Several different designs of hemiarthroplasty are used to treat intracapsular fractures of the proximal femur, with large variations in costs. No clinical benefit of modular over monoblock designs has been reported in the literature. Long-term data are lacking. The aim of this study was to report the ten-year implant survival of commonly used designs of hemiarthroplasty. Patients recorded by the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) between 1 September 1999 and 31 December 2020 who underwent hemiarthroplasty for the treatment of a hip fracture with the following implants were included: a cemented monoblock Exeter Trauma Stem (ETS), cemented Exeter V40 with a bipolar head, a monoblock Thompsons prosthesis (Cobalt/Chromium or Titanium), and an Exeter V40 with a Unitrax head. Overall and age-defined cumulative revision rates were compared over the ten years following surgery.Aims
Methods
Dual-mobility acetabular components (DMCs) have improved total hip arthroplasty (THA) stability in femoral neck fractures (FNFs). In osteoarthritis, the direct anterior approach (DAA) has been promoted for improving early functional results compared with the posterolateral approach (PLA). The aim of this study was to compare these two approaches in FNF using DMC-THA. A prospective continuous cohort study was conducted on patients undergoing operation for FNF using DMC by DAA or PLA. Functional outcome was evaluated using the Harris Hip Score (HHS) and Parker score at three months and one year. Perioperative complications were recorded, and radiological component positioning evaluated.Aims
Methods
Distal femoral osteotomies (DFOs) are commonly used for the correction of valgus deformities and lateral compartment osteoarthritis. However, the impact of a DFO on subsequent total knee arthroplasty (TKA) function remains a subject of debate. Therefore, the purpose of this study was to determine the effect of a unilateral DFO on subsequent TKA function in patients with bilateral TKAs, using the contralateral knee as a self-matched control group. The inclusion criteria consisted of patients who underwent simultaneous or staged bilateral TKA after prior unilateral DFO between 1972 and 2023. The type of osteotomy performed, osteotomy hardware fixation, implanted TKA components, and revision rates were recorded. Postoperative outcomes including the Forgotten Joint Score-12 (FJS-12), Tegner Activity Scale score, and subjective knee preference were also obtained at final follow-up.Aims
Methods
Instability is a common cause of failure after total hip arthroplasty. A novel reverse total hip has been developed, with a femoral cup and acetabular ball, creating enhanced mechanical stability. The purpose of this study was to assess the implant fixation using radiostereometric analysis (RSA), and the clinical safety and efficacy of this novel design. Patients with end-stage osteoarthritis were enrolled in a prospective cohort at a single centre. The cohort consisted of 11 females and 11 males with mean age of 70.6 years (SD 3.5) and BMI of 31.0 kg/m2 (SD 5.7). Implant fixation was evaluated using RSA as well as Western Ontario and McMaster Universities Osteoarthritis Index, Harris Hip Score, Oxford Hip Score, Hip disability and Osteoarthritis Outcome Score, 38-item Short Form survey, and EuroQol five-dimension health questionnaire scores at two-year follow-up. At least one acetabular screw was used in all cases. RSA markers were inserted into the innominate bone and proximal femur with imaging at six weeks (baseline) and six, 12, and 24 months. Independent-samples Aims
Methods
Tribocorrosion at the head–neck taper interface
– so-called ‘taperosis’ – may be a source of metal ions and particulate
debris in metal-on-polyethylene total hip arthroplasty (THA). We examined the effect of femoral head length on fretting and
corrosion in retrieved head–neck tapers in vivo for a
minimum of two years (mean 8.7 years; 2.6 to 15.9). A total of 56
femoral heads ranging from 28 mm to 3 mm to 28 mm + 8 mm, and 17
femoral stems featuring a single taper design were included in the
study. Fretting and corrosion were scored in three horizontally
oriented concentric zones of each taper by stereomicroscopy. Head length was observed to affect fretting (p = 0.03), with
28 mm + 8 mm femoral heads showing greater total fretting scores
than all other head lengths. The central zone of the femoral head
bore taper was subject to increased fretting damage (p = 0.01),
regardless of head length or stem offset. High-offset femoral stems
were associated with greater total fretting of the bore taper (p
= 0.04). Increased fretting damage is seen with longer head lengths and
high-offset
Aseptic loosening is the most common cause of failure following cemented total knee arthroplasty (TKA), and has been linked to poor cementation technique. We aimed to develop a consensus on the optimal technique for component cementation in TKA. A UK-based, three-round, online modified Delphi Expert Consensus Study was completed focusing on cementation technique in TKA. Experts were identified as having a minimum of five years’ consultant experience in the NHS and fulfilling any one of the following criteria: a ‘high volume’ knee arthroplasty practice (> 150 TKAs per annum) as identified from the National joint Registry of England, Wales, Northern Ireland and the Isle of Man; a senior author of at least five peer reviewed articles related to TKA in the previous five years; a surgeon who is named trainer for a post-certificate of comletion of training fellowship in TKA.Aims
Methods
Obtaining solid implant fixation is crucial in revision total knee arthroplasty (rTKA) to avoid aseptic loosening, a major reason for re-revision. This study aims to validate a novel grading system that quantifies implant fixation across three anatomical zones (epiphysis, metaphysis, diaphysis). Based on pre-, intra-, and postoperative assessments, the novel grading system allocates a quantitative score (0, 0.5, or 1 point) for the quality of fixation achieved in each anatomical zone. The criteria used by the algorithm to assign the score include the bone quality, the size of the bone defect, and the type of fixation used. A consecutive cohort of 245 patients undergoing rTKA from 2012 to 2018 were evaluated using the current novel scoring system and followed prospectively. In addition, 100 first-time revision cases were assessed radiologically from the original cohort and graded by three observers to evaluate the intra- and inter-rater reliability of the novel radiological grading system.Aims
Methods
Dual-mobility (DM) components are increasingly used to prevent and treat dislocation after total hip arthroplasty (THA). Intraprosthetic dissociation (IPD) is a rare complication of DM that is believed to have decreased with contemporary implants. This study aimed to report incidence, treatment, and outcomes of contemporary DM IPD. A total of 1,453 DM components were implanted at a single academic institution between January 2010 and December 2021: 695 in primary and 758 in revision THA. Of these, 49 presented with a dislocation of the large DM head and five presented with an IPD. At the time of closed reduction of the large DM dislocation, six additional IPDs occurred. The mean age was 64 years (SD 9.6), 54.5% were female (n = 6), and mean follow-up was 4.2 years (SD 1.8). Of the 11 IPDs, seven had a history of instability, five had abductor insufficiency, four had prior lumbar fusion, and two were conversions for failed fracture management.Aims
Methods
The primary aim of this trial was to compare the subsidence of two similar hydroxyapatite-coated titanium femoral components from different manufacturers. Secondary aims were to compare rotational migration (anteversion/retroversion and varus/valgus tilt) and patient-reported outcome measures between both femoral components. Patients were randomized to receive one of the two femoral components (Avenir or Corail) during their primary total hip arthroplasty between August 2018 and September 2020. Radiostereometric analysis examinations at six, 12, and 24 months were used to assess the migration of each implanted femoral component compared to a baseline assessment. Patient-reported outcome measures were also recorded for these same timepoints. Overall, 50 patients were enrolled (62% male (n = 31), with a mean age of 65.7 years (SD 7.3), and mean BMI of 30.2 kg/m2 (SD 5.2)).Aims
Methods
Iliopsoas impingement occurs in 4% to 30% of patients after undergoing total hip arthroplasty (THA). Despite a relatively high incidence, there are few attempts at modelling impingement between the iliopsoas and acetabular component, and no attempts at modelling this in a representative cohort of subjects. The purpose of this study was to develop a novel computational model for quantifying the impingement between the iliopsoas and acetabular component and validate its utility in a case-controlled investigation. This was a retrospective cohort study of patients who underwent THA surgery that included 23 symptomatic patients diagnosed with iliopsoas tendonitis, and 23 patients not diagnosed with iliopsoas tendonitis. All patients received postoperative CT imaging, postoperative standing radiography, and had minimum six months’ follow-up. 3D models of each patient’s prosthetic and bony anatomy were generated, landmarked, and simulated in a novel iliopsoas impingement detection model in supine and standing pelvic positions. Logistic regression models were implemented to determine if the probability of pain could be significantly predicted. Receiver operating characteristic curves were generated to determine the model’s sensitivity, specificity, and area under the curve (AUC).Aims
Methods
Periprosthetic proximal femoral fractures (PFFs) are a major complication after total hip arthroplasty (THA). Health status after PFF is not specifically investigated. The aim of this study is to evaluate the health status pattern over two years after sustaining a PFF. A cohort of patients with PFF after THA was derived from the Brabant Injury Outcomes Surveillance (BIOS) study. The BIOS study, a prospective, observational, multicentre follow-up cohort study, was conducted to obtain data by questionnaires pre-injury and at one week, and one, three, six, 12, and 24 months after trauma. Primary outcome measures were the EuroQol five-dimension three-level questionnaire (EQ-5D-3L), the Health Utility Index 2 (HUI2), and the Health Utility Index 3 (HUI3). Secondary outcome measures were general measurements such as duration of hospital stay and mortality.Aims
Methods
We reviewed 25 patients in whom a MUTARS megaprosthesis with a conical fluted stem had been implanted. There were three types of stem: a standard stem was used in 17 cases (three in the proximal femur, nine in the distal femur and five proximal tibia), a custom-made proximal
This study aimed to assess the carbon footprint associated with total hip arthroplasty (THA) in a UK hospital setting, considering various components within the operating theatre. The primary objective was to identify actionable areas for reducing carbon emissions and promoting sustainable orthopaedic practices. Using a life-cycle assessment approach, we conducted a prospective study on ten cemented and ten hybrid THA cases, evaluating carbon emissions from anaesthetic room to recovery. Scope 1 and scope 2 emissions were considered, focusing on direct emissions and energy consumption. Data included detailed assessments of consumables, waste generation, and energy use during surgeries.Aims
Methods
Custom triflange acetabular components (CTACs) play an important role in reconstructive orthopaedic surgery, particularly in revision total hip arthroplasty (rTHA) and pelvic tumour resection procedures. Accurate CTAC positioning is essential to successful surgical outcomes. While prior studies have explored CTAC positioning in rTHA, research focusing on tumour cases and implant flange positioning precision remains limited. Additionally, the impact of intraoperative navigation on positioning accuracy warrants further investigation. This study assesses CTAC positioning accuracy in tumour resection and rTHA cases, focusing on the differences between preoperative planning and postoperative implant positions. A multicentre observational cohort study in Australia between February 2017 and March 2021 included consecutive patients undergoing acetabular reconstruction with CTACs in rTHA (Paprosky 3A/3B defects) or tumour resection (including Enneking P2 peri-acetabular area). Of 103 eligible patients (104 hips), 34 patients (35 hips) were analyzed.Aims
Methods
This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.Aims
Methods
Understanding spinopelvic mechanics is important for the success of total hip arthroplasty (THA). Despite significant advancements in appreciating spinopelvic balance, numerous challenges remain. It is crucial to recognize the individual variability and postoperative changes in spinopelvic parameters and their consequential impact on prosthetic component positioning to mitigate the risk of dislocation and enhance postoperative outcomes. This review describes the integration of advanced diagnostic approaches, enhanced technology, implant considerations, and surgical planning, all tailored to the unique anatomy and biomechanics of each patient. It underscores the importance of accurately predicting postoperative spinopelvic mechanics, selecting suitable imaging techniques, establishing a consistent nomenclature for spinopelvic stiffness, and considering implant-specific strategies. Furthermore, it highlights the potential of artificial intelligence to personalize care. Cite this article:
The aims of this study were to determine the incidence and factors for developing periprosthetic joint infection (PJI) following hemiarthroplasty (HA) for hip fracture, and to evaluate treatment outcome and identify factors associated with treatment outcome. A retrospective review was performed of consecutive patients treated for HA PJI at a tertiary referral centre with a mean 4.5 years’ follow-up (1.6 weeks to 12.9 years). Surgeries performed included debridement, antibiotics, and implant retention (DAIR) and single-stage revision. The effect of different factors on developing infection and treatment outcome was determined.Aims
Methods
This study aimed to investigate the risk of postoperative complications in COVID-19-positive patients undergoing common orthopaedic procedures. Using the National Surgical Quality Improvement Programme (NSQIP) database, patients who underwent common orthopaedic surgery procedures from 1 January to 31 December 2021 were extracted. Patient preoperative COVID-19 status, demographics, comorbidities, type of surgery, and postoperative complications were analyzed. Propensity score matching was conducted between COVID-19-positive and -negative patients. Multivariable regression was then performed to identify both patient and provider risk factors independently associated with the occurrence of 30-day postoperative adverse events.Aims
Methods
Revision total hip arthroplasty (THA) is challenging
when there is severe loss of bone in the proximal femur. The purpose
of this study was to evaluate the clinical and radiographic outcomes
of revision THA in patients with severe proximal femoral bone loss
treated with a fluted, tapered, modular femoral component. Between
January 1998 and December 2004, 92 revision THAs were performed
in 92 patients using a single fluted, tapered, modular femoral stem
design. Pre-operative diagnoses included aseptic loosening, infection
and peri-prosthetic fracture. Bone loss was categorised pre-operatively
as Paprosky types III-IV, or Vancouver B3 in patients with a peri-prosthetic
fracture. The mean clinical follow-up was 6.4 years (2 to 12). A
total of 47 patients had peri-operative complications, 27 of whom
required further surgery. However, most of these further operations
involved retention of a well-fixed
This meta-analysis and systematic review aimed to comprehensively investigate the effects of vitamin K supplementation on bone mineral density (BMD) at various sites and bone metabolism in middle-aged and older adults. The databases of PubMed, Web of Science, and Cochrane Library were thoroughly searched from inception to July 2023.Aims
Methods