The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 310 - 315
1 Mar 2009
Olsen M Davis ET Waddell JP Schemitsch EH

We have investigated the accuracy of placement of the femoral component using imageless navigation in 100 consecutive Birmingham Hip Resurfacings. Pre-operative templating determined the native neck-shaft angle and planned stem-shaft angle of the implant. The latter were verified post-operatively using digital anteroposterior unilateral radiographs of the hip.

The mean neck-shaft angle determined before operation was 132.7° (118° to 160°). The mean planned stem-shaft angle was a relative valgus alignment of 9.7° (sd 2.6). The stem-shaft angle after operation differed from that planned by a mean of 2.8° (sd 2.0) and in 86% of cases the final angle measured within ± 5° of that planned. We had no instances of notching of the neck or varus alignment of the implant in our series. A learning curve was observed in the time taken for navigation, but not for accurate placement of the implant.

Navigation in hip resurfacing may afford the surgeon a reliable and accurate method of placement of the femoral component.


Bone & Joint 360
Vol. 5, Issue 4 | Pages 4 - 15
1 Aug 2016
Sehat K

Anatomical total knee arthroplasty alignment versus conventional mechanical alignment; or a combination?


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 902 - 902
1 Jun 2010


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1121 - 1127
1 Sep 2008
Sikorski JM

The advent of computer-assisted knee replacement surgery has focused interest on the alignment of the components. However, there is confusion at times between the alignment of the limb as a whole and that of the components. The interaction between them is discussed in this article. Alignment is expressed relative to some reference axis or plane and measurements will vary depending on what is selected as the reference. The validity of different reference axes is discussed. Varying prosthetic alignment has direct implications for surrounding soft-tissue tension. In this context the interaction between alignment and soft-tissue balance is explored and the current knowledge of the relationship between alignment and outcome is summarised.


Bone & Joint 360
Vol. 5, Issue 1 | Pages 14 - 16
1 Feb 2016


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 629 - 633
1 May 2011
Hirschmann MT Konala P Amsler F Iranpour F Friederich NF Cobb JP

We studied the intra- and interobserver reliability of measurements of the position of the components after total knee replacement (TKR) using a combination of radiographs and axial two-dimensional (2D) and three-dimensional (3D) reconstructed CT images to identify which method is best for this purpose.

A total of 30 knees after primary TKR were assessed by two independent observers (an orthopaedic surgeon and a radiologist) using radiographs and CT scans. Plain radiographs were highly reliable at measuring the tibial slope, but showed wide variability for all other measurements; 2D-CT also showed wide variability. 3D-CT was highly reliable, even when measuring rotation of the femoral components, and significantly better than 2D-CT. Interobserver variability in the measurements on radiographs were good (intraclass correlation coefficient (ICC) 0.65 to 0.82), but rotational measurements on 2D-CT were poor (ICC 0.29). On 3D-CT they were near perfect (ICC 0.89 to 0.99), and significantly more reliable than 2D-CT (p < 0.001).

3D-reconstructed images are sufficiently reliable to enable reporting of the position and orientation of the components. Rotational measurements in particular should be performed on 3D-reconstructed CT images. When faced with a poorly functioning TKR with concerns over component positioning, we recommend 3D-CT as the investigation of choice.


The Bone & Joint Journal
Vol. 96-B, Issue 4 | Pages 467 - 472
1 Apr 2014
Nakamura E Okamoto N Nishioka H Karasugi T Hirose J Mizuta H

We report the long-term outcome of 33 patients (37 knees) who underwent proximal tibial open-wedge osteotomy with hemicallotasis (HCO) for medial osteoarthritis of the knee between 1995 and 2000. Among these, 29 patients with unilateral HCO were enrolled and 19 were available for review at a mean of 14.2 years (10 to 15.7) post-operatively. For these 19 patients, the mean Hospital for Special Surgery knee score was 60 (57 to 62) pre-operatively and 85 (82 to 87) at final follow-up (p < 0.001; paired t-test). The femorotibial angle and tibial inclination angle (IA) were measured at short-term follow-up, one to four years post-operatively, and showed no significant subsequent changes. The clinical scores and radiological measurements showed little change over time. One patient required conversion to total knee replacement during this time. These results suggest that the coronal angle achieved at operation is maintained at long-term follow up after HCO without alteration of the IA, providing a good long-term clinical outcome.

Cite this article: Bone Joint J 2014;96-B:467–72.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1210 - 1216
1 Sep 2011
Mitsuyasu H Matsuda S Fukagawa S Okazaki K Tashiro Y Kawahara S Nakahara H Iwamoto Y

We investigated whether the extension gap in total knee replacement (TKR) would be changed when the femoral component was inserted. The extension gap was measured with and without the femoral component in place in 80 patients with varus osteoarthritis undergoing posterior-stabilised TKR. The effect of a post-operative increase in the size of the femoral posterior condyles was also evaluated. The results showed that placement of the femoral component significantly reduced the medial and lateral extension gaps by means of 1.0 mm and 0.9 mm, respectively (p < 0.0001). The extension gap was reduced when a larger femoral component was selected relative to the thickness of the resected posterior condyle. When the post-operative posterior lateral condyle was larger than that pre-operatively, 17 of 41 knees (41%) showed a decrease in the extension gap of > 2.0 mm. When a specially made femoral trial component with a posterior condyle enlarged by 4 mm was tested, the medial and lateral extension gaps decreased further by means of 2.1 mm and 2.8 mm, respectively.

If the thickness of the posterior condyle is expected to be larger than that pre-operatively, it should be recognised that the extension gap is likely to be altered. This should be taken into consideration when preparing the extension gap.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1064 - 1068
1 Aug 2013
Cavaignac E Lafontan V Reina N Pailhé R Warmy M Laffosse JM Chiron P

The significance of weight in the indications for unicompartmental knee replacement (UKR) is unclear. Our hypothesis was that weight does not affect the long-term rate of survival of UKRs.

We undertook a retrospective study of 212 UKRs at a mean follow-up of 12 years (7 to 22). The patients were distributed according to body mass index (BMI; < vs ≥ 30 kg/m2) and weight (< vs ≥ 82 kg). Kaplan-Meier survivorship analysis was performed and ten-year survival rates were compared between the sub-groups. Multimodal regression analysis determined the impact of the various theoretical contraindications on the long-term rate of survival of UKR.

The ten-year rates of survival were similar in the two weight subgroups (≥ 82 kg: 93.5% (95% confidence interval (CI) 66.5 to 96.3); < 82 kg: 92.5% (95% CI 82.5 to 94.1)) and also in the two BMI subgroups (≥ 30 kg/m2: 92% (95% CI 82.5 to 95.3); < 30 kg/m2: 94% (95% CI 78.4 to 95.9)). Multimodal regression analysis revealed that weight plays a part in reducing the risk of revision with a relative risk of 0.387, although this did not reach statistical significance (p = 0.662). The results relating weight and BMI to the clinical outcome were not statistically significant. Thus, this study confirms that weight does not influence the long-term rate of survival of UKR.

Cite this article: Bone Joint J 2013;95-B:1064–8.


Bone & Joint 360
Vol. 4, Issue 4 | Pages 2 - 7
1 Aug 2015
Nicol S Jackson M Monsell F

This review explores recent advances in fixator design and used in contemporary orthopaedic practice including the management of bone loss, complex deformity and severe isolated limb injury.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 5 | Pages 604 - 611
1 May 2009
Reay E Wu J Holland J Deehan D

We describe a cohort of patients with a high rate of mid-term failure following Kinemax Plus total knee replacement inserted between 1998 and 2001. This implant has been recorded as having a survival rate of 96% at ten years. However, in our series the survival rate was 75% at nine years. This was also significantly lower than that of subsequent consecutive series of PFC Sigma knee replacements performed by the same surgeon. No differences were found in the clinical and radiological parameters between the two groups. At revision the most striking finding was polyethylene wear. An independent analysis of the polyethylene components was therefore undertaken. Scanning electron microscopy revealed type 2 fusion defects in the ultra-high molecular weight polyethylene (UHMWPE), which indicated incomplete boundary fusion. Other abnormalities consistent with weak UHMWPE particle interface strength were present in both the explanted inserts and in unused inserts from the same period.

We consider that these type 2 fusion defects are the cause of the early failure of the Kinemax implants. This may represent a manufacturing defect resulting in a form of programmed polyethylene failure.


Bone & Joint 360
Vol. 1, Issue 1 | Pages 12 - 13
1 Feb 2012


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1448 - 1453
1 Nov 2006
Chen AF Alan RK Redziniak DE Tria AJ

The aim of this study was to compare the results in patients having a quadriceps sparing total knee replacement (TKR) with those undergoing a standard TKR at a minimum follow-up of two years.

All patients who had a TKR with a high-flex posterior-stabilised prosthesis prior to December 2002 were reviewed retrospectively. There were 57 patients available for follow-up. Those with a quadriceps sparing TKR had less pain peri-operatively with a greater degree of flexion at all the post-operative visits and at the final follow-up, but their operations took longer, with less accurate radiological alignment. There was no difference in the complications and in the Knee Society scores between the two groups at the final follow-up.

Total knee replacement through a quadriceps sparing approach has some peri-operative advantages over the standard incision. At a minimum follow-up of two years the clinical results were similar to those with a standard incision, but the radiological outcomes of the quadriceps sparing group were inferior.


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 292 - 299
1 Mar 2015
Karthik K Colegate-Stone T Dasgupta P Tavakkolizadeh A Sinha J

The use of robots in orthopaedic surgery is an emerging field that is gaining momentum. It has the potential for significant improvements in surgical planning, accuracy of component implantation and patient safety. Advocates of robot-assisted systems describe better patient outcomes through improved pre-operative planning and enhanced execution of surgery. However, costs, limited availability, a lack of evidence regarding the efficiency and safety of such systems and an absence of long-term high-impact studies have restricted the widespread implementation of these systems. We have reviewed the literature on the efficacy, safety and current understanding of the use of robotics in orthopaedics.

Cite this article: Bone Joint J 2015; 97-B:292–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1321 - 1329
1 Oct 2012
Sarmah SS Patel S Hossain FS Haddad FS

Radiological assessment of total and unicompartmental knee replacement remains an essential part of routine care and follow-up. Appreciation of the various measurements that can be identified radiologically is important. It is likely that routine plain radiographs will continue to be used, although there has been a trend towards using newer technologies such as CT, especially in a failing knee, where it provides more detailed information, albeit with a higher radiation exposure.

The purpose of this paper is to outline the radiological parameters used to evaluate knee replacements, describe how these are measured or classified, and review the current literature to determine their efficacy where possible.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 475 - 480
1 Apr 2009
Matsumoto T Kuroda R Kubo S Muratsu H Mizuno K Kurosaka M

We have developed a new tensor for total knee replacements which is designed to assist with soft-tissue balancing throughout the full range of movement with a reduced patellofemoral joint. Using this tensor in 40 patients with osteoarthritis we compared the intra-operative joint gap in cruciate-retaining and posterior-stabilised total knee replacements at 0°, 10°, 45°, 90° and 135° of flexion, with the patella both everted and reduced.

While the measurement of the joint gap with a reduced patella in posterior-stabilised knees increased from extension to flexion, it remained constant for cruciate-retaining joints throughout a full range of movement. The joint gaps at deep knee flexion were significantly smaller for both types of prosthetic knee when the patellofemoral joint was reduced (p < 0.05).


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 56 - 61
1 Jan 2012
Kawahara S Matsuda S Fukagawa S Mitsuyasu H Nakahara H Higaki H Shimoto T Iwamoto Y

In posterior stabilised total knee replacement (TKR) a larger femoral component is sometimes selected to manage the increased flexion gap caused by resection of the posterior cruciate ligament. However, concerns remain regarding the adverse effect of the increased anteroposterior dimensions of the femoral component on the patellofemoral (PF) joint. Meanwhile, the gender-specific femoral component has a narrower and thinner anterior flange and is expected to reduce the PF contact force. PF contact forces were measured at 90°, 120°, 130° and 140° of flexion using the NexGen Legacy Posterior Stabilized (LPS)-Flex Fixed Bearing Knee system using Standard, Upsized and Gender femoral components during TKR. Increasing the size of the femoral component significantly increased mean PF forces at 120°, 130° and 140° of flexion (p = 0.005, p < 0.001 and p < 0.001, respectively). No difference was found in contact force between the Gender and the Standard components. Among the patients who had overhang of the Standard component, mean contact forces with the Gender component were slightly lower than those of the Standard component, but no statistical difference was found at 90°, 120°, 130° or 140° of flexion (p = 0.689, 0.615, 0.253 and 0.248, respectively).

Upsized femoral components would increase PF forces in deep knee flexion. Gender-specific implants would not reduce PF forces.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1223 - 1231
1 Sep 2011
Babazadeh S Dowsey MM Swan JD Stoney JD Choong PFM

The role of computer-assisted surgery in maintaining the level of the joint in primary knee joint replacement (TKR) has not been well defined. We undertook a blinded randomised controlled trial comparing joint-line maintenance, functional outcomes, and quality-of-life outcomes between patients undergoing computer-assisted and conventional TKR. A total of 115 patients were randomised (computer-assisted, n = 55; conventional, n = 60).

Two years post-operatively no significant correlation was found between computer-assisted and conventional surgery in terms of maintaining the joint line. Those TKRs where the joint line was depressed post-operatively improved the least in terms of functional scores. No difference was detected in terms of quality-of-life outcomes. Change in joint line was found to be related to change in alignment. Change in alignment significantly affects change in joint line and functional scores.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 61 - 65
1 Jan 2010
Morgan SS Brooke B Harris NJ

We present the outcomes in 38 consecutive patients who had total ankle replacement using the Ankle Evolution System with a minimum follow-up of four years. Pain and function were assessed using the American Orthopaedic Foot and Ankle Society (AOFAS) score and regular standardised anteroposterior and lateral weight-bearing radiographs were obtained. Patient satisfaction and complications were recorded and the survival of the implants was demonstrated by the Kaplan-Meier method.

The mean follow-up was for 57.8 months (48 to 80). The cumulative survival rate at six years was 94.7% (95% confidence interval 80.3 to 98.7). The mean total AOFAS score was 88.1 (53 to 100). The mean score for pain was 35.8 (20 to 40). Ten patients presented with edge-loading of whom nine had corrective surgery. Two ankles were revised, one to an arthrodesis and the other to replace the tibial component. Nine patients showed radiological evidence of osteolysis. They had minimal non-progressive symptoms and further surgery was not undertaken. Nevertheless, the concerns about osteolysis led to the implant being withdrawn by the manufacturer.

The medium-term results of the ankle evolution system ankle replacement are satisfactory with high patient satisfaction, but the rate of osteolysis is of some concern. The long-term benefit of this procedure has yet to be determined.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1232 - 1236
1 Sep 2009
Fahmy MAL Hatata MZ Al-Seesi H

We describe a method of pinning extension supracondylar fractures of the humerus in children. Following closed reduction, a posterior intrafocal wire is inserted and a second lateral wire added when needed for rotational stability. Between May 2002 and November 2005 we performed this technique in 69 consecutive patients. A single posterior wire was used in 29 cases, and two wires in 40. The mean follow-up was two years (21 to 30 months). The results were assessed according to Flynn’s criteria. In the single-wire group there were 21 excellent, five good and one poor result. Two patients were lost to follow-up. In the two-wire group there were 32 excellent, two good and three poor results. Three were lost to follow-up. The poor results were due to a failure to achieve adequate reduction, fixation or both.

We conclude that the intact posterior periosteal hinge can be used successfully in the clinical situation, giving results that compare well with other techniques of pinning. The posterior route offers an attractive alternative method for fixation of supracondylar fractures of the humerus in children.