Aims. Metal particles detached from metal-on-metal hip prostheses (MoM-THA) have been shown to cause inflammation and destruction of tissues. To further explore this, we investigated the histopathology (aseptic lymphocyte-dominated vasculitis-associated lesions (ALVAL) score) and metal concentrations of the periprosthetic tissues obtained from patients who underwent revision knee arthroplasty. We also aimed to investigate whether accumulated metal debris was associated with ALVAL-type reactions in the synovium. Methods. Periprosthetic metal concentrations in the synovia and histopathological samples were analyzed from 230 patients from our institution from October 2016 to December 2019. An ordinal regression model was calculated to investigate the effect of the accumulated metals on the histopathological reaction of the synovia. Results. Median metal concentrations were as follows: cobalt: 0.69 μg/g (interquartile range (IQR) 0.10 to 6.10); chromium: 1.1 μg/g (IQR 0.27 to 4.10); and titanium: 1.6 μg/g (IQR 0.90 to 4.07). Moderate ALVAL scores were found in 30% (n = 39) of the revised knees. There were ten patients with an ALVAL score of 6 or more who were revised for suspected periprosthetic joint infection (PJI), aseptic loosening, or osteolysis. R2 varied between 0.269 and 0.369 for the ordinal regression models. The most important variables were model type, indication for revision, and cobalt and chromium in the ordinal regression models. Conclusion. We found that metal particles released from the
Aims. This study aimed to evaluate the association between the sagittal alignment of the femoral component in total knee arthroplasty (TKA) and new Knee Society Score (2011KSS), under the hypothesis that outliers such as the excessive extended or flexed femoral component were related to worse clinical outcomes. Methods. A group of 156 knees (134 F:22 M) in 133 patients with a mean age 75.8 years (SD 6.4) who underwent TKA with the cruciate-substituting Bi-Surface
A new generation of
We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total
Before proceeding to longer-term studies, we have studied the early clinical results of a new mobile-bearing total
A total of 344 patients underwent bilateral total knee replacement (TKR) using a different prosthesis on each side. Four
The material and design of knee components can have a considerable effect on the contact characteristics of the tibial post. This study aimed to analyze the stress distribution on the tibial post when using different grades of polyethylene for the tibial inserts. In addition, the contact properties of fixed-bearing and mobile-bearing inserts were evaluated. Three different grades of polyethylene were compared in this study; conventional ultra high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (HXLPE), and vitamin E-stabilized polyethylene (VEPE). In addition, tibial baseplates with a fixed-bearing and a mobile-bearing insert were evaluated to understand differences in the contact properties. The inserts were implanted in neutral alignment and with a 10° internal malrotation. The contact stress, von Mises stress, and equivalent plastic strain (PEEQ) on the tibial posts were extracted for comparison.Aims
Methods
Experiments were carried out to determine the optimum conformity between the femoral and tibial condyles in condylar replacement
1. A case of late infection about a total
We examined 86 polyethylene inserts, retrieved from total and unicompartmental
The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs). A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively.Aims
Methods
Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes. A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively.Aims
Methods
We examined the risk of thrombotic and major
bleeding events in patients undergoing total hip and knee replacement
(THR and TKR) treated with thromboprophylaxis, using nationwide
population-based databases. We identified 83 756 primary procedures
performed between 1997 and 2011. The outcomes were symptomatic venous thromboembolism
(VTE), myocardial infarction (MI), stroke, death and major bleeding
requiring hospitalisation within 90 days of surgery. A total of 1114 (1.3%) and 483 (0.6%) patients experienced VTE
and bleeding, respectively. The annual risk of VTE varied between
0.9% and 1.6%, and of bleeding between 0.4% and 0.8%. The risk of
VTE and bleeding was unchanged over a 15-year period. A total of
0.7% of patients died within 90 days, with a decrease from 1% in
1997 to 0.6% in 2011 (p <
0.001). A high level of comorbidity
and general anaesthesia were strong risk factors for both VTE and
bleeding, with no difference between THR and TKR patients. The risk
of both MI and stroke was 0.5%, which remained unchanged during
the study period. In this cohort study of patients undergoing THR and TKR patients
in routine clinical practice, approximately 3% experienced VTE,
MI, stroke or bleeding. These risks did not decline during the 15-year
study period, but the risk of dying fell substantially. Cite this article:
The aim of this study was to assess the effects of transferring patients to a specialized arthroplasty centre between the first and second stages (interstage) of prosthetic joint infection (PJI) of the knee. A search of our institutional database was performed to identify patients having undergone two-stage revision total knee arthroplasty (TKA) for PJI. Two cohorts were created: continuous care (CC) and transferred care (TC). Baseline characteristics and outcomes were collected and compared between cohorts.Aims
Patients and Methods
This single-blinded randomised controlled trial
investigated whether one design of mobile-bearing (MB) total knee replacement
(TKR) has any advantage over a fixed-bearing (FB) design on long-term
fixation as measured by radiostereometry. The amount of wear underneath
the mobile bearing was also evaluated. A series of 42 knees was randomised
to MB or FB tibial components with appropriate polyethylene inserts
and followed for between ten and 12 years, or until the death of
the patient. The polyethylene in the MB group was superior in that
it was gamma-irradiated in inert gas and was calcium-stearate free;
the polyethylene in the FB group was gamma-irradiated in air and
contained calcium stearate. In theory this should be advantageous
to the wear rate of the MB group. At final follow-up the overall
mean migration was 0.75 mm ( For the MB group, the mean linear wear rate on the under-surface
was 0.026 mm/year (
Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene. The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R2 = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart. This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee.
Wear of polyethylene inserts plays an important role in failure
of total knee replacement and can be monitored Before revision, the minimum joint space width values and their
locations on the insert were measured in 15 fully weight-bearing
radiographs. These measurements were compared with the actual minimum
thickness values and locations of the retrieved tibial inserts after
revision. Introduction
Method
Miller-Galante II total knee arthroplasty (MG II TKA) was performed on 32 knees in 30 patients. On both the femoral and tibial components, the fibre-metal area was plasma-sprayed with hydroxyapatite-tricalcium phosphate (HA-TCP). The clinical and radiographic outcome was evaluated. A mean pre-operative knee score of 26.0 ± 18.6 (SD) increased to 97.5 ± 3.5 and a mean pre-operative functional score of 21.7 ± 15.0 (SD) increased to 83.4 ± 12.4 at follow-up of seven years. Clear zones were common around the components at one month post-operatively but had completely disappeared after six months. An autopsy of a patient who underwent MG II TKA with HA-TCP two years previously, showed osteogenesis in all parts of the fibre-metal, and bone tissue comprised 77.7% of the interface. This coated prosthesis has good early fixation which is maintained at seven years with good clinical and radiographic outcomes.
We carried out a prospective study to assess the clinical outcome, complications and survival of the NexGen Legacy posterior-stabilised-Flex total knee replacement (TKR) in a consecutive series of 278 knees between May 2003 and February 2005. Mean follow-up for 259 TKRs (98.2%) was 3.8 years (3.0 to 4.8). Annual follow-up showed improvement in the Knee Society scores (paired This relatively large study indicates that the legacy posterior stabilised-Flex design provides excellent short-term outcome but warrants ongoing evaluation to confirm the long-term durability and functioning of the implant.
We report the incidence of osteolysis in the femur, tibia, and patella of 44 consecutive patients (60 knees) who were followed for more than seven years after cementless knee arthroplasty with a Porous-Coated Anatomic prosthesis. The average age of the patients was 56.5 years (17 to 73); the operative diagnosis was osteoarthritis (33 knees), rheumatoid arthritis (17), tuberculous arthritis (7) and post-traumatic arthritis (3). All patellae were resurfaced. No femoral or tibial component was loose at the final follow-up examination. Thirty patellar components were loose of which six had been revised. Radiographs revealed osteolysis in 90% of the tibial plateaux and in 80% of the 30 intact patellar prostheses. No osteolysis was seen around any femoral component. In 50 knees (83%) the average wear of the polyethylene liner was 2.5 mm in the medial compartment and 1.7 mm in the lateral compartment. Four of 60 knees (6.7%) were revised for complete wear of the polyethylene liner of the tibial component. Fixation of the tibial and patellar components without cement fails to seal the interface between bone and prosthesis and allows the migration of polyethylene particulate debris which causes osteolysis.
We have reviewed 105 Liverpool Mark II knee replacements in 71 patients to assess survival and long-term results. Assessment was both clinical and radiological, using a modification of the British Orthopaedic Association knee function assessment chart, and analysis was by the survivorship method as advocated by Tew and Waugh. The follow-up period was between 13 and 113 months with 42 prostheses being in situ for over six years. Eight knees (7.6%) have been revised or arthrodesed because of infection or loosening, giving a cumulative success rate of 89% after seven years based on prosthesis survival alone. Of the remaining knees, 71.1% were either free of pain or caused only minimal pain. The cumulative success rate as judged by the stricter criteria of the prosthesis being in situ and causing little or no pain suggested a 50% survival between 73 and 96 months. Most patients (77.3%) were enthusiastic or satisfied with their results. Complications included deep wound infection (8.6%), and loosening which needed further surgery (11.4%).
We have reviewed 100 Attenborough total knee replacements in eighty-two patients with a follow-up of one to four years and conclude that this prosthesis has a valuable place in the surgical management of patients suffering from rheumatoid arthritis and osteoarthritis with severe involvement of the knee. In 85 per cent of these knees a good result was obtained with relief from pain, and in 77 per cent a useful range of movement with a stable knee. Only two patients with loosening and three with deep infection were seen in this series.
This is a preliminary report of the results of knee joint replacements using the Liverpool Mark II knee joint system which consists of a bicondylar prosthesis and a set of stereotactic instruments. The prosthesis has been developed from Gunston's concept, and the special instruments ensure its accurate insertion through meniscectomy-type incisions placed on either side of the patella. Particular features of the prosthesis are near-normal articulation, and the simplicity of the operation. Sixty-two knee replacements were implanted in forty-two patients between the spring of 1974 and January 1977. After the operation fifty-six knees were painless and four others produced only slight pain. Full extension was obtained in fifty-eight knees, and none showed a valgus or varus deformity. Collateral laxity was absent in all knees. There were two failures. These early results are most encouraging.
Cadaveric knees replaced with the Geomedic, ICLH, Marmor and Total Condylar prostheses were tested in axial compression, in rotation and in hyperextension in order to observe the strength of fixation of the tibial components. In axial compression the strengths at failure varied widely, both with any one prosthesis and between prostheses. This is attributed largely to the strength of the cancellous bone of the tibia, which was measured in each case and also varied widely. Three natural knees failed at loads of 7300, 7600 and 8300 newtons respectively, whereas the strengths of replaced knees ranged from 3000 to 15750 newtons. At least one example of each design failed at less than 7300 newtons, suggesting little or no reserve of strength. The strength of fixation was greater when the tibial prosthesis was large enough to rest on the whole cross-section of the tibia. In rotation the three prostheses embodying rollers in troughs were stiffer than the Marmor which had a nearly flat tibial-bearing surface. The presence or absence of the cruciate ligaments had a negligible effect on torsional stiffness. In hyperextension, knees replaced with the ICLH, Marmor and Total Condylar prostheses failed by rupture of the posterior capsule at moments of about 60 newton-metres, compared with about 100 for natural knees. With the Marmor prosthesis the anterior cruciate ligament was avulsed at about 20 newton-metres compared with about 75 in natural knees, suggesting that in this respect the retention of the cruciate ligaments contributes little. None of the four knees tested after inserting a Geomedic prosthesis showed strengths as high as those replaced with the other three designs.
Failure of massive knee endoprostheses implanted for malignant tumours of the distal femur in children presents a difficult problem. We present the results of rotationplasty undertaken under these circumstances in four boys. They had been treated initially at a mean age of 9.5 years for a stage-IIB malignant tumour of the distal femur by resection and implantation of a massive knee endoprosthesis. After a mean period of eight years and a mean of four operative procedures, there was failure of the endoprosthesis because of aseptic loosening in two and infection in two. Function was poor with a mean Musculoskeletal Tumor Society score of 7.5/30, and considerable associated psychological problems. At a mean follow-up of 4.5 years after rotationplasty there was excellent function with a mean score of 27.5/30 and resolution of the psychological problems.
The tibial components in 143 patients with total knee replacements performed before 1988 were assessed for micromotion using roentgen stereophotogrammetric analysis (RSA) over a period of 13 years. The fixation of the prostheses remained clinically sound in all cases, although revision had been required for other reasons in seven. In a second group taken from all cases with RSA available on our full database to 1990, 15 tibial components had been followed by RSA from the insertion until, 1 to 11 years after the initial arthroplasty, they were revised for mechanical loosening of the tibial component; 12 of these comprised all the loosenings in the base group, thus making a total of 155 consecutive cases, while an additional three were inserted after the base material had been compiled. The mean migration in the first group was about 1 mm at one year, but subsequent migration was slower, reaching a mean of about 1.5 mm after ten years. About one-third migrated continuously throughout follow-up, while two-thirds ceased to migrate after one to two years. In the revision group, 14 components had migrated continuously and at one year significantly more than those in the first group. One revision case lacked the crucial one-year follow-up and could not be classified. These findings suggest that mechanical loosening begins early in the postoperative period. Clinical symptoms which necessitate revision, seen at this stage in 20% of abnormally migrating tibial components, may not appear until up to ten years after the operation.(ABSTRACT TRUNCATED AT 250 WORDS)
The St Georg modular knee replacement has been studied in 59 cases with an observation period ranging from 28 to 73 months. In 47% of the knees both compartments were replaced; five of the six poor results were in this group. These were revised to a hinge arthroplasty or, in one case, to an arthrodesis. Other complications were few and insignificant. There were no infections. We concluded that unicompartmental knee arthroplasty can be recommended when joint involvement is localised to one compartment only. With more generalised joint disease we prefer a semiconstrained total condylar prosthesis.
Exposure of the prosthesis is one of the problems encountered after arthroplasty of the knee. Eight patients with nine exposed prostheses have been treated by using a gastrocnemius myocutaneous flap to provide cover. This preserved the prosthesis and closed the joint in six of the eight patients.l The factors leading to breakdown of the wound are discussed and the anatomy of the gastrocnemius myocutaneous flap and the operative procedure are described.
Eighty-three Guepar valgus-hinge prostheses and 30 prostheses with collinear femoral and the tibial components were inserted in 97 patients at Vancouver General Hospital between March 1975 and May 1978. One hundred and nine arthroplasties were reviewed between January 1979 and April 1980, after an average follow-up of 19 months. It was found that the amount of bone resected made arthrodesis difficult, that there was a very high incidence of patellar instability and that the disintegration of the Silastic bumper created a severe chemical synovitis. The results were excellent in 32, good in 23, fair in 8, and poor in 30. Patellar subluxation occurred in 28 per cent of the valgus prostheses, and in 10 per cent with the straight prostheses; it did not occur with the straight prostheses in rheumatoid joints. The chemical synovitis led to failure with devastating complications necessitating further operations in some knees.
Aims. Rotating-hinge
Aims. The integrity of the soft tissue envelope is crucial for successful treatment of infected total knee arthroplasty (TKA). The purpose of this study was to evaluate the rate of limb salvage, infection control, and clinical function following microvascular free flap coverage for salvage of the infected TKA. Methods. We retrospectively reviewed 23 microvascular free tissue transfers for management of soft tissue defects in infected TKA. There were 16 men and seven women with a mean age of 61.2 years (39 to 81). The median number of procedures performed prior to soft tissue coverage was five (2 to 9) and all patients had failed at least one two-stage reimplantation procedure. Clinical outcomes were measured using the Knee Society Scoring system for pain and function. Results. In all, one patient was lost to follow-up prior to 12 months. The remaining 22 patients were followed for a mean of 46 months (12 to 92). At latest follow-up, four patients (18%) had undergone amputation for failure of treatment and persistent infection. For the other 18 patients, 11 patients (50%) had maintained a
Aims. Periprosthetic joint infection (PJI) is a challenging complication of any arthroplasty procedure. We reviewed our use of static antibiotic-loaded cement spacers (ABLCSs) for staged management of PJI where segmental bone loss, ligamentous instability, or soft-tissue defects necessitate a static construct. We reviewed factors contributing to their failure and techniques to avoid these complications when using ABLCSs in this context. Methods. A retrospective analysis was conducted of 94 patients undergoing first-stage revision of an infected
Aims. Our objective is to describe our early and mid-term results with the use of a new simple primary
Aims. The use of vitamin E-infused highly crosslinked polyethylene (HXLPE) in total
Aims. Joint registries classify all further arthroplasty procedures to a knee with an existing partial arthroplasty as revision surgery, regardless of the actual procedure performed. Relatively minor procedures, including bearing exchanges, are classified in the same way as major operations requiring augments and stems. A new classification system is proposed to acknowledge and describe the detail of these procedures, which has implications for risk, recovery, and health economics. Methods. Classification categories were proposed by a surgical consensus group, then ranked by patients, according to perceived invasiveness and implications for recovery. In round one, 26 revision cases were classified by the consensus group. Results were tested for inter-rater reliability. In round two, four additional cases were added for clarity. Round three repeated the survey one month later, subject to inter- and intrarater reliability testing. In round four, five additional expert partial knee arthroplasty surgeons were asked to classify the 30 cases according to the proposed revision partial knee classification (RPKC) system. Results. Four classes were proposed: PR1, where no bone-implant interfaces are affected; PR2, where surgery does not include conversion to total knee arthroplasty, for example, a second partial arthroplasty to a native compartment; PR3, when a standard primary total
Aims. In the last decade, interest in partial knee arthroplasties and bicruciate retaining total knee arthroplasties has increased. In addition, patient-related outcomes and functional results such as range of movement and ambulation may be more promising with less invasive procedures such as bicompartmental arthroplasty (BCA). The purpose of this study is to evaluate clinical and radiological outcomes after a third-generation patellofemoral arthroplasty (PFA) combined with a medial or lateral unicompartmental knee arthroplasty (UKA) at mid- to long-term follow-up. Methods. A total of 57 procedures were performed. In 45 cases, a PFA was associated with a medial UKA and, in 12, with a lateral UKA. Patients were followed with validated patient-reported outcome measures (Oxford Knee Score (OKS), EuroQol five-dimension questionnaire (EQ-5D), EuroQoL Visual Analogue Scale (EQ-VAS)), the Knee Society Score (KSS), the Forgotten Joint Score (FJS), and radiological analysis. Results. The mean follow-up was nine years (6 to 13). All scores significantly improved from preoperatively to final follow-up (mean and SD): OKS from 23.2 (8.1) to 42.5 (3.5), EQ-5D from 0.44 (0.25) to 0.815 (0.1), EQ-VAS from 46.7 (24.9) to 89.1 (9.8), KSS (Knee) from 51.4 (8.5) to 94.4 (4.2), and KSS (Function) from 48.7 (5.5) to 88.8 (5.2). The mean FJS at final follow-up was 79.2 (4.2). All failures involved the medial UKA + PFA group. Overall, survival rate was 91.5% for all the combined implants at ten years with 95% confidence intervals and 22 knees at risk. Conclusion. Excellent clinical and radiological outcomes were achieved after a third-generation PFA combined with a medial or lateral UKA. BCA with unlinked partial
Several previously identified patient-, injury-, and treatment-related factors are associated with the development of nonunion in distal femur fractures. However, the predictive value of these factors is not well defined. We aimed to assess the predictive ability of previously identified risk factors in the development of nonunion leading to secondary surgery in distal femur fractures. We conducted a retrospective cohort study of adult patients with traumatic distal femur fracture treated with lateral locking plate between 2009 and 2018. The patients who underwent secondary surgery due to fracture healing problem or plate failure were considered having nonunion. Background knowledge of risk factors of distal femur fracture nonunion based on previous literature was used to form an initial set of variables. A logistic regression model was used with previously identified patient- and injury-related variables (age, sex, BMI, diabetes, smoking, periprosthetic fracture, open fracture, trauma energy, fracture zone length, fracture comminution, medial side comminution) in the first analysis and with treatment-related variables (different surgeon-controlled factors, e.g. plate length, screw placement, and proximal fixation) in the second analysis to predict the nonunion leading to secondary surgery in distal femur fractures.Aims
Methods
Aims. The importance of accurate identification and reporting of surgical
site infection (SSI) is well recognised but poorly defined. Public
Health England (PHE) mandated collection of orthopaedic SSI data
in 2004. Data submission is required in one of four categories (hip
prosthesis,
This study aimed to evaluate if total knee arthroplasty (TKA) femoral components aligned in either mechanical alignment (MA) or kinematic alignment (KA) are more biomimetic concerning trochlear sulcus orientation and restoration of trochlear height. Bone surfaces from 1,012 CT scans of non-arthritic femora were segmented using a modelling and analytics system. TKA femoral components (Triathlon; Stryker) were virtually implanted in both MA and KA. Trochlear sulcus orientation was assessed by measuring the distal trochlear sulcus angle (DTSA) in native femora and in KA and MA prosthetic femoral components. Trochlear anatomy restoration was evaluated by measuring the differences in medial, lateral, and sulcus trochlear height between native femora and KA and MA prosthetic femoral components.Aims
Methods
It is becoming increasingly common for a patient
to have ipsilateral hip and knee replacements. The inter-prosthetic (IP)
distance, the distance between the tips of hip and
Achievement of accurate microbiological diagnosis prior to revision is key to reducing the high rates of persistent infection after revision knee surgery. The effect of change in the microorganism between the first- and second-stage revision of total knee arthroplasty for periprosthetic joint infection (PJI) on the success of management is not clear. A two-centre retrospective cohort study was conducted to review the outcome of patients who have undergone two-stage revision for treatment of knee arthroplasty PJI, focusing specifically on isolated micro-organisms at both the first- and second-stage procedure. Patient demographics, medical, and orthopaedic history data, including postoperative outcomes and subsequent treatment, were obtained from the electronic records and medical notes.Aims
Methods
Sagittal plane imbalance (SPI), or asymmetry between extension and flexion gaps, is an important issue in total knee arthroplasty (TKA). The purpose of this study was to compare SPI between kinematic alignment (KA), mechanical alignment (MA), and functional alignment (FA) strategies. In 137 robotic-assisted TKAs, extension and flexion stressed gap laxities and bone resections were measured. The primary outcome was the proportion and magnitude of medial and lateral SPI (gap differential > 2.0 mm) for KA, MA, and FA. Secondary outcomes were the proportion of knees with severe (> 4.0 mm) SPI, and resection thicknesses for each technique, with KA as reference.Aims
Methods
To evaluate mid-to long-term patient-reported outcome measures (PROMs) of endoprosthetic reconstruction after resection of malignant tumours arising around the knee, and to investigate the risk factors for unfavourable PROMs. The medical records of 75 patients who underwent surgery between 2000 and 2020 were retrospectively reviewed, and 44 patients who were alive and available for follow-up (at a mean of 9.7 years postoperatively) were included in the study. Leg length discrepancy was measured on whole-leg radiographs, and functional assessment was performed with PROMs (Toronto Extremity Salvage Score (TESS) and Comprehensive Outcome Measure for Musculoskeletal Oncology Lower Extremity (COMMON-LE)) with two different aspects. The thresholds for unfavourable PROMs were determined using anchor questions regarding satisfaction, and the risk factors for unfavourable PROMs were investigated.Aims
Methods
The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs). A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with radiostereometric analysis (RSA) at three, six, 12, 24, and 60 months postoperatively. PROMS were collected preoperatively and at all follow-up timepoints.Aims
Methods
Bacteriophages infect, replicate inside bacteria, and are released from the host through lysis. Here, we evaluate the effects of repetitive doses of the For the haematogenous infection, Aims
Methods
This study aimed to investigate the optimal sagittal positioning of the uncemented femoral component in total knee arthroplasty to minimize the risk of aseptic loosening and periprosthetic fracture. Ten different sagittal placements of the femoral component, ranging from -5 mm (causing anterior notch) to +4 mm (causing anterior gap), were analyzed using finite element analysis. Both gait and squat loading conditions were simulated, and Von Mises stress and interface micromotion were evaluated to assess fracture and loosening risk.Aims
Methods
Mid-level constraint designs for total knee arthroplasty (TKA) are intended to reduce coronal plane laxity. Our aims were to compare kinematics and ligament forces of the Zimmer Biomet Persona posterior-stabilized (PS) and mid-level designs in the coronal, sagittal, and axial planes under loads simulating clinical exams of the knee in a cadaver model. We performed TKA on eight cadaveric knees and loaded them using a robotic manipulator. We tested both PS and mid-level designs under loads simulating clinical exams via applied varus and valgus moments, internal-external (IE) rotation moments, and anteroposterior forces at 0°, 30°, and 90° of flexion. We measured the resulting tibiofemoral angulations and translations. We also quantified the forces carried by the medial and lateral collateral ligaments (MCL/LCL) via serial sectioning of these structures and use of the principle of superposition.Aims
Methods
This study compared patient-reported outcomes of three total knee arthroplasty (TKA) designs from one manufacturer: one cruciate-retaining (CR) design, and two cruciate-sacrificing designs, anterior-stabilized (AS) and posterior-stabilized (PS). Patients scheduled for primary TKA were included in a single-centre, prospective, three-armed, blinded randomized trial (n = 216; 72 per group). After intraoperative confirmation of posterior cruciate ligament (PCL) integrity, patients were randomly allocated to receive a CR, AS, or PS design from the same TKA system. Insertion of an AS or PS design required PCL resection. The primary outcome was the mean score of all five subscales of the Knee injury and Osteoarthritis Outcome Score (KOOS) at two-year follow-up. Secondary outcomes included all KOOS subscales, Oxford Knee Score, EuroQol five-dimension health questionnaire, EuroQol visual analogue scale, range of motion (ROM), and willingness to undergo the operation again. Patient satisfaction was also assessed.Aims
Methods
While mechanical alignment (MA) is the traditional technique in total knee arthroplasty (TKA), its potential for altering constitutional alignment remains poorly understood. This study aimed to quantify unintentional changes to constitutional coronal alignment and joint line obliquity (JLO) resulting from MA. A retrospective cohort study was undertaken of 700 primary MA TKAs (643 patients) performed between 2014 and 2017. Lateral distal femoral and medial proximal tibial angles were measured pre- and postoperatively to calculate the arithmetic hip-knee-ankle angle (aHKA), JLO, and Coronal Plane Alignment of the Knee (CPAK) phenotypes. The primary outcome was the magnitude and direction of aHKA, JLO, and CPAK alterations.Aims
Methods
Objectives. The objective of this study was to compare the early migration
characteristics and functional outcome of the Triathlon cemented
knee prosthesis with its predecessor, the Duracon cemented knee
prosthesis (both Stryker). Methods. A total 60 patients were prospectively randomised and tibial
component migration was measured by radiostereometric analysis (RSA)
at three months, one year and two years; clinical outcome was measured
by the American Knee Society score and the Knee Osteoarthritis and
Injury Outcome Score. Results. There were no statistically significant differences in rotation
or translation around or along the three coordinal axes, or in the
maximum total point motion (MTPM) during the two-year follow-up. Conclusions. The Triathlon cemented
We report the results of revision total knee
replacement (TKR) in 26 patients with major metaphyseal osteolytic defects
using 29 trabecular metal cones in conjunction with a rotating hinged
total
To investigate the efficacy of ethylenediaminetetraacetic acid-normal saline (EDTA-NS) in dispersing biofilms and reducing bacterial infections. EDTA-NS solutions were irrigated at different durations (1, 5, 10, and 30 minutes) and concentrations (1, 2, 5, 10, and 50 mM) to disrupt Aims
Methods
The purpose of this study was to compare the clinical outcomes, mortalities, implant survival rates, and complications of total knee arthroplasty (TKA) in patients with or without hepatitis B virus (HBV) infection over at least ten years of follow-up. From January 2008 to December 2010, 266 TKAs were performed in 169 patients with HBV (HBV group). A total of 169 propensity score–matched patients without HBV were chosen for the control group in a one-to-one ratio. Then, the clinical outcomes, mortalities, implant survival rates, and complications of TKA in the two groups were compared. The mean follow-up periods were 11.7 years (10.5 to 13.4) in the HBV group and 11.8 years (11.5 to 12.4) in the control group.Aims
Methods
To fully quantify the effect of posterior tibial slope (PTS) angles on joint kinematics and contact mechanics of intact and anterior cruciate ligament-deficient (ACLD) knees during the gait cycle. In this controlled laboratory study, we developed an original multiscale subject-specific finite element musculoskeletal framework model and integrated it with the tibiofemoral and patellofemoral joints with high-fidelity joint motion representations, to investigate the effects of 2.5° increases in PTS angles on joint dynamics and contact mechanics during the gait cycle.Aims
Methods
This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant. A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life.Aims
Methods
There has been a substantial increase in the
number of hip and
We have designed a prospective study to evaluate
the usefulness of prolonged incubation of cultures from sonicated
orthopaedic implants. During the study period 124 implants from
113 patients were processed (22 osteosynthetic implants, 46 hip
prostheses, 54
We compared the risks of re-revision and mortality between two-stage revision surgery and single-stage revision surgery among patients with infected primary knee arthroplasty. Patients with a periprosthetic joint infection (PJI) of their primary knee arthroplasty, initially revised with a single-stage or a two-stage procedure in England and Wales between 2003 and 2014, were identified from the National Joint Registry. We used Poisson regression with restricted cubic splines to compute hazard ratios (HR) at different postoperative periods. The total number of revisions and re-revisions undergone by patients was compared between the two strategies.Aims
Methods
Wear of polyethylene is associated with aseptic loosening of orthopaedic implants and has been observed in hip and
Total knee arthroplasty (TKA) is a common and safe orthopaedic procedure. Zimmer Biomet's NexGen is the second most popular brand of implant used in the UK. The primary cause of revision after the first year is aseptic loosening. We present our experience of using this implant, with significant concerns around its performance with regards early aseptic loosening of the tibial component. A retrospective, single-surgeon review was carried out of all of the NexGen Legacy Posterior Stabilized (LPS) TKAs performed in this institute. The specific model used for the index procedures was the NexGen Complete Knee System (Legacy Knee-Posterior Stabilized LPS-Flex Articular Surface, LPS-Flex Femoral Component Option, and Stemmed Nonaugmentable Tibial Component Option).Aims
Methods
We aimed to evaluate the utility of 68Ga-citrate positron emission tomography (PET)/CT in the differentiation of periprosthetic joint infection (PJI) and aseptic loosening (AL), and compare it with 99mTc-methylene bisphosphonates (99mTc-MDP) bone scan. We studied 39 patients with suspected PJI or AL. These patients underwent 68Ga-citrate PET/CT, 99mTc-MDP three-phase bone scan and single-photon emission CT (SPECT)/CT. PET/CT was performed at ten minutes and 60 minutes after injection, respectively. Images were evaluated by three nuclear medicine doctors based on: 1) visual analysis of the three methods based on tracer uptake model, and PET images attenuation-corrected with CT and those not attenuation-corrected with CT were analyzed, respectively; and 2) semi-quantitative analysis of PET/CT: maximum standardized uptake value (SUVmax) of lesions, SUVmax of the lesion/SUVmean of the normal bone, and SUVmax of the lesion/SUVmean of the normal muscle. The final diagnosis was based on the clinical and intraoperative findings, and histopathological and microbiological examinations.Aims
Methods
We describe the mid-term results of a prospective study of total knee replacement in severe valgus knees using an osteotomy of the lateral femoral condyle and computer navigation. There were 15 knees with a mean valgus deformity of 21° (17° to 27°) and a mean follow-up of 28 months (24 to 60). A cemented, non-constrained fixed bearing, posterior-cruciate-retaining
We have prospectively studied the outcome of infections associated with implants which were retained and treated using a standardised antimicrobial protocol. Over a period of four years, we studied 24 consecutive patients who had symptoms of infection for less than one year, a stable implant, no sinus tract and a known pathogen which was susceptible to recommended antimicrobial agents. The infections involved hip prostheses (14),
Recently, gender-specific designs of total knee replacement have been developed to accommodate anatomical differences between males and females. We examined a group of male and female distal femora matched for age and height, to determine if there was a difference in the aspect ratio (mediolateral distance versus anteroposterior distance) and the height of the anterior flange between the genders. The Hamann-Todd Collection provided 1207 skeletally mature cadaver femora. The femoral length, the anteroposterior height, height of the lateral and medial flanges and the mediolateral width were measured in all the specimens. The mechanical axis of the femur, the cut articular width and the aspect ratio were assessed. Statistical analysis of the effect of gender upon the aspect ratio and the lateral and medial flanges was undertaken, controlling for age, height and race. The mean aspect ratio of male femora was 1.21 (. sd. 0.07) and of female femora it was 1.16 (. sd. 0.06) (p <
0.001). There was no significant difference between male and female specimens in the mean size of the lateral flange (6.57 mm (. sd. 2.57) and 7.02 mm (. sd. 2.36), respectively; p = 0.099) or of the medial flange (3.03 mm (. sd. 2.47) and 3.56 mm (. sd. 2.32), respectively; p = 0.67). Future work in the design of
In two years we treated four women with ununited stress fractures of their proximal tibial diaphyses. They all had arthritis and valgus deformity. The stress fractures had been treated elsewhere by non-operative means in three patients and by open reduction and internal fixation in one, but had failed to unite. After treatment with a modular total
The main objective of this study is to analyze the penetration of bone cement in four different full cementation techniques of the tibial tray. In order to determine the best tibial tray cementation technique, we applied cement to 40 cryopreserved donor tibiae by four different techniques: 1) double-layer cementation of the tibial component and tibial bone with bone restrictor; 2) metallic cementation of the tibial component without bone restrictor; 3) bone cementation of the tibia with bone restrictor; and 4) superficial bone cementation of the tibia and metallic keel cementation of the tibial component without bone restrictor. We performed CT exams of all 40 subjects, and measured cement layer thickness at both levels of the resected surface of the epiphysis and the endomedular metaphyseal level.Aims
Methods
There is conflicting evidence on the safety of intra-articular injections of hyaluronic acid (HA) or corticosteroids (CSs) before total knee arthroplasty (TKA). We performed a meta-analysis of the relationship between intra-articular injections and subsequent infection rates after TKA. We searched PubMed, EMBASE, and the Cochrane Library for cohort studies that assessed the effect of preoperative injection of drugs into the joint cavity on the infection rate after TKA. The outcomes analyzed included the total infection rate, as well as those for different preoperative injection time periods and different drugs.Aims
Methods
Single-stage revision total knee arthroplasty (rTKA) is gaining popularity in treating chronic periprosthetic joint infections (PJIs). We have introduced this approach to our clinical practice and sought to evaluate rates of reinfection and re-revision, along with predictors of failure of both single- and two-stage rTKA for chronic PJI. A retrospective comparative cohort study of all rTKAs for chronic PJI between 1 April 2003 and 31 December 2018 was undertaken using prospective databases. Patients with acute infections were excluded; rTKAs were classified as single-stage, stage 1, or stage 2 of two-stage revision. The primary outcome measure was failure to eradicate or recurrent infection. Variables evaluated for failure by regression analysis included age, BMI, American Society of Anesthesiologists grade, infecting organisms, and the presence of a sinus. Patient survivorship was also compared between the groups.Aims
Methods
The Blauth prosthesis is a hinged total knee replacement. We have evaluated 422 consecutive primary total knee arthroplasties using this design in 330 patients with a maximum follow-up of 20 years (mean 6). The mean age of the patients at the time of operation was 70 years (29 to 87). Using endpoints of infection, removal because of aseptic loosening, removal for any cause, and the worst case as definitions of failure, the cumulative rates of survival at 20 years were 93.6%, 96.0%, 94.4% and 86.8%, respectively. The cumulative rate of survival of the patients themselves was 14.4% (24% in patients with osteoarthritis and 9% in these with rheumatoid arthritis) after 20 years. Our survivorship analysis shows that hinged
The aim of this study was to compare any differences in the primary outcome (biphasic flexion knee moment during gait) of robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) at one year post-surgery. A total of 76 patients (34 bi-UKA and 42 TKA patients) were analyzed in a prospective, single-centre, randomized controlled trial. Flat ground shod gait analysis was performed preoperatively and one year postoperatively. Knee flexion moment was calculated from motion capture markers and force plates. The same setup determined proprioception outcomes during a joint position sense test and one-leg standing. Surgery allocation, surgeon, and secondary outcomes were analyzed for prediction of the primary outcome from a binary regression model.Aims
Methods
Periprosthetic joint infections (PJIs) are rare, but represent a great burden for the patient. In addition, the incidence of methicillin-resistant For this purpose, sterilized steel implants were implanted into the femur of 77 rats. The metal devices were inoculated with suspensions of two different MRSA strains. The animals were divided into groups and treated with vancomycin, linezolid, cotrimoxazole, or rifampin as monotherapy, or with combination of antibiotics over a period of 14 days. After a two-day antibiotic-free interval, the implant was explanted, and bone, muscle, and periarticular tissue were microbiologically analyzed.Aims
Methods
Periprosthetic joint infection (PJI) occurs in approximately 1% to 2% of total knee arthroplasties (TKA) presenting multiple challenges, such as difficulty in diagnosis, technical complexity, and financial costs. Two-stage exchange is the gold standard for treating PJI but emerging evidence suggests 'two-in-one' single-stage revision as an alternative, delivering comparable outcomes, reduced morbidity, and cost-effectiveness. This study investigates five-year results of modified single-stage revision for treatment of PJI following TKA with bone loss. Patients were identified from prospective data on all TKA patients with PJI following the primary procedure. Inclusion criteria were: revision for PJI with bone loss requiring reconstruction, and a minimum five years’ follow-up. Patients were followed up for recurrent infection and assessment of function. Tools used to assess function were Oxford Knee Score (OKS) and American Knee Society Score (AKSS).Aims
Methods
Simultaneous bilateral total knee arthroplasty (TKA) has been used due to its financial advantages, overall resource usage, and convenience for the patient. The training model where a trainee performs the first TKA, followed by the trainer surgeon performing the second TKA, is a unique model to our institution. This study aims to analyze the functional and clinical outcomes of bilateral simultaneous TKA when performed by a trainee or a supervising surgeon, and also to assess these outcomes based on which side was done by the trainee or by the surgeon. This was a retrospective cohort study of all simultaneous bilateral TKAs performed by a single surgeon in an academic institution between May 2003 and November 2017. Exclusion criteria were the use of partial knee arthroplasty procedures, staged bilateral procedures, and procedures not performed by the senior author on one side and the trainee on another. Primary clinical outcomes of interest included revision and re-revision. Primary functional outcomes included the Oxford Knee Score (OKS) and patient satisfaction scores.Aims
Methods
One of the most controversial issues in total knee replacement is whether or not to resurface the patella. In order to determine the effects of different designs of femoral component on the conformity of the patellofemoral joint, five different
Uncemented mobile bearing designs in medial unicompartmental knee arthroplasty (UKA) have seen an increase over the last decade. However, there are a lack of large-scale studies comparing survivorship of these specific designs to commonly used cemented mobile and fixed bearing designs. The aim of this study was to evaluate the survivorship of these designs. A total of 21,610 medial UKAs from 2007 to 2018 were selected from the Dutch Arthroplasty Register. Multivariate Cox regression analyses were used to compare uncemented mobile bearings with cemented mobile and fixed bearings. Adjustments were made for patient and surgical factors, with their interactions being considered. Reasons and type of revision in the first two years after surgery were assessed.Aims
Methods
Metal allergy in knee arthroplasty patients is a controversial topic. We aimed to conduct a scoping review to clarify the management of metal allergy in primary and revision total knee arthroplasty (TKA). Studies were identified by searching electronic databases: Cochrane Central Register of Controlled Trials, Ovid MEDLINE, and Embase, from their inception to November 2020, for studies evaluating TKA patients with metal hypersensitivity/allergy. All studies reporting on diagnosing or managing metal hypersensitivity in TKA were included. Data were extracted and summarized based on study design, study population, interventions and outcomes. A practical guide is then formulated based on the available evidence.Aims
Methods
Reverse total shoulder arthroplasty (RTSA) using trabecular metal (TM)-backed glenoid implants has been introduced with the aim to increase implant survival. Only short-term reports on the outcomes of TM-RTSA have been published to date. We aim to present the seven-year survival of TM-backed glenoid implants along with minimum five-year clinical and radiological outcomes. All consecutive elective RTSAs performed at a single centre between November 2008 and October 2014 were reviewed. Patients who had primary TM-RTSA for rotator cuff arthropathy and osteoarthritis with deficient cuff were included. A total of 190 shoulders in 168 patients (41 male, 127 female) were identified for inclusion at a mean of 7.27 years (SD 1.4) from surgery. The primary outcome was survival of the implant with all-cause revision and aseptic glenoid loosening as endpoints. Secondary outcomes were clinical, radiological, and patient-related outcomes with a five-year minimum follow-up.Aims
Methods
Infection complicating primary total knee arthroplasty (TKA) is a common reason for revision surgery, hospital readmission, patient morbidity, and mortality. Increasing incidence of methicillin-resistant A retrospective review of 1,060 patients who underwent TKA between May 2016 to July 2020 was performed. There were 572 patients in the IV group and 488 in the IO group, with minimal 30 days of follow-up. Patients were followed up at regularly scheduled intervals (two, six, and 12 weeks). No differences between groups for age, sex, BMI, or baseline comorbidities existed. The IV group received an IV dose of 15 mg/kg vancomycin given over an hour preceding skin incision. The IO group received a 500 mg dose of vancomycin mixed in 150 ml of normal saline, injected into proximal tibia after tourniquet inflation, before skin incision. All patients received an additional dose of first generation cephalosporin. Evaluation included preoperative and postoperative serum creatinine values, tourniquet time, and adverse reactions attributable to vancomycin.Aims
Methods
The prevalence of ipsilateral total hip arthroplasty (THA) and total knee arthroplasty (TKA) is rising in concert with life expectancy, putting more patients at risk for interprosthetic femur fractures (IPFFs). Our study aimed to assess treatment methodologies, implant survivorship, and IPFF clinical outcomes. A total of 76 patients treated for an IPFF from February 1985 to April 2018 were reviewed. Prior to fracture, at the hip/knee sites respectively, 46 femora had primary/primary, 21 had revision/primary, three had primary/revision, and six had revision/revision components. Mean age and BMI were 74 years (33 to 99) and 30 kg/m2 (21 to 46), respectively. Mean follow-up after fracture treatment was seven years (2 to 24).Aims
Methods
Modern total knee arthroplasty (TKA) prostheses are designed to restore near normal kinematics including high flexion. Kneeling is a high flexion, kinematically demanding activity after TKA. The debate about design choice has not yet been informed by six-degrees-of-freedom in vivo kinematics. This prospective randomized clinical trial compared kneeling kinematics in three TKA designs. In total, 68 patients were randomized to either a posterior stabilized (PS-FB), cruciate-retaining (CR-FB), or rotating platform (CR-RP) design. Of these patients, 64 completed a minimum one year follow-up. Patients completed full-flexion kneeling while being imaged using single-plane fluoroscopy. Kinematics were calculated by registering the 3D implant models onto 2D-dynamic fluoroscopic images and exported for analysis.Aims
Methods
Antibiotic resistance represents a threat to human health. It has been suggested that by 2050, antibiotic-resistant infections could cause ten million deaths each year. In orthopaedics, many patients undergoing surgery suffer from complications resulting from implant-associated infection. In these circumstances secondary surgery is usually required and chronic and/or relapsing disease may ensue. The development of effective treatments for antibiotic-resistant infections is needed. Recent evidence shows that bacteriophage (phages; viruses that infect bacteria) therapy may represent a viable and successful solution. In this review, a brief description of bone and joint infection and the nature of bacteriophages is presented, as well as a summary of our current knowledge on the use of bacteriophages in the treatment of bacterial infections. We present contemporary published in vitro and in vivo data as well as data from clinical trials, as they relate to bone and joint infections. We discuss the potential use of bacteriophage therapy in orthopaedic infections. This area of research is beginning to reveal successful results, but mostly in nonorthopaedic fields. We believe that bacteriophage therapy has potential therapeutic value for implant-associated infections in orthopaedics. Cite this article:
Debate remains whether the patella should be resurfaced during total knee replacement (TKR). For non-resurfaced TKRs, we estimated what the revision rate would have been if the patella had been resurfaced, and examined the risk of re-revision following secondary patellar resurfacing. A retrospective observational study of the National Joint Registry (NJR) was performed. All primary TKRs for osteoarthritis alone performed between 1 April 2003 and 31 December 2016 were eligible (n = 842,072). Patellar resurfacing during TKR was performed in 36% (n = 305,844). The primary outcome was all-cause revision surgery. Secondary outcomes were the number of excess all-cause revisions associated with using TKRs without (versus with) patellar resurfacing, and the risk of re-revision after secondary patellar resurfacing.Aims
Methods
Restarting planned surgery during the COVID-19 pandemic is a clinical and societal priority, but it is unknown whether it can be done safely and include high-risk or complex cases. We developed a Surgical Prioritization and Allocation Guide (SPAG). Here, we validate its effectiveness and safety in COVID-free sites. A multidisciplinary surgical prioritization committee developed the SPAG, incorporating procedural urgency, shared decision-making, patient safety, and biopsychosocial factors; and applied it to 1,142 adult patients awaiting orthopaedic surgery. Patients were stratified into four priority groups and underwent surgery at three COVID-free sites, including one with access to a high dependency unit (HDU) or intensive care unit (ICU) and specialist resources. Safety was assessed by the number of patients requiring inpatient postoperative HDU/ICU admission, contracting COVID-19 within 14 days postoperatively, and mortality within 30 days postoperatively.Aims
Methods
Total knee arthroplasty (TKA) using functional alignment aims to implant the components with minimal compromise of the soft-tissue envelope by restoring the plane and obliquity of the non-arthritic joint. The objective of this study was to determine the effect of TKA with functional alignment on mediolateral soft-tissue balance as assessed using intraoperative sensor-guided technology. This prospective study included 30 consecutive patients undergoing robotic-assisted TKA using the Stryker PS Triathlon implant with functional alignment. Intraoperative soft-tissue balance was assessed using sensor-guided technology after definitive component implantation; soft-tissue balance was defined as intercompartmental pressure difference (ICPD) of < 15 psi. Medial and lateral compartment pressures were recorded at 10°, 45°, and 90° of knee flexion. This study included 18 females (60%) and 12 males (40%) with a mean age of 65.2 years (SD 9.3). Mean preoperative hip-knee-ankle deformity was 6.3° varus (SD 2.7°).Aims
Methods
A comprehensive classification for coronal lower limb alignment with predictive capabilities for knee balance would be beneficial in total knee arthroplasty (TKA). This paper describes the Coronal Plane Alignment of the Knee (CPAK) classification and examines its utility in preoperative soft tissue balance prediction, comparing kinematic alignment (KA) to mechanical alignment (MA). A radiological analysis of 500 healthy and 500 osteoarthritic (OA) knees was used to assess the applicability of the CPAK classification. CPAK comprises nine phenotypes based on the arithmetic HKA (aHKA) that estimates constitutional limb alignment and joint line obliquity (JLO). Intraoperative balance was compared within each phenotype in a cohort of 138 computer-assisted TKAs randomized to KA or MA. Primary outcomes included descriptive analyses of healthy and OA groups per CPAK type, and comparison of balance at 10° of flexion within each type. Secondary outcomes assessed balance at 45° and 90° and bone recuts required to achieve final knee balance within each CPAK type.Aims
Methods
Patient-reported outcome measures (PROMs) are being used increasingly in total knee arthroplasty (TKA). We conducted a systematic review aimed at identifying psychometrically sound PROMs by appraising their measurement properties. Studies concerning the development and/or evaluation of the measurement properties of PROMs used in a TKA population were systematically retrieved via PubMed, Web of Science, Embase, and Scopus. Ratings for methodological quality and measurement properties were conducted according to updated COnsensus-based Standards for the selection of health Measurement INstruments (COSMIN) methodology. Of the 155 articles on 34 instruments included, nine PROMs met the minimum requirements for psychometric validation and can be recommended to use as measures of TKA outcome: Oxford Knee Score (OKS); OKS–Activity and Participation Questionnaire (OKS-APQ); 12-item short form Knee Injury and Osteoarthritis Outcome (KOOS-12); KOOS Physical function Short form (KOOS-PS); Western Ontario and McMaster Universities Arthritis Index-Total Knee Replacement function short form (WOMAC-TKR); Lower Extremity Functional Scale (LEFS); Forgotten Joint Score (FJS); Patient’s Knee Implant Performance (PKIP); and University of California Los Angeles (UCLA) activity score. The pain and function subscales in WOMAC, as well as the pain, function, and quality of life subscales in KOOS, were validated psychometrically as standalone subscales instead of as whole instruments. However, none of the included PROMs have been validated for all measurement properties. Thus, further studies are still warranted to evaluate those PROMs. Use of the other 25 scales and subscales should be tempered until further studies validate their measurement properties. Cite this article:
We have developed a new tensor for total knee replacements which is designed to assist with soft-tissue balancing throughout the full range of movement with a reduced patellofemoral joint. Using this tensor in 40 patients with osteoarthritis we compared the intra-operative joint gap in cruciate-retaining and posterior-stabilised total knee replacements at 0°, 10°, 45°, 90° and 135° of flexion, with the patella both everted and reduced. While the measurement of the joint gap with a reduced patella in posterior-stabilised knees increased from extension to flexion, it remained constant for cruciate-retaining joints throughout a full range of movement. The joint gaps at deep knee flexion were significantly smaller for both types of
We describe a patient with multiple sclerosis (MS), who developed recurrent dislocations after total knee arthroplasty. She had both knees replaced using similar mobile-bearing
The purpose of this prospective, randomised study
was to evaluate the clinical and radiological results comparing the
identical cemented or cementless NexGen total
To assess migration of the tibial component we used roentgen stereophotogrammetric analysis in 40 patients who had had a total knee arthroplasty after failure of a closing wedge osteotomy and compared them with 40 matched patients after primary total knee arthroplasty. We found no difference in migration over time or in the tendency for continuous migration between the two groups. There were no differences in alignment or position of the
The mechanisms controlling and limiting movement and serving to transmit load between the femur and the tibia are discussed. Having accounted for the transmission of all components of force and couple across the joint and noted the load-bearing role of the menisci, some principles which might guide the design of
We recovered 23 meniscal bearings from 18 failed bicompartmental Oxford
Although the use of constrained cemented arthroplasty to treat distal femoral fractures in elderly patients has some practical advantages over the use of techniques of fixation, concerns as to a high rate of loosening after implantation of these prostheses has raised doubts about their use. We evaluated the results of hinged total knee replacement in the treatment of 54 fractures in 52 patients with a mean age of 82 years (55 to 98), who were socially dependent and poorly mobile. Within the first year after implantation 22 of the 54 patients had died, six had undergone a further operation and two required a revision of the prosthesis. The subsequent rate of further surgery and revision was low. A constrained
Between 1969 and 1971 20 prototypes of the Stanmore total
Arthrodesis after the removal of a