Whittaker RK Hothi HS Meswania JM Berber R Blunn GW Skinner JA Hart AJ

Aims

Surgeons have commonly used modular femoral heads and stems from different manufacturers, although this is not recommended by orthopaedic companies due to the different manufacturing processes.

We compared the rate of corrosion and rate of wear at the trunnion/head taper junction in two groups of retrieved hips; those with mixed manufacturers (MM) and those from the same manufacturer (SM).

Materials and Methods

We identified 151 retrieved hips with large-diameter cobalt-chromium heads; 51 of two designs that had been paired with stems from different manufacturers (MM) and 100 of seven designs paired with stems from the same manufacturer (SM). We determined the severity of corrosion with the Goldberg corrosion score and the volume of material loss at the head/stem junction. We used multivariable statistical analysis to determine if there was a significant difference between the two groups.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 592 - 600
1 May 2017
Matharu GS Nandra RS Berryman F Judge A Pynsent PB Dunlop DJ

Aims

To determine ten-year failure rates following 36 mm metal-on-metal (MoM) Pinnacle total hip arthroplasty (THA), and identify predictors of failure.

Patients and Methods

We retrospectively assessed a single-centre cohort of 569 primary 36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012 according to Medicines and Healthcare Products Regulation Agency recommendations. All-cause failure rates (all-cause revision, and non-revised cross-sectional imaging failures) were calculated, with predictors for failure identified using multivariable Cox regression.


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 892 - 900
1 Jul 2016
Atrey A Heylen S Gosling O Porteous MJL Haddad FS

Joint replacement of the hip and knee remain very satisfactory operations. They are, however, expensive. The actual manufacturing of the implant represents only 30% of the final cost, while sales and marketing represent 40%. Recently, the patents on many well established and successful implants have expired. Companies have started producing and distributing implants that purport to replicate existing implants with good long-term results.

The aims of this paper are to assess the legality, the monitoring and cost saving implications of such generic implants. We also assess how this might affect the traditional orthopaedic implant companies.

Cite this article: Bone Joint J 2016;98-B:892–900.


Bone & Joint Research
Vol. 7, Issue 4 | Pages 263 - 273
1 Apr 2018
Ferreira E Porter RM

Large bone defects remain a tremendous clinical challenge. There is growing evidence in support of treatment strategies that direct defect repair through an endochondral route, involving a cartilage intermediate. While culture-expanded stem/progenitor cells are being evaluated for this purpose, these cells would compete with endogenous repair cells for limited oxygen and nutrients within ischaemic defects. Alternatively, it may be possible to employ extracellular vesicles (EVs) secreted by culture-expanded cells for overcoming key bottlenecks to endochondral repair, such as defect vascularization, chondrogenesis, and osseous remodelling. While mesenchymal stromal/stem cells are a promising source of therapeutic EVs, other donor cells should also be considered. The efficacy of an EV-based therapeutic will likely depend on the design of companion scaffolds for controlled delivery to specific target cells. Ultimately, the knowledge gained from studies of EVs could one day inform the long-term development of synthetic, engineered nanovesicles. In the meantime, EVs harnessed from in vitro cell culture have near-term promise for use in bone regenerative medicine. This narrative review presents a rationale for using EVs to improve the repair of large bone defects, highlights promising cell sources and likely therapeutic targets for directing repair through an endochondral pathway, and discusses current barriers to clinical translation.

Cite this article: E. Ferreira, R. M. Porter. Harnessing extracellular vesicles to direct endochondral repair of large bone defects. Bone Joint Res 2018;7:263–273. DOI: 10.1302/2046-3758.74.BJR-2018-0006.


The Bone & Joint Journal
Vol. 100-B, Issue 1_Supple_A | Pages 55 - 61
1 Jan 2018
Kim K Elbuluk A Yu S Iorio R

Aims

The aim of this study was to determine the optimal regimen for the management of pain following total knee arthroplasty (TKA) by comparing the outcomes and cost-effectiveness of different protocols implemented at a large, urban, academic medical centre.

Patients and Methods

Between September 2013 and September 2015, we used a series of modifications to our standard regimen for the management of pain after TKA. In May 2014, there was a department-wide transition from protocols focused on femoral nerve blocks (FNB) to periarticular injections of liposomal bupivacaine. In February 2015, patient-controlled analgesia (PCA) was removed from the protocol while continuing liposomal bupivacaine injections. Quality measures and hospital costs were compared between the three protocols.


Bone & Joint 360
Vol. 6, Issue 5 | Pages 33 - 35
1 Oct 2017


Bone & Joint Research
Vol. 7, Issue 1 | Pages 28 - 35
1 Jan 2018
Huang H Nightingale RW Dang ABC

Objectives

Loss of motion following spine segment fusion results in increased strain in the adjacent motion segments. However, to date, studies on the biomechanics of the cervical spine have not assessed the role of coupled motions in the lumbar spine. Accordingly, we investigated the biomechanics of the cervical spine following cervical fusion and lumbar fusion during simulated whiplash using a whole-human finite element (FE) model to simulate coupled motions of the spine.

Methods

A previously validated FE model of the human body in the driver-occupant position was used to investigate cervical hyperextension injury. The cervical spine was subjected to simulated whiplash exposure in accordance with Euro NCAP (the European New Car Assessment Programme) testing using the whole human FE model. The coupled motions between the cervical spine and lumbar spine were assessed by evaluating the biomechanical effects of simulated cervical fusion and lumbar fusion.


Bone & Joint Research
Vol. 5, Issue 11 | Pages 577 - 585
1 Nov 2016
Hase E Sato K Yonekura D Minamikawa T Takahashi M Yasui T

Objectives

This study aimed to evaluate the histological and mechanical features of tendon healing in a rabbit model with second-harmonic-generation (SHG) imaging and tensile testing.

Materials and Methods

A total of eight male Japanese white rabbits were used for this study. The flexor digitorum tendons in their right leg were sharply transected, and then were repaired by intratendinous stitching. At four weeks post-operatively, the rabbits were killed and the flexor digitorum tendons in both right and left legs were excised and used as specimens for tendon healing (n = 8) and control (n = 8), respectively. Each specimen was examined by SHG imaging, followed by tensile testing, and the results of the two testing modalities were assessed for correlation.


Bone & Joint Research
Vol. 6, Issue 10 | Pages 577 - 583
1 Oct 2017
Sallent A Vicente M Reverté MM Lopez A Rodríguez-Baeza A Pérez-Domínguez M Velez R

Objectives

To assess the accuracy of patient-specific instruments (PSIs) versus standard manual technique and the precision of computer-assisted planning and PSI-guided osteotomies in pelvic tumour resection.

Methods

CT scans were obtained from five female cadaveric pelvises. Five osteotomies were designed using Mimics software: sacroiliac, biplanar supra-acetabular, two parallel iliopubic and ischial. For cases of the left hemipelvis, PSIs were designed to guide standard oscillating saw osteotomies and later manufactured using 3D printing. Osteotomies were performed using the standard manual technique in cases of the right hemipelvis. Post-resection CT scans were quantitatively analysed. Student’s t-test and Mann–Whitney U test were used.


Bone & Joint Research
Vol. 6, Issue 8 | Pages 522 - 529
1 Aug 2017
Ali AM Newman SDS Hooper PA Davies CM Cobb JP

Objectives

Unicompartmental knee arthroplasty (UKA) is a demanding procedure, with tibial component subsidence or pain from high tibial strain being potential causes of revision. The optimal position in terms of load transfer has not been documented for lateral UKA. Our aim was to determine the effect of tibial component position on proximal tibial strain.

Methods

A total of 16 composite tibias were implanted with an Oxford Domed Lateral Partial Knee implant using cutting guides to define tibial slope and resection depth. Four implant positions were assessed: standard (5° posterior slope); 10° posterior slope; 5° reverse tibial slope; and 4 mm increased tibial resection. Using an electrodynamic axial-torsional materials testing machine (Instron 5565), a compressive load of 1.5 kN was applied at 60 N/s on a meniscal bearing via a matching femoral component. Tibial strain beneath the implant was measured using a calibrated Digital Image Correlation system.


Bone & Joint Research
Vol. 2, Issue 9 | Pages 186 - 192
1 Sep 2013
Boivin GP Platt KM Corbett J Reeves J Hardy AL Elenes EY Charnigo RJ Hunter SA Pearson KJ

Objectives

The goals of this study were: 1) to determine if high-fat diet (HFD) feeding in female mice would negatively impact biomechanical and histologic consequences on the Achilles tendon and quadriceps muscle; and 2) to investigate whether exercise and branched-chain amino acid (BCAA) supplementation would affect these parameters or attenuate any negative consequences resulting from HFD consumption.

Methods

We examined the effects of 16 weeks of 60% HFD feeding, voluntary exercise (free choice wheel running) and BCAA administration in female C57BL/6 mice. The Achilles tendons and quadriceps muscles were removed at the end of the experiment and assessed histologically and biomechanically.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 793 - 798
1 Jun 2017
Anderson FL Koch CN Elpers ME Wright TM Haas SB Heyse TJ

Aims

We sought to establish whether an oxidised zirconium (OxZr) femoral component causes less loss of polyethylene volume than a cobalt alloy (CoCr) femoral component in total knee arthroplasty.

Materials and Methods

A total of 20 retrieved tibial inserts that had articulated with OxZr components were matched with 20 inserts from CoCr articulations for patient age, body mass index, length of implantation, and revision diagnosis. Changes in dimensions of the articular surfaces were compared with those of pristine inserts using laser scanning. The differences in volume between the retrieved and pristine surfaces of the two groups were calculated and compared.


Bone & Joint 360
Vol. 6, Issue 3 | Pages 14 - 16
1 Jun 2017


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 1028 - 1036
1 Aug 2017
Chawla H Nwachukwu BU van der List JP Eggman AA Pearle AD Ghomrawi HM

Aims

Patellofemoral arthroplasty (PFA) has experienced significant improvements in implant survivorship with second generation designs. This has renewed interest in PFA as an alternative to total knee arthroplasty (TKA) for younger active patients with isolated patellofemoral osteoarthritis (PF OA). We analysed the cost-effectiveness of PFA versus TKA for the management of isolated PF OA in the United States-based population.

Patients and Methods

We used a Markov transition state model to compare cost-effectiveness between PFA and TKA. Simulated patients were aged 60 (base case) and 50 years. Lifetime costs (2015 United States dollars), quality-adjusted life year (QALY) gains and incremental cost-effectiveness ratio (ICER) were calculated from a healthcare payer perspective. Annual rates of revision were derived from the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man. Deterministic and probabilistic sensitivity analysis was performed for all parameters against a $50 000/QALY willingness to pay.


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 865 - 871
1 Jul 2017
Amstutz HC Le Duff MJ Bhaurla SK

Aims

A contact patch to rim (CPR) distance of < 10 mm has been associated with edge-loading and excessive wear. However, not all arthroplasties with a low CPR distance show problems with wear. Therefore, CPR distance may not be the only variable affecting the post-operative metal ion concentrations.

Patients and Methods

We used multiple logistic regression to determine what variables differed between the patients who had high and low cobalt (CoS) and chromium (CrS) serum ion concentrations within a cohort of patients with low (< 10 mm) CPR distances. A total of 56 patients treated with unilateral hip resurfacing arthroplasty (HRA) had CoS and CrS ion studies performed more than one year after surgery. The mean age of the patients at the time of surgery was 51.7 years (29 to 70), with 38 women (68%) and 18 men (32%).


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1499 - 1505
1 Nov 2012
Kim J Kasat NS Bae J Kim S Oh S Lim H

The purpose of this study was to measure the radiological parameters of femoral component alignment of the Oxford Phase 3 unicompartmental knee replacement (UKR), and evaluate their effect on clinical outcome. Multiple regression analysis was used to examine the relative contributions of the radiological assessment of femoral component alignment in 189 consecutive UKRs performed by a single surgeon. The American Knee Society scores were compared between groups, defined as being within or outside recommended tolerances of the position of the femoral component. For the flexion/extension position 21 UKRs (11.1%) lay outside the recommended limits, and for posterior overhang of the femoral component nine (4.8%) lay outside the range. The pre-operative hip/knee/ankle (HKA) angle, narrowest canal distance from the distal femoral entry point of the alignment jig and coronal entry-point position had significant effects on the flexion/extension position. Pre-operative HKA angle had a significant influence on posterior overhang of the femoral component. However, there was no significant difference in American Knee Society scores relative to the position of the femoral component.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 614 - 622
1 May 2017
Grammatopoulos G Bolduc M Atkins BL Kendrick BJL McLardy-Smith P Murray DW Gundle R Taylor AH

Aims

Advocates of debridement, antibiotics and implant retention (DAIR) in hip periprosthetic joint infection (PJI) argue that a procedure not disturbing a sound prosthesis-bone interface is likely to lead to better survival and functional outcome compared with revision. This case-control study aims were to compare outcome of DAIRs for infected primary total hip arthroplasty (THA) with outcomes following primary THA and two-stage revision of infected primary THAs.

Patients and Methods

We retrospectively reviewed all DAIRs, performed for confirmed infected primary hip arthropasty (n = 82) at out institution, between 1997 and 2013. Data recorded included full patient information and type of surgery. Outcome measures included complications, mortality, implant survivorship and functional outcome. Outcome was compared with two control groups matched for gender and age; a cohort of primary THAs (n = 120) and a cohort of two-stage revisions for infection (n = 66).


The Bone & Joint Journal
Vol. 99-B, Issue 3 | Pages 310 - 316
1 Mar 2017
Hothi H Henckel J Shearing P Holme T Cerquiglini A Laura AD Atrey A Skinner J Hart A

Aims

The aim of this study was to compare the design of the generic OptiStem XTR femoral stem with the established Exeter femoral stem.

Materials and Methods

We obtained five boxed, as manufactured, implants of both designs at random (ten in total). Two examiners were blinded to the implant design and independently measured the mass, volume, trunnion surface topography, trunnion roughness, trunnion cone angle, Caput-Collum-Diaphyseal (CCD) angle, femoral offset, stem length, neck length, and the width and roughness of the polished stem shaft using peer-reviewed methods. We then compared the stems using these parameters.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 494 - 502
1 Apr 2017
Simpson AHRW Keenan G Nayagam S Atkins RM Marsh D Clement ND

Aims

The aim of this double-blind prospective randomised controlled trial was to assess whether low intensity pulsed ultrasound (LIPUS) accelerated or enhanced the rate of bone healing in adult patients undergoing distraction osteogenesis.

Patients and Methods

A total of 62 adult patients undergoing limb lengthening or bone transport by distraction osteogenesis were randomised to treatment with either an active (n = 32) or a placebo (n = 30) ultrasound device. A standardised corticotomy was performed in the proximal tibial metaphysis and a circular Ilizarov frame was used in all patients. The rate of distraction was also standardised. The primary outcome measure was the time to removal of the frame after adjusting for the length of distraction in days/cm for both the per protocol (PP) and the intention-to-treat (ITT) groups. The assessor was blinded to the form of treatment. A secondary outcome was to identify covariates affecting the time to removal of the frame.