Advertisement for orthosearch.org.uk
Results 121 - 140 of 152
Results per page:
Bone & Joint Research
Vol. 2, Issue 5 | Pages 84 - 95
1 May 2013
Sidaginamale RP Joyce TJ Lord JK Jefferson R Blain PG Nargol AVF Langton DJ

Objectives

The aims of this piece of work were to: 1) record the background concentrations of blood chromium (Cr) and cobalt (Co) concentrations in a large group of subjects; 2) to compare blood/serum Cr and Co concentrations with retrieved metal-on-metal (MoM) hip resurfacings; 3) to examine the distribution of Co and Cr in the serum and whole blood of patients with MoM hip arthroplasties; and 4) to further understand the partitioning of metal ions between the serum and whole blood fractions.

Methods

A total of 3042 blood samples donated to the local transfusion centre were analysed to record Co and Cr concentrations. Also, 91 hip resurfacing devices from patients who had given pre-revision blood/serum samples for metal ion analysis underwent volumetric wear assessment using a coordinate measuring machine. Linear regression analysis was carried out and receiver operating characteristic curves were constructed to assess the reliability of metal ions to identify abnormally wearing implants. The relationship between serum and whole blood concentrations of Cr and Co in 1048 patients was analysed using Bland-Altman charts. This relationship was further investigated in an in vitro study during which human blood was spiked with trivalent and hexavalent Cr, the serum then separated and the fractions analysed.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1308 - 1313
1 Oct 2011
Hart AJ Sabah SA Bandi AS Maggiore P Tarassoli P Sampson B A. Skinner J

Blood metal ions have been widely used to investigate metal-on-metal hip replacements, but their ability to discriminate between well-functioning and failed hips is not known. The Medicines and Healthcare products Regulatory Agency (MHRA) has suggested a cut-off level of 7 parts per billion (ppb).

We performed a pair-matched, case-control study to investigate the sensitivity and specificity of blood metal ion levels for diagnosing failure in 176 patients with a unilateral metal-on-metal hip replacement. We recruited 88 cases with a pre-revision, unexplained failed hip and an equal number of matching controls with a well-functioning hip. We investigated the 7 ppb cut-off level for the maximum of cobalt or chromium and determined optimal mathematical cut-off levels from receiver-operating characteristic curves.

The 7 ppb cut-off level for the maximum of cobalt or chromium had a specificity of 89% and sensitivity 52% for detecting a pre-operative unexplained failed metal on metal hip replacement. The optimal cut-off level for the maximum of cobalt or chromium was 4.97 ppb and had sensitivity 63% and specificity 86%.

Blood metal ions had good discriminant ability to separate failed from well-functioning hip replacements. The MHRA cut-off level of 7 ppb provides a specific test but has poor sensitivity.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 40 - 48
1 Jan 2016
Matharu GS Mansour R Dada O Ostlere S Pandit HG Murray DW

Aims

The aims of this study were to compare the diagnostic test characteristics of ultrasound alone, metal artefact reduction sequence MRI (MARS-MRI) alone, and ultrasound combined with MARS-MRI for identifying intra-operative pseudotumours in metal-on-metal hip resurfacing (MoMHR) patients undergoing revision surgery.

Methods

This retrospective diagnostic accuracy study involved 39 patients (40 MoMHRs). The time between imaging modalities was a mean of 14.6 days (0 to 90), with imaging performed at a mean of 5.3 months (0.06 to 12) before revision. The prevalence of intra-operative pseudotumours was 82.5% (n = 33).


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 713 - 719
1 Jun 2011
Duckworth AD Ring D McQueen MM

A suspected fracture of the scaphoid remains difficult to manage despite advances in knowledge and imaging methods. Immobilisation and restriction of activities in a young and active patient must be balanced against the risks of nonunion associated with an undiagnosed and undertreated fracture of the scaphoid.

The assessment of diagnostic tests for a suspected fracture of the scaphoid must take into account two important factors. First, the prevalence of true fractures among suspected fractures is low, which greatly reduces the probability that a positive test will correspond with a true fracture, as false positives are nearly as common as true positives. This situation is accounted for by Bayesian statistics. Secondly, there is no agreed reference standard for a true fracture, which necessitates the need for an alternative method of calculating diagnostic performance characteristics, based upon a statistical method which identifies clinical factors tending to associate (latent classes) in patients with a high probability of fracture.

The most successful diagnostic test to date is MRI, but in low-prevalence situations the positive predictive value of MRI is only 88%, and new data have documented the potential for false positive scans. The best strategy for improving the diagnosis of true fractures among suspected fractures of the scaphoid may well be to develop a clinical prediction rule incorporating a set of demographic and clinical factors which together increase the pre-test probability of a fracture of the scaphoid, in addition to developing increasingly sophisticated radiological tests.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 173 - 176
1 Feb 2015
Omar M Ettinger M Reichling M Petri M Guenther D Gehrke T Krettek C Mommsen P

The aim of this study was to assess the role of synovial C-reactive protein (CRP) in the diagnosis of chronic periprosthetic hip infection. We prospectively collected synovial fluid from 89 patients undergoing revision hip arthroplasty and measured synovial CRP, serum CRP, erythrocyte sedimentation rate (ESR), synovial white blood cell (WBC) count and synovial percentages of polymorphonuclear neutrophils (PMN). Patients were classified as septic or aseptic by means of clinical, microbiological, serum and synovial fluid findings. The high viscosity of the synovial fluid precluded the analyses in nine patients permitting the results in 80 patients to be studied. There was a significant difference in synovial CRP levels between the septic (n = 21) and the aseptic (n = 59) cohort. According to the receiver operating characteristic curve, a synovial CRP threshold of 2.5 mg/l had a sensitivity of 95.5% and specificity of 93.3%. The area under the curve was 0.96. Compared with serum CRP and ESR, synovial CRP showed a high diagnostic value. According to these preliminary results, synovial CRP may be a useful parameter in diagnosing chronic periprosthetic hip infection.

Cite this article: Bone Joint J 2015; 97-B:173–6.


The Bone & Joint Journal
Vol. 97-B, Issue 10_Supple_A | Pages 20 - 29
1 Oct 2015
Gehrke T Alijanipour P Parvizi J

Periprosthetic joint infection (PJI) is one of the most feared and challenging complications following total knee arthroplasty. We provide a detailed description of our current understanding regarding the management of PJI of the knee, including diagnostic aids, pre-operative planning, surgical treatment, and outcome.

Cite this article: Bone Joint J 2015;97-B(10 Suppl A):20–9.


The Bone & Joint Journal
Vol. 96-B, Issue 10 | Pages 1366 - 1369
1 Oct 2014
Held M Laubscher M Zar HJ Dunn RN

The lack of an accurate, rapid diagnostic test for mycobacterium tuberculosis (TB) is a major handicap in the management of spinal TB. GeneXpert, a new, rapid molecular diagnostic test is recommended as the first line investigation for suspected pulmonary TB in areas with a high prevalence of HIV or drug resistance, yet it has not been validated for the diagnosis of musculoskeletal TB.

The aim of this study was to assess the accuracy of GeneXpert in diagnosing spinal TB.

A prospective clinical study of 69 consecutive adults with suspected spinal TB was conducted at a tertiary hospital in an area with the highest incidence and prevalence of TB in the world. GeneXpert was used on tissue samples of the enrolled patients and its diagnostic accuracy compared with a reference standard of tissue in liquid culture. A total of 71 spine samples from 69 patients (two re-biopsies) were included in the study.

The GeneXpert test showed a sensitivity of 95.6% and specificity of 96.2% for spinal TB. The results of the GeneXpert test were available within 48 hours compared with a median of 35 days (IQR 15 to 43) for cultures. All cases of multi-drug resistant TB (MDR TB) were diagnosed accurately with the GeneXpert test. The MDR TB rate was 5.8%.

Cite this article: Bone Joint J 2014;96-B:1366–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 961 - 968
1 Jul 2012
Duckworth AD Buijze GA Moran M Gray A Court-Brown CM Ring D McQueen MM

A prospective study was performed to develop a clinical prediction rule that incorporated demographic and clinical factors predictive of a fracture of the scaphoid. Of 260 consecutive patients with a clinically suspected or radiologically confirmed scaphoid fracture, 223 returned for evaluation two weeks after injury and formed the basis of our analysis. Patients were evaluated within 72 hours of injury and at approximately two and six weeks after injury using clinical assessment and standard radiographs. Demographic data and the results of seven specific tests in the clinical examination were recorded.

There were 116 (52%) men and their mean age was 33 years (13 to 95; sd 17.9). In 62 patients (28%) a scaphoid fracture was confirmed. A logistic regression model identified male gender (p = 0.002), sports injury (p = 0.004), anatomical snuff box pain on ulnar deviation of the wrist within 72 hours of injury (p < 0.001), and scaphoid tubercle tenderness at two weeks (p < 0.001) as independent predictors of fracture. All patients with no pain at the anatomical snuff box on ulnar deviation of the wrist within 72 hours of injury did not have a fracture (n = 72, 32%). With four independently significant factors positive, the risk of fracture was 91%.

Our study has demonstrated that clinical prediction rules have a considerable influence on the probability of a suspected scaphoid fracture. This will help improve the use of supplementary investigations where the diagnosis remains in doubt.


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 984 - 988
1 Jul 2014
Pullinger M Southorn T Easton V Hutchinson R Smith RP Sanghrajka AP

Congenital Talipes Equinovarus (CTEV) is one of the most common congenital limb deformities. We reviewed the records of infants who had received treatment for structural CTEV between 1 January 2007 and 30 November 2012. This was cross-referenced with the prenatal scans of mothers over a corresponding period of time. We investigated the sensitivity, specificity, and positive and negative predictive values of the fetal anomaly scan for the detection of CTEV and explored whether the publication of Fetal Anomaly Screening Programme guidelines in 2010 affected the rate of detection.

During the study period there were 95 532 prenatal scans and 34 373 live births at our hospital. A total of 37 fetuses with findings suggestive of CTEV were included in the study, of whom 30 were found to have structural CTEV at birth. The sensitivity of screening for CTEV was 71.4% and the positive predictive value was 81.1%. The negative predictive value and specificity were more than 99.5%. There was no significant difference between the rates of detection before and after publication of the guidelines (p = 0.5).

We conclude that a prenatal fetal anomaly ultrasound screening diagnosis of CTEV has a good positive predictive value enabling prenatal counselling. The change in screening guidance has not affected the proportion of missed cases. This information will aid counselling parents about the effectiveness and accuracy of prenatal ultrasound in diagnosing CTEV.

Cite this article: Bone Joint J 2014;96-B:984–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 584 - 595
1 May 2012
Dartnell J Ramachandran M Katchburian M

A delay in the diagnosis of paediatric acute and subacute haematogenous osteomyelitis can lead to potentially devastating morbidity. There are no definitive guidelines for diagnosis, and recommendations in the literature are generally based on expert opinions, case series and cohort studies.

All articles in the English literature on paediatric osteomyelitis were searched using MEDLINE, CINAHL, EMBASE, Google Scholar, the Cochrane Library and reference lists. A total of 1854 papers were identified, 132 of which were examined in detail. All aspects of osteomyelitis were investigated in order to formulate recommendations.

On admission 40% of children are afebrile. The tibia and femur are the most commonly affected long bones. Clinical examination, blood and radiological tests are only reliable for diagnosis in combination. Staphylococcus aureus is the most common organism detected, but isolation of Kingella kingae is increasing. Antibiotic treatment is usually sufficient to eradicate the infection, with a short course intravenously and early conversion to oral treatment. Surgery is indicated only in specific situations.

Most studies were retrospective and there is a need for large, multicentre, randomised, controlled trials to define protocols for diagnosis and treatment. Meanwhile, evidence-based algorithms are suggested for accurate and early diagnosis and effective treatment.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 781 - 786
1 Jun 2010
Zhao G Yamamoto T Ikemura S Motomura G Mawatari T Nakashima Y Iwamoto Y

We investigated the factors related to the radiological outcome of a transtrochanteric curved varus osteotomy in patients with osteonecrosis of the hip. We reviewed 73 hips in 62 patients with a mean follow-up of 12.4 years (5 to 31.1). There were 28 men and 34 women, with a mean age of 33.3 years (15 to 68) at the time of surgery. The 73 hips were divided into two groups according to their radiological findings: group 1 showed progression of collapse and/or joint-space narrowing; group 2 had neither progressive collapse nor joint-space narrowing. Both of these factors and the radiological outcomes were analysed by a stepwise discriminant analysis. A total of 12 hips were categorised as group 1 and 61 as group 2. Both the post-operative intact ratio and the localisation of the necrotic lesion correlated with the radiological outcome. The cut-off point of the postoperative intact ratio to prevent the progression of collapse was 33.6%, and the cut-off point to prevent both the progression of collapse and joint-space narrowing was 41.9%.

The results of this study indicate that a post-operative intact ratio of 33.0% is necessary if a satisfactory outcome is to be achieved after this varus osteotomy.


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 554 - 557
1 Apr 2013
Dinneen A Guyot A Clements J Bradley N

Cut-off values with highest sensitivity and specificity for the synovial fluid white cell and differential count will facilitate the accurate diagnosis of infection in total knee (TKR) and total hip replacement (THR). All patients undergoing revision TKR or THR for suspected prosthetic joint infection between 2009 and 2011 at two hospitals were identified. A total of 75 patients were included with a mean age of 70.3 years (38 to 89). Synovial fluid was aspirated pre-operatively and peri-prosthetic tissue samples were taken intra-operatively for histological and microbiological examination.

Receiver operating characteristic (ROC) plots were constructed for white cell and differential counts in aspirated fluid. The optimal cut-off for TKR and THR was 1590 white cells/µl and 65% neutrophilia. The white cell count cut-off value identified for THR was notably lower than previously quoted in the literature. A cut-off value for white cell count in synovial aspirate in suspected prosthetic joint infection of between 1100 and 1700 white cells/µl is likely to be applicable to both THR and TKR.

Cite this article: Bone Joint J 2013;95-B:554–7.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 874 - 878
1 Jul 2008
Fink B Makowiak C Fuerst M Berger I Schäfer P Frommelt L

We analysed the serum C-reactive protein level, synovial fluid obtained by joint aspiration and five synovial biopsies from 145 knee replacements prior to revision to assess the value of these parameters in diagnosing late peri-prosthetic infection. Five further synovial biopsies were used for histological analysis. Samples were also obtained during the revision and incubated and analysed in an identical manner for 14 days.

A total of 40 total knee replacements were found to be infected (prevalence 27.6%). The aspiration technique had a sensitivity of 72.5% (95% confidence interval (CI) 58.7 to 86.3), a specificity of 95.2% (95% CI 91.2 to 99.2), a positive predictive value of 85.3% (95% CI 73.4 to 100), a negative predictive value of 90.1% (95% CI 84.5 to 95.7) and an accuracy of 89%. The biopsy technique had a sensitivity of 100%, a specificity of 98.1% (95% CI 95.5 to 100), a positive predictive value of 95.2% (95% CI 88.8 to 100), a negative predictive value of 100% and an accuracy of 98.6%. C-reactive protein with a cut-off-point of 13.5 mg/l had a sensitivity of 72.5% (95% CI 58.7 to 86.3), a specificity of 80.9% (95% CI 73.4 to 88.4), a positive predictive value of 59.2% (95% CI 45.4 to 73.0), a negative predictive value of 88.5% (95% 81.0 to 96.0 CI) and an accuracy of 78.1%.

We found that biopsy was superior to joint aspiration and C-reactive protein in the diagnosis of late peri-prosthetic infection of total knee replacements.


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 563 - 567
1 Apr 2013
İltar S Alemdaroğlu KB Say F Aydoğan NH

Redisplacement is the most common complication of immobilisation in a cast for the treatment of diaphyseal fractures of the forearm in children. We have previously shown that the three-point index (TPI) can accurately predict redisplacement of fractures of the distal radius. In this prospective study we applied this index to assessment of diaphyseal fractures of the forearm in children and compared it with other cast-related indices that might predict redisplacement. A total of 76 children were included. Their ages, initial displacement, quality of reduction, site and level of the fractures and quality of the casting according to the TPI, Canterbury index and padding index were analysed. Logistic regression analysis was used to investigate risk factors for redisplacement. A total of 18 fractures (24%) redisplaced in the cast. A TPI value of > 0.8 was the only significant risk factor for redisplacement (odds ratio 238.5 (95% confidence interval 7.063 to 8054.86); p < 0.001).

The TPI was far superior to other radiological indices, with a sensitivity of 84% and a specificity of 97% in successfully predicting redisplacement. We recommend it for routine use in the management of these fractures in children.

Cite this article: Bone Joint J 2013;95-B:563–7.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 741 - 744
1 May 2005
Beaulé PE Krismer M Mayrhofer P Wanner S Le Duff M Mattesich M Stoeckl B Amstutz HC Biedermann R

Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse - femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers.

The accuracy of the method was found to be ± 1.6 mm for the x-direction and ± 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 versus 32.2 months (p = 0.001) and 24.8 months (p = 0.012), respectively.

EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 571 - 576
1 Apr 2005
Savarino L Granchi D Cenni E Baldini N Greco M Giunti A

There is no diagnostic, non-invasive method for the early detection of loosening after total hip arthroplasty. In a pilot study, we have analysed two serum markers of bone remodelling, procollagen I C-terminal extension peptide (PICP) and cross-linked N-terminal telopeptide (NTx), as well as the diagnostic performance of NTx for the assessment of osteolysis. We recruited 21 patients with loosening (group I), 18 with a well-fixed prosthesis (group II) and 17 at the time of primary arthroplasty for osteoarthritis (OA) (group III). Internal normal reference ranges were obtained from 30 healthy subjects (group IV).

The serum PICP level was found to be significantly lower in patients with OA and those with loosening, when compared with those with stable implants, while the NTx level was significantly increased only in the group with loosening, suggesting that collagen degradation depended on the altered bone turnover induced by the implant. This hypothesis was reinforced by the finding that the values in the pre-surgery patients and stable subjects were comparable with the reference range of younger healthy subjects.

A high specificity and positive predictive value for NTx provided good diagnostic evidence of agreement between the test and the clinical and radiological evaluations. The NTx level could be used to indicate stability of the implant. However, further prospective, larger studies are necessary.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 465 - 470
1 Apr 2005
Reinartz P Mumme T Hermanns B Cremerius U Wirtz DC Schaefer WM Niethard F Buell U

Two major complications of hip replacement are loosening and infection. Reliable differentiation between these pathological processes is difficult since both may be accompanied by similar symptoms. Our aim was to assess the diagnostic ability of triple-phase bone scanning (TPBS) and positron-emission tomography (PET) to detect and differentiate these complications in patients with a hip arthroplasty. Both TPBS and PET were performed in 63 patients (92 prostheses). The radiotracer for PET imaging was 18F-fluorodeoxyglucose (FDG). Image interpretation was performed according to qualitative and quantitative criteria although the final diagnosis was based upon either surgical findings or clinical follow-up.

The sensitivity, specificity and accuracy of PET was 0.94, 0.95 and 0.95 respectively, compared with 0.68, 0.76 and 0.74 for TPBS. We found that an image interpretation based exclusively upon quantitative criteria was inappropriate because of its low selectivity. The histological examination indicated that increased periprosthetic uptake of FDG in patients with aseptic loosening was caused by wear-induced polyethylene particles and the subsequent growth of aggressive granulomatous tissue.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1351 - 1360
1 Oct 2006
Rajasekaran S Babu JN Dheenadhayalan J Shetty AP Sundararajan SR Kumar M Rajasabapathy S

Limb-injury severity scores are designed to assess orthopaedic and vascular injuries. In Gustilo type-IIIA and type-IIIB injuries they have poor sensitivity and specificity to predict salvage or outcome.

We have designed a trauma score to grade the severity of injury to the covering tissues, the bones and the functional tissues, grading the three components from one to five. Seven comorbid conditions known to influence the management and prognosis have been given a score of two each. The score was validated in 109 consecutive open injuries of the tibia, 42 type-IIIA and 67 type-IIIB. The total score was used to assess the possibilities of salvage and the outcome was measured by dividing the injuries into four groups according to their scores as follows: group I scored less than 5, group II 6 to 10, group III 11 to 15 and group IV 16 or more.

A score of 14 to indicate amputation had the highest sensitivity and specificity. Our trauma score compared favourably with the Mangled Extremity Severity score in sensitivity (98% and 99%), specificity (100% and 17%), positive predictive value (100% and 97.5%) and negative predictive value (70% and 50%), respectively. A receiver-operating characteristic curve constructed for 67 type-IIIB injuries to assess the efficiency of the scores to predict salvage, showed that the area under the curve for this score was better (0.988 (± 0.013 sem)) than the Mangled Extremity Severity score (0.938 (± 0.039 sem)). All limbs in group IV and one in group III underwent amputation. Of the salvaged limbs, there was a significant difference in the three groups for the requirement of a flap for wound cover, the time to union, the number of surgical procedures required, the total days as an in-patient and the incidence of deep infection (p < 0.001 for all). The individual scores for covering and functional tissues were also found to offer specific guidelines in the management of these complex injuries.

The scoring system was found to be simple in application and reliable in prognosis for both limb-salvage and outcome measures in type-IIIA and type-IIIB open injuries of the tibia.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 11 | Pages 1433 - 1439
1 Nov 2011
Dias JJ Singh HP

A displaced fracture of the scaphoid is one in which the fragments have moved from their anatomical position or there is movement between them when stressed by physiological loads. Displacement is seen in about 20% of fractures of the waist of the scaphoid, as shown by translation, a gap, angulation or rotation. A CT scan in the true longitudinal axis of the scaphoid demonstrates the shape of the bone and displacement of the fracture more accurately than do plain radiographs. Displaced fractures can be treated in a plaster cast, accepting the risk of malunion and nonunion. Surgically the displacement can be reduced, checked radiologically, arthroscopically or visually, and stabilised with headless screws or wires. However, rates of union and deformity are unknown. Mild malunion is well tolerated, but the long-term outcome of a displaced fracture that healed in malalignment has not been established.

This paper summarises aspects of the assessment, treatment and outcome of displaced fractures of the waist of the scaphoid.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1126 - 1134
1 Aug 2012
Granchi D Cenni E Giunti A Baldini N

We report a systematic review and meta-analysis of the peer-reviewed literature focusing on metal sensitivity testing in patients undergoing total joint replacement (TJR). Our purpose was to assess the risk of developing metal hypersensitivity post-operatively and its relationship with outcome and to investigate the advantages of performing hypersensitivity testing.

We undertook a comprehensive search of the citations quoted in PubMed and EMBASE: 22 articles (comprising 3634 patients) met the inclusion criteria. The frequency of positive tests increased after TJR, especially in patients with implant failure or a metal-on-metal coupling. The probability of developing a metal allergy was higher post-operatively (odds ratio (OR) 1.52 (95% confidence interval (CI) 1.06 to 2.31)), and the risk was further increased when failed implants were compared with stable TJRs (OR 2.76 (95% CI 1.14 to 6.70)).

Hypersensitivity testing was not able to discriminate between stable and failed TJRs, as its predictive value was not statistically proven. However, it is generally thought that hypersensitivity testing should be performed in patients with a history of metal allergy and in failed TJRs, especially with metal-on-metal implants and when the cause of the loosening is doubtful.