We evaluated the rates of survival and cause
of revision of seven different brands of cemented primary total
knee replacement (TKR) in the Norwegian Arthroplasty Register during
the years 1994 to 2009. Revision for any cause, including resurfacing
of the patella, was the primary endpoint. Specific causes of revision
were secondary outcomes. Three posterior cruciate-retaining (PCR) fixed modular-bearing
TKRs, two fixed non-modular bearing PCR TKRs and two mobile-bearing
posterior cruciate-sacrificing TKRs were investigated in a total
of 17 782 primary TKRs. The median follow-up for the implants ranged
from 1.8 to 6.9 years. Kaplan-Meier 10-year survival ranged from
89.5% to 95.3%. Cox’s relative risk (RR) was calculated relative
to the fixed modular-bearing Profix knee (the most frequently used
TKR in Norway), and ranged from 1.1 to 2.6. The risk of revision
for aseptic tibial loosening was higher in the mobile-bearing LCS
Classic (RR 6.8 (95% confidence interval (CI) 3.8 to 12.1)), the
LCS Complete (RR 7.7 (95% CI 4.1 to 14.4)), the fixed modular-bearing
Duracon (RR 4.5 (95% CI 1.8 to 11.1)) and the fixed non-modular
bearing AGC Universal TKR (RR 2.5 (95% CI 1.3 to 5.1)), compared
with the Profix. These implants (except AGC Universal) also had
an increased risk of revision for femoral loosening (RR 2.3
(95% CI 1.1 to 4.8), RR 3.7 (95% CI 1.6 to 8.9), and RR 3.4 (95%
CI 1.1 to 11.0), respectively). These results suggest that aseptic
loosening is related to design in TKR. Cite this article:
In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes. A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone.Aims
Methods
Pelvic discontinuity is a rare but increasingly common complication of total hip arthroplasty (THA). This single-centre study evaluated the performance of custom-made triflange acetabular components in acetabular reconstruction with pelvic discontinuity by determining: 1) revision and overall implant survival rates; 2) discontinuity healing rate; and 3) Harris Hip Score (HHS). Retrospectively collected data of 38 patients (39 hips) with pelvic discontinuity treated with revision THA using a custom-made triflange acetabular component were analyzed. Minimum follow-up was two years (mean 5.1 years (2 to 11)).Aims
Methods
Aims. The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore Modular Individualised Lower Extremity System (SMILES) rotating hinge implant in revision total knee arthroplasty (rTKA) at a tertiary unit. It is hypothesized that this fully cemented construct leads to satisfactory clinical outcomes. Methods. A retrospective consecutive study of all patients who underwent a rTKA using the fully cemented SMILES rotating hinge prosthesis between 2005 to 2018. Outcome measures included aseptic loosening, reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both prospectively collected local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years. Results. Overall, 69 consecutive patients (69 knees) were included with a median age of 78 years (interquartile range 69 to 84), and there were 46 females (66.7%). Indications were septic revisions in 26 (37.7%), and aseptic aetiology in the remining 43 (62.3%). The mean follow-up was 9.7 years (4 to 18), and the overall complication was rate was 7.24%, all with patellofemoral complications. Failure rate with ‘any
Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes. A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively.Aims
Methods
Despite multiple trials and case series on hip hemiarthroplasty designs, guidance is still lacking on which implant to use. One particularly deficient area is long-term outcomes. We present over 1,000 consecutive cemented Thompson’s hemiarthroplasties over a ten-year period, recording all accessible patient and implant outcomes. Patient identifiers for a consecutive cohort treated between 1 January 2003 and 31 December 2011 were linked to radiographs, surgical notes, clinic letters, and mortality data from a national dataset. This allowed charting of their postoperative course, complications, readmissions, returns to theatre, revisions, and deaths. We also identified all postoperative attendances at the Emergency and Outpatient Departments, and recorded any subsequent skeletal injuries.Aims
Methods
Femoral cement-in-cement revision is a well described technique to reduce morbidity and complications in hip revision surgery. Traditional techniques for septic revision of hip arthroplasty necessitate removal of all bone cement from the femur. In our two centres, we have been using a cement-in-cement technique, leaving the distal femoral bone cement in selected patients for septic hip revision surgery, both for single and the first of two-stage revision procedures. A prerequisite for adoption of this technique is that the surgeon considers the cement mantle to be intimately fixed to bone without an intervening membrane between cement and host bone. We aim to report our experience for this technique. We have analyzed patients undergoing this cement-in-cement technique for femoral revision in infection, and present a consecutive series of 89 patients. Follow-up was undertaken at a mean of 56.5 months (24.0 to 134.7) for the surviving cases.Aims
Methods
The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the risk of revision for instability or loosening. A total of 202,511 primary total hip arthroplasties (THAs) with uncemented acetabular components were identified from the National Joint Registry (NJR) dataset between 2003 and 2017. The effect of liner geometry on the risk of revision for instability or loosening was investigated using competing risk regression analyses adjusting for age, sex, American Society of Anesthesiologists grade, indication, side, institution type, surgeon grade, surgical approach, head size, and polyethylene crosslinking. Stratified analyses by surgical approach were performed, including pairwise comparisons of liner geometries.Aims
Methods
Use of the direct anterior approach (DAA) for total hip arthroplasty (THA) has increased in recent years due to proposed benefits, including a lower risk of dislocation and improved early functional recovery. This study investigates the dislocation rate in a non-selective, consecutive cohort undergoing THA via the DAA without any exclusion or bias in patient selection based on habitus, deformity, age, sex, or fixation method. We retrospectively reviewed all patients undergoing THA via the DAA between 2011 and 2017 at our institution. Primary outcome was dislocation at minimum two-year follow-up. Patients were stratified by demographic details and risk factors for dislocation, and an in-depth analysis of dislocations was performed.Aims
Methods
The main aims were to identify risk factors predictive of a radiolucent line (RLL) around the acetabular component with an interface bioactive bone cement (IBBC) technique in the first year after THA, and evaluate whether these risk factors influence the development of RLLs at five and ten years after THA. A retrospective review was undertaken of 980 primary cemented THAs in 876 patients using cemented acetabular components with the IBBC technique. The outcome variable was any RLLs that could be observed around the acetabular component at the first year after THA. Univariate analyses with univariate logistic regression and multivariate analyses with exact logistic regression were performed to identify risk factors for any RLLs based on radiological classification of hip osteoarthritis.Aims
Methods
Bone allografts can store and release high levels of vancomycin. We present our results of a two-stage treatment for infected hip arthroplasty with acetabular and femoral impaction grafting using vancomycin-loaded allografts. We treated 29 patients (30 hips) by removal of the implants, meticulous debridement, parenteral antibiotic therapy and second-stage reconstruction using vancomycin-supplemented impacted bone allografts and a standard cemented Charnley femoral component. The mean follow-up was 32.4 months (24 to 60). Infection control was obtained in 29 cases (re-infection rate of 3.3%; 95% confidence interval 0.08 to 17) without evidence of progressive radiolucent lines, demarcation or graft resorption. One patient had a further infection ten months after
Revision total knee arthroplasty (TKA) is a complex
procedure which carries both a greater risk for patients and greater
cost for the treating hospital than does a primary TKA. As well
as the increased cost of peri-operative investigations, blood transfusions,
surgical instrumentation, implants and operating time, there is
a well-documented increased length of stay which accounts for most
of the actual costs associated with surgery. We compared revision surgery for infection with
During revision procedures for aseptic reasons, there remains a suspicion that failure may have been the result of an undetected subclinical infection. However, there is little evidence available in the literature about unexpected positive results in presumed aseptic revision spine surgery. The aims of our study were to estimate the prevalence of unexpected positive culture using sonication and to evaluate clinical characteristics of these patients. All patients who underwent a revision surgery after instrumented spinal surgery at our institution between July 2014 and August 2016 with spinal implants submitted for sonication were retrospectively analyzed. Only revisions presumed as aseptic are included in the study. During the study period, 204 spinal revisions were performed for diagnoses other than infection. In 38 cases, sonication cultures were not obtained, leaving a study cohort of 166 cases. The mean age of the cohort was 61.5 years (Aims
Patients and Methods
We present the clinical and radiological results at a minimum
follow-up of five years for patients who have undergone multiple
cement-in-cement revisions of their femoral component at revision
total hip arthroplasty (THA). We reviewed the outcome on a consecutive series of 24 patients
(10 men, 14 women) (51 procedures) who underwent more than one cement-in-cement
revision of the same femoral component. The mean age of the patients was
67.5 years (36 to 92) at final follow-up. Function was assessed using the original Harris hip score (HHS),
Oxford Hip Score (OHS) and the Merle D’Aubigné Postel score (MDP).Aims
Patients and Methods
The aims of this study were to evaluate the clinical and radiological
outcomes of the Universal-2 total wrist arthroplasty (TWA) in patients
with rheumatoid arthritis. This was a retrospective review of all 95 Universal-2 TWAs which
were performed in our institution between 2003 to 2012 in patients
with rheumatoid arthritis. A total of six patients were lost to
follow-up and two died of unrelated causes. A total of ten patients
had bilateral procedures. Accordingly, 75 patients (85 TWAs) were
included in the study. There were 59 women and 16 men with a mean
age of 59 years (26 to 86). The mean follow-up was 53 months (24
to 120). Clinical assessment involved recording pain on a visual
analogue score, range of movement, grip strength, the Quick Disabilities
of the Arm, Shoulder and Hand (DASH) and Wrightington wrist scores.
Any adverse effects were documented with particular emphasis on
residual pain, limitation of movement, infection, dislocation and
the need for revision surgery. Radiographic assessment was performed pre-operatively and at
three, six and 12 months post-operatively, and annually thereafter.
Arthroplasties were assessed for distal row intercarpal fusion and
loosening. Radiolucent zones around the components were documented
according to a system developed at our institution.Aims
Patients and Methods
Failed internal fixation of a fracture of the
proximal humerus produces many challenges with limited surgical options.
The aim of this study was to evaluate the clinical outcomes after
the use of a reverse shoulder arthroplasty under these circumstances.
Between 2007 and 2012, 19 patients (15 women and four men, mean
age 66 years; 52 to 82) with failed internal fixation after a proximal
humeral fracture, underwent implant removal and reverse shoulder arthroplasty
(RSA). The mean follow-up was 36 months (25 to 60). The mean American
Shoulder and Elbow Score improved from 27.8 to 50.1 (p = 0.019).
The mean Simple Shoulder Test score improved from 0.7 to 3.2 (p
= 0.020), and the mean visual analogue scale for pain improved from
6.8 to 4.3 (p = 0.012). Mean forward flexion improved from 58.7°
to 101.1° (p <
0.001), mean abduction from 58.7° to 89.1° (p
= 0.012), mean external rotation from 10.7° to 23.1° (p = 0.043)
and mean internal rotation from buttocks to L4 (p = 0.034). A major
complication was recorded in five patients (26%) (one intra-operative
fracture, loosening of the humeral component in two and two peri-prosthetic
fractures). A total of 15 patients (79%) rated their outcome as
excellent or good, one (5%) as satisfactory, and three (16%) as
unsatisfactory. An improvement in outcomes and pain can be expected when performing
a RSA as a salvage procedure after failed internal fixation of a
fracture of the proximal humerus. Patients should be cautioned about
the possibility for major complications following this technically
demanding procedure. Cite this article:
We report a multicentre prospective consecutive
study assessing the long-term outcome of the proximally hydroxyapatite
(HA)-coated ABG II monobloc femoral component in a series of 1148
hips in 1053 patients with a mean age at surgery of 64.77 years
(22 to 80) at a mean follow-up of 10.84 years (10 to 15.25). At
latest follow-up, the mean total Harris hip score was 94.7 points
( Cite this article:
This review summarises the opinions and conclusions
reached from a symposium on infected total knee replacement (TKR)
held at the British Association of Surgery of the Knee (BASK) annual
meeting in 2011. The National Joint Registry for England and Wales
reported 5082 revision TKRs in 2010, of which 1157 (23%) were caused
by infection. The diagnosis of infection beyond the acute post-operative
stage relies on the identification of the causative organism by
aspiration and analysis of material obtained at arthroscopy. Ideal
treatment then involves a two-stage surgical procedure with extensive
debridement and washout, followed by antibiotics. An articulating
or non-articulating drug-eluting cement spacer is used prior to
implantation of the revision prosthesis, guided by the serum level
of inflammatory markers. The use of a single-stage revision is gaining popularity
and we would advocate its use in certain patients where the causative
organism is known, no sinuses are present, the patient is not immunocompromised,
and there is no radiological evidence of component loosening or
osteitis. It is our opinion that single-stage revision produces high-quality
reproducible results and will soon achieve the same widespread acceptance
as it does in infected hip arthroplasty.
The treatment of substantial proximal femoral
bone loss in young patients with developmental dysplasia of the
hip (DDH) is challenging. We retrospectively analysed the outcome
of 28 patients (30 hips) with DDH who underwent revision total hip
replacement (THR) in the presence of a deficient proximal femur,
which was reconstructed with an allograft prosthetic composite.
The mean follow-up was 15 years (8.5 to 25.5). The mean number of
previous THRs was three (1 to 8). The mean age at primary THR and
at the index reconstruction was 41 years (18 to 61) and 58.1 years
(32 to 72), respectively. The indication for revision included mechanical
loosening in 24 hips, infection in three and peri-prosthetic fracture
in three. Six patients required removal and replacement of the allograft
prosthetic composite, five for mechanical loosening and one for
infection. The survivorship at ten, 15 and 20 years was 93% (95%
confidence interval (CI) 91 to 100), 75.5% (95% CI 60 to 95) and
75.5% (95% CI 60 to 95), respectively, with 25, eight, and four
patients at risk, respectively. Additionally, two junctional nonunions
between the allograft and host femur required bone grafting and
plating. An allograft prosthetic composite affords a good long-term outcome
in the management of proximal femoral bone loss in revision THR
in patients with DDH, while preserving distal host bone.
The purpose of this study was twofold: first,
to determine whether the five-year results of hip resurfacing arthroplasty
(HRA) in Canada justified the continued use of HRA; and second,
to identify whether greater refinement of patient selection was
warranted. This was a retrospective cohort study that involved a review
of 2773 HRAs performed between January 2001 and December 2008 at
11 Canadian centres. Cox’s proportional hazards models were used
to analyse the predictors of failure of HRA. Kaplan–Meier survival
analysis was performed to predict the cumulative survival rate at
five years. The factors analysed included age, gender, body mass
index, pre-operative hip pathology, surgeon’s experience, surgical
approach, implant sizes and implant types. The most common modes
of failure were also analysed. The 2773 HRAs were undertaken in 2450 patients: 2127 in men and
646 in women. The mean age at operation was 50.5 years ( The failure rates of HRA at five years justify the ongoing use
of this technique in men. Female gender is an independent predictor
of failure, and a higher failure rate at five years in women leads
the authors to recommend this technique only in exceptional circumstances
for women. Cite this article: