We have assessed the effect of different regimes of
The duration of systemic antibiotic treatment following first-stage revision surgery for periprosthetic joint infection (PJI) after total hip arthroplasty (THA) is contentious. Our philosophy is to perform an aggressive debridement, and to use a high local concentration of targeted antibiotics in cement beads and systemic prophylactic antibiotics alone. The aim of this study was to assess the success of this philosophy in the management of PJI of the hip using our two-stage protocol. The study involved a retrospective review of our prospectively collected database from which we identified all patients who underwent an intended two-stage revision for PJI of the hip. All patients had a diagnosis of PJI according to the major criteria of the Musculoskeletal Infection Society (MSIS) 2013, a minimum five-year follow-up, and were assessed using the MSIS working group outcome-reporting tool. The outcomes were grouped into ‘successful’ or ‘unsuccessful’.Aims
Methods
BoneMaster is a thin electrochemically applied hydroxyapatite (HA) coating for orthopaedic implants that is quickly resorbed during osseointegration. Early stabilization is a surrogacy marker of good survival of femoral stems. The hypothesis of this study was that a BoneMaster coating yields a fast early and lasting fixation of stems. A total of 53 patients were randomized to be treated using Bi-Metric cementless femoral stems with either only a porous titanium plasma-sprayed coating (P group) or a porous titanium plasma-sprayed coating with an additional BoneMaster coating (PBM group). The patients were examined with radiostereometry until five years after surgery.Aims
Methods
Total hip arthroplasty (THA) using the direct anterior approach (DAA) is undertaken with the patient in the supine position, creating an opportunity to replace both hips under one anaesthetic. Few studies have reported simultaneous bilateral DAA-THA. The aim of this study was to characterize a cohort of patients selected for this technique by a single, high-volume arthroplasty surgeon and to investigate their early postoperative clinical outcomes. Using an institutional database, we reviewed 643 patients who underwent bilateral DAA-THA by a single surgeon between 1 January 2010 and 31 December 2018. The demographic characteristics of the 256 patients (39.8%) who underwent simultaneous bilateral DAA-THA were compared with the 387 patients (60.2%) who underwent staged THA during the same period of time. We then reviewed the length of stay, rate of discharge home, 90-day complications, and readmissions for the simultaneous bilateral group.Aims
Methods
There is a paucity of long-term studies analyzing risk factors for failure after single-stage revision for periprosthetic joint infection (PJI) following total hip arthroplasty (THA). We report the mid- to long-term septic and non-septic failure rate of single-stage revision for PJI after THA. We retrospectively reviewed 88 cases which met the Musculoskeletal Infection Society (MSIS) criteria for PJI. Mean follow-up was seven years (1 to 14). Septic failure was diagnosed with a Delphi-based consensus definition. Any reoperation for mechanical causes in the absence of evidence of infection was considered as non-septic failure. A competing risk regression model was used to evaluate factors associated with septic and non-septic failures. A Kaplan-Meier estimate was used to analyze mortality.Aims
Methods
The purpose of this study was to evaluate unexpected positive cultures in total hip arthroplasty (THA) revisions for presumed aseptic loosening, to assess the prevalence of low-grade infection using two definition criteria, and to analyze its impact on implant survival after revision. A total of 274 THA revisions performed for presumed aseptic loosening from 2012 to 2016 were reviewed. In addition to obtaining intraoperative tissue cultures from all patients, synovial and sonication fluid samples of the removed implant were obtained in 215 cases (79%) and 101 cases (37%), respectively. Histopathological analysis was performed in 250 cases (91%). Patients were classified as having low-grade infections according to institutional criteria and Musculoskeletal Infection Society (MSIS) International Consensus Meeting (ICM) 2013 criteria. Low-grade infections according to institutional criteria were treated with targeted antibiotics for six weeks postoperatively. Implant failure was defined as the need for re-revision resulting from periprosthetic joint infection (PJI) and aseptic reasons. The mean follow-up was 68 months (26 to 95).Aims
Methods
Bone stock restoration of acetabular bone defects using impaction bone grafting (IBG) in total hip arthroplasty may facilitate future re-revision in the event of failure of the reconstruction. We hypothesized that the acetabular bone defect during re-revision surgery after IBG was smaller than during the previous revision surgery. The clinical and radiological results of re-revisions with repeated use of IBG were also analyzed. In a series of 382 acetabular revisions using IBG and a cemented component, 45 hips (45 patients) that had failed due to aseptic loosening were re-revised between 1992 and 2016. Acetabular bone defects graded according to Paprosky during the first and the re-revision surgery were compared. Clinical and radiological findings were analyzed over time. Survival analysis was performed using a competing risk analysis.Aims
Methods
We aimed to report the mid- to long-term rates of septic and aseptic failure after two-stage revision surgery for periprosthetic joint infection (PJI) following total hip arthroplasty (THA). We retrospectively reviewed 96 cases which met the Musculoskeletal Infection Society criteria for PJI. The mean follow-up was 90 months (SD 32). Septic failure was assessed using a Delphi-based consensus definition. Any further surgery undertaken for aseptic mechanical causes was considered as aseptic failure. The cumulative incidence with competing risk analysis was used to predict the risk of septic failure. A regression model was used to evaluate factors associated with septic failure. The cumulative incidence of aseptic failure was also analyzed.Aims
Methods
Although periarticular injection plays an important role in multimodal pain management following total hip arthroplasty (THA), there is no consensus on the optimal composition of the injection. In particular, it is not clear whether the addition of a corticosteroid improves the pain relief achieved nor whether it is associated with more complications than are observed without corticosteroid. The aim of this study was to quantify the safety and effectiveness of cortocosteroid use in periarticular injection during THA. We conducted a prospective, two-arm, parallel-group, randomized controlled trial involving patients scheduled for unilateral THA. A total of 187 patients were randomly assigned to receive periarticular injection containing either a corticosteroid (CS group) or without corticosteroid (no-CS group). Other perioperative interventions were identical for all patients. The primary outcome was postoperative pain at rest during the initial 24 hours after surgery. Pain score was recorded every three hours until 24 hours using a 100 mm visual analogue scale (VAS). The primary outcome was assessed based on the area under the curve (AUC).Aims
Methods
Of growing concern in arthroplasty is the emergence of atypical infections, particularly For this non-randomized non-blinded study, 101 adult patients scheduled for hip or knee surgery were recruited. For each, four 3 mm dermal punch biopsies were collected after administration of anaesthesia, but prior to antibiotics. Prebiopsy skin preparation consisted of a standardized preoperative 2% chlorhexidine skin cleansing protocol and an additional 70% isopropyl alcohol mechanical skin scrub immediately prior to biopsy collection. Two skin samples 10 cm apart were collected from a location approximating a standard direct anterior skin incision, and two samples 10 cm apart were collected from a lateral skin incision (suitable for posterior, direct-lateral, or anterolateral approaches). Samples were cultured for two weeks using a protocol optimized for Aims
Methods
During the COVID-19 pandemic, many patients continue to require urgent surgery for hip fractures. However, the impact of COVID-19 on perioperative outcomes in these high-risk patients remains unknown. The objectives of this study were to establish the effects of COVID-19 on perioperative morbidity and mortality, and determine any risk factors for increased mortality in patients with COVID-19 undergoing hip fracture surgery. This multicentre cohort study included 340 COVID-19-negative patients versus 82 COVID-19-positive patients undergoing surgical treatment for hip fractures across nine NHS hospitals in Greater London, UK. Patients in both treatment groups were comparable for age, sex, body mass index, fracture configuration, and type of surgery performed. Predefined perioperative outcomes were recorded within a 30-day postoperative period. Univariate and multivariate analysis were used to identify risk factors associated with increased risk of mortality.Aims
Methods
Aims. The aims of this study were to characterize antibiotic choices for perioperative total knee arthroplasty (TKA) and total hip arthroplasty (THA)
We studied the impact of direct anterior (DA) A total of 6086 consecutive patients undergoing primary total hip arthroplasty (THA) at a single institution between 2013 and 2016 were retrospectively evaluated. Data obtained from electronic patient medical records included age, sex, body mass index (BMI), medical comorbidities, surgical approach, and presence of deep PJI. There were 3053 male patients (50.1%) and 3033 female patients (49.9%). The mean age and BMI of the entire cohort was 62.7 years (18 to 102, Aims
Patients and Methods
We compared implant and patient survival following intraoperative periprosthetic femoral fractures (IOPFFs) during primary total hip arthroplasty (THA) with matched controls. This retrospective cohort study compared 4831 hips with IOPFF and 48 154 propensity score matched primary THAs without IOPFF implanted between 2004 and 2016, which had been recorded on a national joint registry. Implant and patient survival rates were compared between groups using Cox regression.Aims
Patients and Methods
A variety of surgical approaches are used for total hip arthroplasty (THA), all with reported advantages and disadvantages. A number of common complications can occur following THA regardless of the approach used. The purpose of this study was to compare five commonly used surgical approaches with respect to the incidence of surgery-related complications. The electronic medical records of all patients who underwent primary elective THA at a single large-volume arthroplasty centre, between 2011 and 2016, with at least two years of follow-up, were reviewed. After exclusion, 3574 consecutive patients were included in the study. There were 1571 men (44.0%) and 2003 women (56.0%). Their mean age and body mass index (BMI) was 63.0 years (Aims
Patients and Methods
Increasingly, patients with bilateral hip arthritis wish to undergo staged total hip arthroplasty (THA). With the rise in demand for arthroplasty, perioperative risk assessment and counselling is crucial for shared decision making. However, it is unknown if complications that occur after a unilateral hip arthroplasty predict complications following surgery of the contralateral hip. We used nationwide linked discharge data from the Healthcare Cost and Utilization Project between 2005 and 2014 to analyze the incidence and recurrence of complications following the first- and second-stage operations in staged bilateral total hip arthroplasty (BTHAs). Complications included perioperative medical adverse events within 30 to 60 days, and infection and mechanical complications within one year. Conditional probabilities and odds ratios (ORs) were calculated to determine whether experiencing a complication after the first stage of surgery increased the risk of developing the same complication after the second stage.Aims
Patients and Methods
The advent of trabecular metal (TM) augments has revolutionized
the management of severe bone defects during acetabular reconstruction.
The purpose of this study was to evaluate patients undergoing revision
total hip arthroplasty (THA) with the use of TM augments for reconstruction
of Paprosky IIIA and IIIB defects. A retrospective study was conducted at four centres between August
2008 and January 2015. Patients treated with TM augments and TM
shell for a Paprosky grade IIIA or IIIB defect, in the absence of
pelvic discontinuity, and who underwent revision hip arthroplasty
with the use of TM augments were included in the study. A total
of 41 patients with minimum follow-up of two years were included
and evaluated using intention-to-treat analysis.Aims
Patients and Methods
The aim of this study was to compare early functional and health
related quality of life outcomes (HRQoL) in patients who have undergone
total hip arthroplasty (THA) using a bone conserving short stem
femoral component and those in whom a conventional length uncemented
component was used. Outcome was assessed using a validated performance
based outcome instrument as well as patient reported outcome measures
(PROMs). We prospectively analysed 33 patients whose THA involved a contemporary
proximally porous coated tapered short stem femoral component and
53 patients with a standard conventional femoral component, at a
minimum follow-up of two years. The mean follow-up was 31.4 months
(24 to 39). Patients with poor proximal femoral bone quality were
excluded. The mean age of the patients was 66.6 years (59 to 77)
and the mean body mass index was 30.2 kg/m2 (24.1 to
41.0). Outcome was assessed using the Oxford Hip Score (OHS) and
the University College Hospital (UCH) hip score which is a validated
performance based instrument. HRQoL was assessed using the EuroQol
5D (EQ-5D).Aims
Patients and Methods
The World Health Organization (WHO) and the Centre
for Disease Control and Prevention (CDC) recently published guidelines
for the prevention of surgical site infection. The WHO guidelines,
if implemented worldwide, could have an immense impact on our practices
and those of the CDC have implications for healthcare policy in
the United States. Our aim was to review the strategies for prevention of periprosthetic
joint infection in light of these and other recent guidelines. Cite this article:
Advocates of debridement, antibiotics and implant retention (DAIR)
in hip periprosthetic joint infection (PJI) argue that a procedure
not disturbing a sound prosthesis-bone interface is likely to lead
to better survival and functional outcome compared with revision.
This case-control study aims were to compare outcome of DAIRs for
infected primary total hip arthroplasty (THA) with outcomes following
primary THA and two-stage revision of infected primary THAs. We retrospectively reviewed all DAIRs, performed for confirmed
infected primary hip arthropasty (n = 82) at out institution, between
1997 and 2013. Data recorded included full patient information and
type of surgery. Outcome measures included complications, mortality,
implant survivorship and functional outcome. Outcome was compared with
two control groups matched for gender and age; a cohort of primary
THAs (n = 120) and a cohort of two-stage revisions for infection
(n = 66).Aims
Patients and Methods
Our aim was to report survivorship data and lessons learned with
the Corail/Pinnacle cementless total hip arthroplasty (THA) system. Between August 2005 and March 2015, a total of 4802 primary cementless
Corail/Pinnacle THAs were performed in 4309 patients. In March 2016,
we reviewed these hips from a prospectively maintained database.Aims
Patients and Methods
Arthroscopy of the native hip is an established diagnostic and therapeutic procedure. Its application in the symptomatic replaced hip is still being explored. We describe the use of arthroscopy of the hip in 24 symptomatic patients following total hip replacement, resurfacing arthroplasty of the hip and partial resurfacing (study group), and compared it with arthroscopy of the native hip in 24 patients (control group). A diagnosis was made or confirmed at arthroscopy in 23 of the study group and a therapeutic arthroscopic intervention resulted in relief of symptoms in ten of these. In a further seven patients it led to revision hip replacement. In contrast, arthroscopy in the control group was diagnostic in all 24 patients and the resulting arthroscopic therapeutic intervention provided symptomatic relief in 21. The mean operative time in the study group (59.7 minutes (35 to 93)) was less than in the control group (71 minutes (40 to 100), p = 0.04) but the arthroscopic approach was more difficult in the arthroplasty group. We suggest that arthroscopy has a role in the management of patients with a symptomatic arthroplasty when other investigations have failed to provide a diagnosis.
Total hip replacement (THR) has been shown to
be a cost-effective procedure. However, it is not risk-free. Certain conditions,
such as diabetes mellitus, are thought to increase the risk of complications.
In this study we have evaluated the prevalence of diabetes mellitus
in patients undergoing THR and the associated risk of adverse operative
outcomes. A meta-analysis and systematic review were conducted according
to the guidelines of the meta-analysis of observational studies
in epidemiology. Inclusion criteria were observational studies reporting
the prevalence of diabetes in the study population, accompanied
by reports of at least one of the following outcomes: venous thromboembolic
events; acute coronary events; infections of the urinary tract,
lower respiratory tract or surgical site; or requirement for revision
arthroplasty. Altman and Bland’s methods were used to calculate differences
in relative risks. The prevalence of diabetes mellitus was found
to be 5.0% among patients undergoing THR, and was associated with
an increased risk of established surgical site infection (odds ratio
(OR) 2.04 (95% confidence interval (CI) 1.52 to 2.76)), urinary
infection (OR 1.43 (95% CI 1.33 to 1.55)) and lower respiratory
tract infections (OR 1.95 (95% CI 1.61 to 2.26)). Diabetes mellitus
is a relatively common comorbidity encountered in THR. Diabetic
patients have a higher rate of developing both surgical site and
non-surgical site infections following THR. Cite this article:
We prospectively assessed the efficacy of a ceramic-on-metal
(CoM) hip bearing with uncemented acetabular and femoral components
in which cobalt–chrome acetabular liners and alumina ceramic heads
were used. The cohort comprised 94 total hip replacements (THRs) in 83 patients
(38 women and 45 men) with a mean age of 58 years (42 to 70). Minimum
follow-up was two years. All patients had pre- and post-operative
assessment using the Western Ontario and McMaster Universities osteoarthritis
index (WOMAC), Oxford hip score and Short-Form 12 scores. All showed
a statistically significant improvement from three months post-operatively
onwards (all p <
0.001). After two years whole blood metal ion levels were measured and
chromosomal analysis was performed. The levels of all metal ions
were elevated except vanadium. Levels of chromium, cobalt, molybdenum
and titanium were significantly higher in patients who underwent
bilateral THR compared with those undergoing unilateral THR (p <
0.001).
Chromosomal analysis demonstrated both structural and aneuploidy
mutations. There were significantly more breaks and losses than
in the normal population (p <
0.001). There was no significant
difference in chromosomal aberration between those undergoing unilateral
and bilateral procedures (all analyses p ≥ 0.62). The use of a CoM THR is effective clinically in the short-term,
with no concerns, but the significance of high metal ion levels
and chromosomal aberrations in the long-term remains unclear. Cite this article:
Despite the worldwide usage of the cemented Contemporary
acetabular component (Stryker), no published data are available
regarding its use in patients aged <
50 years. We undertook a
mid- to long-term follow-up study, including all consecutive patients
aged
<
50 years who underwent a primary total hip replacement using
the Contemporary acetabular component with the Exeter cemented stem
between January 1999 and January 2006. There were 152 hips in 126
patients, 61 men and 65 women, mean age at surgery 37.6 years (16
to 49 yrs). One patient was lost to follow-up. Mean clinical follow-up of all implants was 7.6 years (0.9 to
12.0). All clinical questionnaire scores, including Harris hip score,
Oxford hip score and several visual analogue scales, were found
to have improved. The eight year survivorship of all acetabular
components for the endpoints revision for any reason or revision
for aseptic loosening was 94.4% (95% confidence interval (CI) 89.2
to 97.2) and 96.4% (95% CI 91.6 to 98.5), respectively. Radiological follow-up
was complete for 146 implants. The eight year survival for the endpoint
radiological loosening was 93.1% (95% CI 86.2 to 96.6). Three surviving
implants were considered radiologically loose but were asymptomatic.
The presence of acetabular osteolysis (n = 17, 11.8%) and radiolucent
lines (n = 20, 13.9%) in the 144 surviving cups indicates a need
for continued observation in the second decade of follow-up in order
to observe their influence on long-term survival. The clinical and radiological data resulting in a ten-year survival
rate >
90% in young patients support the use of the Contemporary
acetabular component in this specific patient group. Cite this article:
The purpose of this study was to examine the
complications and outcomes of total hip replacement (THR) in super-obese
patients (body mass index (BMI) >
50 kg/m2) compared
with class I obese (BMI 30 to 34.9 kg/m2) and normal-weight
patients (BMI 18.5 to 24.9 kg/m2), as defined by the
World Health Organization. A total of 39 THRs were performed in 30 super-obese patients
with a mean age of 53 years (31 to 72), who were followed for a
mean of 4.2 years (2.0 to 11.7). This group was matched with two
cohorts of normal-weight and class I obese patients, each comprising
39 THRs in 39 patients. Statistical analysis was performed to determine differences
among these groups with respect to complications and satisfaction
based on the Western Ontario and McMaster Universities (WOMAC) osteoarthritis
index, the Harris hip score (HHS) and the Short-Form (SF)-12 questionnaire. Super-obese patients experienced significantly longer hospital
stays and higher rates of major complications and readmissions than
normal-weight and class I obese patients. Although super-obese patients
demonstrated reduced pre-operative and post-operative satisfaction
scores, there was no significant difference in improvement, or change in
the score, with respect to HHS or the WOMAC osteoarthritis index. Super-obese patients obtain similar satisfaction outcomes as
class I obese and normal-weight patients with respect to improvement
in their scores. However, they experience a significant increase
in length of hospital stay and major complication and readmission
rates. Cite this article:
Autologous retransfusion and no-drainage are
both blood-saving measures in total hip replacement (THR). A new combined
intra- and post-operative autotransfusion filter system has been
developed especially for primary THR, and we conducted a randomised
controlled blinded study comparing this with no-drainage. A total of 204 THR patients were randomised to autologous blood
transfusion (ABT)
(n = 102) or no-drainage (n = 102). In the ABT group, a mean of
488 ml ( The use of a new intra- and post-operative autologous blood transfusion
filter system results in less total blood loss and a smaller maximum
decrease in haemoglobin levels than no-drainage following primary
THR. Cite this article:
Patients from a randomised trial on resurfacing
hip arthroplasty (RHA) (n = 36, 19 males; median age 57 years, 24
to 65) comparing a conventional 28 mm metal-on-metal total hip arthroplasty
(MoM THA) (n = 28, 17 males; median age 59 years, 37 to 65) and
a matched control group of asymptomatic patients with a 32 mm ceramic-on-polyethylene
(CoP) THA (n = 33, 18 males; median age 63 years, 38 to 71) were
cross-sectionally screened with metal artefact reducing sequence-MRI
(MARS-MRI) for pseudotumour formation at a median of 55 months (23
to 72) post-operatively. MRIs were scored by consensus according
to three different classification systems for pseudotumour formation. Clinical scores were available for all patients and metal ion
levels for MoM bearing patients. Periprosthetic lesions with a median volume of 16 mL (1.5 to
35.9) were diagnosed in six patients in the RHA group (17%), one
in the MoM THA group (4%) and six in the CoP group (18%). The classification
systems revealed no clear differences between the groups. Solid
lesions (n = 3) were exclusively encountered in the RHA group. Two patients
in the RHA group and one in the MoM THA group underwent a revision
for pseudotumour formation. There was no statistically significant
relationship between clinical scoring, metal ion levels and periprosthetic
lesions in any of the groups. Periprosthetic fluid collections are seen on MARS-MRI after conventional
CoP THA and RHA and may reflect a soft-tissue collection or effusion. Currently available MRI classification systems seem to score
these collections as pseudotumours, causing an-overestimatation
of the incidence of pseudotumours. Cite this article:
The aim of this prospective randomised study
was to compare the clinical and radiological results of a cemented
all-polyethylene Ultima acetabular component with those of a cementless
porous-coated acetabular component (PFC) following total hip replacement
(THR). A total of 287 patients received either a polyethylene acetabular
component (group A) or a cobalt–chromium porous-coated component
(group B) with an identical cemented femoral component and 28 mm
cobalt-chromium head, thus making it the largest study of its type.
Patients were evaluated radiologically and clinically using the
Harris hip score (HHS). Group A comprised 183 patients (73 male,
110 female) with a mean age of
71.3 years (55 to 89). Group B comprised 104 patients (48 male,
56 female) with a mean age of 69.8 years (56 to 89). A total of
16 patients (13 in Group A, three in Group B) did not have post-operative
data for analysis. The mean follow-up in group A was 7.52 years
(0.4 to 15.0) and in Group B 7.87 years (0.5 to 14.0). At final follow-up the mean HHS was similar between groups A
and B (74.5 (25 to 100) and 78.0 (37 to 100), respectively; p =
0.068). The total number of revisions for any cause was 28, 17 of
which were in group A and 11 in group B. The ten-year survivorship
was 86.8% (95% confidence interval (CI) 78.4 to 92.1) and 89.2%
(95% CI 78.3 to 94.8) for groups A and B, respectively (log-rank
p-value = 0.938). A total of 20 cemented and two cementless acetabular
components had evidence of acetabular radiolucencies or acetabular
component migration at last follow-up (p = 0.001). These results indicate that patients with a cemented all-polyethylene
and cementless porous-coated polyethylene lined acetabular component
have similar long-term clinical outcomes.
Periprosthetic joint infection (PJI) is a devastating
complication which can follow a total joint arthroplasty (TJA).
Although rare, this ongoing threat undermines the success of TJA,
a historically reputable procedure. It has haunted the orthopedic
community for decades and several ongoing studies have provided
insights and new approaches to effectively battle this multilayered
problem.
We determined the midterm survival, incidence
of peri-prosthetic fracture and the enhancement of the width of
the femur when combining struts and impacted bone allografts in
24 patients (25 hips) with severe femoral bone loss who underwent
revision hip surgery. The pre-operative diagnosis was aseptic loosening
in 16 hips, second-stage reconstruction in seven, peri-prosthetic
fracture in one and stem fracture in one hip. A total of 14 hips
presented with an Endoklinik grade 4 defect and 11 hips a grade
3 defect. The mean pre-operative Merle D’Aubigné and Postel score
was 5.5 points (1 to 8). The survivorship was 96% (95% confidence interval 72 to 98) at
a mean of 54.5 months (36 to 109). The mean functional score was
17.3 points (16 to 18). One patient in which the strut did not completely
bypass the femoral defect was further revised using a long cemented
stem due to peri-prosthetic fracture at six months post-operatively.
The mean subsidence of the stem was 1.6 mm (1 to 3). There was no
evidence of osteolysis, resorption or radiolucencies during follow-up
in any hip. Femoral width was enhanced by a mean of 41% (19% to
82%). A total of 24 hips had partial or complete bridging of the
strut allografts. This combined biological method was associated with a favourable
survivorship, a low incidence of peri-prosthetic fracture and enhancement
of the width of the femur in revision total hip replacement in patients
with severe proximal femoral bone loss.
We describe the results of cemented total hip replacement in 23 patients (23 hips) with active tuberculous arthritis of the hip with a mean follow-up of 4.7 years (4 to 7). In two patients the diagnosis was proved by pre-operative biopsy, whereas all others were diagnosed on a clinicoradiological basis with confirmation obtained by histopathological examination and polymerase chain reaction of tissue samples taken at the time of surgery. All patients received chemotherapy for at least three months before surgery and treatment was continued for a total of 18 months. Post-operative dislocation occurred in one patient and was managed successfully by closed reduction. No reactivation of the infection or loosening of the implant was recorded and function of the hip improved in all patients. Total hip replacement in the presence of active tuberculous arthritis of the hip is a safe procedure when pre-operative chemotherapy is commenced and continued for an extended period after operation.
We report the outcome of total hip replacement in 29 failed metal-on-metal resurfacing hip replacements in which the primary surgery was performed between August 1995 and February 2005. The mean length of follow-up was five years (1.7 to 11.7). Of the 29 hip resurfacings, 19 acetabular components and all the femoral components were revised (28 uncemented stems and one cemented stem). There were no deaths and none of the patients was lost to follow-up. None of the hips underwent any further revision. The results of the revision resurfacing group were compared with those of a control group of age-matched patients. In the latter group there were 236 primary total hip replacements and 523 resurfacings performed during the same period by the same surgeons. The outcome of the revision resurfacing group was comparable with that of the stemmed primary hip replacement group but was less good than that of the primary hip resurfacing group. Long-term follow-up is advocated to monitor the outcome of these cases.
Revision surgery of the hip was performed on 114 hips using an extensively porous-coated femoral component. Of these, 95 hips (94 patients) had a mean follow-up of 10.2 years (5 to 17). No cortical struts were used and the cortical index and the femoral cortical width were measured at different levels. There were two revisions for aseptic loosening. Survivorship at 12 years for all causes of failure was 96.9% (95% confidence interval 93.5 to 100) in the best-case scenario. Fibrous or unstable fixation was associated with major bone defects. The cortical index (p = 0.045) and the lateral cortical thickness (p = 0.008) decreased at the proximal level over time while the medial cortex increased (p = 0.001) at the proximal and distal levels. An increase in the proximal medial cortex was found in patients with an extended transtrochanteric osteotomy (p = 0.026) and in those with components shorter than 25 cm (p = 0.008). The use of the extensively porous-coated femoral component can provide a solution for difficult cases in revision surgery. Radiological bony ingrowth is common. Although without clinical relevance at the end of follow-up, the thickness of the medial femoral cortex often increased while that of the lateral cortex decreased. In cases in which a shorter component was used and in those undertaken using an extended trochanteric osteotomy, there was a greater increase in thickness of the femoral cortex over time.
We undertook a randomised controlled trial to compare the outcomes of skin adhesive and staples for skin closure in total hip replacement. The primary outcome was the cosmetic appearance of the scar at three months using a surgeon-rated visual analogue scale. In all, 90 patients were randomised to skin closure using either skin adhesive (n = 45) or staples (n = 45). Data on demographics, surgical details, infection and oozing were collected during the in-patient stay. Further data on complications, patient satisfaction and evaluation of cosmesis were collected at three-month follow-up, and a photograph of the scar was taken. An orthopaedic and a plastic surgeon independently evaluated the cosmetic appearance of the scars from the photographs. No significant difference was found between groups in the cosmetic appearance of scars at three months (p = 0.172), the occurrence of complications (p = 0.3), or patient satisfaction (p = 0.42). Staples were quicker and easier to use than skin adhesive and also less expensive. Skin adhesive and surgical staples are both effective skin closure methods in total hip replacement.
We report the outcome at a minimum of five years of 110 consecutive metal-on-metal Birmingham Hip Resurfacing arthroplasties in 98 patients. The procedures were performed between October 1999 and June 2002 by one surgeon. All patients were followed up clinically and radiologically. The mean follow-up was 71 months (60 to 93). Revision of either component was defined as failure. The mean Harris Hip score at follow-up was 96.4 (53 to 100). The mean Oxford hip score was 41.9 (16 to 57) pre-operatively and 15.4 (12 to 49) post-operatively (p <
0.001). The mean University of California Los Angeles activity score was 3.91 (1 to 10) pre-operatively and 7.5 (4 to 10) post-operatively (p <
0.001). There were four failures giving a survival at five years of 96.3% (95% confidence interval 92.8 to 99.8). When applying a new method to estimate narrowing of the femoral neck we identified a 10% thinning of the femoral neck in 16 hips (14.5%), but the relevance of this finding to the long-term outcome remains unclear. These good medium-term results from an independent centre confirm the original data from Birmingham.
Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless, polished, tapered femoral stem between 1992 and 1998. The mean follow-up period was 10.5 years (6.3 to 14.1). The median pre-operative bone defect score was 3 (interquartile range (IQR) 2 to 3) using the Endo-Klinik classification. The median subsidence at one year post-operatively was 2 mm (IQR 1 to 3). At the final review the median Harris hip score was 80.6 (IQR 67.6 to 88.9) and the median subsidence 2 mm (IQR 1 to 4). Incorporation of the allograft into trabecular bone and secondary remodelling were noted radiologically at the final follow-up in 87% (393 of 452 zones) and 40% (181 of 452 zones), respectively. Subsidence of the Exeter stem correlated with the pre-operative Endo-Klinik bone loss score (p = 0.037). The degree of subsidence at one year had a strong association with long-term subsidence (p <
0.001). There was a significant correlation between previous revision surgery and a poor Harris Hip score (p = 0.028), and those who had undergone previous revision surgery for infection had a higher risk of complications (p = 0.048). Survivorship at 10.5 years with any further femoral operation as the end-point was 92% (95% confidence interval 82 to 97).
We report the clinical and radiological outcome at ten years of 104 primary total hip replacements (100 patients) using the Metasul metal-on-metal bearing. Of these, 52 had a cemented Stuehmer-Weber polyethylene acetabular component with a Metasul bearing and 52 had an uncemented Allofit acetabular component with a Metasul liner. A total of 15 patients (16 hips) died before their follow-up at ten years and three were lost to follow-up. The study group therefore comprised 82 patients (85 hips). The mean Oxford score at ten years was 20.7 (12 to 42). Six of 85 hips required revision surgery. One was performed because of infection, one for aseptic loosening of the acetabular component and four because of unexplained pain. Histological examination showed an aseptic lymphocytic vasculitis associated lesion-type tissue response in two of these. Continued follow-up is advocated in order to monitor the long-term performance of the Metasul bearing and tissue responses to metal debris.
We have managed 27 patients (16 women and 11 men) with a mean age of 68.4 years (50 to 84), with failed total hip replacement and severe proximal femoral bone loss by revision using a distal fix/proximal wrap prosthesis. The mean follow-up was for 55.3 months (25 to 126). The mean number of previous operations was 2.2 (1 to 4). The mean Oxford hip score decreased from 46.2 (38 to 60) to 28.5 (17 to 42) (paired t-test, p <
0.001) and the mean Harris Hip score increased from 30.4 (3 to 57.7) to 71.7 (44 to 99.7) (paired t-test, p <
0.001). There were two dislocations, and in three patients we failed to eradicate previous infection. None required revision of the femoral stem. This technique allows instant distal fixation while promoting biological integration and restoration of bone stock. In the short term, the functional outcome is encouraging and the complication rates acceptable in this difficult group of patients.
Ultra-high-molecular-weight polyethylene sterilised in the absence of air and highly cross-linked polyethylene have been used to avoid osteolysis and loosening in total hip replacement. Our prospective randomised study has assessed the results using two different polyethylenes associated with the same prosthetic design. We assessed 45 Allofit acetabular components with a Sulene-polyethylene liner of conventional polyethylene gamma sterilised with nitrogen and 45 Allofit acetabular components with a Durasul-polyethylene liner sterilised in ethylene oxide, both matched with an Alloclassic stem with a 28 mm modular femoral head. The prostheses were implanted between May 1999 and December 2001. The mean follow-up was for 66.3 months (60 to 92). The linear penetration of the femoral head was estimated at 6 weeks, at 6 and 12 months and annually thereafter from standardised digitised radiographs using image-analysis software. There was no loosening of any prosthetic component. There were no radiolucent lines or osteolysis. The mean rate of penetration calculated from regression analysis during the first five years was 38 μm/year (
We reviewed 111 hemispherical Duraloc series-500 acetabular components with a minimum follow-up of 12 years. The mean clinical and radiological follow-up was 13.4 years (12 to 15). A Profile hydroxyapatite-coated anatomical femoral component was used in each case. Six patients had a late dislocation, for whom the polyethylene liner was exchanged. Each acetabular component was well fixed and all femoral components showed signs of bone ingrowth. The mean rate of femoral head penetration was 0.10 mm/year (0.021 to 0.481). The probability of not developing femoral cortical hypertrophy and proximal osteopenia by 12 years was 80.2% (95% confidence interval, 72.7 to 87.6) and 77.5% (95% confidence interval, 69.7 to 85.2), respectively. Despite these good clinical results, further follow-up is needed to determine whether these prostheses will loosen with time.
We report the long-term outcome of a modified second-generation cementing technique for fixation of the acetabular component of total hip replacement. An earlier report has shown the superiority of this technique assessed by improved survival compared with first-generation cementing. The acetabular preparation involved reaming only to the subchondral plate, followed by impaction of the bone in the anchorage holes. Between 1978 and 1993, 287 total hip replacements were undertaken in 244 patients with a mean age of 65.3 years (21 to 90) using a hemispherical Weber acetabular component with this modified technique for cementing and a cemented femoral component. The survival with acetabular revision for aseptic loosening as the endpoint was 99.1% (95% confidence interval 97.9 to 100 after ten years and 85.5% (95% confidence interval 74.7 to 96.2) at 20 years. Apart from contributing to a long-lasting fixation of the component, this technique also preserved bone, facilitating revision surgery when necessary.
The dysplasia cup, which was devised as an adjunct to the Birmingham Hip Resurfacing system, has a hydroxyapatite-coated porous surface and two supplementary neutralisation screws to provide stable primary fixation, permit early weight-bearing, and allow incorporation of morcellised autograft without the need for structural bone grafting. A total of 110 consecutive dysplasia resurfacing arthroplasties in 103 patients (55 men and 48 women) performed between 1997 and 2000 was reviewed with a minimum follow-up of six years. The mean age at operation was 47.2 years (21 to 62) and 104 hips (94%) were Crowe grade II or III. During the mean follow-up of 7.8 years (6 to 9.6), three hips (2.7%) were converted to a total hip replacement at a mean of 3.9 years (2 months to 8.1 years), giving a cumulative survival of 95.2% at nine years (95% confidence interval 89 to 100). The revisions were due to a fracture of the femoral neck, a collapse of the femoral head and a deep infection. There was no aseptic loosening or osteolysis of the acetabular component associated with either of the revisions performed for failure of the femoral component. No patient is awaiting a revision. The median Oxford hip score in 98 patients with surviving hips at the final review was 13 and the 10th and the 90th percentiles were 12 and 23, respectively.
Metal-on-metal hip resurfacing was performed for developmental dysplasia in 96 hips in 85 patients, 78 in women and 18 in men, with a mean age at the time of surgery of 43 years (14 to 65). These cases were matched for age, gender, operating surgeon and date of operation with a group of patients with primary osteoarthritis who had been treated by resurfacing, to provide a control group of 96 hips (93 patients). A clinical and radiological follow-up study was performed. The dysplasia group were followed for a mean of 4.4 years (2.0 to 8.5) and the osteoarthritis group for a mean of 4.5 years (2.2 to 9.4). Of the dysplasia cases, 17 (18%) were classified as Crowe grade III or IV. There were five (5.2%) revisions in the dysplasia group and none in the osteoarthritic patients. Four of the failures were due to acetabular loosening and the other sustained a fracture of the neck of femur. There was a significant difference in survival between the two groups (p = 0.02). The five-year survival was 96.7% (95% confidence interval 90.0 to 100) for the dysplasia group and 100% (95% confidence interval 100 to 100) for the osteoarthritic group. There was no significant difference in the median Oxford hip score between the two groups at any time during the study. The medium-term results of metal-on-metal hip resurfacing in all grades of developmental dysplasia are encouraging, although they are significantly worse than in a group of matched patients with osteoarthritis treated in the same manner.
We reviewed the clinical and radiological results of 131 patients who underwent acetabular revision for aseptic loosening with impacted bone allograft and a cemented acetabular component. The mean follow-up was 51.7 months (24 to 156). The mean post-operative Merle D’Aubigné and Postel scores were 5.7 points (4 to 6) for pain, 5.2 (3 to 6) for gait and 4.5 (2 to 6) for mobility. Radiological evaluation revealed migration greater than 5 mm in four acetabular components. Radiological failure matched clinical failure. Asymptomatic radiolucent lines were observed in 31 of 426 areas assessed (7%). Further revision was required in six patients (4.5%), this was due to infection in three and mechanical failure in three. The survival rate for the reconstruction was 95.8% (95% confidence interval 92.3 to 99.1) overall, and 98%, excluding revision due to sepsis. Our study, from an independent centre, has reproduced the results of the originators of the method.
We evaluated the use of a stemmed acetabular component in the treatment of severe acetabular deficiency during revision and complex primary total hip arthroplasty. There were 31 hips of which 24 were revisions (20 for aseptic loosening, four for infection) and the remainder were complex primary arthroplasties. At a mean follow-up of 10.7 years (6 to 12.8), no component had been revised for aseptic loosening; one patient had undergone a revision of the polyethylene liner for wear. There was one failure because of infection. At the latest follow-up, the cumulative survival rate for aseptic loosening, with revision being the end-point, was 100%; for radiographic loosening it was 92% and for infection and radiographic loosening it was 88%. These results justify the continued use of this stemmed component for the reconstruction of severe acetabular deficiency.
We describe the experience with the first consecutive 230 Birmingham hip resurfacings at our centre. At a mean follow-up of three years (25 to 52 months) survivorship was 99.14% with revision in one patient for a loose acetabular component and one death from unrelated causes. One patient developed a fracture of the femoral neck at six weeks which united unremarkably after a period of non-weight-bearing. The Harris hip score improved from a mean of 62.54 (8 to 92) to 97.74 (61 to 100). The mean flexion improved from 91.52° (25 to 140) to 110.41° (80 to 145). Most patients (97%) considered the outcome to be good or excellent. Our preliminary experience with this implant is encouraging and the results are superior to the earlier generation of resurfacings for the same length of follow-up.
Revision arthroplasty after infection can often be complicated by both extensive bone loss and a relatively high rate of re-infection. Using allograft to address the bone loss in such patients is controversial because of the perceived risk of bacterial infection from the use of avascular graft material. We describe 12 two-stage revisions for infection in which segmental allografts were loaded with antibiotics using iontophoresis, a technique using an electrical potential to drive ionised antibiotics into cortical bone. Iontophoresis produced high levels of antibiotic in the allograft, which eluted into the surrounding tissues. We postulate that this offers protection from infection in the high-risk peri-operative period. None of the 12 patients who had two-stage revision with iontophoresed allografts had further infection after a mean period of 47 months (14 to 78).
Bone allografts can store and release high levels of vancomycin. We present our results of a two-stage treatment for infected hip arthroplasty with acetabular and femoral impaction grafting using vancomycin-loaded allografts. We treated 29 patients (30 hips) by removal of the implants, meticulous debridement, parenteral antibiotic therapy and second-stage reconstruction using vancomycin-supplemented impacted bone allografts and a standard cemented Charnley femoral component. The mean follow-up was 32.4 months (24 to 60). Infection control was obtained in 29 cases (re-infection rate of 3.3%; 95% confidence interval 0.08 to 17) without evidence of progressive radiolucent lines, demarcation or graft resorption. One patient had a further infection ten months after revision caused by a different pathogen. Associated post-operative complications were one traumatic periprosthetic fracture at 14 months, a single dislocation in two hips and four displacements of the greater trochanter. Vancomycin-supplemented allografts restored bone stock and provided sound fixation with a low incidence of further infection.
Iontophoresis is a novel technique which may be used to facilitate the movement of antibiotics into the substance of bone using an electrical potential applied externally. We have examined the rate of early infection in allografts following application of this technique in clinical practice. A total of 31 patients undergoing revision arthroplasty or surgery for limb salvage received 34 iontophoresed sequential allografts, of which 26 survived for a minimum of two years. The mean serum antibiotic levels after operation were low (gentamicin 0.37 mg/l (0.2 to 0.5); flucloxacillin 1 mg/l (0 to 1) and the levels in the drains were high (gentamicin 40 mg/l (2.5 to 131); flucloxacillin 17 mg/l (1 to 43). There were no early deep infections. Two late infections were presumed to be haemotogenous; 28 of the 34 allografts were retained. In 12 patients with pre-existing proven infection further infection has not occurred at a mean follow-up of 51 months (24 to 82).