Aims. The aim of this study was to assess the clinical and radiological outcomes of an antiprotrusio acetabular cage (APC) when used in the surgical treatment of periacetabular bone metastases. Methods. This retrospective cohort study using a prospectively collected database involved 56 patients who underwent acetabular reconstruction for periacetabular bone metastases or haematological malignancy using a single APC between January 2009 and 2020. The mean follow-up was 20 months (1 to 143). The primary outcome measure was implant survival. Postoperative radiographs were analyzed for loosening and failure. Patient and implant survival were assessed using a competing risk analysis. Secondary parameters included primary malignancy, oncological treatment, surgical factors, length of stay in hospital, and postoperative complications. Results. A total of 33 patients (59%) died during the study period at a mean of 15 months postoperatively (1 to 63). No patient had radiological evidence of loosening or failure. Acetabular component survival was 100%. Three patients (5.4%) had further surgery; one (1.8%) underwent revision of the
Aims. Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model. Methods. We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the
Dislocation of the hip remains a major complication after periacetabular tumour resection and endoprosthetic reconstruction. The position of the acetabular component is an important modifiable factor for surgeons in determining the risk of postoperative dislocation. We investigated the significance of horizontal, vertical, and sagittal displacement of the hip centre of rotation (COR) on postoperative dislocation using a CT-based 3D model, as well as other potential risk factors for dislocation. A total of 122 patients who underwent reconstruction following resection of periacetabular tumour between January 2011 and January 2020 were studied. The risk factors for dislocation were investigated with univariate and multivariate logistic regression analysis on patient-specific, resection-specific, and reconstruction-specific variables.Aims
Methods
Osteoarticular reconstruction of the distal femur in childhood has the advantage of preserving the tibial physis. However, due to the small size of the distal femur, matching the host bone with an osteoarticular allograft is challenging. In this study, we compared the outcomes and complications of a resurfaced allograft-prosthesis composite (rAPC) with those of an osteoarticular allograft to reconstruct the distal femur in children. A retrospective analysis of 33 skeletally immature children with a malignant tumour of the distal femur, who underwent resection and reconstruction with a rAPC (n = 15) or osteoarticular allograft (n = 18), was conducted. The median age of the patients was ten years (interquartile range (IQR) 9 to 11) in the osteoarticular allograft group and nine years (IQR 8 to 10) in the rAPC group (p = 0.781). The median follow-up of the patients was seven years (IQR 4 to 8) in the osteoarticular allograft group and six years (IQR 3 to 7) in the rAPC group (p = 0.483). Limb function was evaluated using the Musculoskeletal Tumor Society (MSTS) score.Aims
Methods
Proximal femoral endoprosthetic replacements (PFEPRs) are the most common reconstruction option for osseous defects following primary and metastatic tumour resection. This study aimed to compare the rate of implant failure between PFEPRs with monopolar and bipolar hemiarthroplasties and acetabular arthroplasties, and determine the optimum articulation for revision PFEPRs. This is a retrospective review of 233 patients who underwent PFEPR. The mean age was 54.7 years (SD 18.2), and 99 (42.5%) were male. There were 90 patients with primary bone tumours (38.6%), 122 with metastatic bone disease (52.4%), and 21 with haematological malignancy (9.0%). A total of 128 patients had monopolar (54.9%), 74 had bipolar hemiarthroplasty heads (31.8%), and 31 underwent acetabular arthroplasty (13.3%).Aims
Methods
We report our early experience with a new peri-acetabular reconstruction endoprosthesis used for pelvic reconstruction after tumour resection. The outcome of 21 patients who underwent limb salvage following type II pelvic resection and reconstruction using the peri-acetabular reconstruction prosthesis between 2000 and 2006 was retrospectively reviewed. This prosthesis was designed to use the remaining part of the ilium to support a horizontally placed acetabular component secured with internal fixation and bone cement. Into this device a constrained acetabular liner is positioned which is articulates with a conventional
There is an increased risk of dislocation of the hip after the resection of a periacetabular tumour and endoprosthetic reconstruction of the defect in the hemipelvis. The aim of this study was to determine the rate and timing of dislocation and to identify its risk factors. To determine the dislocation rate, we conducted a retrospective single-institution study of 441 patients with a periacetabular tumour who had undergone a standard modular hemipelvic endoprosthetic reconstruction between 2003 and 2019. After excluding ineligible patients, 420 patients were enrolled. Patient-specific, resection-specific, and reconstruction-specific variables were studied using univariate and multivariate analyses.Aims
Methods
We report our results in 24 children with malignant primary bone tumours of the distal femur treated with a Stanmore extendible endoprosthesis (SEER). This consists of a
We aimed to examine the long-term mechanical survivorship, describe the modes of all-cause failure, and identify risk factors for mechanical failure of all-polyethylene tibial components in endoprosthetic reconstruction. This is a retrospective database review of consecutive endoprosthetic reconstructions performed for oncological indications between 1980 and 2019. Patients with all-polyethylene tibial components were isolated and analyzed for revision for mechanical failure. Outcomes included survival of the all-polyethylene tibial component, revision surgery categorized according to the Henderson Failure Mode Classification, and complications and functional outcome, as assessed by the Musculoskeletal Tumor Society (MSTS) score at the final follow-up.Aims
Methods
The early mortality in patients with hip fractures from bony metastases is unknown. The objectives of this study were to quantify 30- and 90-day mortality in patients with proximal femoral metastases, and to create a mortality prediction tool based on biomarkers associated with early death. This was a retrospective cohort study of consecutive patients referred to the orthopaedic department at a UK trauma centre with a proximal femoral metastasis (PFM) over a seven-year period (2010 to 2016). The study group were compared to a matched control group of non-metastatic hip fractures. Minimum follow-up was one year.Aims
Methods
Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique. A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1).Aims
Patients and Methods
The aim of this study was to evaluate the prosthesis characteristics and associated conditions that may modify the survival of total femoral endoprosthetic replacements (TFEPR). In all, 81 patients treated with TFEPR from 1976 to 2017 were retrospectively evaluated and failures were categorized according to the Henderson classification. There were 38 female patients (47%) and 43 male patients (53%) with a mean age at diagnosis of 43 years (12 to 86). The mean follow-up time was 10.3 years (0 to 31.7). A survival analysis was performed followed by univariate and multivariate Cox regression to identify independent implant survival factors.Aims
Patients and Methods
Pelvic reconstruction after the resection of a tumour around
the acetabulum is a challenging procedure due to the complex anatomy
and biomechanics. Several pelvic endoprostheses have been introduced,
but the rates of complication remain high. Our aim was to review
the use of a stemmed acetabular pedestal cup in the management of
these patients. The study involved 48 patients who underwent periacetabular reconstruction
using a stemmed pedestal cup (Schoellner cup; Zimmer Biomet Inc.,
Warsaw, Indiana) between 2000 and 2013. The indications for treatment included
a primary bone tumour in 27 patients and metastatic disease in 21
patients. The mean age of the patients at the time of surgery was
52 years (16 to 83).Aims
Patients and Methods
Peri-acetabular tumour resections and their subsequent
reconstruction are among the most challenging procedures in orthopaedic
oncology. Despite the fact that a number of different pelvic endoprostheses
have been introduced, rates of complication remain high and long-term
results are mostly lacking. In this retrospective study, we aimed to evaluate the outcome
of reconstructing a peri-acetabular defect with a pedestal cup endoprosthesis
after a type 2 or type 2/3 internal hemipelvectomy. A total of 19 patients (11M:8F) with a mean age of 48 years (14
to 72) were included, most of whom had been treated for a primary
bone tumour (n = 16) between 2003 and 2009. After a mean follow-up
of 39 months (28 days to 8.7 years) seven patients had died. After
a mean follow-up of 7.9 years (4.3 to 10.5), 12 patients were alive,
of whom 11 were disease-free. Complications occurred in 15 patients.
Three had recurrent dislocations and three experienced aseptic loosening.
There were no mechanical failures. Infection occurred in nine patients,
six of whom required removal of the prosthesis. Two patients underwent
hindquarter amputation for local recurrence. The implant survival rate at five years was 50% for all reasons,
and 61% for non-oncological reasons. The mean Musculoskeletal Tumor
Society score at final follow-up was 49% (13 to 87). Based on these poor results, we advise caution if using the pedestal
cup for reconstruction of a peri-acetabular tumour resection. Cite this article:
Giant cell tumour is the most common aggressive
benign tumour of the musculoskeletal system and has a high rate of
local recurrence. When it occurs in proximity to the hip, reconstruction
of the joint is a challenge. Options for reconstruction after wide
resection include the use of a megaprosthesis or an allograft-prosthesis
composite. We performed a clinical and radiological study to evaluate
the functional results of a proximal femoral allograft-prosthesis
composite in the treatment of proximal femoral giant cell tumour
after wide resection. This was an observational study, between 2006
and 2012, of 18 patients with a mean age of 32 years (28 to 42)
and a mean follow-up of 54 months (18 to 79). We achieved excellent
outcomes using Harris Hip Score in 13 patients and a good outcome
in five. All allografts united. There were no complications such
as infection, failure, fracture or resorption of the graft, or recurrent
tumour. Resection and reconstruction of giant cell tumours with
proximal femoral allograft–prosthesis composite is a better option
than using a prosthesis considering preservation of bone stock and excellent
restoration of function. A good result requires demanding bone banking techniques, effective
measures to prevent infection and stability at the allograft-host
junction. Cite this article:
We report our early experience with the use of
a new prosthesis, the Modular Hemipelvic Prosthesis II, for reconstruction
of the hemipelvis after resection of a primary malignant peri-acetabular
tumour involving the sacroiliac joint. We retrospectively reviewed the outcome of 17 patients who had
undergone resection of a pelvic tumour and reconstruction with this
prosthesis between July 2002 and July 2010. One patient had a type I+II+III+IV resection (ilium + peri-acetabulum
+ pubis/ischium + sacrum) and 16 had a type I+II+IV resection (ilium
+ acetabulum + sacrum). The outcome was assessed at a mean follow-up
of 33 months (15 to 59). One patient was alive with disease, 11
were alive without disease and five had died of disease. The overall
five-year survival rate was 62.4%. Six patients had a local recurrence.
The mean Musculoskeletal Tumour Society score was 58% (33 to 77).
Deep infection occurred in two patients, problems with wound healing
in five and dislocation in one. For patients with a primary malignant peri-acetabular sarcoma
involving the sacroiliac joint, we believe that this new prosthesis
is a viable option for reconstruction of the bony defect left following
resection of the tumour. It results in a satisfactory functional
outcome with an acceptable rate of complications. Cite this article:
We report our early experience with the use of a non-invasive distal femoral expandable endoprosthesis in seven skeletally immature patients with osteosarcoma of the distal femur. The patients had a mean age of 12.1 years (9 to 15) at the time of surgery. The prosthesis was lengthened at appropriate intervals in outpatient clinics, without anaesthesia, using the principle of electromagnetic induction. The patients were functionally evaluated using the Musculoskeletal Tumour Society scoring system. The mean follow-up was 20.2 months (14 to 30). The prostheses were lengthened by a mean of 25 mm (4.25 to 55) and maintained a mean knee flexion of 110° (100° to 120°). The mean Musculoskeletal Tumour Society score was 68% (11 to 29). Complications developed in two patients; one developed a flexion deformity of 25° at the knee joint, which was subsequently overcome and one died of disseminated disease. The early results from patients treated with this device have been encouraging. The implant avoids multiple surgical procedures, general anaesthesia and assists in maintaining leg-length equality.
We report positive and negative factors associated with the most commonly-used methods of reconstruction after pathological fracture of the proximal femur. The study was based on 142 patients treated surgically for 146 metastatic lesions between 1996 and 2003. The local rate of failure was 10.3% (15 of 146). Of 37 operations involving osteosynthetic devices, six failed (16.2%) compared with nine (8.3%) in 109 operations involving endoprostheses. Of nine cases of prosthetic failure, four were due to periprosthetic fractures and three to recurrent dislocation. In the osteosynthesis group, three (13.6%) of 22 reconstruction nails failed. The two-year risk of re-operation after any type of osteosynthesis was 0.35 compared with 0.18 after any type of endoprosthetic reconstruction (p = 0.07). Endoprosthetic reconstructions are preferable to the use of reconstruction nails and other osteosynthetic devices for the treatment of metastatic lesions in the proximal third of the femur.
In skeletally immature patients, resection of
bone tumours and reconstruction of the lower limb often results
in leg-length discrepancy. The Stanmore non-invasive extendible
endoprosthesis, which uses electromagnetic induction, allows post-operative
lengthening without anaesthesia. Between 2002 and 2009, 55 children
with a mean age of 11.4 years (5 to 16) underwent reconstruction
with this prosthesis; ten patients (18.2%) died of disseminated
disease and one child underwent amputation due to infection. We
reviewed 44 patients after a mean follow-up of 41.2 months (22 to
104). The mean Musculoskeletal Tumor Society score was 24.7 (8 to
30) and the Toronto Extremity Salvage score was 92.3% (55.2% to
99.0%). There was no local recurrence of tumour. Complications developed
in 16 patients (29.1%) and ten (18.2%) underwent revision. The mean length gained per patient was 38.6 mm (3.5 to 161.5),
requiring a mean of 11.3 extensions (1 to 40), and ten component
exchanges were performed in nine patients (16.4%) after attaining
the maximum lengthening capacity of the implant. There were 11 patients
(20%) who were skeletally mature at follow-up, ten of whom had equal
leg lengths and nine had a full range of movement of the hip and
knee. This is the largest reported series using non-invasive extendible
endoprostheses after excision of primary bone tumours in skeletally
immature patients. The technique produces a good functional outcome,
with prevention of limb-length discrepancy at skeletal maturity.
Endoprosthetic replacement of the pelvis is one of the most challenging types of limb-salvage surgery, with a high rate of complications. In an attempt to reduce this and build greater versatility into the reconstruction process, a new type of pelvic endoprosthesis was developed in 2003, based on the old McKee-Farrar prosthesis. This study reviews the outcomes in 27 patients who had an ice-cream cone pelvic prosthesis inserted at two different specialist bone tumour centres in the United Kingdom over the past six years. The indications for treatment included primary bone tumours in 19 patients and metastatic disease in two, and six implants were inserted following failure of a previous pelvic reconstruction. Most of the patients had a P2+P3 resection as classified by Enneking, and most had resection of the ilium above the sciatic notch. The mean age of the patients at operation was 49 years (13 to 81). Complications occurred in ten patients (37.0%), of which dislocation was the most common, affecting four patients (14.8%). A total of three patients (11.1%) developed a deep infection around the prosthesis but all were successfully controlled by early intervention and two patients (7.4%) developed a local recurrence, at the same time as widespread metastases appeared. In one patient the prosthesis was removed for severe pain. This method of treatment is still associated with high morbidity, but early results are promising. Complications are diminishing with increasing experience.
We treated 98 patients with peri-acetabular tumours by resection and reconstruction with a custom-made pelvic endoprosthesis. The overall survival of the patients was 67% at five years, 54% at ten years and 51% at 30 years. One or more complications occurred in 58.1% of patients (54), of which infection was the most common, affecting 30% (28 patients). The rate of local recurrence was 31% (29 patients) after a mean follow-up of 71 months (11 to 147). Dislocation occurred in 20% of patients (19). Before 1996 the rate was 40.5% (17 patients) but this was reduced to 3.9% (two patients) with the introduction of a larger femoral head. There were six cases of palsy of the femoral nerve with recovery in only two. Revision or excision arthroplasty was performed in 23.7% of patients (22), principally for uncontrolled infection or aseptic loosening. Higher rates of death, infection and revision occurred in men. This method of treatment is still associated with high morbidity. Patients should be carefully selected and informed of this pre-operatively.
Pathological fractures due to metastasis with destruction of the acetabulum and central dislocation of the hip present a difficult surgical challenge. We describe a series using a single technique in which a stable and long-lasting reconstruction was obtained using standard primary hip replacement implants augmented by strong, fully-threaded steel rods with cement and steel mesh, where required. Between 1997 and 2006, 19 patients with a mean age of 66 years (48 to 83) were treated using a modified Harrington technique. Acetabular destruction was graded as Harrington class II in six cases and class III in 13. Reconstruction was achieved using three 6.5 mm rods inserted through a separate incision in the iliac crest followed by augmentation with cement and a conventional cemented Charnley or Exeter primary hip replacement. There were no peri-operative deaths. At the final follow-up (mean 25 months (5 to 110)) one rod had fractured and one construct required revision. Of the 18 patients who did not require revision, 13 had died. The mean time to death was 16 months (5 to 55). The mean follow-up of the five survivors was 31 months (18 to 47). There were no cases of dislocation, deep infection or injury to a nerve, the blood vessels or the bladder.
We set out to determine the impact of surgery on quality of life and function in patients who had undergone surgery for symptomatic peri-acetabular metastases. From a prospective database we retrospectively reviewed 46 consecutive patients who had been treated operatively between June 2003 and June 2009. The mean age of the patients was 56.4 years (20 to 73) and the mean post-operative follow-up was 19.2 months (4 to 70). Functional evaluation and quality-of-life assessments were performed. At the most recent follow-up, 26 patients (56.5%) were alive. Their median survival time was 25.0 months. Ten major postoperative complications had occurred in eight patients (17.4%). The mean post-operative Musculoskeletal Tumor Society score (MSTS 93) was 56.3% (6.7% to 90.0%). Improvement in the Eastern Cooperative Oncology Group (ECOG) performance status was seen in 32 patients (69.6%). On the European Organisation for Research and Treatment of Cancer core quality-of-life questionnaire (QLQ-C30) measure of global health status there was a statistically significant improvement from the patients’ pre-operative status (42.8 ( Surgery for patients with peri-acetabular metastases reduces pain and improves their quality of life, and has a low rate of surgical complications.
We investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the distal femur after resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database, review of the design of the implant and case records. In total, 335 patients underwent a distal femoral replacement, 162 having a fixed-hinge design and 173 a rotating-hinge. The median age of the patients was 24 years (interquartile range 17 to 48). A total of 192 patients remained alive with a mean follow-up of 12 years (5 to 30). The risk of revision for any reason was 17% at five years, 33% at ten years and 58% at 20 years. Aseptic loosening was the main reason for revision of the fixed-hinge knees while infection and fracture of the stem were the most common for the rotating-hinge implant. The risk of revision for aseptic loosening was 35% at ten years with the fixed-hinge knee, which has, however, been replaced by the rotating-hinge knee with a hydroxyapatite collar. The overall risk of revision for any reason fell by 52% when the rotating-hinge implant was used. Improvements in the design of distal femoral endoprostheses have significantly decreased the need for revision operations, but infection remains a serious problem. We believe that a cemented, rotating-hinge prosthesis with a hydroxyapatite collar offers the best chance of long-term survival of the prosthesis.
We undertook a cemental unipolar proximal femoral endoprosthetic replacement in 131 patients with a mean age of 50 years (2 to 84). Primary malignant tumours were present in 54 patients and 67 had metastatic disease. In addition, eight patients had either lymphoma or myeloma and two had non-oncological disorders. The mean follow-up was 27 months (0 to 180). An acetabular revision was required later in 14 patients, 12 of whom had been under the age of 21 years at the time of insertion of their original prosthesis. The risk of acetabular revision in patients over 21 years of age was 8% at five years compared with 36% in those aged under 21 years. All the unipolar hips in this younger age group required revision within 11 years of the initial operation. We conclude that unipolar replacement should not be used in younger patients and should be avoided in patients with a life expectancy of more than five years.
We undertook a retrospective review of 33 patients who underwent total femoral endoprosthetic replacement as limb salvage following excision of a malignant bone tumour. In 22 patients this was performed as a primary procedure following total femoral resection for malignant disease. Revision to a total femoral replacement was required in 11 patients following failed segmental endoprosthetic or allograft reconstruction. There were 33 patients with primary malignant tumours, and three had metastatic lesions. The mean age of the patients was 31 years (5 to 68). The mean follow-up was 4.2 years (9 months to 16.4 years). At five years the survival of the implants was 100%, with removal as the endpoint and 56% where the endpoint was another surgical intervention. At five years the patient survival was 32%. Complications included dislocation of the hip in six patients (18%), local recurrence in three (9%), peri-prosthetic fracture in two and infection in one. One patient subsequently developed pulmonary metastases. There were no cases of aseptic loosening or amputation. Four patients required a change of bushings. The mean Musculoskeletal Tumour Society functional outcome score was 67%, the mean Harris Hip Score was 70, and the mean Oxford Knee Score was 34. Total femoral endoprosthetic replacement can provide good functional outcome without compromising patient survival, and in selected cases provides an effective alternative to amputation.
We have investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the proximal tibia following resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database. A total of 194 patients underwent a proximal tibial replacement, with 95 having a fixed-hinge design and 99 a rotating-hinge with a hydroxyapatite collar; their median age was 21.5 years (10 to 74). At a mean follow-up of 14.7 years (5 to 29), 115 patients remain alive. The risk of revision for any reason in the fixed-hinge group was 32% at five years, 61% at ten years and 75% at 15 and 20 years, and in the rotating-hinge group 12% at five years, 25% at ten years and 30% at 15 years. Aseptic loosening was the most common reason for revision in the fixed-hinge knees, fracture of the implant in the early design of rotating hinges and infection in the current version. The risk of revision for aseptic loosening in the fixed-hinge knees was 46% at ten years. This was reduced to 3% in the rotating-hinge knee with a hydroxyapatite collar. The cemented, rotating hinge design currently offers the best chance of long-term survival of the prosthesis.
A retrospective analysis was performed of eight patients with an open triradiate cartilage, who underwent resection for osteosarcoma and reconstruction of the proximal femur with a hemiarthroplasty, in order to identify changes of acetabular development. An analysis of the centre-edge angle, teardrop-to-medial prosthesis distance, superior joint space, teardrop-to-superior prosthesis distance, degree of lateral translation, and arthritic changes, was performed on serial radiographs. The median age at the time of the initial surgery was 11 years (5 to 14). All patients developed progressive superior and lateral migration of the prosthetic femoral head. Following hemiarthroplasty in the immature acetabulum, the normal deepening and enlargement of the acetabulum is arrested. The degree of superior and lateral migration of the prosthetic head depends on the age at diagnosis and the length of follow-up.
We reviewed retrospectively the results in 211 consecutive patients who had undergone limb salvage for bone neoplasia with endoprosthetic reconstruction of the proximal femur (96), distal femur (78), proximal tibia (30) and total femur (7). Their mean age was 50 years (11 to 86) and the mean follow-up period was 37.3 months (1 to 204). A total of 35 (16.6%) prostheses failed. Overall, implant survival was 78% (95% confidence interval (CI) 0.29 to 0.54) at five years, 60% (95% CI 0.93 to 2.35) at ten years and 60% (95% CI 1.27 to 3.88) at 15 years. Survivorship of the limb was 97.6% (95% CI 1.73 to 3.35) at ten years. The gender, age, diagnosis and location of the tumour were not prognostic variables for failure. Modular endoprosthetic replacement in the lower limb is a durable long-term reconstructive option, with the implants generally outlasting the patient.