We report a prospective study of the use of intramedullary bone blocks to improve the fixation of a matt-finish femoral stem in Charnley low-friction arthroplasties. There were 379 patients (441 hips), but at a minimum follow-up of ten years there were 258 arthroplasties in 221 patients including some which had been revised. The mean age at surgery was 41 years (17 to 51) and the mean follow-up was 13.4 years (1 to 20 including the early revisions). Nine stems (3.5%) had been revised for aseptic loosening, but there were no stem fractures. Survivorship of stems was 99.2% at ten years and 94.35% at 15 and 20 years. We found that the patient’s gender, the position of the stem and the experience of the surgeon all influenced the outcome. Our findings suggest that using our method of
In order to improve the
Both the femoral and tibial component are usually cemented at revision total knee arthroplasty (rTKA), while stems can be added with either cemented or press-fit (hybrid) fixation. The aim of this study was to compare the long-term stability of rTKA with cemented and press-fitted stems, using radiostereometric analysis (RSA). This is a follow-up of a randomized controlled trial, initially involving 32 patients, of whom 19 (nine cemented, ten hybrid) were available for follow-up ten years postoperatively, when further RSA measurements were made. Micromotion of the femoral and tibial components was assessed using model-based RSA software (RSAcore). The clinical outcome was evaluated using the Knee Society Score (KSS), the Knee injury and Osteoarthritis Outcome Score (KOOS), and visual analogue scale (pain and satisfaction).Aims
Methods
To achieve the functional benefits of the direct anterior (DA) approach and the fixation benefits of cemented replacement, this study combined the two techniques posing the following questions: does the limited access of the DA approach adversely affect the cement technique?; and does such a cementing technique reduce the incidence of cementless complications? A consecutive series of 341 patients (360 hips) receiving the DA approach between 2016 and 2018 were reviewed. There were 203 cementless stems and 157 cemented stems. Mean age was 75 years (70 to 86) in the cementless group and 76 years (52 to 94) in the cemented group, with 239 (70%) females in the whole series. Femoral complications were compared between the two groups. Mean follow-up was 1.5 years (0.1 to 4.4) for patients in the cementless group and 1.3 years (0.0 to 3.9) for patients in the cemented group.Aims
Methods
The excellent long-term results for the first-generation Charnley stem may not apply to later versions with flanges. It seems possible that the early design functioned as a taper-slip system, as accepted in the Exeter prosthesis. Comparison with the requirements for the alternative composite-beam system for the femoral component shows considerable differences that have important implications. These include design, surface finish, cementing technique and the interpretation of radiological signs of loosening. A distinction should be made between the requirements for the successful use of the two different engineering systems.s
The calcar femorale is a vertical plate of bone lying deep to the lesser trochanter and is formed as a result of traction of the iliopsoas which separates the femoral cortex into two distinct layers, the calcar femorale and the medial femoral cortex. They fuse together proximally to form the medial femoral neck. A stem placed centrally will abut against the calcar femorale with little or no space for cement. Clearing of the calcar will offer space for a cement layer, which will support the stem proximally on the posterior aspect. We compared two consecutive groups of Charnley low-friction arthroplasties, with and without clearing of the calcar. In 330 patients who had an arthroplasty without clearing the calcar, there were ten revisions for aseptic loosening of the stem and six other stems were considered ‘definitely loose’, giving a rate of failure of 4.8%. In 111 patients in whom the calcar was cleared there was only one revision for aseptic loosening and no stems were classed as ‘definitely loose’, giving a rate of failure of 0.9%. Survivorship analysis has again shown the need for long-term follow-up; the differences became clear after ten years but because of the relatively small numbers, statistical analysis is not yet applicable. We now clear the calcar femorale routinely and advocate optimal access to the medullary canal and insertion of the stem in the area of the piriform fossa.
A leading problem with cementless hip replacement is thigh pain, probably due to some degree of loosening. This is most pronounced during activities such as stair climbing or getting up from a chair. Our study compares the immediate fixation of three stems in widespread clinical use, using a cadaver test simulation in which the femur was horizontal, and cyclic forces acted vertically downward on the prosthetic head with gradual increments of load. We implanted and tested 18 pairs of femurs, making analyses between stem types by right to left comparisons. We found no statistically significant differences in loosening between cementless AML and PCA prostheses, but cemented stems withstood greater loads than uncemented stems by a factor of five or more.
Aims. To investigate the effect of polyethylene manufacturing characteristics and irradiation dose on the survival of cemented and reverse hybrid total hip arthroplasties (THAs). Methods. In this registry study, data from the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man (NJR) were linked with manufacturing data supplied by manufacturers. The primary endpoint was revision of any component. Cox proportional hazard regression was a primary analytic approach adjusting for competing risk of death, patient characteristics, head composition, and
Aims. Osteoporosis can determine surgical strategy for total hip arthroplasty (THA), and perioperative fracture risk. The aims of this study were to use hip CT to measure femoral bone mineral density (BMD) using CT X-ray absorptiometry (CTXA), determine if systematic evaluation of preoperative femoral BMD with CTXA would improve identification of osteopenia and osteoporosis compared with available preoperative dual-energy X-ray absorptiometry (DXA) analysis, and determine if improved recognition of low BMD would affect the use of cemented
Aims. Revision total hip arthroplasty in patients with Vancouver type B3 fractures with Paprosky type IIIA, IIIB, and IV femoral defects are difficult to treat. One option for Paprovsky type IIIB and IV defects involves modular cementless, tapered, revision femoral components in conjunction with distal interlocking screws. The aim of this study was to analyze the rate of reoperations and complications and union of the fracture, subsidence of the stem, mortality, and the clinical outcomes in these patients. Methods. A total of 46 femoral components in patients with Vancouver B3 fractures (23 with Paprosky type IIIA, 19 with type IIIB, and four with type IV defects) in 46 patients were revised with a transfemoral approach using a modular, tapered, cementless revision Revitan curved femoral component with distal cone-in-cone fixation and prospectively followed for a mean of 48.8 months (SD 23.9; 24 to 112). The mean age of the patients was 80.4 years (66 to 100). Additional distal interlocking was also used in 23 fractures in which distal cone-in-cone fixation in the isthmus was < 3 cm. Results. One patient (2.2%) died during the first postoperative year. After six months, 43 patients (93.5%) had osseous, and three had fibrous consolidation of the fracture and the bony flap, 42 (91.3%) had bony ingrowth and four had stable fibrous
Aims. This study evaluates risk factors influencing fracture characteristics for postoperative periprosthetic femoral fractures (PFFs) around cemented stems in total hip arthroplasty. Methods. Data were collected for PFF patients admitted to eight UK centres between 25 May 2006 and 1 March 2020. Radiographs were assessed for Unified Classification System (UCS) grade and AO/OTA type. Statistical comparisons investigated relationships by age, gender, and
Aims. The aim of this study was to describe the management and associated outcomes of patients sustaining a femoral hip periprosthetic fracture (PPF) in the UK population. Methods. This was a multicentre retrospective cohort study including adult patients who presented to 27 NHS hospitals with 539 new PPFs between 1 January 2018 and 31 December 2018. Data collected included: management strategy (operative and nonoperative), length of stay, discharge destination, and details of post-treatment outcomes (reoperation, readmission, and 30-day and 12-month mortality). Descriptive analysis by fracture type was performed, and predictors of PPF management and outcomes were assessed using mixed-effects logistic regression. Results. In all, 417 fractures (77%) were managed operatively and 122 (23%) conservatively. The median time to surgery was four days (interquartile range (IQR) 2 to 7). Of those undergoing surgery, 246 (59%) underwent revision and/or fixation and 169 (41%) fixation alone. The surgical strategy used differed by Unified Classification System for PPF type, with the highest rate of revision in B2/B3 fractures (both 77%, 176/228 and 24/31, respectively) and the highest rate of fixation alone in B1- (55/78; 71%) and C-type (49/65; 75%) fractures. Cemented
Aims. Metaphyseal cones with cemented stems are frequently used in revision total knee arthroplasty (TKA). However, if the diaphysis has been previously violated, the resultant sclerotic canal can impair cemented
Aims. BoneMaster is a thin electrochemically applied hydroxyapatite (HA) coating for orthopaedic implants that is quickly resorbed during osseointegration. Early stabilization is a surrogacy marker of good survival of femoral stems. The hypothesis of this study was that a BoneMaster coating yields a fast early and lasting fixation of stems. Methods. A total of 53 patients were randomized to be treated using Bi-Metric cementless femoral stems with either only a porous titanium plasma-sprayed coating (P group) or a porous titanium plasma-sprayed coating with an additional BoneMaster coating (PBM group). The patients were examined with radiostereometry until five years after surgery. Results. At three months, the mean total translation (TT) was 0.95 mm (95% confidence interval (CI) 0.68 to 1.22) in the P group and 0.57 mm (95% CI 0.31 to 0.83) in the PBM group (p = 0.047). From two to five years, the TT increased by a mean of 0.14 mm (95% CI 0.03 to 0.25) more in the P group than in the PBM group (p = 0.021). In osteopenic patients (n = 20), the mean TT after three months was 1.61 mm (95% CI 1.03 to 2.20) in the P group and 0.73 mm (95% CI 0.25 to 1.21) in the PBM group (p = 0.023). After 60 months, the mean TT in osteopenic patients was 1.87 mm (95% CI 1.24 to 2.50) in the P group and 0.82 mm (95% CI 0.30 to 1.33) in the PBM group (p = 0.011). Conclusion. There was less early and midterm migration of cementless stems with BoneMaster coating compared with those with only a porous titanium plasma-sprayed coating. Although a BoneMaster coating seems to be important for
Aims. We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur. Methods. A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the femoral revisions were amenable to standard cemented or uncemented
Aims. The goals of this study were to define the risk factors, characteristics,
and chronology of fractures in 5417 revision total hip arthroplasties
(THAs). . Patients and Methods. From our hospital’s prospectively collected database we identified
all patients who had undergone a revision THA between 1969 and 2011
which involved the femoral stem. The patients’ medical records and
radiographs were examined and the relevant data extracted. Post-operative
periprosthetic fractures were classified using the Vancouver system.
A total of 5417 revision THAs were identified. Results. There were 668 intra-operative fractures, giving an incidence
of 12%. Fractures were three times more common with uncemented stems
(19%) than with cemented stems (6%) (p <
0.001). The incidence
of intra-operative femoral fracture varied by uncemented stem type:
fully-coated (20%); proximally-coated (19%); modular fluted tapered
(16%) (p <
0.05). Most fractures occurred during the insertion
of the femoral component (35%). One-third involved the diaphysis
and 26% were of the calcar: 69% were undisplaced. There were 281 post-operative fractures of the femur (20-year
probability = 11%). There was no difference in risk for cemented
and uncemented stems. Post-operative fractures were more common
in men <
70 years (p = 0.02). Periprosthetic fractures occurred
earlier after uncemented revision of the femoral component, but
later after a cemented revision. The most common fracture type was
a Vancouver B. 1. (31%). Of all post-operative fractures,
24% underwent open reduction and internal fixation and 15% revision
arthroplasty. Conclusion. In revision THA, intra-operative fractures occurred three times
more often with an uncemented stem. Many were undisplaced diaphyseal
fractures treated with cerclage fixation. . While the risk of post-operative fracture is similar between
uncemented and cemented components, they occur at notably different
times depending on the type of
We undertook a review of the literature relating to the two basic stem designs in use in cemented hip replacement, namely loaded tapers or force-closed femoral stems, and the composite beam or shape-closed designs. The associated
The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore Modular Individualised Lower Extremity System (SMILES) rotating hinge implant in revision total knee arthroplasty (rTKA) at a tertiary unit. It is hypothesized that this fully cemented construct leads to satisfactory clinical outcomes. A retrospective consecutive study of all patients who underwent a rTKA using the fully cemented SMILES rotating hinge prosthesis between 2005 to 2018. Outcome measures included aseptic loosening, reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both prospectively collected local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.Aims
Methods
Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes. This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively.Aims
Methods
The aim of this study was to evaluate the survival of a collarless, straight, hydroxyapatite-coated femoral stem in total hip arthroplasty (THA) at a minimum follow-up of 20 years. We reviewed the results of 165 THAs using the Omnifit HA system in 138 patients, performed between August 1993 and December 1999. The mean age of the patients at the time of surgery was 46 years (20 to 77). Avascular necrosis was the most common indication for THA, followed by ankylosing spondylitis and primary osteoarthritis. The mean follow-up was 22 years (20 to 31). At 20 and 25 years, 113 THAs in 91 patients and 63 THAs in 55 patients were available for review, respectively, while others died or were lost to follow-up. Kaplan-Meier analysis was performed to evaluate the survival of the stem. Radiographs were reviewed regularly, and the stability of the stem was evaluated using the Engh classification.Aims
Methods
The December 2023 Hip & Pelvis Roundup360 looks at: Early hip fracture surgery is safe for patients on direct oral anticoagulants; Time to return to work by occupational class after total hip or knee arthroplasty; Is there a consensus on air travel following hip and knee arthroplasty?; Predicting whether patients will achieve minimal clinically important differences following hip or knee arthroplasty; High-dose dual-antibiotic-loaded cement for hip hemiarthroplasty in the UK (WHiTE 8): a randomized controlled trial; Vitamin E – a positive thing in your poly?; Hydroxapatite-coated femoral stems: is there a difference in fixation?
When performing revision total hip arthroplasty using diaphyseal-engaging titanium tapered stems (TTS), the recommended 3 to 4 cm of stem-cortical diaphyseal contact may not be available. In challenging cases such as these with only 2 cm of contact, can sufficient axial stability be achieved and what is the benefit of a prophylactic cable? This study sought to determine, first, whether a prophylactic cable allows for sufficient axial stability when the contact length is 2 cm, and second, if differing TTS taper angles (2° vs 3.5°) impact these results. A biomechanical matched-pair cadaveric study was designed using six matched pairs of human fresh cadaveric femora prepared so that 2 cm of diaphyseal bone engaged with 2° (right femora) or 3.5° (left femora) TTS. Before impaction, three matched pairs received a single 100 lb-tensioned prophylactic beaded cable; the remaining three matched pairs received no cable adjuncts. Specimens underwent stepwise axial loading to 2600 N or until failure, defined as stem subsidence > 5 mm.Aims
Methods
The August 2023 Hip & Pelvis Roundup360 looks at: Using machine learning to predict venous thromboembolism and major bleeding events following total joint arthroplasty; Antibiotic length in revision total hip arthroplasty; Preoperative colonization and worse outcomes; Short stem cemented total hip arthroplasty; What are the outcomes of one- versus two-stage revisions in the UK?; To cement or not to cement? The best approach in hemiarthroplasty; Similar re-revisions in cemented and cementless femoral revisions for periprosthetic femoral fractures in total hip arthroplasty; Are hip precautions still needed?
Obtaining solid implant fixation is crucial in revision total knee arthroplasty (rTKA) to avoid aseptic loosening, a major reason for re-revision. This study aims to validate a novel grading system that quantifies implant fixation across three anatomical zones (epiphysis, metaphysis, diaphysis). Based on pre-, intra-, and postoperative assessments, the novel grading system allocates a quantitative score (0, 0.5, or 1 point) for the quality of fixation achieved in each anatomical zone. The criteria used by the algorithm to assign the score include the bone quality, the size of the bone defect, and the type of fixation used. A consecutive cohort of 245 patients undergoing rTKA from 2012 to 2018 were evaluated using the current novel scoring system and followed prospectively. In addition, 100 first-time revision cases were assessed radiologically from the original cohort and graded by three observers to evaluate the intra- and inter-rater reliability of the novel radiological grading system.Aims
Methods
Cementing in arthroplasty for hip fracture is associated with improved postoperative function, but may have an increased risk of early mortality compared to uncemented fixation. Quantifying this mortality risk is important in providing safe patient care. This study investigated the association between cement use in arthroplasty and mortality at 30 days and one year in patients aged 50 years and over with hip fracture. This retrospective cohort study used linked data from the Australian Hip Fracture Registry and the National Death Index. Descriptive analysis and Kaplan-Meier survival curves tested the unadjusted association of mortality between cemented and uncemented procedures. Multilevel logistic regression, adjusted for covariates, tested the association between cement use and 30-day mortality following arthroplasty. Given the known institutional variation in preference for cemented fixation, an instrumental variable analysis was also performed to minimize the effect of unknown confounders. Adjusted Cox modelling analyzed the association between cement use and mortality at 30 days and one year following surgery.Aims
Methods
Despite higher rates of revision after total hip arthroplasty (THA) being reported for uncemented stems in patients aged > 75 years, they are frequently used in this age group. Increased mortality after cemented fixation is often used as a justification, but recent data do not confirm this association. The aim of this study was to investigate the influence of the design of the stem and the type of fixation on the rate of revision and immediate postoperative mortality, focusing on the age and sex of the patients. A total of 333,144 patients with primary osteoarthritis (OA) of the hip who underwent elective THA between November 2012 and September 2022, using uncemented acetabular components without reconstruction shells, from the German arthroplasty registry were included in the study. The revision rates three years postoperatively for four types of stem (uncemented, uncemented with collar, uncemented short, and cemented) were compared within four age groups: < 60 years (Young), between 61 and 70 years (Mid-I), between 71 and 80 years (Mid-II), and aged > 80 years (Old). A noninferiority analysis was performed on the most frequently used designs of stem.Aims
Methods
Refobacin Bone Cement R and Palacos Overall, 75 patients were included in the study and 71 were available at two years postoperatively. Prior to surgery, they were randomized to one of the three combinations studied: Palacos cement with use of the Optivac mixing system, Refobacin with use of the Optivac system, and Refobacin with use of the Optipac system. Cemented MS30 stems and cemented Exceed acetabular components were used in all hips. Postoperative radiographs were used to assess the quality of the cement mantle according to Barrack et al, and the position and migration of the femoral stem. Harris Hip Score, Oxford Hip Score, Forgotten Joint Score, and University of California, Los Angeles Activity Scale were collected.Aims
Methods
The aim of this investigation was to compare risk of infection in both cemented and uncemented hemiarthroplasty (HA) as well as in total hip arthroplasty (THA) following femoral neck fracture. Data collection was performed using the German Arthroplasty Registry (EPRD). In HA and THA following femoral neck fracture, fixation method was divided into cemented and uncemented prostheses and paired according to age, sex, BMI, and the Elixhauser Comorbidity Index using Mahalanobis distance matching.Aims
Methods
Breast cancer survivors have known risk factors that might influence the results of total hip arthroplasty (THA) or total knee arthroplasty (TKA). This study evaluated clinical outcomes of patients with breast cancer history after primary THA and TKA. Our total joint registry identified patients with breast cancer history undergoing primary THA (n = 423) and TKA (n = 540). Patients were matched 1:1 based upon age, sex, BMI, procedure (hip or knee), and surgical year to non-breast cancer controls. Mortality, implant survival, and complications were assessed via Kaplan-Meier methods. Clinical outcomes were evaluated via Harris Hip Scores (HHSs) or Knee Society Scores (KSSs). Mean follow-up was six years (2 to 15).Aims
Methods
The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail). Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using radiostereometric analysis (RSA), change of bone mineral density (BMD) around the stem, the development of radiolucent lines, and additional patient-reported outcome measures (PROMs).Aims
Methods
Porous metaphyseal cones can be used for fixation in revision total knee arthroplasty (rTKA) and complex TKAs. This metaphyseal fixation has led to some surgeons using shorter cemented stems instead of diaphyseal engaging cementless stems with a potential benefit of ease of obtaining proper alignment without being beholden to the diaphysis. The purpose of this study was to evaluate short term clinical and radiographic outcomes of a series of TKA cases performed using 3D-printed metaphyseal cones. A retrospective review of 86 rTKAs and nine complex primary TKAs, with an average age of 63.2 years (SD 8.2) and BMI of 34.0 kg/m2 (SD 8.7), in which metaphyseal cones were used for both femoral and tibial fixation were compared for their knee alignment based on the type of stem used. Overall, 22 knees had cementless stems on both sides, 52 had cemented stems on both sides, and 15 had mixed stems. Postoperative long-standing radiographs were evaluated for coronal and sagittal plane alignment. Adjusted logistic regression models were run to assess malalignment hip-knee-ankle (HKA) alignment beyond ± 3° and sagittal alignment of the tibial and femoral components ± 3° by stem type.Aims
Methods
The aim of this study was to compare the cost-effectiveness of cemented hemiarthroplasty (HA) versus hydroxyapatite-coated uncemented HA for the treatment of displaced intracapsular hip fractures in older adults. A within-trial economic evaluation was conducted based on data collected from the World Hip Trauma Evaluation 5 (WHiTE 5) multicentre randomized controlled trial in the UK. Resource use was measured over 12 months post-randomization using trial case report forms and participant-completed questionnaires. Cost-effectiveness was reported in terms of incremental cost per quality-adjusted life year (QALY) gained from the NHS and personal social service perspective. Methodological uncertainty was addressed using sensitivity analysis, while decision uncertainty was represented graphically using confidence ellipses and cost-effectiveness acceptability curves.Aims
Methods
The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression.Aims
Methods
Total hip arthroplasty (THA) is a very successful and cost-effective operation, yet debate continues about the optimum fixation philosophy in different age groups. The concept of the 'cementless paradox' and the UK 'Getting it Right First Time' initiative encourage increased use of cemented fixation due to purported lower revision rates, especially in elderly patients, and decreased cost. In a high-volume, tertiary referral centre, we identified 10,112 THAs from a prospectively collected database, including 1,699 cemented THAs, 5,782 hybrid THAs, and 2,631 cementless THAs. The endpoint was revision for any reason. Secondary analysis included examination of implant survivorship in patients aged over 70 years, over 75 years, and over 80 years at primary THA.Aims
Methods
The aim of this study was to determine whether total hip arthroplasty (THA) for chronic hip pain due to unilateral primary osteoarthritis (OA) has a beneficial effect on cognitive performance. A prospective cohort study was conducted with 101 patients with end-stage hip OA scheduled for THA (mean age 67.4 years (SD 9.5), 51.5% female (n = 52)). Patients were assessed at baseline as well as after three and months. Primary outcome was cognitive performance measured by d2 Test of Attention at six months, Trail Making Test (TMT), FAS-test, Rivermead Behavioural Memory Test (RBMT; story recall subtest), and Rey-Osterrieth Complex Figure Test (ROCF). The improvement of cognitive performance was analyzed using repeated measures analysis of variance.Aims
Methods
Although the short stem concept in hip arthroplasty procedure shows acceptable clinical performance, we sometimes get unexplainable radiological findings. The aim of this retrospective study was to evaluate changes of radiological findings up to three years postoperatively, and to assess any potential contributing factors on such radiological change in a Japanese population. This is a retrospective radiological study conducted in Japan. Radiological assessment was done in accordance with predetermined radiological review protocol. A total of 241 hips were included in the study and 118 hips (49.0%) revealed radiological change from immediately after surgery to one year postoperatively; these 118 hips were eligible for further analyses. Each investigator screened whether either radiolucent lines (RLLs), cortical hypertrophy (CH), or atrophy (AT) appeared or not on the one-year radiograph. Further, three-year radiographs of eligible cases were reviewed to determine changes such as, disappeared (D), improved (I), stable (S), and progression (P). Additionally, bone condensation (BC) was assessed on the three-year radiograph.Aims
Methods
Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model. We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation.Aims
Methods
Varus-valgus constrained (VVC) devices are typically used in revision settings, often with stems to mitigate the risk of aseptic loosening. However, in at least one system, the VVC insert is compatible with the primary posterior-stabilized (PS) femoral component, which may be an option in complex primary situations. We sought to determine the implant survivorship, radiological and clinical outcomes, and complications when this VVC insert was coupled with a PS femur without stems in complex primary total knee arthroplasties (TKAs). Through our institution’s total joint registry, we identified 113 primary TKAs (103 patients) performed between 2007 and 2017 in which a VVC insert was coupled with a standard cemented PS femur without stems. Mean age was 68 years (SD 10), mean BMI was 32 kg/m2 (SD 7), and 59 patients (50%) were male. Mean follow-up was four years (2 to 10).Aims
Methods
Stemmed tibial components are frequently used in revision total knee arthroplasty (TKA). The purpose of this study was to evaluate patient satisfaction, overall pain, and diaphyseal tibial pain in patients who underwent revision TKA with cemented or uncemented stemmed tibial components. This is a retrospective cohort study involving 110 patients with revision TKA with cemented versus uncemented stemmed tibial components. Patients who underwent revision TKA with stemmed tibial components over a 15-year period at a single institution with at least two-year follow-up were assessed. Pain was evaluated through postal surveys. There were 63 patients with cemented tibial stems and 47 with uncemented stems. Radiographs and Knee Society Scores were used to evaluate for objective findings associated with pain or patient dissatisfaction. Postal surveys were analyzed using Fisher’s exact test and the independent-samples Aims
Methods
Femoral bone preparation using compaction technique has been shown to preserve bone and improve implant fixation in animal models. No long-term clinical outcomes are available. There are no significant long-term differences between compaction and broaching techniques for primary total hip arthroplasty (THA) in terms of migration, clinical, and radiological outcomes. A total of 28 patients received one-stage bilateral primary THA with cementless femoral stems (56 hips). They were randomized to compaction on one femur and broaching on the contralateral femur. Overall, 13 patients were lost to the ten-year follow-up leaving 30 hips to be evaluated in terms of stem migration (using radiostereometry), radiological changes, Harris Hip Score, Oxford Hip Score, and complications.Aims
Methods
At the end of this short study we have to sum up our views about the use of the acrylic prosthesis for arthroplasty of the hip. Some fatalities and a proportion of bad or poor results make this operation one to be undertaken only by surgeons well trained in the surgery of the hip and only on patients who really need it. However, the tolerance of the tissues to acrylic resin and the
This aim of this study was to assess the reliability and validity of the Unified Classification System (UCS) for postoperative periprosthetic femoral fractures (PFFs) around cemented polished taper-slip (PTS) stems. Radiographs of 71 patients with a PFF admitted consecutively at two centres between 25 February 2012 and 19 May 2020 were collated by an independent investigator. Six observers (three hip consultants and three trainees) were familiarized with the UCS. Each PFF was classified on two separate occasions, with a mean time between assessments of 22.7 days (16 to 29). Interobserver reliability for more than two observers was assessed using percentage agreement and Fleiss’ kappa statistic. Intraobserver reliability between two observers was calculated with Cohen kappa statistic. Validity was tested on surgically managed UCS type B PFFs where stem stability was documented in operation notes (n = 50). Validity was assessed using percentage agreement and Cohen kappa statistic between radiological assessment and intraoperative findings. Kappa statistics were interpreted using Landis and Koch criteria. All six observers were blinded to operation notes and postoperative radiographs.Aims
Methods
Between 1990 and 1992, we implanted 71 hybrid alumina-on-alumina hip arthroplasties in 62 consecutive patients under the age of 55 years, with a mean age of 46 years at surgery. There were 56 primary and 15 secondary procedures. The prostheses involved a cemented titanium alloy stem, a 32 mm alumina head, and a press-fit metal-backed socket with an alumina insert. Three patients (four hips) died from unrelated causes. Four hips had revision surgery for either deep infection, unexplained persistent pain, fracture of the alumina head, or aseptic loosening of the socket. The nine-year survival rate was 93.7% with revision for any cause as the end-point and 98.4% with revision for aseptic loosening as the end-point. The outcome in the surviving patients (50 patients, 57 hips) with a minimum five-year follow-up (mean eight years) was excellent in 47 hips (82.5%), very good in eight (14%), good in one and fair in one. A thin, partial, lucent line, mainly in zone III was present in 38% of the sockets and one socket had a complete lucency less than 1 mm thick. One stem had isolated femoral osteolysis. There was no detectable component migration nor acetabular osteolysis. This hybrid arthroplasty gave satisfactory medium-term results in active patients. The press-fit metal-backed socket appeared to have reliable fixation in alumina-on-alumina hip arthroplasty. The excellent results using cemented
The prevalence of ipsilateral total hip arthroplasty (THA) and total knee arthroplasty (TKA) is rising in concert with life expectancy, putting more patients at risk for interprosthetic femur fractures (IPFFs). Our study aimed to assess treatment methodologies, implant survivorship, and IPFF clinical outcomes. A total of 76 patients treated for an IPFF from February 1985 to April 2018 were reviewed. Prior to fracture, at the hip/knee sites respectively, 46 femora had primary/primary, 21 had revision/primary, three had primary/revision, and six had revision/revision components. Mean age and BMI were 74 years (33 to 99) and 30 kg/m2 (21 to 46), respectively. Mean follow-up after fracture treatment was seven years (2 to 24).Aims
Methods
Impacted morcellised bone allograft and a Charnley stem was used to revise 59 loose femoral components in 57 consecutive patients. Femoral bone loss was rated as Endo-Klinik grade 2 in nine patients, grade 3 in 41, and grade 4 in nine. The immediate postoperative radiographs and those taken at the most recent follow-up were compared for radiolucencies, subsidence and incorporation of the graft. One patient was lost to follow-up and two were not available for radiological analysis. The mean clinical follow-up in 58 procedures was 56.7 months (24 to 144) and the mean radiological review of 56 reconstructions was 54.4 months (24 to 144). An intraoperative femoral fracture occurred in one patient (1.7%) and was successfully treated by strut grafting and cerclage wiring. Extrusion of cement through perforations or incomplete hoop fractures was detected in the postoperative radiographs of ten procedures (17%); none of these patients sustained a complete fracture. Three patients had dislocations (5%) and two (3.5%) developed painful subsidence of the stem which required a further revision. The latest follow-up radiographs in 56 reconstructions showed a well fixed stem and radiological healing of the graft in 52 (93%), and definite loosening in four (7%). Of these four, two were revised again and two were asymptomatic after a follow-up of 120 months each. The mean subsidence in the 52 successful revisions was 0.38 mm (0 to 4). Impaction allografting with a Charnley stem restored bone stock and provided adequate
There are concerns regarding initial stability and early periprosthetic fractures in cementless hip arthroplasty using short stems. This study aimed to investigate stress on the cortical bone around the stem and micromotions between the stem and cortical bone according to femoral stem length and positioning. In total, 12 femoral finite element models (FEMs) were constructed and tested in walking and stair-climbing. Femoral stems of three different lengths and two different positions were simulated, assuming press-fit fixation within each FEM. Stress on the cortical bone and micromotions between the stem and bone were measured in each condition.Aims
Methods
We compared implant and patient survival following intraoperative periprosthetic femoral fractures (IOPFFs) during primary total hip arthroplasty (THA) with matched controls. This retrospective cohort study compared 4831 hips with IOPFF and 48 154 propensity score matched primary THAs without IOPFF implanted between 2004 and 2016, which had been recorded on a national joint registry. Implant and patient survival rates were compared between groups using Cox regression.Aims
Patients and Methods
We aimed to report the mid- to long-term rates of septic and aseptic failure after two-stage revision surgery for periprosthetic joint infection (PJI) following total hip arthroplasty (THA). We retrospectively reviewed 96 cases which met the Musculoskeletal Infection Society criteria for PJI. The mean follow-up was 90 months (SD 32). Septic failure was assessed using a Delphi-based consensus definition. Any further surgery undertaken for aseptic mechanical causes was considered as aseptic failure. The cumulative incidence with competing risk analysis was used to predict the risk of septic failure. A regression model was used to evaluate factors associated with septic failure. The cumulative incidence of aseptic failure was also analyzed.Aims
Methods
The aim of this study was to determine whether fixation, as opposed to revision arthroplasty, can be safely used to treat reducible Vancouver B type fractures in association with a cemented collarless polished tapered femoral stem (the Exeter). This retrospective cohort study assessed 152 operatively managed consecutive unilateral Vancouver B fractures involving Exeter stems; 130 were managed with open reduction and internal fixation (ORIF) and 22 with revision arthroplasty. Mean follow-up was 6.5 years (SD 2.6; 3.2 to 12.1). The primary outcome measure was revision of at least one component. Kaplan–Meier survival analysis was performed. Regression analysis was used to identify risk factors for revision following ORIF. Secondary outcomes included any reoperation, complications, blood transfusion, length of hospital stay, and mortality.Aims
Methods
The aim of this study was to evaluate the differences in revision and complication rates, functional outcomes, and radiological outcomes between cemented and press-fit humeral stems in primary anatomical total shoulder arthroplasty (TSA). A comprehensive systematic review and meta-analysis was conducted searching for studies that included patients who underwent primary anatomical TSA for primary osteoarthritis or rheumatoid arthritis.Aims
Materials and Methods
In the 1990s, a bioactive bone cement (BABC) containing apatite-wollastonite glass-ceramic (AW-GC) powder and bisphenol-a-glycidyl methacrylate resin was developed at our hospital. In 1996, we used BABC to fix the acetabular component in primary total hip arthroplasty (THA) in 20 patients as part of a clinical trial. The purpose of this study was to investigate the long-term results of primary THA using BABC. A total of 20 patients (three men and 17 women) with a mean age of 57.4 years (40 to 71), a mean body weight of 52.3 kg (39 to 64), and a mean body mass index (BMI) of 23.0 kg/m2 (19.8 to 28.6) were evaluated clinically and radiologically. Survival analyses were undertaken, and wear analyses were carried out using a computer-aided method.Aims
Patients and Methods
The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA). Prospective cohort study using linked National Joint Registry (NJR) and manufacturer data. The primary endpoint was revision for aseptic loosening. Cox proportional hazard regression was the primary analytical approach. Manufacturing variables included resin type, crosslinking radiation dose, terminal sterilization method, terminal sterilization radiation dose, stabilization treatment, total radiation dose, packaging, and face asymmetry. Total radiation dose was further divided into G1 (no radiation), G2 (> 0 Mrad to < 5 Mrad), G3 (≥ 5 Mrad to < 10 Mrad), and G4 (≥ 10 Mrad).Aims
Methods
The consensus is that bipolar hemiarthroplasty (BHA) in allograft-prosthesis composite (APC) reconstruction of the proximal femur following primary tumour resection provides more stability than total hip arthroplasty (THA). However, no comparative study has been performed. In this study, we have compared the outcome and complication rates of these two methods. In a retrospective study, 57 patients who underwent APC reconstruction of proximal femur following the primary tumour resection, either using BHA (29) or THA (28), were included. Functional outcome was assessed using the Musculoskeletal Tumour Society (MSTS) scoring system and Harris Hip Score (HHS). Postoperative complications of the two techniques were also compared.Aims
Methods
Although good clinical outcomes have been reported for monolithic tapered, fluted, titanium stems (TFTS), early results showed high rates of subsidence. Advances in stem design may mitigate these concerns. This study reports on the use of a current monolithic TFTS for a variety of indications. A multi-institutional retrospective study of all consecutive total hip arthroplasty (THA) and revision total hip arthroplasty (rTHA) patients who received the monolithic TFTS was conducted. Surgery was performed by eight fellowship-trained arthroplasty surgeons at four institutions. A total of 157 hips in 153 patients at a mean follow-up of 11.6 months (SD7.8) were included. Mean patient age at the time of surgery was 67.4 years (SD 13.3) and mean body mass index (BMI) was 28.9 kg/m2 (SD 6.5). Outcomes included intraoperative complications, one-year all-cause re-revisions, and subsidence at postoperative time intervals (two weeks, six weeks, six months, nine months, and one year).Aims
Methods
Knee osteonecrosis in advanced stages may lead to joint degeneration. Total knee arthroplasty (TKA) for osteonecrosis has traditionally been associated with suboptimal results. We analyzed outcomes of contemporary TKAs for osteonecrosis, with particular emphasis on: survivorship free from aseptic loosening, any revision, and any reoperation plus the clinical outcomes, complications, and radiological results. In total, 156 patients undergoing 167 primary TKAs performed for osteonecrosis between 2004 and 2014 at a single institution were reviewed. The mean age at index TKA was 61 years (14 to 93) and the mean body mass index (BMI) was 30 kg/m2 (18 to 51) The mean follow-up was six years (2 to 12). A total of 110 TKAs (66%) were performed for primary osteonecrosis and 57 TKAs (34%) for secondary osteonecrosis. Overall, 15 TKAs (9%) had tibial stems, while 12 TKAs (7%) had femoral stems. Posterior-stabilized designs were used in 147 TKAs (88%) of TKAs. Bivariate Cox regression analysis was conducted to identify risk factors for revision and reoperation.Aims
Patients and Methods
A number of methods have been described to remove a well-fixed humeral implant as part of revision shoulder arthroplasty. These include the use of cortical windows and humeral osteotomies. The router bit extraction technique uses a high-speed router bit to disrupt the bone-implant interface. The implant is then struck in a retrograde fashion with a square-tip impactor and mallet. The purpose of this study was to determine the characteristics and frequency of the different techniques needed for the removal of a well-fixed humeral stem in revision shoulder arthroplasty. Between 2010 and 2018, 288 revision shoulder arthroplasty procedures requiring removal of a well-fixed humeral component were carried out at a tertiary referral centre by a single surgeon. The patient demographics, indications for surgery, and method of extraction were collected.Aims
Patients and Methods
To evaluate the hypothesis that failed osteosynthesis of periprosthetic
Vancouver type B1 fractures can be treated successfully with stem
revision using a transfemoral approach and a cementless, modular,
tapered revision stem with reproducible rates of fracture healing,
stability of the revision stem, and clinically good results. A total of 14 patients (11 women, three men) with a mean age
of 72.4 years (65 to 90) undergoing revision hip arthroplasty after
failed osteosynthesis of periprosthetic fractures of Vancouver type
B1 were treated using a transfemoral approach to remove the well-fixed
stem before insertion of a modular, fluted titanium stem which obtained
distal fixation. These patients were clinically and radiologically
followed up for a mean 52.2 months (24 to 144).Aims
Patients and Methods
Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique. A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1).Aims
Patients and Methods
Aims
Patients and Methods
Few studies have assessed outcomes following non-metal-on-metal hip arthroplasty (non-MoMHA) revision surgery performed for adverse reactions to metal debris (ARMD). We assessed outcomes following non-MoMHA revision surgery performed for ARMD, and identified predictors of re-revision. We performed a retrospective observational study using data from the National Joint Registry for England and Wales. All non-MoMHAs undergoing revision surgery for ARMD between 2008 and 2014 were included (185 hips in 185 patients). Outcome measures following ARMD revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using Cox regression.Objectives
Methods
Cement thickness of at least 2 mm is generally associated with more favorable results for the femoral component in cemented hip arthroplasty. However, French-designed stems have shown favorable outcomes even with thin cement mantle. The biomechanical behaviors of a French stem, Charnley-Marcel-Kerboull (CMK) and cement were researched in this study. Six polished CMK stems were implanted into a composite femur, and one million times dynamic loading tests were performed. Stem subsidence and the compressive force at the bone-cement interface were measured. Tantalum ball (ball) migration in the cement was analyzed by micro CTObjective
Methods
To date, there is insufficient evidence available to compare
the outcome of cemented and uncemented fixation of the humeral stem
in reverse shoulder arthroplasty (RSA). A systemic review comprising 41 clinical studies was performed
to compare the functional outcome and rate of complications of cemented
and uncemented stems in RSA. These included 1455 cemented and 329
uncemented shoulders. The clinical characteristics of the two groups
were similar. Variables were compared using pooled frequency-weighted means
and relative risk ratios (RR).Aims
Methods
The Corail stem has good long-term results. After four years
of using this stem, we have detected a small group of patients who
have presented with symptomatic metaphyseal debonding. The aim of
this study was to quantify the incidence of this complication, to
delineate the characteristics of patients presenting with this complication
and to compare these patients with asymptomatic controls to determine
any important predisposing factors. Of 855 Corail collarless cementless stems implanted for osteoarthritis,
18 presented with symptomatic metaphyseal debonding. A control group
of 74 randomly selected patients was assembled. Clinical and radiological
parameters were measured and a logistic regression model was created
to evaluate factors associated with metaphyseal debonding.Aims
Patients and Methods
In the initial development of total shoulder arthroplasty (TSA),
the humeral component was usually fixed with cement. Cementless
components were subsequently introduced. The aim of this study was
to compare the long-term outcome of cemented and cementless humeral
components in arthroplasty of the shoulder. All patients who underwent primary arthroplasty of the shoulder
at our institution between 1970 and 2012 were included in the study.
There were 4636 patients with 1167 cemented humeral components and
3469 cementless components. Patients with the two types of fixation
were matched for nine different covariates using a propensity score
analysis. A total of 551 well-balanced pairs of patients with cemented
and cementless components were available after matching for comparison
of the outcomes. The clinical outcomes which were analysed included loosening
of the humeral component determined at revision surgery, periprosthetic
fractures, post-operative infection and operating time.Aims
Patients and Methods
Favourable results for collarless polished tapered stems have been reported, and cement creep due to taper slip may be a contributing factor. However, the ideal cement thickness around polished stems remains unknown. We investigated the influence of cement thickness on stem subsidence and cement creep. We cemented six collarless polished tapered (CPT) stems (two stems each of small, medium and large sizes) into composite femurs that had been reamed with a large CPT rasp to achieve various thicknesses of the cement mantle. Two or three tantalum balls were implanted in the proximal cement in each femur. A cyclic loading test was then performed for each stem. The migration of the balls was measured three-dimensionally, using a micro-computed tomography (CT) scanner, before and after loading. A digital displacement gauge was positioned at the stem shoulder, and stem subsidence was measured continuously by the gauge. Final stem subsidence was measured at the balls at the end of each stem.Objectives
Methods
We report on the outcome of the Synergy cementless femoral stem
with a minimum follow-up of 15 years (15 to 17). A retrospective review was undertaken of a consecutive series
of 112 routine primary cementless total hip arthroplasties (THAs)
in 102 patients (112 hips). There were 60 female and 42 male patients
with a mean age of 61 years (18 to 82) at the time of surgery. A
total of 78 hips in the 69 patients remain Aims
Patients and Methods
Studies reporting specifically on squeaking in total hip arthroplasty have focused on cementless, and not on hybrid, fixation. We hypothesised that the cement mantle of the femur might have a damping effect on the sound transmitted through the metal stem. The objective of this study was to test the effect of cement on sound propagation along different stem designs and under different fixation conditions. An Objectives
Methods
In total hip arthroplasty (THA), the cementless, tapered-wedge stem design contributes to achieving initial stability and providing optimal load transfer in the proximal femur. However, loading conditions on the femur following THA are also influenced by femoral structure. Therefore, we determined the effects of tapered-wedge stems on the load distribution of the femur using subject-specific finite element models of femurs with various canal shapes. We studied 20 femurs, including seven champagne flute-type femurs, five stovepipe-type femurs, and eight intermediate-type femurs, in patients who had undergone cementless THA using the Accolade TMZF stem at our institution. Subject–specific finite element (FE) models of pre- and post-operative femurs with stems were constructed and used to perform FE analyses (FEAs) to simulate single-leg stance. FEA predictions were compared with changes in bone mineral density (BMD) measured for each patient during the first post-operative year.Objectives
Patients and Methods
The treatment of bone loss in revision total
knee arthroplasty has evolved over the past decade. While the management
of small to moderate sized defects has demonstrated good results
with a variety of traditional techniques (cement and screws, small
metal augments, impaction bone grafting or modular stems), the treatment of
severe defects continues to be problematic. The use of a structural
allograft has declined in recent years due to an increased failure
rate with long-term follow-up and with the introduction of highly
porous metal augments that emphasise biological metaphyseal fixation.
Recently published mid-term results on the use of tantalum cones
in patients with severe bone loss has reaffirmed the success of
this treatment strategy. Cite this article:
We describe our technique and rationale using
hybrid fixation for primary total hip arthroplasty (THA) at the Hospital
for Special Surgery. Modern uncemented acetabular components have
few screw holes, or no holes, polished inner surfaces, improved
locking mechanisms, and maximised thickness and shell-liner conformity. Uncemented
sockets can be combined with highly cross-linked polyethylene liners,
which have demonstrated very low wear and osteolysis rates after
ten to 15 years of implantation. The results of cement fixation
with a smooth or polished surface finished stem have been excellent,
virtually eliminating complications seen with cementless fixation
like peri-operative femoral fractures and thigh pain. Although mid-term
results of modern cementless stems are encouraging, the long-term
data do not show reduced revision rates for cementless stems compared
with cemented smooth stems. In this paper we review the conduct
of a hybrid THA, with emphasis on pre-operative planning, surgical
technique, hypotensive epidural anaesthesia, and intra-operative
physiology. Cite this article:
The goals of this study were to define the risk factors, nature,
chronology, and treatment strategies adopted for periprosthetic
femoral fractures in 32 644 primary total hip arthroplasties (THAs). There were 564 intra-operative fractures (1.7%); 529 during uncemented
stem placement (3.0%) and 35 during cemented stem placement (0.23%).
Intra-operative fractures were more common in females and patients
over 65 years (p <
0.001). The majority occurred during placement
of the femoral component (60%), and involved the calcar (69%). There
were 557 post-operative fractures (20-year probability: 3.5%; 95%
confidence interval (CI) 3.2 to 3.9); 335 fractures after placement
of an uncemented stem (20-year probability: 7.7%; 95% CI 6.2 to
9.1) and 222 after placement of a cemented stem (20-year probability:
2.1%; 95% CI 1.8 to 2.5). The probability of a post-operative fracture
within 30 days after an uncemented stem was ten times higher than
a cemented stem. The most common post-operative fracture type was
a Vancouver AG (32%; n = 135), with 67% occurring after
a fall. In all, 36% (n = 152) were treated with revision arthroplasty. Aim and Methods
Results
The purpose of this study was to examine the
complications and outcomes of total hip replacement (THR) in super-obese
patients (body mass index (BMI) >
50 kg/m2) compared
with class I obese (BMI 30 to 34.9 kg/m2) and normal-weight
patients (BMI 18.5 to 24.9 kg/m2), as defined by the
World Health Organization. A total of 39 THRs were performed in 30 super-obese patients
with a mean age of 53 years (31 to 72), who were followed for a
mean of 4.2 years (2.0 to 11.7). This group was matched with two
cohorts of normal-weight and class I obese patients, each comprising
39 THRs in 39 patients. Statistical analysis was performed to determine differences
among these groups with respect to complications and satisfaction
based on the Western Ontario and McMaster Universities (WOMAC) osteoarthritis
index, the Harris hip score (HHS) and the Short-Form (SF)-12 questionnaire. Super-obese patients experienced significantly longer hospital
stays and higher rates of major complications and readmissions than
normal-weight and class I obese patients. Although super-obese patients
demonstrated reduced pre-operative and post-operative satisfaction
scores, there was no significant difference in improvement, or change in
the score, with respect to HHS or the WOMAC osteoarthritis index. Super-obese patients obtain similar satisfaction outcomes as
class I obese and normal-weight patients with respect to improvement
in their scores. However, they experience a significant increase
in length of hospital stay and major complication and readmission
rates. Cite this article:
Revision knee arthroplasty presents a number
of challenges, not least of which is obtaining solid primary fixation
of implants into host bone. Three anatomical zones exist within
both femur and tibia which can be used to support revision implants.
These consist of the joint surface or epiphysis, the metaphysis
and the diaphysis. The methods by which fixation in each zone can
be obtained are discussed. The authors suggest that solid fixation
should be obtained in at least two of the three zones and emphasise
the importance of pre-operative planning and implant selection. Cite this article:
Endoprosthetic reconstruction following resection of 31 tumours of the proximal femur in 30 patients was performed using a Wagner SL femoral revision stem. The mean follow-up was 25.6 months (0.6 to 130.0). Of the 28 patients with a metastasis, 27 died within a mean follow-up period of 18.1 months (0.6 to 56.3) after the operation, and the remaining patient was excluded from the study 44.4 months post-operatively when the stem was removed. The two patients with primary bone tumours were still alive at the latest follow-up of 81.0 and 130.0 months, respectively. One stem only was removed for suspected low-grade infection 44.4 months post-operatively. The worst-case survival rate with removal of the stem for any cause and/or loss to follow-up was 80.0% (95% confidence interval 44.9 to 100) at 130.0 months. The mean Karnofsky index increased from 44.2% (20% to 70%) pre-operatively to 59.7% (0% to 100%) post-operatively, and the mean Merle d’Aubigné score improved from 4.5 (0 to 15) to 12.0 (0 to 18). The mean post-operative Musculoskeletal Tumour Society score was 62.4% (3.3% to 100%). The Wagner SL femoral revision stem offers an alternative to special tumour prostheses for the treatment of primary and secondary tumours of the proximal femur. The mid-term results are very promising, but long-term experience is necessary.
Total hip replacement for high dislocation of the hip joint remains technically difficult in terms of preparation of the true acetabulum and restoration of leg length. We describe our experience of cementless total hip replacement combined with a subtrochanteric femoral shortening osteotomy in 20 hips with Crowe grade IV dislocation with a mean follow-up of 8.1 years (4 to 11.5). There was one man and 17 women with a mean age of 55 years (44 to 69) at the time of the operation. After placment of the acetabular component at the site of the natural acetabulum, a cementless porous-coated cylindrical femoral component was implanted following a subtrochanteric femoral shortening osteotomy. The mean Japanese Orthopedic Association hip score improved from a mean of 38 (22 to 62) to a mean of 83 points (55 to 98) at the final follow-up. The mean lengthening of the leg was 14.8 mm (−9 to 34) in patients with iliofemoral osteoarthritis and 35.3 mm (15 to 51) in patients with no arthritic changes. No nerve palsy was observed. Total hip replacement combined with subtrochanteric shortening femoral osteotomy in this situation is beneficial in avoiding nerve injury and still permits valuable improvement in inequality of leg length.
The purpose of this prospective study was to
evaluate the long-term clinical and radiological outcomes of revision of
the femoral component of a total hip replacement using impaction
bone grafting. Femoral revision with an impacted allograft was performed
on 29 patients (31 hips). In all, 21 hips (68%) had grade III or
IV femoral defects according to the Endo-Klinik classification.
A total of 11 patients (12 hips) died before the ten-year follow-up
period. Of the remaining patients, 18 patients (19 hips) were followed
for 10 to 15 years; three further patients died during this time.
None of the 31 stems underwent further revision of their stem. However,
four stems showed extensive subsidence (>
15 mm). One of these patients
had a femoral fracture that required fixation. Three other patients
had a femoral fracture, two of which required fixation and the other
was treated conservatively. Patients with a femoral fracture and/or
severe subsidence had significantly more grade IV defects (six of
seven hips; p = 0.004). One patient needed a closed reduction for
dislocation. Impaction allografting in revision hip surgery gives good long-term
results for femora with grades I, II and III Endo-Klinik-classified
defects. Extensive subsidence and femoral fractures were seen mainly
in patients with grade IV damaged femora.
Stems may improve fixation and stability of components
during revision total knee replacement. However, the choice between
cemented and cementless stems is not a clear one. Cemented stems
offer several advantages in terms of versatility, mechanical stability, surgical
technique and clinical outcome over their cementless counterpart. Cite this article:
Tribocorrosion at the head–neck taper interface
– so-called ‘taperosis’ – may be a source of metal ions and particulate
debris in metal-on-polyethylene total hip arthroplasty (THA). We examined the effect of femoral head length on fretting and
corrosion in retrieved head–neck tapers Head length was observed to affect fretting (p = 0.03), with
28 mm + 8 mm femoral heads showing greater total fretting scores
than all other head lengths. The central zone of the femoral head
bore taper was subject to increased fretting damage (p = 0.01),
regardless of head length or stem offset. High-offset femoral stems
were associated with greater total fretting of the bore taper (p
= 0.04). Increased fretting damage is seen with longer head lengths and
high-offset femoral stems, and occurs within a central concentric
zone of the femoral head bore taper. Further investigation is required
to determine the effect of increased head size, and variations in
head–neck taper design. Cite this article:
Periprosthetic joint infection (PJI) is one of
the most feared and challenging complications following total knee arthroplasty.
We provide a detailed description of our current understanding regarding
the management of PJI of the knee, including diagnostic aids,
pre-operative planning, surgical treatment, and outcome. Cite this article:
Metaphyseal bone loss is common with revision
total knee replacement (RTKR). Using the Anderson Orthopaedic Research
Institute (AORI) classification, type 2-B and type 3 defects usually
require large metal blocks, bulk structural allograft or highly
porous metal cones. Tibial and femoral trabecular metal metaphyseal
cones are a unique solution for large bone defects. These cones
substitute for bone loss, improve metaphyseal fixation, help correct
malalignment, restore the joint line and may permit use of a shorter
stem. The technique for insertion involves sculpturing of the remaining
bone with a high speed burr and rasp, followed by press-fit of the
cone into the metaphysis. The fixation and osteoconductive properties
of the porous cone outer surface allow ingrowth and encourage long-term
biological fixation. The revision knee component is then cemented
into the porous cone inner surface, which provides superior fixation
compared with cementing into native but deficient metaphyseal bone.
The advantages of the cone compared with allograft include: technical
ease, biological fixation, no resorption, and possibly a lower risk
of infection. The disadvantages include: difficult extraction and
relatively short-term follow-up. Several studies using cones report
promising short-term results for the reconstruction of large bone
defects in RTKR. Cite this article:
We report the effectiveness of revision of total
elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows
in 52 patients were treated with re-cementing of a total elbow replacement
into part or all of the existing cement mantle or into the debrided
host-bone interface, without the use of structural bone augmentation
or a custom prosthesis. The original implant revision was still
We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total knee prostheses in 32 young patients with primary malignant bone tumours. Initial stable fixation, especially in rotation, was achieved by precise fit of the stem into the reamed endosteum, before cementation. Complementary fixation, especially in migration and rotation, was obtained by pressurised antibiotic-loaded cement. The mean age at operation was 33 years (13 to 61). No patient was lost to follow-up; 13 patients died and the 19 survivors were examined at a mean follow-up of 12.5 years (4 to 21). Standard revision press-fit cemented stems were used on the side of the joint which was not involved with tumour (26 tibial and six femoral), on this side there was no loosening or osteolysis and stem survival was 100%. On the reconstruction side, custom-made press-fit stems were used and the survival rate, with any cause for revision as an end point, was 88%, but 97% for loosening or osteolysis. This longevity is similar to that achieved at 20 years with the Charnley-Kerboull primary total hip replacement with press-fit cemented femoral components. We recommend this type of fixation when extensive reconstruction of the knee is required. It may also be suitable for older patients requiring revision of a total knee replacement or in difficult situations such as severe deformity and complex articular fractures.
This review examines the future of total hip arthroplasty, aiming to avoid past mistakes
We report our experience of revision total hip
replacement (THR) using the Revitan curved modular titanium fluted revision
stem in patients with a full spectrum of proximal femoral defects.
A total of 112 patients (116 revisions) with a mean age of 73.4
years (39 to 90) were included in the study. The mean follow-up
was 7.5 years (5.3 to 9.1). A total of 12 patients (12 hips) died
but their data were included in the survival analysis, and four
patients (4 hips) were lost to follow-up. The clinical outcome,
proximal bone regeneration and subsidence were assessed for 101
hips. The mean Harris Hip Score was 88.2 (45.8 to 100) after five years
and there was an increase of the mean Barnett and Nordin-Score,
a measure of the proximal bone regeneration, of 20.8 (-3.1 to 52.7).
Five stems had to be revised (4.3%), three (2.9%) showed subsidence,
five (4.3%) a dislocation and two of 85 aseptic revisions (2.3%)
a periprosthetic infection. At the latest follow-up, the survival with revision of the stem
as the endpoint was 95.7% (95% confidence interval 91.9% to 99.4%)
and with aseptic loosening as the endpoint, was 100%. Peri-prosthetic
fractures were not observed. We report excellent results with respect to subsidence, the risk
of fracture, and loosening after femoral revision using a modular
curved revision stem with distal cone-in-cone fixation. A successful
outcome depends on careful pre-operative planning and the use of
a transfemoral approach when the anatomy is distorted or a fracture
is imminent, or residual cement or a partially-secured existing
stem cannot be removed. The shortest appropriate stem should, in
our opinion, be used and secured with >
3 cm fixation at the femoral
isthmus, and distal interlocking screws should be used for additional
stability when this goal cannot be realised. Cite this article:
Conventional uncemented femoral implants provide
dependable long-term fixation in patients with a wide range of functional
requirements. Yet challenges associated with proximal–distal femoral
dimensional mismatch, preservation of bone stock, and minimally
invasive approaches have led to exploration into alternative implant designs.
Short stem designs focusing on a stable metaphyseal fit have emerged
to address these issues in total hip replacement (THR). Uncemented
metaphyseal-engaging short stem implants are stable and are associated
with proximal bone remodeling closer to the metaphysis when compared
with conventional stems and they also have comparable clinical performances.
Short stem metaphyseal-engaging implants can meet the goals of a
successful THR, including tolerating a high level of patient function,
as well as durable fixation. Cite this article:
We reviewed 25 patients in whom a MUTARS megaprosthesis with a conical fluted stem had been implanted. There were three types of stem: a standard stem was used in 17 cases (three in the proximal femur, nine in the distal femur and five proximal tibia), a custom-made proximal femoral stem in four cases and a custom-made distal femoral stem in four cases. The mean age of the patients was 40.1 years (17 to 70) and the mean follow-up was for 2.5 years (0.9 to 7.4). At follow-up two patients had died from their disease: one was alive with disease and 22 were disease-free. One of 23 prostheses had been removed for infection and another revised to a cemented stem. The mean Musculoskeletal Tumor Society score was 24.9 (12 to 30) and the mean Karnofsky index was 82% (60% to 100%). There was no radiological evidence of loosening or subsidence. Stem stress shielding was seen in 11 patients and was marked in five of these. There were five complications, rupture of the extensor mechanism of the knee after extra-articular resection in two patients, deep venous thrombosis in one, septic loosening in one, and dislocation of the hip in one. The survival rate after seven years was 87% (95% confidence interval (CI) 83 to 91) for the patients and 95% (95% CI 91 to 99) for the megaprosthesis. A longer follow-up is needed to confirm these encouraging results.
Large-head metal-on-metal (MoM) total hip replacements
(THR) have given rise to concern. Comparative studies of small-head
MoM THRs over a longer follow-up period are lacking. Our objective
was to compare the incidence of complications such as infection,
dislocation, revision, adverse local tissue reactions, mortality
and radiological and clinical outcomes in small-head (28 mm) MoM
and ceramic-on-polyethylene (CoP) THRs up to 12 years post-operatively. A prospective cohort study included 3341 THRs in 2714 patients.
The mean age was 69.1 years (range 24 to 98) and 1848 (55.3%) were
performed in women, with a mean follow-up of 115 months (18 to 201).
There were 883 MoM and 2458 CoP bearings. Crude incidence rates
(cases/1000 person-years) were: infection 1.3 In conclusion, we found similar results for small-head MoM and
CoP bearings up to ten years post-operatively, but after ten years
MoM THRs had a higher risk of all-cause revision. Furthermore, the
presence of an adverse response to metal debris seen in the small-head
MOM group at revision is a cause for concern. Cite this article:
We present the extended follow-up (≥ 20 years)
of a series of fully hydroxyapatite-coated femoral components used in
72 primary total hip replacements (THRs). Earlier results of this
cohort have been previously published. All procedures were performed
between 1986 and 1991. The series involved 45 women and 15 men with
12 bilateral procedures. Their mean age at the time of surgery was
60 years (46 to 80) and the mean duration of follow-up was 22.5
years (20 to 25). At final follow-up, the mean Merle d’Aubigné and
Postel hip scores were 5.5 (4.5 to 6), 3.8 (3.5 to 5) and 3.3 (3.0
to 5.0) for pain, mobility and function, respectively. Of the patients
92% were very satisfied at the time of final follow-up. There were seven revisions: six of the acetabular component for
aseptic loosening and one of both the stem and the acetabular component
for loosening due to deep infection. The survival of this prosthesis
at 22.5 years with revision for any reason as the endpoint was 91.7%
(95% confidence interval (CI) 84 to 99). Survival with aseptic loosening
of the stem as the endpoint was 100% (95% CI 90 to 100). This prosthesis provides pain relief in the long term. Survival
of this component is comparable to the best results for primary
THR with any means of fixation. Cite this article:
This study reviewed the results of a cementless anatomical femoral component to give immediate post-operative stability, and with a narrow distal section in order not to contact the femoral cortex in the diaphysis, ensuring exclusively metaphyseal loading. A total of 471 patients (601 hips) who had a total hip replacement between March 1995 and February 2002 were included in the study. There were 297 men and 174 women. The mean age at the time of operation was 52.7 years (28 to 63). Clinical and radiological evaluation were performed at each follow-up. Bone densitometry was carried out on all patients two weeks after operation and at the final follow-up examination. The mean follow-up was 8.8 years (5 to 12). The mean pre-operative Harris hip score was 41 points (16 to 54), which improved to a mean of 96 (68 to 100) at the final follow-up. No patient complained of thigh pain at any stage. No acetabular or femoral osteolysis was observed and no hip required revision for aseptic loosening of either component. Deep infection occurred in two hips (0.3%) which required revision. One hip (0.2%) required revision of the acetabular component for recurrent dislocation. Bone mineral densitometry revealed a minimal bone loss in the proximal femur. This cementless anatomical femoral component with metaphyseal loading but without distal fixation produced satisfactory fixation and encourages proximal femoral loading.
The August 2012 Hip &
Pelvis Roundup360 looks at: whether cemented hip replacement might be bad for your health; highly cross-linked polyethylene; iHOT-33 - a new hip outcome measure; hamstring injuries; total hip replacement; stemmed metal-on-metal THR; bipolar hemiarthroplasty, neuromuscular disease and dislocation; the high risk of secondary hemiarthroplasty; and whether we have to repair the labrum after all?
Stems improve the mechanical stability of tibial
components in total knee replacement (TKR), but come at a cost of stress
shielding along their length. Their advantages include resistance
to shear, reduced tibial lift-off and increased stability by reducing
micromotion. Longer stems may have disadvantages including stress
shielding along the length of the stem with associated reduction
in bone density and a theoretical risk of subsidence and loosening, peri-prosthetic
fracture and end-of-stem pain. These features make long stems unattractive
in the primary TKR setting, but often desirable in revision surgery
with bone loss and instability. In the revision scenario, stems
are beneficial in order to convey structural stability to the construct
and protect the reconstruction of bony defects. Cemented and uncemented
long stemmed implants have different roles depending on the nature
of the bone loss involved. This review discusses the biomechanics of the design of tibial
components and stems to inform the selection of the component and
the technique of implantation.
We reviewed 59 bone graft substitutes marketed
by 17 companies currently available for implantation in the United Kingdom,
with the aim of assessing the peer-reviewed literature to facilitate
informed decision-making regarding their use in clinical practice.
After critical analysis of the literature, only 22 products (37%)
had any clinical data. Norian SRS (Synthes), Vitoss (Orthovita),
Cortoss (Orthovita) and Alpha-BSM (Etex) had Level I evidence. We question
the need for so many different products, especially with limited
published clinical evidence for their efficacy, and conclude that
there is a considerable need for further prospective randomised
trials to facilitate informed decision-making with regard to the
use of current and future bone graft substitutes in clinical practice. Cite this article:
The purpose of this study was to define immediate post-operative ‘quality’ in total hip replacements and to study prospectively the occurrence of failure based on these definitions of quality. The evaluation and assessment of failure were based on ten radiological and clinical criteria. The cumulative summation (CUSUM) test was used to study 200 procedures over a one-year period. Technical criteria defined failure in 17 cases (8.5%), those related to the femoral component in nine (4.5%), the acetabular component in 32 (16%) and those relating to discharge from hospital in five (2.5%). Overall, the procedure was considered to have failed in 57 of the 200 total hip replacements (28.5%). The use of a new design of acetabular component was associated with more failures. For the CUSUM test, the level of adequate performance was set at a rate of failure of 20% and the level of inadequate performance set at a failure rate of 40%; no alarm was raised by the test, indicating that there was no evidence of inadequate performance. The use of a continuous monitoring statistical method is useful to ensure that the quality of total hip replacement is maintained, especially as newer implants are introduced.
Segmental resection of malignant bone disease in the femoral diaphysis with subsequent limb reconstruction is a major undertaking. This is a retrospective review of 23 patients who had undergone limb salvage by endoprosthetic replacement of the femoral diaphysis for a primary bone tumour between 1989 and 2005. There were 16 males and seven females, with a mean age of 41.3 years (10 to 68). The mean overall follow-up was for 97 months (3 to 240), and 120 months (42 to 240) for the living patients. The cumulative patient survival was 77% (95% confidence interval 63% to 95%) at ten years. Survival of the implant, with failure of the endoprosthesis as an endpoint, was 85% at five years and 68% (95% confidence interval 42% to 92%) at ten years. The revision rate was 22% and the overall rate of re-operation was 26%. Complications included deep infection (4%), breakage of the prosthesis (8%), periprosthetic fracture (4%), aseptic loosening (4%), local recurrence (4%) and metastases (17%). The 16 patients who retained their diaphyseal endoprosthesis had a mean Musculoskeletal Tumour Society score of 87% (67% to 93%). They were all able to comfortably perform most activities of daily living. Femoral diaphyseal endoprosthetic replacement is a viable option for reconstruction following segmental resection of malignant bone disease. It allows immediate weight-bearing, is associated with a good long-term functional outcome, has an acceptable complication and revision rate and, most importantly, does not appear to compromise patient survival.