We assessed the reliability, accuracy and variability of closed-wedge high tibial osteotomy (HTO) using computer-assisted surgery compared to the conventional technique. A total of 50 closed-wedge HTO procedures were performed using the navigation system, and compared with 50 HTOs that had been performed with the conventional technique. In the navigation group, the mean mechanical axis prior to osteotomy was varus 8.2°, and the mean mechanical axis following fixation was valgus 3.6°. On the radiographs the mean pre-operative mechanical axis was varus 7.3°, and the mean post-operative mechanical axis was valgus 2.1°. There was a positive correlation between the measured data taken under navigation and by radiographs (r >
0.3, p <
0.05). The mean correction angle was significantly more accurate in the navigation group (p <
0.002). The variability of the correction was significantly lower in the navigation group (2.3°
The April 2012 Knee Roundup360 looks at the torn ACL, ACL reconstruction, the risk of ACL rupture, the benefit of warm-ups before exercise, glucosamine and tibiofemoral osteoarthritis, sensitisation and sporting tendinopathy, pain relief after TKR, the long-term results of the Genesis I, the gender specific recovery times after TKR, and the accuracy of the orthopaedic eyeball
We compared lower limb coronal alignment measurements
obtained pre- and post-operatively with long-leg radiographs and
computer navigation in patients undergoing primary total knee replacement
(TKR). A series of 185 patients had their pre- and post-implant
radiological and computer-navigation system measurements of coronal alignment
compared using the Bland-Altman method. The study included 81 men
and 104 women with a mean age of 68.5 years (32 to 87) and a mean
body mass index of 31.7 kg/m2 (19 to 49). Pre-implant
Bland–Altman limits of agreement were -9.4° to 8.6° with a repeatability
coefficient of 9.0°. The Bland–Altman plot showed a tendency for the
radiological measurement to indicate a higher level of pre-operative
deformity than the corresponding navigation measurement. Post-implant
limits of agreement were -5.0° to 5.4° with a repeatability coefficient
of 5.2°. The tendency for valgus knees to have greater deformity
on the radiograph was still seen, but was weaker for varus knees. The alignment seen or measured intra-operatively during TKR is
not necessarily the same as the deformity seen on a standing long-leg
radiograph either pre- or post-operatively. Further investigation
into the effect of weight-bearing and surgical exposure of the joint
on the mechanical femorotibial angle is required to enable the most appropriate
intra-operative alignment to be selected.
This single-blinded randomised controlled trial
investigated whether one design of mobile-bearing (MB) total knee replacement
(TKR) has any advantage over a fixed-bearing (FB) design on long-term
fixation as measured by radiostereometry. The amount of wear underneath
the mobile bearing was also evaluated. A series of 42 knees was randomised
to MB or FB tibial components with appropriate polyethylene inserts
and followed for between ten and 12 years, or until the death of
the patient. The polyethylene in the MB group was superior in that
it was gamma-irradiated in inert gas and was calcium-stearate free;
the polyethylene in the FB group was gamma-irradiated in air and
contained calcium stearate. In theory this should be advantageous
to the wear rate of the MB group. At final follow-up the overall
mean migration was 0.75 mm ( For the MB group, the mean linear wear rate on the under-surface
was 0.026 mm/year (
Evaluation of patients with painful total knee replacement requires a thorough clinical examination and relevant investigations in order to reach a diagnosis. Awareness of the common and uncommon problems leading to painful total knee replacement is useful in the diagnostic approach. This review article aims to act as a guide to the evaluation of patients with painful total knee replacement.
Peri-prosthetic patellar fracture following resurfacing
as part of total knee replacement (TKR) is an infrequent yet challenging
complication. This case-control study was performed to identify
clinical, radiological and surgical factors that increase the risk
of developing a spontaneous patellar fracture after TKR. Patellar
fractures were identified in 74 patients (88 knees) from a series
of 7866 consecutive TKRs conducted between 1998 and 2009. After excluding
those with a previous history of extensor mechanism realignment
or a clear traumatic event, a metal-backed patella, any uncemented
component or subsequent infection, the remaining 64 fractures were
compared with a matched group of TKRs with an excellent outcome
defined by the Knee Society score. The mean age of patients with
a fracture was 70 years (51 to 81) at the time of TKR. Patellar
fractures were detected at a mean of 13.4 months (2 to 84) after
surgery. The incidence of patellar fracture was found to be strongly
associated with the number of previous knee operations, greater
pre-operative mechanical malalignment, smaller post-operative patellar
tendon length, thinner post-resection patellar thickness, and a
lower post-operative Insall-Salvati ratio. An understanding of the risk factors associated with spontaneous
patellar fracture following TKR provides a valuable insight into
prevention of this challenging complication.
Robots have been used in surgery since the late
1980s. Orthopaedic surgery began to incorporate robotic technology
in 1992, with the introduction of ROBODOC, for the planning and
performance of total hip replacement. The use of robotic systems
has subsequently increased, with promising short-term radiological
outcomes when compared with traditional orthopaedic procedures.
Robotic systems can be classified into two categories: autonomous
and haptic (or surgeon-guided). Passive surgery systems, which represent
a third type of technology, have also been adopted recently by orthopaedic
surgeons. While autonomous systems have fallen out of favour, tactile systems
with technological improvements have become widely used. Specifically,
the use of tactile and passive robotic systems in unicompartmental
knee replacement (UKR) has addressed some of the historical mechanisms
of failure of non-robotic UKR. These systems assist with increasing
the accuracy of the alignment of the components and produce more
consistent ligament balance. Short-term improvements in clinical
and radiological outcomes have increased the popularity of robot-assisted
UKR. Robot-assisted orthopaedic surgery has the potential for improving
surgical outcomes. We discuss the different types of robotic systems
available for use in orthopaedics and consider the indication, contraindications
and limitations of these technologies.
This study compared the outcome of total knee
replacement (TKR) in adult patients with fixed- and mobile-bearing prostheses
during the first post-operative year and at five years’ follow-up,
using gait parameters as a new objective measure. This double-blind
randomised controlled clinical trial included 55 patients with mobile-bearing (n
= 26) and fixed-bearing (n = 29) prostheses of the same design,
evaluated pre-operatively and post-operatively at six weeks, three
months, six months, one year and five years. Each participant undertook
two walking trials of 30 m and completed the EuroQol questionnaire,
Western Ontario and McMaster Universities osteoarthritis index,
Knee Society score, and visual analogue scales for pain and stiffness.
Gait analysis was performed using five miniature angular rate sensors
mounted on the trunk (sacrum), each thigh and calf. The study population
was divided into two groups according to age (≤ 70 years Improvements in most gait parameters at five years’ follow-up
were greater for fixed-bearing TKRs in older patients (>
70 years),
and greater for mobile-bearing TKRs in younger patients (≤ 70 years).
These findings should be confirmed by an extended age controlled
study, as the ideal choice of prosthesis might depend on the age
of the patient at the time of surgery.
Between April 2004 and July 2007, we performed 241 primary total knee replacements in 204 patients using the e.motion posterior cruciate-retaining, multidirectional mobile-bearing prosthesis. Of these, 100 were carried out using an image-free navigation system, and the remaining 141 with the conventional technique. We conducted a retrospective study from the prospectively collected data of these patients to assess the early results of this new mobile-bearing design. At a mean follow-up of 49 months (32 to 71), 18 knees (7.5%) had mechanical complications of which 13 required revision. Three of these had a peri-prosthetic fracture, and were removed from the study. The indication for revision in the remaining ten was loosening of the femoral component in two, tibiofemoral dislocation in three, disassociation of the polyethylene liner in four, and a broken polyethyene liner in one. There were eight further mechanically unstable knees which presented with recurrent disassociation of the polyethylene liner. There was no significant difference in the incidence of mechanical instability between the navigation-assisted procedures (8 of 99, 8.1%) and the conventionally implanted knees (10 of 139, 7.2%). In our view, the relatively high rate of mechanical complications and revision within 30 months precludes the further use of new design of knee replacement.
This conversation represents an attempt by several
arthroplasty surgeons to critique several abstracts presented over
the last year as well as to use them as a jumping off point for trying
to figure out where they fit in into our current understanding of
multiple issues in modern hip and knee arthroplasty.
Medial open-wedge high tibial osteotomy has been gaining popularity in recent years, but adequate supporting material is required in the osteotomy gap for early weight-bearing and rapid union. The purpose of this study was to investigate whether the implantation of a polycaprolactone-tricalcium phosphate composite scaffold wedge would enhance healing of the osteotomy in a micro pig model. We carried out open-wedge high tibial osteotomies in 12 micro pigs aged from 12 to 16 months. A scaffold wedge was inserted into six of the osteotomies while the other six were left open. Bone healing was evaluated after three and six months using plain radiographs, CT scans, measurement of the bone mineral density and histological examination. Complete bone union was obtained at six months in both groups. There was no collapse at the osteotomy site, loss of correction or failure of fixation in either group. Staining with haematoxylin and eosin demonstrated that there was infiltration of new bone tissue into the macropores and along the periphery of the implanted scaffold in the scaffold group. The CT scans and measurement of the bone mineral density showed that at six months specimens in the scaffold group had a higher bone mineral density than in the control group, although the implantation of the polycaprolactone-tricalcium phosphate composite scaffold wedge did not enhance healing of the osteotomy.
We conducted a retrospective study to investigate the effect of femoral bowing on the placement of components in total knee replacement (TKR), with regard to its effect on reestablishing the correct mechanical axis, as we hypothesised that computer-assisted total knee replacement (CAS-TKR) would produce more accurate alignment than conventional TKR. Between January 2006 and December 2009, 212 patients (306 knees) underwent TKR. The conventional TKR was compared with CAS-TKR for accuracy of placement of the components and post-operative alignment, as determined by five radiological measurements. There were significant differences in the reconstructed mechanical axes between the bowed and the non-bowed group after conventional TKR (176.2° ( For patients with significant femoral bowing, the reconstructed mechanical axes were significantly closer to normal in the CAS group than in the conventional group (179.2° (
The Morquio syndrome is a rare disorder which presents with a number of musculoskeletal problems. The literature describing total knee replacement in these patients is sparse. We describe the management of a patient with bilateral instability and pain in the knees using bilateral constrained knee replacements, and followed up for five years with pre- and postoperative knee scores. We highlight the difficulties encountered and discuss the end results.
Mechanical failure because of wear or fracture of the polyethylene tibial post in posteriorly-stabilised total knee replacements has been extensively described. In this study of 12 patients with a clinically and radiologically successful NexGen LPS posteriorly-stabilised prosthesis impingement of the anterior tibial post was evaluated in vivo in three dimensions during gait using radiologically-based image-matching techniques. Impingement was observed in all images of the patients during the stance phase, although the NexGen LPS was designed to accommodate 14° of hyperextension of the component before impingement occurred. Impingement arises as a result of posterior translation of the femur during the stance phase. Further attention must therefore be given to the configuration of the anterior portion of the femoral component and the polyethylene post when designing posteriorly-stabilised total knee replacements.
The records of patients aged 50 years or over who underwent primary reconstruction of the anterior cruciate ligament between 1990 and 2002 were reviewed. There were 35 knees in 34 patients that met the inclusion criteria. The mean age of the patients was 57 years (50 to 66) and the mean clinical follow-up was for 72 months (25 to 173). A total of 23 knees were reconstructed with patellar tendon allograft, and 12 with patellar tendon autograft. The mean pre-operative knee extension was 1° (−5° to 10°) and flexion was 129° (125° to 150°) and at follow-up these values were 0° (−5° to 5°) and 135° (120° to 150°), respectively. Pre-operatively there were 31 knees (89%) with a Lachman grade 2+ or 3+. Post-operatively, 33 knees (94%) were Lachman grade 0 or 1+. The mean pre- and post-operative International Knee Documentation Committee scores were 39 (23 to 72) and 90 (33 to 100) respectively. The mean pre- and post-operative Lysholm scores were 50 (18 to 68) and 92 (28 to 100) respectively and the mean University of California Los Angeles activity scores were 8.5 before injury (4 to 10), 4.3 (3 to 6) after injury and 8.3 (4 to 10) post-operatively. There were three graft failures (8.6%) requiring revision. We conclude that reconstruction of the anterior cruciate ligament in carefully-selected patients aged 50 years or over can achieve similar results to those in younger patients, with no increased risk of complications.
We wished to determine whether simultaneous bilateral sequential total knee replacement (TKR) carried increased rates of mortality and complications compared with unilateral TKR in low- and high-risk patients. Our study included 2385 patients who had undergone bilateral sequential TKR under one anaesthetic and 719 who had unilateral TKR. There were no significant pre-operative differences between the groups in terms of age, gender, height, weight, body mass index, diagnosis, comorbidity and duration of follow-up, which was a mean of 10.2 years (5 to 14) in the bilateral and 10.4 years (5 to 14) in the unilateral group. The peri-operative mortality rate (eight patients, 0.3%) of patients who had bilateral sequential TKR was similar to that (five patients, 0.7%) of those undergoing unilateral TKR. In bilateral cases the peri-operative mortality rate (three patients, 0.4%) of patients at high risk was similar to that (five patients, 0.3%) of patients at low risk as it was also in unilateral cases (two patients, 1.0% vs three patients, 0.6%). There was no significant difference (p = 0.735) in either the overall number of major complications between bilateral and unilateral cases or between low- (p = 0.57) and high-risk (p = 0.61) patients. Also, the overall number of minor complications was not significantly different between the bilateral and unilateral group (p = 0.143). Simultaneous bilateral sequential TKR can be offered to patients at low and high risk and has an expected rate of complications similar to that of unilateral TKR.
We carried out a prospective study to assess the clinical outcome, complications and survival of the NexGen Legacy posterior-stabilised-Flex total knee replacement (TKR) in a consecutive series of 278 knees between May 2003 and February 2005. Mean follow-up for 259 TKRs (98.2%) was 3.8 years (3.0 to 4.8). Annual follow-up showed improvement in the Knee Society scores (paired This relatively large study indicates that the legacy posterior stabilised-Flex design provides excellent short-term outcome but warrants ongoing evaluation to confirm the long-term durability and functioning of the implant.
Surgeons need to be able to measure angles and distances in three dimensions in the planning and assessment of knee replacement. Computed tomography (CT) offers the accuracy needed but involves greater radiation exposure to patients than traditional long-leg standing radiographs, which give very little information outside the plane of the image. There is considerable variation in CT radiation doses between research centres, scanning protocols and individual scanners, and ethics committees are rightly demanding more consistency in this area. By refining the CT scanning protocol we have reduced the effective radiation dose received by the patient down to the equivalent of one long-leg standing radiograph. Because of this, it will be more acceptable to obtain the three-dimensional data set produced by CT scanning. Surgeons will be able to document the impact of implant position on outcome with greater precision.
We present a retrospective study of patients suffering from a variety of benign tumours in whom external fixators were used to treat deformity and limb-length discrepancy, and for the reconstruction of bone defects. A total of 43 limbs in 31 patients (12 male and 19 female) with a mean age of 14 years (2 to 54) were treated. The diagnosis was Ollier’s disease in 12 limbs, fibrous dysplasia in 11, osteochondroma in eight, giant cell tumour in five, osteofibrous dysplasia in five and non-ossifying fibroma in two. The lesions were treated in the tibia in 19 limbs, in the femur in 16, and in the forearm in eight. The Ilizarov frame was used in 25 limbs, the Taylor Spatial Frame in seven, the Orthofix fixator in six, the Monotube in four and the Heidelberg fixator in one. The mean follow-up was 72 months (22 to 221). The mean external fixation period was 168 days (71 to 352). The mean external fixation index was 42 days/cm (22.2 to 102.0) in the 22 patients who required limb lengthening. The mean correction angle for those with angular deformity was 23° (7° to 45°). At final follow-up all patients had returned to normal activities. Four patients required a second operation for recurrent deformity of further limb lengthening. Local recurrence occurred in one patient, requiring further surgery.
We performed a prospective, randomised study to compare the results and rates of complications of primary total knee replacement performed using a quadriceps-sparing technique or a standard arthrotomy in 120 patients who had bilateral total knee replacements carried out under the same anaesthetic. The clinical results, pain scales, surgical and hospital data, post-operative complications and radiological results were compared. No significant differences were found between the two groups with respect to the blood loss, knee score, function score, pain scale, range of movement or radiological findings. In contrast, the operating time (p = 0.0001) and the tourniquet time (p <
0.0001) were significantly longer in the quadriceps-sparing group, as was the rate of complications (p = 0.0468). We therefore recommend the use of a standard arthrotomy with the shortest possible skin incision for total knee replacement.