osteoblastic bone activity is expected in aseptic loosening and painful unicompartmental knee arthroplasty (UKA). However, insights into normal bone activity patterns after medial UKAs are lacking. The aim of this study was to identify the evolution in bone activity pattern in well-functioning medial mobile-bearing UKAs.

Methods

In total, 34 patients (13 female, 21 male; mean age 62 years (41 to 79); BMI 29.7 kg/m2 (23.6 to 42.1)) with 38 medial Oxford partial UKAs (20 left, 18 right; 19 cementless, 14 cemented, and five hybrid) were prospectively followed with sequential 99mTc-hydroxymethane diphosphonate single photon emission CT (SPECT)/CT preoperatively, and at one and two years postoperatively. Changes in mean osteoblastic activity were investigated using a tracer localization scheme with volumes of interest (VOIs), reported by normalized mean tracer values. A SPECT/CT registration platform additionally explored cortical tracer evolution in zones of interest identified by previous experimental research.


Bone & Joint Research
Vol. 10, Issue 7 | Pages 388 - 400
8 Jul 2021
Dall’Ava L Hothi H Henckel J Di Laura A Tirabosco R Eskelinen A Skinner J Hart A

Aims

The main advantage of 3D-printed, off-the-shelf acetabular implants is the potential to promote enhanced bony fixation due to their controllable porous structure. In this study we investigated the extent of osseointegration in retrieved 3D-printed acetabular implants.

Methods

We compared two groups, one made via 3D-printing (n = 7) and the other using conventional techniques (n = 7). We collected implant details, type of surgery and removal technique, patient demographics, and clinical history. Bone integration was assessed by macroscopic visual analysis, followed by sectioning to allow undecalcified histology on eight sections (~200 µm) for each implant. The outcome measures considered were area of bone attachment (%), extent of bone ingrowth (%), bone-implant contact (%), and depth of ingrowth (%), and these were quantified using a line-intercept method.


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims

The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs).

Methods

At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.


Bone & Joint Open
Vol. 2, Issue 8 | Pages 569 - 575
1 Aug 2021
Bouguennec N Robinson J Douiri A Graveleau N Colombet PD

Aims

MRI has been suggested as an objective method of assessing anterior crucate ligament (ACL) graft “ligamentization” after reconstruction. It has been proposed that the MRI appearances could be used as an indicator of graft maturity and used as part of a return-to-sport assessment. The aim of this study was to evaluate the correlation between MRI graft signal and postoperative functional scores, anterior knee laxity, and patient age at operation.

Methods

A consecutive cohort of 149 patients who had undergone semitendinosus autograft ACL reconstruction, using femoral and tibial adjustable loop fixations, were evaluated retrospectively postoperatively at two years. All underwent MRI analysis of the ACL graft, performed using signal-to-noise quotient (SNQ) and the Howell score. Functional outcome scores (Lysholm, Tegner, International Knee Documentation Committee (IKDC) subjective, and IKDC objective) were obtained and all patients underwent instrumented side-to-side anterior laxity differential laxity testing.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 33 - 40
1 Jul 2020
Gustafson JA Pourzal R Levine BR Jacobs JJ Lundberg HJ

Aims

The aim of this study was to develop a novel computational model for estimating head/stem taper mechanics during different simulated assembly conditions.

Methods

Finite element models of generic cobalt-chromium (CoCr) heads on a titanium stem taper were developed and driven using dynamic assembly loads collected from clinicians. To verify contact mechanics at the taper interface, comparisons of deformed microgroove characteristics (height and width of microgrooves) were made between model estimates with those measured from five retrieved implants. Additionally, these models were used to assess the role of assembly technique—one-hit versus three-hits—on the taper interlock mechanical behaviour.


Bone & Joint Open
Vol. 2, Issue 1 | Pages 48 - 57
19 Jan 2021
Asokan A Plastow R Kayani B Radhakrishnan GT Magan AA Haddad FS

Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes.

Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty.

This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice.

Cite this article: Bone Jt Open 2021;2(1):48–57.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1292 - 1300
1 Jul 2021
Märtens N Heinze M Awiszus F Bertrand J Lohmann CH Berth A

Aims

The purpose of this study was to compare clinical results, long-term survival, and complication rates of stemless shoulder prosthesis with stemmed anatomical shoulder prostheses for treatment of osteoarthritis and to analyze radiological bone changes around the implants during follow-up.

Methods

A total of 161 patients treated with either a stemmed or a stemless shoulder arthroplasty for primary osteoarthritis of the shoulder were evaluated with a mean follow-up of 118 months (102 to 158). The Constant score (CS), the Disabilities of the Arm, Shoulder and Hand (DASH) score, and active range of motion (ROM) were recorded. Radiological analysis for bone adaptations was performed by plain radiographs. A Kaplan-Meier survivorship analysis was calculated and complications were noted.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 165 - 170
1 Jun 2021
Larson DJ Rosenberg JH Lawlor MA Garvin KL Hartman CW Lyden E Konigsberg BS

Aims

Stemmed tibial components are frequently used in revision total knee arthroplasty (TKA). The purpose of this study was to evaluate patient satisfaction, overall pain, and diaphyseal tibial pain in patients who underwent revision TKA with cemented or uncemented stemmed tibial components.

Methods

This is a retrospective cohort study involving 110 patients with revision TKA with cemented versus uncemented stemmed tibial components. Patients who underwent revision TKA with stemmed tibial components over a 15-year period at a single institution with at least two-year follow-up were assessed. Pain was evaluated through postal surveys. There were 63 patients with cemented tibial stems and 47 with uncemented stems. Radiographs and Knee Society Scores were used to evaluate for objective findings associated with pain or patient dissatisfaction. Postal surveys were analyzed using Fisher’s exact test and the independent-samples t-test. Logistic regression was used to adjust for age, sex, and preoperative bone loss.


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1016 - 1024
1 Aug 2020
Hasan S van Hamersveld KT Marang-van de Mheen PJ Kaptein BL Nelissen RGHH Toksvig-Larsen S

Aims

Although bone cement is the primary mode of fixation in total knee arthroplasty (TKA), cementless fixation is gaining interest as it has the potential of achieving lasting biological fixation. By 3D printing an implant, highly porous structures can be manufactured, promoting osseointegration into the implant to prevent aseptic loosening. This study compares the migration of cementless, 3D-printed TKA to cemented TKA of a similar design up to two years of follow-up using radiostereometric analysis (RSA) known for its ability to predict aseptic loosening.

Methods

A total of 72 patients were randomized to either cementless 3D-printed or a cemented cruciate retaining TKA. RSA and clinical scores were evaluated at baseline and postoperatively at three, 12, and 24 months. A mixed model was used to analyze the repeated measurements.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 12 | Pages 1602 - 1609
1 Dec 2011
Malviya A Ramaskandhan JR Bowman R Hashmi M Holland JP Kometa S Lingard E

The aim of this study was to investigate the possible benefit of large-head metal-on-metal bearing on a stem for primary hip replacement compared with a 28 mm diameter conventional metal-on-polyethylene bearing in a prospective randomised controlled trial. We investigated cemented stem behaviour between these two different bearings using Einzel-Bild-Röntgen-Analyse, clinical and patient reported measures (Harris hip score, Western Ontario and McMaster Universities osteoarthritis index, Short Form-36 and satisfaction) and whole blood metal ion levels at two years. A power study indicated that 50 hips were needed in each group to detect subsidence of > 5 mm at two years with a p-value of < 0.05. . Significant improvement (p < 0.001) was found in the mean clinical and patient reported outcomes at two years for both groups. Comparison of outcomes between the groups at two years showed no statistically significant difference for mean stem migration, clinical and patient reported outcomes; except overall patient satisfaction which was higher for metal-on-metal group (p = 0.05). Metal ion levels were raised above the Medicines and Healthcare products Regulatory Agency advised safety level (7 µg per litre) in 20% of the metal-on-metal group and in one patient in metal-on-polyethylene group (who had a metal-on-metal implant on the contralateral side). Two patients in the metal-on-metal group were revised, one for pseudotumour and one for peri-prosthetic fracture. . Use of large modular heads is associated with a risk of raised whole blood metal ion levels despite using a proven bearing from resurfacing. The head-neck junction or excess stem micromotion are possibly the weak links warranting further research


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 51 - 58
1 Jun 2021
Yang J Heckmann ND Nahhas CR Salzano MB Ruzich GP Jacobs JJ Paprosky WG Rosenberg AG Nam D

Aims

Recent total knee arthroplasty (TKA) designs have featured more anatomical morphologies and shorter tibial keels. However, several reports have raised concerns about the impact of these modifications on implant longevity. The aim of this study was to report the early performance of a modern, cemented TKA design.

Methods

All patients who received a primary, cemented TKA between 2012 and 2017 with a minimum two-year follow-up were included. The implant investigated features an asymmetrical tibial baseplate and shortened keel. Patient demographic details, Knee Society Scores (KSS), component alignment, and the presence of radiolucent lines at final follow-up were recorded. Kaplan-Meier analyses were performed to estimate survivorship.


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 830 - 839
1 May 2021
Ahmed I Chawla A Underwood M Price AJ Metcalfe A Hutchinson CE Warwick J Seers K Parsons H Wall PDH

Aims

Many surgeons choose to perform total knee arthroplasty (TKA) surgery with the aid of a tourniquet. A tourniquet is a device that fits around the leg and restricts blood flow to the limb. There is a need to understand whether tourniquets are safe, and if they benefit, or harm, patients. The aim of this study was to determine the benefits and harms of tourniquet use in TKA surgery.

Methods

We searched MEDLINE, EMBASE, Cochrane Central Register of Controlled trials, and trial registries up to 26 March 2020. We included randomized controlled trials (RCTs), comparing TKA with a tourniquet versus without a tourniquet. Outcomes included: pain, function, serious adverse events (SAEs), blood loss, implant stability, duration of surgery, and length of hospital stay.


Bone & Joint Open
Vol. 2, Issue 6 | Pages 443 - 456
28 Jun 2021
Thompson JW Corbett J Bye D Jones A Tissingh EK Nolan J

Aims

The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems.

Methods

A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 417 - 425
1 May 1998
Önsten I Nordqvist A Carlsson ÅS Besjakov J Shott S

In a single-blind, randomised series of knee replacements in 116 patients, we used radiostereometric analysis (RSA) to measure micromotion in three types of tibial implant fixation for two years after knee replacement. We compared hydroxyapatite-augmented porous coating, porous coating, and cemented fixation of the same design of tibial component. At one to two years, porous-coated implants migrated at a statistically significantly higher rate than hydroxyapatite-augmented or cemented implants. There was no significant difference between hydroxyapatite-coated and cemented implants. We conclude that hydroxyapatite augmentation may offer a clinically relevant advantage over a simple porous coating for tibial component fixation, but is no better than cemented fixation


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 644 - 649
1 Apr 2021
Alsousou J Oragu E Martin A Strickland L Newman S Kendrick B Taylor A Glyn-Jones S

Aims

The aim of this prospective cohort study was to evaluate the early migration of the TriFit cementless proximally coated tapered femoral stem using radiostereometric analysis (RSA).

Methods

A total of 21 patients (eight men and 13 women) undergoing primary total hip arthroplasty (THA) for osteoarthritis of the hip were recruited in this study and followed up for two years. Two patients were lost to follow-up. All patients received a TriFit stem and Trinity Cup with a vitamin E-infused highly cross-linked ultra-high molecular weight polyethylene liner. Radiographs for RSA were taken postoperatively and then at three, 12, and 24 months. Oxford Hip Score (OHS), EuroQol five-dimension questionnaire (EQ-5D), and adverse events were reported.


Bone & Joint 360
Vol. 9, Issue 5 | Pages 4 - 9
1 Oct 2020
Matthews E Waterson HB Phillips JR Toms AD


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 1 | Pages 85 - 91
1 Jan 1996
Önsten I Carlsson ÅS Sanzén L Besjakov J

A consecutive series of 30 total hip replacements using a hydroxyapatite (HA)-coated, modular implant (Omnifit) was followed clinically and by roentgen stereophotogrammetric analysis for two years and compared with two control groups, one of 27 cemented Charnley sockets and one of 40 cemented Charnley stems. Omnifit sockets with a central gap between the dome of the socket and the acetabular bone in the postoperative radiographs, migrated less than sockets without such gaps (p = 0.01). After adjustment for patient-related factors (age, gender and weight), no significant difference was found between the two prostheses with respect to micromotion and wear. We conclude that the early fixation of the HA-coated Omnifit prosthesis compares with that of the cemented Charnley prosthesis


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 650 - 658
1 Apr 2021
Konow T Baetz J Melsheimer O Grimberg A Morlock M

Aims

Periprosthetic femoral fractures (PPF) are a serious complication of total hip arthroplasty (THA) and are becoming an increasingly common indication for revision arthroplasty with the ageing population. This study aimed to identify potential risk factors for PPF based on an analysis of registry data.

Methods

Cases recorded with PPF as the primary indication for revision arthroplasty in the German Arthroplasty Registry (Endoprothesenregister Deutschland (EPRD)), as well as those classified as having a PPF according to the International Classification of Diseases (ICD) codes in patients’ insurance records were identified from the complete datasets of 249,639 registered primary hip arthroplasties in the EPRD and included in the analysis.


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 1 | Pages 67 - 73
1 Jan 1992
Kwong L Jasty M Mulroy R Maloney W Bragdon C Harris W

The radiographic and histological features of radiolucent areas at the cement-bone interface were correlated in 15 specimens retrieved at post-mortem from patients who had undergone cemented total hip arthroplasty, two weeks to 15 years prior to death. All but one of the components were securely fixed, as demonstrated by direct measurements of micromotion. Extensive radiolucencies were present in all but one case. In 11 of the 14 specimens with radiolucencies, histological examination showed that the radiolucent areas represented regions of osteoporosis and bone remodelling. The remodelling changes were characterised by osteoporosis, cancellisation and thinning of the endosteal cortex, and osteopenia of the trabecular bone. In two specimens the appearance of radiolucency was found to be due to fibrous tissue at the cement-bone interface and in one specimen there was a mixed picture of osteolysis and fibrosis. The study demonstrates that radiolucent lines can occur with well-fixed components and that they may commonly represent osteoporosis rather than the presence of a fibrous membrane at the cement-bone interface


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1025 - 1032
1 Aug 2020
Hampton M Mansoor J Getty J Sutton PM

Aims

Total knee arthroplasty is an established treatment for knee osteoarthritis with excellent long-term results, but there remains controversy about the role of uncemented prostheses. We present the long-term results of a randomized trial comparing an uncemented tantalum metal tibial component with a conventional cemented component of the same implant design.

Methods

Patients under the age of 70 years with symptomatic osteoarthritis of the knee were randomized to receive either an uncemented tantalum metal tibial monoblock component or a standard cemented modular component. The mean age at time of recruitment to the study was 63 years (50 to 70), 46 (51.1%) knees were in male patients, and the mean body mass index was 30.4 kg/m2 (21 to 36). The same cruciate retaining total knee system was used in both groups. All patients received an uncemented femoral component and no patients had their patella resurfaced. Patient outcomes were assessed preoperatively and postoperatively using the modified Oxford Knee Score, Knee Society Score, and 12-Item Short-Form Health Survey questionnaire (SF-12) score. Radiographs were analyzed using the American Knee Society Radiograph Evaluation score. Operative complications, reoperations, or revision surgery were recorded. A total of 90 knees were randomized and at last review 77 knees were assessed. In all, 11 patients had died and two were lost to follow-up.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 20 - 26
1 Jul 2020
Romero J Wach A Silberberg S Chiu Y Westrich G Wright TM Padgett DE

Aims

This combined clinical and in vitro study aimed to determine the incidence of liner malseating in modular dual mobility (MDM) constructs in primary total hip arthroplasties (THAs) from a large volume arthroplasty centre, and determine whether malseating increases the potential for fretting and corrosion at the modular metal interface in malseated MDM constructs using a simulated corrosion chamber.

Methods

For the clinical arm of the study, observers independently reviewed postoperative radiographs of 551 primary THAs using MDM constructs from a single manufacturer over a three-year period, to identify the incidence of MDM liner-shell malseating. Multivariable logistic regression analysis was performed to identify risk factors including age, sex, body mass index (BMI), cup design, cup size, and the MDM case volume of the surgeon. For the in vitro arm, six pristine MDM implants with cobalt-chrome liners were tested in a simulated corrosion chamber. Three were well-seated and three were malseated with 6° of canting. The liner-shell couples underwent cyclic loading of increasing magnitudes. Fretting current was measured throughout testing and the onset of fretting load was determined by analyzing the increase in average current.


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 3 | Pages 377 - 382
1 May 1996
Lind M Overgaard S Ongpipattanakul B Nguyen T Bünger C Søballe K

Bone growth into cementless prosthetic components is compromised by osteoporosis, by any gap between the implant and the bone, by micromotion, and after the revision of failed prostheses. Recombinant human transforming growth factor-β1 (rhTGF-β1) has recently been shown to be a potent stimulator of bone healing and bone formation in various models in vivo. We have investigated the potential of rhTGF-β1, adsorbed on to weight-loaded tricalcium phosphate (TCP) coated implants, to enhance bone ongrowth and mechanical fixation. We inserted cylindrical grit-blasted titanium alloy implants bilaterally into the weight-bearing part of the medial femoral condyles of ten skeletally mature dogs. The implants were mounted on special devices which ensured stable weight-loading during each gait cycle. All implants were initially surrounded by a 0.75 mm gap and were coated with TCP ceramic. Each animal received two implants, one with 0.3 μg rhTGF-β1 adsorbed on the ceramic surface and the other without growth factor. Histological analysis showed that bone ongrowth was significantly increased from 22 ± 5.6% bone-implant contact in the control group to 36 ± 2.9% in the rhTGF-β stimulated group, an increase of 59%. The volume of bone in the gap was increased by 16% in rhTGF-β1-stimulated TCP-coated implants, but this difference was not significant. Mechanical push-out tests showed no difference in fixation of the implant between the two groups. Our study suggests that rhTGF-β1 adsorbed on TCP-ceramic-coated implants can enhance bone ongrowth


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 6 | Pages 900 - 905
1 Aug 2003
Shardlow DL Stone MH Ingham E Fisher J

Proponents of the biological theory of aseptic loosening have in recent years tended to concentrate on the production and distribution of particulate ultra-high-molecular-weight polyethylene (UHMWPE) debris around the potential joint space. However, mechanical loading of cemented implants with the differing elastic moduli of metal stems, polymethylmethacrylate (PMMA) cement and bone can result in relative micromotion, implying the potential for production of metal and PMMA particles from the stem-cement interface by fretting wear. In order to investigate the production and biological reactivity of debris from this interface, PMMA and metal particulate debris was produced by sliding wear of PMMA pins containing barium sulphate and zirconium dioxide against a Vaquasheened stainless steel counterface. This debris was characterised by SEM, energy-dispersive analysis by X-ray (EDAX) and image analysis, then added to cell cultures of a human monocytic cell line, U937, and stimulation of pro-osteolytic cytokines measured by ELISA. Large quantities of PMMA cement debris were generated by the sliding wear of PMMA pins against Vaquasheened stainless steel plates in the method developed for this study. Both cements stimulated the release of pro-osteolytic TNFα from the U937 monocytic cell line, in a dose-dependent fashion. There was a trend towards greater TNFα release with Palacos cement than CMW cement at the same dose. Palacos particles also caused significant release of IL-6, another pro-osteolytic cytokine, while CMW did not. The particulate cement debris produced did not stimulate the release of GM-CSF or IL1β from the U937 cells. These results may explain the cytokine pathway responsible for bone resorption caused by particulate PMMA debris. Radio-opaque additives are of value in surgical practice and clinical studies to quantify the relevance of these in vitro findings are required before the use of cement containing radio-opacifier is constrained


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 3 | Pages 377 - 383
1 May 1995
Ryd L Albrektsson B Carlsson L Dansgard F Herberts P Lindstrand A Regner L Toksvig-Larsen S

The tibial components in 143 patients with total knee replacements performed before 1988 were assessed for micromotion using roentgen stereophotogrammetric analysis (RSA) over a period of 13 years. The fixation of the prostheses remained clinically sound in all cases, although revision had been required for other reasons in seven. In a second group taken from all cases with RSA available on our full database to 1990, 15 tibial components had been followed by RSA from the insertion until, 1 to 11 years after the initial arthroplasty, they were revised for mechanical loosening of the tibial component; 12 of these comprised all the loosenings in the base group, thus making a total of 155 consecutive cases, while an additional three were inserted after the base material had been compiled. The mean migration in the first group was about 1 mm at one year, but subsequent migration was slower, reaching a mean of about 1.5 mm after ten years. About one-third migrated continuously throughout follow-up, while two-thirds ceased to migrate after one to two years. In the revision group, 14 components had migrated continuously and at one year significantly more than those in the first group. One revision case lacked the crucial one-year follow-up and could not be classified. These findings suggest that mechanical loosening begins early in the postoperative period. Clinical symptoms which necessitate revision, seen at this stage in 20% of abnormally migrating tibial components, may not appear until up to ten years after the operation.(ABSTRACT TRUNCATED AT 250 WORDS)


The Bone & Joint Journal
Vol. 102-B, Issue 12 | Pages 1654 - 1661
1 Dec 2020
Perelgut ME Polus JS Lanting BA Teeter MG

Aims

The direct anterior (DA) approach has been associated with rapid patient recovery after total hip arthroplasty (THA) but may be associated with more frequent femoral complications including implant loosening. The objective of this study was to determine whether the addition of a collar to the femoral stem affects implant migration, patient activity, and patient function following primary THA using the DA approach.

Methods

Patients were randomized to either a collared (n = 23) or collarless (n = 26) cementless femoral stem implanted using the DA approach. Canal fill ratio (CFR) was measured on the first postoperative radiographs. Patients underwent a supine radiostereometric analysis (RSA) exam postoperatively on the day of surgery and at two, four, six, 12, 26, and 52 weeks postoperatively. Patient-reported outcome measures (Western Ontario and McMaster Universities Osteoarthritis (WOMAC) Index, the 12-item Short Form Health Survey Mental and Physical Score, and University of California, Los Angeles (UCLA) Activity Score) were measured preoperatively and at each post-surgery clinic visit. Activity and function were also measured as the weekly average step count recorded by an activity tracker, and an instrumented timed up-and-go (TUG) test in clinic, respectively.


Bone & Joint Research
Vol. 8, Issue 10 | Pages 489 - 494
1 Oct 2019
Klasan A Bäumlein M Dworschak P Bliemel C Neri T Schofer MD Heyse TJ

Objectives

Periprosthetic femoral fractures (PFFs) have a higher incidence with cementless stems. The highest incidence among various cementless stem types was observed with double-wedged stems. Short stems have been introduced as a bone-preserving alternative with a higher incidence of PFF in some studies. The purpose of this study was a direct load-to-failure comparison of a double-wedged cementless stem and a short cementless stem in a cadaveric fracture model.

Methods

Eight hips from four human cadaveric specimens (age mean 76 years (60 to 89)) and eight fourth-generation composite femurs were used. None of the cadaveric specimens had compromised quality (mean T value 0.4 (-1.0 to 5.7)). Each specimen from a pair randomly received either a double-wedged stem or a short stem. A materials testing machine was used for lateral load-to-failure test of up to a maximal load of 5000 N.


Aims

Mobile-bearing unicompartmental knee arthroplasty (UKA) with a flat tibial plateau has not performed well in the lateral compartment, leading to a high rate of dislocation. For this reason, the Domed Lateral UKA with a biconcave bearing was developed. However, medial and lateral tibial plateaus have asymmetric anatomical geometries, with a slightly dished medial and a convex lateral plateau. Therefore, the aim of this study was to evaluate the extent at which the normal knee kinematics were restored with different tibial insert designs using computational simulation.

Methods

We developed three different tibial inserts having flat, conforming, and anatomy-mimetic superior surfaces, whereas the inferior surface in all was designed to be concave to prevent dislocation. Kinematics from four male subjects and one female subject were compared under deep knee bend activity.


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 832 - 837
1 Jul 2020
Dover C Kuiper JH Craig P Shaylor P

Aims

We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice.

Methods

Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time.


Bone & Joint Research
Vol. 8, Issue 7 | Pages 349 - 356
1 Jul 2019
Starlinger J Kaiser G Thomas A Sarahrudi K

Objectives

The osteoprotegerin (OPG) and receptor activator of nuclear factor kappa-B ligand (RANKL) balance is of the utmost importance in fracture healing. The aim of this study was therefore to investigate the impact of nonosteogenic factors on OPG and RANKL levels.

Methods

Serum obtained from 51 patients with long bone fractures was collected over 48 weeks. The OPG and serum sRANKL (soluble RANKL) concentrations were measured using enzyme-linked immunosorbent assay (ELISA). Smoking habit, diabetes, and alcohol consumption were recorded.


The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 709 - 715
1 Jun 2020
Abdelsamie KR Elhawary I Ali H Ali M EL-Shafie M Dominic Meek RM

Aims

Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs.

Methods

A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 107 - 115
1 Jun 2020
Tetreault MW Perry KI Pagnano MW Hanssen AD Abdel MP

Aims

Metaphyseal fixation during revision total knee arthroplasty (TKA) is important, but potentially difficult when using historical designs of cone. Material and manufacturing innovations have improved the size and shape of the cones which are available, and simplified the required bone preparation. In a large series, we assessed the implant survivorship, radiological results, and clinical outcomes of new porous 3D-printed titanium metaphyseal cones featuring a reamer-based system.

Methods

We reviewed 142 revision TKAs in 139 patients using 202 cones (134 tibial, 68 femoral) which were undertaken between 2015 and 2016. A total of 60 involved tibial and femoral cones. Most cones (149 of 202; 74%) were used for Type 2B or 3 bone loss. The mean age of the patients was 66 years (44 to 88), and 76 (55 %) were female. The mean body mass index (BMI) was 34 kg/m2 (18 to 60). The patients had a mean of 2.4 (1 to 8) previous operations on the knee, and 68 (48%) had a history of prosthetic infection. The mean follow-up was 2.4 years (2 to 3.6).


The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 699 - 708
1 Jun 2020
Nieuwenhuijse MJ Vehmeijer SBW Mathijsen NMC Keizer SB

Aims

Short, bone-conserving femoral components are increasingly used in total hip arthroplasty (THA). They are expected to allow tissue-conserving implantation and to render future revision surgery more straightforward but the long-term data on such components is limited. One such component is the global tissue-sparing (GTS) stem. Following the model for stepwise introduction of new orthopaedic implants, we evaluated early implant fixation and clinical outcome of this novel short-stem THA and compared it to that of a component with established good long-term clinical outcome.

Methods

In total, 50 consecutive patients ≤ 70 years old with end-stage symptomatic osteo-arthritis were randomized to receive THA with the GTS stem or the conventional Taperloc stem using the anterior supine intermuscular approach by two experienced hip surgeons in two hospitals in the Netherlands. Primary outcome was implant migration. Patients were followed using routine clinical examination, patient reported outcome using Harris Hip Score (HHS), Hip Disability And Osteoarthritis Outcome Score (HOOS), EuroQol five-dimension questionnaire (EQ5D), and Roentgen Stereophotogrammetric Analysis (RSA) at three, six, 12, and 24 months. This study evaluated the two-year follow-up results.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1107 - 1114
1 Sep 2019
Uy M Wang J Horner NS Bedi A Leroux T Alolabi B Khan M

Aims

The aim of this study was to evaluate the differences in revision and complication rates, functional outcomes, and radiological outcomes between cemented and press-fit humeral stems in primary anatomical total shoulder arthroplasty (TSA).

Materials and Methods

A comprehensive systematic review and meta-analysis was conducted searching for studies that included patients who underwent primary anatomical TSA for primary osteoarthritis or rheumatoid arthritis.


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 530 - 538
1 Apr 2020
Rollick NC Gadinsky NE Klinger CE Kubik JF Dyke JP Helfet DL Wellman DS

Aims

Dual plating of distal femoral fractures with medial and lateral implants has been performed to improve construct mechanics and alignment, in cases where isolated lateral plating would be insufficient. This may potentially compromise vascularity, paradoxically impairing healing. This study investigates effects of single versus dual plating on distal femoral vascularity.

Methods

A total of eight cadaveric lower limb pairs were arbitrarily assigned to either 1) isolated lateral plating, or 2) lateral and medial plating of the distal femur, with four specimens per group. Contralateral limbs served as matched controls. Pre- and post-contrast MRI was performed to quantify signal intensity enhancement in the distal femur. Further evaluation of intraosseous vascularity was done with barium sulphate infusion with CT scan imaging. Specimens were then injected with latex medium and dissection was completed to assess extraosseous vasculature.


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 365 - 370
1 Mar 2020
Min KS Fox HM Bedi A Walch G Warner JJP

Aims

Patient-specific instrumentation has been shown to increase a surgeon’s precision and accuracy in placing the glenoid component in shoulder arthroplasty. There is, however, little available information about the use of patient-specific planning (PSP) tools for this operation. It is not known how these tools alter the decision-making patterns of shoulder surgeons. The aim of this study was to investigate whether PSP, when compared with the use of plain radiographs or select static CT images, influences the understanding of glenoid pathology and surgical planning.

Methods

A case-based survey presented surgeons with a patient’s history, physical examination, and, sequentially, radiographs, select static CT images, and PSP with a 3D imaging program. For each imaging modality, the surgeons were asked to identify the Walch classification of the glenoid and to propose the surgical treatment. The participating surgeons were grouped according to the annual volume of shoulder arthroplasties that they undertook, and responses were compared with the recommendations of two experts.


The Bone & Joint Journal
Vol. 102-B, Issue 2 | Pages 232 - 238
1 Feb 2020
Javed S Hadi S Imam MA Gerogiannis D Foden P Monga P

Aims

Accurate measurement of the glenoid version is important in performing total shoulder arthroplasty (TSA). Our aim was to evaluate the Ellipse method, which involves formally defining the vertical mid-point of the glenoid prior to measuring the glenoid version and comparing it with the ‘classic’ Friedman method.

Methods

This was a retrospective study which evaluated 100 CT scans for patients who underwent a primary TSA. The glenoid version was measured using the Friedman and Ellipse methods by two senior observers. Statistical analyses were performed using the paired t-test for significance and the Bland-Altman plot for agreement.


The Bone & Joint Journal
Vol. 102-B, Issue 2 | Pages 191 - 197
1 Feb 2020
Gabor JA Padilla JA Feng JE Schnaser E Lutes WB Park KJ Incavo S Vigdorchik J Schwarzkopf R

Aims

Although good clinical outcomes have been reported for monolithic tapered, fluted, titanium stems (TFTS), early results showed high rates of subsidence. Advances in stem design may mitigate these concerns. This study reports on the use of a current monolithic TFTS for a variety of indications.

Methods

A multi-institutional retrospective study of all consecutive total hip arthroplasty (THA) and revision total hip arthroplasty (rTHA) patients who received the monolithic TFTS was conducted. Surgery was performed by eight fellowship-trained arthroplasty surgeons at four institutions. A total of 157 hips in 153 patients at a mean follow-up of 11.6 months (SD7.8) were included. Mean patient age at the time of surgery was 67.4 years (SD 13.3) and mean body mass index (BMI) was 28.9 kg/m2 (SD 6.5). Outcomes included intraoperative complications, one-year all-cause re-revisions, and subsidence at postoperative time intervals (two weeks, six weeks, six months, nine months, and one year).


The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1442 - 1448
1 Nov 2018
Hipfl C Janz V Löchel J Perka C Wassilew GI

Aims

Severe acetabular bone loss and pelvic discontinuity (PD) present particular challenges in revision total hip arthroplasty. To deal with such complex situations, cup-cage reconstruction has emerged as an option for treating this situation. We aimed to examine our success in using this technique for these anatomical problems.

Patients and Methods

We undertook a retrospective, single-centre series of 35 hips in 34 patients (seven male, 27 female) treated with a cup-cage construct using a trabecular metal shell in conjunction with a titanium cage, for severe acetabular bone loss between 2011 and 2015. The mean age at the time of surgery was 70 years (42 to 85) and all patients had an acetabular defect graded as Paprosky Type 2C through to 3B, with 24 hips (69%) having PD. The mean follow-up was 47 months (25 to 84).


The Bone & Joint Journal
Vol. 101-B, Issue 3 | Pages 311 - 316
1 Mar 2019
Löchel J Janz V Hipfl C Perka C Wassilew GI

Aims

The use of trabecular metal (TM) shells supported by augments has provided good mid-term results after revision total hip arthroplasty (THA) in patients with a bony defect of the acetabulum. The aim of this study was to assess the long-term implant survivorship and radiological and clinical outcomes after acetabular revision using this technique.

Patients and Methods

Between 2006 and 2010, 60 patients (62 hips) underwent acetabular revision using a combination of a TM shell and augment. A total of 51 patients (53 hips) had complete follow-up at a minimum of seven years and were included in the study. Of these patients, 15 were men (29.4%) and 36 were women (70.6%). Their mean age at the time of revision THA was 64.6 years (28 to 85). Three patients (5.2%) had a Paprosky IIA defect, 13 (24.5%) had a type IIB defect, six (11.3%) had a type IIC defect, 22 (41.5%) had a type IIIA defect, and nine (17%) had a type IIIB defect. Five patients (9.4%) also had pelvic discontinuity.


Bone & Joint Research
Vol. 8, Issue 12 | Pages 593 - 600
1 Dec 2019
Koh Y Lee J Lee H Kim H Chung H Kang K

Aims

Commonly performed unicompartmental knee arthroplasty (UKA) is not designed for the lateral compartment. Additionally, the anatomical medial and lateral tibial plateaus have asymmetrical geometries, with a slightly dished medial plateau and a convex lateral plateau. Therefore, this study aims to investigate the native knee kinematics with respect to the tibial insert design corresponding to the lateral femoral component.

Methods

Subject-specific finite element models were developed with tibiofemoral (TF) and patellofemoral joints for one female and four male subjects. Three different TF conformity designs were applied. Flat, convex, and conforming tibial insert designs were applied to the identical femoral component. A deep knee bend was considered as the loading condition, and the kinematic preservation in the native knee was investigated.


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1192 - 1198
1 Oct 2019
Sköldenberg OG Rysinska AD Chammout G Salemyr M Mukka SS Bodén H Eisler T

Aims

Radiostereometric analysis (RSA) studies of vitamin E-doped, highly crosslinked polyethylene (VEPE) liners show low head penetration rates in cementless acetabular components. There is, however, currently no data on cemented VEPE acetabular components in total hip arthroplasty (THA). The aim of this study was to evaluate the safety of a new cemented VEPE component, compared with a conventional polyethylene (PE) component regarding migration, head penetration, and clinical results.

Patients and Methods

We enrolled 42 patients (21 male, 21 female) with osteoarthritis and a mean age of 67 years (sd 5), in a double-blinded, noninferiority, randomized controlled trial. The subjects were randomized in a 1:1 ratio to receive a reverse hybrid THA with a cemented component of either argon-gas gamma-sterilized PE component (controls) or VEPE, with identical geometry. The primary endpoint was proximal implant migration of the component at two years postoperatively measured with RSA. Secondary endpoints included total migration of the component, penetration of the femoral head into the component, and patient-reported outcome measurements.


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 365 - 371
1 Apr 2019
Nam D Salih R Nahhas CR Barrack RL Nunley RM

Aims

Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell (vs a monoblock acetabular component) have the advantage of allowing supplementary screw fixation, but the potential for corrosion between the liner and acetabulum has raised concerns. While DM prostheses have shown improved stability in patients deemed ‘high-risk’ for dislocation undergoing total hip arthroplasty (THA), their performance in young, active patients has not been reported. This study’s purpose was to assess clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) in young, active patients receiving a modular DM acetabulum and recently introduced titanium, proximally coated, tapered femoral stem design.

Patients and Methods

This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m2 and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene mobile bearing, and cementless titanium femoral stem for their primary THA. Patients with a history of renal disease and metal hardware elsewhere in the body were excluded. A total of 43 patients (30 male, 13 female; mean age 52.6 years (sd 6.5)) were enrolled. All patients had a minimum of two years’ clinical follow-up. Patient-reported outcome measures, whole blood metal ion levels (ug/l), and periprosthetic femoral BMD were measured at baseline, as well as at one and two years postoperatively. Power analysis indicated 40 patients necessary to demonstrate a five-fold increase in cobalt levels from baseline (alpha = 0.05, beta = 0.80). A mixed model with repeated measures was used for statistical analysis.


Bone & Joint Research
Vol. 8, Issue 3 | Pages 136 - 145
1 Mar 2019
Cerquiglini A Henckel J Hothi H Allen P Lewis J Eskelinen A Skinner J Hirschmann MT Hart AJ

Objectives

The Attune total knee arthroplasty (TKA) has been used in over 600 000 patients worldwide. Registry data show good clinical outcome; however, concerns over the cement-tibial interface have been reported. We used retrieval analysis to give further insight into this controversial topic.

Methods

We examined 12 titanium (Ti) PFC Sigma implants, eight cobalt-chromium (CoCr) PFC Sigma implants, eight cobalt-chromium PFC Sigma rotating platform (RP) implants, and 11 Attune implants. We used a peer-reviewed digital imaging method to quantify the amount of cement attached to the backside of each tibial tray. We then measured: 1) the size of tibial tray thickness, tray projections, peripheral lips, and undercuts; and 2) surface roughness (Ra) on the backside and keel of the trays. Statistical analyses were performed to investigate differences between the two designs.