Advertisement for orthosearch.org.uk
Results 141 - 160 of 222
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1446 - 1451
1 Nov 2007
Biring GS Masri BA Greidanus NV Duncan CP Garbuz DS

A prospective cohort of 222 patients who underwent revision hip replacement between April 2001 and March 2004 was evaluated to determine predictors of function, pain and activity level between one and two years post-operatively, and to define quality of life outcomes using validated patient reported outcome tools. Predictive models were developed and proportional odds regression analyses were performed to identify factors that predict quality of life outcomes at one and two years post-operatively. The dependent outcome variables were the Western Ontario and McMaster Osteoarthritis Index (WOMAC) function and pain scores, and University of California Los Angeles activity scores. The independent variables included patient demographics, operative factors, and objective quality of life parameters, including pre-operative WOMAC, and the Short Form-12 mental component score.

There was a significant improvement (t-test, p < 0.001) in all patient quality of life scores. In the predictive model, factors predictive of improved function (original regression analyses, p < 0.05) included a higher pre-operative WOMAC function score (p < 0.001), age between 60 and 70 years (p < 0.037), male gender (p = 0.017), lower Charnley class (p < 0.001) and aseptic loosening being the indication for revision (p < 0.003). Using the WOMAC pain score as an outcome variable, factors predictive of improvement included the pre-operative WOMAC function score (p = 0.001), age between 60 and 70 years (p = 0.004), male gender (p = 0.005), lower Charnley class (p = 0.001) and no previous revision procedure (p = 0.023). The pre-operative WOMAC function score (p = 0.001), the indication for the operation (p = 0.007), and the operating surgeon (p = 0.008) were significant predictors of the activity assessment at follow-up.

Predictors of quality of life outcomes after revision hip replacement were established. Although some patient-specific and surgery-specific variables were important, age, gender, Charnley class and pre-operative WOMAC function score had the most robust associations with outcome.


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 259 - 265
1 Feb 2013
Dhawale AA Karatas AF Holmes L Rogers KJ Dabney KW Miller F

We reviewed the long-term radiological outcome, complications and revision operations in 19 children with quadriplegic cerebral palsy and hip dysplasia who underwent combined peri-iliac osteotomy and femoral varus derotation osteotomy. They had a mean age of 7.5 years (1.6 to 10.9) and comprised 22 hip dislocations and subluxations. We also studied the outcome for the contralateral hip. At a mean follow-up of 11.7 years (10 to 15.1) the Melbourne cerebral palsy (CP) hip classification was grade 2 in 16 hips, grade 3 in five, and grade 5 in one. There were five complications seen in four hips (21%, four patients), including one dislocation, one subluxation, one coxa vara with adduction deformity, one subtrochanteric fracture and one infection. A recurrent soft-tissue contracture occurred in five hips and ten required revision surgery.

In pre-adolescent children with quadriplegic cerebral palsy good long-term outcomes can be achieved after reconstruction of the hip; regular follow-up is required.

Cite this article: Bone Joint J 2013;95-B:259–65.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1011 - 1021
1 Aug 2013
Krishnan H Krishnan SP Blunn G Skinner JA Hart AJ

Following the recall of modular neck hip stems in July 2012, research into femoral modularity will intensify over the next few years. This review aims to provide surgeons with an up-to-date summary of the clinically relevant evidence. The development of femoral modularity, and a classification system, is described. The theoretical rationale for modularity is summarised and the clinical outcomes are explored. The review also examines the clinically relevant problems reported following the use of femoral stems with a modular neck.

Joint replacement registries in the United Kingdom and Australia have provided data on the failure rates of modular devices but cannot identify the mechanism of failure. This information is needed to determine whether modular neck femoral stems will be used in the future, and how we should monitor patients who already have them implanted.

Cite this article: Bone Joint J 2013;95-B:1011–21.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1487 - 1491
1 Nov 2006
Ahlmann ER Menendez LR

Custom-made intercalary endoprostheses may be used for the reconstruction of diaphyseal defects following the resection of bone tumours. The aim of this study was to determine the survival of intercalary endoprostheses with a lap joint design, and to evaluate the clinical results, complications and functional outcome. We retrospectively reviewed six consecutive patients, three of whom underwent limb salvage with intercalary endoprostheses of the tibia, two of the femur, and one of the humerus. Their mean age was 42 years (28 to 64). The mean follow-up was 21.6 months (9 to 58). The humeral prosthesis required revision at 14 months owing to aseptic loosening. There were no implant-related failures. Musculoskeletal Tumour Society functional outcome scores indicated that patients achieved 90% of premorbid function.

Custom intercalary endoprostheses result in reconstructions comparable with, if not better than, those of allografts. Using this design of implant reduces the incidence of early complications and difficulties experienced with previous versions.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 16 - 22
1 Jan 2008
Yates PJ Burston BJ Whitley E Bannister GC

We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known.

At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%).

The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1561 - 1566
1 Dec 2007
Nowicki P Chaudhary H

Avascular necrosis of the femoral head creates considerable morbidity in successful renal transplant recipients who are generally young and expect active lifestyles. Total hip replacement is considered the treatment of choice in these patients, but surgeons may be wary because of a supposed increase in the risk of infection and other complications.

A review of the literature reveals that cemented hip arthroplasty provides good to excellent functional outcomes for renal transplant patients. Most authors have found that the risk of infection is not increased despite chronic immunosuppression, but the rates of general complications are and should be anticipated and treated. There is a high rate of early failure in these patients because of their young age and diffuse osteopenia as a result of secondary hyperparathyroidism related to the underlying renal disease and chronic steroid use. Recent studies have found that despite decreased bone stock in these patients, porous-coated prostheses are not contraindicated.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 336 - 342
1 Mar 2008
Cuff DJ Virani NA Levy J Frankle MA Derasari A Hines B Pupello DR Cancio M Mighell M

We retrospectively reviewed 21 patients (22 shoulders) who presented with deep infection after surgery to the shoulder, 17 having previously undergone hemiarthroplasty and five open repair of the rotator cuff. Nine shoulders had undergone previous surgical attempts to eradicate their infection. The diagnosis of infection was based on a combination of clinical suspicion (16 shoulders), positive frozen sections (> 5 polymorphonuclear leukocytes per high-power field) at the time of revision (15 shoulders), positive intra-operative cultures (18 shoulders) or the pre-operative radiological appearances. The patients were treated by an extensive debridement, intravenous antibiotics, and conversion to a reverse shoulder prosthesis in either a single- (10 shoulders) or a two-stage (12 shoulders) procedure.

At a mean follow-up of 43 months (25 to 66) there was no evidence of recurrent infection. All outcome measures showed statistically significant improvements. Mean abduction improved from 36.1° (sd 27.8) pre-operatively to 75.7° (sd 36.0) (p < 0.0001), the mean forward flexion from 43.1° (sd 33.5) to 79.5° (sd 43.2) (p = 0.0003), and mean external rotation from 10.2° (sd 18.7) to 25.4° (sd 23.5) (p = 0.0037). There was no statistically significant difference in any outcome between the single-stage and the two-stage group.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 2 | Pages 160 - 165
1 Feb 2007
Blomfeldt R Törnkvist H Eriksson K Söderqvist A Ponzer S Tidermark J

The best treatment for the active and lucid elderly patient with a displaced intracapsular fracture of the femoral neck is still controversial. Randomised controlled trials have shown that a primary total hip replacement is superior to internal fixation as regards the need for secondary surgery, hip function and health-related quality of life. Despite good results achieved with total hip replacement in this group, most orthopaedic surgeons still advocate hemiarthroplasty for this injury. We studied 120 patients with a mean age of 81 years (70 to 90) with an acute displaced intracapsular fracture of the femoral neck. They were randomly allocated to be treated with either a bipolar hemiarthroplasty or total hip replacement. Outcome measurements included peri-operative data, general and hip-specific complications, hip function and health-related quality of life. The patients were reviewed at four and 12 months.

The duration of surgery was longer in the total hip replacement group (102 minutes (70 to 151)) versus 78 minutes (43 to 131) (p < 0.001), and the intra-operative blood loss was increased 460 ml (100 to 1100) versus 320 ml (50 to 850) (p < 0.001), but there were no differences between the groups regarding any complications or mortality. There were no dislocations in either group. Hip function measured by the Harris hip score was significantly better in the total hip replacement group at both follow-up periods (p = 0.011 and p < 0.001, respectively). The health-related quality of life measure was in favour of the total hip replacement group but did not reach statistical significance (p = 0.818 at four months and p = 0.636 at 12 months).

These results indicate that a total hip replacement provides better function than a bipolar hemiarthroplasty as soon as one year post-operatively, without increasing the complication rate. We recommend total hip replacement as the primary treatment for this group of patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 297 - 301
1 Mar 2012
Haddad FS Konan S

An international faculty of orthopaedic surgeons presented their work on the current challenges in hip surgery at the London Hip Meeting which was attended by over 400 delegates. The topics covered included femoroacetabular impingement, thromboembolic phenomena associated with hip surgery, bearing surfaces (including metal-on-metal articulations), outcomes of hip replacement surgery and revision hip replacement. We present a concise report of the current opinions on hip surgery from this meeting with appropriate references to the current literature.


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 436 - 444
1 Apr 2013
Scott CEH Nutton RW Biant LC

The lateral compartment is predominantly affected in approximately 10% of patients with osteoarthritis of the knee. The anatomy, kinematics and loading during movement differ considerably between medial and lateral compartments of the knee. This in the main explains the relative protection of the lateral compartment compared with the medial compartment in the development of osteoarthritis. The aetiology of lateral compartment osteoarthritis can be idiopathic, usually affecting the femur, or secondary to trauma commonly affecting the tibia. Surgical management of lateral compartment osteoarthritis can include osteotomy, unicompartmental knee replacement and total knee replacement. This review discusses the biomechanics, pathogenesis and development of lateral compartment osteoarthritis and its management.

Cite this article: Bone Joint J 2013;95-B:436–44.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1574 - 1580
1 Dec 2007
Hallan G Lie SA Furnes O Engesaeter LB Vollset SE Havelin LI

Primary uncemented femoral stems reported to the Norwegian arthroplasty register between 1987 and 2005 were included in this prospective observational study. There were 11 516 hips (9679 patients) and 14 different designs of stem. Kaplan-Meier survival probabilities and Cox regression were used to analyse the data.

With aseptic loosening as the end-point, all currently used designs performed excellently with survival of 96% to 100% at ten years. With the end-point as stem revision for any cause, the long-term results of the different designs varied from poor to excellent, with survival at 15 years ranging between 29% and 97%. Follow-up for longer than seven years was needed to identify some of the poorly-performing designs. There were differences between the stems; the Corail, used in 5456 hips, was the most frequently used stem with a survival of 97% at 15 years. Male gender was associated with an increased risk of revision of × 1.3 (95% confidence interval 1.05 to 1.52), but age and diagnosis had no influence on the results. Overall, modern uncemented femoral stems performed well.

Moderate differences in survival between well-performing stems should be interpreted with caution since the differences may be caused by factors other than the stem itself.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 932 - 936
1 Jul 2012
Lin P Hsu C Huang C Chen W Wang J

Tranexamic acid (TEA), an inhibitor of fibrinolysis, reduces blood loss after routine total knee replacement (TKR). However, controversy persists regarding the dosage and timing of administration of this drug during surgery. We performed a prospective randomised controlled study to examine the optimum blood-saving effect of TEA in minimally invasive TKR.

We randomly assigned 151 patients who underwent unilateral minimally invasive TKR to three groups: 1) a placebo group (50 patients); 2) a one-dose TEA group (52 patients), who received one injection of TEA (10 mg/kg) intra-operatively on deflation of the tourniquet; and 3) a two-dose TEA group (49 patients), who received two injections of TEA (10 mg/kg) given pre-operatively and intra-operatively. Total blood loss was calculated from the maximum loss of haemoglobin. All patients were followed clinically for the presence of venous thromboembolism (VTE).

The mean total blood loss was significantly higher in the placebo group than in the other two groups (1222 ml (845 to 2043) versus 1035 ml (397 to 1934) and 986 ml (542 to 1811), respectively (both p < 0.0001)). The mean blood loss was not significantly different between the one- and two-TEA groups (p = 0.148). The mean transfusion rate was higher in the placebo group than in the other two groups (22% versus 3.8% (p = 0.006) and 6.1% (p = 0.041), respectively) and there was no statistically significant difference in the mean transfusion rate between the one- and two-TEA groups (p = 0.672). Only one patient, in the two-dose group, had a radiologically confirmed deep venous thrombosis.

Our prospective randomised controlled study showed that one intra-operative injection of TEA is effective for blood conservation after minimally invasive TKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 889 - 894
1 Jul 2012
Burston BJ Barnett AJ Amirfeyz R Yates PJ Bannister GC

We prospectively followed 191 consecutive collarless polished tapered (CPT) femoral stems, implanted in 175 patients who had a mean age at operation of 64.5 years (21 to 85). At a mean follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were still alive. The fate of all original stems is known. The 16-year survivorship with re-operation for any reason was 80.7% (95% confidence interval 72 to 89.4). There was no loss to follow-up, with clinical data available on all 95 hips and radiological assessment performed on 90 hips (95%). At latest follow-up, the mean Harris hip score was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48). Stems subsided within the cement mantle, with a mean subsidence of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem (0.5%) has been revised due to aseptic loosening. In total seven stems were revised for any cause, of which four revisions were required for infection following revision of the acetabular component. A total of 21 patients (11%) required some sort of revision procedure; all except three of these resulted from failure of the acetabular component. Cemented acetabular components had a significantly lower revision burden (three hips, 2.7%) than Harris Galante uncemented components (17 hips, 21.8%) (p < 0.001).

The CPT stem continues to provide excellent radiological and clinical outcomes at 15 years following implantation. Its results are consistent with other polished tapered stem designs.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 609 - 614
1 May 2012
Flecher X Blanc G Sainsous B Parratte S Argenson J

We describe the results of 81 consecutive revision total hip replacements with impaction grafting in 79 patients using a collared polished chrome–cobalt stem, customised in length according to the extent of distal bone loss. Our hypothesis was that the features of this stem would reduce the rate of femoral fracture and subsidence of the stem.

The mean follow-up was 12 years (8 to 15). No intra-operative fracture or significant subsidence occurred. Only one patient suffered a post-operative diaphyseal fracture, which was associated with a fall. All but one femur showed incorporation of the graft. No revision for aseptic loosening was recorded.

The rate of survival of the femoral component at 12 years, using further femoral revision as the endpoint, was 100% (95% confidence interval (CI) 95.9 to 100), and at nine years using re-operation for any reason as the endpoint, was 94.6% (95% CI 92.0 to 97.2).

These results suggest that a customised cemented polished stem individually adapted to the extent of bone loss and with a collar may reduce subsidence and the rate of fracture while maintaining the durability of the fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1511 - 1516
1 Nov 2012
Chang CB Cho W

In a prospective multicentre study we investigated variations in pain management used by knee arthroplasty surgeons in order to compare the differences in pain levels among patients undergoing total knee replacements (TKR), and to compare the effectiveness of pain management protocols. The protocols, peri-operative levels of pain and patient satisfaction were investigated in 424 patients who underwent TKR in 14 hospitals. The protocols were highly variable and peri-operative pain levels varied substantially, particularly during the first two post-operative days. Differences in levels of pain were greatest during the night after TKR, when visual analogue scores ranged from 16.9 to 94.3 points.

Of the methods of managing pain, the combined use of peri-articular infiltration and nerve blocks provided better pain relief than other methods during the first two post-operative days. Patients managed with peri-articular injection plus nerve block, and epidural analgesia were more likely to have higher satisfaction at two weeks after TKR. This study highlights the need to establish a consistent pain management strategy after TKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1475 - 1481
1 Nov 2012
Berton C Puskas GJ Christofilopoulos P Stern R Hoffmeyer P Lübbeke A

There are no recent studies comparing cable with wire for the fixation of osteotomies or fractures in total hip replacement (THR). Our objective was to evaluate the five-year clinical and radiological outcomes and complication rates of the two techniques. We undertook a review including all primary and revision THRs performed in one hospital between 1996 and 2005 using cable or wire fixation. Clinical and radiological evaluation was performed five years post-operatively. Cables were used in 51 THRs and wires in 126, and of these, 36 THRs with cable (71%) and 101 with wire (80%) were evaluated at follow-up. The five-year radiographs available for 33 cable and 91 wire THRs revealed rates of breakage of fixation of 12 of 33 (36%) and 42 of 91 (46%), respectively. With cable there was a significantly higher risk of metal debris (68% vs 9%; adjusted relative risk (RR) 6.6; 95% confidence interval (CI) 3.0 to 14.1), nonunion (36% vs 21%; adjusted RR 2.0; 95% CI 1.0 to 3.9) and osteolysis around the material, acetabulum or femur (61% vs 19%; adjusted RR 3.9; 95% CI 2.3 to 6.5). Cable breakage increased the risk of osteolysis to 83%. There was a trend towards foreign-body reaction and increased infection with cables. Clinical results did not differ between the groups.

In conclusion, we found a higher incidence of complications and a trend towards increased infection and foreign-body reaction with the use of cables.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1045 - 1050
1 Aug 2012
Malek IA King A Sharma H Malek S Lyons K Jones S John A

Plasma levels of cobalt and chromium ions and Metal Artefact Reduction Sequence (MARS)-MRI scans were performed on patients with 209 consecutive, unilateral, symptomatic metal-on-metal (MoM) hip arthroplasties. There was wide variation in plasma cobalt and chromium levels, and MARS-MRI scans were positive for adverse reaction to metal debris (ARMD) in 84 hips (40%). There was a significant difference in the median plasma cobalt and chromium levels between those with positive and negative MARS-MRI scans (p < 0.001). Compared with MARS-MRI as the potential reference standard for the diagnosis of ARMD, the sensitivity of metal ion analysis for cobalt or chromium with a cut-off of > 7 µg/l was 57%. The specificity was 65%, positive predictive value was 52% and the negative predictive value was 69% in symptomatic patients. A lowered threshold of > 3.5 µg/l for cobalt and chromium ion levels improved the sensitivity and negative predictive value to 86% and 74% but at the expense of specificity (27%) and positive predictive value (44%).

Metal ion analysis is not recommended as a sole indirect screening test in the surveillance of symptomatic patients with a MoM arthroplasty. The investigating clinicians should have a low threshold for obtaining cross-sectional imaging in these patients, even in the presence of low plasma metal ion levels.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 908 - 913
1 Jul 2012
Seo JG Moon YW Park SH Lee JH Kang HM Kim SM

Peri-prosthetic patellar fracture following resurfacing as part of total knee replacement (TKR) is an infrequent yet challenging complication. This case-control study was performed to identify clinical, radiological and surgical factors that increase the risk of developing a spontaneous patellar fracture after TKR. Patellar fractures were identified in 74 patients (88 knees) from a series of 7866 consecutive TKRs conducted between 1998 and 2009. After excluding those with a previous history of extensor mechanism realignment or a clear traumatic event, a metal-backed patella, any uncemented component or subsequent infection, the remaining 64 fractures were compared with a matched group of TKRs with an excellent outcome defined by the Knee Society score. The mean age of patients with a fracture was 70 years (51 to 81) at the time of TKR. Patellar fractures were detected at a mean of 13.4 months (2 to 84) after surgery. The incidence of patellar fracture was found to be strongly associated with the number of previous knee operations, greater pre-operative mechanical malalignment, smaller post-operative patellar tendon length, thinner post-resection patellar thickness, and a lower post-operative Insall-Salvati ratio.

An understanding of the risk factors associated with spontaneous patellar fracture following TKR provides a valuable insight into prevention of this challenging complication.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 901 - 907
1 Jul 2012
Esposito CI Walter WL Roques A Tuke MA Zicat BA Walsh WR Walter WK

We analysed 54 alumina ceramic-on-ceramic bearings from total hip replacements retrieved at one centre after a mean duration of 3.5 years (0.2 to 10.6) in situ. These implants were obtained from 54 patients (16 men and 38 women) with a mean age of 67 years (33 to 88) who underwent revision for a variety of reasons. Posterior edge loading was found in the majority of these retrievals (32 out of 54). Anterosuperior edge loading occurred less often but produced a higher rate of wear. Stripe wear on the femoral heads had a median volumetric wear rate of 0.2 mm3/year (0 to 7.2). The wear volume on the femoral heads corresponded to the width of edge wear on the matching liner. Anteversion of the acetabular component was found to be a more important determinant than inclination for wear in ceramic bearings. Posterior edge loading may be considered to be a normal occurrence in ceramic-on-ceramic bearings, with minimal clinical consequences. Edge loading should be defined as either anterosuperior or posterior, as each edge loading mechanism may result in different clinical implications.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1557 - 1566
1 Nov 2012
Jameson SS Kyle J Baker PN Mason J Deehan DJ McMurtry IA Reed MR

United Kingdom National Institute for Health and Clinical Excellence guidelines recommend the use of total hip replacement (THR) for displaced intracapsular fractures of the femoral neck in cognitively intact patients, who were independently mobile prior to the injury. This study aimed to analyse the risk factors associated with revision of the implant and mortality following THR, and to quantify risk. National Joint Registry data recording a THR performed for acute fracture of the femoral neck between 2003 and 2010 were analysed. Cox proportional hazards models were used to investigate the extent to which risk of revision was related to specific covariates. Multivariable logistic regression was used to analyse factors affecting peri-operative mortality (< 90 days). A total of 4323 procedures were studied. There were 80 patients who had undergone revision surgery at the time of censoring (five-year revision rate 3.25%, 95% confidence interval 2.44 to 4.07) and 137 patients (3.2%) patients died within 90 days. After adjusting for patient and surgeon characteristics, an increased risk of revision was associated with the use of cementless prostheses compared with cemented (hazard ratio (HR) 1.33, p = 0.021). Revision was independent of bearing surface and head size. The risk of mortality within 90 days was significantly increased with higher American Society of Anesthesiologists (ASA) grade (grade 3: odds ratio (OR) 4.04, p < 0.001; grade 4/5: OR 20.26, p < 0.001; both compared with grades 1/2) and older age (≥ 75 years: OR 1.65, p = 0.025), but reduced over the study period (9% relative risk reduction per year).

THR is a good option in patients aged < 75 years and with ASA 1/2. Cementation of the femoral component does not adversely affect peri-operative mortality but improves survival of the implant in the mid-term when compared with cementless femoral components. There are no benefits of using head sizes > 28 mm or bearings other than metal-on-polyethylene. More research is required to determine the benefits of THR over hemiarthroplasty in older patients and those with ASA grades > 2.