Aims. The aim of this study was to compare the
Aims. Early implant migration measured with radiostereometric analysis (RSA) has been proposed as a useful predictor of long-term fixation of tibial components in total knee arthroplasty. Evaluation of actual long-term fixation is of interest for cemented components, as well as for cementless fixation, which may offer long-term advantages once osseointegration has occurred. The objective of this study was to compare the long-term
Aims. The objective of this study was to compare the two-year
Aims. The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs). Methods. A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with radiostereometric analysis (RSA) at three, six, 12, 24, and 60 months postoperatively. PROMS were collected preoperatively and at all follow-up timepoints. Results. There was no clinically relevant difference in terms of tibial component migration, insert wear, and PROMs between the HXLPE and PE groups. The mean difference in tibial component migration (maximal total point
Aims. Thresholds of acceptable early
Aims. Medial pivot (MP) total knee arthroplasties (TKAs) were designed to mimic native knee kinematics with their deep medial congruent fitting of the tibia to the femur almost like a ball-on-socket, and a flat lateral part. GMK Sphere is a novel MP implant. Our primary aim was to study the
Aims. The primary aim of this study was to compare the
Aims. The primary objective of this study was to compare
Aims. Proliferation,
Aims. Although bone cement is the primary mode of fixation in total knee arthroplasty (TKA), cementless fixation is gaining interest as it has the potential of achieving lasting biological fixation. By 3D printing an implant, highly porous structures can be manufactured, promoting osseointegration into the implant to prevent aseptic loosening. This study compares the
Aims. Hydroxyapatite coatings for uncemented fixation in total knee
arthroplasty can theoretically provide a long-lasting biological
interface with the host bone. The objective of this study was to
test this hypothesis with propriety hydroxyapatite, peri-apatite,
coated tibial components using component migration measured with radiostereometric
analysis over two years as an indicator of long-term fixation. Patients and Methods. A total of 29 patients at two centres received uncemented PA-coated
tibial components and were followed for two years with radiostereometric
analysis exams to quantify the
Aims. Patient-specific instrumentation of total knee arthroplasty (TKA) is a technique permitting the targeting of individual kinematic alignment, but deviation from a neutral mechanical axis may have implications on implant fixation and therefore survivorship. The primary objective of this randomized controlled study was to compare the fixation of tibial components implanted with patient-specific instrumentation targeting kinematic alignment (KA+PSI) versus components placed using computer-assisted surgery targeting neutral mechanical alignment (MA+CAS). Tibial component migration measured by radiostereometric analysis was the primary outcome measure (compared longitudinally between groups and to published acceptable thresholds). Secondary outcome measures were inducible displacement after one year and patient-reported outcome measures (PROMS) over two years. The secondary objective was to assess the relationship between alignment and both tibial component migration and inducible displacement. Patients and Methods. A total of 47 patients due to undergo TKA were randomized to KA+PSI (n = 24) or MA+CAS (n = 23). In the KA+PSI group, there were 16 female and eight male patients with a mean age of 64 years (. sd. 8). In the MA+CAS group, there were 17 female and six male patients with a mean age of 63 years (. sd. 7). Surgery was performed using cemented, cruciate-retaining Triathlon total knees with patellar resurfacing, and patients were followed up for two years. The effect of alignment on tibial component migration and inducible displacement was analyzed irrespective of study group. Results. There was no difference over two years in longitudinal
Aims. The purpose of the present study was to examine the long-term
fixation of a cemented fixed-bearing polished titanium tibial baseplate
(Genesis ll). . Patients and Methods. Patients enrolled in a previous two-year prospective trial (n
= 35) were recalled at ten years. Available patients (n = 15) underwent
radiostereometric analysis (RSA) imaging in a supine position using
a conventional RSA protocol.
Accurate quantitative measurements of micromovement immediately after operation would be a reliable indicator of the stability of an individual component. We have therefore developed a system for measuring micromovement of the tibial component using three non-contact displacement transducers attached to the tibial cortex during total knee arthroplasty (TKA). Using this system we measured the initial stability in 31 uncemented TKAs. All the tibial components were fixed by a stem and four screws. The initial stability was defined as the amount of displacement when a load of 20 kg was applied. The mean subsidence was 60.7 μm and the mean lift-off was 103.3 μm. We also studied the
Aims. Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes. Methods. A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively. Results. At five-year follow-up, RSA data from 61 patients were available and the mean maximum total point motion (MTPM) of the femoral components were: ATTUNE: 0.96 mm (95% confidence interval (CI) 0.79 to 1.14) and PFC-Sigma 1.37 mm (95% CI 1.18 to 1.59) (p < 0.001). The PFC-Sigma femoral component migrated more in the first postoperative year, but stabilized thereafter. MPTM of the tibial components were comparable at five-year follow-up: ATTUNE 1.12 mm (95% CI 0.95 to 1.31) and PFC-Sigma 1.25 mm (95% CI 1.07 to 1.44) (p = 0.438). RLL at the medial tibial implant-cement interface remained more prevalent for the ATTUNE at five-year follow-up compared to the PFC-Sigma (20% vs 3%). RLL did not progress over time, and varied between patients at different timepoints for both TKA systems. Clinical outcomes and PROMs improved compared with preoperative scores, and were not different between groups. Conclusion. MTPM
Aims. The aim of this study was to compare a bicruciate-retaining (BCR) total knee arthroplasty (TKA) with a posterior cruciate-retaining (CR) TKA design in terms of kinematics, measured using fluoroscopy and stability as micromotion using radiostereometric analysis (RSA). Methods. A total of 40 patients with end-stage osteoarthritis were included in this randomized controlled trial. All patients performed a step-up and lunge task in front of a monoplane fluoroscope one year postoperatively. Femorotibial contact point (CP) locations were determined at every flexion angle and compared between the groups. RSA images were taken at baseline, six weeks, three, six, 12, and 24 months postoperatively. Clinical and functional outcomes were compared postoperatively for two years. Results. The BCR-TKA demonstrated a kinematic pattern comparable to the natural knee’s screw-home mechanism in the step-up task. In the lunge task, the medial CP of the BCR-TKA was more anterior in the early flexion phase, while laterally the CP was more posterior during the entire movement cycle. The BCR-TKA group showed higher tibial
Aims. Micromotion of the polyethylene (PE) inlay may contribute to backside PE wear in addition to articulate wear of total knee arthroplasty (TKA). Using radiostereometric analysis (RSA) with tantalum beads in the PE inlay, we evaluated PE micromotion and its relationship to PE wear. Methods. A total of 23 patients with a mean age of 83 years (77 to 91), were available from a RSA study on cemented TKA with Maxim tibial components (Zimmer Biomet). PE inlay
Aims. A novel enhanced cement fixation (EF) tibial implant with deeper cement pockets and a more roughened bonding surface was released to market for an existing total knee arthroplasty (TKA) system.This randomized controlled trial assessed fixation of the both the EF (ATTUNE S+) and standard (Std; ATTUNE S) using radiostereometric analysis. Methods. Overall, 50 subjects were randomized (21 EF-TKA and 23 Std-TKA in the final analysis), and had follow-up visits at six weeks, and six, 12, and 24 months to assess
Aims. Meniscal injuries are common and often induce knee pain requiring surgical intervention. To develop effective strategies for meniscus regeneration, we hypothesized that a minced meniscus embedded in an atelocollagen gel, a firm gel-like material, may enhance meniscus regeneration through cell
Aims. Our intention was to investigate if the highly porous biological fixation surfaces of a new 3D-printed total knee arthroplasty (TKA) achieved adequate fixation of the tibial and patellar components to the underlying bone. Patients and Methods. A total of 29 patients undergoing primary TKA consented to participate in this prospective cohort study. All patients received a highly porous tibial baseplate and metal-backed patella. Patient-reported outcomes measures were recorded and implant migration was assessed using radiostereometric analysis. Results. Patient function significantly improved by three months postoperatively (p < 0.001). Mean difference in maximum total point motion between 12 and 24 months was 0.021 mm (-0.265 to 0.572) for the tibial implant and 0.089 mm (-0.337 to 0.758) for the patellar implant. The rate of tibial and patellar
Aims. The optimal method of tibial component fixation remains uncertain
in total knee arthroplasty (TKA). Hydroxyapatite coatings have been
applied to improve bone ingrowth in uncemented designs, but may
only coat the directly accessible surface. As peri-apatite (PA)
is solution deposited, this may increase the coverage of the implant
surface and thereby fixation. We assessed the tibial component fixation
of uncemented PA-coated TKAs versus cemented TKAs. Patients and Methods. Patients were randomised to PA-coated or cemented TKAs. In 60
patients (30 in each group), radiostereometric analysis of tibial
component
Aims. For many designs of total knee arthroplasty (TKA) it remains unclear whether cemented or uncemented fixation provides optimal long-term survival. The main limitation in most studies is a retrospective or non-comparative study design. The same is true for comparative trials looking only at the survival rate as extensive sample sizes are needed to detect true differences in fixation and durability. Studies using radiostereometric analysis (RSA) techniques have shown to be highly predictive in detecting late occurring aseptic loosening at an early stage. To investigate the difference in predicted long-term survival between cemented, uncemented, and hybrid fixation of TKA, we performed a randomized controlled trial using RSA. Methods. A total of 105 patients were randomized into three groups (cemented, uncemented, and hybrid fixation of the ACS Mobile Bearing (ACS MB) knee system, implantcast). RSA examinations were performed on the first day after surgery and at scheduled follow-up visits at three months, six months, one year, and two years postoperatively. Patient-reported outcome measures (PROMs) were obtained preoperatively and after two years follow-up. Patients and follow-up investigators were blinded for the result of randomization. Results. RSA secondary stabilization did not show a significant difference between the three types of fixation. A maximum total point motion of less than 0.2 mm in the second postoperative year was shown in each group, which suggests stabilization of the implant. At 24 months after surgery, PROMs significantly improved compared to baseline in all treatment groups. No significant difference was observed between the three groups. Conclusion. Secondary stabilization measurements in this study demonstrated no significant difference between the groups. In all groups
Aims. The purpose of the present study was to compare patient-specific instrumentation (PSI) and conventional surgical instrumentation (CSI) for total knee arthroplasty (TKA) in terms of early implant migration, alignment, surgical resources, patient outcomes, and costs. . Patients and Methods. The study was a prospective, randomized controlled trial of 50 patients undergoing TKA. There were 25 patients in each of the PSI and CSI groups. There were 12 male patients in the PSI group and seven male patients in the CSI group. The patients had a mean age of 69.0 years (. sd. 8.4) in the PSI group and 69.4 years (. sd. 8.4) in the CSI group. All patients received the same TKA implant. Intraoperative surgical resources and any surgical waste generated were recorded. Patients underwent radiostereometric analysis (RSA) studies to measure femoral and tibial component migration over two years. Outcome measures were recorded pre- and postoperatively. Overall costs were calculated for each group. Results. There were no differences (p > 0.05) in any measurement of
Tibial tubercle osteotomy (TTO) facilitates surgical exposure and protects the extensor mechanism during revision total knee arthroplasty (rTKA). The purpose of this study was to determine the rates of bony union, complications, and reoperations following TTO during rTKA, to assess the functional outcomes of rTKA with TTO at two years’ minimum follow-up, and to identify the risk factors of failure. Between January 2010 and September 2020, 695 rTKAs were performed and data were entered into a prospective database. Inclusion criteria were rTKAs with concomitant TTO, without extensor mechanism allograft, and a minimum of two years’ follow-up. A total of 135 rTKAs were included, with a mean age of 65 years (SD 9.0) and a mean BMI of 29.8 kg/m2 (SD 5.7). The most frequent indications for revision were infection (50%; 68/135), aseptic loosening (25%; 34/135), and stiffness (13%; 18/135). Patients had standardized follow-up at six weeks, three months, six months, and annually thereafter. Complications and revisions were evaluated at the last follow-up. Functional outcomes were assessed using the Knee Society Score (KSS) and range of motion.Aims
Methods
The aim of this study was to assess factors associated with the estimated lifetime risk of revision surgery after primary knee arthroplasty (KA). All patients from the Scottish Arthroplasty Project dataset undergoing primary KA during the period 1 January 1998 to 31 December 2019 were included. The cumulative incidence function for revision and death was calculated up to 20 years. Adjusted analyses used cause-specific Cox regression modelling to determine the influence of patient factors. The lifetime risk was calculated as a percentage for patients aged between 45 and 99 years using multiple-decrement life table methodology.Aims
Methods
The most common reasons for revision of unicompartmental
knee arthroplasty (UKA) are loosening and pain. Cementless components
may reduce the revision rate. The aim of this study was to compare
the fixation and clinical outcome of cementless and cemented Oxford
UKAs. . A total of 43 patients were randomised to receive either a cemented
or a cementless Oxford UKA and were followed for two years with
radiostereometric analysis (RSA), radiographs aligned with the bone–implant
interfaces and clinical scores. The femoral components migrated significantly during the first
year (mean 0.2 mm) but not during the second. There was no significant
difference in the extent of
This study aimed to explore the biological and clinical importance of dysregulated key genes in osteoarthritis (OA) patients at the cartilage level to find potential biomarkers and targets for diagnosing and treating OA. Six sets of gene expression profiles were obtained from the Gene Expression Omnibus database. Differential expression analysis, weighted gene coexpression network analysis (WGCNA), and multiple machine-learning algorithms were used to screen crucial genes in osteoarthritic cartilage, and genome enrichment and functional annotation analyses were used to decipher the related categories of gene function. Single-sample gene set enrichment analysis was performed to analyze immune cell infiltration. Correlation analysis was used to explore the relationship among the hub genes and immune cells, as well as markers related to articular cartilage degradation and bone mineralization.Aims
Methods
Both the femoral and tibial component are usually cemented at revision total knee arthroplasty (rTKA), while stems can be added with either cemented or press-fit (hybrid) fixation. The aim of this study was to compare the long-term stability of rTKA with cemented and press-fitted stems, using radiostereometric analysis (RSA). This is a follow-up of a randomized controlled trial, initially involving 32 patients, of whom 19 (nine cemented, ten hybrid) were available for follow-up ten years postoperatively, when further RSA measurements were made. Micromotion of the femoral and tibial components was assessed using model-based RSA software (RSAcore). The clinical outcome was evaluated using the Knee Society Score (KSS), the Knee injury and Osteoarthritis Outcome Score (KOOS), and visual analogue scale (pain and satisfaction).Aims
Methods
The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design. This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.Aims
Methods
Implantation of ultra-purified alginate (UPAL) gel is safe and effective in animal osteochondral defect models. This study aimed to examine the applicability of UPAL gel implantation to acellular therapy in humans with cartilage injury. A total of 12 patients (12 knees) with symptomatic, post-traumatic, full-thickness cartilage lesions (1.0 to 4.0 cm2) were included in this study. UPAL gel was implanted into chondral defects after performing bone marrow stimulation technique, and assessed for up to three years postoperatively. The primary outcomes were the feasibility and safety of the procedure. The secondary outcomes were self-assessed clinical scores, arthroscopic scores, tissue biopsies, and MRI-based estimations.Aims
Methods
The aim of this study was to investigate the distribution of phenotypes in Asian patients with end-stage osteoarthritis (OA) and assess whether the phenotype affected the clinical outcome and survival of mechanically aligned total knee arthroplasty (TKA). We also compared the survival of the group in which the phenotype unintentionally remained unchanged with those in which it was corrected to neutral. The study involved 945 TKAs, which were performed in 641 patients with primary OA, between January 2000 and January 2009. These were classified into 12 phenotypes based on the combined assessment of four categories of the arithmetic hip-knee-ankle angle and three categories of actual joint line obliquity. The rates of survival were analyzed using Kaplan-Meier methods and the log-rank test. The Hospital for Special Surgery score and survival of each phenotype were compared with those of the reference phenotype with neutral alignment and a parallel joint line. We also compared long-term survival between the unchanged phenotype group and the corrected to neutral alignment-parallel joint line group in patients with Type IV-b (mild to moderate varus alignment-parallel joint line) phenotype.Aims
Methods
This multicentre retrospective observational study’s aims were to investigate whether there are differences in the occurrence of radiolucent lines (RLLs) following total knee arthroplasty (TKA) between the conventional Attune baseplate and its successor, the novel Attune S+, independent from other potentially influencing factors; and whether tibial baseplate design and presence of RLLs are associated with differing risk of revision. A total of 780 patients (39% male; median age 70.7 years (IQR 62.0 to 77.2)) underwent cemented TKA using the Attune Knee System) at five centres, and with the latest radiograph available for the evaluation of RLL at between six and 36 months from surgery. Univariate and multivariate logistic regression models were performed to assess associations between patient and implant-associated factors on the presence of tibial and femoral RLLs. Differences in revision risk depending on RLLs and tibial baseplate design were investigated with the log-rank test.Aims
Methods
Patient dissatisfaction is not uncommon following primary total knee arthroplasty. One proposed method to alleviate this is by improving knee kinematics. Therefore, we aimed to answer the following research question: are there significant differences in knee kinematics based on the design of the tibial insert (cruciate-retaining (CR), ultra-congruent (UC), or medial congruent (MC))? Overall, 15 cadaveric knee joints were examined with a CR implant with three different tibial inserts (CR, UC, and MC) using an established knee joint simulator. The effects on coronal alignment, medial and lateral femoral roll back, femorotibial rotation, bony rotations (femur, tibia, and patella), and patellofemoral length ratios were determined.Aims
Methods
Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening. A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.Aims
Methods
The role of modular tibial implants in total knee replacement is not fully defined. We performed a prospective randomised controlled clinical trial using radiostereophotogrammetric analysis to compare the performance of an all-polyethylene tibia with a metal-backed cruciate-retaining condylar design, PFC-∑ total knee replacement for up to 24 months. There were 51 patients who were randomised into two treatment groups. There were 10 subsequent withdrawals, leaving 21 all-polyethylene and 20 metal-backed tibial implants. No patient was lost to follow-up. There were no significant demographic differences between the groups. At two years one metal-backed implant showed
Aims . To assess the effect of high tibial and distal femoral osteotomies
(HTO and DFO) on the pressure characteristics of the ankle joint. Materials and Methods. Varus and valgus malalignment of the knee was simulated in human
cadaver full-length legs. Testing included four measurements: baseline
malalignment, 5° and 10° re-aligning osteotomy, and control baseline
malalignment. For HTO, testing was rerun with the subtalar joint
fixed. In order to represent half body weight, a 300 N force was applied
onto the femoral head. Intra-articular sensors captured ankle pressure. Results. In the absence of restriction of subtalar movement, insignificant
migration of the centre of force and changes of maximal pressure
were seen at the ankle joint. With restricted subtalar motion, more
significant lateralisation of the centre of force were seen with
the subtalar joint in varus than in valgus position. Changes in
maximum pressure were again not significant. . Conclusion. The re-alignment of coronal plane knee deformities by HTO and
DFO altered ankle pressure characteristics. When the subtalar joint
was fixed in the varus position,
One of the main causes of tibial revision surgery for total knee arthroplasty is aseptic loosening. Therefore, stable fixation between the tibial component and the cement, and between the tibial component and the bone, is essential. A factor that could influence the implant stability is the implant design, with its different variations. In an existing implant system, the tibial component was modified by adding cement pockets. The aim of this experimental in vitro study was to investigate whether additional cement pockets on the underside of the tibial component could improve implant stability. The relative motion between implant and bone, the maximum pull-out force, the tibial cement mantle, and a possible path from the bone marrow to the metal-cement interface were determined. A tibial component with (group S: Attune S+) and without (group A: Attune) additional cement pockets was implanted in 15 fresh-frozen human leg pairs. The relative motion was determined under dynamic loading (extension-flexion 20° to 50°, load-level 1,200 to 2,100 N) with subsequent determination of the maximum pull-out force. In addition, the cement mantle was analyzed radiologically for possible defects, the tibia base cement adhesion, and preoperative bone mineral density (BMD).Aims
Methods
Platelet-rich plasma (PRP) intra-articular injections may provide a simple and minimally invasive treatment for early-stage knee osteoarthritis (OA). This has led to an increase in its adoption as a treatment for knee OA, although there is uncertainty about its efficacy and benefit. We hypothesized that patients with early-stage symptomatic knee OA who receive multiple PRP injections will have better clinical outcomes than those receiving single PRP or placebo injections. A double-blinded, randomized placebo-controlled trial was performed with three groups receiving either placebo injections (Normal Saline), one PRP injection followed by two placebo injections, or three PRP injections. Each injection was given one week apart. Outcomes were prospectively collected prior to intervention and then at six weeks, three months, six months, and 12 months post-intervention. Primary outcome measures were Knee Injury and Osteoarthritis Outcome Score (KOOS) and EuroQol five-dimension five-level index (EQ-5D-5L). Secondary outcomes included visual analogue scale for pain and patient subjective assessment of the injections.Aims
Methods
The kinematic alignment (KA) approach to total knee arthroplasty (TKA) has recently increased in popularity. Accordingly, a number of derivatives have arisen and have caused confusion. Clarification is therefore needed for a better understanding of KA-TKA. Calipered (or true, pure) KA is performed by cutting the bone parallel to the articular surface, compensating for cartilage wear. In soft-tissue respecting KA
This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant. A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life.Aims
Methods
Objectives. The objective of this study was to compare the early migration
characteristics and functional outcome of the Triathlon cemented
knee prosthesis with its predecessor, the Duracon cemented knee
prosthesis (both Stryker). Methods. A total 60 patients were prospectively randomised and tibial
component
To assess
Tibial cones are often utilized in revision total knee arthroplasty (TKA) with metaphyseal defects. Because there are few studies evaluating mid-term outcomes with a sufficient cohort, the purpose of this study was to evaluate tibial cone survival and complications in revision TKAs with tibial cones at minimum follow-up of five years. A retrospective review was completed from September 2006 to March 2015, evaluating 67 revision TKAs (64 patients) that received one specific porous tibial cone during revision TKA. The final cohort was composed of 62 knees (59 patients) with five years of clinical follow-up or reoperation. The mean clinical follow-up of the TKAs with minimum five-year clinical follow-up was 7.6 years (5.0 to 13.3). Survivorship analysis was performed with the endpoints of tibial cone revision for aseptic loosening, tibial cone revision for any reason, and reoperation. We also evaluated periprosthetic joint infection (PJI), risk factors for failure, and performed a radiological review.Aims
Methods
The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.Aims
Methods
This single-blinded randomised controlled trial
investigated whether one design of mobile-bearing (MB) total knee replacement
(TKR) has any advantage over a fixed-bearing (FB) design on long-term
fixation as measured by radiostereometry. The amount of wear underneath
the mobile bearing was also evaluated. A series of 42 knees was randomised
to MB or FB tibial components with appropriate polyethylene inserts
and followed for between ten and 12 years, or until the death of
the patient. The polyethylene in the MB group was superior in that
it was gamma-irradiated in inert gas and was calcium-stearate free;
the polyethylene in the FB group was gamma-irradiated in air and
contained calcium stearate. In theory this should be advantageous
to the wear rate of the MB group. At final follow-up the overall
mean
Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes. Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty. This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice. Cite this article:
Total knee arthroplasty (TKA) is known to lead
to a reduction in periprosthetic bone mineral density (BMD). In theory,
this may lead to
To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period. This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending.Aims
Methods
We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component. All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting
Objectives. Static radiostereometric analysis (RSA) using implanted markers is considered the most accurate system for the evaluation of prosthesis
Clinical experience of impaction bone grafting for revision knee arthroplasty is limited, with initial stability of the tibial tray emerging as a major concern. The length of the stem and its diameter have been altered to improve stability. Our aim was to investigate the effect of the type of stem, support of the rim and graft impaction on early stability of the tray. We developed a system for impaction grafting of trays which we used with morsellised bone in artificial tibiae. Trays with short, long thick or long thin stems were implanted, with or without support of the rim. They were cyclically loaded while measuring relative movement. Long-stemmed trays migrated 4.5 times less than short-stemmed trays, regardless of diameter. Those with support migrated 2.8 times less than those without. The
Uncemented mobile bearing designs in medial unicompartmental knee arthroplasty (UKA) have seen an increase over the last decade. However, there are a lack of large-scale studies comparing survivorship of these specific designs to commonly used cemented mobile and fixed bearing designs. The aim of this study was to evaluate the survivorship of these designs. A total of 21,610 medial UKAs from 2007 to 2018 were selected from the Dutch Arthroplasty Register. Multivariate Cox regression analyses were used to compare uncemented mobile bearings with cemented mobile and fixed bearings. Adjustments were made for patient and surgical factors, with their interactions being considered. Reasons and type of revision in the first two years after surgery were assessed.Aims
Methods
Stress shielding resulting in diminished bone
density following total knee replacement (TKR) may increase the
risk of
We describe the survivorship of the Medial Rotation total knee replacement (TKR) at ten years in 228 cemented primary replacements implanted between October 1994 and October 2006, with their clinical and radiological outcome. This implant has a highly congruent medial compartment, with the femoral component represented by a portion of a sphere which articulates with a matched concave surface on the medial side of the tibial insert. There were 78 men (17 bilateral TKRs) and 111 women (22 bilateral TKRs) with a mean age of 67.9 years (28 to 90). All the patients were assessed clinically and radiologically using the American Knee Society scoring systems. The mean follow-up was for six years (1 to 13) with only two patients lost to follow-up and 34 dying during the period of study, one of whom had required revision for infection. There were 11 revisions performed in total, three for aseptic loosening, six for infection, one for a periprosthetic fracture and one for a painful but well-fixed replacement performed at another centre. With revision for any cause as the endpoint, the survival at ten years was 94.5% (95% CI 85.1 to 100), and with aseptic loosening as the endpoint 98.4% (95% CI 93 to 100). The mean American Knee Society score improved from 47.6 (0 to 88) to 72.2 (26 to 100) and for function from 45.1 (0 to 100) to 93.1 (45 to 100). Radiological review failed to detect
We have examined the results obtained with 72 NexGen legacy posterior stabilised-flex fixed total knee replacements in 47 patients implanted by a single surgeon between March 2003 and September 2004. Aseptic loosening of the femoral component was found in 27 (38%) of the replacements at a mean follow-up of 32 months (30 to 48) and 15 knees (21%) required revision at a mean of 23 months (11 to 45). We compared the radiologically-loose and revised knees with those which had remained well-fixed to identify the factors which had contributed to this high rate of aseptic loosening. Post-operatively, the mean maximum flexion was 136° (110° to 140°) in the loosened group and 125° (95° to 140°) in the well-fixed group (independent t-test, p = 0.022). Squatting, kneeling, or sitting cross-legged could be achieved by 23 (85%) of the loosened knees, but only 22 (49%) of the well-fixed knees (chi-squared test, p = 0.001). The loosened femoral components were found to migrate into a more flexed position, but no
This study aimed to identify the tibial component and femoral component coronal angles (TCCAs and FCCAs), which concomitantly are associated with the best outcomes and survivorship in a cohort of fixed-bearing, cemented, medial unicompartmental knee arthroplasties (UKAs). We also investigated the potential two-way interactions between the TCCA and FCCA. Prospectively collected registry data involving 264 UKAs from a single institution were analyzed. The TCCAs and FCCAs were measured on postoperative radiographs and absolute angles were analyzed. Clinical assessment at six months, two years, and ten years was undertaken using the Knee Society Knee score (KSKS) and Knee Society Function score (KSFS), the Oxford Knee Score (OKS), the 36-Item Short-Form Health Survey questionnaire (SF-36), and range of motion (ROM). Fulfilment of expectations and satisfaction was also recorded. Implant survivorship was reviewed at a mean follow-up of 14 years (12 to 16). Multivariate regression models included covariates, TCCA, FCCA, and two-way interactions between them. Partial residual graphs were generated to identify angles associated with the best outcomes. Kaplan-Meier analysis was used to compare implant survivorship between groups.Aims
Methods
The anterior cruciate ligament (ACL) is known to have a poor wound healing capacity, whereas other ligaments outside of the knee joint capsule such as the medial collateral ligament (MCL) apparently heal more easily. Plasmin has been identified as a major component in the synovial fluid that varies among patients. The aim of this study was to test whether plasmin, a component of synovial fluid, could be a main factor responsible for the poor wound healing capacity of the ACL. The effects of increasing concentrations of plasmin (0, 0.1, 1, 10, and 50 µg/ml) onto the wound closing speed (WCS) of primary ACL-derived ligamentocytes (ACL-LCs) were tested using wound scratch assay and time-lapse phase-contrast microscopy. Additionally, relative expression changes (quantitative PCR (qPCR)) of major LC-relevant genes and catabolic genes were investigated. The positive controls were 10% fetal calf serum (FCS) and platelet-derived growth factor (PDGF).Aims
Methods
Total knee arthroplasty is an established treatment for knee osteoarthritis with excellent long-term results, but there remains controversy about the role of uncemented prostheses. We present the long-term results of a randomized trial comparing an uncemented tantalum metal tibial component with a conventional cemented component of the same implant design. Patients under the age of 70 years with symptomatic osteoarthritis of the knee were randomized to receive either an uncemented tantalum metal tibial monoblock component or a standard cemented modular component. The mean age at time of recruitment to the study was 63 years (50 to 70), 46 (51.1%) knees were in male patients, and the mean body mass index was 30.4 kg/m2 (21 to 36). The same cruciate retaining total knee system was used in both groups. All patients received an uncemented femoral component and no patients had their patella resurfaced. Patient outcomes were assessed preoperatively and postoperatively using the modified Oxford Knee Score, Knee Society Score, and 12-Item Short-Form Health Survey questionnaire (SF-12) score. Radiographs were analyzed using the American Knee Society Radiograph Evaluation score. Operative complications, reoperations, or revision surgery were recorded. A total of 90 knees were randomized and at last review 77 knees were assessed. In all, 11 patients had died and two were lost to follow-up.Aims
Methods
Metaphyseal fixation during revision total knee arthroplasty (TKA) is important, but potentially difficult when using historical designs of cone. Material and manufacturing innovations have improved the size and shape of the cones which are available, and simplified the required bone preparation. In a large series, we assessed the implant survivorship, radiological results, and clinical outcomes of new porous 3D-printed titanium metaphyseal cones featuring a reamer-based system. We reviewed 142 revision TKAs in 139 patients using 202 cones (134 tibial, 68 femoral) which were undertaken between 2015 and 2016. A total of 60 involved tibial and femoral cones. Most cones (149 of 202; 74%) were used for Type 2B or 3 bone loss. The mean age of the patients was 66 years (44 to 88), and 76 (55 %) were female. The mean body mass index (BMI) was 34 kg/m2 (18 to 60). The patients had a mean of 2.4 (1 to 8) previous operations on the knee, and 68 (48%) had a history of prosthetic infection. The mean follow-up was 2.4 years (2 to 3.6).Aims
Methods
Aims
Patients and Methods
We reviewed 100 cemented Endo-model rotating-hinge (Waldemar Link GMBH &
Co, Hamburg, Germany) total knee replacements in 80 patients with a mean age of 70 years (56 to 85) at a mean post-operative follow-up of 11 years (7 to 15). Good or excellent results were seen in 91% of knees and survival at 15 years was 96.1%. There were two deep infections, one dislocation and one supracondylar fracture. There were no peri-operative deaths and there was no loosening, malalignment,
We have found poor mid-term results in a multisurgeon series of 94 Johnson-Elloy (Accord) total knee replacements. A total of 27 knees (29%) has required revision, in 26 for aseptic loosening. Only 18 (19%) remain in situ, and these give poor function, are painful and most show radiological evidence of early failure. At 12 to 13 years the survival rate is 43% (confidence interval 29 to 57) with failure requiring revision as the endpoint. Proximal
We studied the quality of fixation of the tibial component using radiostereometric analysis (RSA) in 40 patients who had undergone a cemented Freeman-Samuelson total knee arthroplasty. They were prospectively randomised to either a stemmed metal-backed (MB) or non-stemmed all-polyethylene (AP) tibial component. The articulating geometry of the implants was identical, as was the operative technique and the postoperative regime. The study showed no complications of fixation using AP tibial components, and the
We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total knee prostheses in 32 young patients with primary malignant bone tumours. Initial stable fixation, especially in rotation, was achieved by precise fit of the stem into the reamed endosteum, before cementation. Complementary fixation, especially in
The treatment of patients with allergies to metal in total joint arthroplasty is an ongoing debate. Possibilities include the use of hypoallergenic prostheses, as well as the use of standard cobalt-chromium (CoCr) alloy. This non-designer study was performed to evaluate the clinical outcome and survival rates of unicondylar knee arthroplasty (UKA) using a standard CoCr alloy in patients reporting signs of a hypersensitivity to metal. A consecutive series of patients suitable for UKA were screened for symptoms of metal hypersensitivity by use of a questionnaire. A total of 82 patients out of 1737 patients suitable for medial UKA reporting cutaneous metal hypersensitivity to cobalt, chromium, or nickel were included into this study and prospectively evaluated to determine the functional outcome, possible signs of hypersensitivity, and short-term survivorship at a minimum follow-up of 1.5 years.Aims
Patients and Methods
The aim of this study was to evaluate the long-term inducible
displacement of cemented tibial components ten years after total
knee arthroplasty (TKA). A total of 15 patients from a previously reported prospective
trial of fixation using radiostereometric analysis (RSA) were examined
at a mean of 11 years (10 to 11) postoperatively. Longitudinal supine
RSA examinations were acquired at one week, one year, and two years
postoperatively and at final follow-up. Weight-bearing RSA examinations
were also undertaken with the operated lower limb in neutral and
in maximum internal rotation positions. Maximum total point motion
(MTPM) was calculated for the longitudinal and inducible displacement examinations
(supine Aims
Patients and Methods
This study compares the PFC total knee arthroplasty (TKA) system in a prospective randomized control trial (RCT) of the mobile-bearing rotating-platform (RP) TKA against the fixed-bearing (FB) TKA. This is the largest RCT with the longest follow-up where cruciate-retaining PFC total knee arthroplasties are compared in a non-bilateral TKA study. A total of 167 patients (190 knees with 23 bilateral cases), were recruited prospectively and randomly assigned, with 91 knees receiving the RP and 99 knees receiving FB. The mean age was 65.5 years (48 to 82), the mean body mass index (BMI) was 29.7 kg/m2 (20 to 52) and 73 patients were female. The Knee Society Score (KSS), Knee Society Functional Score (KSFS), Oxford Knee Score (OKS), Western Ontario and McMaster Universities Arthritis Index (WOMAC), and 12-Item Short-Form Health Survey Physical and Mental Component Scores (SF-12 PCS, SF-12 MCS) were gathered and recorded preoperatively, at five-years’ follow-up, and at ten years’ follow-up. Additionally, Knee Injury and Osteoarthritis Outcome Scores (KOOS) were collected at five- and ten-year follow-ups. The prevalence of radiolucent lines (RL) on radiographs and implant survival were recorded at five- and ten-year follow-ups.Aims
Patients and Methods
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
The Advance Medial-Pivot total knee arthroplasty (TKA) was designed
to reflect contemporary data regarding the kinematics of the knee.
We wished to examine the long-term results obtained with this prosthesis
by extending a previous evaluation. We retrospectively evaluated prospectively collected data from
225 consecutive patients (41 men and 184 women; mean age at surgery
71 years, 52 to 84) who underwent 284 TKAs with a mean follow-up
of 13.4 years (11 to 15). Implant failure, complication rate, clinical (both
subjective and objective) and radiological outcome were assessed.
Pre- and post-operative clinical and radiographic data were available
at regular intervals for all patients. A total of ten patients (4.4%;
ten TKAs) were lost to follow-up.Aims
Patients and Methods
The relationship between post-operative bone
density and subsequent failure of total knee replacement (TKR) is
not known. This retrospective study aimed to determine the relationship
between bone density and failure, both overall and according to
failure mechanism. All 54 aseptic failures occurring in 50 patients
from 7760 consecutive primary cemented TKRs between 1983 and 2004
were matched with non-failing TKRs, and 47 failures in 44 patients
involved tibial failures with the matching characteristics of age
(65.1 for failed and 69.8 for non-failed), gender (70.2% female), diagnosis
(93.6% OA), date of operation, bilaterality, pre-operative alignment
(0.4 and 0.3 respectively), and body mass index (30.2 and 30.0 respectively).
In each case, the density of bone beneath the tibial component was assessed
at each follow-up interval using standardised, calibrated radiographs.
Failing knees were compared with controls both overall and, as a
subgroup analysis, by failure mechanism. Knees were compared with
controls using univariable linear regression. Significant and continuous elevation in tibial density was found
in knees that eventually failed by medial collapse (p <
0.001)
and progressive radiolucency (p <
0.001) compared with controls,
particularly in the medial region of the tibia. Knees failing due
to ligamentous instability demonstrated an initial decline in density
(p = 0.0152) followed by a non-decreasing density over time (p =
0.034 for equivalence). Non-failing knees reported a decline in
density similar to that reported previously using dual-energy x-ray
absorptiometry (DEXA). Differences between failing and non-failing
knees were observable as early as two months following surgery.
This tool may be used to identify patients at risk of failure following
TKR, but more validation work is needed. Cite this article:
The purpose of this study was to develop an accurate, reliable and easily applicable method for determining the anatomical location of the joint line during revision knee arthroplasty. The transepicondylar width (TEW), the perpendicular distance between the medial and lateral epicondyles and the distal articular surfaces (DMAD, DLAD) and the distance between the medial and lateral epicondyles and the posterior articular surfaces (PMAD, DLAD) were measured in 40 knees from 20 formalin-fixed adult cadavers (11 male and nine female; mean age at death 56.9 years, Objectives
Methods
A key to the success of revision total knee arthroplasty
(TKA) is a safe surgical approach using an exposure that minimises
complications. In most patients, a medial parapatellar arthrotomy
with complete synovectomy is sufficient. If additional exposure
is needed, a quadriceps snip performed through the quadriceps tendon
often provides the additional exposure required. It is simple to
perform and does not alter the post-operative rehabilitative protocol.
In rare cases, in which additional exposure is needed, or when removal
of a cemented long-stemmed tibial component is required, a tibial
tubercle osteotomy (TTO) may be used. Given the risk of post-operative
extensor lag, a V-Y quadricepsplasty is rarely indicated and usually
considered only if TTO is not possible. Cite this article:
The term mid-flexion instability has entered
the orthopaedic literature as a concept, but has not been confirmed
as a distinct clinical entity. The term is used freely, sometimes
as a synonym for flexion instability. However, the terms need to
be clearly separated. A cadaver study published in 1990 associated
joint line elevation with decreased stability at many angles of
flexion, but that model was not typical of clinical scenarios. The
literature is considered and it is proposed that the more common
entity of an uncorrected flexion contracture after a measured resection arthroplasty
technique is more likely to produce clinical findings that suggest
instability mid-flexion. It is proposed that the clinical scenario encountered is generalised
instability, with the appearance of stability in full extension
from tight posterior structures. This paper seeks to clarify whether mid-flexion instability exists
as an entity distinct from other commonly recognised forms of instability. Cite this article:
This randomised trial evaluated the outcome of
a single design of unicompartmental arthroplasty of the knee (UKA) with
either a cemented all-polyethylene or a metal-backed modular tibial
component. A total of 63 knees in 45 patients (17 male, 28 female)
were included, 27 in the all-polyethylene group and 36 in the metal-backed
group. The mean age was 57.9 years (39.6 to 76.9). At a mean follow-up
of 6.4 years (5 to 9.9), 11 all-polyethylene components (41%) were
revised (at a mean of 5.8 years; 1.4 to 8.0) post-operatively and
two metal-backed components were revised (at one and five years).
One revision in both groups was for unexplained pain, one in the
metal-backed group was for progression of osteoarthritis. The others
in the all-polyethylene group were for aseptic loosening. The survivorship
at seven years calculated by the Kaplan–Meier method for the all-polyethylene
group was 56.5% (95% CI 31.9 to 75.2, number at risk 7) and for
the metal-backed group was 93.8% (95% CI 77.3 to 98.4, number at
risk 16) This difference was statistically significant (p <
0.001).
At the most recent follow-up, significantly better mean Western
Ontario and McMaster Universities Arthritis Index Scores were found
in the all-polyethylene group (13.4 This randomised study demonstrates that all-polyethylene components
in this design of fixed bearing UKA had unsatisfactory results with
significantly higher rates of failure before ten years compared
with the metal-back components. Cite this article:
Stems improve the mechanical stability of tibial
components in total knee replacement (TKR), but come at a cost of stress
shielding along their length. Their advantages include resistance
to shear, reduced tibial lift-off and increased stability by reducing
micromotion. Longer stems may have disadvantages including stress
shielding along the length of the stem with associated reduction
in bone density and a theoretical risk of subsidence and loosening, peri-prosthetic
fracture and end-of-stem pain. These features make long stems unattractive
in the primary TKR setting, but often desirable in revision surgery
with bone loss and instability. In the revision scenario, stems
are beneficial in order to convey structural stability to the construct
and protect the reconstruction of bony defects. Cemented and uncemented
long stemmed implants have different roles depending on the nature
of the bone loss involved. This review discusses the biomechanics of the design of tibial
components and stems to inform the selection of the component and
the technique of implantation.
Valgus high tibial osteotomy for osteoarthritis of the medial compartment of the knee can be performed using medial opening- and lateral closing-wedge techniques. The latter have been thought to offer greater initial stability. We measured and compared the stability of opening- and closing-wedge osteotomies fixed by TomoFix plates using radiostereometry in a series of 42 patients in a prospective, randomised clinical trial. There were no differences between the opening- and closing-wedge groups in the time to regain knee function and full weight-bearing. Pain and knee function were significantly improved in both groups without any differences between them. All the osteotomies united within one year. Radiostereometry showed no clinically relevant movement of bone or differences between either group. Medial opening-wedge high tibial osteotomy secured by a TomoFix plate offers equal stability to a lateral closing-wedge technique. Both give excellent initial stability and provide significantly improved knee function and reduction in pain, although the opening-wedge technique was more likely to produce the intended correction.
The cementless Oxford unicompartmental knee replacement
has been demonstrated to have superior fixation on radiographs and
a similar early complication rate compared with the cemented version.
However, a small number of cases have come to our attention where,
after an apparently successful procedure, the tibial component subsides into
a valgus position with an increased posterior slope, before becoming
well-fixed. We present the clinical and radiological findings of
these six patients and describe their natural history and the likely
causes. Two underwent revision in the early post-operative period,
and in four the implant stabilised and became well-fixed radiologically with
a good functional outcome. This situation appears to be avoidable by minor modifications
to the operative technique, and it appears that it can be treated
conservatively in most patients. Cite this article:
The aim of this study was to evaluate the risk
factors for dislocation of the bearing after a mobile-bearing Oxford medial
unicompartmental knee replacement (UKR) and to test the hypothesis
that surgical factors, as measured from post-operative radiographs,
are associated with its dislocation From a total of 480 UKRs performed between 2001 and 2012, in
391 patients with a mean age of 66.5 years (45 to 82) (316 female,
75 male), we identified 17 UKRs where bearing dislocation occurred.
The post-operative radiological measurements of the 17 UKRs and
51 matched controls were analysed using conditional logistic regression analysis.
The post-operative radiological measurements included post-operative
change in limb alignment, the position of the femoral and tibial
components, the resection depth of the proximal tibia, and the femoral component-posterior
condyle classification. We concluded that a post-operative decrease in the posterior
tibial slope relative to the pre-operative value was the only significant
determinant of dislocation of the bearing after medial Oxford UKR
(odds ratio 1.881; 95% confidence interval 1.272 to 2.779). A post-operative
posterior tibial slope <
8.45° and a difference between the pre-operative
and post-operative posterior tibial slope of >
2.19° may increase
the risk of dislocation. Cite this article:
This prospective study compares the outcome of
157 hydroxyapatite (HA)-coated tibial components with 164 cemented
components in the ROCC Rotating Platform total knee replacement
in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There
were two revisions for loosening: one for an HA-coated and one for
a cemented tibial component. Radiological evaluation demonstrated
no radiolucent lines with the HA-coated femoral components. A total
of three HA-coated tibial components exhibited radiolucent lines
at three months post-operatively and these disappeared after three
further months of protected weight-bearing. With HA-coated components
the operating time was shorter (p <
0.006) and the radiological
assessment of the tibial interface was more stable (p <
0.01).
Using revision for aseptic loosening of the tibial component as
the end point, the survival rates at nine years was identical for
both groups at 99.1%. Our results suggest that HA-coated components perform at least
as well as the same design with cemented components and compare
favourably with those of series describing cemented or porous-coated
knee replacements, suggesting that fixation of both components with
hydroxyapatite is a reliable option in primary total knee replacement.
We have carried out a radiostereometric study of 50 patients (54 knees) with osteoarthritis of the knee who were randomly allocated to receive a cemented or a hydroxyapatite-coated femoral component for total knee replacement. The patients were also stratified to receive one of three types of articulating surface (standard, rotating platform, Freeman-Samuelson (FS)1000) all based on the Freeman-Samuelson design. The tibial components were cemented in all cases. Radiostereometry was performed post-operatively and at 3, 12 and 24 months. The analysis was restricted to rotation of the femoral component over time. After two years, rotation of the femoral components in the transverse, longitudinal and sagittal planes did not differ between the cemented and the hydroxyapatite-coated implants (p = 0.2 to 0.9). In total knee replacements with a rotating platform, the femoral component tended to tilt more posteriorly than in the other two designs, regardless of the choice of fixation (cemented or hydroxyapatite-coated, p = 0.04). The standard version of the femoral component, whether cemented or hydroxyapatite-coated, rotated more into valgus than was observed with the rotating-platform and FS1000 designs (p = 0.005). The increased constraint provided by the FS1000 component did not appear to have any adverse effect on fixation of the femoral component.
We report the results of revision total knee
replacement (TKR) in 26 patients with major metaphyseal osteolytic defects
using 29 trabecular metal cones in conjunction with a rotating hinged
total knee prosthesis. The osteolytic defects were types II and
III (A or B) according to the Anderson Orthopaedic Research Institute
(AORI) classification. The mean age of the patients was 72 years
(62 to 84) and there were 15 men and 11 women. In this series patients had
undergone a mean of 2.34 previous total knee arthroplasties. The
main objective was to restore anatomy along with stability and function
of the knee joint to allow immediate full weight-bearing and active
knee movement. Outcomes were measured using Knee Society scores,
Oxford knee scores, range of movement of the knee and serial radiographs.
Patients were followed for a mean of 36 months (24 to 49). The mean
Oxford knee clinical scores improved from 12.83 (10 to 15) to 35.20
(32 to 38) (p <
0.001) and mean American Knee Society scores
improved from 33.24 (13 to 36) to 81.12 (78 to 86) (p <
0.001).
No radiolucent lines suggestive of loosening were seen around the trabecular
metal cones, and by one year all the radiographs showed good osteo-integration.
There was no evidence of any collapse or implant migration. Our
early results confirm the findings of others that trabecular metal
cones offer a useful way of managing severe bone loss in revision
TKR. Cite this article:
Structural allografts may be used to manage uncontained
bone defects in revision total knee replacement (TKR). However,
the availability of cadaver grafts is limited in some areas of Asia.
The aim of this study was to evaluate the mid-term outcome of the
use of femoral head allografts for the reconstruction of uncontained
defects in revision TKR, focusing on complications related to the
graft. We retrospectively reviewed 28 patients (30 TKRs) with Anderson
Orthopaedic Research Institute (AORI) type 3 bone defects, who underwent
revision using femoral head allografts and stemmed components. The
mean number of femoral heads used was 1.7 (1 to 3). The allograft–host
junctions were packed with cancellous autograft. At a mean follow-up of 76 months (38 to 136) the mean American
Knee Society knee score improved from 37.2 (17 to 60) pre-operatively
to 90 (83 to 100) (p <
0.001). The mean function score improved
from 26.5 (0 to 50) pre-operatively to 81 (60 to 100) (p <
0.001).
All the grafts healed to the host bone. The mean time to healing
of the graft was 6.6 months
(4 to 16). There have been no complications of collapse of the graft,
nonunion, infection or implant loosening. No revision surgery was
required. The use of femoral head allografts in conjunction with a stemmed
component and autogenous bone graft in revision TKR in patients
with uncontained bone defects resulted in a high rate of healing
of the graft with minimal complications and a satisfactory outcome.
Longer follow-up is needed to observe the evolution of the graft. Cite this article:
The purpose of this study was to undertake a
meta-analysis to determine whether there is lower polyethylene wear and
longer survival when using mobile-bearing implants in total knee
replacement when compared with fixed-bearing implants. Of 975 papers
identified, 34 trials were eligible for data extraction and meta-analysis
comprising 4754 patients (6861 knees). We found no statistically
significant differences between the two designs in terms of the incidence
of radiolucent lines, osteolysis, aseptic loosening or survival.
There is thus currently no evidence to suggest that the use of mobile-bearing
designs reduce polyethylene wear and prolong survival after total
knee replacement. Cite this article:
Using meta-analysis we compared the survival and clinical outcomes of cemented and uncemented techniques in primary total knee replacement. We reviewed randomised controlled trials and observational studies comparing cemented and uncemented fixation. Our primary outcome was survival of the implant free of aseptic loosening. Our secondary outcome was joint function as measured by the Knee Society score. We identified 15 studies that met our final eligibility criteria. The combined odds ratio for failure of the implant due to aseptic loosening for the uncemented group was 4.2 (95% confidence interval (CI) 2.7 to 6.5) (p <
0.0001). Subgroup analysis of data only from randomised controlled trials showed no differences between the groups for odds of aseptic loosening (odds ratio 1.9, 95% CI 0.55 to 6.40, p = 0.314). The weighted mean difference for the Knee Society score was 0.005 (95% CI −0.26 to 0.26) (p = 0.972). There was improved survival of the cemented compared to uncemented implants, with no statistically significant difference in the mean Knee Society score between groups for all pooled data.
In this study we present our experience with
four generations of uncemented total knee arthroplasty (TKA) from Smith
&
Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing
on the failure rates correlating with each design change. Beginning
in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix
were implanted by the senior author. The rate of revision for loosening
was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the
Tricon 2 with a HA coated tibial component, and 1.3% for the Profix
TKA. No loosening of the femoral component was seen with the Tricon
M, Tricon LS or Tricon 2, with no loosening seen of the tibial component
with the Profix TKA. Regarding revision for wear, the incidence
was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the
Tricon 2, and 0% for the Profix. These results demonstrate that
improvements in the design of uncemented components, including increased
polyethylene thickness, improved polyethylene quality, and the introduction
of hydroxyapatite coating, has improved the outcomes of uncemented
TKA over time.
Matrix-assisted autologous chondrocyte transplantation (MACT)
has been developed and applied in the clinical practice in the last
decade to overcome most of the disadvantages of the first generation
procedures. The purpose of this systematic review is to document
and analyse the available literature on the results of MACT in the
treatment of chondral and osteochondral lesions of the knee. All studies published in English addressing MACT procedures were
identified, including those that fulfilled the following criteria:
1) level I-IV evidence, 2) measures of functional or clinical outcome,
3) outcome related to cartilage lesions of the knee cartilage.Objectives
Methods
This was a retrospective analysis of the medium-
to long-term results of 46 TC3 Sigma revision total knee replacements
using long uncemented stems in press-fit mode. Clinical and radiological analysis took place pre-operatively,
at two years post-operatively, and at a mean follow-up of 8.5 years
(4 to 12). The mean pre-operative International Knee Society (IKS)
clinical score was 42 points (0 to 74), improving to 83.7 (52 to
100) by the final follow-up. The mean IKS score for function improved
from 34.3 points (0 to 80) to 64.2 (15 to 100) at the final follow-up.
At the final follow-up 30 knees (65.2%) had an excellent result, seven
(15.2%) a good result, one (2.2%) a medium and eight (17.4%) a poor
result. There were two failures, one with anteroposterior instability
and one with aseptic loosening. The TC3 revision knee system, when used with press-fit for long
intramedullary stems and cemented femoral and tibial components,
in both septic and aseptic revisions, results in a satisfactory
clinical and radiological outcome, and has a good medium- to long-term
survival rate.
Medium-term survivorship of the Oxford phase
3 unicompartmental knee replacement (UKR) has not yet been established
in an Asian population. We prospectively evaluated the outcome of
400 phase 3 Oxford UKRs in 320 Korean patients with a mean age at
the time of operation of 69 years (48 to 82). The mean follow-up
was 5.2 years (1 to 10). Clinical and radiological assessment was
carried out pre- and post-operatively. At five years, the mean Knee
Society knee and functional scores had increased significantly from
56.2 (30 to 91) pre-operatively to 87.2 (59 to 98) (p = 0.034) and
from 59.2 (30 to 93) to 88.3 (50 to 100) (p = 0.021), respectively.
The Oxford knee score increased from a mean of 25.8 (12 to 39) pre-operatively
to 39.8 (25 to 58) at five years (p = 0.038). The ten-year survival
rate was 94% (95% confidence interval 90.1 to 98.0). A total of
14 UKRs (3.5%) required revision. The most common reason for revision
was dislocation of the bearing in 12 (3%). Conversion to a total
knee replacement was required in two patients who developed osteoarthritis
of the lateral compartment. This is the largest published series of UKR in Korean patients.
It shows that the mid-term results after a minimally invasive Oxford
phase 3 UKR can yield satisfactory clinical and functional results
in this group of patients.
We evaluated the survival of moulded monoblock and modular tibial components of the AGC total knee replacement in patients with rheumatoid arthritis. Between 1985 and 1995, 751 knees with this diagnosis were replaced at our institution. A total of 256 tibial components were of the moulded design and 495 of the modular design. The mean follow-up of the moulded subgroup was 9.6 years (0.5 to 14.7), and that of the modular group 7.0 years (0.1 to 14.7). The groups differed significantly from each other in Larsen grade, cementing of components and patellar resurfacing, but no statistically significant difference between the survival of the components was found (Log rank test, p = 0.91). The cumulative success rate of the moulded group was 96.8% (95% confidence interval 93.6% to 98.4%) at five years and 94.4% (95% confidence interval 90.4% to 96.7%) at ten years, and of the modular group 96.2% (95% confidence interval 94% to 97.6%) and 93.6% (95% confidence interval 89.7% to 96%), respectively. Revision was required in 37 total knee replacements, the main causes were infection, pain, loosening of the tibial component and patellar problems. Survival rates for both components were satisfactory.
We examined whether enamel matrix derivative
(EMD) could improve healing of the tendon–bone interface following
reconstruction of the anterior cruciate ligament (ACL) using a hamstring
tendon in a rat model. ACL reconstruction was performed in both
knees of 30 Sprague-Dawley rats using the flexor digitorum tendon.
The effect of commercially available EMD (EMDOGAIN), a preparation
of matrix proteins from developing porcine teeth, was evaluated.
In the left knee joint the space around the tendon–bone interface
was filled with 40 µl of EMD mixed with propylene glycol alginate
(PGA). In the right knee joint PGA alone was used. The ligament
reconstructions were evaluated histologically and biomechanically
at four, eight and 12 weeks (n = 5 at each time point). At eight weeks,
EMD had induced a significant increase in collagen fibres connecting
to bone at the tendon–bone interface (p = 0.047), whereas the control
group had few fibres and the tendon–bone interface was composed
of cellular and vascular fibrous tissues. At both eight and 12 weeks,
the mean load to failure in the treated specimens was higher than
in the controls (p = 0.009). EMD improved histological tendon–bone
healing at eight weeks and biomechanical healing at both eight and
12 weeks. EMD might therefore have a human application to enhance
tendon–bone repair in ACL reconstruction.
We carried out a prospective study to assess the clinical outcome, complications and survival of the NexGen Legacy posterior-stabilised-Flex total knee replacement (TKR) in a consecutive series of 278 knees between May 2003 and February 2005. Mean follow-up for 259 TKRs (98.2%) was 3.8 years (3.0 to 4.8). Annual follow-up showed improvement in the Knee Society scores (paired This relatively large study indicates that the legacy posterior stabilised-Flex design provides excellent short-term outcome but warrants ongoing evaluation to confirm the long-term durability and functioning of the implant.
Bone mineral density (BMD) around the femoral component has been reported to decrease after total knee replacement (TKR) because of stress shielding. Our aim was to determine whether a cemented mobile-bearing component reduced the post-operative loss of BMD. In our study 28 knees receiving a cemented fixed-bearing TKR were matched with 28 receiving a cemented mobile-bearing TKR. They underwent dual-energy x-ray absorptiometry, pre-operatively and at three weeks and at three, six, 12, 18 and 24 months post-operatively. The patients were not taking medication to improve the BMD. The pre-operative differences in the BMD of the femoral neck, wrist, lumbar spine and knee in the two groups were not significant. The BMD of the femur decreased postoperatively in the fixed-bearing group, but not the mobile-bearing group. The difference in the post-operative change in the BMD in the two groups was statistically significant (p <
0.05) at 18 and 24 months. Our findings show that a cemented mobile-bearing TKR has a favourable effect on the BMD of the distal femur after TKR in the short term. Further study is required to determine the long-term effects.
Autologous chondrocyte implantation (ACI) is used widely as a treatment for symptomatic chondral and osteochondral defects of the knee. Variations of the original periosteum-cover technique include the use of porcine-derived type I/type III collagen as a cover (ACI-C) and matrix-induced autologous chondrocyte implantation (MACI) using a collagen bilayer seeded with chondrocytes. We have performed a prospective, randomised comparison of ACI-C and MACI for the treatment of symptomatic chondral defects of the knee in 91 patients, of whom 44 received ACI-C and 47 MACI grafts. Both treatments resulted in improvement of the clinical score after one year. The mean modified Cincinnati knee score increased by 17.6 in the ACI-C group and 19.6 in the MACI group (p = 0.32). Arthroscopic assessments performed after one year showed a good to excellent International Cartilage Repair Society score in 79.2% of ACI-C and 66.6% of MACI grafts. Hyaline-like cartilage or hyaline-like cartilage with fibrocartilage was found in the biopsies of 43.9% of the ACI-C and 36.4% of the MACI grafts after one year. The rate of hypertrophy of the graft was 9% (4 of 44) in the ACI-C group and 6% (3 of 47) in the MACI group. The frequency of re-operation was 9% in each group. We conclude that the clinical, arthroscopic and histological outcomes are comparable for both ACI-C and MACI. While MACI is technically attractive, further long-term studies are required before the technique is widely adopted.
A total of 218 patients with unilateral anterior cruciate ligament deficiency were randomly assigned to one of four groups. In group A an anatomical double bundle anterior cruciate ligament reconstruction was performed; group B were treated by a single bundle using an Endobutton for femoral fixation; in group C by a single bundle using RigidFix cross pins for femoral fixation; and in group D by a single bundle using a bioabsorbable TransFix II screw for femoral fixation. For tibial fixation a bioabsorbable Intrafix interference screw was used for all the groups and the graft was fashioned from the semitendinosus and gracilis tendons in all patients. In all, 18 patients were lost to follow-up. The remaining 200 were subjected to a clinical evaluation, with assessment of the anterior drawer, Lachman’s and the pivot-shift tests, and KT-1000 arthrometer measurement. They also completed the International Knee Documentation Committee, Lysholm knee and Tegner activity scores. At a mean of 29 months (25 to 38) follow-up there were no significant differences concerning time between injury and range of movement and Lysholm knee scores among the four groups. However, the double bundle method showed significantly better results for the pivot-shift test (p = 0.002). The KT 1000 measurements showed a mean difference between the reconstructed knee and the patients’ normal knee of 1.4 mm in the double bundle group and 2.4 mm in the single bundle group; which was statistically significant. The Lachman and anterior drawer tests also showed superior results for the double bundle method. The International Knee Documentation Committee scale showed no significant difference among the groups (p <
0.001). On clinical evaluation the double bundle group showed less laxity than the single bundle groups. However, regardless of the technique, all knees were improved by anterior cruciate ligament reconstruction compared with their pre-operative status.
The effect of weight-bearing on the height of the patellar using four radiological indices was studied in 25 healthy men using lateral radiographs of the knee in 30° of flexion non-weight-bearing and weight-bearing. The position of the patella was quantified using the Insall-Salvati, the modified Insall-Salvati, the Blackburne-Peel and the Caton indices. The contraction of the quadriceps on weight-bearing resulted in statistically significant proximal displacement of the patella with all four indices studied. The mean Insall-Salvati index was 0.919 ( The effect of quadriceps contraction should be considered in clinical studies where the patellar position indices are reported.
We present the operative technique and clinical results of concomitant reconstruction of the medial collateral ligament (MCL) and the posterior oblique ligament for medial instability of the knee using autogenous semitendinosus tendon with preservation of the tibial attachment. The semitendinosus tendon graft between the screw on the medial epicondyle and the tibial attachment of the graft was overlapped by the MCL, while the graft between the screw and the insertion of the direct head of the semimembranosus tendon was overlapped by the central arm of the posterior oblique ligament. Assessment was by stress radiograph and the Lysholm knee scoring scale. After a mean follow-up of 52.6 months (25 to 92), the medial joint opening of the knee was within 2 mm in 22 of 24 patients. The mean Lysholm score was 91.9 (80 to 100). Concomitant reconstruction of the MCL and posterior oblique ligament using autogenous semitendinosus tendon provides a good solution to medial instability.
We identified 148 patients who had undergone a revision total knee replacement using a single implant system between 1990 and 2000. Of these 18 patients had died, six had developed a peri-prosthetic fracture and ten had incomplete records or radiographs. This left 114 with prospectively-collected radiographs and Bristol knee scores available for study. The height of the joint line before and after revision total knee replacement was measured and classified as either restored to within 5 mm of the pre-operative height or elevated if it was positioned more than 5 mm above the pre-operative height. The joint line was elevated in 41 knees (36%) and restored in 73 (64%). Revision surgery significantly improved the mean Bristol knee score from 41.1 ( Our findings show that restoration of the joint line at revision total knee replacement gives a significantly better result than leaving it unrestored by more than 5 mm. We recommend the greater use of distal femoral augments to help to achieve this goal.
The Oxford medial unicompartmental knee replacement was designed to reproduce normal mobility and forces in the knee, but its detailed effect on the patellofemoral joint has not been studied previously. We have examined the effect on patellofemoral mechanics of the knee by simultaneously measuring patellofemoral kinematics and forces in 11 cadaver knee specimens in a supine leg-extension rig. Comparison was made between the intact normal knee and sequential unicompartmental and total knee replacement. Following medial mobile-bearing unicompartmental replacement in 11 knees, patellofemoral kinematics and forces did not change significantly from those in the intact knee across any measured parameter. In contrast, following posterior cruciate ligament retaining total knee replacement in eight knees, there were significant changes in patellofemoral movement and forces. The Oxford device appears to produce near-normal patellofemoral mechanics, which may partly explain the low incidence of complications with the extensor mechanism associated with clinical use.
We examined the placement of the stem in relation to the medial tibial cortex when using total knee replacements (TKRs) with medially-offset tibial stems in Korean patients. Measurements were performed on the pre- and post-operative radiographs of 246 osteoarthritic knees replaced between January 2005 and May 2006 using the Genesis II or E-motion TKR with a medially-offset stem. Pre-operatively, we measured the distance between the mechanical axis and that of the tibial shaft and post-operatively, that between the midline of the tibial stem and the axis of the shaft. Knees were identified in which there was radiological contact between the tip of the stem and the medial tibial cortex. The mechanical axis was located medial to the axis of the shaft in 203 knees (82.5%). Post-operatively, the midline of the tibial stem was located medial to the tibial shaft axis in 196 knees (79.7%). In 16 knees (6.5%) there was radiological contact between the tibial stem or cement mantle and the medial tibial cortex. Our study has shown that the medially-offset stem in the tibial component may not be a good option for knees undergoing replacement for advanced osteoarthritis in some Korean patients.
We present a comparison of the results of the Oxford unicompartmental knee arthroplasty in patients younger and older than 60 years of age. The ten-year all-cause survival of the <
60 years of age group (52) was 91% (95% confidence interval (CI) 12), while in the ≥ 60 years of age group (512), the figure was 96% (95% CI 3). For the younger group, the mean Hospital for Special Surgery score at ten-year follow-up (n = 21) was 94 of 100, compared with a mean of 86 of 100 for the older group (n = 135). The results show that the Oxford unicompartmental arthroplasty can achieve ten-year results that are comparable to total knee arthroplasty in patients <
60 years of age. We conclude that for patients aged over 50, age should not be considered a contraindication for this procedure.