Aims. Unicompartmental knee arthroplasty (UKA) has a higher
Aims. This multicentre retrospective observational study’s aims were to investigate whether there are differences in the occurrence of radiolucent lines (RLLs) following total knee arthroplasty (TKA) between the conventional Attune baseplate and its successor, the novel Attune S+, independent from other potentially influencing factors; and whether tibial baseplate design and presence of RLLs are associated with differing
Aims. This study aimed to determine whether lateral femoral wall thickness (LWT) < 20.5 mm was associated with increased
Aims. The success of total knee arthroplasty (TKA) is usually measured using functional outcome scores and revision-free survivorship. However, reporting the lifetime
Aims. Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). Methods. We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model. Results. Median follow-up was 868 days (interquartile range 418 to 1,364). The offset liner design reduced the
Aims. The aim of this study was to describe the pattern of revision indications for unicompartmental knee arthroplasty (UKA) and total knee arthroplasty (TKA) and any change to this pattern for UKA patients over the last 20 years, and to investigate potential associations to changes in surgical practice over time. Methods. All primary knee arthroplasty surgeries performed due to primary osteoarthritis and their revisions reported to the Danish Knee Arthroplasty Register from 1997 to 2017 were included. Complex surgeries were excluded. The data was linked to the National Patient Register and the Civil Registration System for comorbidity, mortality, and emigration status. TKAs were propensity score matched 4:1 to UKAs.
Aims. Unicompartmental knee arthroplasty (UKA) has higher revision rates than total knee arthroplasty (TKA). As revision of UKA may be less technically demanding than revision TKA, UKA patients with poor functional outcomes may be more likely to be offered revision than TKA patients with similar outcomes. The aim of this study was to compare clinical thresholds for revisions between TKA and UKA using revision incidence and patient-reported outcomes, in a large, matched cohort at early, mid-, and late-term follow-up. Methods. Analyses were performed on propensity score-matched patient cohorts of TKAs and UKAs (2:1) registered in the New Zealand Joint Registry between 1 January 1999 and 31 December 2019 with an Oxford Knee Score (OKS) response at six months (n, TKA: 16,774; UKA: 8,387), five years (TKA: 6,718; UKA: 3,359), or ten years (TKA: 3,486; UKA: 1,743). Associations between OKS and revision within two years following the score were examined. Thresholds were compared using receiver operating characteristic analysis. Reasons for aseptic revision were compared using cumulative incidence with competing risk. Results. Fewer TKA patients with ‘poor’ outcomes (≤ 25) subsequently underwent revision compared with UKA at six months (5.1% vs 19.6%; p < 0.001), five years (4.3% vs 12.5%; p < 0.001), and ten years (6.4% vs 15.0%; p = 0.024). Compared with TKA, the relative risk for UKA was 2.5-times higher for ‘unknown’ reasons, bearing dislocations, and disease progression. Conclusion. Compared with TKA, more UKA patients with poor outcomes underwent revision from early to long-term follow-up, and were more likely to undergo revision for ‘unknown’ reasons, which suggest a lower clinical threshold for UKA. For UKA,
Aims. Corticosteroid injections are often used to manage glenohumeral arthritis in patients who may be candidates for future total shoulder arthroplasty (TSA) or reverse shoulder arthroplasty (rTSA). In the conservative management of these patients, corticosteroid injections are often provided for symptomatic relief. The purpose of this study was to determine if the timing of corticosteroid injections prior to TSA or rTSA is associated with changes in rates of revision and periprosthetic joint infection (PJI) following these procedures. Methods. Data were collected from a national insurance database from January 2006 to December 2017. Patients who underwent shoulder corticosteroid injection within one year prior to ipsilateral TSA or rTSA were identified and stratified into the following cohorts: < three months, three to six months, six to nine months, and nine to 12 months from time of corticosteroid injection to TSA or rTSA. A control cohort with no corticosteroid injection within one year prior to TSA or rTSA was used for comparison. Univariate and multivariate analyses were conducted to determine the association between specific time intervals and outcomes. Results. In total, 4,252 patients were included in this study. Among those, 1,632 patients (38.4%) received corticosteroid injection(s) within one year prior to TSA or rTSA and 2,620 patients (61.6%) did not. On multivariate analysis, patients who received corticosteroid injection < three months prior to TSA or rTSA were at significantly increased
Aims. The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the
Aims. Periprosthetic joint infection (PJI) is a challenging complication of any arthroplasty procedure. We reviewed our use of static antibiotic-loaded cement spacers (ABLCSs) for staged management of PJI where segmental bone loss, ligamentous instability, or soft-tissue defects necessitate a static construct. We reviewed factors contributing to their failure and techniques to avoid these complications when using ABLCSs in this context. Methods. A retrospective analysis was conducted of 94 patients undergoing first-stage revision of an infected knee prosthesis between September 2007 and January 2020 at a single institution. Radiographs and clinical records were used to assess and classify the incidence and causes of static spacer failure. Of the 94 cases, there were 19 primary total knee arthroplasties (TKAs), ten revision TKAs (varus-valgus constraint), 20 hinged TKAs, one arthrodesis (nail), one failed spacer (performed elsewhere), 21 distal femoral endoprosthetic arthroplasties, and 22 proximal tibial arthroplasties. Results. A total of 35/94 patients (37.2%) had spacer-related complications, of which 26/35 complications (74.3%) were because of mechanical failure of the spacer construct, while 9/35 (25.7%) were due to recurrence of infection. Risk factors for internal failure were a construct where the total intramedullary spacer length was less than twice the length of the central osseous defect (p = 0.009), where proximal or distal intraosseous spacer contact was < 10%, and after tibial tubercle osteotomy (p = 0.005). The incidence of spacer complications significantly increased the time to second stage: mean 157 days (42 to 458) in those without complications versus 227 days (11 to 528) with complications (p = 0.014). Conclusion. The failure rate of static antibiotic-loaded cement spacers is much higher than anticipated. Complications of the spacer significantly increased the time to second-stage
Aims. The aim of this study was to measure the effect of hospital case volume on the survival of revision total knee arthroplasty (RTKA). Methods. This is a retrospective analysis of Scottish Arthroplasty Project data, a nationwide audit which prospectively collects data on all arthroplasty procedures performed in Scotland. The primary outcome was RTKA survival at ten years. The primary explanatory variable was the effect of hospital case volume per year on RTKA survival. Kaplan-Meier survival curves were plotted with 95% confidence intervals (CIs) to determine the lifespan of RTKA. Multivariate Cox proportional hazards were used to estimate relative
Aims. The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA). Methods. Prospective cohort study using linked National Joint Registry (NJR) and manufacturer data. The primary endpoint was revision for aseptic loosening. Cox proportional hazard regression was the primary analytical approach. Manufacturing variables included resin type, crosslinking radiation dose, terminal sterilization method, terminal sterilization radiation dose, stabilization treatment, total radiation dose, packaging, and face asymmetry. Total radiation dose was further divided into G1 (no radiation), G2 (> 0 Mrad to < 5 Mrad), G3 (≥ 5 Mrad to < 10 Mrad), and G4 (≥ 10 Mrad). Results. A total of 5,329 THAs were revised, 1,290 of which were due to aseptic loosening. Total radiation dose, face asymmetry, and stabilization treatments were found to significantly affect implant survival. G1 had the highest
Aims. Proximal femoral endoprosthetic replacements (PFEPRs) are the most common reconstruction option for osseous defects following primary and metastatic tumour resection. This study aimed to compare the rate of implant failure between PFEPRs with monopolar and bipolar hemiarthroplasties and acetabular arthroplasties, and determine the optimum articulation for revision PFEPRs. Methods. This is a retrospective review of 233 patients who underwent PFEPR. The mean age was 54.7 years (SD 18.2), and 99 (42.5%) were male. There were 90 patients with primary bone tumours (38.6%), 122 with metastatic bone disease (52.4%), and 21 with haematological malignancy (9.0%). A total of 128 patients had monopolar (54.9%), 74 had bipolar hemiarthroplasty heads (31.8%), and 31 underwent acetabular arthroplasty (13.3%). Results. At a mean 74.4 months follow-up, the overall revision rate was 15.0%. Primary malignancy (p < 0.001) and age < 50 years (p < 0.001) were risk factors for
The aim of this study was to identify variables associated with time to revision, demographic details associated with revision indication, and type of prosthesis employed, and to describe the survival of hinge knee arthroplasty (HKA) when used for first-time knee revision surgery and factors that were associated with re-revision. Patient demographic details, BMI, American Society of Anesthesiologists (ASA) grade, indication for revision, surgical approach, surgeon grade, implant type (fixed and rotating), time of revision from primary implantation, and re-revision if undertaken were obtained from the National Joint Registry data for England, Wales, Northern Ireland, and the Isle of Man over an 18-year period (2003 to 2021).Aims
Methods
Aims. The aim of this study was to assess factors associated with the estimated lifetime risk of revision surgery after primary knee arthroplasty (KA). Methods. All patients from the Scottish Arthroplasty Project dataset undergoing primary KA during the period 1 January 1998 to 31 December 2019 were included. The cumulative incidence function for revision and death was calculated up to 20 years. Adjusted analyses used cause-specific Cox regression modelling to determine the influence of patient factors. The lifetime risk was calculated as a percentage for patients aged between 45 and 99 years using multiple-decrement life table methodology. Results. The estimated lifetime
Aims. The aims of the study were to report for a cohort aged younger than 40 years: 1) indications for HRA; 2) patient-reported outcomes in terms of the modified Harris Hip Score (HHS); 3) dislocation rate; and 4) revision rate. Methods. This retrospective analysis identified 267 hips from 224 patients who underwent an hip resurfacing arthroplasty (HRA) from a single fellowship-trained surgeon using the direct lateral approach between 2007 and 2019. Inclusion criteria was minimum two-year follow-up, and age younger than 40 years. Patients were followed using a prospectively maintained institutional database. Results. A total of 217 hips (81%) were included for follow-up analysis at a mean of 3.8 years. Of the 23 females who underwent HRA, none were revised, and the median head size was 46 mm (compared to 50 mm for males). The most common indication for HRA was femoroacetabular impingement syndrome (n = 133), and avascular necrosis ( (n = 53). Mean postoperative HHS was 100 at two and five years. No dislocations occurred. A total of four hips (1.8%) required reoperation for resection of heterotopic ossification, removal of components for infection, and subsidence with loosening. The overall revision rate was 0.9%. Conclusion. For younger patients with higher functional expectations and increased lifetime
The aims were to assess whether joint-specific outcome after total knee arthroplasty (TKA) was influenced by implant design over a 12-year follow-up period, and whether patient-related factors were associated with loss to follow-up and mortality risk. Long-term follow-up of a randomized controlled trial was undertaken. A total of 212 patients were allocated a Triathlon or a Kinemax TKA. Patients were assessed preoperatively, and one, three, eight, and 12 years postoperatively using the Oxford Knee Score (OKS). Reasons for patient lost to follow-up, mortality, and revision were recorded.Aims
Methods
With up to 40% of patients having patellofemoral joint osteoarthritis (PFJ OA), the two arthroplasty options are to replace solely the patellofemoral joint via patellofemoral arthroplasty (PFA), or the entire knee via total knee arthroplasty (TKA). The aim of this study was to assess postoperative success of second-generation PFAs compared to TKAs for patients treated for PFJ OA using patient-reported outcome measures (PROMs) and domains deemed important by patients following a patient and public involvement meeting. MEDLINE, EMBASE via OVID, CINAHL, and EBSCO were searched from inception to January 2022. Any study addressing surgical treatment of primary patellofemoral joint OA using second generation PFA and TKA in patients aged above 18 years with follow-up data of 30 days were included. Studies relating to OA secondary to trauma were excluded. ROB-2 and ROBINS-I bias tools were used.Aims
Methods
To perform an incremental cost-utility analysis and assess the impact of differential costs and case volume on the cost-effectiveness of robotic arm-assisted unicompartmental knee arthroplasty (rUKA) compared to manual (mUKA). This was a five-year follow-up study of patients who were randomized to rUKA (n = 64) or mUKA (n = 65). Patients completed the EuroQol five-dimension questionnaire (EQ-5D) preoperatively, and at three months and one, two, and five years postoperatively, which was used to calculate quality-adjusted life years (QALYs) gained. Costs for the primary and additional surgery and healthcare costs were calculated.Aims
Methods
Polished taper-slip (PTS) cemented stems have an excellent clinical track record and are the most common stem type used in primary total hip arthroplasty (THA) in the UK. Due to low rates of aseptic loosening, they have largely replaced more traditional composite beam (CB) cemented stems. However, there is now emerging evidence from multiple joint registries that PTS stems are associated with higher rates of postoperative periprosthetic femoral fracture (PFF) compared to their CB stem counterparts. The risk of both intraoperative and postoperative PFF remains greater with uncemented stems compared to either of these cemented stem subtypes. PFF continues to be a devastating complication following primary THA and is associated with high complication and mortality rates. Recent efforts have focused on identifying implant-related risk factors for PFF in order to guide preventative strategies, and therefore the purpose of this article is to present the current evidence on the effect of cemented femoral stem design on the risk of PFF. Cite this article:
Aims. Few studies have compared survivorship of total shoulder arthroplasty (TSA) with hemiarthroplasty (HA). This observational study compared survivorship of TSA with HA while controlling for important covariables and accounting for death as a competing risk. Patients and Methods. All patients who underwent shoulder arthroplasty in Ontario, Canada between April 2002 and March 2012 were identified using population-based health administrative data. We used the Fine–Gray sub-distribution hazard model to measure the association of arthroplasty type with time to revision surgery (accounting for death as a competing risk) controlling for age, gender, Charlson Comorbidity Index, income quintile, diagnosis, and surgeon factors. Results. During the study period, 5777 patients underwent shoulder arthroplasty (4079 TSA, 70.6%; 1698 HA, 29.4%), 321 (5.6%) underwent revision, and 1090 (18.9%) died. TSA patients were older (TSA mean age 68.4 years (. sd. 10.2) vs HA mean age 66.5 years (. sd. 12.7); p = 0.001). The proportion of female patients was slightly lower in the TSA group (58.0% vs 58.4%). The adjusted association between surgery type and time to shoulder revision interacted significantly with patient age. Compared with TSA patients, revision was more common in the HA group (adjusted-health ratio (HR) 1.214, 95% confidence interval (CI) 0.96 to 1.53) but this did not reach statistical significance. Conclusion. Although there was a trend towards higher
The optimal bearing surface design for medial unicompartmental knee arthroplasty (UKA) remains controversial. The aim of this study was to compare outcomes of fixed-bearing (FB) and mobile-bearing (MB) UKAs from a single high-volume institution. Prospectively collected data were reviewed for all primary cemented medial UKAs performed by seven surgeons from January 2006 to December 2022. A total of 2,999 UKAs were identified, including 2,315 FB and 684 MB cases. The primary outcome measure was implant survival. Secondary outcomes included 90-day and cumulative complications, reoperations, component revisions, conversion arthroplasties, range of motion, and patient-reported outcome measures. Overall mean age at surgery was 65.7 years (32.9 to 94.3), 53.1% (1,593/2,999) of UKAs were implanted in female patients, and demographics between groups were similar (p > 0.05). The mean follow-up for all UKAs was 3.7 years (0.0 to 15.6).Aims
Methods
The December 2022 Knee Roundup360 looks at: Effect of physical therapy versus arthroscopic partial meniscectomy: the ESCAPE trial at five years; Patellofemoral arthroplasty or total knee arthroplasty: a randomized controlled trial; Rehabilitation versus surgical reconstruction for anterior cruciate ligament injury; End-stage knee osteoarthritis in Australia: the effect of obesity; Do poor patient-reported outcome measures at six months relate to knee revision?; What is the cost of nonoperative interventions for knee osteoarthritis?
The February 2023 Knee Roundup. 360. looks at: Machine-learning models: are all complications predictable?; Positive cultures can be safely ignored in revision arthroplasty patients that do not meet the 2018 International Consensus Meeting Criteria; Spinal versus general anaesthesia in contemporary primary total knee arthroplasty; Preoperative pain and early arthritis are associated with poor outcomes in total knee arthroplasty; Risk factors for infection and revision surgery following patellar tendon and quadriceps tendon repairs; Supervised versus unsupervised rehabilitation following total knee arthroplasty; Kinematic alignment has similar outcomes to mechanical alignment: a systematic review and meta-analysis; Lifetime
Achilles tendon re-rupture (ATRR) poses a significant risk of postoperative complication, even after a successful initial surgical repair. This study aimed to identify risk factors associated with Achilles tendon re-rupture following operative fixation. This retrospective cohort study analyzed a total of 43,287 patients from national health claims data spanning 2008 to 2018, focusing on patients who underwent surgical treatment for primary Achilles tendon rupture. Short-term ATRR was defined as cases that required revision surgery occurring between six weeks and one year after the initial surgical repair, while omitting cases with simultaneous infection or skin necrosis. Variables such as age, sex, the presence of Achilles tendinopathy, and comorbidities were systematically collected for the analysis. We employed multivariate stepwise logistic regression to identify potential risk factors associated with short-term ATRR.Aims
Methods
The December 2022 Hip & Pelvis Roundup360 looks at: Fix and replace: simultaneous fracture fixation and hip arthroplasty for acetabular fractures in older patients; Is the revision rate for femoral neck fracture lower for total hip arthroplasty than for hemiarthroplasty?; Femoral periprosthetic fractures: data from the COMPOSE cohort study; Dual-mobility cups and fracture of the femur; What’s the deal with outcomes for hip and knee arthroplasty outcomes internationally?; Osteochondral lesions of the femoral head: is costal cartilage the answer?
Dual-mobility (DM) components are increasingly used to prevent and treat dislocation after total hip arthroplasty (THA). Intraprosthetic dissociation (IPD) is a rare complication of DM that is believed to have decreased with contemporary implants. This study aimed to report incidence, treatment, and outcomes of contemporary DM IPD. A total of 1,453 DM components were implanted at a single academic institution between January 2010 and December 2021: 695 in primary and 758 in revision THA. Of these, 49 presented with a dislocation of the large DM head and five presented with an IPD. At the time of closed reduction of the large DM dislocation, six additional IPDs occurred. The mean age was 64 years (SD 9.6), 54.5% were female (n = 6), and mean follow-up was 4.2 years (SD 1.8). Of the 11 IPDs, seven had a history of instability, five had abductor insufficiency, four had prior lumbar fusion, and two were conversions for failed fracture management.Aims
Methods
The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.Aims
Methods
This study evaluates the association between consultant and hospital volume and the risk of re-revision and 90-day mortality following first-time revision of primary hip arthroplasty for aseptic loosening. We conducted a cohort study of first-time, single-stage revision hip arthroplasties (RHAs) performed for aseptic loosening and recorded in the National Joint Registry (NJR) data for England, Wales, Northern Ireland, and the Isle of Man between 2003 and 2019. Patient identifiers were used to link records to national mortality data, and to NJR data to identify subsequent re-revision procedures. Multivariable Cox proportional hazard models with restricted cubic splines were used to define associations between volume and outcome.Aims
Methods
The aim of this study was to perform a systematic review and bias evaluation of the current literature to create an overview of risk factors for re-revision following revision total knee arthroplasty (rTKA). A systematic search of MEDLINE and Embase was completed in accordance with the Preferred Reporting Items for Systematic reviews and Meta-Analyses (PRISMA) guidelines. The studies were required to include a population of index rTKAs. Primary or secondary outcomes had to be re-revision. The association between preoperative factors and the effect on the risk for re-revision was also required to be reported by the studies.Aims
Methods
Surgical limb sparing for knee-bearing paediatric bone sarcoma is considered to have a clinically significant influence on postoperative function due to complications and leg-length discrepancies. However, researchers have not fully evaluated the long-term postoperative functional outcomes. Therefore, in this study, we aimed to elucidate the risk factors and long-term functional prognosis associated with paediatric limb-sparing surgery. We reviewed 40 patients aged under 14 years who underwent limb-sparing surgery for knee bone sarcoma (15 cases in the proximal tibia and 25 in the distal femur) between January 2000 and December 2013, and were followed up for a minimum of five years. A total of 35 patients underwent reconstruction using artificial materials, and five underwent biological reconstruction. We evaluated the patients’ postoperative complications, survival rate of reconstruction material, and limb, limb function, and leg-length discrepancy at the final follow-up, as well as the risk factors for each.Aims
Methods
The advent of modular porous metal augments has ushered in a new form of treatment for acetabular bone loss. The function of an augment can be seen as reducing the size of a defect or reconstituting the anterosuperior/posteroinferior columns and/or allowing supplementary fixation. Depending on the function of the augment, the surgeon can decide on the sequence of introduction of the hemispherical shell, before or after the augment. Augments should always, however, be used with cement to form a unit with the acetabular component. Given their versatility, augments also allow the use of a hemispherical shell in a position that restores the centre of rotation and biomechanics of the hip. Progressive shedding or the appearance of metal debris is a particular finding with augments and, with other radiological signs of failure, should be recognized on serial radiographs. Mid- to long-term outcomes in studies reporting the use of augments with hemispherical shells in revision total hip arthroplasty have shown rates of survival of > 90%. However, a higher risk of failure has been reported when augments have been used for patients with chronic pelvic discontinuity. Cite this article:
Implant failure has become more common as the number of primary total ankle arthroplasties (TAAs) performed has increased. Although revision arthroplasty has gained attention for functional preservation, the long-term results remain unclear. This study aimed to assess the long-term outcomes of revision TAA using a mobile-bearing prosthesis in a considerably large cohort; the risk factors for failure were also determined. This single-centre retrospective cohort study included 116 patients (117 ankles) who underwent revision TAA for failed primary TAA between July 2000 and March 2010. Survival analysis and risk factor assessment were performed, and clinical performance and patient satisfaction were evaluated preoperatively and at last follow-up.Aims
Methods
Histology is widely used for diagnosis of persistent infection during reimplantation in two-stage revision hip and knee arthroplasty, although data on its utility remain scarce. Therefore, this study aims to assess the predictive value of permanent sections at reimplantation in relation to reinfection risk, and to compare results of permanent and frozen sections. We retrospectively collected data from 226 patients (90 hips, 136 knees) with periprosthetic joint infection who underwent two-stage revision between August 2011 and September 2021, with a minimum follow-up of one year. Histology was assessed via the SLIM classification. First, we analyzed whether patients with positive permanent sections at reimplantation had higher reinfection rates than patients with negative histology. Further, we compared permanent and frozen section results, and assessed the influence of anatomical regions (knee versus hip), low- versus high-grade infections, as well as first revision versus multiple prior revisions on the histological result at reimplantation. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), chi-squared tests, and Kaplan-Meier estimates were calculated.Aims
Methods
Despite multiple trials and case series on hip hemiarthroplasty designs, guidance is still lacking on which implant to use. One particularly deficient area is long-term outcomes. We present over 1,000 consecutive cemented Thompson’s hemiarthroplasties over a ten-year period, recording all accessible patient and implant outcomes. Patient identifiers for a consecutive cohort treated between 1 January 2003 and 31 December 2011 were linked to radiographs, surgical notes, clinic letters, and mortality data from a national dataset. This allowed charting of their postoperative course, complications, readmissions, returns to theatre, revisions, and deaths. We also identified all postoperative attendances at the Emergency and Outpatient Departments, and recorded any subsequent skeletal injuries.Aims
Methods
The aim of this study was to explore the functional results in a fitter subgroup of participants in the Hip Fracture Evaluation with Alternatives of Total Hip Arthroplasty versus Hemiarthroplasty (HEALTH) trial to determine whether there was an advantage of total hip arthroplasty (THA) versus hemiarthroplasty (HA) in this population. We performed a post hoc exploratory analysis of a fitter cohort of patients from the HEALTH trial. Participants were aged over 50 years and had sustained a low-energy displaced femoral neck fracture (FNF). The fittest participant cohort was defined as participants aged 70 years or younger, classified as American Society of Anesthesiologists grade I or II, independent walkers prior to fracture, and living at home prior to fracture. Multilevel models were used to estimate the effect of THA versus HA on functional outcomes. In addition, a sensitivity analysis of the definition of the fittest participant cohort was performed.Aims
Methods
The aim of this study was to explore the relationship between reason for revision total hip arthroplasty (rTHA) and outcomes in terms of patient-reported outcome measures (PROMs). We reviewed a prospective cohort of 647 patients undergoing full or partial rTHA at a single high-volume centre with a minimum of two years’ follow-up. The reasons for revision were classified as: infection; aseptic loosening; dislocation; structural failure; and painful THA for other reasons. PROMs (modified Oxford Hip Score (mOHS), EuroQol five-dimension three-level health questionnaire (EQ-5D-3L) score, and visual analogue scales for pain during rest and activity), complication rates, and failure rates were compared among the groups.Aims
Methods
This systematic review aims to compare the precision of component positioning, patient-reported outcome measures (PROMs), complications, survivorship, cost-effectiveness, and learning curves of MAKO robotic arm-assisted unicompartmental knee arthroplasty (RAUKA) with manual medial unicompartmental knee arthroplasty (mUKA). Searches of PubMed, MEDLINE, and Google Scholar were performed in November 2021 according to the Preferred Reporting Items for Systematic Review and Meta-Analysis statement. Search terms included “robotic”, “unicompartmental”, “knee”, and “arthroplasty”. Published clinical research articles reporting the learning curves and cost-effectiveness of MAKO RAUKA, and those comparing the component precision, functional outcomes, survivorship, or complications with mUKA, were included for analysis.Aims
Methods
Aims. This study aimed to describe the use of revision knee arthroplasty in Australia and examine changes in lifetime risk over a decade. Methods. De-identified individual-level data on all revision knee arthroplasties performed in Australia from 2007 to 2017 were obtained from the Australian Orthopaedic Association National Joint Replacement Registry. Population data and life tables were obtained from the Australian Bureau of Statistics. The lifetime risk of revision surgery was calculated for each year using a standardized formula. Separate calculations were undertaken for males and females. Results. In total, 43,188 revision knee arthroplasty procedures were performed in Australia during the study period, with a median age at surgery of 69 years (interquartile range (IQR) 62 to 76). In 2017, revision knee arthroplasty rates were highest for males aged 70 to 79 years (102.9 procedures per 100,000 population). Lifetime
The aim of this study was to determine the outcome of all primary total hip arthroplasties (THAs) and their subsequent revision procedures in patients aged under 50 years performed at our institution. All 1,049 primary THAs which were undertaken in 860 patients aged under 50 years between 1988 and 2018 in our tertiary care institution were included. We used cemented implants in both primary and revision surgery. Impaction bone grafting was used in patients with acetabular or femoral bone defects. Kaplan-Meier analyses were used to determine the survival of primary and revision THA with the endpoint of revision for any reason, and of revision for aseptic loosening.Aims
Methods
Aims. To determine if primary cemented acetabular component geometry (long posterior wall (LPW), hooded, or offset reorientating) influences the risk of revision total hip arthroplasty (THA) for instability or loosening. Methods. The National Joint Registry (NJR) dataset was analyzed for primary THAs performed between 2003 and 2017. A cohort of 224,874 cemented acetabular components were included. The effect of acetabular component geometry on the
The distal radius is a major site of osteoporotic bone loss resulting in a high risk of fragility fracture. This study evaluated the capability of a cortical index (CI) at the distal radius to predict the local bone mineral density (BMD). A total of 54 human cadaver forearms (ten singles, 22 pairs) (19 to 90 years) were systematically assessed by clinical radiograph (XR), dual-energy X-ray absorptiometry (DXA), CT, as well as high-resolution peripheral quantitative CT (HR-pQCT). Cortical bone thickness (CBT) of the distal radius was measured on XR and CT scans, and two cortical indices mean average (CBTavg) and gauge (CBTg) were determined. These cortical indices were compared to the BMD of the distal radius determined by DXA (areal BMD (aBMD)) and HR-pQCT (volumetric BMD (vBMD)). Pearson correlation coefficient (r) and intraclass correlation coefficient (ICC) were used to compare the results and degree of reliability.Aims
Methods
One-stage revision hip arthroplasty for periprosthetic joint infection (PJI) has several advantages; however, resection of the proximal femur might be necessary to achieve higher success rates. We investigated the risk factors for resection and re-revisions, and assessed complications and subsequent re-revisions. In this single-centre, case-control study, 57 patients who underwent one-stage revision arthroplasty for PJI of the hip and required resection of the proximal femur between 2009 and 2018 were identified. The control group consisted of 57 patients undergoing one-stage revision without bony resection. Logistic regression analysis was performed to identify any correlation with resection and the risk factors for re-revisions. Rates of all-causes re-revision, reinfection, and instability were compared between groups.Aims
Methods
Aims. Periprosthetic fracture (PF) after primary total hip arthroplasty
(THA) is an uncommon but potentially devastating complication. This
study aims to investigate the influence of cemented stem designs
on the risk of needing a revision for a PF. Patients and Methods. We analysed data on 257 202 primary THAs with cemented stems
and 390 linked first revisions for PF recorded in the National Joint
Registry (NJR) of England, Wales and Northern Ireland to determine
if a cemented femoral stem brand was associated with the
Aims. Joint registries typically use revision of an implant as an endpoint and report survival rates after a defined number of years. However, reporting lifetime
Aims. Debate remains whether the patella should be resurfaced during total knee replacement (TKR). For non-resurfaced TKRs, we estimated what the revision rate would have been if the patella had been resurfaced, and examined the risk of re-revision following secondary patellar resurfacing. Methods. A retrospective observational study of the National Joint Registry (NJR) was performed. All primary TKRs for osteoarthritis alone performed between 1 April 2003 and 31 December 2016 were eligible (n = 842,072). Patellar resurfacing during TKR was performed in 36% (n = 305,844). The primary outcome was all-cause revision surgery. Secondary outcomes were the number of excess all-cause revisions associated with using TKRs without (versus with) patellar resurfacing, and the risk of re-revision after secondary patellar resurfacing. Results. The cumulative risk of all-cause revision at ten years was higher (p < 0.001) in primary TKRs without patellar resurfacing (3.54% (95% confidence interval (CI) 3.47 to 3.62)) compared to those with resurfacing (3.00% (95% CI 2.91 to 3.11)). Using flexible parametric survival modelling, we estimated one ‘excess’ revision per 189 cases performed where the patella was not resurfaced by ten years (equivalent to 2,842 excess revisions in our cohort). The risk of all-cause re-revision following secondary patellar resurfacing was 4.6 times higher than the
Unicompartmental knee arthroplasty (UKA) is a bone-preserving treatment option for osteoarthritis localized to a single compartment in the knee. The success of the procedure is sensitive to patient selection and alignment errors. Robotic arm-assisted UKA provides technological assistance to intraoperative bony resection accuracy, which is thought to improve ligament balancing. This paper presents the five-year outcomes of a comparison between manual and robotically assisted UKAs. The trial design was a prospective, randomized, parallel, single-centre study comparing surgical alignment in patients undergoing UKA for the treatment of medial compartment osteoarthritis (ISRCTN77119437). Participants underwent surgery using either robotic arm-assisted surgery or conventional manual instrumentation. The primary outcome measure (surgical accuracy) has previously been reported, and, along with secondary outcomes, were collected at one-, two-, and five-year timepoints. Analysis of five-year results and longitudinal analysis for all timepoints was performed to compare the two groups.Aims
Methods
The aims of this systematic review were to assess the learning curve of semi-active robotic arm-assisted total hip arthroplasty (rTHA), and to compare the accuracy, patient-reported functional outcomes, complications, and survivorship between rTHA and manual total hip arthroplasty (mTHA). Searches of PubMed, Medline, and Google Scholar were performed in April 2020 in line with the Preferred Reporting Items for Systematic Review and Meta-Analysis statement. Search terms included “robotic”, “hip”, and “arthroplasty”. The criteria for inclusion were published clinical research articles reporting the learning curve for rTHA (robotic arm-assisted only) and those comparing the implantation accuracy, functional outcomes, survivorship, or complications with mTHA.Aims
Methods
There is little evidence on the cost effectiveness
of different brands of hip prostheses. We compared lifetime cost effectiveness
of frequently used brands within types of prosthesis including cemented
(Exeter V40 Contemporary, Exeter V40 Duration and Exeter V40 Elite
Plus Ogee), cementless (Corail Pinnacle, Accolade Trident, and Taperloc Exceed)
and hybrid (Exeter V40 Trilogy, Exeter V40 Trident, and CPT Trilogy).
We used data from three linked English national databases to estimate
the lifetime
To investigate whether chronic kidney disease (CKD) is associated with the risk of all-cause revision or revision due to a periprosthetic joint infection (PJI) after primary hip or knee arthroplasty. This retrospective cohort study comprised 18,979 consecutive hip and knee arthroplasties from a single high-volume academic hospital. At a median of 5.6 years (interquartile range (IQR) 3.5 to 8.1), all deaths and revisions were counted. To overcome the competing risk of death, competing risk analysis using the cumulative incidence function (CIF) was applied to analyze the association between different stages of CKD and revisions. Confounding factors such as diabetes and BMI were considered using either a stratified CIF or the Fine and Gray model.Aims
Methods
Surgical treatment of hip fracture is challenging; the bone is porotic and fixation failure can be catastrophic. Novel implants are available which may yield superior clinical outcomes. This study compared the clinical effectiveness of the novel X-Bolt Hip System (XHS) with the sliding hip screw (SHS) for the treatment of fragility hip fractures. We conducted a multicentre, superiority, randomized controlled trial. Patients aged 60 years and older with a trochanteric hip fracture were recruited in ten acute UK NHS hospitals. Participants were randomly allocated to fixation of their fracture with XHS or SHS. A total of 1,128 participants were randomized with 564 participants allocated to each group. Participants and outcome assessors were blind to treatment allocation. The primary outcome was the EuroQol five-dimension five-level health status (EQ-5D-5L) utility at four months. The minimum clinically important difference in utility was pre-specified at 0.075. Secondary outcomes were EQ-5D-5L utility at 12 months, mortality, residential status, mobility, revision surgery, and radiological measures.Aims
Methods
Aims. To investigate the effect of polyethylene manufacturing characteristics and irradiation dose on the survival of cemented and reverse hybrid total hip arthroplasties (THAs). Methods. In this registry study, data from the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man (NJR) were linked with manufacturing data supplied by manufacturers. The primary endpoint was revision of any component. Cox proportional hazard regression was a primary analytic approach adjusting for competing risk of death, patient characteristics, head composition, and stem fixation. Results. A total of 290,770 primary THAs were successfully linked with manufacturing characteristics. Overall 4,708 revisions were analyzed, 1,260 of which were due to aseptic loosening. Total radiation dose was identified as a risk factor and included in the Cox model. For statistical modelling of aseptic loosening, THAs were grouped into three categories: G1 (no radiation); G2 ( > 0 to < 5 Mrad); and G3 ( ≥ 5 Mrad). G1 had the worst survivorship. The Cox regression hazard ratio for revision due to aseptic loosening for G2 was 0.7 (95% confidence interval (CI) 0.58 to 0.83), and for G3 0.4 (95% CI 0.30 to 0.53). Male sex and uncemented stem fixation were associated with higher
Aims. There is evidence that prior lumbar fusion increases the risk of dislocation and revision after total hip arthroplasty (THA). The relationship between prior lumbar fusion and the effect of femoral head diameter on THA dislocation has not been investigated. We examined the relationship between prior lumbar fusion or discectomy and the risk of dislocation or revision after THA. We also examined the effect of femoral head component diameter on the risk of dislocation or revision. Methods. Data used in this study were compiled from several Finnish national health registers, including the Finnish Arthroplasty Register (FAR) which was the primary source for prosthesis-related data. Other registers used in this study included the Finnish Health Care Register (HILMO), the Social Insurance Institutions (SII) registers, and Statistics Finland. The study was conducted as a prospective retrospective cohort study. Cox proportional hazards regression and Kaplan-Meier survival analysis were used for analysis. Results. Prior lumbar fusion surgery was associated with increased risk of prosthetic dislocation (hazard ratio (HR) = 2.393, p < 0.001) and revision (HR = 1.528, p < 0.001). Head components larger than 28 mm were associated with lower dislocation rates compared to the 28 mm head (32 mm: HR = 0.712, p < 0.001; 36 mm: HR = 0.700, p < 0.001; 38 mm: HR = 0.808, p < 0.140; and 40 mm: HR = 0.421, p < 0.001). Heads of 38 mm (HR = 1.288, p < 0.001) and 40 mm (HR = 1.367, p < 0.001) had increased
To assess the effect of age on clinical outcome and revision rates in patients who underwent total ankle arthroplasty (TAA) for end-stage ankle osteoarthritis (OA). A consecutive series of 811 ankles (789 patients) that underwent TAA between May 2003 and December 2013 were enrolled. The influence of age on clinical outcome, including the American Orthopaedic Foot and Ankle Society (AOFAS) hindfoot score, and pain according to the visual analogue scale (VAS) was assessed. In addition, the risk for revision surgery that includes soft tissue procedures, periarticular arthrodeses/osteotomies, ankle joint debridement, and/or inlay exchange (defined as minor revision), as well as the risk for revision surgery necessitating the exchange of any of the metallic components or removal of implant followed by ankle/hindfoot fusion (defined as major revision) was calculated.Aims
Methods
Aims. Antibiotic-loaded bone cements (ALBCs) may offer early protection against the formation of bacterial biofilm after joint arthroplasty. Use in hip arthroplasty is widely accepted, but there is a lack of evidence in total knee arthroplasty (TKA). The objective of this study was to evaluate the use of ALBC in a large population of TKA patients. Materials and Methods. Data from the National Joint Registry (NJR) of England and Wales were obtained for all primary cemented TKAs between March 2003 and July 2016. Patient, implant, and surgical variables were analyzed. Cox proportional hazards models were used to assess the influence of ALBC on
It is unknown whether kinematic alignment (KA) objectively improves knee balance in total knee arthroplasty (TKA), despite this being the biomechanical rationale for its use. This study aimed to determine whether restoring the constitutional alignment using a restrictive KA protocol resulted in better quantitative knee balance than mechanical alignment (MA). We conducted a randomized superiority trial comparing patients undergoing TKA assigned to KA within a restrictive safe zone or MA. Optimal knee balance was defined as an intercompartmental pressure difference (ICPD) of 15 psi or less using a pressure sensor. The primary endpoint was the mean intraoperative ICPD at 10° of flexion prior to knee balancing. Secondary outcomes included balance at 45° and 90°, requirements for balancing procedures, and presence of tibiofemoral lift-off.Aims
Methods
The aim of this study was to investigate mortality and risk of intraoperative medical complications depending on delay to hip fracture surgery by using data from the Norwegian Hip Fracture Register (NHFR) and the Norwegian Patient Registry (NPR). A total of 83 727 hip fractures were reported to the NHFR between 2008 and 2017. Pathological fractures, unspecified type of fractures or treatment, patients less than 50 years of age, unknown delay to surgery, and delays to surgery of greater than four days were excluded. We studied total delay (fracture to surgery, n = 38 754) and hospital delay (admission to surgery, n = 73 557). Cox regression analyses were performed to calculate relative risks (RRs) adjusted for sex, age, American Society of Anesthesiologists (ASA) classification, type of surgery, and type of fracture. Odds ratio (OR) was calculated for intraoperative medical complications. We compared delays of 12 hours or less, 13 to 24 hours, 25 to 36 hours, 37 to 48 hours, and more than 48 hours.Aims
Patients and Methods
Aims. We compared implant and patient survival following intraoperative periprosthetic femoral fractures (IOPFFs) during primary total hip arthroplasty (THA) with matched controls. Patients and Methods. This retrospective cohort study compared 4831 hips with IOPFF and 48 154 propensity score matched primary THAs without IOPFF implanted between 2004 and 2016, which had been recorded on a national joint registry. Implant and patient survival rates were compared between groups using Cox regression. Results. Ten-year stem survival was worse in the IOPFF group (p < 0.001).
Vitamin E-diffused, highly crosslinked polyethylene (VEPE) and porous titanium-coated (PTC) shells were introduced in total hip arthroplasty (THA) to reduce the risk of aseptic loosening. The purpose of this study was: 1) to compare the wear properties of VEPE to moderately crosslinked polyethylene; 2) to assess the stability of PTC shells; and 3) to report their clinical outcomes at seven years. A total of 89 patients were enrolled into a prospective study. All patients received a PTC shell and were randomized to receive a VEPE liner (n = 44) or a moderately crosslinked polyethylene (ModXLPE) liner (n = 45). Radiostereometric analysis (RSA) was used to measure polyethylene wear and component migration. Differences in wear were assessed while adjusting for body mass index, activity level, acetabular inclination, anteversion, and head size. Plain radiographs were assessed for radiolucency and patient-reported outcome measures (PROMs) were administered at each follow-up.Aims
Patients and Methods
Aims. The aim of this study was to estimate the 90-day
The aims of this study were to report the efficacy of revision surgery for patients with co-infective bacterial and fungal prosthetic joint infections (PJIs) presenting to a single institution, and to identify prognostic factors that would guide management. A total of 1189 patients with a PJI were managed in our bone infection service between 2006 and 2015; 22 (1.85%) with co-infective bacterial and fungal PJI were included in the study. There were nine women and 13 men, with a mean age at the time of diagnosis of 64.5 years (47 to 83). Their mean BMI was 30.9 kg/m2 (24 to 42). We retrospectively reviewed the outcomes of these PJIs, after eight total hip arthroplasties and 14 total knee arthroplasties. The mean clinical follow-up was 4.1 years (1.4 to 8.8).Aims
Patients and Methods
Between 1999 and 2005, 10 264 patients who had undergone total hip replacement (THR) for subcapital fracture of the hip were compared with 76 520 in whom THR had been performed for other reasons. All the cases were identified through the Swedish Hip Arthroplasty Register. The THRs performed as primary treatment for fracture were also compared with those done after failure of internal fixation. After seven years the rate of revision was higher in THR after fracture (4.4% vs 2.9%). Dislocation and periprosthetic fracture were the most common causes of
During the 11-year-period from 1970 to 1980, 799 total hip arthroplasties with metal-on-plastic prostheses were performed for primary osteoarthritis. In all cases high viscosity cement was used, but not under pressure; acetabular cartilage was removed, but the femoral canal was not plugged. The rate of
To present a surgically relevant update of trunnionosis. Systematic review performed April 2017.Aims
Materials and Methods
The aim of this study was to determine whether the rates of revision
for metal-on-metal (MoM) total hip arthroplasties (THAs) with Pinnacle
components varied according to the year of the initial operation,
and compare these with the rates of revision for other designs of
MoM THA. Data from the National Joint Registry for England and Wales included
36 mm MoM THAs with Pinnacle acetabular components which were undertaken
between 2003 and 2012 with follow-up for at least five years (n
= 10 776) and a control group of other MoM THAs (n = 13 817). The
effect of the year of the primary operation on all-cause rates of revision
was assessed using Cox regression and interrupted time-series analysis.Aims
Patients and Methods
Patellofemoral arthroplasty (PFA) has experienced significant
improvements in implant survivorship with second generation designs.
This has renewed interest in PFA as an alternative to total knee
arthroplasty (TKA) for younger active patients with isolated patellofemoral
osteoarthritis (PF OA). We analysed the cost-effectiveness of PFA We used a Markov transition state model to compare cost-effectiveness
between PFA and TKA. Simulated patients were aged 60 (base case)
and 50 years. Lifetime costs (2015 United States dollars), quality-adjusted
life year (QALY) gains and incremental cost-effectiveness ratio
(ICER) were calculated from a healthcare payer perspective. Annual rates
of revision were derived from the National Joint Registry for England,
Wales, Northern Ireland and the Isle of Man. Deterministic and probabilistic
sensitivity analysis was performed for all parameters against a
$50 000/QALY willingness to pay. Aims
Patients and Methods
Few studies have assessed outcomes following non-metal-on-metal hip arthroplasty (non-MoMHA) revision surgery performed for adverse reactions to metal debris (ARMD). We assessed outcomes following non-MoMHA revision surgery performed for ARMD, and identified predictors of re-revision. We performed a retrospective observational study using data from the National Joint Registry for England and Wales. All non-MoMHAs undergoing revision surgery for ARMD between 2008 and 2014 were included (185 hips in 185 patients). Outcome measures following ARMD revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using Cox regression.Objectives
Methods
Aims. Our aim was to report survivorship data and lessons learned with
the Corail/Pinnacle cementless total hip arthroplasty (THA) system. Patients and Methods. Between August 2005 and March 2015, a total of 4802 primary cementless
Corail/Pinnacle THAs were performed in 4309 patients. In March 2016,
we reviewed these hips from a prospectively maintained database. Results . A total of 80 hips (1.67%) have been revised which is equivalent
to a cumulative
Our aim was to compare kinematic with mechanical alignment in
total knee arthroplasty (TKA). We performed a prospective blinded randomised controlled trial
to compare the functional outcome of patients undergoing TKA in
mechanical alignment (MA) with those in kinematic alignment (KA).
A total of 71 patients undergoing TKA were randomised to either
kinematic (n = 36) or mechanical alignment (n = 35). Pre- and post-operative
hip-knee-ankle radiographs were analysed. The knee injury and osteoarthritis
outcome score (KOOS), American Knee Society Score, Short Form-36,
Euro-Qol (EQ-5D), range of movement (ROM), two minute walk, and timed
up and go tests were assessed pre-operatively and at six weeks,
three and six months and one year post-operatively.Aims
Patients and Methods
The aim of this study was to inform a definitive trial which
could determine the clinical effectiveness of the X-Bolt Dynamic
Hip Plating System compared with the sliding hip screw for patients
with complex pertrochanteric fragility fractures of the femur. This was a single centre, participant blinded, randomised, standard-of-care
controlled pilot trial. Patients aged 60 years and over with AO/ASIF
A2 and A3 type femoral pertrochanteric fractures were eligible.Aims
Patients and Methods
The February 2015 Knee Roundup360 looks at: Intra-operative sensors for knee balance; Mobile bearing no advantage; Death and knee replacement: a falling phenomenon; The swings and roundabouts of unicompartmental arthroplasty; Regulation, implants and innovation; The weight of arthroplasty responsibility!; BMI in arthroplasty
A number of studies have reported satisfactory
results from the isolated revision of an acetabular component. However,
many of these studies reported only the short- to intermediate-term
results of heterogeneous bearing surfaces in a mixed age group. We present our experience of using a ceramic-on-ceramic (CoC)
bearing for isolated revision of an uncemented acetabular component
in 166 patients (187 hips) who were under the age of 50 years at
the time of revision. There were 78 men and 88 women with a mean
age of 47.4 years (28 to 49). The most common reason for revision
was polyethylene wear and acetabular osteolysis in 123 hips (66%),
followed by aseptic loosening in 49 hips (26%). We report the clinical and radiological outcome, complication
rate, and survivorship of this group. The mean duration of follow-up
was 15.6 years (11 to 19). The mean pre-operative Harris hip score was 33 points (1 to 58),
and improved to a mean of 88 points (51 to 100) at follow-up. The
mean pre-operative total Western Ontario and McMaster Universities
Osteoarthritis Index score was 63.2 (43 to 91) and improved to 19.8
points (9 to 61) post-operatively. Overall, 153 of 166 patients
(92%) were satisfied with their outcome. Kaplan–Meier survivorship
analysis, with revision or radiological evidence of implant failure
(13 patients, 8%) as end-points, was 92% at 15 years (95% confidence
interval 0.89 to 0.97). Isolated revision of a cementless acetabular component using
a CoC bearing gives good results in patients under 50 years of age. Cite this article:
The June 2015 Hip &
Pelvis Roundup360 looks at: neuraxial anaesthesia and large joint arthroplasty; revision total hip arthoplasty: factors associated with re-revision surgery; acetabular version and clinical outcomes in impingement surgery; hip precautions may be ineffective; implant selection and cost effectiveness; femoroacetabular impingement in the older age group; multiple revision in hip arthroplasty
The August 2014 Shoulder &
Elbow Roundup360 looks at: Myofibroblasts perhaps not implicated in post-traumatic elbow stiffness; olecranon tip biomechanically sound for coranoid reconstruction; obesity and elbow replacement don’t mix; single column plating successful for extra-articular distal humeral fractures; satisfaction not predictable in frozen shoulder; tenodesis and repair both acceptable in Grade II SLAP tears; glenoid bone grafting is effective and glenohumeral articular lesions best seen with an arthroscope.
We compared a new fixation system, the Targon
Femoral Neck (TFN) hip screw, with the current standard treatment of
cannulated screw fixation. This was a single-centre, participant-blinded,
randomised controlled trial. Patients aged 65 years and over with
either a displaced or undisplaced intracapsular fracture of the
hip were eligible. The primary outcome was the risk of revision
surgery within one year of fixation. A total of 174 participants were included in the trial. The absolute
reduction in
Fractures of the proximal femur are one of the
greatest challenges facing the medical community, constituting a
heavy socioeconomic burden worldwide. Controversy exists regarding
the optimal treatment for independent patients with displaced intracapsular fractures
of the proximal femur. The recognised alternatives are hemiarthroplasty
and total hip replacement. At present there is no established standard
of care, with both types of arthroplasty being used in many centres.
The principal advantages of total hip replacement are a functional
benefit over hemiarthroplasty and a reduced risk of revision surgery.
The principal criticism is the increased risk of dislocation. We
believe that an alternative acetabular component may reduce the
risk of dislocation but still provide the functional benefit of
total hip replacement in these patients. We therefore propose to
investigate the dislocation risk of a dual-mobility acetabular component
compared with standard polyethylene component in total hip replacement
for independent patients with displaced intracapsular fractures
of the proximal femur within the framework of the larger WHiTE (Warwick
Hip Trauma Evaluation) Comprehensive Cohort Study. Cite this article:
Total hip replacement (THR) has been shown to
be a cost-effective procedure. However, it is not risk-free. Certain conditions,
such as diabetes mellitus, are thought to increase the risk of complications.
In this study we have evaluated the prevalence of diabetes mellitus
in patients undergoing THR and the associated risk of adverse operative
outcomes. A meta-analysis and systematic review were conducted according
to the guidelines of the meta-analysis of observational studies
in epidemiology. Inclusion criteria were observational studies reporting
the prevalence of diabetes in the study population, accompanied
by reports of at least one of the following outcomes: venous thromboembolic
events; acute coronary events; infections of the urinary tract,
lower respiratory tract or surgical site; or requirement for revision
arthroplasty. Altman and Bland’s methods were used to calculate differences
in relative risks. The prevalence of diabetes mellitus was found
to be 5.0% among patients undergoing THR, and was associated with
an increased risk of established surgical site infection (odds ratio
(OR) 2.04 (95% confidence interval (CI) 1.52 to 2.76)), urinary
infection (OR 1.43 (95% CI 1.33 to 1.55)) and lower respiratory
tract infections (OR 1.95 (95% CI 1.61 to 2.26)). Diabetes mellitus
is a relatively common comorbidity encountered in THR. Diabetic
patients have a higher rate of developing both surgical site and
non-surgical site infections following THR. Cite this article:
Our objective was to examine the rate of revision
and its predictive factors in patients undergoing total shoulder arthroplasty
(TSA). We used prospectively collected data from the Mayo Clinic
Total Joint Registry to examine five-, ten- and 20-year revision-free
survival following TSA and the predictive factors. We examined patient
characteristics (age, gender, body mass index, comorbidity), implant
fixation (cemented
Fractures of the proximal femur are one of the
greatest challenges facing the medical community, constituting a
heavy socioeconomic burden worldwide. Controversy exists regarding
the optimal treatment for patients with unstable trochanteric proximal
femoral fractures. The recognised treatment alternatives are extramedullary
fixation usually with a sliding hip screw and intramedullary fixation
with a cephalomedullary nail. Current evidence suggests that best
results and lowest complication rates occur using a sliding hip screw.
Complications in these difficult fractures are relatively common
regardless of type of treatment. We believe that a novel device,
the X-Bolt dynamic plating system, may offer superior fixation over
a sliding hip screw with lower reoperation risk and better function.
We therefore propose to investigate the clinical effectiveness of
the X-bolt dynamic plating system compared with standard sliding
hip screw fixation within the framework of a the larger WHiTE (Warwick
Hip Trauma Evaluation) Comprehensive Cohort Study. Cite this article:
The popularity of cementless total hip replacement
(THR) has surpassed cemented THR in England and Wales. This retrospective
cohort study records survival time to revision following primary
cementless THR with the most common combination (accounting for
almost a third of all cementless THRs), and explores risk factors independently
associated with failure, using data from the National Joint Registry
for England and Wales. Patients with osteoarthritis who had a DePuy
Corail/Pinnacle THR implanted between the establishment of the registry
in 2003 and 31 December 2010 were included within analyses. There
were 35 386 procedures. Cox proportional hazard models were used
to analyse the extent to which the
The purpose of this study was twofold: first,
to determine whether the five-year results of hip resurfacing arthroplasty
(HRA) in Canada justified the continued use of HRA; and second,
to identify whether greater refinement of patient selection was
warranted. This was a retrospective cohort study that involved a review
of 2773 HRAs performed between January 2001 and December 2008 at
11 Canadian centres. Cox’s proportional hazards models were used
to analyse the predictors of failure of HRA. Kaplan–Meier survival
analysis was performed to predict the cumulative survival rate at
five years. The factors analysed included age, gender, body mass
index, pre-operative hip pathology, surgeon’s experience, surgical
approach, implant sizes and implant types. The most common modes
of failure were also analysed. The 2773 HRAs were undertaken in 2450 patients: 2127 in men and
646 in women. The mean age at operation was 50.5 years ( The failure rates of HRA at five years justify the ongoing use
of this technique in men. Female gender is an independent predictor
of failure, and a higher failure rate at five years in women leads
the authors to recommend this technique only in exceptional circumstances
for women. Cite this article:
The June 2013 Knee Roundup360 looks at: knee arthroplasty in diabetic patients; whether TKR is a timebomb; the use of antidepressants for knee OA; trochleoplasty; articulated spacers; mental health and joint replacement; and the use of physiotherapy for meniscal tear.
Despite excellent results, the use of cemented
total hip replacement (THR) is declining. This retrospective cohort study
records survival time to revision following primary cemented THR
using the most common combination of components that accounted for
almost a quarter of all cemented THRs, exploring risk factors independently associated
with failure. All patients with osteoarthritis who had an Exeter
V40/Contemporary THR (Stryker) implanted before 31 December 2010
and recorded in the National Joint Registry for England and Wales
were included in the analysis. Cox’s proportional hazard models
were used to analyse the extent to which
Using data from the Norwegian Hip Fracture Register,
8639 cemented and 2477 uncemented primary hemiarthroplasties for
displaced fractures of the femoral neck in patients aged > 70 years
were included in a prospective observational study. A total of 218
re-operations were performed after cemented and 128 after uncemented
procedures. Survival of the hemiarthroplasties was calculated using
the Kaplan-Meier method and hazard rate ratios (HRR) for revision
were calculated using Cox regression analyses. At five years the
implant survival was 97% (95% confidence interval (CI) 97 to 97)
for cemented and 91% (95% CI 87 to 94) for uncemented hemiarthroplasties.
Uncemented hemiarthroplasties had a 2.1 times increased risk of
revision compared with cemented prostheses (95% confidence interval
1.7 to 2.6, p < 0.001). The increased risk was mainly caused
by revisions for peri-prosthetic fracture (HRR = 17), aseptic loosening
(HRR = 17), haematoma formation (HRR = 5.3), superficial infection
(HRR = 4.6) and dislocation (HRR = 1.8). More intra-operative complications,
including intra-operative death, were reported for the cemented
hemiarthroplasties. However, in a time-dependent analysis, the HRR
for re-operation in both groups increased as follow-up increased. This study showed that the