Advertisement for orthosearch.org.uk
Results 161 - 180 of 1053
Results per page:
Bone & Joint 360
Vol. 11, Issue 2 | Pages 47 - 49
1 Apr 2022


Bone & Joint Research
Vol. 11, Issue 2 | Pages 82 - 90
7 Feb 2022
Eckert JA Bitsch RG Sonntag R Reiner T Schwarze M Jaeger S

Aims

The cemented Oxford unicompartmental knee arthroplasty (OUKA) features two variants: single and twin peg OUKA. The purpose of this study was to assess the stability of both variants in a worst-case scenario of bone defects and suboptimal cementation.

Methods

Single and twin pegs were implanted randomly allocated in 12 pairs of human fresh-frozen femora. We generated 5° bone defects at the posterior condyle. Relative movement was simulated using a servohydraulic pulser, and analyzed at 70°/115° knee flexion. Relative movement was surveyed at seven points of measurement on implant and bone, using an optic system.


Bone & Joint Research
Vol. 1, Issue 6 | Pages 104 - 110
1 Jun 2012
Swinteck BJ Phan DL Jani J Owen JR Wayne JS Mounasamy V

Objectives. The use of two implants to manage concomitant ipsilateral femoral shaft and proximal femoral fractures has been indicated, but no studies address the relationship of dynamic hip screw (DHS) side plate screws and the intramedullary nail where failure might occur after union. This study compares different implant configurations in order to investigate bridging the gap between the distal DHS and tip of the intramedullary nail. Methods. A total of 29 left synthetic femora were tested in three groups: 1) gapped short nail (GSN); 2) unicortical short nail (USN), differing from GSN by the use of two unicortical bridging screws; and 3) bicortical long nail (BLN), with two angled bicortical and one unicortical bridging screws. With these findings, five matched-pairs of cadaveric femora were tested in two groups: 1) unicortical long nail (ULN), with a longer nail than USN and three bridging unicortical screws; and 2) BLN. Specimens were axially loaded to 22.7 kg (50 lb), and internally rotated 90°/sec until failure. Results. For synthetic femora, a difference was detected between GSN and BLN in energy to failure (p = 0.04) and torque at failure (p = 0.02), but not between USN and other groups for energy to failure (vs GSN, p = 0.71; vs BLN, p = 0.19) and torque at failure (vs GSN, p = 0.55; vs BLN, p = 0.15). For cadaveric femora, ULN and BLN performed similarly because of the improvement provided by the bridging screws. Conclusions. Our study shows that bicortical angled screws in the DHS side plate are superior to no screws at all in this model and loading scenario, and suggests that adding unicortical screws to a gapped construct is probably beneficial


Bone & Joint 360
Vol. 11, Issue 1 | Pages 38 - 41
1 Feb 2022


Bone & Joint 360
Vol. 11, Issue 3 | Pages 14 - 17
1 Jun 2022


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1522 - 1527
1 Nov 2008
Davis ET Olsen M Zdero R Waddell JP Schemitsch EH

A total of 20 pairs of fresh-frozen cadaver femurs were assigned to four alignment groups consisting of relative varus (10° and 20°) and relative valgus (10° and 20°), 75 composite femurs of two neck geometries were also used. In both the cadaver and the composite femurs, placing the component in 20° of valgus resulted in a significant increase in load to failure. Placing the component in 10° of valgus had no appreciable effect on increasing the load to failure except in the composite femurs with varus native femoral necks. Specimens in 10° of varus were significantly weaker than the neutrally-aligned specimens. The results suggest that retention of the intact proximal femoral strength occurs at an implant angulation of ≥ 142°. However, the benefit of extreme valgus alignment may be outweighed in clinical practice by the risk of superior femoral neck notching, which was avoided in this study


Bone & Joint Open
Vol. 3, Issue 6 | Pages 475 - 484
13 Jun 2022
Jang SJ Vigdorchik JM Windsor EW Schwarzkopf R Mayman DJ Sculco PK

Aims

Navigation devices are designed to improve a surgeon’s accuracy in positioning the acetabular and femoral components in total hip arthroplasty (THA). The purpose of this study was to both evaluate the accuracy of an optical computer-assisted surgery (CAS) navigation system and determine whether preoperative spinopelvic mobility (categorized as hypermobile, normal, or stiff) increased the risk of acetabular component placement error.

Methods

A total of 356 patients undergoing primary THA were prospectively enrolled from November 2016 to March 2018. Clinically relevant error using the CAS system was defined as a difference of > 5° between CAS and 3D radiological reconstruction measurements for acetabular component inclination and anteversion. Univariate and multiple logistic regression analyses were conducted to determine whether hypermobile (Δsacral slope(SS)stand-sit > 30°), or stiff (SSstand-sit < 10°) spinopelvic mobility contributed to increased error rates.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1281 - 1286
1 Oct 2009
Olsen M Davis ET Chiu M Gamble P Tumia N Boyle RA Schemitsch EH

The computed neck-shaft angle and the size of the femoral component were recorded in 100 consecutive hip resurfacings using imageless computer-navigation and compared with the angle measured before operation and with actual component implanted. The reliability of the registration was further analysed using ten cadaver femora. The mean absolute difference between the measured and navigated neck-shaft angle was 16.3° (0° to 52°). Navigation underestimated the measured neck-shaft angle in 38 patients and the correct implant size in 11. Registration of the cadaver femora tended to overestimate the correct implant size and provided a low level of repeatability in computing the neck-shaft angle. Prudent pre-operative planning is advisable for use in conjunction with imageless navigation since misleading information may be registered intraoperatively, which could lead to inappropriate sizing and positioning of the femoral component in hip resurfacing


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1060 - 1065
1 Aug 2018
Hwang J Shields MN Berglund LJ Hooke AW Fitzsimmons JS O’Driscoll SW

Aims. The aim of this study was to evaluate two hypotheses. First, that disruption of posterior bundle of the medial collateral ligament (PMCL) has to occur for the elbow to subluxate in cases of posteromedial rotatory instability (PMRI) and second, that ulnohumeral contact pressures increase after disruption of the PMCL. Materials and Methods. Six human cadaveric elbows were prepared on a custom-designed apparatus which allowed muscle loading and passive elbow motion under gravitational varus. Joint contact pressures were measured sequentially in the intact elbow (INTACT), followed by an anteromedial subtype two coronoid fracture (COR), a lateral collateral ligament (LCL) tear (COR + LCL), and a PMCL tear (COR + LCL + PMCL). Results. There was no subluxation or joint incongruity in the INTACT, COR, and COR + LCL specimens. All specimens in the COR + LCL + PMCL group subluxated under gravity-varus loads. The mean articular contact pressure of the COR + LCL group was significantly higher than those in the INTACT and the COR groups. The mean articular contact pressure of the COR + LCL + PMCL group was significantly higher than that of the INTACT group, but not higher than that of the COR + LCL group. Conclusion. In the presence of an anteromedial fracture and disruption of the LCL, the posterior bundle of the MCL has to be disrupted for gross subluxation of the elbow to occur. However, elevated joint contact pressures are seen after an anteromedial fracture and LCL disruption even in the absence of such subluxation. Cite this article: Bone Joint J 2018;100-B:1060–5


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1638 - 1640
1 Dec 2009
Pichler W Weinberg AM Grechenig S Tesch NP Heidari N Grechenig W

Intra-articular punctures and injections are performed routinely on patients with injuries to and chronic diseases of joints, to release an effusion or haemarthrosis, or to inject drugs. The purpose of this study was to investigate the accuracy of placement of the needle during this procedure. A total of 76 cadaver acromioclavicular joints were injected with a solution containing methyl blue and subsequently dissected to distinguish intra- from peri-articular injection. In order to assess the importance of experience in achieving accurate placement, half of the injections were performed by an inexperienced resident and half by a skilled specialist. The specialist injected a further 20 cadaver acromioclavicular joints with the aid of an image intensifier. The overall frequency of peri-articular injection was much higher than expected at 43% (33 of 76) overall, with 42% (16 of 38) by the specialist and 45% (17 of 38) by the resident. The specialist entered the joint in all 20 cases when using the image intensifier. Correct positioning of the needle in the joint should be facilitated by fluoroscopy, thereby guaranteeing an intra-articular injection


Bone & Joint Research
Vol. 11, Issue 5 | Pages 270 - 277
6 May 2022
Takegami Y Seki T Osawa Y Imagama S

Aims

Periprosthetic hip fractures (PPFs) after total hip arthroplasty are difficult to treat. Therefore, it is important to identify modifiable risk factors such as stem selection to reduce the occurrence of PPFs. This study aimed to clarify differences in fracture torque, surface strain, and fracture type analysis between three different types of cemented stems.

Methods

We conducted biomechanical testing of bone analogues using six cemented stems of three different types: collarless polished tapered (CPT) stem, Versys Advocate (Versys) stem, and Charnley-Marcel-Kerboull (CMK) stem. Experienced surgeons implanted each of these types of stems into six bone analogues, and the analogues were compressed and internally rotated until failure. Torque to fracture and fracture type were recorded. We also measured surface strain distribution using triaxial rosettes.


Bone & Joint Open
Vol. 3, Issue 5 | Pages 390 - 397
1 May 2022
Hiranaka T Suda Y Saitoh A Tanaka A Arimoto A Koide M Fujishiro T Okamoto K

The kinematic alignment (KA) approach to total knee arthroplasty (TKA) has recently increased in popularity. Accordingly, a number of derivatives have arisen and have caused confusion. Clarification is therefore needed for a better understanding of KA-TKA. Calipered (or true, pure) KA is performed by cutting the bone parallel to the articular surface, compensating for cartilage wear. In soft-tissue respecting KA, the tibial cutting surface is decided parallel to the femoral cutting surface (or trial component) with in-line traction. These approaches are categorized as unrestricted KA because there is no consideration of leg alignment or component orientation. Restricted KA is an approach where the periarthritic joint surface is replicated within a safe range, due to concerns about extreme alignments that have been considered ‘alignment outliers’ in the neutral mechanical alignment approach. More recently, functional alignment and inverse kinematic alignment have been advocated, where bone cuts are made following intraoperative planning, using intraoperative measurements acquired with computer assistance to fulfill good coordination of soft-tissue balance and alignment. The KA-TKA approach aims to restore the patients’ own harmony of three knee elements (morphology, soft-tissue balance, and alignment) and eventually the patients’ own kinematics. The respective approaches start from different points corresponding to one of the elements, yet each aim for the same goal, although the existing implants and techniques have not yet perfectly fulfilled that goal.


Bone & Joint 360
Vol. 10, Issue 6 | Pages 21 - 24
1 Dec 2021


The Bone & Joint Journal
Vol. 104-B, Issue 5 | Pages 559 - 566
1 May 2022
Burden EG Batten T Smith C Evans JP

Aims

Arthroplasty is being increasingly used for the management of distal humeral fractures (DHFs) in elderly patients. Arthroplasty options include total elbow arthroplasty (TEA) and hemiarthroplasty (HA); both have unique complications and there is not yet a consensus on which implant is superior. This systematic review asked: in patients aged over 65 years with unreconstructable DHFs, what differences are there in outcomes, as measured by patient-reported outcome measures (PROMs), range of motion (ROM), and complications, between distal humeral HA and TEA?

Methods

A systematic review of the literature was performed via a search of MEDLINE and Embase. Two reviewers extracted data on PROMs, ROM, and complications. PROMs and ROM results were reported descriptively and a meta-analysis of complications was conducted. Quality of methodology was assessed using Wylde’s non-summative four-point system. The study was registered with PROSPERO (CRD42021228329).


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1389 - 1394
1 Oct 2011
Soubeyrand M Ciais G Wassermann V Kalouche I Biau D Dumontier C Gagey O

Disruption of the interosseous membrane is easily missed in patients with Essex-Lopresti syndrome. None of the imaging techniques available for diagnosing disruption of the interosseous membrane are completely dependable. We undertook an investigation to identify whether a simple intra-operative test could be used to diagnose disruption of the interosseous membrane during surgery for fracture of the radial head and to see if the test was reproducible. We studied 20 cadaveric forearms after excision of the radial head, ten with and ten without disruption of the interosseous membrane. On each forearm, we performed the radius joystick test: moderate lateral traction was applied to the radial neck with the forearm in maximal pronation, to look for lateral displacement of the proximal radius indicating that the interosseous membrane had been disrupted. Each of six surgeons (three junior and three senior) performed the test on two consecutive days. Intra-observer agreement was 77% (95% confidence interval (CI) 67 to 85) and interobserver agreement was 97% (95% CI 92 to 100). Sensitivity was 100% (95% CI 97 to 100), specificity 88% (95% CI 81 to 93), positive predictive value 90% (95% CI 83 to 94), and negative predictive value 100%). This cadaveric study suggests that the radius joystick test may be useful for detecting disruption of the interosseous membrane in patients undergoing open surgery for fracture of the radial head and is reproducible. A confirmatory study in vivo is now required


Bone & Joint 360
Vol. 11, Issue 2 | Pages 34 - 37
1 Apr 2022


Bone & Joint 360
Vol. 11, Issue 2 | Pages 27 - 30
1 Apr 2022


The Bone & Joint Journal
Vol. 98-B, Issue 8 | Pages 1036 - 1042
1 Aug 2016
Amanatullah DF Masini MA Roger DJ Pagnano MW

Aims. We wished to quantify the extent of soft-tissue damage sustained during minimally invasive total hip arthroplasty through the direct anterior (DA) and direct superior (DS) approaches. Materials and Methods. In eight cadavers, the DA approach was performed on one side, and the DS approach on the other, a single brand of uncemented hip prosthesis was implanted by two surgeons, considered expert in their surgical approaches. Subsequent reflection of the gluteus maximus allowed the extent of muscle and tendon damage to be measured and the percentage damage to each anatomical structure to be calculated. Results. The DA approach caused substantially greater damage to the gluteus minimus muscle and tendon when compared with the DS approach (t-test, p = 0.049 and 0.003, respectively). The tensor fascia lata and rectus femoris muscles were damaged only in the DA approach. There was no difference in the amount of damage to the gluteus medius muscle and tendon, piriformis tendon, obturator internus tendon, obturator externus tendon or quadratus femoris muscle between approaches. The posterior soft-tissue releases of the DA approach damaged the gluteus minimus muscle and tendon, piriformis tendon and obturator internus tendon. Conclusion. The DS approach caused less soft-tissue damage than the DA approach. However the clinical relevance is unknown. Further clinical outcome studies, radiographic evaluation of component position, gait analyses and serum biomarker levels are necessary to evaluate and corroborate the safety and efficacy of the DS approach. Cite this article: Bone Joint J 2016;98-B1036–42


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 9 | Pages 1252 - 1256
1 Sep 2006
Mayr E Krismer M Ertl M Kessler O Thaler M Nogler M

A complete cement mantle is important for the longevity of a total hip replacement. In the minimally-invasive direct anterior approach used at the Innsbruck University hospital, the femoral component has to be inserted into the femoral canal by an angulated movement. In a cadaver study, the quality and the extent of the cement mantle surrounding 13 Exeter femoral components implanted straight through a standard anterolateral transgluteal approach were compared with those of 13 similar femoral components implanted in an angulated fashion through a direct anterior approach. A third-generation cementing technique was used. The inner and outer contours of the cement mantles was traced from CT scans and the thickness and cross-sectional area determined. In no case was the cement mantle incomplete. The total mean thickness of the cement mantle was 3.62 mm (95% confidence interval 3.59 to 3.65). The mean thickness in the group using the minimally-invasive approach was 0.16 mm less than that in the anterolateral group. The distribution of the thickness was similar in the two groups. The mean thickness was less on the anteromedial and anterolateral aspect than on the posterior aspect of the femur. There is no evidence that the angulated introduction of Exeter femoral components in the direct anterior approach in cadavers compromises the quality, extent or thickness of the cement mantle


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 416 - 423
1 Apr 2022
Mourkus H Phillips NJ Rangan A Peach CA

Aims

The aim of this study was to investigate the outcome of periprosthetic fractures of the humerus and to assess the uniformity of the classifications used for these fractures (including those around elbow and/or shoulder arthroplasties) by performing a systematic review of the literature.

Methods

A systematic search was conducted using the National Institute for Health and Care Excellence Healthcare Databases Advance Search. For inclusion, studies had to report clinical outcomes following the management of periprosthetic fractures of the humerus. The protocol was registered on the PROSPERO database.