Burgo FJ Mengelle DE Ozols A Fernandez C Autorino CM

Objectives

Studies reporting specifically on squeaking in total hip arthroplasty have focused on cementless, and not on hybrid, fixation. We hypothesised that the cement mantle of the femur might have a damping effect on the sound transmitted through the metal stem. The objective of this study was to test the effect of cement on sound propagation along different stem designs and under different fixation conditions.

Methods

An in vitro model for sound detection, composed of a mechanical suspension structure and a sound-registering electronic assembly, was designed. A pulse of sound in the audible range was propagated along bare stems and stems implanted in cadaveric bone femurs with and without cement. Two stems of different alloy and geometry were compared.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1529 - 1533
1 Nov 2012
Taniguchi A Takakura Y Sugimoto K Hayashi K Ouchi K Kumai T Tanaka Y

The purpose of this study was to evaluate the clinical results of a newly designed prosthesis to replace the body of the talus in patients with aseptic necrosis. Between 1999 and 2006, 22 tali in 22 patients were replaced with a ceramic prosthesis. A total of eight patients were treated with the first-generation prosthesis, incorporating a peg to fix into the retained neck and head of the talus, and the remaining 14 were treated with the second-generation prosthesis, which does not have the peg. The clinical results were assessed by the American Orthopaedic Foot and Ankle Society ankle/hindfoot scale.

The mean follow-up was 98 months (18 to 174). The clinical results of the first-generation prostheses were excellent in three patients, good in one, fair in three and poor in one. There were, however, radiological signs of loosening, prompting a change in design. The clinical results of the second-generation prostheses were excellent in three patients, good in five, fair in four and poor in two, with more favourable radiological appearances. Revision was required using a total talar implant in four patients, two in each group.

Although the second-generation prosthesis produced better results, we cannot recommend the use of a talar body prosthesis. We now recommend the use of a total talar implant in these patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1655 - 1659
1 Dec 2007
Anwar HA Aldam CH Visuvanathan S Hart AJ

The biological significance of cobalt-chromium wear particles from metal-on-metal hip replacements may be different to the effects of the constituent metal ions in solution. Bacteria may be able to discriminate between particulate and ionic forms of these metals because of a transmembrane nickel/cobalt-permease. It is not known whether wear particles are bacteriocidal.

We compared the doubling time of coagulase negative staphylococcus, Staphylococcus aureus and methicillin resistant S. aureus when cultured in either wear particles from a metal-on-metal hip simulator, wear particles from a metal-on-polyethylene hip simulator, metal ions in solution or a control.

Doubling time halved in metal-on-metal (p = 0.003) and metal-on-polyethylene (p = 0.131) particulate debris compared with the control.

Bacterial nickel/cobalt-transporters allow metal ions but not wear particles to cross bacterial membranes. This may be useful for testing the biological characteristics of different wear debris. This experiment also shows that metal-on-metal hip wear debris is not bacteriocidal.


Bone & Joint 360
Vol. 5, Issue 5 | Pages 2 - 7
1 Oct 2016
Forward DP Ollivere BJ Ng JWG Coughlin TA Rollins KE

Rib fracture fixation by orthopaedic and cardiothoracic surgeons has become increasingly popular for the treatment of chest injuries in trauma. The literature, though mainly limited to Level II and III evidence, shows favourable results for operative fixation. In this paper we review the literature and discuss the indications for rib fracture fixation, surgical approaches, choice of implants and the future direction for management. With the advent of NICE guidance and new British Orthopaedic Association Standards for Trauma (BOAST) guidelines in production, the management of rib fractures is going to become more and more commonplace.


Bone & Joint Research
Vol. 6, Issue 2 | Pages 113 - 122
1 Feb 2017
Scholes SC Hunt BJ Richardson VM Langton DJ Smith E Joyce TJ

Objectives

The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of metal-on-metal (MoM) hip joints. Some designs of MoM hip joint do, however, have reasonable mid-term performance when implanted in appropriate patients. Investigations into the reasons for implant failure are important to offer help with the choice of implants and direction for future implant designs. One way to assess the performance of explanted hip prostheses is to measure the wear (in terms of material loss) on the joint surfaces.

Methods

In this study, a coordinate measuring machine (CMM) was used to measure the wear on five failed cementless Biomet Magnum/ReCap/ Taperloc large head MoM THAs, along with one Biomet ReCap resurfacing joint. Surface roughness measurements were also taken. The reason for revision of these implants was pain and/or adverse reaction to metal debris (ARMD) and/or elevated blood metal ion levels.


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 579 - 584
1 May 2016
Osman K Panagiotidou AP Khan M Blunn G Haddad FS

There is increasing global awareness of adverse reactions to metal debris and elevated serum metal ion concentrations following the use of second generation metal-on-metal total hip arthroplasties. The high incidence of these complications can be largely attributed to corrosion at the head-neck interface. Severe corrosion of the taper is identified most commonly in association with larger diameter femoral heads. However, there is emerging evidence of varying levels of corrosion observed in retrieved components with smaller diameter femoral heads. This same mechanism of galvanic and mechanically-assisted crevice corrosion has been observed in metal-on-polyethylene and ceramic components, suggesting an inherent biomechanical problem with current designs of the head-neck interface.

We provide a review of the fundamental questions and answers clinicians and researchers must understand regarding corrosion of the taper, and its relevance to current orthopaedic practice.

Cite this article: Bone Joint J 2016;98-B:579–84.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1427 - 1430
1 Oct 2011
Lindgren JU Brismar BH Wikstrom AC

A 70-year-old man with an uncemented metal-on-polyethylene total hip prosthesis underwent revision arthroplasty 33 months later because of pain, swelling and recurrent dislocation. There appeared to be corrosion and metal release from the prosthetic head, resulting in pseudotumour formation and severe local soft-tissue destruction. The corrosion occurred at the junction between the titanium-molybdenum-zirconium-iron taper and the cobalt-chrome-molybdenum head, but the mechanism was unproven.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 315 - 320
1 Mar 2006
Field RE Singh PJ Latif AMH Cronin MD Matthews DJ

We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different registrars. There were three withdrawals from the study because of psychiatric illness, a deep infection and a recurrent dislocation. Five deaths occurred prior to five-year follow-up and one patient withdrew from clinical review.

In the remaining 51 hips the mean pre-operative Oxford hip score was 47 points which decreased to 19 points at five years (45 hips). Of the stems 49 (98%) were implanted within 1° of neutral in the femoral canal. The mean migration of the stem at five years was 1.9 mm and the survivorship for aseptic loosening was 100%. There was no significant difference in outcome between the consultant and registrar groups. At five years, the results were comparable with those of other polished, tapered, cemented stems. Long-term surveillance continues.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1253 - 1259
1 Sep 2012
Sewell MD Higgs DS Al-Hadithy N Falworth M Bayley I Lambert SM

Scapulothoracic fusion (STF) for painful winging of the scapula in neuromuscular disorders can provide effective pain relief and functional improvement, but there is little information comparing outcomes between patients with dystrophic and non-dystrophic conditions. We performed a retrospective review of 42 STFs in 34 patients with dystrophic and non-dystrophic conditions using a multifilament trans-scapular, subcostal cable technique supported by a dorsal one-third semi-tubular plate. There were 16 males and 18 females with a mean age of 30 years (15 to 75) and a mean follow-up of 5.0 years (2.0 to 10.6). The mean Oxford shoulder score improved from 20 (4 to 39) to 31 (4 to 48). Patients with non-dystrophic conditions had lower overall functional scores but achieved greater improvements following STF. The mean active forward elevation increased from 59° (20° to 90°) to 97° (30° to 150°), and abduction from 51° (10° to 90°) to 83° (30° to 130°) with a greater range of movement achieved in the dystrophic group. Revision fusion for nonunion was undertaken in five patients at a mean time of 17 months (7 to 31) and two required revision for fracture. There were three pneumothoraces, two rib fractures, three pleural effusions and six nonunions. The main risk factors for nonunion were smoking, age and previous shoulder girdle surgery.

STF is a salvage procedure that can provide good patient satisfaction in 82% of patients with both dystrophic and non-dystrophic pathologies, but there was a relatively high failure rate (26%) when poor outcomes were analysed. Overall function was better in patients with dystrophic conditions which correlated with better range of movement; however, patients with non-dystrophic conditions achieved greater functional improvement.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 10 | Pages 1439 - 1444
1 Oct 2005
Davies AP Sood A Lewis AC Newson R Learmonth ID Case CP

Previous research has shown an increase in chromosomal aberrations in patients with worn implants. The type of aberration depended on the type of metal alloy in the prosthesis. We have investigated the metal-specific difference in the level of DNA damage (DNA stand breaks and alkali labile sites) induced by culturing human fibroblasts in synovial fluid retrieved at revision arthroplasty.

All six samples from revision cobalt-chromium metal-on-metal and four of six samples from cobalt-chromium metal-on-polyethylene prostheses caused DNA damage. By contrast, none of six samples from revision stainless-steel metal-on-polyethylene prostheses caused significant damage. Samples of cobalt-chromium alloy left to corrode in phosphate-buffered saline also caused DNA damage and this depended on a synergistic effect between the cobalt and chromium ions.

Our results further emphasise that epidemiological studies of orthopaedic implants should take account of the type of metal alloy used.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1475 - 1481
1 Nov 2012
Berton C Puskas GJ Christofilopoulos P Stern R Hoffmeyer P Lübbeke A

There are no recent studies comparing cable with wire for the fixation of osteotomies or fractures in total hip replacement (THR). Our objective was to evaluate the five-year clinical and radiological outcomes and complication rates of the two techniques. We undertook a review including all primary and revision THRs performed in one hospital between 1996 and 2005 using cable or wire fixation. Clinical and radiological evaluation was performed five years post-operatively. Cables were used in 51 THRs and wires in 126, and of these, 36 THRs with cable (71%) and 101 with wire (80%) were evaluated at follow-up. The five-year radiographs available for 33 cable and 91 wire THRs revealed rates of breakage of fixation of 12 of 33 (36%) and 42 of 91 (46%), respectively. With cable there was a significantly higher risk of metal debris (68% vs 9%; adjusted relative risk (RR) 6.6; 95% confidence interval (CI) 3.0 to 14.1), nonunion (36% vs 21%; adjusted RR 2.0; 95% CI 1.0 to 3.9) and osteolysis around the material, acetabulum or femur (61% vs 19%; adjusted RR 3.9; 95% CI 2.3 to 6.5). Cable breakage increased the risk of osteolysis to 83%. There was a trend towards foreign-body reaction and increased infection with cables. Clinical results did not differ between the groups.

In conclusion, we found a higher incidence of complications and a trend towards increased infection and foreign-body reaction with the use of cables.


Bone & Joint Research
Vol. 1, Issue 11 | Pages 289 - 296
1 Nov 2012
Savaridas T Wallace RJ Muir AY Salter DM Simpson AHRW

Objectives

Small animal models of fracture repair primarily investigate indirect fracture healing via external callus formation. We present the first described rat model of direct fracture healing.

Methods

A rat tibial osteotomy was created and fixed with compression plating similar to that used in patients. The procedure was evaluated in 15 cadaver rats and then in vivo in ten Sprague-Dawley rats. Controls had osteotomies stabilised with a uniaxial external fixator that used the same surgical approach and relied on the same number and diameter of screw holes in bone.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 28 - 32
1 Jan 2016
Hanna SA Somerville L McCalden RW Naudie DD MacDonald SJ

Aims

The purpose of this study was to compare the long-term results of primary total hip arthroplasty (THA) in young patients using either a conventional (CPE) or a highly cross-linked (HXLPE) polyethylene liner in terms of functional outcome, incidence of osteolysis, radiological wear and rate of revision.

Methods

We included all patients between the ages of 45 and 65 years who, between January 2000 and December 2001, had undergone a primary THA for osteoarthritis at our hospital using a CPE or HXLPE acetabular liner and a 28 mm cobalt-chrome femoral head.

From a total of 160 patients, 158 (177 hips) were available for review (CPE 89; XLPE 88). The mean age, body mass index (BMI) and follow-up in each group were: CPE: 56.8 years (46 to 65); 30.7 kg/m2 (19 to 58); 13.2 years (2.1 to 14.7) and HXLPE: 55.6 years (45 to 65); BMI: 30 kg/m2 (18 to 51); 13.1 years (5.7 to 14.4).


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1194 - 1200
1 Sep 2011
Akiyama H Yamamoto K Tsukanaka M Kawanabe K Otsuka H So K Goto K Nakamura T

We retrospectively reviewed 40 hips in 36 patients who had undergone acetabular reconstruction using a titanium Kerboull-type acetabular reinforcement device with bone allografts between May 2001 and April 2006. Impacted bone allografts were used for the management of American Academy of Orthopaedic Surgeons Type II defects in 17 hips, and bulk bone allografts together with impacted allografts were used for the management of Type III defects in 23 hips. A total of five hips showed radiological failure at a mean follow-up of 6.7 years (4.5 to 9.3), two of which were infected. The mean pre-operative Merle d’Aubigné score was 10 (5 to 15) vs 13.6 (9 to 18) at the latest follow-up. The Kaplan-Meier survival rate at ten years, calculated using radiological failure or revision of the acetabular component for any reason as the endpoint, was 87% (95% confidence interval 76.3 to 97.7). A separate experimental analysis of the mechanical properties of the device and the load-displacement properties of bone grafts showed that a structurally hard allograft resected from femoral heads of patients with osteoarthritis should be preferentially used in any type of defect. If impacted bone allografts were used, a bone graft thickness of < 25 mm was acceptable in Type II defects.

This clinical study indicates that revision total hip replacement using the Kerboull-type acetabular reinforcement device with bone allografts yielded satisfactory mid-term results.


The Bone & Joint Journal
Vol. 98-B, Issue 3 | Pages 301 - 306
1 Mar 2016
Reidy MJ Faulkner A Shitole B Clift B

Aims

The long-term functional outcome of total hip arthroplasty (THA) performed by trainees is not known. A multicentre retrospective study of 879 THAs was undertaken to investigate any differences in outcome between those performed by trainee surgeons and consultants.

Patients and Methods

A total of 879 patients with a mean age of 69.5 years (37 to 94) were included in the study; 584 THAs (66.4%) were undertaken by consultants, 138 (15.7%) by junior trainees and 148 (16.8%) by senior trainees. Patients were scored using the Harris Hip Score (HHS) pre-operatively and at one, three, five, seven and ten years post-operatively. Surgical outcome, complications and survival were compared between groups. The effect of supervision was determined by comparing supervised and unsupervised trainees. A primary univariate analysis was used to select variables for inclusion in multivariate analysis.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 241 - 248
1 Feb 2012
Firoozabadi R McDonald E Nguyen T Buckley JM Kandemir U

Filling the empty holes in peri-articular locking plates may improve the fatigue strength of the fixation. The purpose of this in vitro study was to investigate the effect of plugging the unused holes on the fatigue life of peri-articular distal femoral plates used to fix a comminuted supracondylar fracture model.

A locking/compression plate was applied to 33 synthetic femurs and then a 6 cm metaphyseal defect was created (AO Type 33-A3). The specimens were then divided into three groups: unplugged, plugged with locking screw only and fully plugged holes. They were then tested using a stepwise or run-out fatigue protocol, each applying cyclic physiological multiaxial loads.

All specimens in the stepwise group failed at the 770 N load level. The mean number of cycles to failure for the stepwise specimen was 25 500 cycles (sd 1500), 28 800 cycles (sd 6300), and 26 400 cycles (sd 2300) cycles for the unplugged, screw only and fully plugged configurations, respectively (p = 0.16). The mean number of cycles to failure for the run-out specimens was 42 800 cycles (sd 10 700), 36 000 cycles (sd 7200), and 36 600 cycles (sd 10 000) for the unplugged, screw only and fully plugged configurations, respectively (p = 0.50). There were also no differences in axial or torsional stiffness between the constructs. The failures were through the screw holes at the level of comminution.

In conclusion, filling the empty combination locking/compression holes in peri-articular distal femur locking plates at the level of supracondylar comminution does not increase the fatigue life of the fixation in a comminuted supracondylar femoral fracture model (AO 33-A3) with a 6 cm gap.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1466 - 1470
1 Oct 2010
Didden K Luyckx T Bellemans J Labey L Innocenti B Vandenneucker H

The biomechanics of the patellofemoral joint can become disturbed during total knee replacement by alterations induced by the position and shape of the different prosthetic components. The role of the patella and femoral trochlea has been well studied. We have examined the effect of anterior or posterior positioning of the tibial component on the mechanisms of patellofemoral contact in total knee replacement. The hypothesis was that placing the tibial component more posteriorly would reduce patellofemoral contact stress while providing a more efficient lever arm during extension of the knee.

We studied five different positions of the tibial component using a six degrees of freedom dynamic knee simulator system based on the Oxford rig, while simulating an active knee squat under physiological loading conditions. The patellofemoral contact force decreased at a mean of 2.2% for every millimetre of posterior translation of the tibial component. Anterior positions of the tibial component were associated with elevation of the patellofemoral joint pressure, which was particularly marked in flexion > 90°.

From our results we believe that more posterior positioning of the tibial component in total knee replacement would be beneficial to the patellofemoral joint.


Bone & Joint Research
Vol. 5, Issue 2 | Pages 26 - 32
1 Feb 2016
Wendling A Mar D Wischmeier N Anderson D McIff T

Objectives

The objective of this study was to determine if combining variations in mixing technique of antibiotic-impregnated polymethylmethacrylate (PMMA) cement with low frequency ultrasound (LFUS) improves antibiotic elution during the initial high phase (Phase I) and subsequent low phase (Phase II) while not diminishing mechanical strength.

Methods

Three batches of vancomycin-loaded PMMA were prepared with different mixing techniques: a standard technique; a delayed technique; and a control without antibiotic. Daily elution samples were analysed using flow injection analysis (FIA). Beginning in Phase II, samples from each mix group were selected randomly to undergo either five, 15, 45, or 0 minutes of LFUS treatment. Elution amounts between LFUS treatments were analysed. Following Phase II, compression testing was done to quantify strength. A-priorit-tests and univariate ANOVAs were used to compare elution and mechanical test results between the two mix groups and the control group.


The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1519 - 1524
1 Nov 2015
Salih S Paskins A Nichol T Smith T Hamer A

We investigated whether the indentation of bone cement spacers used in revision of infected joint arthroplasty with a MacDonald dissector increased the elution of antibiotic in vitro. A total of 24 cement discs containing either 0.17 g (0.88% w/w), 0.25 g (1.41% w/w), or 0.33 g (1.75% w/w) gentamicin of constant size were made. Of these, 12 were indented with the dissector. Each disc was immersed in ammonium acetate buffer in a sealed container, and fluid from each container was sampled at zero, one, three, six, 24, 48 and 72 hours and at one, and two weeks. The concentration of gentamicin in the fluid was analysed using high performance liquid chromatography mass spectrometry.

The fluid sampled at 72 hours from the indented discs containing 0.17 g gentamicin (0.88% w/w) contained a mean of 113 mcg/ml (90.12 to 143.5) compared with 44.5 mcg/ml (44.02 to 44.90) in the fluid sampled from the plain discs (p = 0.012). In discs containing 0.33 g gentamicin (1.75% w/w), the concentration eluted from the indented discs at 72 hours was a mean of 316 mcg/ml (223 to 421) compared with a mean of 118 mcg/ml (100 to 140) from the plain discs (p < 0.001).

At two weeks, these significant differences persisted. At nine weeks the indented discs eluted a greater concentration for all gentamicin doses, but the difference was only significant for the discs containing 0.17 g (0.88% w/w, p = 0.006). However if the area under the curve is taken as a measure of the total antibiotic eluted, the indented discs eluted more gentamicin than the plain discs for the 0.17 g (0.88% w/w, p = 0.031), the 0.25 g (1.41% w/w, p < 0.001) and the 0.33 g (1.75% w/w, p < 0.001) discs.

When preparing antibiotic spacers for use in staged revision arthroplasty surgery we recommend indenting the spacer with a MacDonald dissector to increase the elution of antibiotic.

Cite this article: Bone Joint J 2015;97-B:1519–24.