We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur. A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the femoral revisions were amenable to standard cemented or uncemented stem fixation. Patient and disease characteristics, surgical history, final ambulatory statusAims
Methods
Previously, we showed that case-specific non-linear
finite element (FE) models are better at predicting the load to failure
of metastatic
Surgical limb sparing for knee-bearing paediatric bone sarcoma is considered to have a clinically significant influence on postoperative function due to complications and leg-length discrepancies. However, researchers have not fully evaluated the long-term postoperative functional outcomes. Therefore, in this study, we aimed to elucidate the risk factors and long-term functional prognosis associated with paediatric limb-sparing surgery. We reviewed 40 patients aged under 14 years who underwent limb-sparing surgery for knee bone sarcoma (15 cases in the proximal tibia and 25 in the distal femur) between January 2000 and December 2013, and were followed up for a minimum of five years. A total of 35 patients underwent reconstruction using artificial materials, and five underwent biological reconstruction. We evaluated the patients’ postoperative complications, survival rate of reconstruction material, and limb, limb function, and leg-length discrepancy at the final follow-up, as well as the risk factors for each.Aims
Methods
Following the resection of an extensive amount of bone in the
treatment of a tumour, the residual segment may be insufficient
to accept a standard length intramedullary cemented stem. Short-stemmed
endoprostheses conceivably have an increased risk of aseptic loosening.
Extra-cortical plates have been added to minimise this risk by supplementing
fixation. The aim of this study was to investigate the survivorship
of short-stemmed endoprostheses and extra-cortical plates. The study involved 37 patients who underwent limb salvage surgery
for a primary neoplasm of bone between 1998 and 2013. Endoprosthetic
replacement involved the proximal humerus in nine, the proximal
femur in nine, the distal femur in 13 and the proximal tibia in
six patients. There were 12 primary (32%) and 25 revision procedures (68%).
Implant survivorship was compared with matched controls. The amount
of bone that was resected was >
70% of its length and statistically
greater than the standard control group at each anatomical site.Aims
Patients and Methods
We retrospectively reviewed 101 consecutive patients
with 114 femoral tumours treated by massive bone allograft at our
institution between 1986 and 2005. There were 49 females and 52
males with a mean age of 20 years (4 to 74). At a median follow-up
of 9.3 years (2 to 19.8), 36 reconstructions (31.5%) had failed.
The allograft itself failed in 27 reconstructions (24%). Mechanical complications such as delayed union, fracture and
failure of fixation were studied. The most adverse factor on the
outcome was the use of intramedullary nails, followed by post-operative
chemotherapy, resection length >
17 cm and age >
18 years at the
time of intervention. The simultaneous use of a vascularised fibular
graft to protect the allograft from mechanical complications improved
the outcome, but the use of intramedullary cementing was not as
successful. In order to improve the strength of the reconstruction and to
advance the biology of host–graft integration, we suggest avoiding
the use of intramedullary nails and titanium plates, but instead
using stainless steel plates, as these gave better results. The
use of a supplementary vascularised fibular graft should be strongly
considered in adult patients with resection >
17 cm and in those
who require post-operative chemotherapy.
Deformity of the proximal femur in fibrous dysplasia
leads to deviation of the mechanical axis of the hip, which may lead
to the development of secondary osteoarthritis (OA). This study
investigated the prevalence and predisposing factors for the development
of OA in patients with fibrous dysplasia of the proximal femur.
We reviewed the records of 209 patients from our institutional database
with fibrous dysplasia of the proximal femur, investigating possible predisposing
factors including patient demographics, the extent of the coxa vara
deformity, the presence of peri-articular disease, and the overall
burden of skeletal disease. Of the 209 patients, 24 (12%) had radiological
evidence of OA in the ipsilateral hip. The prevalence was significantly
higher in patients with polyostotic fibrous dysplasia compared with
those with monostotic disease (p <
0.001). In a subgroup analysis of
patients with polyostotic disease, the extent of deformity (quantified
using the neck–shaft angle), and the presence of peri-articular
disease (whether in the head of the femur or the acetabulum) were
significant predictors of osteoarthritis (neck–shaft angle likelihood
ratio (LR) = 0.847 per 1° increase, p = 0.004; presence of lesion
in the head of the femur LR = 9.947, p = 0.027; presence of lesion
in the acetabulum LR = 11.231, p = 0.014). Our data suggest that patients with polyostotic fibrous dysplasia
have a high risk of developing secondary OA of the hips. This risk
is higher in patients with peri-articular disease, and those with
a more severe deformity of proximal femur. Cite this article:
Cancellous allograft bone chips are commonly
used in the reconstruction of defects in bone after removal of benign tumours.
We investigated the MRI features of grafted bone chips and their
change over time, and compared them with those with recurrent tumour.
We retrospectively reviewed 66 post-operative MRIs from 34 patients
who had undergone curettage and grafting with cancellous bone chips
to fill the defect after excision of a tumour. All grafts showed
consistent features at least six months after grafting: homogeneous
intermediate or low signal intensities with or without scattered
hyperintense foci (speckled hyperintensities) on T1 images; high
signal intensities with scattered hypointense foci (speckled hypointensities)
on T2 images, and peripheral rim enhancement with or without central
heterogeneous enhancements on enhanced images. Incorporation of
the graft occurred from the periphery to the centre, and was completed
within three years. Recurrent lesions consistently showed the same signal
intensities as those of pre-operative MRIs of the primary lesions.
There were four misdiagnoses, three of which were chondroid tumours. We identified typical MRI features and clarified the incorporation
process of grafted cancellous allograft bone chips. The most important
characteristics of recurrent tumours were that they showed the same
signal intensities as the primary tumours. It might sometimes be
difficult to differentiate grafted cancellous allograft bone chips
from a recurrent chondroid tumour. Cite this article:
We present a retrospective study of patients suffering from a variety of benign tumours in whom external fixators were used to treat deformity and limb-length discrepancy, and for the reconstruction of bone defects. A total of 43 limbs in 31 patients (12 male and 19 female) with a mean age of 14 years (2 to 54) were treated. The diagnosis was Ollier’s disease in 12 limbs, fibrous dysplasia in 11, osteochondroma in eight, giant cell tumour in five, osteofibrous dysplasia in five and non-ossifying fibroma in two. The lesions were treated in the tibia in 19 limbs, in the femur in 16, and in the forearm in eight. The Ilizarov frame was used in 25 limbs, the Taylor Spatial Frame in seven, the Orthofix fixator in six, the Monotube in four and the Heidelberg fixator in one. The mean follow-up was 72 months (22 to 221). The mean external fixation period was 168 days (71 to 352). The mean external fixation index was 42 days/cm (22.2 to 102.0) in the 22 patients who required limb lengthening. The mean correction angle for those with angular deformity was 23° (7° to 45°). At final follow-up all patients had returned to normal activities. Four patients required a second operation for recurrent deformity of further limb lengthening. Local recurrence occurred in one patient, requiring further surgery.
We investigated the clinical outcome of internal
fixation for pathological fracture of the femur after primary excision of
a soft-tissue sarcoma that had been treated with adjuvant radiotherapy. A review of our database identified 22 radiation-induced fractures
of the femur in 22 patients (seven men, 15 women). We noted the
mechanism of injury, fracture pattern and any complications after
internal fixation, including nonunion, hardware failure, secondary
fracture or deep infection. The mean age of the patients at primary excision of the tumour
was 58.3 years (39 to 86). The mean time from primary excision to
fracture was 73.2 months (2 to 195). The mean follow-up after fracture
fixation was 65.9 months (12 to 205). Complications occurred in
19 patients (86%). Nonunion developed in 18 patients (82%), of whom
11 had a radiological nonunion at 12 months, five a nonunion and
hardware failure and two an infected nonunion. One patient developed
a second radiation-associated fracture of the femur after internal
fixation and union of the initial fracture. A total of 13 patients
(59%) underwent 24 revision operations. Internal fixation of a pathological fracture of the femur after
radiotherapy for a soft-tissue sarcoma has an extremely high rate
of complication and requires specialist attention. Cite this article:
We describe the management of nonunion combined with limb-length discrepancy following vascularised fibular grafting for the reconstruction of long-bone defects in the lower limb after resection of a tumour in skeletally immature patients. We operated on nine patients with a mean age of 13.1 years (10.5 to 14.5) who presented with a mean limb-length discrepancy of 7 cm (4 to 9) and nonunion at one end of a vascularised fibular graft, which had been performed previously, to reconstruct a bone defect after resection of an osteosarcoma. Reconstruction was carried out using a ring fixator secured with correction by half pins of any malalignment, compression of the site of nonunion and lengthening through a metaphyseal parafocal osteotomy without bone grafting. The expected limb-length discrepancy at maturity was calculated using the arithmetic method. Solid union and the intended leg length were achieved in all the patients. Excessive scarring and the distorted anatomy from previous surgery in these patients required other procedures to be performed with minimal exposures and dissection in order to avoid further compromise to the vascularity of the graft or damage to neurovascular structures. The methods which we chose were simple and effective in addressing these complex problems.