Aims. The aims of the study were to report for a cohort aged younger than 40 years: 1) indications for HRA; 2) patient-reported outcomes in terms of the modified Harris Hip Score (HHS); 3) dislocation rate; and 4) revision rate. Methods. This retrospective analysis identified 267 hips from 224 patients who underwent an
Aims. Professional dancers represent a unique patient population in the setting of hip arthroplasty, given the high degree of hip strength and mobility required by their profession. We sought to determine the clinical outcomes and ability to return to professional dance after total hip arthroplasty (THA) or
Aims. Surgical approaches that claim to be minimally invasive, such as the direct anterior approach (DAA), are reported to have a clinical advantage, but are technically challenging and may create more injury to the soft-tissues during joint exposure. Our aim was to quantify the effect of soft-tissue releases on the joint torque and femoral mobility during joint exposure for
Aims. The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR)
The effects of the method of fixation and interface conditions on the biomechanics of the femoral component of the Birmingham
Objectives. Modern metal-on-metal (MoM)
Aims.
Despite the increasing interest and subsequent published literature on
Aims. Adverse local tissue reactions associated with abnormal wear considerably slowed down the general use of metal-on-metal (MoM)
The aim of this study was to evaluate the ten-year
clinical and functional outcome of
Pseudotumours are a rare complication of
Aims.
This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties:
We undertook a randomised prospective follow-up study of changes in peri-prosthetic bone mineral density (BMD) after
Aims. This study investigates the use of the metabolic equivalent of task (MET) score in a young hip arthroplasty population, and its ability to capture additional benefit beyond the ceiling effect of conventional patient-reported outcome measures. Methods. From our electronic database of 751 hip arthroplasty procedures, 221 patients were included. Patients were excluded if they had revision surgery, an alternative hip procedure, or incomplete data either preoperatively or at one-year follow-up. Included patients had a mean age of 59.4 years (SD 11.3) and 54.3% were male, incorporating 117 primary total hip and 104
Modern metal-on-metal hip resurfacing has been
widely performed in the United Kingdom for over a decade. However,
the literature reports conflicting views of the benefits: excellent
medium- to long-term results with some brands in specific subgroups,
but high failure rates and local soft-tissue reactions in others.
The National Joint Registry for England and Wales (NJR) has collected
data on all
The cause of fracture of the femoral neck after
We report the survival, radiological and functional outcomes of a single surgeon series of his first 144 consecutive Birmingham
There have been no large comparative studies of the blood levels of metal ions after implantation of commercially available
We report the findings of an independent review
of 230 consecutive Birmingham
Aims. This study presents the long-term survivorship, risk factors for prosthesis survival, and an assessment of the long-term effects of changes in surgical technique in a large series of patients treated by metal-on-metal (MoM)
Metal-on-metal (MoM)
We report an independent prospective review of the first 230 Birmingham
The major advantage of
The painful subluxed or dislocated hip in adults
with cerebral palsy presents a challenging problem. Prosthetic dislocation
and heterotopic ossification are particular concerns. We present
the first reported series of 19 such patients (20 hips) treated
with
Because the femoral head/neck junction is preserved in
We report the survival at five years of 144 consecutive metal-on-metal resurfacings of the hip implanted between August 1997 and May 1998. Failure was defined as revision of either the acetabular or femoral component for any reason during the study period. The survival at the end of five years was 98% overall and 99% for aseptic revisions only. The mean age of the patients at implantation was 52.1 years. Three femoral components failed during the first two years, two were infected and one fractured. A single stage revision was carried out in each case. No other revisions were performed or are impending. No patients were lost to follow-up. Four died from unrelated causes during the study period. This study confirms that
Objectives. To study the vascularity and bone metabolism of the femoral head/neck
following
The purpose of this study was twofold: first,
to determine whether the five-year results of
The aim of our study was to investigate whether placing of the femoral component of a
The computed neck-shaft angle and the size of the femoral component were recorded in 100 consecutive
We describe the experience with the first consecutive 230 Birmingham
The Birmingham
We inserted an electrode up the femoral neck into the femoral head of ten patients undergoing a metal-on-metal hip resurfacing arthroplasty through a posterior surgical approach and measured the oxygen concentration during the operation. In every patient the blood flow was compromised during surgery, but the extent varied. In three patients, the oxygen concentration was zero at the end of the procedure. The surgical approach caused a mean 60% drop (p <
0.005) in oxygen concentration while component insertion led to a further 20% drop (p <
0.04). The oxygen concentration did not improve significantly on wound closure. This study demonstrates that during
Aims. A contact patch to rim (CPR) distance of <
10 mm has been
associated with edge-loading and excessive wear. However, not all
arthroplasties with a low CPR distance show problems with wear.
Therefore, CPR distance may not be the only variable affecting the
post-operative metal ion concentrations. Patients and Methods. We used multiple logistic regression to determine what variables
differed between the patients who had high and low cobalt (CoS)
and chromium (CrS) serum ion concentrations within a cohort of patients
with low (<
10 mm) CPR distances. A total of 56 patients treated
with unilateral
Early failure associated with adverse reactions to metal debris is an emerging problem after
In 12 patients, we measured the oxygen concentration in the femoral head-neck junction during
Fracture of the neck of the femur after resurfacing arthroplasty usually leads to failure and conversion to a total hip replacement. We describe an intracapsular fracture of the femoral neck sustained after
We compared the medium-term clinical and radiological results of hybrid total hip replacement (THR) with metal-on-metal Birmingham
The Birmingham
This is a longitudinal study of the daily urinary output and the concentrations in whole blood of cobalt and chromium in patients with metal-on-metal resurfacings over a period of four years. Twelve-hour urine collections and whole blood specimens were collected before and periodically after a Birmingham
We report the outcome at a minimum of five years of 110 consecutive metal-on-metal Birmingham Hip Resurfacing arthroplasties in 98 patients. The procedures were performed between October 1999 and June 2002 by one surgeon. All patients were followed up clinically and radiologically. The mean follow-up was 71 months (60 to 93). Revision of either component was defined as failure. The mean Harris Hip score at follow-up was 96.4 (53 to 100). The mean Oxford hip score was 41.9 (16 to 57) pre-operatively and 15.4 (12 to 49) post-operatively (p <
0.001). The mean University of California Los Angeles activity score was 3.91 (1 to 10) pre-operatively and 7.5 (4 to 10) post-operatively (p <
0.001). There were four failures giving a survival at five years of 96.3% (95% confidence interval 92.8 to 99.8). When applying a new method to estimate narrowing of the femoral neck we identified a 10% thinning of the femoral neck in 16 hips (14.5%), but the relevance of this finding to the long-term outcome remains unclear. These good medium-term results from an independent centre confirm the original data from Birmingham.
A total of 20 pairs of fresh-frozen cadaver femurs were assigned to four alignment groups consisting of relative varus (10° and 20°) and relative valgus (10° and 20°), 75 composite femurs of two neck geometries were also used. In both the cadaver and the composite femurs, placing the component in 20° of valgus resulted in a significant increase in load to failure. Placing the component in 10° of valgus had no appreciable effect on increasing the load to failure except in the composite femurs with varus native femoral necks. Specimens in 10° of varus were significantly weaker than the neutrally-aligned specimens. The results suggest that retention of the intact proximal femoral strength occurs at an implant angulation of ≥ 142°. However, the benefit of extreme valgus alignment may be outweighed in clinical practice by the risk of superior femoral neck notching, which was avoided in this study.
This study compared the demographic, clinical and patient-reported outcomes after total hip replacement (THR) and Birmingham Hip Resurfacing (BHR) carried out by a single surgeon. Patients completed a questionnaire that included the WOMAC, SF-36 scores and comorbid medical conditions. Data were collected before operation and one year after. The outcome scores were adjusted for age, gender, comorbid conditions and, at one year, for the pre-operative scores. There were 214 patients with a THR and 132 with a BHR. Patients with a BHR were significantly younger (49 vs 67 years, p <
0.0001), more likely to be male (68% vs 42% of THR, p <
0.0001) and had fewer comorbid conditions (1.3 vs 2.0, p <
0.0001). Before operation there was no difference in WOMAC and SF-36 scores, except for function, in which patients awaiting THR were worse than those awaiting a BHR. At one year patients with a BHR reported significantly better WOMAC pain scores (p = 0.04) and in all SF-36 domains (p <
0.05). Patients undergoing BHR report a significantly greater improvement in general health compared with those with a THR.
We sought to establish the incidence of joint failure secondary to adverse reaction to metal debris (ARMD) following metal-on-metal hip resurfacing in a large, three surgeon, multicentre study involving 4226 hips with a follow-up of 10 to 142 months. Three implants were studied: the Articular Surface Replacement; the Birmingham Hip Resurfacing; and the Conserve Plus. Retrieved implants underwent analysis using a co-ordinate measuring machine to determine volumetric wear. There were 58 failures associated with ARMD. The median chromium and cobalt concentrations in the failed group were significantly higher than in the control group (p <
0.001). Survival analysis showed a failure rate in the patients with Articular Surface Replacement of 9.8% at five years, compared with <
1% at five years for the Conserve Plus and 1.5% at ten years for the Birmingham Hip Resurfacing. Two ARMD patients had relatively low wear of the retrieved components. Increased wear from the metal-on-metal bearing surface was associated with an increased rate of failure secondary to ARMD. However, the extent of tissue destruction at revision surgery did not appear to be dose-related to the volumetric wear.
Bone surface strains were measured in cadaver femora during loading prior to and after resurfacing of the hip and total hip replacement using an uncemented, tapered femoral component. In vitro loading simulated the single-leg stance phase during walking. Strains were measured on the medial and the lateral sides of the proximal aspect and the mid-diaphysis of the femur. Bone surface strains following femoral resurfacing were similar to those in the native femur, except for proximal shear strains, which were significantly less than those in the native femur. Proximomedial strains following total hip replacement were significantly less than those in the native and the resurfaced femur. These results are consistent with previous clinical evidence of bone loss after total hip replacement, and provide support for claims of bone preservation after resurfacing arthroplasty of the hip.
Between April 1999 and April 2004, 3497 Birmingham hips were inserted by 89 surgeons. Fracture of the neck of the femur occurred in 50 patients, an incidence of 1.46%. Complete data were available for 45 patients in whom the incidence was 1.91% for women and 0.98% for men. The relative risk of fracture for women The mean time to fracture was 15.4 weeks. In women the fractures occurred at a mean of 18.5 weeks and were more likely to have been preceded by a pro-dromal phase of pain and limping. In men the mean time to fracture was 13.5 weeks. Significant varus placement of the femoral component, intra-operative notching of the femoral neck and technical problems were common factors in 85% of cases.
Resurfacing arthroplasty of the hip is being used increasingly as an alternative to total hip replacement, especially for young active patients. There is concern about necrosis of the femoral head after resurfacing which can result in fracture and loosening. Most systems use a cemented femoral component, with the potential for thermal necrosis of the cancellous bone of the reamed femoral head. We used thermal probes to record temperatures close to the cement-bone interface during resurfacing arthroplasty. The maximum temperature recorded at the cement-bone interface in four cases was approximately 68°C which was higher than that reported to kill osteocytes. A modified surgical technique using insertion of a suction cannula into the lesser trochanter, generous pulsed lavage and early reduction of the joint significantly reduced the maximum recorded cancellous bone temperature to approximately 36°C in five cases (p = 0.014). We recommend the modified technique since it significantly reduces temperatures at the cement-bone interface.
Metal-on-metal hip resurfacing is commonly performed for osteoarthritis in young active patients. We have observed cystic or solid masses, which we have called inflammatory pseudotumours, arising around these devices. They may cause soft-tissue destruction with severe symptoms and a poor outcome after revision surgery. The aim of this study was to determine the incidence of and risk factors for pseudotumours that are serious enough to require revision surgery. Since 1999, 1419 metal-on-metal hip resurfacings have been implanted by our group in 1224 patients; 1.8% of the patients had a revision for pseudotumour. In this series the Kaplan-Meier cumulative revision rate for pseudotumour increased progressively with time. At eight years, in all patients, it was 4% (95% confidence interval (CI) 2.2 to 5.8). Factors significantly associated with an increase in revision rate were female gender (p <
0.001), age under 40 (p = 0.003), small components (p = 0.003), and dysplasia (p = 0.019), whereas implant type was not (p = 0.156). These factors were inter-related, however, and on fitting a Cox proportional hazard model only gender (p = 0.002) and age (p = 0.024) had a significant independent influence on revision rate; size nearly reached significance (p = 0.08). Subdividing the cohort according to significant factors, we found that the revision rate for pseudotumours in men was 0.5% (95% CI 0 to 1.1) at eight years wheras in women over 40 years old it was 6% (95% CI 2.3 to 10.1) at eight years and in women under 40 years it was 13.1% at six years (95% CI 0 to 27) (p <
0.001). We recommend that resurfacings are undertaken with caution in women, particularly those under 40 years of age but they remain a good option in young men. Further work is required to understand the aetiology of pseudotumours so that this complication can be avoided.
A 60-year-old man developed severe neuropathic pain and foot-drop in his left leg following resurfacing arthroplasty of the left hip. The pain was refractory to all analgesics for 16 months. At exploration, a PDS suture was found passing through the sciatic nerve at several points over 6 cm and terminating in a large knot. After release of the suture and neurolysis there was dramatic and rapid improvement of the neuropathic pain and of motor function. This case represents the human equivalent of previously described nerve ligation in an animal model of neuropathic pain. It emphasises that when neuropathic pain is present after an operation, the nerve related to the symptoms must be inspected, and that removal of a suture or irritant may lead to relief of pain, even after many months.
Inflammatory pseudotumours occasionally occur after metal-on-metal hip resurfacing and often lead to revision. Our aim was to determine the severity of this complication by assessing the outcome of revision in these circumstances and by comparing this with the outcome of other metal-on-metal hip resurfacing revisions as well as that of matched primary total hip replacements. We identified 53 hips which had undergone metal-on-metal hip resurfacing and required revision at a mean of 1.59 years (0.01 to 6.69) after operation. Of these, 16 were revised for pseudotumours, 21 for fracture and 16 for other reasons. These were matched by age, gender and diagnosis with 103 patients undergoing primary total hip replacement with the Exeter implant. At a mean follow-up of three years (0.8 to 7.2) the outcome of metal-on-metal hip resurfacing revision for pseudotumour was poor with a mean Oxford hip score of 20.9 ( The outcome of revision for pseudotumour is poor and consideration should be given to early revision to limit the extent of the soft-tissue destruction. The outcome of resurfacing revision for other causes is good.
We have studied the relationship between metal ion levels and lymphocyte counts in patients with metal-on-metal hip resurfacings. Peripheral blood samples were analysed for lymphocyte subtypes and whole blood cobalt and chromium ion levels in 68 patients (34 with metal-on-metal hip resurfacings and 34 with standard metal-on-polyethylene total hip replacements). All hip components were radiologically well-fixed and the patients were asymptomatic. Cobalt and chromium levels were significantly elevated in the patients with metal-on-metal hip resurfacings, compared with the patients with standard metal-on-polyethylene designs (p <
0.0001). There was a statistically significant decrease in the level of CD8+ cells (T-cytotoxic/suppressor) (p = 0.005) in the metal-on-metal hip resurfacing group. A threshold level of blood cobalt and chromium ions was associated with reduced CD8+ T-cell counts. We have no evidence that our patients suffered as a result of this reduced level of CD8+ T-cells.
Aims.
The October 2024 Hip & Pelvis Roundup. 360. looks at: Does the primary surgical approach matter when choosing the approach for revision total hip arthroplasty?; Time to achieve the minimal clinically important difference in primary total hip arthroplasty: comparison of anterior and posterior surgical approaches; To scope or not to scope: arthroscopy as an adjunct to PAO does not provide better clinical outcomes at one year than PAO alone; Re-exploring horizons in
Aims. The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. Methods. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken. Results. Overall implant survival was 83.50% (95% confidence interval (CI) 0.79 to 0.90) at 25 years, and the number at risk was 79. Survival in male patients at 25 years was 89.5% (95% CI 0.83 to 0.96) compared to 66.9% for female patients (95% CI 0.51 to 0.83). Ten additional failures occurred in the period of ten to 25 years. These involved an adverse reaction to metal debris in four patients, a periprosthetic femoral neck fracture affecting five patients, and aseptic loosening in one patient. The median chromium levels were 49.50 nmol/l (interquartile range (IQR) 34 to 70), and the median cobalt serum levels were 42 nmol/l (IQR 24.50 to 71.25). The median OHS at last follow-up was 35 (IQR 10 to 48). During the 25-year study period, 29 patients died. Patient survival at 25 years was 75.10% (95% CI 0.67 to 0.83). Conclusion. This study demonstrates that MoM
Aims. The hip’s capsular ligaments passively restrain extreme range of movement (ROM) by wrapping around the native femoral head/neck. We determined the effect of
Aims. The proportion of arthroplasties performed in the ambulatory setting has increased considerably. However, there are concerns whether same-day discharge may increase the risk of complications. The aim of this study was to compare 90-day outcomes between inpatient arthroplasties and outpatient arthroplasties performed at an ambulatory surgery centre (ASC), and determine whether there is a learning curve associated with performing athroplasties in an ASC. Methods. Among a single-surgeon cohort of 970 patients who underwent arthroplasty at an ASC, 854 (88.0%) were matched one-to-one with inpatients based on age, sex, American Society of Anesthesiologists (ASA) grade, BMI, and procedure (105 could not be adequately matched and 11 lacked 90-day follow-up). The cohort included 281 total hip arthroplasties (THAs) (32.9%), 267 unicompartmental knee arthroplasties (31.3%), 242 primary total knee arthroplasties (TKAs) (28.3%), 60
We compared the five- to seven-year clinical and radiological results of the metal-on-metal Birmingham
We review the history and literature of
We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/Spotorno femoral components and titanium AlloFit acetabular components with Metasul bearings, with Durom
Femoroacetabular impingement (FAI) is commonly
associated with early hip arthritis. We reviewed our series of 1300
hip resurfacing procedures. More than 90% of our male patients,
with an average age of 53 years, had cam impingement lesions. In
this condition, there are anterior femoral neck osteophytes, and
a retroverted femoral head on a normally anteverted neck. It is
postulated that FAI results in collision of the anterior neck of
the femur against the rim of the acetabulum, causing damage to the
acetabular labrum and articular cartilage, resulting in osteoarthritis.
Early treatment of FAI involves arthroscopic or open removal of
bone from the anterior femoral neck, as well as repair or removal
of labral tears. However, once osteoarthritis has developed, hip
replacement or
The aim of this study was to establish the natural
course of unrevised asymptomatic pseudotumours after metal-on-metal
(MoM)
Aims. The aim of this study was to assess the functional gain achieved following
Aims. In previous studies, we identified multiple factors influencing the survivorship of
Narrowing of the femoral neck after resurfacing arthroplasty of the hip has been described previously in both cemented and uncemented
We report the outcome of total hip replacement in 29 failed metal-on-metal resurfacing hip replacements in which the primary surgery was performed between August 1995 and February 2005. The mean length of follow-up was five years (1.7 to 11.7). Of the 29
The early designs of
In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component. Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).Aims
Methods
Objectives. We have encountered patients who developed large joint fluid collections with massive elevations in chromium (Cr) and cobalt (Co) concentrations following metal-on-metal (MoM) hip arthroplasties. In some cases, retrieval analysis determined that these ion concentrations could not be explained simply by the wear rates of the components. We hypothesized that these effects may be associated with aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL). Patients and Methods. We examined the influence of the ALVAL grade on synovial fluid Co and Cr concentrations following adjustment for patient and device variables, including volumetric wear rates. Initially restricting the analysis to include only patients with one MoM
Patients considered suitable for total
The rate and mode of early failure in 463 Birmingham
Iliopsoas impingement occurs in 4% to 30% of patients after undergoing total hip arthroplasty (THA). Despite a relatively high incidence, there are few attempts at modelling impingement between the iliopsoas and acetabular component, and no attempts at modelling this in a representative cohort of subjects. The purpose of this study was to develop a novel computational model for quantifying the impingement between the iliopsoas and acetabular component and validate its utility in a case-controlled investigation. This was a retrospective cohort study of patients who underwent THA surgery that included 23 symptomatic patients diagnosed with iliopsoas tendonitis, and 23 patients not diagnosed with iliopsoas tendonitis. All patients received postoperative CT imaging, postoperative standing radiography, and had minimum six months’ follow-up. 3D models of each patient’s prosthetic and bony anatomy were generated, landmarked, and simulated in a novel iliopsoas impingement detection model in supine and standing pelvic positions. Logistic regression models were implemented to determine if the probability of pain could be significantly predicted. Receiver operating characteristic curves were generated to determine the model’s sensitivity, specificity, and area under the curve (AUC).Aims
Methods
Elevated blood cobalt levels secondary to metal-on-metal (MoM) hip arthroplasties are a suggested risk factor for developing cardiovascular complications including cardiomyopathy. Clinical studies assessing patients with MoM hips using left ventricular ejection fraction (LVEF) have found conflicting evidence of cobalt-induced cardiomyopathy. Global longitudinal strain (GLS) is an echocardiography measurement known to be more sensitive than LVEF when diagnosing early cardiomyopathies. The extent of cardiovascular injury, as measured by GLS, in patients with elevated blood cobalt levels has not previously been examined. A total of 16 patients with documented blood cobalt ion levels above 13 µg/l (13 ppb, 221 nmol/l) were identified from a regional arthroplasty database. They were matched with eight patients awaiting hip arthroplasty. All patients underwent echocardiography, including GLS, investigating potential signs of cardiomyopathy.Aims
Methods
Resurfacing arthroplasty of the hip is being performed more frequently in the United Kingdom. The majority of these patients are younger than 55 years of age, and in this group the key benefits include conservation of femoral bone stock and the potential reduction in the rate of dislocation afforded by the larger resurfacing head. Early aseptic loosening is well recognised in patients younger than 55 years of age, and proponents of resurfacing believe that the improved wear characteristics of the metal-on-metal bearing may improve the long-term survival of this implant. There has been some concern, however, that resurfacing may not be conservative of acetabular bone. We compared a series of 33 consecutive patients who had a hybrid total hip arthroplasty with an uncemented acetabular component and a cemented femoral implant, with 35 patients undergoing a Birmingham
Metal-on-metal hip resurfacing (MoM-HR) has seen decreased usage due to safety and longevity concerns. Joint registries have highlighted the risks in females, smaller hips, and hip dysplasia. This study aimed to identify if reported risk factors are linked to revision in a long-term follow-up of MoM-HR performed by a non-designer surgeon. A retrospective review of consecutive MoM hip arthroplasties (MoM-HRAs) using Birmingham Hip Resurfacing was conducted. Data on procedure side, indication, implant sizes and orientation, highest blood cobalt and chromium ion concentrations, and all-cause revision were collected from local and UK National Joint Registry records.Aims
Methods
The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.Aims
Methods
The first aim of this study was to evaluate whether preoperative renal function is associated with postoperative changes in whole blood levels of metal ions in patients who have undergone a Birmingham Hip Resurfacing (BHR) arthroplasty with a metal-on-metal bearing. The second aim was to evaluate whether exposure to increased cobalt (Co) and chromium (Cr) levels for ten years adversely affected renal function. As part of a multicentre, prospective post-approval study, whole blood samples were sent to a single specialized laboratory to determine Co and Cr levels, and the estimated glomerular filtration rate (eGFR). The study included patients with 117 unrevised unilateral BHRs. There were 36 females (31%). The mean age of the patients at the time of surgery of 51.3 years (SD 6.5), and they all had preoperative one-, four-, five-, and ten-year laboratory data. The mean follow-up was 10.1 years (SD 0.2).Aims
Methods
Aims. Routine surveillance of primary hip and knee arthroplasties has traditionally been performed with office follow-up visits at one year postoperatively. The value of these visits is unclear. The present study aims to determine the utility and burden of routine clinical follow-up at one year after primary arthroplasty to patients and providers. Methods. All patients (473) who underwent primary total hip (280),
The primary objective of this study was to develop a validated classification system for assessing iatrogenic bone trauma and soft-tissue injury during total hip arthroplasty (THA). The secondary objective was to compare macroscopic bone trauma and soft-tissues injury in conventional THA (CO THA) versus robotic arm-assisted THA (RO THA) using this classification system. This study included 30 CO THAs versus 30 RO THAs performed by a single surgeon. Intraoperative photographs of the osseous acetabulum and periacetabular soft-tissues were obtained prior to implantation of the acetabular component, which were used to develop the proposed classification system. Interobserver and intraobserver variabilities of the proposed classification system were assessed.Aims
Methods
The primary purpose of this study of metal-on-metal
(MoM)
Metal-on-metal resurfacing of the hip (MoMHR)
has enjoyed a resurgence in the last decade, but is now again in question
as a routine option for osteoarthritis of the hip. Proponents of
hip resurfacing suggest that its survival is superior to that of
conventional hip replacement (THR), and that
In this paper, we will consider the current role
of metal-on-metal bearings by looking at three subtypes of MoM hip
arthroplasty separately:
Aims. Our first aim was to determine whether there are significant
changes in the level of metal ions in the blood at mid-term follow-up,
in patients with an Articular Surface Replacement (ASR) arthroplasty.
Secondly, we sought to identify risk factors for any increases. Patients and Methods. The study involved 435 patients who underwent unilateral, metal-on-metal
(MoM)
The April 2024 Hip & Pelvis Roundup360 looks at: Impaction bone grafting for femoral revision hip arthroplasty with the Exeter stem; Effect of preoperative corticosteroids on postoperative glucose control in total joint replacement; Tranexamic acid in patients with a history of venous thromboembolism; Bisphosphonate use may be associated with an increased risk of periprosthetic hip fracture; A balanced approach: exploring the impact of surgical techniques on hip arthroplasty outcomes; A leap forward in hip arthroplasty: dual-mobility bearings reduce groin pain; A new perspective on complications: the link between blood glucose and joint infection risks.
The primary aim of this prospective, multicentre study is to describe the rates of returning to golf following hip, knee, ankle, and shoulder arthroplasty in an active golfing population. Secondary aims will include determining the timing of return to golf, changes in ability, handicap, and mobility, and assessing joint-specific and health-related outcomes following surgery. This is a multicentre, prospective, longitudinal study between the Hospital for Special Surgery, (New York City, New York, USA) and Edinburgh Orthopaedics, Royal Infirmary of Edinburgh, (Edinburgh, UK). Both centres are high-volume arthroplasty centres, specializing in upper and lower limb arthroplasty. Patients undergoing hip, knee, ankle, or shoulder arthroplasty at either centre, and who report being golfers prior to arthroplasty, will be included. Patient-reported outcome measures will be obtained at six weeks, three months, six months, and 12 months. A two-year period of recruitment will be undertaken of arthroplasty patients at both sites.Aims
Methods
Contemporary outcomes of primary total hip arthroplasties (THAs) with highly cross-linked polyethylene (HXLPE) liners in patients with inflammatory arthritis have not been well studied. This study examined the implant survivorship, complications, radiological results, and clinical outcomes of THA in patients with inflammatory arthritis. We identified 418 hips (350 patients) with a primary diagnosis of inflammatory arthritis who underwent primary THA with HXLPE liners from January 2000 to December 2017. Of these hips, 68% had rheumatoid arthritis (n = 286), 13% ankylosing spondylitis (n = 53), 7% juvenile rheumatoid arthritis (n = 29), 6% psoriatic arthritis (n = 24), 5% systemic lupus erythematosus (n = 23), and 1% scleroderma (n = 3). Mean age was 58 years (SD 14.8), 66.3% were female (n = 277), and mean BMI was 29 kg/m2 (SD 7). Uncemented femoral components were used in 77% of cases (n = 320). Uncemented acetabular components were used in all patients. Competing risk analysis was used accounting for death. Mean follow-up was 4.5 years (2 to 18).Aims
Methods
Mechanical impingement of the iliopsoas (IP) tendon accounts for 2% to 6% of persistent postoperative pain after total hip arthroplasty (THA). The most common initiator is anterior acetabular component protrusion, where the anterior margin is not covered by anterior acetabular wall. A CT scan can be used to identify and measure this overhang; however, no threshold exists for determining symptomatic anterior IP impingement due to overhang. A case-control study was conducted in which CT scan measurements were used to define a threshold that differentiates patients with IP impingement from asymptomatic patients after THA. We analyzed the CT scans of 622 patients (758 THAs) between May 2011 and May 2020. From this population, we identified 136 patients with symptoms suggestive of IP impingement. Among them, six were subsequently excluded: three because the diagnosis was refuted intraoperatively, and three because they had another obvious cause of impingement, leaving 130 hips (130 patients) in the study (impingement) group. They were matched to a control group of 138 asymptomatic hips (138 patients) after THA. The anterior acetabular component overhang was measured on an axial CT slice based on anatomical landmarks (orthogonal to the pelvic axis).Aims
Methods
Modular dual-mobility (DM) articulations are increasingly used during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS)-MRI abnormalities with serum metal ion levels in patients with DM articulations. A total of 45 patients (50 hips) with a modular DM articulation were included with mean follow-up of 3.7 years (SD 1.2). Enrolled patients with an asymptomatic, primary THA and DM articulation with over two years’ follow-up underwent MARS-MRI. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.Aims
Methods
The aim of this study was to compare open reduction and internal fixation (ORIF) with revision surgery for the surgical management of Unified Classification System (UCS) type B periprosthetic femoral fractures around cemented polished taper-slip femoral components following primary total hip arthroplasty (THA). Data were collected for patients admitted to five UK centres. The primary outcome measure was the two-year reoperation rate. Secondary outcomes were time to surgery, transfusion requirements, critical care requirements, length of stay, two-year local complication rates, six-month systemic complication rates, and mortality rates. Comparisons were made by the form of treatment (ORIF vs revision) and UCS type (B1 vs B2/B3). Kaplan-Meier survival analysis was performed with two-year reoperation for any reason as the endpoint.Aims
Methods