Aims. One goal of total hip arthroplasty is to restore normal hip anatomy.
The aim of this study was to compare displacement of the centre
of rotation (COR) using a standard reaming technique with a technique
in which the acetabulum was reamed immediately peripherally and
referenced off the rim. Patients and Methods. In the first cohort the acetabulum was reamed to the floor followed
by sequentially larger reamers. In the second cohort the acetabulum
was only reamed peripherally, starting with a reamer the same size
as the native femoral head. Anteroposterior pelvic radiographs were
analysed for acetabular floor depth and vertical and horizontal position
of the COR. Results. Horizontally, the mean medial displacement of the COR was 0.8
mm (standard deviation (. sd. ) 1.4) in the peripheral remaing
group and 5.0 mm (. sd. 3.30) in the standard reaming group
(p <
0.001). Vertically, the mean superior displacement of the
COR was 0.7 mm (. sd. 1.3) in the peripheral reaming group
and 3.7 mm (. sd. 2.6) in the standard reaming group (p <
0.001).
In the standard reaming group, there was a strong correlation between
the pre-operative acetabular floor depth and displacement of the
COR (p <
0.001). Conclusion.
We present three young men who sustained closed diaphyseal fracture of the tibia and later developed severe osteocutaneous necrosis induced by heat during intramedullary reaming. They all had a narrow medullary cavity and in all a tourniquet had been used. Each developed a pretibial cutaneous blister soon after operation. In the following month severe osteomyelitis ensued, requiring segmental resection and osteocutaneous reconstruction.
Intramedullary infection in long bones represents
a complex clinical challenge, with an increasing incidence due to the
increasing use of intramedullary fixation. We report a prospective
case series using an intramedullary reaming device, the Reamer–Irrigator–Aspirator
(RIA) system, in association with antibiotic cement rods for the
treatment of lower limb long bone infections. A total of 24 such
patients, 16 men and eight women, with a mean age of 44.5 years
(17 to 75), 14 with femoral and 10 with tibial infection, were treated
in a staged manner over a period of 2.5 years in a single referral
centre. Of these, 21 patients had had previous surgery, usually
for fixation of a fracture (seven had sustained an open fracture
originally and one had undergone fasciotomies). According to the
Cierny–Mader classification system, 18 patients were classified
as type 1A, four as 3A (discharging sinus tract), one as type 4A
and one as type 1B. Cite this article:
1. The Küntscher method of intramedullary reaming and nail fixation was applied to thirty-five cases of non-union of the lower extremity, twenty femurs and fifteen tibias. A bone graft was generally not used. 2. Most of the patients were allowed to bear weight and to exercise the joints within a few days of the operation. Additional measures such as pre-operative surgery for infection or distraction of shortening were used in some cases. 3. The pseudarthrosis healed in all cases, even when infection was present. The nails were generally not removed until after healing had occurred. 4. It is concluded that the Küntscher method gives remarkably rapid consolidation and restoration of function even in difficult cases of non-union of shaft fracture, particularly of the femur.
Orthopaedic surgery uses many varied instruments with high-speed, high-impact, thermal energy and sometimes heavy instruments, all of which potentially result in aerosolization of contaminated blood, tissue, and bone, raising concerns for clinicians’ health. This study quantifies the aerosol exposure by measuring the number and size distribution of the particles reaching the lead surgeon during key orthopaedic operations. The aerosol yield from 17 orthopaedic open surgeries (on the knee, hip, and shoulder) was recorded at the position of the lead surgeon using an Aerodynamic Particle Sizer (APS; 0.5 to 20 μm diameter particles) sampling at 1 s time resolution. Through timestamping, detected aerosol was attributed to specific procedures.Aims
Methods
Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA). A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion.Aims
Methods
The aim of the study was to investigate whether the primary stability of press-fit acetabular components can be improved by altering the impaction procedure. Three impaction procedures were used to implant acetabular components into human cadaveric acetabula using a powered impaction device. An impaction frequency of 1 Hz until complete component seating served as reference. Overimpaction was simulated by adding ten strokes after complete component seating. High-frequency implantation was performed at 6 Hz. The lever-out moment of the acetabular components was used as measure for primary stability. Permanent bone deformation was assessed by comparison of double micro-CT (µCT) measurements before and after impaction. Acetabular component deformation and impaction forces were recorded, and the extent of bone-implant contact was determined from 3D laser scans.Aims
Methods
The aims of this study were to determine the incidence and factors for developing periprosthetic joint infection (PJI) following hemiarthroplasty (HA) for hip fracture, and to evaluate treatment outcome and identify factors associated with treatment outcome. A retrospective review was performed of consecutive patients treated for HA PJI at a tertiary referral centre with a mean 4.5 years’ follow-up (1.6 weeks to 12.9 years). Surgeries performed included debridement, antibiotics, and implant retention (DAIR) and single-stage revision. The effect of different factors on developing infection and treatment outcome was determined.Aims
Methods
We report the use of a new method of locked intramedullary flexible osteosynthesis (LIFO) in the treatment of 118 unstable fractures of the femur and tibia. The implant utilises a set of flexible pins with a separate locking device for their proximal ends. The LIFO system proved capable of stabilising unstable fractures, and most of the complications occurred during the early testing. At follow-up, 19 of 21 femoral fractures had healed; one become infected and one showed defective callus. Of the 78 tibial shaft fractures, five failed to consolidate and five had inadequate callus.
We have carried out a prospective, randomised trial to measure the rise of temperature during reaming of the tibia before intramedullary nailing. We studied 34 patients with a mean age of 35.1 years (18 to 63) and mean injury severity score of 10 (9 to 13). The patients were randomised into two groups: group 1 included 18 patients whose procedure was undertaken without a tourniquet and group 2, 16 patients in whom a tourniquet was used. The temperature in the bone was measured directly by two thermocouples inserted into the cortical bone near the isthmus of the tibial diaphysis.
The direct anterior approach (DAA) for total hip arthroplasty (THA) has potential advantages over other approaches and is most commonly performed with the patient in the supine position. We describe a technique for DAA THA with the patient in the lateral decubitus position and report the early clinical and radiological outcomes, the characteristics of the learning curve, and perioperative complications. All primary DAA THAs performed in the lateral position by a single surgeon over a four-year period from the surgeon’s first case using the technique were identified from a prospectively collected database. Modified Harris Hip Scores (mHHS) were collected to assess clinical outcome, and routine radiological analysis was performed. Retrospective review of the medical records identified perioperative complications, the characteristics of the learning curve, and revisions.Aims
Methods
The purpose of this study was to compare clinical results, long-term survival, and complication rates of stemless shoulder prosthesis with stemmed anatomical shoulder prostheses for treatment of osteoarthritis and to analyze radiological bone changes around the implants during follow-up. A total of 161 patients treated with either a stemmed or a stemless shoulder arthroplasty for primary osteoarthritis of the shoulder were evaluated with a mean follow-up of 118 months (102 to 158). The Constant score (CS), the Disabilities of the Arm, Shoulder and Hand (DASH) score, and active range of motion (ROM) were recorded. Radiological analysis for bone adaptations was performed by plain radiographs. A Kaplan-Meier survivorship analysis was calculated and complications were noted.Aims
Methods
Existing literature indicates that inferiorly inclined glenoid baseplates following reverse total shoulder arthroplasty (RSA) produce better outcomes compared to superiorly inclined baseplates. We aim to compare clinical outcomes for RSAs with superiorly and neutrally/inferiorly inclined lateralized glenospheres. We retrospectively reviewed 154 consecutive patients undergoing RSA between July 2015 and July 2017 by one single-fellowship trained surgeon (AJ). Two raters (KAM and MVS) independently measured glenoid inclination in preoperative and minimum two year follow-up radiographs (anteroposterior/Grashey) using the RSA angle. Inclination was then compared to patient-reported outcomes, range of motion (ROM), and independently assessed degree of scapular notching and staging of heterotopic ossification at two year follow-up.Aims
Methods
A lack of supporting clinical studies have been published to determine the ideal length of intramedullary nail in fixation of trochanteric fractures of the hip. Nevertheless, there has been a trend to use shorter intramedullary nails for the internal fixation of trochanteric hip fractures. Our aim was to determine if the length of nail affected the outcome. We randomized 229 patients with a trochanteric hip fracture between two implants: a ‘standard’ nail of 220 mm and a shorter nail of 175 mm, which had decreased proximal angulation (4° vs 7°) and a reduced diameter at the level of the lesser trochanter. Patients were followed up for one year by a nurse blinded to the type of implant used to determine if there were differences in mobility and pain with two nail designs. Pain was assessed on a scale of 1 (none) to 8 (severe and constant) and mobility on a scale of 1 (full mobility) to 9 (immobile).Aims
Methods
The Fassier Duval (FD) rod is a third-generation telescopic implant for children with osteogenesis imperfecta (OI). Threaded fixation enables proximal insertion without opening the knee or ankle joint. We have reviewed our combined two-centre experience with this implant. In total, 34 children with a mean age of five years (1 to 14) with severe OI have undergone rodding of 72 lower limb long bones (27 tibial, 45 femoral) for recurrent fractures with progressive deformity despite optimized bone health and bisphosphonate therapy. Data were collected prospectively, with 1.5 to 11 years follow-up.Aims
Methods
We analyzed the acetabular morphology of Crowe type IV hips using CT data to identify a landmark for the ideal placement of the centre of the acetabular component, as assessed by morphometric geometrical analysis, and its reliability. A total of 52 Crowe IV hips (42 patients; seven male, 35 female; mean age 68.5 years (32 to 82)) and 50 normal hips (50 patients; eight male, 42 female; mean age 60.7 years (34 to 86)) undergoing total hip arthroplasty were retrospectively identified. In this CT-based simulation study, the acetabular component was positioned at the true acetabulum with a radiological inclination of 40° and anteversion of 20°. Acetabular shape and the position of the centre of the acetabular component were analyzed by morphometric geometrical analysis using the generalized Procrustes analysis.Aims
Patients and Methods
Double-level lengthening, bone transport, and bifocal compression-distraction are commonly undertaken using Ilizarov or other fixators. We performed double-level fixator-assisted nailing, mainly for the correction of deformity and lengthening in the same segment, using a straight intramedullary nail to reduce the time in a fixator. A total of 23 patients underwent this surgery, involving 27 segments (23 femora and four tibiae), over a period of ten years. The most common indication was polio in ten segments and rickets in eight; 20 nails were inserted retrograde and seven antegrade. A total of 15 lengthenings were performed in 11 femora and four tibiae, and 12 double-level corrections of deformity without lengthening were performed in the femur. The mean follow-up was 4.9 years (1.1 to 11.4). Four patients with polio had tibial lengthening with arthrodesis of the ankle. We compared the length of time in a fixator and the external fixation index (EFI) with a control group of 27 patients (27 segments) who had double-level procedures with external fixation. The groups were matched for the gain in length, age, and level of difficulty score.Aims
Patients and Methods
This study compared the primary stability of two commercially
available acetabular components from the same manufacturer, which
differ only in geometry; a hemispherical and a peripherally enhanced
design (peripheral self-locking (PSL)). The objective was to determine
whether altered geometry resulted in better primary stability. Acetabular components were seated with 0.8 mm to 2 mm interference
fits in reamed polyethylene bone substrate of two different densities
(0.22 g/cm3 and 0.45 g/cm3). The primary stability
of each component design was investigated by measuring the peak
failure load during uniaxial pull-out and tangential lever-out tests.Objective
Methods