Advertisement for orthosearch.org.uk
Results 1 - 65 of 65
Results per page:
The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 41 - 46
1 Jul 2020
Ransone M Fehring K Fehring T

Aims

Patients with abnormal spinopelvic mobility are at increased risk for instability. Measuring the change in sacral slope (ΔSS) can help determine spinopelvic mobility preoperatively. Sacral slope (SS) should decrease at least 10° to demonstrate adequate posterior pelvic tilt. There is potential for different ΔSS measurements in the same patient based on sitting posture. The purpose of this study was to determine the effect of sitting posture on the ΔSS in patients undergoing total hip arthroplasty (THA).

Methods

In total, 51 patients undergoing THA were reviewed to quantify the variability in preoperative spinopelvic mobility when measuring two different sitting positions using SS for planning.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 25 - 32
1 Jul 2021
Amstutz HC Le Duff MJ

Aims. Adverse local tissue reactions associated with abnormal wear considerably slowed down the general use of metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), now limited to a few specialized centres. In this study, we provide the clinical results of 400 consecutive MoM HRAs implanted more than 20 years ago in one such centre. Methods. A total of 355 patients (400 hips) were treated with Conserve Plus HRA between November 1996 and November 2000. There were 96 female (27%) and 259 male patients (73%). Their mean age was 48.2 years (SD 10.9). The University of California, Los Angeles (UCLA) hip scores and 12-item Short Form Survey (SF-12) quality of life scores were reported. Survivorship was assessed using Kaplan-Meier analyses. Results. The mean follow-up was 16.5 years (0.1 to 24.0), including 34 patients (37 hips) who died. The mean UCLA hip scores were 9.3 (2 to 10), 9.1 (3 to 10), 9.0 (3 to 10), and 6.9 (2 to 10) for pain, walking, function, and activity, respectively. The mean SF-12 scores were 48.4 (16.0 to 62.1) for the physical component and 48.5 (10.5 to 66.5) for the mental component, and did not differ from those of the general population of the USA. A total of 60 hips in 55 patients were revised. Using revision for any indication as the endpoint, the Kaplan-Meier survivorship was 83.5% at 20 years. A diagnosis of developmental dysplasia (hazard ratio (HR) 2.199 (95% confidence interval (CI) 1.140 to 4.239); p = 0.019) and a low BMI (HR 0.931 (95% CI 0.873 to 0.994); p = 0.032) were risk factors for revision. Female sex was a risk factor only because of hip dysplasia and small component size. There were no cases of metal sensitivity associated with revision surgery. Radiological analysis showed persistent fixation in all but one hip. Conclusion. The 83.5% 20-year survivorship of this initial series surpasses that of total hip arthroplasties in use 20 years ago in these young patients, suggesting satisfactory lifelong durability of the device for almost all of the remaining patients. Cite this article: Bone Joint J 2021;103-B(7 Supple B):25–32


The Bone & Joint Journal
Vol. 99-B, Issue 12 | Pages 1590 - 1595
1 Dec 2017
Atrey A Ward SE Khoshbin A Hussain N Bogoch E Schemitsch EH Waddell JP

Aims. We present the ten-year data of a cohort of patients, aged between 18 and 65 years (mean age 52.7 years; 19 to 64), who underwent total hip arthroplasty. Patients were randomised to be treated with a cobalt-chrome (CoCr) femoral head with an ultra-high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (XLPE) or ceramic-on-ceramic (CoC) bearing surface. Patients and Methods. A total of 102 hips (91 patients) were randomised into the three groups. At ten years, 97 hips were available for radiological and functional follow-up. Two hips (two patients) had been revised (one with deep infection and one for periprosthetic fracture) and three were lost to follow-up. Radiological analysis was performed using a validated digital assessment programme to give linear, directional and volumetric wear of the two polyethylene groups. Results. There was a significantly reduced rate of steady-state linear wear with XLPE (0.07 mm/yr) compared with UHMWPE (0.37 mm/yr) (p = 0.001). Volumetric wear was also significantly reduced in the XLPE group (29.29 mm. 3. /yr) compared with the UHMWPE group (100.75mm. 3. /yr) (p = 0.0001). There were six patients with UHMWPE who had non-progressive osteolysis and none in the XLPE group. All three bearing groups had significant improvements in 12-item short form health survey scores, Western Ontario and McMaster Universities Osteoarthritis Index score and Harris Hip Score. However, the improvement in HSS was significantly less in the UHMWPE group (p = 0.0188) than in the other two groups. At ten years, the rates of volumetric and linear wear in the XLPE group remain low and predominantly below the estimated threshold for osteolysis (1 mm/yr). The rate of linear wear in the XLPE group was three times less than in the UHMWPE group at five-year follow-up and five times less at ten years. The rate of volumetric wear was also three times less in the XLPE group at ten years. Conclusion. While CoC also performs well, XLPE at ten years remains a safe and excellent bearing option in young patients, with low rates of wear and no evidence of osteolysis. Cite this article: Bone Joint J 2017;99-B:1590–5


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 768 - 774
1 Jul 2023
Wooster BM Kennedy NI Dugdale EM Sierra RJ Perry KI Berry DJ Abdel MP

Aims

Contemporary outcomes of primary total hip arthroplasties (THAs) with highly cross-linked polyethylene (HXLPE) liners in patients with inflammatory arthritis have not been well studied. This study examined the implant survivorship, complications, radiological results, and clinical outcomes of THA in patients with inflammatory arthritis.

Methods

We identified 418 hips (350 patients) with a primary diagnosis of inflammatory arthritis who underwent primary THA with HXLPE liners from January 2000 to December 2017. Of these hips, 68% had rheumatoid arthritis (n = 286), 13% ankylosing spondylitis (n = 53), 7% juvenile rheumatoid arthritis (n = 29), 6% psoriatic arthritis (n = 24), 5% systemic lupus erythematosus (n = 23), and 1% scleroderma (n = 3). Mean age was 58 years (SD 14.8), 66.3% were female (n = 277), and mean BMI was 29 kg/m2 (SD 7). Uncemented femoral components were used in 77% of cases (n = 320). Uncemented acetabular components were used in all patients. Competing risk analysis was used accounting for death. Mean follow-up was 4.5 years (2 to 18).


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 424 - 429
1 Apr 2008
Andrew JG Palan J Kurup HV Gibson P Murray DW Beard DJ

A prospective, multi-centre study was carried out on 1421 total hip replacements between January 1999 and July 2007 to examine if obesity has an effect on clinical outcomes. The patients were categorised into three groups: non-obese (body mass index (BMI) < 30 kg/m. 2. ), obese (BMI 30 to 40 kg/m. 2. ) and morbidly obese (BMI > 40 kg/m. 2. ). The primary outcome measure was the change in Oxford hip score at five years. Secondary outcome measures included dislocation and revision rates, increased haemorrhage, deep infection, deep-vein thrombosis and pulmonary embolism, mean operating time and length of hospital stay. Radiological analysis assessing heterotopic ossification, femoral osteolysis and femoral stem positioning was performed. Data were incomplete for 362 hips (25.5%). There was no difference in the change in the Oxford hip score, complication rates or radiological changes at five years between the groups. The morbidly obese group was significantly younger and required a significantly longer operating time. Obese and morbidly obese patients have as much to gain from total hip replacement as non-obese patients


Bone & Joint Open
Vol. 3, Issue 10 | Pages 795 - 803
12 Oct 2022
Liechti EF Attinger MC Hecker A Kuonen K Michel A Klenke FM

Aims

Traditionally, total hip arthroplasty (THA) templating has been performed on anteroposterior (AP) pelvis radiographs. Recently, additional AP hip radiographs have been recommended for accurate measurement of the femoral offset (FO). To verify this claim, this study aimed to establish quantitative data of the measurement error of the FO in relation to leg position and X-ray source position using a newly developed geometric model and clinical data.

Methods

We analyzed the FOs measured on AP hip and pelvis radiographs in a prospective consecutive series of 55 patients undergoing unilateral primary THA for hip osteoarthritis. To determine sample size, a power analysis was performed. Patients’ position and X-ray beam setting followed a standardized protocol to achieve reproducible projections. All images were calibrated with the KingMark calibration system. In addition, a geometric model was created to evaluate both the effects of leg position (rotation and abduction/adduction) and the effects of X-ray source position on FO measurement.


Bone & Joint Open
Vol. 4, Issue 2 | Pages 79 - 86
10 Feb 2023
McLaughlin JR Johnson MA Lee KR

Aims

The purpose of this study is to report our updated results at a minimum follow-up of 30 years using a first generation uncemented tapered femoral component in primary total hip arthroplasty (THA).

Methods

The original cohort consisted of 145 consecutive THAs performed by a single surgeon in 138 patients. A total of 37 patients (40 hips) survived a minimum of 30 years, and are the focus of this review. The femoral component used in all cases was a first-generation Taperloc with a non-modular 28 mm femoral head. Clinical follow-up at a minimum of 30 years was obtained on every living patient. Radiological follow-up at 30 years was obtained on all but four.


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 352 - 358
1 Apr 2024
Wilson JM Trousdale RT Bedard NA Lewallen DG Berry DJ Abdel MP

Aims

Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct.

Methods

We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97).


Bone & Joint Open
Vol. 4, Issue 8 | Pages 551 - 558
1 Aug 2023
Thomas J Shichman I Ohanisian L Stoops TK Lawrence KW Ashkenazi I Watson DT Schwarzkopf R

Aims

United Classification System (UCS) B2 and B3 periprosthetic fractures in total hip arthroplasties (THAs) have been commonly managed with modular tapered stems. No study has evaluated the use of monoblock fluted tapered titanium stems for this indication. This study aimed to evaluate the effects of a monoblock stems on implant survivorship, postoperative outcomes, radiological outcomes, and osseointegration following treatment of THA UCS B2 and B3 periprosthetic fractures.

Methods

A retrospective review was conducted of all patients who underwent revision THA (rTHA) for periprosthetic UCS B2 and B3 periprosthetic fracture who received a single design monoblock fluted tapered titanium stem at two large, tertiary care, academic hospitals. A total of 72 patients met inclusion and exclusion criteria (68 UCS B2, and four UCS B3 fractures). Primary outcomes of interest were radiological stem subsidence (> 5 mm), radiological osseointegration, and fracture union. Sub-analysis was also done for 46 patients with minimum one-year follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1296 - 1300
1 Oct 2009
Jackson MP Sexton SA Yeung E Walter WL Walter WK Zicat BA

The outcome of total hip replacement (THR) is potentially affected by the body mass index (BMI) of the patient. We studied the outcome of 2026 consecutive primary cementless THRs performed for osteoarthritis. The mean follow-up was 6.3 years (0 to 11.71) and no patient was lost to follow-up for survival analysis. The patients were divided into two groups according to their BMI as follows: non-obese (BMI < 30 kg/m. 2. ) and obese (BMI ≥ 30 kg/m. 2. ). The obese patient undergoing surgery was found to be significantly younger (p < 0.001). The log-rank test for equality of survival showed no difference in the mid-term survival (p = 0.552) with an estimated survival at 11 years of 95.2% (95% CI 92.5 to 98.0) in the non-obese and 96.7% (95% CI 94.9 to 98.5) in the obese groups. The clinical and radiological outcome was determined in a case-matched study performed on 134 obese individuals closely matched with 134 non-obese controls. The non-obese group was found to have a significantly higher post-operative Harris hip score (p < 0.001) and an increased range of movement, but overall satisfaction with surgery was comparable with that of the obese patients. Radiological analysis of the acetabular and femoral components showed no significant differences with regard to radiolucent lines, osteolysis, ingrowth of the femoral component, the acetabular inclination angle or alignment of the femoral component. Our results suggest that the survival of cementless THR is not adversely affected by obesity. Obese patients can therefore be counselled that despite a lower clinical score, they should expect to be satisfied with the result of their THR with a mid-term survival rate equivalent to that of non-obese patients


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 540 - 547
1 Jun 2024
Nandra RS Elnahal WA Mayne A Brash L McBryde CW Treacy RBC

Aims

The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years.

Methods

The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 898 - 906
1 Sep 2024
Kayani B Wazir MUK Mancino F Plastow R Haddad FS

Aims

The primary objective of this study was to develop a validated classification system for assessing iatrogenic bone trauma and soft-tissue injury during total hip arthroplasty (THA). The secondary objective was to compare macroscopic bone trauma and soft-tissues injury in conventional THA (CO THA) versus robotic arm-assisted THA (RO THA) using this classification system.

Methods

This study included 30 CO THAs versus 30 RO THAs performed by a single surgeon. Intraoperative photographs of the osseous acetabulum and periacetabular soft-tissues were obtained prior to implantation of the acetabular component, which were used to develop the proposed classification system. Interobserver and intraobserver variabilities of the proposed classification system were assessed.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1013 - 1018
1 Aug 2008
Goto K Kawanabe K Akiyama H Morimoto T Nakamura T

We reviewed 44 consecutive revision hip replacements in 38 patients performed using the cement-in-cement technique. All were performed for acetabular loosening in the presence of a well-fixed femoral component. The mean follow-up was 5.1 years (2 to 10.1). Radiological analysis at final follow-up indicated no loosening of the femoral component, except for one case with a continuous radiolucent line in all zones and peri-prosthetic fracture which required further revision. Peri-operative complications included nine proximal femoral fractures (20.4%) and perforation of the proximal femur in one hip. In five hips wiring or fixation with a braided suture was undertaken but no additional augmentation was required. There was an improvement in the mean Japanese Orthopaedic Association score from 55.5 (28 to 81) pre-operatively to 77.8 (40 to 95) at final follow-up (p < 0.001). Revision using a cement-in-cement technique allows increased exposure for acetabular revision and is effective in the medium term. Further follow-up is required to assess the long-term results in the light of in vitro studies which have questioned the quality of the cement-in-cement bond


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 396 - 403
1 May 1998
Böhm P Bösche R

We reviewed 264 consecutive primary total hip replacements in 244 patients in which the Harris-Galante I porous-coated acetabular component had been used. The mean follow-up was 95 months (69 to 132). In 221 arthroplasties a 32 mm ceramic head had been used, and in the other 43 one of cobalt-chrome alloy. There were 124 women and 120 men with a mean age at operation of 56.8 years (21 to 83). Survival analysis of the acetabular components was performed using the following definitions of failure: 1) infection; 2) removal because of aseptic loosening; 3) removal for any cause; and 4) a worst-case analysis including removal of the cup or infection or loss to follow-up. Two well-fixed cups had been removed because of deep infection and three, similarly sound cups had been exchanged at the time of femoral revision. None of the acetabular components had required revision for aseptic loosening. After 11 years survival was 97.7% using the worst-case criterion. Radiological analysis of 187 acetabular components at a mean of 82 months (41 to 132) revealed no case of aseptic loosening. Pelvic osteolysis was seen in only one patient, but was not progressive and showed signs of regression. The results using these acetabular components are encouraging but the need for regular follow-up remains since complications such as significant wear of the liner, massive osteolysis of the pelvis, aseptic loosening and late infection may increase in the second decade


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 7 | Pages 952 - 958
1 Sep 2000
Morrey BF Adams RA Kessler M

Between 1985 and 1993, 146 patients (162 hips) had total hip replacement (THR) using a conservative uncemented femoral component. The mean age of the patients was 50.8 years and the mean follow-up was 6.2 years (2 to 13). One patient was lost to follow-up, one died within two years of surgery and one had a revision procedure after a fracture sustained in a road-traffic accident. For the remaining 159, Kaplan-Meier survival analysis was calculated for the incidence of revision because of mechanical loosening or osteolysis. Survival without mechanical loosening at both five and ten years was 98.2%. Survival without osteolysis was 99% at five and 91% at ten years. The Harris hip score improved from a mean of 66.3 before to 90.4 at follow-up. Of particular note is the lack of thigh pain in this group. Radiological analysis showed that 139 stems (88%) had no measurable subsidence, 8 (5%) had less than 2 mm and 12 (7%) had more than 2 mm. Two of the eight and one of the 12 were revised for mechanical loosening. Nine hips were revised for late loosening associated with osteolysis. No reaming of the femoral canal was associated with statistically significant less blood loss compared with a comparable control group of uncemented implants (p < 0.0001). Our study suggests that using a conservative femoral implant does not protect against wear debris but the reliable mechanical stability (98.2%) makes this an attractive design of implant particularly for young patients


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 16 - 22
1 Jan 2008
Yates PJ Burston BJ Whitley E Bannister GC

We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1766 - 1773
1 Dec 2021
Sculco PK Windsor EN Jerabek SA Mayman DJ Elbuluk A Buckland AJ Vigdorchik JM

Aims

Spinopelvic mobility plays an important role in functional acetabular component position following total hip arthroplasty (THA). The primary aim of this study was to determine if spinopelvic hypermobility persists or resolves following THA. Our second aim was to identify patient demographic or radiological factors associated with hypermobility and resolution of hypermobility after THA.

Methods

This study investigated patients with preoperative posterior hypermobility, defined as a change in sacral slope (SS) from standing to sitting (ΔSSstand-sit) ≥ 30°. Radiological spinopelvic parameters, including SS, pelvic incidence (PI), lumbar lordosis (LL), PI-LL mismatch, anterior pelvic plane tilt (APPt), and spinopelvic tilt (SPT), were measured on preoperative imaging, and at six weeks and a minimum of one year postoperatively. The severity of bilateral hip osteoarthritis (OA) was graded using Kellgren-Lawrence criteria.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1215 - 1221
1 Jul 2021
Kennedy JW Ng NYB Young D Kane N Marsh AG Meek RMD

Aims

Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components.

Methods

From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 46 - 52
1 Jul 2021
McGoldrick NP Fischman D Nicol GM Kreviazuk C Grammatopoulos G Beaulé PE

Aims

The aim of this study was to radiologically evaluate the quality of cement mantle and alignment achieved with a polished tapered cemented femoral stem inserted through the anterior approach and compared with the posterior approach.

Methods

A comparative retrospective study of 115 consecutive hybrid total hip arthroplasties or cemented hemiarthroplasties in 110 patients, performed through anterior (n = 58) or posterior approach (n = 57) using a collarless polished taper-slip femoral stem, was conducted. Cement mantle quality and thickness were assessed in both planes. Radiological outcomes were compared between groups.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 122 - 128
1 Jul 2021
Tibbo ME Limberg AK Gausden EB Huang P Perry KI Yuan BJ Berry DJ Abdel MP

Aims

The prevalence of ipsilateral total hip arthroplasty (THA) and total knee arthroplasty (TKA) is rising in concert with life expectancy, putting more patients at risk for interprosthetic femur fractures (IPFFs). Our study aimed to assess treatment methodologies, implant survivorship, and IPFF clinical outcomes.

Methods

A total of 76 patients treated for an IPFF from February 1985 to April 2018 were reviewed. Prior to fracture, at the hip/knee sites respectively, 46 femora had primary/primary, 21 had revision/primary, three had primary/revision, and six had revision/revision components. Mean age and BMI were 74 years (33 to 99) and 30 kg/m2 (21 to 46), respectively. Mean follow-up after fracture treatment was seven years (2 to 24).


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 492 - 499
1 Mar 2021
Garcia-Rey E Saldaña L Garcia-Cimbrelo E

Aims

Bone stock restoration of acetabular bone defects using impaction bone grafting (IBG) in total hip arthroplasty may facilitate future re-revision in the event of failure of the reconstruction. We hypothesized that the acetabular bone defect during re-revision surgery after IBG was smaller than during the previous revision surgery. The clinical and radiological results of re-revisions with repeated use of IBG were also analyzed.

Methods

In a series of 382 acetabular revisions using IBG and a cemented component, 45 hips (45 patients) that had failed due to aseptic loosening were re-revised between 1992 and 2016. Acetabular bone defects graded according to Paprosky during the first and the re-revision surgery were compared. Clinical and radiological findings were analyzed over time. Survival analysis was performed using a competing risk analysis.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 321 - 328
1 Feb 2021
Vandeputte F Vanbiervliet J Sarac C Driesen R Corten K

Aims

Optimal exposure through the direct anterior approach (DAA) for total hip arthroplasty (THA) conducted on a regular operating theatre table is achieved with a standardized capsular releasing sequence in which the anterior capsule can be preserved or resected. We hypothesized that clinical outcomes and implant positioning would not be different in case a capsular sparing (CS) technique would be compared to capsular resection (CR).

Methods

In this prospective trial, 219 hips in 190 patients were randomized to either the CS (n = 104) or CR (n = 115) cohort. In the CS cohort, a medial based anterior flap was created and sutured back in place at the end of the procedure. The anterior capsule was resected in the CR cohort. Primary outcome was defined as the difference in patient-reported outcome measures (PROMs) after one year. PROMs (Harris Hip Score (HHS), Hip disability and Osteoarthritis Outcome Score (HOOS), and Short Form 36 Item Health Survey (SF-36)) were collected preoperatively and one year postoperatively. Radiological parameters were analyzed to assess implant positioning and implant ingrowth. Adverse events were monitored.


Bone & Joint Open
Vol. 2, Issue 1 | Pages 33 - 39
14 Jan 2021
McLaughlin JR Lee KR Johnson MA

Aims

We present the clinical and radiological results at a minimum follow-up of 20 years using a second-generation uncemented total hip arthroplasty (THA). These results are compared to our previously published results using a first-generation hip arthroplasty followed for 20 years.

Methods

A total of 62 uncemented THAs in 60 patients were performed between 1993 and 1994. The titanium femoral component used in all cases was a Taperloc with a reduced distal stem. The acetabular component was a fully porous coated threaded hemispheric titanium shell (T-Tap ST). The outcome of every femoral and acetabular component with regard to retention or revision was determined for all 62 THAs. Complete clinical follow-up at a minimum of 20 years was obtained on every living patient. Radiological follow-up was obtained on all but one.


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1319 - 1323
3 Oct 2020
Khatun F Gill DF Atrey A Porteous M

Aims

We present the results, in terms of survival, clinical outcome, and radiological appearance at 20 years, in a cohort of 225 cemented Exeter Universal femoral components (Stryker, Newbury, UK) implanted in 207 patients, at a district general hospital.

Methods

All patients in this study had a total hip arthroplasty (THA) using an Exeter Universal femoral component with a cemented (n = 215) or cementless (n = 10) acetabular component. Clinical and radiological data were collected prospectively at one year, five years, and every five years thereafter. Patients lost to radiological and clinical follow-up (five) were cross-referenced with National Joint Registry (NJR) data and general practitioner (GP) records to assess whether they had undergone revision for any reason.


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 390 - 395
1 Apr 2019
Yasunaga Y Tanaka R Mifuji K Shoji T Yamasaki T Adachi N Ochi M

Aims

The aim of this study was to report the long-term results of rotational acetabular osteotomy (RAO) for symptomatic hip dysplasia in patients aged younger than 21 years at the time of surgery.

Patients and Methods

We evaluated 31 patients (37 hips) aged younger than 21 years at the time of surgery retrospectively. There were 29 female and two male patients. Their mean age at the time of surgery was 17.4 years (12 to 21). The mean follow-up was 17.9 years (7 to 30). The RAO was combined with a varus or valgus femoral osteotomy or a greater trochanteric displacement in eight hips, as instability or congruence of the hip could not be corrected adequately using RAO alone.


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 378 - 385
1 Apr 2019
García-Rey E Carbonell-Escobar R Cordero-Ampuero J García-Cimbrelo E

Aims

We previously reported the long-term results of the cementless Duraloc-Profile total hip arthroplasty (THA) system in a 12- to 15-year follow-up study. In this paper, we provide an update on the clinical and radiological results of a previously reported cohort of patients at 23 to 26 years´ follow-up.

Patients and Methods

Of the 99 original patients (111 hips), 73 patients (82 hips) with a mean age of 56.8 years (21 to 70) were available for clinical and radiological study at a minimum follow-up of 23 years. There were 40 female patients (44 hips) and 33 male patients (38 hips).


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 787 - 792
1 Jul 2019
Goto K Kuroda Y Kawai T Kawanabe K Matsuda S

Aims

In the 1990s, a bioactive bone cement (BABC) containing apatite-wollastonite glass-ceramic (AW-GC) powder and bisphenol-a-glycidyl methacrylate resin was developed at our hospital. In 1996, we used BABC to fix the acetabular component in primary total hip arthroplasty (THA) in 20 patients as part of a clinical trial. The purpose of this study was to investigate the long-term results of primary THA using BABC.

Patients and Methods

A total of 20 patients (three men and 17 women) with a mean age of 57.4 years (40 to 71), a mean body weight of 52.3 kg (39 to 64), and a mean body mass index (BMI) of 23.0 kg/m2 (19.8 to 28.6) were evaluated clinically and radiologically. Survival analyses were undertaken, and wear analyses were carried out using a computer-aided method.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 749 - 758
1 Jun 2017
García-Rey E Cruz-Pardos A García-Cimbrelo E

Aims

To determine the effect of a change in design of a cementless ceramic acetabular component in fixation and clinical outcome after total hip arthroplasty

Patients and Methods

We compared 342 hips (302 patients) operated between 1999 and 2005 with a relatively smooth hydroxyapatite coated acetabular component (group 1), and 337 hips (310 patients) operated between 2006 and 2011 using a similar acetabular component with a macrotexture on the entire outer surface of the component (group 2). The mean age of the patients was 53.5 (14 to 70) in group 1 and 53.0 (15 to 70) in group 2. The mean follow-up was 12.7 years (10 to 17) for group 1 and 7.2 years (4 to 10) for group 2.


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 192 - 198
1 Feb 2017
Schmitz MWJL Bronsema E de Kam DCJ Gardeniers JWM Veth RPH Schreurs BW

Aims

We carried out a further study of the long-term results of the cemented Exeter femoral component in patients under the age of 40 with a mean follow-up of 13.6 years (10 to 20).

Patients and Methods

We reviewed our original cohort of 104 cemented Exeter stems in 78 consecutive patients with a mean age of 31 years (16 to 39). Only one patient was lost to radiological follow-up.


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1597 - 1603
1 Dec 2016
Meermans G Doorn JV Kats J

Aims

One goal of total hip arthroplasty is to restore normal hip anatomy. The aim of this study was to compare displacement of the centre of rotation (COR) using a standard reaming technique with a technique in which the acetabulum was reamed immediately peripherally and referenced off the rim.

Patients and Methods

In the first cohort the acetabulum was reamed to the floor followed by sequentially larger reamers. In the second cohort the acetabulum was only reamed peripherally, starting with a reamer the same size as the native femoral head. Anteroposterior pelvic radiographs were analysed for acetabular floor depth and vertical and horizontal position of the COR.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 617 - 622
1 May 2015
Haddad FS Konan S Tahmassebi J

The aim of this study was to evaluate the ten-year clinical and functional outcome of hip resurfacing and to compare it with that of cementless hip arthroplasty in patients under the age of 55 years.

Between 1999 and 2002, 80 patients were enrolled into the study: 24 were randomised (11 to hip resurfacing, 13 to total hip arthroplasty), 18 refused hip resurfacing and chose cementless total hip arthroplasty with a 32 mm bearing, and 38 insisted on resurfacing. The mean follow-up for all patients was 12.1 years (10 to 14).

Patients were assessed clinically and radiologically at one year, five years and ten years. Outcome measures included EuroQol EQ5D, Oxford, Harris hip, University of California Los Angeles and University College Hospital functional scores.

No differences were seen between the two groups in the Oxford or Harris hip scores or in the quality of life scores. Despite a similar aspiration to activity pre-operatively, a higher proportion of patients with a hip resurfacing were running and involved in sport and heavy manual labour after ten years.

We found significantly higher function scores in patients who had undergone hip resurfacing than in those with a cementless hip arthroplasty at ten years. This suggests a functional advantage for hip resurfacing. There were no other attendant problems.

Cite this article: Bone Joint J 2015; 97-B:617–22.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1172 - 1177
1 Sep 2013
Matharu GS McBryde CW Pynsent WB Pynsent PB Treacy RBC

We report the long-term survival and functional outcome of the Birmingham Hip Resurfacing (BHR) in patients aged < 50 years at operation, and explore the factors affecting survival. Between 1997 and 2006, a total of 447 BHRs were implanted in 393 patients (mean age 41.5 years (14.9 to 49.9)) by one designing surgeon. The mean follow-up was 10.1 years (5.2 to 14.7), with no loss to follow-up. In all, 16 hips (3.6%) in 15 patients were revised, giving an overall cumulative survival of 96.3% (95% confidence interval (CI) 93.7 to 98.3) at ten years and 94.1% (95% CI 84.9 to 97.3) at 14 years. Using aseptic revision as the endpoint, the survival for men with primary osteoarthritis (n = 195) was 100% (95% CI 100 to 100) at both ten years and 14 years, and in women with primary osteoarthritis (n = 109) it was 96.1% (95% CI 90.1 to 99.9) at ten years and 91.2% (95% CI 68.6 to 98.7) at 14 years. Female gender (p = 0.047) and decreasing femoral head size (p = 0.044) were significantly associated with an increased risk of revision. The median Oxford hip score (OHS, modified as a percentage with 100% indicating worst outcome) at last follow-up was 4.2% (46 of 48; interquartile range (IQR) 0% to 24%) and the median University of California, Los Angeles (UCLA) score was 6.0 (IQR 5 to 8). Men had significantly better OHS (p = 0.02) and UCLA scores (p = 0.01) than women. The BHR provides excellent survival and functional results in men into the second decade, with good results achieved in appropriately selected women.

Cite this article: Bone Joint J 2013;95-B:1172–7.


The Bone & Joint Journal
Vol. 96-B, Issue 9 | Pages 1207 - 1213
1 Sep 2014
te Stroet MAJ Bronsema E Rijnen WHC Gardeniers JWM Schreurs BW

In this retrospective study, we investigated the results of revision total hip replacement (THR) using a cemented long-stemmed Exeter femoral component, with a minimum length of 205 mm in patients with extensive femoral bone defects. The study included 37 consecutive patients with a mean age of 76 years (39 to 93) and a mean follow-up of nine years (5 to 16). A total of 26 patients (70%) had a pre-operative Endo-Klinik score of 3 or 4. Impaction bone grafting was used in 24 patients (65%). At the time of evaluation, 22 patients (59%) were still alive and were evaluated clinically and radiologically. A total of 14 patients died during follow-up and their data were included until the time of their death. One reconstruction failed after five years and five months owing to recurrent dislocation: the hip was converted to an excision arthroplasty. Intra-operative fractures or fissures were encountered in nine patients (24%), but none occurred during impaction of the bone graft. Post-operative peri-prosthetic fractures occurred in two patients (5%); both were treated with plate fixation. At nine years, survival with the endpoint of all-cause re-revision was 96.3% (95% CI 76.4 to 99.5); including re-operations for any reason, it was 80.7% (95% CI 56.3 to 92.3%). There were no re-revisions for aseptic loosening.

The survival of long stem cemented femoral components following revision THR is satisfactory in a fragile population with extensive femoral defects.

Cite this article: Bone Joint J 2014; 96-B:1207–13.


The Bone & Joint Journal
Vol. 96-B, Issue 4 | Pages 455 - 461
1 Apr 2014
Evola FR Evola G Graceffa A Sessa A Pavone V Costarella L Sessa G Avondo S

In 2012 we reviewed a consecutive series of 92 uncemented THRs performed between 1986 and 1991 at our institution using the CLS Spotorno stem, in order to assess clinical outcome and radiographic data at a minimum of 21 years. The series comprised 92 patients with a mean age at surgery of 59.6 years (39 to 77) (M:F 43;49).

At the time of this review, seven (7.6%) patients had died and two (2.2%) were lost to follow-up. The 23-year Kaplan–Meier survival rates were 91.5% (95% confidence intervals (CI) 85.4% to 97.6%; 55 hips at risk) and 80.3% (95% CI, 71.8% to 88.7%; 48 hips at risk) respectively, with revision of the femoral stem or of any component as endpoints. At the time of this review, 76 patients without stem revision were assessed clinically and radiologically (mean follow-up 24.0 years (21.5 to 26.5)). For the 76 unrevised hips the mean Harris hip score was 87.1 (65 to 97). Femoral osteolysis was detected in five hips (6.6%) only in Gruen zone 7. Undersized stems were at higher risk of revision owing to aseptic loosening (p = 0.0003). Patients implanted with the stem in a varus position were at higher risk of femoral cortical hypertrophy and thigh pain (p = 0.0006 and p = 0.0007, respectively).

In our study, survival, clinical outcome and radiographic data remained excellent in the third decade after implantation. Nonetheless, undersized stems were at higher risk of revision owing to aseptic loosening.

Cite this article: Bone Joint J 2014;96-B:455–61.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1035 - 1039
1 Aug 2013
Ebreo D Bell PJ Arshad H Donell ST Toms A Nolan JF

Metal artefact reduction (MAR) MRI is now widely considered to be the standard for imaging metal-on-metal (MoM) hip implants. The Medicines and Healthcare Products Regulatory Agency (MHRA) has recommended cross-sectional imaging for all patients with symptomatic MoM bearings. This paper describes the natural history of MoM disease in a 28 mm MoM total hip replacement (THR) using MAR MRI. Inclusion criteria were patients with MoM THRs who had not been revised and had at least two serial MAR MRI scans. All examinations were reported by an experienced observer and classified as A (normal), B (infection) or C1–C3 (mild, moderate, severe MoM-related abnormalities). Between 2002 and 2011 a total of 239 MRIs were performed on 80 patients (two to four scans per THR); 63 initial MRIs (61%) were normal. On subsequent MRIs, six initially normal scans (9.5%) showed progression to a disease state; 15 (15%) of 103 THRs with sequential scans demonstrated worsening disease on subsequent imaging.

Most patients with a MoM THR who do not undergo early revision have normal MRI scans. Late progression (from normal to abnormal, or from mild to more severe MoM disease) is not common and takes place over several years.

Cite this article: Bone Joint J 2013;95-B:1035–9.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1464 - 1473
1 Nov 2013
Vendittoli P Rivière C Roy AG Barry J Lusignan D Lavigne M

A total of 219 hips in 192 patients aged between 18 and 65 years were randomised to 28-mm metal-on-metal uncemented total hip replacements (THRs, 107 hips) or hybrid hip resurfacing (HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3) there was no significant difference between the THR and HR groups regarding rate of revision (4.0% (4 of 99) vs 5.8% (6 of 104), p = 0.569) or re-operation rates without revision (5.1% (5 of 99) vs 2.9% (3 of 104), p = 0.428). In the THR group one recurrent dislocation, two late deep infections and one peri-prosthetic fracture required revision, whereas in the HR group five patients underwent revision for femoral head loosening and one for adverse reaction to metal debris. The mean University of California, Los Angeles activity scores were significantly higher in HR (7.5 (sd 1.7) vs 6.9 (sd 1.7), p = 0.035), but similar mean Western Ontario and McMaster Universities Osteoarthritis Index scores were obtained (5.8 (sd 9.5) in HR vs 5.1 (sd 8.9) in THR, p = 0.615) at the last follow-up. Osteolysis was found in 30 of 81 THR patients (37.4%), mostly in the proximal femur, compared with two of 83 HR patients (2.4%) (p < 0.001). At five years the mean metal ion levels were < 2.5 μg/l for cobalt and chromium in both groups; only titanium was significantly higher in the HR group (p = 0.001). Although revision rates and functional scores were similar in both groups at mid-term, long-term survival analysis is necessary to determine whether one procedure is more advantageous than the other.

Cite this article: Bone Joint J 2013;95-B:1464–73.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1522 - 1526
1 Nov 2010
McGrory BJ McGrory CP Barbour L Barbour B

Dislocation is a common and well-studied complication after total hip replacement. However, subluxation, which we define as a clinically recognised episode of incomplete movement of the femoral head outside the acetabulum with spontaneous reduction, has not been studied previously. Out of a total of 2521 hip replacements performed over 12 years by one surgeon, 30 patients experienced subluxations which occurred in 31 arthroplasties. Data were collected prospectively with a minimum follow-up of two years. Subluxation occurred significantly more frequently after revision than after primary hip replacement, and resolved in 19 of 31 cases (61.3%). In six of the 31 hips (19.4%) the patient subsequently dislocated the affected hip, and in six hips (19.4%) intermittent subluxation continued. Four patients had a revision operation for instability, three for recurrent dislocation and one for recurrent subluxation. Clinical and radiological comparisons with a matched group of stable total hips showed no correlation with demographic or radiological parameters.

Patients with subluxing hips reported significantly more concern that their hip would dislocate, more often changed their behaviour to prevent instability and had lower postoperative Harris hip scores than patients with stable replacements.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 608 - 615
1 May 2011
Bolland BJRF Culliford DJ Langton DJ Millington JPS Arden NK Latham JM

This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83).

Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem.

The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4).

This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1475 - 1481
1 Nov 2012
Berton C Puskas GJ Christofilopoulos P Stern R Hoffmeyer P Lübbeke A

There are no recent studies comparing cable with wire for the fixation of osteotomies or fractures in total hip replacement (THR). Our objective was to evaluate the five-year clinical and radiological outcomes and complication rates of the two techniques. We undertook a review including all primary and revision THRs performed in one hospital between 1996 and 2005 using cable or wire fixation. Clinical and radiological evaluation was performed five years post-operatively. Cables were used in 51 THRs and wires in 126, and of these, 36 THRs with cable (71%) and 101 with wire (80%) were evaluated at follow-up. The five-year radiographs available for 33 cable and 91 wire THRs revealed rates of breakage of fixation of 12 of 33 (36%) and 42 of 91 (46%), respectively. With cable there was a significantly higher risk of metal debris (68% vs 9%; adjusted relative risk (RR) 6.6; 95% confidence interval (CI) 3.0 to 14.1), nonunion (36% vs 21%; adjusted RR 2.0; 95% CI 1.0 to 3.9) and osteolysis around the material, acetabulum or femur (61% vs 19%; adjusted RR 3.9; 95% CI 2.3 to 6.5). Cable breakage increased the risk of osteolysis to 83%. There was a trend towards foreign-body reaction and increased infection with cables. Clinical results did not differ between the groups.

In conclusion, we found a higher incidence of complications and a trend towards increased infection and foreign-body reaction with the use of cables.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 609 - 614
1 May 2012
Flecher X Blanc G Sainsous B Parratte S Argenson J

We describe the results of 81 consecutive revision total hip replacements with impaction grafting in 79 patients using a collared polished chrome–cobalt stem, customised in length according to the extent of distal bone loss. Our hypothesis was that the features of this stem would reduce the rate of femoral fracture and subsidence of the stem.

The mean follow-up was 12 years (8 to 15). No intra-operative fracture or significant subsidence occurred. Only one patient suffered a post-operative diaphyseal fracture, which was associated with a fall. All but one femur showed incorporation of the graft. No revision for aseptic loosening was recorded.

The rate of survival of the femoral component at 12 years, using further femoral revision as the endpoint, was 100% (95% confidence interval (CI) 95.9 to 100), and at nine years using re-operation for any reason as the endpoint, was 94.6% (95% CI 92.0 to 97.2).

These results suggest that a customised cemented polished stem individually adapted to the extent of bone loss and with a collar may reduce subsidence and the rate of fracture while maintaining the durability of the fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 304 - 309
1 Mar 2009
Kerboull L Hamadouche M Kerboull M

We describe 129 consecutive revision total hip replacements using a Charnley-Kerboull femoral component of standard length with impaction allografting. The mean follow-up was 8.2 years (2 to 16). Additionally, extramedullary reinforcement was performed using struts of cortical allograft in 49 hips and cerclage wires in 30.

There was one intra-operative fracture of the femur but none later. Two femoral components subsided by 5 mm and 8 mm respectively, and were considered to be radiological failures. No further revision of a femoral component was required. The rate of survival of the femoral component at nine years, using radiological failure as the endpoint, was 98%. Our study showed that impaction grafting in association with a Charnley-Kerboull femoral component has a low rate of subsidence. Reconstruction of deficiencies of distal bone with struts of cortical allograft appeared to be an efficient way of preventing postoperative femoral fracture for up to 16 years.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 145 - 151
1 Feb 2012
Henderson RA Lachiewicz PF

Persistent groin pain after seemingly successful total hip replacement (THR) appears to have become more common. Recent studies have indicated a high incidence after metal-on-polyethylene and metal-on-metal conventional THR and it has been documented in up to 18% of patients after metal-on-metal resurfacing. There are many causes, including acetabular loosening, stress fracture, and iliopsoas tendonitis and impingement. The evaluation of this problem requires a careful history and examination, plain radiographs and an algorithmic approach to special diagnostic imaging and tests. Non-operative treatment is not usually successful. Specific operative treatment depending on the cause of the pain usually involves revision of the acetabular component, iliopsoas tenotomy or other procedures, and is usually successful. Here, an appropriate algorithm is described.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1017 - 1020
1 Aug 2011
Stafford GH Islam SU Witt JD

Ceramic-on-ceramic bearings in hip replacement have low rates of wear and are increasingly being used in young adults. Our aim was to determine the incidence of audible phenomena or other bearing-related complications.

We retrospectively analysed 250 ceramic-on-ceramic hip replacements in 224 patients which had been implanted between April 2000 and December 2007. The mean age of the patients at operation was 44 years (14 to 83) and all the operations were performed using the same surgical technique at a single centre.

At a mean follow-up of 59 months (24 to 94), the mean Oxford hip score was 40.89 (11 to 48). There were six revisions, three of which were for impingement-related complications. No patient reported squeaking, but six described grinding or clicking, which was usually associated with deep flexion. No radiological evidence of osteolysis or migration of the components was observed in any hip.

The early to mid-term results of contemporary ceramic-on-ceramic hip replacement show promising results with few concerns in terms of noise and squeaking. Positioning of the acetabular component remains critical in regard to the reduction of other impingement-related complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 6 | Pages 731 - 737
1 Jun 2008
Kim Y Kim J

It has been suggested that the wear of ultra-high molecular weight polyethylene (UHMWPE) in total hip replacement is substantially reduced when the femoral head is ceramic rather than metal. However, studies of alumina and zirconia ceramic femoral heads on the penetration of an UHMWPE liner in vivo have given conflicting results.

The purpose of this study was to examine the surface characteristics of 30 alumina and 24 zirconia ceramic femoral heads and to identify any phase transformation in the zirconia heads. We also studied the penetration rate of alumina and zirconia heads into contemporary UHMWPE liners. The alumina heads had been implanted for a mean of 11.3 years (8.1 to 16.2) and zirconia heads for a mean of 9.8 years (7.5 to 15).

The mean surface roughness values of the explanted alumina heads (Ra 40.12 nm and Rpm 578.34 nm) were similar to those for the explanted zirconia heads (Ra 36.21 nm and Rpm 607.34 nm). The mean value of the monoclinic phase of two control zirconia heads was 1% (0.8% to 1.5%) and 1.2% (0.9% to 1.3%), respectively. The mean value of the monoclinic phase of 24 explanted zirconia heads was 7.3% (1% to 26%).

In the alumina group, the mean linear penetration rate of the UMWPE liner was 0.10 mm/yr (0.09 to 0.12) in hips with low Ra and Rpm values (13.22 nm and 85.91 nm, respectively). The mean linear penetration rate of the UHMWPE liner was 0.13 mm/yr (0.07 to 0.23) in hips with high Ra and Rpm values (198.72 nm and 1329 nm, respectively). This difference was significant (p = 0.041).

In the zirconia head group, the mean linear penetration rate of the UHMWPE liner was 0.09 mm/yr (0.07 to 0.14) in hips with low Ra and Rpm values (12.78 nm and 92.99 nm, respectively). The mean linear penetration rate of the UHMWPE liner was 0.12 mm/yr (0.08 to 0.22) in hips with high Ra and Rpm values (199.21 nm and 1381 nm, respectively). This difference was significant (p = 0.039).

The explanted zirconia heads which had a minimal phase transformation had similar surface roughness and a similar penetration rate of UHMWPE liner as the explanted alumina head.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 315 - 320
1 Mar 2006
Field RE Singh PJ Latif AMH Cronin MD Matthews DJ

We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different registrars. There were three withdrawals from the study because of psychiatric illness, a deep infection and a recurrent dislocation. Five deaths occurred prior to five-year follow-up and one patient withdrew from clinical review.

In the remaining 51 hips the mean pre-operative Oxford hip score was 47 points which decreased to 19 points at five years (45 hips). Of the stems 49 (98%) were implanted within 1° of neutral in the femoral canal. The mean migration of the stem at five years was 1.9 mm and the survivorship for aseptic loosening was 100%. There was no significant difference in outcome between the consultant and registrar groups. At five years, the results were comparable with those of other polished, tapered, cemented stems. Long-term surveillance continues.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 12 | Pages 1636 - 1641
1 Dec 2010
McLaughlin JR Lee KR

We reviewed 123 second-generation uncemented total hip replacements performed on 115 patients by a single surgeon between 1993 and 1994. The acetabular component used in all cases was a fully porous-coated threaded hemispheric titanium shell (T-Tap ST) with a calcium ion stearate-free, isostatically compression-moulded polyethylene liner. The titanium femoral component used was a Taperloc with a reduced distal stem. No patient was lost to follow-up. Complete clinical and radiological follow-up was obtained for all 123 hips at a mean of 14 years (12 to 16). One femoral component was revised after a fracture, and three acetabular components for aseptic loosening. No additional femoral or acetabular components were judged loose by radiological criteria. Mild proximal femoral osteolysis was identified in two hips and minor acetabular osteolysis was present in four. The mean rate of penetration of the femoral head was 0.036 mm/year (0.000 to 0.227).

These findings suggest that refinements in component design may be associated with excellent long-term fixation in cementless primary total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 456 - 463
1 Apr 2011
Lübbeke A Garavaglia G Barea C Stern R Peter R Hoffmeyer P

We conducted a longitudinal study including patients with the same type of primary hybrid total hip replacement and evaluated patient activity and femoral osteolysis at either five or ten years post-operatively. Activity was measured using the University of California, Los Angeles scale. The primary outcome was the radiological assessment of femoral osteolysis. Secondary outcomes were revision of the femoral component for aseptic loosening and the patients’ quality of life. Of 503 hip replacements in 433 patients with a mean age of 67.7 years (30 to 91), 241 (48%) were seen at five and 262 (52%) at ten years post-operatively. Osteolytic lesions were identified in nine of 166 total hip replacements (5.4%) in patients with low activity, 21 of 279 (7.5%) with moderate activity, and 14 of 58 (24.1%) patients with high activity. The risk of osteolysis increased with participation in a greater number of sporting activities. In multivariate logistic regression adjusting for age, gender, body mass index and the inclination angle of the acetabular component, the adjusted odds ratio for osteolysis comparing high vs moderate activity was 3.6 (95% confidence interval 1.6 to 8.3). Stratification for the cementing technique revealed that lower quality cementing increased the effect of high activity on osteolysis. Revision for aseptic loosening was most frequent with high activity. Patients with the highest activity had the best outcome and highest satisfaction.

In conclusion, of patients engaged in high activity, 24% had developed femoral osteolysis five to ten years post-operatively.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1501 - 1508
1 Nov 2010
Donell ST Darrah C Nolan JF Wimhurst J Toms A Barker THW Case CP Tucker JK

Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell.

Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%).

At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis.

Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 20 - 27
1 Jan 2010
Daniel J Ziaee H Kamali A Pradhan C Band T McMinn DJW

Second-generation metal-on-metal bearings were introduced as a response to the considerable incidence of wear-induced failures associated with conventional replacements, especially in young patients. We present the results at ten years of a consecutive series of patients treated using a metal-on-metal hip resurfacing. A distinct feature of the bearings used in our series was that they had been subjected to double-heat treatments during the post-casting phase of their manufacture. In the past these bearings had not been subjected to thermal treatments, making this a unique metal-on-metal bearing which had not been used before in clinical practice.

We report the outcome of 184 consecutive hips (160 patients) treated using a hybrid-fixed metal-on-metal hip resurfacing during 1996. Patients were invited for a clinicoradiological follow-up at a minimum of ten years. The Oxford hip score and anteroposterior and lateral radiographs were obtained. The mean age at operation was 54 years (21 to 75). A series of 107 consecutive hips (99 patients) who received the same prosthesis, but subjected to a single thermal treatment after being cast, between March 1994 and December 1995, were used as a control group for comparison.

In the 1994 to 1995 group seven patients (seven hips) died from unrelated causes and there were four revisions (4%) for osteolysis and aseptic loosening. In the 1996 group nine patients died at a mean of 6.9 years after operation because of unrelated causes. There were 30 revisions (16%) at a mean of 7.3 years (1.2 to 10.9), one for infection at 1.2 years and 29 for osteolysis and aseptic loosening. Furthermore, in the latter group there were radiological signs of failure in 27 (24%) of the 111 surviving hips.

The magnitude of the problem of osteolysis and aseptic loosening in the 1996 cohort did not become obvious until five years after the operation. Our results indicate that double-heat treatments of metal-on-metal bearings can lead to an increased incidence of wear-induced osteolysis.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 170 - 175
1 Feb 2009
Young L Duckett S Dunn A

We describe the survivorship of the Exeter femoral component in a District General Hospital. Between 1994 and 1996, 230 Exeter Universal cemented femoral components were implanted in 215 patients who were reviewed at a mean of 11.2 years (10 to 13). We used one acetabular implant, the Elite Ogee component, in 218 of the 230 hips.

During the period of this study 76 patients (79 hips) died. Of the remaining 139 patients (151 hips), 121 were able to attend for radiological analysis at a minimum of ten years. One patient was lost to follow-up.

No femoral component was revised for aseptic loosening. Three hips were revised for deep infection and six acetabular components required revision, four for loosening and two for recurrent dislocation. Taking the ‘worst-case scenario’ including the one patient lost to follow-up, the overall survival rate was 94.4% at 13 years.

Our results confirm excellent medium-term results for the Exeter Universal femoral component, implanted in a general setting. The excellent survival of this femoral component, when used in combination with the Ogee acetabular component, suggests that this is a successful pairing.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 28 - 37
1 Jan 2010
Jameson SS Langton DJ Nargol AVF

We present the early clinical and radiological results of Articular Surface Replacement (ASR) resurfacings in 214 hips (192 patients) with a mean follow-up of 43 months (30 to 57). The mean age of the patients was 56 years (28 to 74) and 85 hips (40%) were in 78 women.

The mean Harris hip score improved from 52 (11 to 81) to 95 (27 to 100) at two years and the mean University of California, Los Angeles activity score from 3.9 (1 to 10) to 7.4 (2 to 10) in the same period. Narrowing of the neck (to a maximum of 9%) was noted in 124 of 209 hips (60%). There were 12 revisions (5.6%) involving four (1.9%) early fractures of the femoral neck and two (0.9%) episodes of collapse of the femoral head secondary to avascular necrosis. Six patients (2.8%) had failure related to metal wear debris. The overall survival for our series was 93% (95% confidence interval 80 to 98) and 89% (95% confidence interval 82 to 96) for hips with acetabular components smaller than 56 mm in diameter.

The ASR implant has a lower diametrical clearance and a subhemispherical acetabular component when compared with other more frequently implanted metal-on-metal hip resurfacings. These changes may contribute to the higher failure rate than in other series, compared with other designs. Given our poor results with the small components we are no longer implanting the smaller size.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 327 - 332
1 Mar 2009
García-Rey E García-Cimbrelo E Cordero-Ampuero J

We reviewed 111 hemispherical Duraloc series-500 acetabular components with a minimum follow-up of 12 years. The mean clinical and radiological follow-up was 13.4 years (12 to 15). A Profile hydroxyapatite-coated anatomical femoral component was used in each case. Six patients had a late dislocation, for whom the polyethylene liner was exchanged. Each acetabular component was well fixed and all femoral components showed signs of bone ingrowth. The mean rate of femoral head penetration was 0.10 mm/year (0.021 to 0.481). The probability of not developing femoral cortical hypertrophy and proximal osteopenia by 12 years was 80.2% (95% confidence interval, 72.7 to 87.6) and 77.5% (95% confidence interval, 69.7 to 85.2), respectively. Despite these good clinical results, further follow-up is needed to determine whether these prostheses will loosen with time.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1422 - 1427
1 Nov 2008
Utting MR Raghuvanshi M Amirfeyz R Blom AW Learmonth ID Bannister GC

We have reviewed 70 Harris-Galante uncemented acetabular components implanted as hybrid hip replacements with cemented stems between 1991 and 1995 in 53 patients whose mean age was 40 years (19 to 49). The mean follow-up was for 13.6 years (12 to 16) with no loss to follow-up. We assessed the patients both clinically and radiologically.

The mean Oxford hip score was 20 (12 to 46) and the mean Harris hip score 81 (37 to 100) at the final review. Radiologically, 27 hips (39%) had femoral osteolysis, 13 (19%) acetabular osteolysis, and 31 (44%) radiolucent lines around the acetabular component. Kaplan-Meier survival curves were constructed for the outcomes of revision of the acetabular component, revision of the component and polyethylene liner, and impending revision for progressive osteolysis.

The cumulative survival for revision of the acetabular component was 94% (95% confidence interval 88.4 to 99.7), for the component and liner 84% (95% confidence interval 74.5 to 93.5) and for impending revision 55.3% (95% confidence interval 40.6 to 70) at 16 years.

Uncemented acetabular components with polyethylene liners undergo silent lysis and merit regular long-term radiological review.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 31 - 36
1 Jan 2009
de Jong PT de Man FHR Haverkamp D Marti RK

We report the long-term outcome of a modified second-generation cementing technique for fixation of the acetabular component of total hip replacement. An earlier report has shown the superiority of this technique assessed by improved survival compared with first-generation cementing. The acetabular preparation involved reaming only to the subchondral plate, followed by impaction of the bone in the anchorage holes.

Between 1978 and 1993, 287 total hip replacements were undertaken in 244 patients with a mean age of 65.3 years (21 to 90) using a hemispherical Weber acetabular component with this modified technique for cementing and a cemented femoral component.

The survival with acetabular revision for aseptic loosening as the endpoint was 99.1% (95% confidence interval 97.9 to 100 after ten years and 85.5% (95% confidence interval 74.7 to 96.2) at 20 years. Apart from contributing to a long-lasting fixation of the component, this technique also preserved bone, facilitating revision surgery when necessary.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 296 - 300
1 Mar 2007
van Haaren EH Heyligers IC Alexander FGM Wuisman PIJM

We reviewed the results of 71 revisions of the acetabular component in total hip replacement, using impaction of bone allograft. The mean follow-up was 7.2 years (1.6 to 9.7). All patients were assessed according to the American Academy of Orthopedic Surgeons (AAOS) classification of bone loss, the amount of bone graft required, thickness of the graft layer, signs of graft incorporation and use of augmentation.

A total of 20 acetabular components required re-revision for aseptic loosening, giving an overall survival of 72% (95% CI, 54.4 to 80.5). Of these failures, 14 (70%) had an AAOS type III or IV bone defect. In the failed group, poor radiological and histological graft incorporation was seen.

These results suggest that impaction allografting in acetabular revision with severe bone defects may have poorer results than have previously been reported.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 786 - 789
1 Jun 2005
Grobler GP Learmonth ID Bernstein BP Dower BJ

We retrospectively reviewed, ten years after surgery, 100 consecutive total hip replacements in which the Duraloc 300 cup had been used. Post-operative radiographs were analysed for placement of the cup and interface gaps and follow-up radiographs for lucent lines, osteolysis, wear and migration.

All the components were found to be stable with no evidence of loosening. The mean rate of wear was 0.12 mm/year. Three hips developed acetabular osteolysis at the level of the apex hole. Two have successfully undergone bone grafting without removal of the implants and one patient is awaiting surgery. The Duraloc 300 cup has a survival of 100% at ten years with no aseptic loosening and a low incidence of pelvic osteolysis.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1431 - 1438
1 Nov 2007
Hing CB Back DL Bailey M Young DA Dalziel RE Shimmin AJ

We report an independent prospective review of the first 230 Birmingham hip resurfacings in 212 patients at a mean follow-up of five years (4 to 6).

Two patients, one with a loose acetabular component and the other with suspected avascular necrosis of the femoral head, underwent revision. There were two deaths from unrelated causes and one patient was lost to follow-up. The survivorship with the worst-case scenario was 97.8% (95% confidence interval 95.8 to 99.5). The mean Harris hip score improved significantly (paired t-test, p < 0.05) from 62.54 (8 to 92) pre-operatively to 97.7 (61 to 100) at a mean of three years (2.1 to 4.3), then deteriorated slightly to a mean of 95.2 (47 to 100) at a mean of five years. The mean flexion improved from 91.5° (25° to 140°) to 110.4° (80° to 145°) at a mean of three years with no further improvement at five years (111.2°; 70° to 160°).

On radiological review at five years, one patient had a progressive lucent line around the acetabular component and six had progressive lucent lines around the femoral component. A total of 18 femoral components (8%) had migrated into varus and those with lucent lines present migrated a mean of 3.8° (1.02° to 6.54°) more than the rest. Superolateral notching of the femoral neck and reactive sclerosis at the tip of the peg of the femoral component were associated with the presence of lucent lines (chi-squared test, p < 0.05), but not with migration of the femoral component, and are of unknown significance.

Our results with the Birmingham hip resurfacing continue to be satisfactory at a mean follow-up of five years.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 879 - 886
1 Jul 2007
Howie DW Wimhurst JA McGee MA Carbone TA Badaruddin BS

We investigated 219 revisions of total hip replacement (THR) in 211 patients using a collarless double-taper cemented femoral component. The mean age of the patients was 72 years (30 to 90). The 137 long and 82 standard length stems were analysed separately. The mean follow-up was six years (2 to 18), and no patient was lost to follow-up.

Survival of the long stems to re-revision for aseptic loosening at nine years was 98% (95% confidence interval (CI) 94 to 100), and for the standard stems was 93% (95% CI 85 to 100). At five years, one long stem was definitely loose radiologically and one standard stem was probably loose. Pre-operative femoral bone deficiency did not influence the results for the long stems, and corrective femoral osteotomy was avoided, as were significant subsidence, major stress shielding and persistent thigh pain.

Because of these reliable results, cemented long collarless double-taper femoral components are recommended for routine revision THR in older patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1567 - 1573
1 Dec 2006
Kovac S Trebse R Milosev I Pavlovcic V Pisot V

We present a retrospective series of 170 cemented titanium straight-stem femoral components combined with two types of femoral head: cobalt-chromium (CoCr) alloy (114 heads) and alumina ceramic (50 heads). Of the study group, 55 patients (55 stems) had died and six (six stems) were lost to follow-up. At a mean of 13.1 years (3 to 15.3) 26 stems had been revised for aseptic loosening. The mean follow-up time for stable stems was 15.1 years (12.1 to 16.6).

Survival of the stem at 15 years was 75.4% (95% confidence interval (CI) 67.3 to 83.5) with aseptic failure (including radiological failure) as the end-point, irrespective of the nature of the head and the quality of the cement mantle. Survival of the stem at 15 years was 79.1% (95% CI 69.8 to 88.4) and 67.1% (95% CI 51.3 to 82.9) with the CoCr alloy and ceramic heads, respectively. The quality of the cement mantle was graded as a function of stem coverage: stems with complete tip coverage (type 1) had an 84.9% (95% CI 77.6 to 92.2) survival at 15 years, compared with those with a poor tip coverage (type 2) which had a survival of only 22.4% (95% CI 2.4 to 42.4). The poor quality of the cement mantle and the implantation of an alumina head substantially lowered the survival of the stem.

In our opinion, further use of the cemented titanium alloy straight-stem femoral components used in our series is undesirable.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 2 | Pages 173 - 178
1 Feb 2006
de Jong PT Haverkamp D van der Vis HM Marti RK

We evaluated the long-term results of 116 total hip replacements with a superolateral shelfplasty in 102 patients with osteoarthritis secondary to developmental dysplasia of the hip. After a mean follow-up of 19.5 years (11.5 to 26.0), 14 acetabular components (12%) had been revised. The cumulative survival at 20 years was 78%, with revision for loosening of the acetabular component as the end-point. All grafts were well integrated and showed remodelling. In six grafts some resorption had occurred under the heads of the screws where the graft was not supporting the socket. Apart from these 14 revisions, seven acetabular components had possible radiological signs of loosening at a mean follow-up of 14.5 years, one had signs of probable loosening, and five had signs of definite loosening. These results indicate that this technique of bone grafting for acetabular reconstruction in hip dysplasia is a durable solution for cemented acetabular components.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1192 - 1196
1 Sep 2005
Argenson J Ryembault E Flecher X Brassart N Parratte S Aubaniac J

Using radiography and computer tomography (CT) we studied the morphology of 83 hips in 69 Caucasian adults with osteoarthritis secondary to developmental dysplasia of the hip (DDH). A previously published series of 310 hips with primary osteoarthritis was used as a control group. According to the Crowe classification, 33 of the dysplastic hips were graded as class I, 27 as class II and 23 as class III or class IV.

The intramedullary femoral canal had reduced mediolateral and anteroposterior dimensions in all groups compared with the control group. Only in Crowe class II hips was the femoral neck-shaft angle increased. The proximal femur had more anteversion in all the developmental dysplasia of the hip groups, ranging from 2° to 80°. Templated measurement of acetabular dimensions for plain radiography closely matched measurements taken by CT.

The results of our study confirm the observations previously confined to the Japanese population.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 997 - 1002
1 Aug 2006
Vendittoli P Lavigne M Girard J Roy AG

We have undertaken a prospective, randomised study to compare conservation of acetabular bone after total hip replacement and resurfacing arthroplasty of the hip. We randomly assigned 210 hips to one of the two treatment groups. Uncemented, press-fit acetabular components were used for both.

No significant difference was found in the mean diameter of acetabular implant inserted in the groups (54.74 mm for total hip replacement and 54.90 mm for resurfacing arthroplasty). In seven resurfacing procedures (6.8%), the surgeon used a larger size of component in order to match the corresponding diameter of the femoral component.

With resurfacing arthroplasty, conservation of bone is clearly advantageous on the femoral side. Our study has shown that, with a specific design of acetabular implant and by following a careful surgical technique, removal of bone on the acetabular side is comparable with that of total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 776 - 780
1 Jun 2005
Kim Y Kim S Kim S Park B Kim P Ihn J

We performed 114 consecutive primary total hip arthroplasties with a cementless expansion acetabular component in 101 patients for advanced osteonecrosis of the femoral head. The mean age of the patients at surgery was 51 years (36 to 62) and the mean length of follow-up was 110 months (84 to 129).

The mean pre-operative Harris hip score of 47 points improved to 93 points at final follow-up. The polyethylene liner was exchanged in two hips during this period and one broken acetabular component was revised. The mean linear wear rate of polyethylene was 0.07 mm/year and peri-acetabular osteolysis was seen in two hips (1.9%). Kaplan-Meier analysis indicated that the survival of the acetabular component without revision was 97.8% (95% confidence interval 0.956 to 1.000) at ten years.

Our study has shown that the results of THA with a cementless expansion acetabular component and an alumina-polyethylene bearing surface are good.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1068 - 1072
1 Aug 2005
Morag G Zalzal P Liberman B Safir O Flint M Gross AE

Our aim was to determine if the height of the cup, lateralisation or the abduction angle correlated with functional outcome or survivorship in revision total hip replacement in patients with a previous diagnosis of developmental dysplasia of the hip. A retrospective investigation of 51 patients (63 hips) who had undergone revision total hip replacement was performed. The mean duration of follow-up was 119 months. Forty-one patients (52 hips) were available for both determination of functional outcome and survivorship analysis. Ten patients (11 hips) were only available for survivorship analysis.

The height of the cup was found to have a statistically significant correlation with functional outcome and a high hip centre correlated with a worse outcome score. Patients with a hip centre of less than 3.5 cm above the anatomical level had a statistically better survivorship of the cup than those with centres higher than this. Restoration of the height of the centre of the hip to as near the anatomical position as possible improved functional outcome and survivorship of the cup.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 762 - 769
1 Jun 2005
Biedermann R Tonin A Krismer M Rachbauer F Eibl G Stöckl B

Malposition of the acetabular component is a risk factor for post-operative dislocation after total hip replacement (THR). We have investigated the influence of the orientation of the acetabular component on the probability of dislocation. Radiological anteversion and abduction of the component of 127 hips which dislocated post-operatively were measured by Einzel-Bild-Röentgen-Analysis and compared with those in a control group of 342 patients.

In the control group, the mean value of anteversion was 15° and of abduction 44°. Patients with anterior dislocation after primary THR showed significant differences in the mean angle of anteversion (17°), and abduction (48°) as did patients with posterior dislocation (anteversion 11°, abduction 42°). After revision patients with posterior dislocation showed significant differences in anteversion (12°) and abduction (40°).

Our results demonstrate the importance of accurate positioning of the acetabular component in order to reduce the frequency of subsequent dislocations. Radiological anteversion of 15° and abduction of 45° are the lowest at-risk values for dislocation.