We reviewed nine patients at a mean period of 11 years (6 to 16) after curettage and cementing of a giant-cell tumour around the knee to determine if there were any long-term adverse effects on the cartilage. Plain radiography, MRI, delayed gadolinium-enhanced MRI of the cartilage and measurement of the serum level of cartilage oligomeric matrix protein were carried out. The functional outcome was evaluated using the Lysholm knee score. Each patient was physically active and had returned to their previous occupation. Most participated in recreational sports or exercise. The mean Lysholm knee score was 92 (83 to 100). Only one patient was found to have cartilage damage adjacent to the cement. This patient had a history of intra-articular fracture and local recurrence, leading to degenerative changes. Interpretation of the data obtained from delayed gadolinium-enhanced MRI of the cartilage was difficult, with variation in the T1 values which did not correlate with the clinical or radiological findings. We did not find it helpful in the early diagnosis of degeneration of cartilage. We also found no obvious correlation between the serum cartilage oligomeric matrix protein level and the radiological and MR findings, function, time after surgery and the age of the patient. In summary, we found no evidence that the long-term presence of cement close to the knee joint was associated with the development of degenerative osteoarthritis.
We reviewed the results of 71 revisions of the acetabular component in total hip replacement, using impaction of bone allograft. The mean follow-up was 7.2 years (1.6 to 9.7). All patients were assessed according to the American Academy of Orthopedic Surgeons (AAOS) classification of bone loss, the amount of bone graft required, thickness of the graft layer, signs of graft incorporation and use of augmentation. A total of 20 acetabular components required re-revision for aseptic loosening, giving an overall survival of 72% (95% CI, 54.4 to 80.5). Of these failures, 14 (70%) had an AAOS type III or IV bone defect. In the failed group, poor radiological and histological graft incorporation was seen. These results suggest that impaction allografting in acetabular revision with severe bone defects may have poorer results than have previously been reported.
We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total knee prostheses in 32 young patients with primary malignant bone tumours. Initial stable fixation, especially in rotation, was achieved by precise fit of the stem into the reamed endosteum, before cementation. Complementary fixation, especially in migration and rotation, was obtained by pressurised antibiotic-loaded cement. The mean age at operation was 33 years (13 to 61). No patient was lost to follow-up; 13 patients died and the 19 survivors were examined at a mean follow-up of 12.5 years (4 to 21). Standard revision press-fit cemented stems were used on the side of the joint which was not involved with tumour (26 tibial and six femoral), on this side there was no loosening or osteolysis and stem survival was 100%. On the reconstruction side, custom-made press-fit stems were used and the survival rate, with any cause for revision as an end point, was 88%, but 97% for loosening or osteolysis. This longevity is similar to that achieved at 20 years with the Charnley-Kerboull primary total hip replacement with press-fit cemented femoral components. We recommend this type of fixation when extensive reconstruction of the knee is required. It may also be suitable for older patients requiring revision of a total knee replacement or in difficult situations such as severe deformity and complex articular fractures.
Our aim was to determine the success rate of repeated debridement and two-stage cementless revision arthroplasty according to the type of infected total hip replacement (THR). We enrolled 294 patients (294 hips) with an infected THR in the study. There were 222 men and 72 women with a mean age of 55.1 years (24.0 to 78.0). The rate of control of infection after the initial treatment and after repeated debridement and two-stage revisions was determined. The clinical (Harris hip score) and radiological results were evaluated. The mean follow-up was 10.4 years (5.0 to 14.0). The eventual rate of control of infection was 100.0% for early superficial post-operative infection, 98.4% for early deep post-operative infection, 98.5% for late chronic infection and 91.0% for acute haematogenous infection. Overall, 288 patients (98%) maintained a functioning THR at the latest follow-up. All the allografts appeared to be united and there were no failures. These techniques effectively controlled infection and maintained a functional THR with firm fixation in most patients. Repeated debridement and two-stage or repeated two-stage revisions further improved the rate of control of infection after the initial treatment and increased the likelihood of maintaining a functional THR.
We investigated 219 revisions of total hip replacement (THR) in 211 patients using a collarless double-taper cemented femoral component. The mean age of the patients was 72 years (30 to 90). The 137 long and 82 standard length stems were analysed separately. The mean follow-up was six years (2 to 18), and no patient was lost to follow-up. Survival of the long stems to re-revision for aseptic loosening at nine years was 98% (95% confidence interval (CI) 94 to 100), and for the standard stems was 93% (95% CI 85 to 100). At five years, one long stem was definitely loose radiologically and one standard stem was probably loose. Pre-operative femoral bone deficiency did not influence the results for the long stems, and corrective femoral osteotomy was avoided, as were significant subsidence, major stress shielding and persistent thigh pain. Because of these reliable results, cemented long collarless double-taper femoral components are recommended for routine revision THR in older patients.
Clinical experience of impaction bone grafting for revision knee arthroplasty is limited, with initial stability of the tibial tray emerging as a major concern. The length of the stem and its diameter have been altered to improve stability. Our aim was to investigate the effect of the type of stem, support of the rim and graft impaction on early stability of the tray. We developed a system for impaction grafting of trays which we used with morsellised bone in artificial tibiae. Trays with short, long thick or long thin stems were implanted, with or without support of the rim. They were cyclically loaded while measuring relative movement. Long-stemmed trays migrated 4.5 times less than short-stemmed trays, regardless of diameter. Those with support migrated 2.8 times less than those without. The migration of short-stemmed trays correlated inversely with the density of the impacted groups. That of impaction-grafted tibial trays was in the range reported for uncemented primary trays. Movements of short-stemmed trays without cortical support were largest and sensitive to the degree of compaction of the graft. If support of the rim was sufficient or a long stem was used, impacted morsellised bone graft achieved adequate initial stability.
We describe the mid-term results of a prospective study of total knee replacement in severe valgus knees using an osteotomy of the lateral femoral condyle and computer navigation. There were 15 knees with a mean valgus deformity of 21° (17° to 27°) and a mean follow-up of 28 months (24 to 60). A cemented, non-constrained fixed bearing, posterior-cruciate-retaining knee prosthesis of the same design was used in all cases (Columbus-B. Braun; Aesculap, Tuttlingen, Germany). All the knees were corrected to a mean of 0.5° of valgus (0° to 2°). Flexion of the knee had been limited to a mean of 85° (75° to 110°) pre-operatively and improved to a mean of 105° (90° to 130°) after operation. The mean Knee Society score improved from 37 (30 to 44) to 90 points (86 to 94). Osteotomy of the lateral femoral condyle combined with computer-assisted surgery gave an excellent mid-term outcome in patients undergoing total knee replacement in the presence of severe valgus deformity.
Metal meshes are used in revision surgery of the hip to contain impacted bone grafts in cases with cortical or calcar defects in order to provide rotational stability to the stem. However, the viability of bone allografts under these metal meshes has been uncertain. We describe the histological appearances of biopsies obtained from impacted bone allografts to the calcar contained by a metal mesh in two femoral reconstructions which needed further surgery at 24 and 33 months after the revision procedure. A line of osteoid and viable new bone was observed on the surface of necrotic trabeculae. Active bone marrow between these trabeculae showed necrotic areas in some medullary spaces with reparative fibrous tissue and isolated reactive lymphocytes. This is interpreted as reparative changes after revascularisation of the cancellous allografts. These pathological findings are similar to those reported in allografts contained by cortical host bone and support the hypothesis that incorporation of morcellised bone under metal meshes is not affected by these devices.
Between 1980 and 2000, 63 support rings were used in the management of acetabular deficiency in a series of 60 patients, with a mean follow-up of 8.75 years (2 months to 23.8 years). There was a minimum five-year follow-up for successful reconstructions. The indication for revision surgery was aseptic loosening in 30 cases and infection in 33. All cases were Paprosky III defects; IIIA in 33 patients (52.4%) and IIIB in 30 (47.6%), including four with pelvic dissociation. A total of 26 patients (43.3%) have died since surgery, and 34 (56.7%) remain under clinical review. With acetabular revision for infection or aseptic loosening as the definition of failure, we report success in 53 (84%) of the reconstructions. A total of 12 failures (19%) required further surgery, four (6.3%) for aseptic loosening of the acetabular construct, six (9.5%) for recurrent infection and two (3.2%) for recurrent dislocation requiring captive components. Complications, seen in 11 patients (18.3%), included six femoral or sciatic neuropraxias which all resolved, one grade III heterotopic ossification, one on-table acetabular revision for instability, and three early post-operative dislocations managed by manipulation under anaesthesia, with no further instability. We recommend support rings and morcellised bone graft for significant acetabular bone deficiency that cannot be reconstructed using mesh.
Failure of fixation is a common problem in the treatment of osteoporotic fractures around the hip. The reinforcement of bone stock or of fixation of the implant may be a solution. Our study assesses the existing evidence for the use of bone substitutes in the management of these fractures in osteoporotic patients. Relevant publications were retrieved through Medline research and further scrutinised. Of 411 studies identified, 22 met the inclusion criteria, comprising 12 experimental and ten clinical reports. The clinical studies were evaluated with regard to their level of evidence. Only four were prospective and randomised. Polymethylmethacrylate and calcium-phosphate cements increased the primary stability of the implant-bone construct in all experimental and clinical studies, although there was considerable variation in the design of the studies. In randomised, controlled studies, augmentation of intracapsular fractures of the neck of the femur with calcium-phosphate cement was associated with poor long-term results. There was a lack of data on the long-term outcome for trochanteric fractures. Because there were only a few, randomised, controlled studies, there is currently poor evidence for the use of
Hip resurfacing is a bone-conserving procedure with respect to proximal femoral resection, but there is debate in the literature as to whether the same holds true for the acetabulum. We have investigated whether the Birmingham hip resurfacing conserves acetabular bone. Between 1998 and 2005, 500 Birmingham hip resurfacings were performed by two surgeons. Between 1996 and 2005 they undertook 700 primary hip replacements, with an uncemented acetabular component. These patients formed the clinical material to compare acetabular component sizing. The Birmingham hip resurfacing group comprised 350 hips in men and 150 hips in women. The uncemented total hip replacement group comprised 236 hips in men and 464 hips in women. Age- and gender-matched analysis of a cohort of patients for the sizes of the acetabular components required for the two types of replacement was also undertaken. Additionally, an analysis of the sizes of the components used by each surgeon was performed. For age-matched women, the mean outside diameter of the Birmingham hip resurfacing acetabular components was 2.03 mm less than that of the acetabular components in the uncemented total hip replacements (p <
0.0001). In similarly matched men there was no significant difference (p = 0.77). A significant difference was also found between the size of acetabular components used by the two surgeons for Birmingham hip resurfacing for both men (p = 0.0015) and women (p = 0.001). In contrast, no significant difference was found between the size of acetabular components used by the two surgeons for uncemented total hip replacement in either men or women (p = 0.06 and p = 0.14, respectively). This suggests that variations in acetabular preparation also influence acetabular component size in hip resurfacing.
Interfacial defects between the cement mantle and a hip implant may arise from constrained shrinkage of the cement or from air introduced during insertion of the stem. Shrinkage-induced interfacial porosity consists of small pores randomly located around the stem, whereas introduced interfacial gaps are large, individual and less uniformly distributed areas of stem-cement separation. Using a validated CT-based technique, we investigated the extent, morphology and distribution of interfacial gaps for two types of stem, the Charnley-Kerboul and the Lubinus SPII, and for two techniques of implantation, line-to-line and undersized. The interfacial gaps were variable and involved a mean of 6.43% (
Resurfacing arthroplasty of the hip is being used increasingly as an alternative to total hip replacement, especially for young active patients. There is concern about necrosis of the femoral head after resurfacing which can result in fracture and loosening. Most systems use a cemented femoral component, with the potential for thermal necrosis of the cancellous bone of the reamed femoral head. We used thermal probes to record temperatures close to the cement-bone interface during resurfacing arthroplasty. The maximum temperature recorded at the cement-bone interface in four cases was approximately 68°C which was higher than that reported to kill osteocytes. A modified surgical technique using insertion of a suction cannula into the lesser trochanter, generous pulsed lavage and early reduction of the joint significantly reduced the maximum recorded cancellous bone temperature to approximately 36°C in five cases (p = 0.014). We recommend the modified technique since it significantly reduces temperatures at the cement-bone interface.
Polymethylmethacrylate remains one of the most enduring materials in orthopaedic surgery. It has a central role in the success of total joint replacement and is also used in newer techniques such as percutaneous vertebroplasty and kyphoplasty. This article describes the current uses and limitations of polymethylmethacrylate in orthopaedic surgery. It focuses on its mechanical and chemical properties and links these to its clinical performance. The behaviour of antibiotic-loaded
Finite element analysis was used to examine the initial stability after hip resurfacing and the effect of the procedure on the contact mechanics at the articulating surfaces. Models were created with the components positioned anatomically and loaded physiologically through major muscle forces. Total micromovement of less than 10 μm was predicted for the press-fit acetabular components models, much below the 50 μm limit required to encourage osseointegration. Relatively high compressive acetabular and contact stresses were observed in these models. The press-fit procedure showed a moderate influence on the contact mechanics at the bearing surfaces, but produced marked deformation of the acetabular components. No edge contact was predicted for the acetabular components studied. It is concluded that the frictional compressive stresses generated by the 1 mm to 2 mm interference-fit acetabular components, together with the minimal micromovement, would provide adequate stability for the implant, at least in the immediate post-operative situation.
The effects of the method of fixation and interface conditions on the biomechanics of the femoral component of the Birmingham hip resurfacing arthroplasty were examined using a highly detailed three-dimensional computer model of the hip. Stresses and strains in the proximal femur were compared for the natural femur and for the femur resurfaced with the Birmingham hip resurfacing. A comparison of cemented
A modular layered acetabular component (metal-polyethylene-ceramic) was developed in Japan for use in alumina ceramic-on-ceramic total hip replacement. Between May 1999 and July 2000, we performed 35 alumina ceramic-on-ceramic total hip replacements in 30 consecutive patients, using this layered component and evaluated the clinical and radiological results over a mean follow-up of 5.8 years (5 to 6.5). A total of six hips underwent revision, one for infection, two for dislocation with loosening of the acetabular component, two for alumina liner fractures and one for component dissociation with pelvic osteolysis. There were no fractures of the ceramic heads, and no loosening of the femoral or acetabular component in the unrevised hips was seen at final follow-up. Osteolysis was not observed in any of the unrevised hips. The survivorship analysis at six years after surgery was 83%. The layered acetabular component in our experience, has poor durability because of unexpected mechanical failures including alumina liner fracture and component dissociation.
We reviewed 142 consecutive primary total hip replacements implanted into 123 patients between 1988 and 1993 using the Exeter Universal femoral stem. A total of 74 patients (88 hips) had survived for ten years or more and were reviewed at a mean of 12.7 years (10 to 17). There was no loss to follow-up. The rate of revision of the femoral component for aseptic loosening and osteolysis was 1.1% (1 stem), that for revision for any cause was 2.2% (2 stems), and for re-operation for any cause was 21.6% (19 hips). Re-operation was because of failure of the acetabular component in all but two hips. All but one femoral component subsided within the cement mantle to a mean of 1.52 mm (0 to 8.3) at the final follow-up. One further stem had subsided excessively (8 mm) and had lucent lines at the cement-stem and cement-bone interfaces. This was classified as a radiological failure and is awaiting revision. One stem was revised for deep infection and one for excessive peri-articular osteolysis. Defects of the cement mantle (Barrack grade C and D) were found in 28% of stems (25 hips), associated with increased subsidence (p = 0.01), but were not associated with endosteal lysis or failure. Peri-articular osteolysis was significantly related to the degree of polyethylene wear (p <
0.001), which was in turn associated with a younger age (p = 0.01) and male gender (p <
0.001). The use of the Exeter metal-backed acetabular component was a notable failure with 12 of 32 hips (37.5%) revised for loosening. The Harris-Galante components failed with excessive wear, osteolysis and dislocation with 15% revised (5 of 33 hips). Only one of 23 hips with a cemented Elite component (4%) was revised for loosening and osteolysis. Our findings show that the Exeter Universal stem implanted outside the originating centre has excellent medium-term results.
The use of impaction bone grafting during revision arthroplasty of the hip in the presence of cortical defects has a high risk of post-operative fracture. Our laboratory study addressed the effect of extramedullary augmentation and length of femoral stem on the initial stability of the prosthesis and the risk of fracture. Cortical defects in plastic femora were repaired using either surgical mesh without extramedullary augmentation, mesh with a strut graft or mesh with a plate. After bone impaction, standard or long-stem Exeter prostheses were inserted, which were tested by cyclical loading while measuring defect strain and migration of the stem. Compared with standard stems without extramedullary augmentation, defect strains were 31% lower with longer stems, 43% lower with a plate and 50% lower with a strut graft. Combining extramedullary augmentation with a long stem showed little additional benefit (p = 0.67). The type of repair did not affect the initial stability. Our results support the use of impaction bone grafting and extramedullary augmentation of diaphyseal defects after mesh containment.