Aims. When performing revision total hip arthroplasty using diaphyseal-engaging
Tapered fluted
Cemented
This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.Aims
Methods
Aims. United Classification System (UCS) B2 and B3 periprosthetic fractures in total hip arthroplasties (THAs) have been commonly managed with modular tapered stems. No study has evaluated the use of monoblock fluted tapered
Aims. Femoral stems with exchangeable (modular) necks were introduced
to offer surgeons an increased choice when determining the version,
offset and length of the femoral neck during total hip arthroplasty
(THA). It was hoped that this would improve outcomes and reduce
complications, particularly dislocation. In 2010, the Australian Orthopaedic
Association National Joint Replacement Registry (AOANJRR) first
reported an increased rate of revision after primary THA using femoral
stems with an exchangeable neck. The aim of this study was to provide
a more comprehensive up-to-date analysis of primary THA using femoral
stems with exchangeable and fixed necks. Materials and Methods. The data included all primary THA procedures performed for osteoarthritis
(OA), reported to the AOANJRR between 01 September 1999 and 31 December
2014. There were 9289 femoral stems with an exchangeable neck and
253 165 femoral stems with a fixed neck. The characteristics of
the patients and prostheses including the bearing surface and stem/neck
metal combinations were examined using Cox proportional hazard ratios
(HRs) and Kaplan-Meier estimates of survivorship. . Results. It was found that prostheses with an exchangeable neck had a
higher rate of revision and this was evident regardless of the bearing
surface or the size of the femoral head. Exchangeable neck prostheses
with a
Aims. Although good clinical outcomes have been reported for monolithic tapered, fluted,
Non-modular tapered fluted,
Aims. The aim of this study was to develop a novel computational model for estimating head/stem taper mechanics during different simulated assembly conditions. Methods. Finite element models of generic cobalt-chromium (CoCr) heads on a
Early failures have seen titanium fall from favour as a material for cemented femoral stems. Between 1989 and 1994, we performed a prospective review of a consecutive series of 122 cemented total hip replacements using the Ultima straight textured
Aims. There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs. Patients and Methods. We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling. Results. A total of 95 Exeter MoP and 249 MoM THAs were examined. The median volumetric loss from the MoM cohort was over four times larger than that from the MoP cohort (1.01 mm. 3. vs 0.23 mm. 3. , p < 0.001), despite a significantly shorter median period in vivo for the MoM group (48 months vs 90 months, p < 0.001). Multiple regression modelling indicated that the dominant variables leading to greater female taper material loss were bearing diameter (p < 0.001), larger female taper angles (p < 0.001), and male
Tapered, fluted, modular,
Aims. Dual mobility (DM) implants have been shown to reduce the dislocation rate after total hip arthroplasty (THA), but there remain concerns about the use of cobalt chrome liners inserted into titanium shells. The aim of this study was to assess the clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) at mid-term follow-up in young, active patients receiving a modular DM THA. Methods. This was a prospective study involving patients aged < 65 years, with a BMI of < 35 kg/m. 2. , and University of California, Los Angeles activity score of > 6 who underwent primary THA with a modular cobalt chrome acetabular liner, highly cross-linked polyethylene mobile bearing, and a cementless
There are limited long-term studies reporting on outcomes of the Zimmer Modular Revision (ZMR) stem, and concerns remain regarding failure. Our primary aim was to determine long-term survival free from all-cause revision and stem-related failure for this modular revision stem in revision total hip arthroplasty (THA). Secondary aims included evaluating radiological and functional outcomes. We retrospectively identified all patients in our institutional database who underwent revision THA using the ZMR system from January 2000 to December 2007. We included 106 patients (108 hips) with a mean follow-up of 14.5 years (2.3 to 22.3). Mean patient age was 69.2 years (37.0 to 89.4), and 51.9% were female (n = 55). Indications for index revision included aseptic loosening (73.1%), infection (16.7%), fracture (9.3%), and stem fracture (0.9%). Kaplan-Meier analysis was used to determine the all-cause and stem-related failure revision-free survival. At most recent follow-up, Oxford Hip Scores (OHS) were collected, and radiological stem stability was determined using the Engh classification.Aims
Methods
Peri-prosthetic femoral fracture after total hip replacement (THR) is associated with a poor outcome and high mortality. However, little is known about its long-term incidence after uncemented THR. We retrospectively reviewed a consecutive series of 326 patients (354 hips) who had received a CLS Spotorno replacement with an uncemented, straight, collarless tapered
Aims. To evaluate the hypothesis that failed osteosynthesis of periprosthetic
Vancouver type B1 fractures can be treated successfully with stem
revision using a transfemoral approach and a cementless, modular,
tapered revision stem with reproducible rates of fracture healing,
stability of the revision stem, and clinically good results. Patients and Methods. A total of 14 patients (11 women, three men) with a mean age
of 72.4 years (65 to 90) undergoing revision hip arthroplasty after
failed osteosynthesis of periprosthetic fractures of Vancouver type
B1 were treated using a transfemoral approach to remove the well-fixed
stem before insertion of a modular, fluted
The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore Modular Individualised Lower Extremity System (SMILES) rotating hinge implant in revision total knee arthroplasty (rTKA) at a tertiary unit. It is hypothesized that this fully cemented construct leads to satisfactory clinical outcomes. A retrospective consecutive study of all patients who underwent a rTKA using the fully cemented SMILES rotating hinge prosthesis between 2005 to 2018. Outcome measures included aseptic loosening, reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both prospectively collected local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.Aims
Methods
The December 2023 Hip & Pelvis Roundup360 looks at: Early hip fracture surgery is safe for patients on direct oral anticoagulants; Time to return to work by occupational class after total hip or knee arthroplasty; Is there a consensus on air travel following hip and knee arthroplasty?; Predicting whether patients will achieve minimal clinically important differences following hip or knee arthroplasty; High-dose dual-antibiotic-loaded cement for hip hemiarthroplasty in the UK (WHiTE 8): a randomized controlled trial; Vitamin E – a positive thing in your poly?; Hydroxapatite-coated femoral stems: is there a difference in fixation?
Tapered, fluted, modular,
Objectives. The use of ceramic femoral heads in total hip arthroplasty (THA) has increased due to their proven low bearing wear characteristics. Ceramic femoral heads are also thought to reduce wear and corrosion at the head-stem junction with
The aim of this study was to analyze how proximal radial neck resorption (PRNR) starts and progresses radiologically in two types of press-fit radial head arthroplasties (RHAs), and to investigate its clinical relevance. A total of 97 patients with RHA were analyzed: 56 received a bipolar RHA (Group 1) while 41 received an anatomical implant (Group 2). Radiographs were performed postoperatively and after three, six, nine, and 12 weeks, six, nine, 12, 18, and 24 months, and annually thereafter. PRNR was measured in all radiographs in the four radial neck quadrants. The Mayo Elbow Performance Score (MEPS), the abbreviated version of the Disabilities of the Arm, Shoulder, and Hand questionnaire (QuickDASH), and the patient-assessed American Shoulder and Elbow Surgeons score - Elbow (pASES-E) were used for the clinical assessment. Radiological signs of implant loosening were investigated.Aims
Methods
This study aimed to assess the carbon footprint associated with total hip arthroplasty (THA) in a UK hospital setting, considering various components within the operating theatre. The primary objective was to identify actionable areas for reducing carbon emissions and promoting sustainable orthopaedic practices. Using a life-cycle assessment approach, we conducted a prospective study on ten cemented and ten hybrid THA cases, evaluating carbon emissions from anaesthetic room to recovery. Scope 1 and scope 2 emissions were considered, focusing on direct emissions and energy consumption. Data included detailed assessments of consumables, waste generation, and energy use during surgeries.Aims
Methods
This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.Aims
Methods
Revision after failed femoral components may
be technically demanding due to loss of peri-prosthetic bone. This retrospective
study evaluated the long-term results of femoral revision using
the cementless Wagner Self-Locking stem. Between 1992 and 1998,
68 consecutive hips in 66 patients underwent femoral revision using
this implant. A total of 25 patients died from unrelated causes
without further revision; the remaining 41 hips in 41 patients (12 men
and 29 women) with a mean age of 61 years (29 to 80) were reviewed
at a mean follow-up of 13.9 years (10.4 to 15.8). A transfemoral
approach was used in 32 hips. A total of five stems required further
revision because of infection in two, progressive subsidence in
two and recurrent dislocation in one. Four hips had dislocated and
eight stems had subsided ≥ 10 mm. The mean Harris hip score improved
from 33 points pre-operatively to 75 points at final follow-up (p
<
0.001). In all, 33 stems (91.7%) showed radiological signs
of stable bone fixation. The cumulative survival rates at 15.8 years
with femoral revision for any reason and for stem failure as the
endpoints were 92.0% (95% confidence interval (CI) 86.0% to 98.4%)
and 96.6% (95% CI 92.2% to 100%), respectively. The survivorship
with revision and ≥ 10 mm migration of the stem as the endpoint
was 83.6% (95% CI 76.6% to 91.4%). This study shows quite good survival and moderate clinical outcome
when using a monoblock tapered
We describe the development and early clinical application of a ported, proximally-cemented
Aims. Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell (vs a monoblock acetabular component) have the advantage of allowing supplementary screw fixation, but the potential for corrosion between the liner and acetabulum has raised concerns. While DM prostheses have shown improved stability in patients deemed ‘high-risk’ for dislocation undergoing total hip arthroplasty (THA), their performance in young, active patients has not been reported. This study’s purpose was to assess clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) in young, active patients receiving a modular DM acetabulum and recently introduced titanium, proximally coated, tapered femoral stem design. Patients and Methods. This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m. 2. and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene mobile bearing, and cementless
We present the clinical and radiological results at a minimum follow-up of 20 years using a second-generation uncemented total hip arthroplasty (THA). These results are compared to our previously published results using a first-generation hip arthroplasty followed for 20 years. A total of 62 uncemented THAs in 60 patients were performed between 1993 and 1994. The titanium femoral component used in all cases was a Taperloc with a reduced distal stem. The acetabular component was a fully porous coated threaded hemispheric titanium shell (T-Tap ST). The outcome of every femoral and acetabular component with regard to retention or revision was determined for all 62 THAs. Complete clinical follow-up at a minimum of 20 years was obtained on every living patient. Radiological follow-up was obtained on all but one.Aims
Methods
Highly cross-linked polyethylene (HXLPE) has greatly improved the durability of total hip arthroplasty (THA) in young patients because of its improved wear characteristics. Few studies have followed this population into the second decade, and therefore the purpose of this investigation was to evaluate the clinical outcome for THA patients 50 years of age and younger at a minimum of 15 years postoperatively. The second purpose was to evaluate the radiological findings secondary to wear or mechanical failure of the implant. Between October 1999 and December 2005, 105 THAs were performed in 95 patients (53 female, 42 male) aged 50 years and younger (mean 42 years (20 to 50)). There were 87 patients (96 hips) that were followed for a minimum of 15 years (mean 17.3 years (15 to 21)) for analysis. Posterior approach was used with cementless fixation with a median head size of 28 mm. HXLPE was the acetabular bearing for all hips. Radiographs were evaluated for polyethylene wear, radiolucent lines, and osteolysis.Aims
Methods
It has been hypothesized that proximal radial neck resorption (PRNR) following press-fit radial head arthroplasty (RHA) is due to stress-shielding. We compared two different press-fit stems by means of radiographs to investigate whether the shape and size of the stems are correlated with the degree of PRNR. The radiographs of 52 RHAs were analyzed both at 14 days postoperatively and after two years. A cylindrical stem and a conical stem were implanted in 22 patients (group 1) and 30 patients (group 2), respectively. The PRNR was measured in the four quadrants of the radial neck and the degree of stem filling was calculated by analyzing the ratio between the prosthetic stem diameter (PSD) and the medullary canal diameter (MCD) at the proximal portion of the stem (level A), halfway along the stem length (level B), and distally at the stem tip (level C).Aims
Methods
The purpose of this study was to report bone adaptive changes after anatomical total shoulder arthroplasty (TSA) using a standard-length hydroxyapatite (HA)-coated humeral component, and to report on a computer-based analysis of radiographs to determine changes in peri-implant bone density objectively. A total of 44 TSAs, performed between 2011 and 2014 using a cementless standard-length humeral component proximally coated with HA, were included. There were 23 males and 21 females with a mean age of 65 years (17 to 65). All shoulders had good quality radiographs at six weeks and five years postoperatively. Three observers graded bone adaptive changes. All radiographs were uploaded into a commercially available photographic software program. The grey value density of humeral radiological areas was corrected to the grey value density of the humeral component and compared over time.Aims
Methods
We aimed to evaluate the long-term outcome of highly cross-linked polyethylene (HXLPE) cemented acetabular components and assess whether any radiolucent lines (RLLs) which arose were progressive. We retrospectively reviewed 170 patients who underwent 187 total hip arthroplasties at two hospitals with a minimum follow-up of ten years. All interventions were performed using the same combination of HXLPE cemented acetabular components with femoral stems made of titanium alloy. Kaplan-Meier survival analysis was performed for the primary endpoint of acetabular component revision surgery for any reason and secondary endpoint of the appearance of RLLs. RLLs that had appeared once were observed over time. We statistically assessed potential relationships between RLLs and a number of factors, including the technique of femoral head autografting and the Japanese Orthopaedic Association score.Aims
Methods
To present a surgically relevant update of trunnionosis. Systematic review performed April 2017.Aims
Materials and Methods
We examined radiographic polyethylene wear in 233 cemented total hip arthroplasties (201 patients) with either a metal-backed or a non-metal-backed acetabular cup. All patients had identical cemented one-piece
The aim of this study was to report the mid-term clinical outcome
of cemented unlinked J-alumina ceramic elbow (JACE) arthroplasties
when used in patients with rheumatoid arthritis (RA). We retrospectively reviewed 87 elbows, in 75 patients with RA,
which was replaced using a cemented JACE total elbow arthroplasty
(TEA) between August 2003 and December 2012, with a follow-up of
96%. There were 72 women and three men, with a mean age of 62 years
(35 to 79). The mean follow-up was nine years (2 to 14). The clinical condition
of each elbow before and after surgery was assessed using the Mayo
Elbow Performance Index (MEPI, 0 to 100 points). Radiographic loosening
was defined as a progressive radiolucent line of >1 mm that was
completely circumferential around the prosthesis.Aims
Patients and Methods
Taper junctions between modular hip arthroplasty femoral heads and stems fail by wear or corrosion which can be caused by relative motion at their interface. Increasing the assembly force can reduce relative motion and corrosion but may also damage surrounding tissues. The purpose of this study was to determine the effects of increasing the impaction energy and the stiffness of the impactor tool on the stability of the taper junction and on the forces transmitted through the patient’s surrounding tissues. A commercially available impaction tool was modified to assemble components in the laboratory using impactor tips with varying stiffness at different applied energy levels. Springs were mounted below the modular components to represent the patient. The pull-off force of the head from the stem was measured to assess stability, and the displacement of the springs was measured to assess the force transmitted to the patient’s tissues.Objectives
Methods
We conducted a prospective study of a delta ceramic total hip
arthroplasty (THA) to determine the rate of ceramic fracture, to
characterise post-operative noise, and to evaluate the mid-term
results and survivorship. Between March 2009 and March 2011, 274 patients (310 hips) underwent
cementless THA using a delta ceramic femoral head and liner. At
each follow-up, clinical and radiological outcomes were recorded.
A Kaplan-Meier analysis was undertaken to estimate survival.Aims
Patients and Methods
The aim of this study was to investigate the outcomes of Vancouver
type B2 and B3 fractures by performing a systematic review of the
methods of surgical treatment which have been reported. A systematic search was performed in Ovid MEDLINE, Embase and
the Cochrane Central Register of Controlled Trials. For inclusion,
studies required a minimum of ten patients with a Vancouver type
B2 and/or ten patients with a Vancouver type B3 fracture, a minimum
mean follow-up of two years and outcomes which were matched to the type
of fracture. Studies were also required to report the rate of re-operation
as an outcome measure. The protocol was registered in the PROSPERO
database. Aims
Materials and Methods
Cementless femoral stems are currently preferred
for total hip replacement (THR) in the United States. Improvements
in stem design, instrumentation and surgical technique have made
this technology highly successful, reproducible, and applicable
to the vast majority of patients requiring a THR. However, there
are ongoing developments in some aspects of stem design that influence
clinical results, the incidence of complications and their inherent adaptability
in accommodating the needs of individual patients. Here we examine
some of these design features. Cite this article:
We wished to investigate the influence of metal debris exposure
on the subsequent immune response and resulting soft-tissue injury
following metal-on-metal (MoM) hip arthroplasty. Some reports have
suggested that debris generated from the head-neck taper junction
is more destructive than equivalent doses from metal bearing surfaces. We investigated the influence of the source and volume of metal
debris on chromium (Cr) and cobalt (Co) concentrations in corresponding
blood and hip synovial fluid samples and the observed agglomerated
particle sizes in excised tissues using multiple regression analysis
of prospectively collected data. A total of 199 explanted MoM hips
(177 patients; 132 hips female) were analysed to determine rates
of volumetric wear at the bearing surfaces and taper junctions. Aims
Patients and Methods
Previous studies have suggested that metal-on-metal (MoM) Pinnacle (DePuy Synthes, Warsaw, Indiana) hip arthroplasties implanted after 2006 exhibit higher failure rates. This was attributed to the production of implants with reduced diametrical clearances between their bearing surfaces, which, it was speculated, were outside manufacturing tolerances. This study aimed to better understand the performance of Pinnacle Systems manufactured before and after this event. A total of 92 retrieved MoM Pinnacle hips were analyzed, of which 45 were implanted before 2007, and 47 from 2007 onwards. The ‘pre-2007’ group contained 45 implants retrieved from 21 male and 24 female patients, with a median age of 61.3 years (interquartile range (IQR) 57.1 to 65.5); the ‘2007 onwards’ group contained 47 implants retrieved from 19 male and 28 female patients, with a median age of 61.8 years (IQR 58.5 to 67.8). The volume of material lost from their bearing and taper surfaces was measured using coordinate and roundness measuring machines. These outcomes were then compared statistically using linear regression models, adjusting for potentially confounding factors.Objectives
Methods
We have encountered patients who developed large joint fluid collections with massive elevations in chromium (Cr) and cobalt (Co) concentrations following metal-on-metal (MoM) hip arthroplasties. In some cases, retrieval analysis determined that these ion concentrations could not be explained simply by the wear rates of the components. We hypothesized that these effects may be associated with aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL). We examined the influence of the ALVAL grade on synovial fluid Co and Cr concentrations following adjustment for patient and device variables, including volumetric wear rates. Initially restricting the analysis to include only patients with one MoM hip resurfacing device, we performed multiple regression analyses of prospectively collected data. We then repeated the same statistical approach using results from a larger cohort with different MoM designs, including total hip arthroplasties.Objectives
Patients and Methods
The aim of this study was to compare the survivorship and radiographic
outcomes at ten-year follow-up of three prospective consecutive
series of patients each of which received a different design of
cementless femoral components for total hip arthroplasty (THA). In Cohort 1, 91 consecutive patients (100 hips) underwent THA
with a cementless porous-coated anatomic femoral stem (PCA) between
October 1983 and January 1986. In Cohort 2, 86 consecutive patients
(100 hips) underwent THA with an extensively porous-coated cementless
femoral stem (Prodigy) between June 1994 and October 1997. In Cohort
3, 88 consecutive patients (100 hips) underwent THA with a proximally
porous-coated triple-tapered cementless stem (Summit) between April
2002 and October 2003. All three groups underwent prospective clinical
and radiographic evaluation.Aims
Patients and Methods
Patients with neuromuscular imbalance who require
total hip arthroplasty (THA) present particular technical problems
due to altered anatomy, abnormal bone stock, muscular imbalance
and problems of rehabilitation. In this systematic review, we studied articles dealing with THA
in patients with neuromuscular imbalance, published before April
2017. We recorded the demographics of the patients and the type
of neuromuscular pathology, the indication for surgery, surgical
approach, concomitant soft-tissue releases, the type of implant
and bearing, pain and functional outcome as well as complications
and survival. Recent advances in THA technology allow for successful outcomes
in these patients. Our review suggests excellent benefits for pain
relief and good functional outcome might be expected with a modest
risk of complication. Cite this article:
The use of large-diameter metal-on-metal (MoM)
components in total hip arthroplasty (THA) is associated with an increased
risk of early failure due to adverse local tissue reaction to metal
debris (ARMD) in response to the release of metal ions from the
bearing couple and/or head-neck taper corrosion. The aim of this
paper was to present a review of the incidence and natural history
of ARMD and the forms of treatment, with a focus on the need for
and extent of resection or debulking of the pseudotumour. An illustrative
case report is presented of a patient with an intra-pelvic pseudotumour
associated with a large diameter MoM THA, which was treated successfully
with revision of the bearing surface to a dual mobility couple and
retention of the well-fixed acetabular and femoral components. The
pseudotumour was left Cite this article:
The aim of this study was to assess the effect
of frictional torque and bending moment on fretting corrosion at
the taper interface of a modular femoral component and to investigate
whether different combinations of material also had an effect. The
combinations we examined were 1) cobalt–chromium (CoCr) heads on
CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic
heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem
in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and
8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14
Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending
moment by offsetting the application of axial load from the midline
in the mediolateral plane. Increments of offset equivalent to head
+ 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing
torque for all combinations of material. However, Ti stems showed
the highest corrosion currents. Increased bending moments associated
with using larger offset heads produced more corrosion: Ti stems
generally performed worse than CoCr stems. Using ceramic heads did
not prevent corrosion, but reduced it significantly in all loading
configurations. Cite this article:
We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid recurrent dislocation. Aims
Methods