A retrospective analysis was performed of eight patients with an open triradiate cartilage, who underwent resection for osteosarcoma and reconstruction of the proximal femur with a hemiarthroplasty, in order to identify changes of acetabular development. An analysis of the centre-edge angle, teardrop-to-medial prosthesis distance, superior joint space, teardrop-to-superior prosthesis distance, degree of lateral translation, and arthritic changes, was performed on serial radiographs. The median age at the time of the initial surgery was 11 years (5 to 14). All patients developed progressive superior and lateral migration of the
Periprosthetic joint infection (PJI) is a difficult complication requiring a comprehensive eradication protocol. Cure rates have essentially stalled in the last two decades, using methods of antimicrobial cement joint spacers and parenteral antimicrobial agents. Functional spacers with higher-dose antimicrobial-loaded cement and antimicrobial-loaded calcium sulphate beads have emphasized local antimicrobial delivery on the premise that high-dose local antimicrobial delivery will enhance eradication. However, with increasing antimicrobial pressures, microbiota have responded with adaptive mechanisms beyond traditional antimicrobial resistance genes. In this review we describe adaptive resistance mechanisms that are relevant to the treatment of PJI. Some mechanisms are well known, but others are new. The objective of this review is to inform clinicians of the known adaptive resistance mechanisms of microbes relevant to PJI. We also discuss the implications of these adaptive mechanisms in the future treatment of PJI. Cite this article:
The differential diagnosis of the painful total
hip arthroplasty (resurfacing or total hip) includes infection,
failure of fixation (loosening), tendinitis, bursitis, synovitis,
adverse local tissue reaction (ALTR) to cobalt-chromium alloys,
and non-hip issues, such as spinal disorders, hernia, gynecologic,
and other pelvic pain. Assuming that the hip is the source of pain,
the first level question is
Adverse spinal motion or balance (spine mobility) and adverse pelvic mobility, in combination, are often referred to as adverse spinopelvic mobility (SPM). A stiff lumbar spine, large posterior standing pelvic tilt, and severe sagittal spinal deformity have been identified as risk factors for increased hip instability. Adverse SPM can create functional malposition of the acetabular components and hence is an instability risk. Adverse pelvic mobility is often, but not always, associated with abnormal spinal motion parameters. Dislocation rates for dual-mobility articulations (DMAs) have been reported to be between 0% and 1.1%. The aim of this study was to determine the early survivorship from the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) of patients with adverse SPM who received a DMA. A multicentre study was performed using data from 227 patients undergoing primary total hip arthroplasty (THA), enrolled consecutively. All the patients who had one or more adverse spine or pelvic mobility parameter had a DMA inserted at the time of their surgery. The mean age was 76 years (22 to 93) and 63% were female (n = 145). At a mean of 14 months (5 to 31) postoperatively, the AOANJRR was analyzed for follow-up information. Reasons for revision and types of revision were identified.Aims
Methods
Progressive retroversion of a cemented stem is
predictive of early loosening and failure. We assessed the relationship
between direct post-operative stem anteversion, measured with CT,
and the resulting rotational stability, measured with repeated radiostereometric
analysis over ten years. The study comprised 60 cemented total hip
replacements using one of two types of matt collared stem with a
rounded cross-section. The patients were divided into three groups
depending on their measured post-operative anteversion (<
10°,
10° to 25°, >
25°). There was a strong correlation between direct
post-operative anteversion and later posterior rotation. At one
year the <
10° group showed significantly more progressive retroversion
together with distal migration, and this persisted to the ten-year
follow-up. In the <
10° group four of ten stems (40%) had been
revised at ten years, and an additional two stems (20%) were radiologically
loose. In the ‘normal’ (10° to 25°) anteversion group there was
one revised (3%) and one loose stem (3%) of a total of 30 stems,
and in the >
25° group one stem (5%) was revised and another loose (5%)
out of 20 stems. This poor outcome is partly dependent on the design
of this prosthesis, but the results strongly suggest that the initial
rotational position of cemented stems during surgery affects the
subsequent progressive retroversion, subsidence and eventual loosening.
The degree of retroversion may be sensitive to
Aims. The aim of this to study was to compare the previously unreported
long-term survival outcome of the Oxford medial unicompartmental
knee arthroplasty (UKA) performed by trainee surgeons and consultants. . Patients and Methods. We therefore identified a previously unreported cohort of 1084
knees in 947 patients who had a UKA inserted for anteromedial knee
arthritis by consultants and surgeons in training, at a tertiary
arthroplasty centre and performed survival analysis on the group
with revision as the endpoint. Results. The ten-year cumulative survival rate for revision or exchange
of any part of the
Total knee arthroplasty (TKA) is a common and safe orthopaedic procedure. Zimmer Biomet's NexGen is the second most popular brand of implant used in the UK. The primary cause of revision after the first year is aseptic loosening. We present our experience of using this implant, with significant concerns around its performance with regards early aseptic loosening of the tibial component. A retrospective, single-surgeon review was carried out of all of the NexGen Legacy Posterior Stabilized (LPS) TKAs performed in this institute. The specific model used for the index procedures was the NexGen Complete Knee System (Legacy Knee-Posterior Stabilized LPS-Flex Articular Surface, LPS-Flex Femoral Component Option, and Stemmed Nonaugmentable Tibial Component Option).Aims
Methods
Giant cell tumour of bone (GCTB) treatment changed since the introduction of denosumab from purely surgical towards a multidisciplinary approach, with recent concerns of higher recurrence rates after denosumab. We evaluated oncological, surgical, and functional outcomes for distal radius GCTB, with a critically appraised systematic literature review. We included 76 patients with distal radius GCTB in three sarcoma centres (1990 to 2019). Median follow-up was 8.8 years (2 to 23). Seven patients underwent curettage, 38 curettage with adjuvants, and 31 resection; 20 had denosumab.Aims
Methods
Aims. The aim of this paper was to present the clinical features of
patients with musculoskeletal sources of methicillin-sensitive Staphylococcus
aureus (MSSA) septicaemia. Patients and Methods. A total of 137 patients presented with MSSA septicaemia between
2012 and 2015. The primary source of infection was musculoskeletal
in 48 patients (35%). Musculoskeletal infection was considered the
primary source of septicaemia when endocarditis and other obvious
sources were excluded. All patients with an arthroplasty at the time
were evaluated for any
Femoral cement-in-cement revision is a well described technique to reduce morbidity and complications in hip revision surgery. Traditional techniques for septic revision of hip arthroplasty necessitate removal of all bone cement from the femur. In our two centres, we have been using a cement-in-cement technique, leaving the distal femoral bone cement in selected patients for septic hip revision surgery, both for single and the first of two-stage revision procedures. A prerequisite for adoption of this technique is that the surgeon considers the cement mantle to be intimately fixed to bone without an intervening membrane between cement and host bone. We aim to report our experience for this technique. We have analyzed patients undergoing this cement-in-cement technique for femoral revision in infection, and present a consecutive series of 89 patients. Follow-up was undertaken at a mean of 56.5 months (24.0 to 134.7) for the surviving cases.Aims
Methods
Hip arthroplasty aims to accurately recreate joint biomechanics. Considerable attention has been paid to vertical and horizontal offset, but femoral head centre in the anteroposterior (AP) plane has received little attention. This study investigates the accuracy of restoration of joint centre of rotation in the AP plane. Postoperative CT scans of 40 patients who underwent unilateral uncemented total hip arthroplasty were analyzed. Anteroposterior offset (APO) and femoral anteversion were measured on both the operated and non-operated sides. Sagittal tilt of the femoral stem was also measured. APO measured on axial slices was defined as the perpendicular distance between a line drawn from the anterior most point of the proximal femur (anterior reference line) to the centre of the femoral head. The anterior reference line was made parallel to the posterior condylar axis of the knee to correct for rotation.Aims
Methods
The August 2015 Oncology Roundup. 360 . looks at: Glasgow prognostic score in soft-tissue sarcoma; Denosumab in giant cell tumour; Timing, complications and radiotherapy; Pigmented villonodular synovitis and arthroscopy; PATHFx: estimating survival in pathological cancer;
The preoperative diagnosis of periprosthetic joint infection (PJI) remains a challenge due to a lack of biomarkers that are both sensitive and specific. We investigated the performance characteristics of polymerase chain reaction (PCR), interleukin-6 (IL6), and calprotectin of synovial fluid in the diagnosis of PJI. We performed systematic search of PubMed, Embase, The Cochrane Library, Web of Science, and Science Direct from the date of inception of each database through to 31 May 2021. Studies which described the diagnostic accuracy of synovial fluid PCR, IL6, and calprotectin using the Musculoskeletal Infection Society criteria as the reference standard were identified.Aims
Methods
Periprosthetic hip fractures (PPFs) after total hip arthroplasty are difficult to treat. Therefore, it is important to identify modifiable risk factors such as stem selection to reduce the occurrence of PPFs. This study aimed to clarify differences in fracture torque, surface strain, and fracture type analysis between three different types of cemented stems. We conducted biomechanical testing of bone analogues using six cemented stems of three different types: collarless polished tapered (CPT) stem, Versys Advocate (Versys) stem, and Charnley-Marcel-Kerboull (CMK) stem. Experienced surgeons implanted each of these types of stems into six bone analogues, and the analogues were compressed and internally rotated until failure. Torque to fracture and fracture type were recorded. We also measured surface strain distribution using triaxial rosettes.Aims
Methods
The complications of impaction bone grafting in revision hip replacement includes fracture of the femur and subsidence of the prosthesis. In this in vitro study we aimed to investigate whether the use of vibration, combined with a perforated tamp during the compaction of morsellised allograft would reduce peak loads and hoop strains in the femur as a surrogate marker of the risk of fracture and whether it would also improve graft compaction and
Aims. To evaluate the effectiveness of an institutionally developed
algorithm for evaluation and diagnosis of
Excision of the proximal femur for tumour with
prosthetic reconstruction using a bipolar femoral head places a considerable
load on the unreplaced acetabulum. . We retrospectively reviewed the changes which occur around the
affected hip joint by evaluating the post-operative radiographs
of 65 consecutive patients who underwent proximal
Many aspects of total knee arthroplasty have
changed since its inception. Modern