A 70-year-old man with an uncemented metal-on-polyethylene
total
Aims. Large-diameter metal-on-metal (MoM) total hip arthroplasty (THA) has demonstrated unexpected high failure rates and pseudotumour formation. The purpose of this prospective cohort study is to report ten-year results in order to establish revision rate, prevalence of pseudotumour formation, and relation with whole blood cobalt levels. Methods. All patients were recalled according to the guidelines of the Dutch Orthopaedic Association. They underwent clinical and radiographical assessments (radiograph and CT scan) of the
Nine cases of acetabular erosion following insertion of the Monk "hard top"
We report the clinical and operative details of seven cases of fracture of the femoral stem of the Ring TiMESH cementless
The Kent hip is a distally-locked femoral stem which was developed to address severe proximal bone loss, severe bony deformity and peri-prosthetic fracture. We reviewed the results of 145 consecutive Kent hips implanted into 141 patients between 1987 and 2000. The indications for implantation were aseptic loosening (75 hips), septic loosening (two), peri-prosthetic and prosthetic fracture (37), severe bony deformity (24), and fracture through a proximal femoral metastasis (seven). The median time to full weight-bearing after surgery was two days and the mean length of follow-up was 5.1 years (2 to 15). Further revisions were required for 13 femoral stems. With removal of the stem for any reason as an end-point, the cumulative survival at five, ten and 15 years was 93%, 89% and 77%, respectively. In patients aged ≥ 70 years, the cumulative survival at 15 years was 92%, compared with 68% in those aged <
70 years. Because of these findings, we recommend the use of interlocking stems in patients aged ≥ 70 years, particularly in those with a peri-prosthetic fracture, for whom alternative methods are limited. Outcome scores and survival data, compared with other systems, indicate that the Kent hip should be used with caution in younger patients.
The aim of this study was to evaluate blood metal ion levels, leucocyte profiles, and serum cytokines in patients with a total hip arthroplasty (THA) involving modular dual-mobility components. A total of 39 patients were recruited, with clinical follow-up of up to two years. Outcome was assessed using the Harris Hip Score (HHS, the 12-Item Short-Form Health Survey (SF-12), the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and a visual analogue scale (VAS) for pain. Blood concentrations of cobalt (Co), chromium (Cr), and serum cytokines were measured. Subpopulations of leucocytes were analyzed by flow cytometry.Aims
Patients and Methods
A consecutive series of 30 total hip replacements using a hydroxyapatite (HA)-coated, modular implant (Omnifit) was followed clinically and by roentgen stereophotogrammetric analysis for two years and compared with two control groups, one of 27 cemented Charnley sockets and one of 40 cemented Charnley stems. Omnifit sockets with a central gap between the dome of the socket and the acetabular bone in the postoperative radiographs, migrated less than sockets without such gaps (p = 0.01). After adjustment for patient-related factors (age, gender and weight), no significant difference was found between the two prostheses with respect to micromotion and wear. We conclude that the early fixation of the HA-coated Omnifit prosthesis compares with that of the cemented Charnley prosthesis.
We reviewed 82 primary arthroplasties (in 71 patients) in which cementless porous-coated hip prostheses were used. The mean age of the patients at operation was 52 years (24 to 86); they were followed up for an average of 62.1 months (60 to 66). The diagnosis was avascular necrosis of the femoral head in 35%, fracture of the femoral neck in 24%, primary osteoarthritis in 16% and miscellaneous in 25%. The average preoperative Harris hip score was 56.7 points and the average postoperative score was 83.3 points. Eight hips (10%) had component loosening; four had been revised and four were awaiting revision. In 27 hips (33%) there was a radiolucent line wider than 2 mm in zones 1 and 7. In 55 hips (67%) there was calcar resorption of more than 10 mm. Twenty patients (28%) complained of thigh pain although they had no radiographic evidence of loosening of a component. Factors that may have contributed to the poor clinical and radiographic results were: 1) inadequate surface area for bone ingrowth, particularly on the lateral aspect of the upper part of the prosthesis, 2) poor initial fit of the stem in the metaphysis, which resulted in cantilever motion of the proximal part of the stem about the well-fixed distal stem, and 3) the collar of the prosthesis, which prevented it from subsiding to a naturally stable position and caused damage to the calcar.
Between January 1984 and January 1986, 131 porous-coated anatomic total hip replacements were performed in 119 consecutive patients. Of these, 100 patients (110 hips) who were alive at a minimum of 18 years after replacement were included in the study. The mean age of the patients at surgery was 48.4 years and that of surviving patients at the latest follow-up was 67.8 years. The mean duration of the clinical and radiological follow-up was 19.4 years (18 to 20). The mean Harris hip score initially improved from 55 points before to 95 points at two years after operation, but subsequently decreased to 91 points after six years, 87 points after 11.2 years, and 85 points after 19.4 years. At the final follow-up, 23 hips (18%) of the entire cohort and 21% of survivors had undergone revision because of loosening or osteolysis of the acetabular component and eight hips (6%) of the entire cohort and 7% of survivors for loosening or osteolysis of the femoral component. Only four femoral components (4%) were revised for isolated aseptic loosening without osteolysis and two (2%) for recurrent dislocation. On the basis of these long-term results, the porous-coated anatomic femoral component survived for a minimum of 18 years after operation while the acetabular component was less durable. The findings identify the principles of uncemented joint replacement which can be applied to current practice.
In a frail elderly patient a pathological fracture of the femur at the lower end of a loose femoral stem was treated by the retrograde insertion of an intramedullary nail over the tip of the prosthesis.
A prospective, randomised and independently assessed trial of the Ring UPM total hip replacement showed that the quality of the early result was better if the femoral prosthesis was cemented than if it was not. More patients with cemented prostheses were painfree at four months (58% cemented:42% uncemented) and at one year (63% cemented:50% uncemented), but at two years pain relief was equal in both groups. At two years significantly more patients with cemented prostheses could walk without support (96% cemented:62% uncemented, p = 0.01 to 0.05). There is a need for more similar trials to compare the results of contemporary designs of cemented and uncemented total hip prostheses.
We report the clinical outcome and survival of 100 total hip arthroplasties with Link RS cementless hip prostheses after a mean of 63 months (51 to 93). The average preoperative Merle d'Aubigne hip score was 9.4 points and the average score at the latest follow-up was 15.3 points. Thirteen prostheses have been revised and seven await revision. Survivorship analysis based on intention to revise showed only 68% survival at seven years (95% CI 53% to 81%). The survivals of femoral and acetabular components verified to be loose at surgery were 78% and 89% respectively at seven years. A number of other patients have pain and it seems likely that there will be substantial reduction in these survival figures in the future.
Seven hips with protrusio acetabuli which showed complete or partial resolution of the protrusion after replacement with Ring's prosthesis are reported. It is suggested that this prosthesis offers a simple and effective method of treating painful protrusion into the pelvis; it allows healing of the medial wall of the acetabulum while avoiding many of the hazards of other methods of treatment.
Sixty-five total hip arthroplastics were reinserted after sepsis around the hip, positive cultures being obtained from fifty-six. Although 65 per cent of patients still have their implant in position, only sixteen of sixty-five (25 per cent) show an excellent or good result on a Harris rating. Twenty-three of sixty-five (35 per cent) subsequently required an excision arthroplasty. The indications and contraindications for this procedure are given.
We have followed up for a period of seven to nine years 100 consecutive arthroplasties of the hip in which an entirely HA-coated implant had been used. The clinical results were excellent and bony incorporation was extensive in all components. No stem became loose or subsided but five cups were revised because of loosening after 3.8 to 5.5 years, having functioned painlessly and shown radiological ingrowth. Revision procedures because of excessive polyethylene wear have been performed on 18 hips and are planned for six more. Two eroded metal backings with worn-through polyethylene were exchanged; six hips showed metallosis without polyethylene wear-through. There were two cases of granulomatous cysts in the groin and 66 hips had osteolysis located periarticularly, in the greater trochanter or in the acetabulum.
We enrolled 98 patients (107 hips) with a mean age of 47 years (SD 8.6) into a prospective study of the Madreporic Lord THR; 34 hips had primary and 73 secondary osteoarthritis. After ten years, the survival rate using revision as the endpoint for failure was 70% (±9) for the cup and 98% (±0.3) for the stem. The combined clinical and radiological survival rates were 46% (±11) and 81% (±10), respectively. Osteoporosis due to stress-shielding was observed in the proximal femur. Hips with radiologically dense bone postoperatively showed the most pronounced bone loss. We recommend continued radiological follow-up of patients with this type of implant to allow revision to be performed before there is severe bony destruction of the pelvis.
Twenty-nine patients with thirty pseudarthroses after removal of infected total hip prostheses have been reviewed one to six years later. Six hips still had a discharging sinus. Complete removal of all cement was found to be essential for healing, and various points in the technique of its clearance have been made. In this respect radio-opaque cement was a great advantage, and lateral guttering of the femur was most effective. Compared with the situation before replacement considerable relief of pain was obtained in most patients but there was much less improvement in function.
In this meta-analysis we included 32 English-language articles published between January 1975 and June 2004 on the diagnostic performance of plain radiography, subtraction arthrography, nuclear arthrography and bone scintigraphy in detecting aseptic loosening of the femoral component, using criteria based on the Cochrane systematic review of screening and diagnostic tests. The mean sensitivity and specificity were, respectively, 82% (95% confidence interval (CI) 76 to 87) and 81% (95% CI 73 to 87) for plain radiography and 85% (95% CI 75 to 91) and 83% (95% CI 75 to 89) for nuclear arthrography. Pooled sensitivity and specificity were, respectively, 86% (95% CI 74 to 93) and 85% (95% CI 77 to 91) for subtraction arthrography and 85% (95% CI 79 to 89) and 72% (95% CI 64 to 79) for bone scintigraphy. Although the diagnostic performance of the imaging techniques was not significantly different, plain radiography and bone scintigraphy are preferred for the assessment of a femoral component because of their efficacy and lower risk of patient morbidity.
We compared and quantified the modes of failure and patterns of wear of 11 Mittelmeier and 11 Ceraver-Ostal retrieved alumina-alumina hip prostheses with reference to the corresponding clinical and radiological histories. Macroscopic wear was assessed using a three-dimensional co-ordinate measuring machine. Talysurf contacting profilometry was used to measure surface roughness on a microscopic scale and SEM to determine mechanisms of wear at the submicron level. The components were classified into one of three categories of wear: low (no visible/measurable wear), stripe (elliptical wear stripe on the heads and larger worn areas on the cups) and severe (macroscopic wear, large volumes of material lost). Overall, the volumetric wear of the alumina-alumina prostheses was substantially less than the widely used metal and ceramic-on-polyethylene combinations. By identifying and eliminating the factors which accelerate wear, it is expected that the lifetime of these devices can be further increased.
The early failure and revision of bimodular primary
total hip arthroplasty prostheses requires the identification of the
risk factors for material loss and wear at the taper junctions through
taper wear analysis. Deviations in taper geometries between revised
and pristine modular neck tapers were determined using high resolution
tactile measurements. A new algorithm was developed and validated
to allow the quantitative analysis of material loss, complementing
the standard visual inspection currently used. The algorithm was applied to a sample of 27 retrievals ( Cite this article:
Aims. There is evidence that prior lumbar fusion increases the risk of dislocation and revision after total hip arthroplasty (THA). The relationship between prior lumbar fusion and the effect of femoral head diameter on THA dislocation has not been investigated. We examined the relationship between prior lumbar fusion or discectomy and the risk of dislocation or revision after THA. We also examined the effect of femoral head component diameter on the risk of dislocation or revision. Methods. Data used in this study were compiled from several Finnish national health registers, including the Finnish Arthroplasty Register (FAR) which was the primary source for prosthesis-related data. Other registers used in this study included the Finnish Health Care Register (HILMO), the Social Insurance Institutions (SII) registers, and Statistics Finland. The study was conducted as a prospective retrospective cohort study. Cox proportional hazards regression and Kaplan-Meier survival analysis were used for analysis. Results. Prior lumbar fusion surgery was associated with increased risk of prosthetic dislocation (hazard ratio (HR) = 2.393, p < 0.001) and revision (HR = 1.528, p < 0.001). Head components larger than 28 mm were associated with lower dislocation rates compared to the 28 mm head (32 mm: HR = 0.712, p < 0.001; 36 mm: HR = 0.700, p < 0.001; 38 mm: HR = 0.808, p < 0.140; and 40 mm: HR = 0.421, p < 0.001). Heads of 38 mm (HR = 1.288, p < 0.001) and 40 mm (HR = 1.367, p < 0.001) had increased risk of revision compared to the 28 mm head. Conclusion. Lumbar fusion surgery was associated with higher rate of
Post-traumatic periprosthetic acetabular fractures are rare but serious. Few studies carried out on small cohorts have reported them in the literature. The aim of this work is to describe the specific characteristics of post-traumatic periprosthetic acetabular fractures, and the outcome of their surgical treatment in terms of function and complications. Patients with this type of fracture were identified retrospectively over a period of six years (January 2016 to December 2021). The following data were collected: demographic characteristics, date of insertion of the prosthesis, details of the intervention, date of the trauma, characteristics of the fracture, and type of treatment. Functional results were assessed with the Harris Hip Score (HHS). Data concerning complications of treatment were collected.Aims
Methods
The objectives of this study were to investigate the patient characteristics and mortality of Vancouver type B periprosthetic femoral fractures (PFF) subgroups divided into two groups according to femoral component stability and to compare postoperative clinical outcomes according to treatment in Vancouver type B2 and B3 fractures. A total of 126 Vancouver type B fractures were analyzed from 2010 to 2019 in 11 associated centres' database (named TRON). We divided the patients into two Vancouver type B subtypes according to implant stability. Patient demographics and functional scores were assessed in the Vancouver type B subtypes. We estimated the mortality according to various patient characteristics and clinical outcomes between the open reduction internal fixation (ORIF) and revision arthroplasty (revision) groups in patients with unstable subtype.Aims
Methods
Arthroplasty surgery of the knee and hip is performed in two to three million patients annually. Periprosthetic joint infections occur in 4% of these patients. Debridement, antibiotics, and implant retention (DAIR) surgery aimed at cleaning the infected prosthesis often fails, subsequently requiring invasive revision of the complete prosthetic reconstruction. Infection-specific imaging may help to guide DAIR. In this study, we evaluated a bacteria-specific hybrid tracer (99mTc-UBI29-41-Cy5) and its ability to visualize the bacterial load on femoral implants using clinical-grade image guidance methods.
99mTc-UBI29-41-Cy5 specificity for Aims
Methods
Several previously identified patient-, injury-, and treatment-related factors are associated with the development of nonunion in distal femur fractures. However, the predictive value of these factors is not well defined. We aimed to assess the predictive ability of previously identified risk factors in the development of nonunion leading to secondary surgery in distal femur fractures. We conducted a retrospective cohort study of adult patients with traumatic distal femur fracture treated with lateral locking plate between 2009 and 2018. The patients who underwent secondary surgery due to fracture healing problem or plate failure were considered having nonunion. Background knowledge of risk factors of distal femur fracture nonunion based on previous literature was used to form an initial set of variables. A logistic regression model was used with previously identified patient- and injury-related variables (age, sex, BMI, diabetes, smoking, periprosthetic fracture, open fracture, trauma energy, fracture zone length, fracture comminution, medial side comminution) in the first analysis and with treatment-related variables (different surgeon-controlled factors, e.g. plate length, screw placement, and proximal fixation) in the second analysis to predict the nonunion leading to secondary surgery in distal femur fractures.Aims
Methods
Young adults undergoing total hip arthroplasty (THA) largely have different indications for surgery, preoperative function, and postoperative goals compared to a standard patient group. The aim of our study was to describe young adult THA preoperative function and quality of life, and to assess postoperative satisfaction and compare this with functional outcome measures. A retrospective cohort analysis of young adults (aged < 50 years) undergoing THA between May 2018 and May 2023 in a single tertiary centre was undertaken. Median follow-up was 31 months (12 to 61). Oxford Hip Score (OHS) and focus group-designed questionnaires were distributed. Searches identified 244 cases in 225 patients. Those aged aged under 30 years represented 22.7% of the cohort. Developmental dysplasia of the hip (50; 45.5%) and Perthes’ disease (15; 13.6%) were the commonest indications for THA.Aims
Methods
Fracture-related infection (FRI) is commonly classified based on the time of onset of symptoms. Early infections (< two weeks) are treated with debridement, antibiotics, and implant retention (DAIR). For late infections (> ten weeks), guidelines recommend implant removal due to tolerant biofilms. For delayed infections (two to ten weeks), recommendations are unclear. In this study we compared infection clearance and bone healing in early and delayed FRI treated with DAIR in a rabbit model.
Aims
Methods
The aim of this study was to investigate the association between fracture displacement and survivorship of the native hip joint without conversion to a total hip arthroplasty (THA), and to determine predictors for conversion to THA in patients treated nonoperatively for acetabular fractures. A multicentre cross-sectional study was performed in 170 patients who were treated nonoperatively for an acetabular fracture in three level 1 trauma centres. Using the post-injury diagnostic CT scan, the maximum gap and step-off values in the weightbearing dome were digitally measured by two trauma surgeons. Native hip survival was reported using Kaplan-Meier curves. Predictors for conversion to THA were determined using Cox regression analysis.Aims
Methods
The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate. A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.Aims
Methods
Literature surrounding artificial intelligence (AI)-related applications for hip and knee arthroplasty has proliferated. However, meaningful advances that fundamentally transform the practice and delivery of joint arthroplasty are yet to be realized, despite the broad range of applications as we continue to search for meaningful and appropriate use of AI. AI literature in hip and knee arthroplasty between 2018 and 2021 regarding image-based analyses, value-based care, remote patient monitoring, and augmented reality was reviewed. Concerns surrounding meaningful use and appropriate methodological approaches of AI in joint arthroplasty research are summarized. Of the 233 AI-related orthopaedics articles published, 178 (76%) constituted original research, while the rest consisted of editorials or reviews. A total of 52% of original AI-related research concerns hip and knee arthroplasty (n = 92), and a narrative review is described. Three studies were externally validated. Pitfalls surrounding present-day research include conflating vernacular (“AI/machine learning”), repackaging limited registry data, prematurely releasing internally validated prediction models, appraising model architecture instead of inputted data, withholding code, and evaluating studies using antiquated regression-based guidelines. While AI has been applied to a variety of hip and knee arthroplasty applications with limited clinical impact, the future remains promising if the question is meaningful, the methodology is rigorous and transparent, the data are rich, and the model is externally validated. Simple checkpoints for meaningful AI adoption include ensuring applications focus on: administrative support over clinical evaluation and management; necessity of the advanced model; and the novelty of the question being answered. Cite this article:
Our objective was describing an algorithm to identify and prevent vascular injury in patients with intrapelvic components. Patients were defined as at risk to vascular injuries when components or cement migrated 5 mm or more beyond the ilioischial line in any of the pelvic incidences (anteroposterior and Judet view). In those patients, a serial investigation was initiated by a CT angiography, followed by a vascular surgeon evaluation. The investigation proceeded if necessary. The main goal was to assure a safe tissue plane between the hardware and the vessels.Aims
Methods
Periprosthetic fractures (PPFs) following hip arthroplasty are complex injuries. This study evaluates patient demographic characteristics, management, outcomes, and risk factors associated with PPF subtypes over a decade. Using a multicentre collaborative study design, independent of registry data, we identified adults from 29 centres with PPFs around the hip between January 2010 and December 2019. Radiographs were assessed for the Unified Classification System (UCS) grade. Patient and injury characteristics, management, and outcomes were compared between UCS grades. A multinomial logistic regression was performed to estimate relative risk ratios (RRR) of variables on UCS grade.Aims
Methods
Factors associated with high mortality rates in geriatric hip fracture patients are frequently unmodifiable. Time to surgery, however, might be a modifiable factor of interest to optimize clinical outcomes after hip fracture surgery. This study aims to determine the influence of postponement of surgery due to non-medical reasons on clinical outcomes in acute hip fracture surgery. This observational cohort study enrolled consecutively admitted patients with a proximal femoral fracture, for which surgery was performed between 1 January 2018 and 11 January 2021 in two level II trauma teaching hospitals. Patients with medical indications to postpone surgery were excluded. A total of 1,803 patients were included, of whom 1,428 had surgery < 24 hours and 375 had surgery ≥ 24 hours after admission.Aims
Methods
Dislocation of the hip remains a major complication after periacetabular tumour resection and endoprosthetic reconstruction. The position of the acetabular component is an important modifiable factor for surgeons in determining the risk of postoperative dislocation. We investigated the significance of horizontal, vertical, and sagittal displacement of the hip centre of rotation (COR) on postoperative dislocation using a CT-based 3D model, as well as other potential risk factors for dislocation. A total of 122 patients who underwent reconstruction following resection of periacetabular tumour between January 2011 and January 2020 were studied. The risk factors for dislocation were investigated with univariate and multivariate logistic regression analysis on patient-specific, resection-specific, and reconstruction-specific variables.Aims
Methods
The aim of this study was to describe the management and associated outcomes of patients sustaining a femoral hip periprosthetic fracture (PPF) in the UK population. This was a multicentre retrospective cohort study including adult patients who presented to 27 NHS hospitals with 539 new PPFs between 1 January 2018 and 31 December 2018. Data collected included: management strategy (operative and nonoperative), length of stay, discharge destination, and details of post-treatment outcomes (reoperation, readmission, and 30-day and 12-month mortality). Descriptive analysis by fracture type was performed, and predictors of PPF management and outcomes were assessed using mixed-effects logistic regression.Aims
Methods
We aimed to evaluate the utility of 68Ga-citrate positron emission tomography (PET)/CT in the differentiation of periprosthetic joint infection (PJI) and aseptic loosening (AL), and compare it with 99mTc-methylene bisphosphonates (99mTc-MDP) bone scan. We studied 39 patients with suspected PJI or AL. These patients underwent 68Ga-citrate PET/CT, 99mTc-MDP three-phase bone scan and single-photon emission CT (SPECT)/CT. PET/CT was performed at ten minutes and 60 minutes after injection, respectively. Images were evaluated by three nuclear medicine doctors based on: 1) visual analysis of the three methods based on tracer uptake model, and PET images attenuation-corrected with CT and those not attenuation-corrected with CT were analyzed, respectively; and 2) semi-quantitative analysis of PET/CT: maximum standardized uptake value (SUVmax) of lesions, SUVmax of the lesion/SUVmean of the normal bone, and SUVmax of the lesion/SUVmean of the normal muscle. The final diagnosis was based on the clinical and intraoperative findings, and histopathological and microbiological examinations.Aims
Methods
Two-stage exchange revision total hip arthroplasty (THA) performed in case of periprosthetic joint infection (PJI) has been considered for many years as being the gold standard for the treatment of chronic infection. However, over the past decade, there have been concerns about its safety and its effectiveness. The purposes of our study were to investigate our practice, collecting the overall spacer complications, and then to analyze their risk factors. We retrospectively included 125 patients with chronic hip PJI who underwent a staged THA revision performed between January 2013 and December 2019. All spacer complications were systematically collected, and risk factors were analyzed. Statistical evaluations were performed using the Student's Aims
Methods
Total hip arthroplasty (THA) in patients with post-polio residual paralysis (PPRP) is challenging. Despite relief in pain after THA, pre-existing muscle imbalance and altered gait may cause persistence of difficulty in walking. The associated soft tissue contractures not only imbalances the pelvis, but also poses the risk of dislocation, accelerated polyethylene liner wear, and early loosening. In all, ten hips in ten patients with PPRP with fixed pelvic obliquity who underwent THA as per an algorithmic approach in two centres from January 2014 to March 2018 were followed-up for a minimum of two years (2 to 6). All patients required one or more additional soft tissue procedures in a pre-determined sequence to correct the pelvic obliquity. All were invited for the latest clinical and radiological assessment.Aims
Methods
Objectives. The cytotoxicity induced by cobalt ions (Co. 2+. ) and cobalt nanoparticles (Co-NPs) which released following the insertion of a total
A 66-year-old man presented with an infected resurfacing hip replacement in the immediate post-operative period following an uneventful transurethral resection of the prostate. Prophylactic antibiotics had been administered on induction of anaesthesia. The
The aims of this study were to develop an in vivo model of periprosthetic joint infection (PJI) in cemented hip hemiarthroplasty, and to monitor infection and biofilm formation in real-time. Sprague-Dawley rats underwent cemented hip hemiarthroplasty via the posterior approach with pre- and postoperative gait assessments. Infection with Aims
Methods
Antibiotic resistance represents a threat to human health. It has been suggested that by 2050, antibiotic-resistant infections could cause ten million deaths each year. In orthopaedics, many patients undergoing surgery suffer from complications resulting from implant-associated infection. In these circumstances secondary surgery is usually required and chronic and/or relapsing disease may ensue. The development of effective treatments for antibiotic-resistant infections is needed. Recent evidence shows that bacteriophage (phages; viruses that infect bacteria) therapy may represent a viable and successful solution. In this review, a brief description of bone and joint infection and the nature of bacteriophages is presented, as well as a summary of our current knowledge on the use of bacteriophages in the treatment of bacterial infections. We present contemporary published in vitro and in vivo data as well as data from clinical trials, as they relate to bone and joint infections. We discuss the potential use of bacteriophage therapy in orthopaedic infections. This area of research is beginning to reveal successful results, but mostly in nonorthopaedic fields. We believe that bacteriophage therapy has potential therapeutic value for implant-associated infections in orthopaedics. Cite this article:
In a time of limited resources, the debate continues
over which types of
The aim of this prospective cohort study was to evaluate the early migration of the TriFit cementless proximally coated tapered femoral stem using radiostereometric analysis (RSA). A total of 21 patients (eight men and 13 women) undergoing primary total hip arthroplasty (THA) for osteoarthritis of the hip were recruited in this study and followed up for two years. Two patients were lost to follow-up. All patients received a TriFit stem and Trinity Cup with a vitamin E-infused highly cross-linked ultra-high molecular weight polyethylene liner. Radiographs for RSA were taken postoperatively and then at three, 12, and 24 months. Oxford Hip Score (OHS), EuroQol five-dimension questionnaire (EQ-5D), and adverse events were reported.Aims
Methods
Resection of malignant bony tumours of the pelvis
creates large bone and soft-tissue defects, and is frequently associated
with complications such as wound dehiscence and deep infection.
We present the results of six patients in whom a rectus abdominis
myocutaneous (RAM) flap was used following resection of a malignant
tumour of the pelvis. Bony reconstruction was performed using a
constrained
The purpose of this study was to evaluate the cost of reverse shoulder arthroplasty (RSA) for patients with a proximal humerus fracture, using time-driven activity based costing (TDABC), and to compare treatment costs with reimbursement under the Healthcare Resource Groups (HRGs). TDABC analysis based on the principles outlined by Kaplan and a clinical pathway that has previously been validated for this institution was used. Staffing cost, consumables, implants, and overheads were updated to reflect 2019/2020 costs. This was compared with the HRG reimbursements.Aims
Methods
Second-generation metal-on-metal (MoM) articulations in total hip arthroplasty (THA) were introduced in order to reduce wear-related complications. The current study reports on the serum cobalt levels and the clinical outcome at a minimum of 20 years following THA with a MoM (Metasul) or a ceramic-on-polyethylene (CoP) bearing. The present study provides an update of a previously published prospective randomized controlled study, evaluating the serum cobalt levels of a consecutive cohort of 100 patients following THA with a MoM or a CoP articulation. A total of 31 patients were available for clinical and radiological follow-up examination. After exclusion of 11 patients because of other cobalt-containing implants, 20 patients (MoM (n = 11); CoP (n = 9)) with a mean age of 69 years (42 to 97) were analyzed. Serum cobalt levels were compared to serum cobalt levels five years out of surgery.Aims
Methods
The Thompson hemiarthroplasty is a popular
The aim of this study was to describe variation in hip fracture treatment in Norway expressed as adherence to international and national evidence-based treatment guidelines, to study factors influencing deviation from guidelines, and to analyze consequences of non-adherence. International and national guidelines were identified and treatment recommendations extracted. All 43 hospitals routinely treating hip fractures in Norway were characterized. From the Norwegian Hip Fracture Register (NHFR), hip fracture patients aged > 65 years and operated in the period January 2014 to December 2018 for fractures with conclusive treatment guidelines were included (n = 29,613: femoral neck fractures (n = 21,325), stable trochanteric fractures (n = 5,546), inter- and subtrochanteric fractures (n = 2,742)). Adherence to treatment recommendations and a composite indicator of best practice were analyzed. Patient survival and reoperations were evaluated for each recommendation.Aims
Methods
We report a patient with a peri-acetabular chondrosarcoma which was treated by resection of part of the pelvis and the femoral head. This was followed by replacement with a custom-made titanium alloy pelvic portion and a conventional total
This paper reports a singular case of pathological fracture through a large osteolytic lesion of the mid-femur consequent on reaction to a
We performed thermal analysis of polyethylene samples obtained from 73 new cups of the Weller
Four en-bloc resections for malignant tumours of the hip, the peri-acetabular region and the iliac wing were reconstructed using an irradiated hemipelvic allograft together with a total
The ABG I cementless
Deep infection, the most serious local complication of total hip replacement, prompted a study of the records of 135 patients (137 hips) thus afflicted in a nationwide survey of Canada. Particular attention has been paid to the natural history of the infection, and the problems of diagnosis are described. Twenty-one patients died after the insertion, or removal, of the prosthesis, and of the survivors of the original 135 patients only eighteen have been able to retain the prosthesis without further problems with the wound. The remaining patients had the prosthesis removed, and most dry wounds. Certain suggestions are made on management. The advice that a second total
In acetabular dysplasia, fixation of the acetabular component of a cemented total
We investigated the circulating levels of the main cytokines involved in bone resorption (IL-1β, IL-6, TNF-α), prostaglandins (PGE. 2. ) and metalloproteases (MMP-1), as possible early markers of osteolysis, in the serum of eight patients with periprosthetic osteolysis and ten patients without osteolysis. All had received a cementless
Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs. A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared.Aims
Methods
We have studied the effect of hydroxyapatite (HA) coating in 15 ovariectomised and 15 normal rats which had had a sham procedure. Twenty-four weeks after operation, HA-coated implants were inserted into the intramedullary canal of the right femur and uncoated implants into the left femur. The prostheses were removed four weeks after implantation. Twelve specimens in each group had mechanical push-out tests. Sagittal sections of the other three were evaluated by SEM. The bone mineral density (BMD) of the dissected left tibia was measured by dual-energy x-ray absorptiometry. The difference in BMD between the control and ovariectomised tibiae was 35.01 mg/cm. 2. (95% CI, 26.60 to 43.42). The push-out strength of the HA-coated implants was higher than that of the uncoated implants in both groups (p <
0.0001), but the HA-coated implants of the ovariectomised group had a reduction in push-out strength of 40.3% compared with the control group (p <
0.0001). Our findings suggest that HA-coated implants may improve the fixation of a cementless total
We investigated patient characteristics and outcomes of Vancouver type B periprosthetic fractures treated with femoral component revision and/or osteosynthesis. The study utilized data from the Swedish Hip Arthroplasty Register (SHAR) and information from patient records. We included all primary total hip arthroplasties (THAs) performed in Sweden since 1979, and undergoing further surgery due to Vancouver type B periprosthetic femoral fracture between 2001 and 2011. The primary outcome measure was any further reoperation between 2001 and 2013. Cross-referencing with the National Patient Register was performed in two stages, in order to identify all surgical procedures not recorded on the SHAR.Aims
Patients and Methods
We describe the survival at ten years of 100 femoral components of the Freeman
Our aim was to determine if the serum levels of bone-resorbing cytokines (IL-1β, TNF-α, IL-6, GM-CSF) are altered in patients with aseptic loosening of a total
Biopsy of the periprosthetic tissue is an important diagnostic tool for prosthetic joint infection (PJI) as it enables the detection of the responsible microorganism with its sensitivity to antibiotics. We aimed to investigate how often the bacteria identified in the tissue analysis differed between samples obtained from preoperative biopsy and intraoperative revision surgery in cases of late PJI; and whether there was a therapeutic consequence. A total of 508 patients who required revision surgery of total hip arthroplasty (THA) (n = 231) or total knee arthroplasty (TKA) (n = 277) because of component loosening underwent biopsy before revision surgery. The tissue samples collected at biopsy and during revision surgery were analyzed according to the criteria of the Musculoskeletal Infection Society (MSIS).Aims
Methods
1. Currently available total replacement hip and knee prostheses were tested in a machine enabling flexion-extension movements to be applied whilst the prostheses were surrounded with Ringer's solution or other liquid and loaded within the physiological range. 2. Prostheses of which both components were made in cobalt-chromium-molybdenum alloy produced visible quantities of alloy particles, whose sizes ranged down to about 0·1 microns, and cobalt and molybdenum ions in solution. 3. No metallic or plastic particles were detected during tests on a
Preclinical data showed poly(methyl methacrylate) (PMMA) loaded with microsilver to be effective against a variety of bacteria. The purpose of this study was to assess patient safety of PMMA spacers with microsilver in prosthetic hip infections in a prospective cohort study. A total of 12 patients with prosthetic hip infections were included for a three-stage revision procedure. All patients received either a gentamicin-PMMA spacer (80 g to 160 g PMMA depending on hip joint dimension) with additional loading of 1% (w/w) of microsilver (0.8 g to 1.6 g per spacer) at surgery 1 followed by a gentamicin-PMMA spacer without microsilver at surgery 2 or vice versa. Implantation of the revision prosthesis was carried out at surgery 3.Objectives
Methods
This single-centre observational study aimed to describe the results of extensive bone impaction grafting of the whole acetabular cavity in combination with an uncemented component in acetabular revisions performed in a standardized manner since 1993. Between 1993 and 2013, 370 patients with a median age of 72 years (interquartile range (IQR) 63 to 79 years) underwent acetabular revision surgery. Of these, 229 were more than ten years following surgery and 137 were more than 15 years. All revisions were performed with extensive use of morcellized allograft firmly impacted into the entire acetabular cavity, followed by insertion of an uncemented component with supplementary screw fixation. All types of reoperation were captured using review of radiographs and medical charts, combined with data from the local surgical register and the Swedish Hip Arthroplasty Register.Aims
Methods
We have compared the energy expenditure during walking in three patients, aged between 51 and 55 years, with unilateral disarticulation of the hip when using the mechanical-controlled stance-phase control knee (Otto Bock 3R15) and the microprocessor-controlled pneumatic swing-phase control knee (Intelligent Prosthesis, IP). All had an endoskeletal
The present study investigated the five-year interval changes in pseudotumours and measured serum metal ions at long-term follow-up of a previous report of 28 mm diameter metal-on-metal (MoM) total hip arthroplasty (THA). A total of 72 patients (mean age 46.6 years (37 to 55); 43 men, 29 women; 91 hips) who underwent cementless primary MoM THA with a 28 mm modular head were included. The mean follow-up duration was 20.3 years (18 to 24). All patients had CT scans at a mean 15.1 years (13 to 19) after the index operation and subsequent follow-up at a mean of 20.2 years (18 to 24). Pseudotumour volume, type of mass, and new-onset pseudotumours were evaluated using CT scanning. Clinical outcomes were assessed by Harris Hip Score (HHS) and the presence of groin pain. Serum metal ion (cobalt (Co) and chromium (Cr)) levels were measured at the latest follow-up.Aims
Patients and Methods
We studied the impact of direct anterior (DA) A total of 6086 consecutive patients undergoing primary total hip arthroplasty (THA) at a single institution between 2013 and 2016 were retrospectively evaluated. Data obtained from electronic patient medical records included age, sex, body mass index (BMI), medical comorbidities, surgical approach, and presence of deep PJI. There were 3053 male patients (50.1%) and 3033 female patients (49.9%). The mean age and BMI of the entire cohort was 62.7 years (18 to 102, Aims
Patients and Methods
Implant-related infection is one of the leading reasons for failure in orthopaedics and trauma, and results in high social and economic costs. Various antibacterial coating technologies have proven to be safe and effective both in preclinical and clinical studies, with post-surgical implant-related infections reduced by 90% in some cases, depending on the type of coating and experimental setup used. Economic assessment may enable the cost-to-benefit profile of any given antibacterial coating to be defined, based on the expected infection rate with and without the coating, the cost of the infection management, and the cost of the coating. After reviewing the latest evidence on the available antibacterial coatings, we quantified the impact caused by delaying their large-scale application. Considering only joint arthroplasties, our calculations indicated that for an antibacterial coating, with a final user’s cost price of €600 and able to reduce post-surgical infection by 80%, each year of delay to its large-scale application would cause an estimated 35 200 new cases of post-surgical infection in Europe, equating to additional hospital costs of approximately €440 million per year. An adequate reimbursement policy for antibacterial coatings may benefit patients, healthcare systems, and related research, as could faster and more affordable regulatory pathways for the technologies still in the pipeline. This could significantly reduce the social and economic burden of implant-related infections in orthopaedics and trauma.
Prosthetic joint infection (PJI) is a devastating complication following total joint arthroplasty. Non-contact induction heating of metal implants is a new and emerging treatment for PJI. However, there may be concerns for potential tissue necrosis. It is thought that segmental induction heating can be used to control the thermal dose and to limit collateral thermal injury to the bone and surrounding tissues. The purpose of this study was to determine the thermal dose, for commonly used metal implants in orthopaedic surgery, at various distances from the heating centre (HC). Commonly used metal orthopaedic implants (hip stem, intramedullary nail, and locking compression plate (LCP)) were heated segmentally using an induction heater. The thermal dose was expressed in cumulative equivalent minutes at 43°C (CEM43) and measured with a thermal camera at several different distances from the HC. A value of 16 CEM43 was used as the threshold for thermal damage in bone.Objectives
Methods
Prosthetic joint infection (PJI) diagnosis is a major challenge in orthopaedics, and no reliable parameters have been established for accurate, preoperative predictions in the differential diagnosis of aseptic loosening or PJI. This study surveyed factors in synovial fluid (SF) for improving PJI diagnosis. We enrolled 48 patients (including 39 PJI and nine aseptic loosening cases) who required knee/hip revision surgery between January 2016 and December 2017. The PJI diagnosis was established according to the Musculoskeletal Infection Society (MSIS) criteria. SF was used to survey factors by protein array and enzyme-linked immunosorbent assay to compare protein expression patterns in SF among three groups (aseptic loosening and first- and second-stage surgery). We compared routine clinical test data, such as C-reactive protein level and leucocyte number, with potential biomarker data to assess the diagnostic ability for PJI within the same patient groups.Objectives
Methods
Deep infection was identified as a serious complication in the earliest days of total hip arthroplasty. It was identified that airborne contamination in conventional operating theatres was the major contributing factor. As progress was made in improving the engineering of operating theatres, airborne contamination was reduced. Detailed studies were carried out relating airborne contamination to deep infection rates. In a trial conducted by the United Kingdom Medical Research Council (MRC), it was found that the use of ultra-clean air (UCA) operating theatres was associated with a significant reduction in deep infection rates. Deep infection rates were further reduced by the use of a body exhaust system. The MRC trial also included a detailed microbiology study, which confirmed the relationship between airborne contamination and deep infection rates. Recent observational evidence from joint registries has shown that in contemporary practice, infection rates remain a problem, and may be getting worse. Registry observations have also called into question the value of “laminar flow” operating theatres. Observational evidence from joint registries provides very limited evidence on the efficacy of UCA operating theatres. Although there have been some changes in surgical practice in recent years, the conclusions of the MRC trial remain valid, and the use of UCA is essential in preventing deep infection. There is evidence that if UCA operating theatres are not used correctly, they may have poor microbiological performance. Current UCA operating theatres have limitations, and further research is required to update them and improve their microbiological performance in contemporary practice. Cite this article:
The aim of this study was to assess the current trends in the estimation of survival and the preferred forms of treatment of pathological fractures among national and international general and oncological orthopaedic surgeons, and to explore whether improvements in the management of these patients could be identified in this way. All members of the Dutch Orthopaedic Society (DOS) and the European Musculoskeletal Oncology Society (EMSOS) were invited to complete a web-based questionnaire containing 12 cases.Aims
Materials and Methods
Increasing innovation in rapid prototyping (RP)
and additive manufacturing (AM), also known as 3D printing, is bringing
about major changes in translational surgical research. This review describes the current position in the use of additive
manufacturing in orthopaedic surgery. Cite this article:
Favourable results for collarless polished tapered stems have been reported, and cement creep due to taper slip may be a contributing factor. However, the ideal cement thickness around polished stems remains unknown. We investigated the influence of cement thickness on stem subsidence and cement creep. We cemented six collarless polished tapered (CPT) stems (two stems each of small, medium and large sizes) into composite femurs that had been reamed with a large CPT rasp to achieve various thicknesses of the cement mantle. Two or three tantalum balls were implanted in the proximal cement in each femur. A cyclic loading test was then performed for each stem. The migration of the balls was measured three-dimensionally, using a micro-computed tomography (CT) scanner, before and after loading. A digital displacement gauge was positioned at the stem shoulder, and stem subsidence was measured continuously by the gauge. Final stem subsidence was measured at the balls at the end of each stem.Objectives
Methods
Computed tomography (CT) plays an important role in evaluating wear and periacetabular osteolysis (PAO) in total hip replacements. One concern with CT is the high radiation exposure since standard pelvic CT provides approximately 3.5 millisieverts (mSv) of radiation exposure, whereas a planar radiographic examination with three projections totals approximately 0.5 mSv. The objective of this study was to evaluate the lowest acceptable radiation dose for dual-energy CT (DECT) images when measuring wear and periacetabular osteolysis in uncemented metal components. A porcine pelvis with bilateral uncemented hip prostheses and with known linear wear and acetabular bone defects was examined in a third-generation multidetector DECT scanner. The examinations were performed with four different radiation levels both with and without iterative reconstruction techniques. From the high and low peak kilo voltage acquisitions, polychrmoatic images were created together with virtual monochromatic images of energies 100 kiloelectron volts (keV) and 150 keV.Objectives
Materials and Methods
The aim of this study was to compare the survivorship and radiographic
outcomes at ten-year follow-up of three prospective consecutive
series of patients each of which received a different design of
cementless femoral components for total hip arthroplasty (THA). In Cohort 1, 91 consecutive patients (100 hips) underwent THA
with a cementless porous-coated anatomic femoral stem (PCA) between
October 1983 and January 1986. In Cohort 2, 86 consecutive patients
(100 hips) underwent THA with an extensively porous-coated cementless
femoral stem (Prodigy) between June 1994 and October 1997. In Cohort
3, 88 consecutive patients (100 hips) underwent THA with a proximally
porous-coated triple-tapered cementless stem (Summit) between April
2002 and October 2003. All three groups underwent prospective clinical
and radiographic evaluation.Aims
Patients and Methods
The diagnosis of periprosthetic joint infection (PJI) remains
demanding due to limitations of all the available diagnostic tests.
The synovial fluid marker, α-defensin, is a promising adjunct for
the assessment of potential PJI. The purpose of this study was to
investigate the qualitative assessment of α-defensin, using Synovasure
to detect or exclude periprosthetic infection in total joint arthroplasty. We studied 50 patients (28 women, 22 men, mean age 65 years;
20 to 89) with a clinical indication for revision arthroplasty who
met the inclusion criteria of this prospective diagnostic study.
The presence of α-defensin was determined using the qualitative
Synovasure test and compared with standard diagnostic methods for
PJI. Based on modified Musculoskeletal Infection Society (MSIS)
criteria, 13 cases were categorised as septic and 36 as aseptic revisions.
One test was inconclusive.Aims
Patients and Methods
The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of metal-on-metal (MoM) hip joints. Some designs of MoM hip joint do, however, have reasonable mid-term performance when implanted in appropriate patients. Investigations into the reasons for implant failure are important to offer help with the choice of implants and direction for future implant designs. One way to assess the performance of explanted hip prostheses is to measure the wear (in terms of material loss) on the joint surfaces. In this study, a coordinate measuring machine (CMM) was used to measure the wear on five failed cementless Biomet Magnum/ReCap/ Taperloc large head MoM THAs, along with one Biomet ReCap resurfacing joint. Surface roughness measurements were also taken. The reason for revision of these implants was pain and/or adverse reaction to metal debris (ARMD) and/or elevated blood metal ion levels.Objectives
Methods
Alarm over the reported high failure rates for metal-on-metal (MoM) hip implants as well as their potential for locally aggressive Adverse Reactions to Metal Debris (ARMDs) has prompted government agencies, internationally, to recommend the monitoring of patients with MoM hip implants. Some have advised that a blood ion level >7 µg/L indicates potential for ARMDs. We report a systematic review and meta-analysis of the performance of metal ion testing for ARMDs. We searched MEDLINE and EMBASE to identify articles from which it was possible to reconstruct a 2 × 2 table. Two readers independently reviewed all articles and extracted data using explicit criteria. We computed a summary receiver operating curve using a Bayesian random-effects hierarchical model.Objectives
Methods
The widely used and well-proven Palacos R (a.k.a. Refobacin Palacos
R) bone cement is no longer commercially available and was superseded
by Refobacin bone cement R and Palacos R + G in 2005. However, the
performance of these newly introduced bone cements have not been
tested in a phased evidence-based manner, including roentgen stereophotogrammetric
analysis (RSA). In this blinded, randomised, clinical RSA study, the migration
of the Stanmore femoral component was compared between Refobacin
bone cement R and Palacos R + G in 62 consecutive total hip arthroplasties.
The primary outcome measure was femoral component migration measured
using RSA and secondary outcomes were Harris hip score (HHS), Hip
disability and Osteoarthritis Outcome Score (HOOS), EuroQol 5D (EQ-5D)
and Short Form 36 (SF-36).Aims
Patients and Methods
We wanted to investigate regional variations in the organisms
reported to be causing peri-prosthetic infections and to report
on prophylaxis regimens currently in use across England. Analysis of data routinely collected by Public Health England’s
(PHE) national surgical site infection database on elective primary
hip and knee arthroplasty procedures between April 2010 and March
2013 to investigate regional variations in causative organisms.
A separate national survey of 145 hospital Trusts (groups of hospitals
under local management) in England routinely performing primary
hip and/or knee arthroplasty was carried out by standard email questionnaire.Objectives
Methods
Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection. Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences.Objectives
Materials and Methods
The long-term survival of the cementless Spotorno
CLS femoral component in patients aged >
50 years at the time of
arthroplasty was investigated. Survivorship analysis of a consecutive
series of 85 patients (100 hips; under 50 years of age at a mean
follow-up of 18.4 years (16.3 to 20.8)) was performed. The clinical
and radiographic outcomes were satisfactory. The overall rate of
survival of the femoral component was 93.5% (95% confidence interval
(CI), 90.9 to 96.1) after 19 years. Survival with revision for aseptic
loosening as the end point was 95.7% (95% CI 93.6 to 97.8%) at 19
years. This study demonstrates an excellent long-term survival of the
Spotorno CLS femoral component after 16 to 20 years in young patients
undergoing total hip arthroplasty. Cite this article:
High failure rates of metal-on-metal hip arthroplasty implants have highlighted the need for more careful introduction and monitoring of new implants and for the evaluation of the safety of medical devices. The National Joint Registry and other regulatory services are unable to detect failing implants at an early enough stage. We aimed to identify validated surrogate markers of long-term outcome in patients undergoing primary total hip arthroplasty (THA). We conducted a systematic review of studies evaluating surrogate markers for predicting long-term outcome in primary THA. Long-term outcome was defined as revision rate of an implant at ten years according to National Institute of Health and Care Excellence guidelines. We conducted a search of Medline and Embase (OVID) databases. Separate search strategies were devised for the Cochrane database and Google Scholar. Each search was performed to include articles from the date of their inception to June 8, 2015.Objectives
Methods
The August 2015 Research Roundup360 looks at: Lightbulbs, bleeding and procedure durations; Infection and rheumatoid agents; Infection rates and ‘bundles of care’ revisited; ACI: new application for a proven technology?; Hydrogel coating given the thumbs up; Hydroxyapatite as a smart coating?
We compared the incidence of pseudotumours after
large head metal-on-metal (MoM) total hip arthroplasty (THA) with
that after conventional metal-on-polyethylene (MoP) THA and assessed
the predisposing factors to pseudotumour formation. From a previous randomised controlled trial which compared large
head (38 mm to 60 mm) cementless MoM THA with conventional head
(28 mm) cementless MoP THA, 93 patients (96 THAs: 41 MoM (21 males,
20 females, mean age of 64 years, standard deviation ( The incidence of pseudotumours, measured using a standardised
CT protocol was 22 (53.7%) after MoM THA and 12 (21.8%) after MoP
THA. Women with a MoM THA were more likely to develop a pseudotumour
than those with a MoP THA (15 Contrary to popular belief, pseudotumours occur frequently around
MoP THAs. Women with a MoM THA and an elevated cobalt level are
at greatest risk. In this study, pseudotumours had no effect on
the functional outcome after either large head MoM or conventional
MoP THA. Cite this article:
Hip resurfacing arthroplasty (HRA) is an alternative to conventional
total hip arthroplasty for patients with osteonecrosis (ON) of the
femoral head. Our aim was to report the long-term outcome of HRA,
which is not currently known. Long-term survivorship, clinical scores and radiographic results
for 82 patients (99 hips) treated with HRA for ON over a period
of 18 years were reviewed retrospectively. The mean age of the 67
men and 15 women at the time of surgery was 40.8 years (14 to 64).
Patients were resurfaced regardless of the size of the osteonecrotic
lesion.Aims
Patients and Methods
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
We report the clinical and radiographic outcomes
of 208 consecutive femoral revision arthroplasties performed in 202
patients (119 women, 83 men) between March 1991 and December 2007
using the X-change Femoral Revision System, fresh-frozen morcellised
allograft and a cemented polished Exeter stem. All patients were
followed prospectively. The mean age of the patients at revision
was 65 years (30 to 86). At final review in December 2013 a total
of 130 patients with 135 reconstructions (64.9%) were alive and
had a non re-revised femoral component after a mean follow-up of
10.6 years (4.7 to 20.9). One patient was lost to follow-up at six
years, and their data were included up to this point.
Re-operation for any reason was performed in 33 hips (15.9%), in
13 of which the femoral component was re-revised (6.3%). The mean
pre-operative Harris hip score was 52 (19 to 95) (n = 73) and improved
to 80 (22 to 100) (n = 161) by the last follow-up. Kaplan–Meier
survival with femoral re-revision for any reason as the endpoint
was 94.9% (95% confidence intervals (CI) 90.2 to 97.4) at ten years;
with femoral re-revision for aseptic loosening as the endpoint it was
99.4% (95% CI 95.7 to 99.9); with femoral re-operation for any reason
as the endpoint it was 84.5% (95% CI 78.3 to 89.1); and with subsidence ≥ 5
mm it was 87.3% (95% CI 80.5 to 91.8). Femoral revision with the
use of impaction allograft bone grafting and a cemented polished
stem results in a satisfying survival rate at a mean of ten years’ follow-up. Cite this article:
It is becoming increasingly common for a patient
to have ipsilateral hip and knee replacements. The inter-prosthetic (IP)
distance, the distance between the tips of hip and knee prostheses,
has been thought to be associated with an increased risk of IP fracture.
Small gap distances are generally assumed to act as stress risers,
although there is no real biomechanical evidence to support this. The purpose of this study was to evaluate the influence of IP
distance, cortical thickness and bone mineral density on the likelihood
of an IP femoral fracture. A total of 18 human femur specimens were randomised into three
groups by bone density and cortical thickness. For each group, a
defined IP distance of 35 mm, 80 mm or 160 mm was created by choosing
the appropriate lengths of component. The maximum fracture strength
was determined using a four-point bending test. The fracture force of all three groups was similar (p = 0.498).
There was a highly significant correlation between the cortical
area and the fracture strength (r = 0.804, p <
0.001), whereas
bone density showed no influence. This study suggests that the IP distance has little influence
on fracture strength in IP femoral fractures: the thickness of the
cortex seems to be the decisive factor. Cite this article: