Advertisement for orthosearch.org.uk
Results 121 - 140 of 178
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 12 | Pages 1704 - 1709
1 Dec 2015
Drexler M Gortzak Y Sternheim A Kollender Y Amar E Bickels J

Excision of the proximal femur for tumour with prosthetic reconstruction using a bipolar femoral head places a considerable load on the unreplaced acetabulum.

We retrospectively reviewed the changes which occur around the affected hip joint by evaluating the post-operative radiographs of 65 consecutive patients who underwent proximal prosthetic arthroplasty of the femur, and in whom an acetabular component had not been used. There were 37 men and 28 women with a mean age of 57.3 years (17 to 93). Radiological assessment included the extent of degenerative change in the acetabulum, heterotopic ossification, and protrusio acetabuli.

The mean follow-up was 9.1 years (2 to 11.8). Degenerative changes in the acetabulum were seen in three patients (4.6%), Brooker grade 1 or 2 heterotopic ossification in 17 (26%) and protrusion of the prosthetic head in nine (13.8%).

A total of eight patients (12.3%) needed a revision. Five were revised to the same type of prosthesis and three (4.6%) were converted to a total hip arthroplasty.

We conclude that radiological evidence of degenerative change, heterotopic ossification and protrusion occur in a few patients who undergo prosthetic arthroplasty of the proximal femur for tumour. The limited extent of these changes and the lack of associated symptoms do not justify the routine arthroplasty of the acetabulum in these patients.

Cite this article: Bone Joint J 2015;97-B:1704–9


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 58 - 64
1 Jan 2016
Ahmed I Salmon LJ Waller A Watanabe H Roe JP Pinczewski LA

Aims

Oxidised zirconium was introduced as a material for femoral components in total knee arthroplasty (TKA) as an attempt to reduce polyethylene wear. However, the long-term survival of this component is not known.

Methods

We performed a retrospective review of a prospectively collected database to assess the ten year survival and clinical and radiological outcomes of an oxidised zirconium total knee arthroplasty with the Genesis II prosthesis.

The Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), Knee Injury and Osteoarthritis Outcome Score (KOOS) and a patient satisfaction scale were used to assess outcome.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1275 - 1279
1 Sep 2013
Liu T Liu Z Zhang Q Zhang X

The aim of this study was to assess a specific protocol for the treatment of patients with a parosteal osteosarcoma of the distal femur with limb salvage involving hemicortical resection and reconstruction using recycled pasteurised autograft and internal fixation. Between January 2000 and January 2010, 13 patients with a mean age of 26.5 years (17 to 39) underwent this procedure. All the tumours were staged according to Enneking’s criteria: there were eight stage IA tumours and five stage IB tumours. The mean follow-up was 101.6 months (58 to 142), and mean post-operative Musculoskeletal Tumour Society functional score was 88.6% (80% to 100%) at the final follow-up. All the patients had achieved bony union; the mean time to union was 11.2 months (6 to 18). Local recurrence occurred in one patient 27 months post-operatively. No patient had a pulmonary metastasis.

A hemicortical procedure for the treatment of a parosteal osteosarcoma is safe and effective. Precise pre-operative planning using MRI is essential in order to define the margins of resection. Although it is a technically demanding procedure, gratifying results make it worthwhile for selected patients.

Cite this article: Bone Joint J 2013;95-B:1275–9.


The Bone & Joint Journal
Vol. 97-B, Issue 12 | Pages 1608 - 1614
1 Dec 2015
Tice A Kim P Dinh† L Ryu JJ Beaulé PE

The primary purpose of this study of metal-on-metal (MoM) hip resurfacing was to compare the effect of using a cementless or cemented femoral component on the subsequent bone mineral density (BMD) of the femoral neck.

This was a single-centre, prospective, double-blinded control trial which randomised 120 patients (105 men and 15 women) with a mean age of 49.4 years (21 to 68) to receive either a cemented or cementless femoral component. Follow-up was to two years. Outcome measures included total and six-point region-of-interest BMD of the femoral neck, radiological measurements of acetabular inclination, neck-shaft and stem-shaft angles, and functional outcome scores including the Harris hip score, the Western Ontario and McMaster Universities Osteoarthritis Index and the University of California at Los Angeles activity scale.

In total, 17 patients were lost to follow-up leaving 103 patients at two years. There were no revisions in the cementless group and three revisions (5%) in the cemented group (two because of hip pain and one for pseudotumour).

The total BMD was significantly higher in the cementless group at six months (p < 0.001) and one year (p = 0.01) than in the cemented group, although there was a loss of statistical significance in the difference at two years (p = 0.155).

All patient outcomes improved significantly: there were no significant differences between the two groups.

The results show better preservation of femoral neck BMD with a cementless femoral component after two years of follow-up. Further investigation is needed to establish whether this translates into improved survivorship.

Cite this article: Bone Joint J 2015;97-B:1608–14.


The Bone & Joint Journal
Vol. 97-B, Issue 10_Supple_A | Pages 3 - 8
1 Oct 2015
Murray DW Liddle AD Dodd CAF Pandit H

There is a large amount of evidence available about the relative merits of unicompartmental and total knee arthroplasty (UKA and TKA). Based on the same evidence, different people draw different conclusions and as a result, there is great variability in the usage of UKA.

The revision rate of UKA is much higher than TKA and so some surgeons conclude that UKA should not be performed. Other surgeons believe that the main reason for the high revision rate is that UKA is easy to revise and, therefore, the threshold for revision is low. They also believe that UKA has many advantages over TKA such as a faster recovery, lower morbidity and mortality and better function. They therefore conclude that UKA should be undertaken whenever appropriate.

The solution to this argument is to minimise the revision rate of UKA, thereby addressing the main disadvantage of UKA. The evidence suggests that this will be achieved if surgeons use UKA for at least 20% of their knee arthroplasties and use implants that are appropriate for these broad indications.

Cite this article: Bone Joint J 2015;97-B(10 Suppl A):3–8.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1500 - 1507
1 Nov 2013
Zaidi R Cro S Gurusamy K Sivanadarajah N Macgregor A Henricson A Goldberg A

We performed a systematic review and meta-analysis of modern total ankle replacements (TARs) to determine the survivorship, outcome, complications, radiological findings and range of movement, in patients with end-stage osteoarthritis (OA) of the ankle who undergo this procedure. We used the methodology of the Cochrane Collaboration, which uses risk of bias profiling to assess the quality of papers in favour of a domain-based approach. Continuous outcome scores were pooled across studies using the generic inverse variance method and the random-effects model was used to incorporate clinical and methodological heterogeneity. We included 58 papers (7942 TARs) with an interobserver reliability (Kappa) for selection, performance, attrition, detection and reporting bias of between 0.83 and 0.98. The overall survivorship was 89% at ten years with an annual failure rate of 1.2% (95% confidence interval (CI) 0.7 to 1.6). The mean American Orthopaedic Foot and Ankle Society score changed from 40 (95% CI 36 to 43) pre-operatively to 80 (95% CI 76 to 84) at a mean follow-up of 8.2 years (7 to 10) (p < 0.01). Radiolucencies were identified in up to 23% of TARs after a mean of 4.4 years (2.3 to 9.6). The mean total range of movement improved from 23° (95% CI 19 to 26) to 34° (95% CI 26 to 41) (p = 0.01).

Our study demonstrates that TAR has a positive impact on patients’ lives, with benefits lasting ten years, as judged by improvement in pain and function, as well as improved gait and increased range of movement. However, the quality of evidence is weak and fraught with biases and high quality randomised controlled trials are required to compare TAR with other forms of treatment such as fusion.

Cite this article: Bone Joint J 2013;95-B:1500–7.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 40 - 48
1 Jan 2016
Matharu GS Mansour R Dada O Ostlere S Pandit HG Murray DW

Aims

The aims of this study were to compare the diagnostic test characteristics of ultrasound alone, metal artefact reduction sequence MRI (MARS-MRI) alone, and ultrasound combined with MARS-MRI for identifying intra-operative pseudotumours in metal-on-metal hip resurfacing (MoMHR) patients undergoing revision surgery.

Methods

This retrospective diagnostic accuracy study involved 39 patients (40 MoMHRs). The time between imaging modalities was a mean of 14.6 days (0 to 90), with imaging performed at a mean of 5.3 months (0.06 to 12) before revision. The prevalence of intra-operative pseudotumours was 82.5% (n = 33).


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 786 - 792
1 Jun 2015
Hutt JRB Farhadnia P Massé V Lavigne M Vendittoli P

This randomised trial evaluated the outcome of a single design of unicompartmental arthroplasty of the knee (UKA) with either a cemented all-polyethylene or a metal-backed modular tibial component. A total of 63 knees in 45 patients (17 male, 28 female) were included, 27 in the all-polyethylene group and 36 in the metal-backed group. The mean age was 57.9 years (39.6 to 76.9). At a mean follow-up of 6.4 years (5 to 9.9), 11 all-polyethylene components (41%) were revised (at a mean of 5.8 years; 1.4 to 8.0) post-operatively and two metal-backed components were revised (at one and five years). One revision in both groups was for unexplained pain, one in the metal-backed group was for progression of osteoarthritis. The others in the all-polyethylene group were for aseptic loosening. The survivorship at seven years calculated by the Kaplan–Meier method for the all-polyethylene group was 56.5% (95% CI 31.9 to 75.2, number at risk 7) and for the metal-backed group was 93.8% (95% CI 77.3 to 98.4, number at risk 16) This difference was statistically significant (p <  0.001). At the most recent follow-up, significantly better mean Western Ontario and McMaster Universities Arthritis Index Scores were found in the all-polyethylene group (13.4 vs 23.0, p = 0.03) but there was no difference in the mean Knee injury and Osteoarthritis Outcome scores (68.8; 41.4 to 99.0 vs 62.6; 24.0 to 100.0), p = 0.36). There were no significant differences for range of movement (p = 0.36) or satisfaction (p = 0.23).

This randomised study demonstrates that all-polyethylene components in this design of fixed bearing UKA had unsatisfactory results with significantly higher rates of failure before ten years compared with the metal-back components.

Cite this article: Bone Joint J 2015;97-B:786–92.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 160 - 163
1 Feb 2015
Biemond JE Venkatesan S van Hellemondt GG

The long-term survival of the cementless Spotorno CLS femoral component in patients aged > 50 years at the time of arthroplasty was investigated. Survivorship analysis of a consecutive series of 85 patients (100 hips; under 50 years of age at a mean follow-up of 18.4 years (16.3 to 20.8)) was performed. The clinical and radiographic outcomes were satisfactory. The overall rate of survival of the femoral component was 93.5% (95% confidence interval (CI), 90.9 to 96.1) after 19 years. Survival with revision for aseptic loosening as the end point was 95.7% (95% CI 93.6 to 97.8%) at 19 years.

This study demonstrates an excellent long-term survival of the Spotorno CLS femoral component after 16 to 20 years in young patients undergoing total hip arthroplasty.

Cite this article: Bone Joint J 2015;97-B:160–3.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1175 - 1182
1 Sep 2015
Bisseling P de Wit BWK Hol AM van Gorp MJ van Kampen A van Susante JLC

Patients from a randomised trial on resurfacing hip arthroplasty (RHA) (n = 36, 19 males; median age 57 years, 24 to 65) comparing a conventional 28 mm metal-on-metal total hip arthroplasty (MoM THA) (n = 28, 17 males; median age 59 years, 37 to 65) and a matched control group of asymptomatic patients with a 32 mm ceramic-on-polyethylene (CoP) THA (n = 33, 18 males; median age 63 years, 38 to 71) were cross-sectionally screened with metal artefact reducing sequence-MRI (MARS-MRI) for pseudotumour formation at a median of 55 months (23 to 72) post-operatively. MRIs were scored by consensus according to three different classification systems for pseudotumour formation.

Clinical scores were available for all patients and metal ion levels for MoM bearing patients.

Periprosthetic lesions with a median volume of 16 mL (1.5 to 35.9) were diagnosed in six patients in the RHA group (17%), one in the MoM THA group (4%) and six in the CoP group (18%). The classification systems revealed no clear differences between the groups. Solid lesions (n = 3) were exclusively encountered in the RHA group. Two patients in the RHA group and one in the MoM THA group underwent a revision for pseudotumour formation. There was no statistically significant relationship between clinical scoring, metal ion levels and periprosthetic lesions in any of the groups.

Periprosthetic fluid collections are seen on MARS-MRI after conventional CoP THA and RHA and may reflect a soft-tissue collection or effusion.

Currently available MRI classification systems seem to score these collections as pseudotumours, causing an-overestimatation of the incidence of pseudotumours.

Cite this article: Bone Joint J 2015;97-B:1175–82.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1604 - 1610
1 Dec 2012
Angadi DS Brown S Crawfurd EJP

The aim of this prospective randomised study was to compare the clinical and radiological results of a cemented all-polyethylene Ultima acetabular component with those of a cementless porous-coated acetabular component (PFC) following total hip replacement (THR). A total of 287 patients received either a polyethylene acetabular component (group A) or a cobalt–chromium porous-coated component (group B) with an identical cemented femoral component and 28 mm cobalt-chromium head, thus making it the largest study of its type. Patients were evaluated radiologically and clinically using the Harris hip score (HHS). Group A comprised 183 patients (73 male, 110 female) with a mean age of 71.3 years (55 to 89). Group B comprised 104 patients (48 male, 56 female) with a mean age of 69.8 years (56 to 89). A total of 16 patients (13 in Group A, three in Group B) did not have post-operative data for analysis. The mean follow-up in group A was 7.52 years (0.4 to 15.0) and in Group B 7.87 years (0.5 to 14.0).

At final follow-up the mean HHS was similar between groups A and B (74.5 (25 to 100) and 78.0 (37 to 100), respectively; p = 0.068). The total number of revisions for any cause was 28, 17 of which were in group A and 11 in group B. The ten-year survivorship was 86.8% (95% confidence interval (CI) 78.4 to 92.1) and 89.2% (95% CI 78.3 to 94.8) for groups A and B, respectively (log-rank p-value = 0.938). A total of 20 cemented and two cementless acetabular components had evidence of acetabular radiolucencies or acetabular component migration at last follow-up (p = 0.001).

These results indicate that patients with a cemented all-polyethylene and cementless porous-coated polyethylene lined acetabular component have similar long-term clinical outcomes.


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 939 - 944
1 Jul 2015
McArthur BA Abdel MP Taunton MJ Osmon DR Hanssen AD

The aim of our study was to describe the characteristics, treatment, and outcomes of patients with periprosthetic joint infection (PJI) and normal inflammatory markers after total knee arthroplasty (TKA) and total hip arthroplasty (THA).

In total 538 TKAs and 414 THAs underwent surgical treatment for PJI and met the inclusion criteria. Pre-operative erythrocyte sedimentation rate (ESR) and C-reactive protein level (CRP) were reviewed to identify the seronegative cohort. An age- and gender-matched cohort was identified from the remaining patients for comparison. Overall, 4% of confirmed infections were seronegative (21 TKA and 17 THA). Of those who underwent pre-operative aspiration, cultures were positive in 76% of TKAs (n = 13) and 64% of THAs (n = 7). Cell count and differential were suggestive of infection in 85% of TKA (n = 11) and all THA aspirates (n = 5). The most common organism was coagulase-negative Staphylococcus. Seronegative infections were associated with a lower aspirate cell count and a lower incidence of Staphylococcus aureus infection. Two-stage revision was performed in 35 cases (95%). At a mean of five years (14 to 162 months) following revision, re-operation for infection occurred in two TKAs, and one THA. From our study we estimate around 4% of patients with PJI may present with normal ESR and CRP. When performed, pre-operative aspirate is useful in delivering a definitive diagnosis. When treated, similar outcomes can be obtained compared with patients with positive serology.

Cite this article: Bone Joint J 2015;97-B:939–44.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 591 - 594
1 May 2007
Lie SA Hallan G Furnes O Havelin LI Engesæter LB

We analysed the results of different strategies in the revision of primary uncemented acetabular components reported to the Norwegian Arthroplasty Register. The aim was to compare the risk of further acetabular revision after isolated liner exchange and complete component revision. The results of exchanging well-fixed components were also compared with those of exchanging loose acetabular components. The period studied was between September 1987 and April 2005. The following groups were compared: group 1, exchange of liner only in 318 hips; group 2, exchange of well-fixed components in 398; and group 3, exchange of loose components in 933. We found that the risk of a further cup revision was lower after revision of well-fixed components (relative risk from a Cox model (RR) = 0.56, 95% confidence interval 0.37% to 0.87%) and loose components (RR = 0.56, 95% confidence interval 0.39% to 0.80%), compared with exchange of the liner in isolation. The most frequent reason for a further acetabular revision was dislocation, accounting for 61 (28%) of the re-revisions. Other reasons for further revision included pain in 27 (12%), loosening in 24 (11%) and infection in 20 (9%). Re-revisions because of pain were less frequent when complete component (fixed or loose) revision was undertaken compared with isolated exchange of the liner (RR = 0.20 (95% confidence interval 0.06% to 0.65%) and RR = 0.10 (95% confidence interval 0.03% to 0.30%), respectively). The risk of further acetabular revision for infection, however, did not differ between the groups.

In this study, exchange of the liner only had a higher risk of further cup revision than revision of the complete acetabular component. Our results suggest that the threshold for revising well-fixed components in the case of liner wear and osteolysis should be lowered.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1015 - 1018
1 Aug 2007
Wroblewski BM Siney PD Fleming PA

We studied survival to 38 years after Charnley low-friction arthroplasty of the hip. We used revision as an end-point, while adopting a policy of regular follow-up and early revision for radiological changes alone if indicated.

Between November 1962 and June 2005, 22 066 primary low-friction arthroplasties (17 409 patients) had been performed at Wrightington Hospital by more than 330 surgeons. By June 2006, 1001 (4.5%) hips had been revised and 1490 patients (2662 hips, 12%) had died. At 31 years, where a minimum of 40 hips were still attending follow-up, survival with revision for infection as an endpoint was 95%, for dislocation 98%, for a fractured stem 88.6%, for a loose stem 72.5% and for a loose acetabular component 53.7%. Wear and loosening of the ultra-high-molecular-weight polyethylene acetabular component were the main long-term problems.

We conclude that regular follow-up after hip replacement is essential and that all operative findings should be recorded at revision.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1657 - 1662
1 Dec 2014
Stambough JB Clohisy JC Barrack RL Nunley RM Keeney JA

The aims of this retrospective study were to compare the mid-term outcomes following revision total knee replacement (TKR) in 76 patients (81 knees) < 55 years of age with those of a matched group of primary TKRs based on age, BMI, gender and comorbid conditions. We report the activity levels, functional scores, rates of revision and complications. Compared with patients undergoing primary TKR, those undergoing revision TKR had less improvement in the mean Knee Society function scores (8.14 (–55 to +60) vs 23.3 points (–40 to +80), p < 0.001), a similar improvement in UCLA activity level (p = 0.52), and similar minor complication rates (16% vs 13%, p = 0.83) at a mean follow-up of 4.6 years (2 to 13.4). Further revision surgery was more common among revised TKRs (17% vs 5%, p = 0.02), with deep infection and instability being the most common reasons for failure. As many as one-third of patients aged < 55 years in the revision group had a complication or failure requiring further surgery.

Young patients undergoing revision TKR should be counselled that they can expect somewhat less improvement and a higher risk of complications than occur after primary TKR.

Cite this article: Bone Joint J 2014; 96-B:1657–62


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1158 - 1163
1 Sep 2008
De Haan R Campbell PA Su EP De Smet KA

We have reviewed 42 patients who had revision of metal-on-metal resurfacing procedures, mostly because of problems with the acetabular component. The revisions were carried out a mean of 26.2 months (1 to 76) after the initial operation and most of the patients (30) were female.

Malpositioning of the acetabular component resulted in 27 revisions, mostly because of excessive abduction (mean 69.9°; 56° to 98°) or insufficient or excessive anteversion. Seven patients had more than one reason for revision. The mean increase in the diameter of the component was 1.8 mm (0 to 4) when exchange was needed.

Malpositioning of the components was associated with metallosis and a high level of serum ions. The results of revision of the femoral component to a component with a modular head were excellent, but four patients had dislocation after revision and four required a further revision.


Bone & Joint 360
Vol. 3, Issue 1 | Pages 23 - 24
1 Feb 2014

The February 2014 Wrist & Hand Roundup360 looks at: simple debridement and ulnar-sided wrist pain; needle fasciotomy or collagenase injection; joint replacement in osteoarthritic knuckles; the Mannerfelt arthrodesis; scaphoid union rates with conservative treatment; the benefits of atorvastatin for muscle re-innervation after sciatic nerve transection; and complications of trapeziectomy.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 88 - 91
1 Nov 2013
Su EP Su SL

Surface hip replacement (SHR) is generally used in younger, active patients as an alternative conventional total hip replacement in part because of the ability to preserve femoral bone. This major benefit of surface replacement will only hold true if revision procedures of SHRs are found to provide good clinical results.

A retrospective review of SHR revisions between 2007 and 2012 was presented, and the type of revision and aetiologies were recorded. There were 55 SHR revisions, of which 27 were in women. At a mean follow-up of 2.3 years (0.72 to 6.4), the mean post-operative Harris hip score (HHS) was 94.8 (66 to 100). Overall 23 were revised for mechanical reasons, nine for impingement, 13 for metallosis, nine for unexplained pain and one for sepsis. Of the type of revision surgery performed, 14 were femoral-only revisions; four were acetabular-only revisions, and 37 were complete revisions.

We did not find that clinical scores were significantly different between gender or different types of revisions. However, the mean post-operative HHS was significantly lower in patients revised for unexplained pain compared with patients revised for mechanical reasons (86.9 (66 to 100) versus 99 (96 to 100); p = 0.029). There were two re-revisions for infection in the entire cohort.

Based on the overall clinical results, we believe that revision of SHR can have good or excellent results and warrants a continued use of the procedure in selected patients. Close monitoring of these patients facilitates early intervention, as we believe that tissue damage may be related to the duration of an ongoing problem. There should be a low threshold to revise a surface replacement if there is component malposition, rising metal ion levels, or evidence of soft-tissue abnormalities.

Cite this article: Bone Joint J 2013;95-B, Supple A:88–91.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1035 - 1039
1 Aug 2013
Ebreo D Bell PJ Arshad H Donell ST Toms A Nolan JF

Metal artefact reduction (MAR) MRI is now widely considered to be the standard for imaging metal-on-metal (MoM) hip implants. The Medicines and Healthcare Products Regulatory Agency (MHRA) has recommended cross-sectional imaging for all patients with symptomatic MoM bearings. This paper describes the natural history of MoM disease in a 28 mm MoM total hip replacement (THR) using MAR MRI. Inclusion criteria were patients with MoM THRs who had not been revised and had at least two serial MAR MRI scans. All examinations were reported by an experienced observer and classified as A (normal), B (infection) or C1–C3 (mild, moderate, severe MoM-related abnormalities). Between 2002 and 2011 a total of 239 MRIs were performed on 80 patients (two to four scans per THR); 63 initial MRIs (61%) were normal. On subsequent MRIs, six initially normal scans (9.5%) showed progression to a disease state; 15 (15%) of 103 THRs with sequential scans demonstrated worsening disease on subsequent imaging.

Most patients with a MoM THR who do not undergo early revision have normal MRI scans. Late progression (from normal to abnormal, or from mild to more severe MoM disease) is not common and takes place over several years.

Cite this article: Bone Joint J 2013;95-B:1035–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1019 - 1024
1 Aug 2009
Grammatopolous G Pandit H Kwon Y Gundle R McLardy-Smith P Beard DJ Murray DW Gill HS

Inflammatory pseudotumours occasionally occur after metal-on-metal hip resurfacing and often lead to revision. Our aim was to determine the severity of this complication by assessing the outcome of revision in these circumstances and by comparing this with the outcome of other metal-on-metal hip resurfacing revisions as well as that of matched primary total hip replacements.

We identified 53 hips which had undergone metal-on-metal hip resurfacing and required revision at a mean of 1.59 years (0.01 to 6.69) after operation. Of these, 16 were revised for pseudotumours, 21 for fracture and 16 for other reasons. These were matched by age, gender and diagnosis with 103 patients undergoing primary total hip replacement with the Exeter implant.

At a mean follow-up of three years (0.8 to 7.2) the outcome of metal-on-metal hip resurfacing revision for pseudotumour was poor with a mean Oxford hip score of 20.9 (sd 9.3) and was significantly worse (p < 0.001) than the outcome for fracture with a mean Oxford hip score of 40.2 (sd 9.2) or that for other causes with a mean Oxford hip score of 37.8 (sd 9.4). The clinical outcome of revision for pseudotumour was also significantly worse (p < 0.001) than the outcome of matched primary total hip replacements. By contrast, the outcome for fracture and other causes was not significantly different from that of matched primary total hip replacements (p = 0.065). After revision for pseudotumour there were three cases of recurrent dislocation, three of palsy of the femoral nerve, one of stenosis of the femoral artery and two of loosening of the component. Five hips required further revision. In three of these there was evidence of recurrent pseudotumour, and one is currently awaiting further revision. The incidence of major complications after revision for pseudotumour (50%) was significantly higher (p = 0.018) than that after revision for other causes (14%).

The outcome of revision for pseudotumour is poor and consideration should be given to early revision to limit the extent of the soft-tissue destruction. The outcome of resurfacing revision for other causes is good.