Aims. This study aims to assess the relationship between history of
Peri-articular soft-tissue masses or ‘pseudotumours’
can occur after large-diameter metal-on-metal (MoM) resurfacing
of the hip and conventional total hip replacement (THR). Our aim
was to assess the incidence of
Aims. The purpose of this study was to compare two different types
of metal-on-metal (MoM) bearing for total hip arthroplasty (THA):
one with a large femoral head (38 mm to 52 mm) and the other with
a conventional femoral head (28 mm or 32 mm). We compared clinical
outcome, blood metal ion levels, and the incidence of
The use of large-diameter metal-on-metal (MoM)
components in total hip arthroplasty (THA) is associated with an increased
risk of early failure due to adverse local tissue reaction to metal
debris (ARMD) in response to the release of metal ions from the
bearing couple and/or head-neck taper corrosion. The aim of this
paper was to present a review of the incidence and natural history
of ARMD and the forms of treatment, with a focus on the need for
and extent of resection or debulking of the
We present a series of 35 patients (19 men and
16 women) with a mean age of 64 years (36.7 to 75.9), who underwent
total hip replacement using the ESKA dual-modular short stem with
metal on-polyethylene bearing surfaces. This implant has a modular
neck section in addition to the modular head. Of these patients,
three presented with increasing post-operative pain due to pseudotumour
formation that resulted from corrosion at the modular neck-stem
junction. These patients underwent further surgery and aseptic lymphocytic
vaculitis associated lesions were demonstrated on histological analysis. Retrieval analysis of two modular necks showed corrosion at the
neck-stem taper. Blood cobalt and chromium levels were measured
at a mean of nine months (3 to 28) following surgery. These were
compared with the levels in seven control patients (three men and
four women) with a mean age of 53.4 years (32.1 to 64.1), who had
an identical prosthesis and articulation but with a prosthesis that
had no modularity at neck-stem junction. The mean blood levels of
cobalt in the study group were raised at 50.75 nmol/l (5 to 145)
compared with 5.6 nmol/l (2 to 13) in control patients. Corrosion at neck-stem tapers has been identified as an important
source of metal ion release and
Aims. Hip resurfacing remains a potentially valuable surgical procedure for appropriately-selected patients with optimised implant choices. However, concern regarding high early failure rates continues to undermine confidence in use. A large contributor to failure is adverse local tissue reactions around metal-on-metal (MoM) bearing surfaces. Such phenomena have been well-explored around MoM total hip arthroplasties, but comparable data in equivalent hip resurfacing procedures is lacking. In order to define genetic predisposition, we performed a case-control study investigating the role of human leucocyte antigen (HLA) genotype in the development of
Aims. The present study investigated the five-year interval changes in
Inflammatory
The presence of
Metal-on-metal hip resurfacing is commonly performed for osteoarthritis in young active patients. We have observed cystic or solid masses, which we have called inflammatory
We compared the incidence of
The aim of this study was to establish the natural
course of unrevised asymptomatic
Aims. The aims of this study were to compare the diagnostic test characteristics
of ultrasound alone, metal artefact reduction sequence MRI (MARS-MRI)
alone, and ultrasound combined with MARS-MRI for identifying intra-operative
pseudotumours in metal-on-metal hip resurfacing (MoMHR) patients
undergoing revision surgery. . Methods. This retrospective diagnostic accuracy study involved 39 patients
(40 MoMHRs). The time between imaging modalities was a mean of 14.6
days (0 to 90), with imaging performed at a mean of 5.3 months (0.06
to 12) before revision. The prevalence of intra-operative pseudotumours
was 82.5% (n = 33). Results. Agreement with the intra-operative findings was 82.5% (n = 33)
for ultrasound alone, 87.5% (n = 35) for MARS-MRI alone, and 92.5%
(n = 37) for ultrasound and MARS-MRI combined. The diagnostic characteristics
for ultrasound alone and MARS-MRI alone reached similar sensitivities
(90.9% vs 93.9%) and positive predictive values
(PPVs; 88.2% vs 91.2%), but higher specificities
(57.1% vs 42.9%) and negative predictive values
(NPVs; 66.7% vs 50.0%) were achieved with MARS-MRI.
Ultrasound and MARS-MRI combined produced 100% sensitivity and 100%
NPV, whilst maintaining both specificity (57.1%) and PPV (91.7%). For the identification of a
Aims. The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Methods. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis. Results. Mean follow-up was 20.9 years (19.3 to 22.4). Registry data revealed that 19 hips (8.1%) had been revised and 26 patients (12%) had died from causes unrelated to the BHR. Among the remaining 189 hips, 61% were available for clinical follow-up at 20 years (n = 115) and 70% of patients had biochemical follow-up (n = 132). The cumulative implant survival rate at 20 years for male patients was 96.5% (95% confidence interval (CI) 93.5 to 99.6), and for female patients 87% (95% CI 79.7 to 94.9). The difference was statistically significant (p = 0.029). The mean Oxford Hip Score, Hip disability and Osteoarthritis Outcome Score, and Forgotten Joint Score were 45 (29 to 48), 89 (43 to 100), and 84 (19 to 100), respectively. The mean scores for each of the five domains of the EuroQol five-dimension three-level questionnaire were 1.2, 1.0, 1.2, 1.3, and 1.1, and mean overall score 82.6 (50 to 100). Ultrasound showed no
1. The case history of a haemophiliac in whom a large haematoma of the thigh was treated by amputation of the limb is described. 2. Examination of the available radiographs and of the histology led to the conclusion that the cyst was subperiosteal in origin. 3. Evidence is presented to support the hypothesis that haemophilic
Modular dual-mobility (DM) articulations are increasingly used during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS)-MRI abnormalities with serum metal ion levels in patients with DM articulations. A total of 45 patients (50 hips) with a modular DM articulation were included with mean follow-up of 3.7 years (SD 1.2). Enrolled patients with an asymptomatic, primary THA and DM articulation with over two years’ follow-up underwent MARS-MRI. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.Aims
Methods
We retrospectively studied the possibility that direct trauma to the biceps muscle might be the cause of poor elbow flexion and supination in 18 consecutive children with birth lesions of the brachial plexus who had delayed or impaired biceps recovery despite neurophysiological evidence of reinnervation. All had good shoulder and hand function at three months of age. Eight recovered a strong biceps after six months, but nine required a pectoralis minor to biceps transfer to augment elbow flexion and supination. One had a delayed but good recovery of the biceps after microsurgical reconstruction of the plexus. All had a clinical ‘pseudotumour’ in the biceps muscle, which was biopsied during pectoralis minor transfer in two patients and showed rupture and degeneration of muscle fibres with a fibro-fatty infiltrate, suggesting previous muscle trauma. Direct muscle trauma is an uncommon but important cause of delayed or impaired biceps recovery after brachial plexus birth injuries. Surgery to reinnervate the biceps muscle will not work if substantial muscle damage is present when a suitable muscle transfer should be considered.
The purpose of this study was twofold: first,
to determine whether the five-year results of hip resurfacing arthroplasty
(HRA) in Canada justified the continued use of HRA; and second,
to identify whether greater refinement of patient selection was
warranted. This was a retrospective cohort study that involved a review
of 2773 HRAs performed between January 2001 and December 2008 at
11 Canadian centres. Cox’s proportional hazards models were used
to analyse the predictors of failure of HRA. Kaplan–Meier survival
analysis was performed to predict the cumulative survival rate at
five years. The factors analysed included age, gender, body mass
index, pre-operative hip pathology, surgeon’s experience, surgical
approach, implant sizes and implant types. The most common modes
of failure were also analysed. The 2773 HRAs were undertaken in 2450 patients: 2127 in men and
646 in women. The mean age at operation was 50.5 years ( The failure rates of HRA at five years justify the ongoing use
of this technique in men. Female gender is an independent predictor
of failure, and a higher failure rate at five years in women leads
the authors to recommend this technique only in exceptional circumstances
for women. Cite this article:
We report a case of a male patient presenting
with bilateral painful but apparently well-positioned and -fixed
large-diameter metal-on-metal hip replacements four years post-operatively.
Multiple imaging modes revealed a thick-walled, cystic expansile
mass in communication with the hip joint (a pseudotumour). Implant
retrieval analysis and tissue culture eliminated high bearing wear
or infection as causes for the soft-tissue reaction, but noted marked corrosion
of the modular neck taper adaptor and corrosion products in the
tissues. Therefore, we believe corrosion products from the taper
caused by mismatch of the implant components led to pseudotumour
formation requiring revision.