Reconstruction of the acetabulum after failed total hip arthroplasty
(THA) can be a surgical challenge in the presence of severe bone
loss. We report the long-term survival of a porous tantalum revision
acetabular component, its radiological appearance and quality of
life outcomes. We reviewed the results of 46 patients who had undergone revision
of a failed acetabular component with a Paprosky II or III bone
defect and reconstruction with a hemispherical, tantalum acetabular
component, supplementary screws and a cemented polyethylene liner.Aims
Patients and Methods
This study reports the results of 38 total hip
arthroplasties (THAs) in 33 patients aged <
50 years, using the
JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component.
This represents an update of previous reports of the same cohort
at ten and 16 years, which were reported in 2004 and 2009, respectively.
We describe the survival, radiological and functional outcomes at
a mean follow-up of 21 years (17 to 25). Of the surviving 34 THAs,
one underwent femoral revision for peri-prosthetic fracture after
21 years, and one patient (one hip) was lost to follow-up. Using
aseptic loosening as the end-point, 12 hips (31.5%) needed acetabular
revision but none needed femoral revision, demonstrating 100% survival
(95% confidence interval 89 to 100). In young patients with high demands, the Furlong HAC–coated femoral
component gives excellent long-term results. Cite this article:
A possible solution for the management of proximal femoral bone
loss is a modular femoral endoprosthesis (EPR). Although the outcome
of EPRs in tumour surgery has been well described, the outcome of
their use in revision hip surgery has received less attention. The
aim of this study was to describe the outcome of using EPR for non-neoplastic
indications. A retrospective review of 79 patients who underwent 80 EPRs for
non-neoplastic indications was performed, including the rates of
complication and survival and the mean Oxford Hip Scores (OHS),
at a mean of five years post-operatively. The mean age at the time
of surgery was 69 years (28 to 93) and the mean number of previous operations
on the hip was 2.4 (0 to 17). The most common indications for EPR
implantation were periprosthetic joint infection (PJI) (n = 40),
periprosthetic fracture (n = 12) and failed osteosynthesis of a
proximal femoral fracture or complex trauma (n = 11).Objectives
Methods
The February 2015 Hip &
Pelvis Roundup360 looks at: Hip arthroplasty in Down syndrome; Bulk femoral autograft successful in acetabular reconstruction; Arthroplasty follow-up: is the internet the solution?; Total hip arthroplasty following acetabular fracture; Salvage arthroplasty following failed hip internal fixation; Bone banking sensible financially and clinically; Allogenic blood transfusion in arthroplasty.
We evaluated the accuracy with which a custom-made
acetabular component could be positioned at revision arthroplasty
of the hip in patients with a Paprosky type 3 acetabular defect. A total of 16 patients with a Paprosky type 3 defect underwent
revision surgery using a custom-made trabecular titanium implant.
There were four men and 12 women with a median age of 67 years (48
to 79). The planned inclination (INCL), anteversion (AV), rotation
and centre of rotation (COR) of the implant were compared with the post-operative
position using CT scans. A total of seven implants were malpositioned in one or more parameters:
one with respect to INCL, three with respect to AV, four with respect
to rotation and five with respect to the COR. To the best of our knowledge, this is the first study in which
CT data acquired for the pre-operative planning of a custom-made
revision acetabular implant have been compared with CT data on the
post-operative position. The results are encouraging. Cite this article:
To determine the effect of a change in design of a cementless
ceramic acetabular component in fixation and clinical outcome after
total hip arthroplasty We compared 342 hips (302 patients) operated between 1999 and
2005 with a relatively smooth hydroxyapatite coated acetabular component
(group 1), and 337 hips (310 patients) operated between 2006 and
2011 using a similar acetabular component with a macrotexture on
the entire outer surface of the component (group 2). The mean age of
the patients was 53.5 (14 to 70) in group 1 and 53.0 (15 to 70)
in group 2. The mean follow-up was 12.7 years (10 to 17) for group
1 and 7.2 years (4 to 10) for group 2.Aims
Patients and Methods
We compared the incidence of pseudotumours after
large head metal-on-metal (MoM) total hip arthroplasty (THA) with
that after conventional metal-on-polyethylene (MoP) THA and assessed
the predisposing factors to pseudotumour formation. From a previous randomised controlled trial which compared large
head (38 mm to 60 mm) cementless MoM THA with conventional head
(28 mm) cementless MoP THA, 93 patients (96 THAs: 41 MoM (21 males,
20 females, mean age of 64 years, standard deviation ( The incidence of pseudotumours, measured using a standardised
CT protocol was 22 (53.7%) after MoM THA and 12 (21.8%) after MoP
THA. Women with a MoM THA were more likely to develop a pseudotumour
than those with a MoP THA (15 Contrary to popular belief, pseudotumours occur frequently around
MoP THAs. Women with a MoM THA and an elevated cobalt level are
at greatest risk. In this study, pseudotumours had no effect on
the functional outcome after either large head MoM or conventional
MoP THA. Cite this article:
Data on early morbidity and complications after
revision total hip replacement (THR) are limited. The aim of this nationwide
study was to describe and quantify early morbidity after aseptic
revision THR and relate the morbidity to the extent of the revision
surgical procedure. We analysed all aseptic revision THRs from 1st
October 2009 to 30th September 2011 using the Danish National Patient
Registry, with additional information from the Danish Hip Arthroplasty
Registry. There were 1553 procedures (1490 patients) performed in
40 centres and we divided them into total revisions, acetabular
component revisions, femoral stem revisions and partial revisions.
The mean age of the patients was 70.4 years (25 to 98) and the median
hospital stay was five days (interquartile range 3 to 7). Within 90
days of surgery, the readmission rate was 18.3%, mortality rate
1.4%, re-operation rate 6.1%, dislocation rate 7.0% and infection
rate 3.0%. There were no differences in these outcomes between high-
and low-volume centres. Of all readmissions, 255 (63.9%) were due
to ‘surgical’ complications Cite this article:
A number of studies have reported satisfactory
results from the isolated revision of an acetabular component. However,
many of these studies reported only the short- to intermediate-term
results of heterogeneous bearing surfaces in a mixed age group. We present our experience of using a ceramic-on-ceramic (CoC)
bearing for isolated revision of an uncemented acetabular component
in 166 patients (187 hips) who were under the age of 50 years at
the time of revision. There were 78 men and 88 women with a mean
age of 47.4 years (28 to 49). The most common reason for revision
was polyethylene wear and acetabular osteolysis in 123 hips (66%),
followed by aseptic loosening in 49 hips (26%). We report the clinical and radiological outcome, complication
rate, and survivorship of this group. The mean duration of follow-up
was 15.6 years (11 to 19). The mean pre-operative Harris hip score was 33 points (1 to 58),
and improved to a mean of 88 points (51 to 100) at follow-up. The
mean pre-operative total Western Ontario and McMaster Universities
Osteoarthritis Index score was 63.2 (43 to 91) and improved to 19.8
points (9 to 61) post-operatively. Overall, 153 of 166 patients
(92%) were satisfied with their outcome. Kaplan–Meier survivorship
analysis, with revision or radiological evidence of implant failure
(13 patients, 8%) as end-points, was 92% at 15 years (95% confidence
interval 0.89 to 0.97). Isolated revision of a cementless acetabular component using
a CoC bearing gives good results in patients under 50 years of age. Cite this article:
Total hip replacement (THR) still is a rare intervention
in many African countries. In Burkina Faso it is not performed on
a regular basis. A visiting programme for THR was started in a district
hospital with no previous relevant experience. In this paper we
present an analysis of the surgical technical problems and peri-operative
complications of 152 THRs in 136 patients and three bipolar hemiarthroplasties
in three patients undertaken in this new programme with limited
orthopaedic equipment. There were 86 male and 53 female patients
with a mean age of 49 years (21 to 78). We identified 77 intra-operative
technical problems in 51 operations. There were 24 peri-operative complications
in 21 patients, 17 of which were bony in nature. So far, ten revision
THRs have been performed in nine patients. Regular analysis of the technical problems and complications
was used to improve quality, and we identified patient selection
adapted to the local circumstances as important to avoid complications.
Our reflections on the problems encountered in initiating such a
programme may be of help to other teams planning similar projects. Cite this article:
The increasing need for total hip replacement
(THR) in an ageing population will inevitably generate a larger number
of revision procedures. The difficulties encountered in dealing
with the bone deficient acetabulum are amongst the greatest challenges
in hip surgery. The failed acetabular component requires reconstruction
to restore the hip centre and improve joint biomechanics. Impaction
bone grafting is successful in achieving acetabular reconstruction
using both cemented and cementless techniques. Bone graft incorporation
restores bone stock whilst providing good component stability. We
provide a summary of the evidence and current literature regarding impaction
bone grafting using both cemented and cementless techniques in revision
THR. Cite this article:
We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid recurrent dislocation. Aims
Methods
Revision total hip replacement (THR) for young
patients is challenging because of technical complexity and the potential
need for subsequent further revisions. We have assessed the survivorship,
functional outcome and complications of this procedure in patients
aged <
50 years through a large longitudinal series with consistent treatment
algorithms. Of 132 consecutive patients (181 hips) who underwent
revision THR, 102 patients (151 hips) with a mean age of 43 years
(22 to 50) were reviewed at a mean follow-up of 11 years (2 to 26)
post-operatively. We attempted to restore bone stock with allograft
where indicated. Using further revision for any reason as an end point,
the survival of the acetabular component was 71% ( This overall perspective on the mid- to long-term results is
valuable when advising young patients on the prospects of revision
surgery at the time of primary replacement. Cite this article:
The practice of removing a well-fixed cementless
femoral component is associated with high morbidity. Ceramic bearing
couples are low wearing and their use minimises the risk of subsequent
further revision due to the production of wear debris. A total of
165 revision hip replacements were performed, in which a polyethylene-lined acetabular
component was revised to a new acetabular component with a ceramic
liner, while retaining the well-fixed femoral component. A titanium
sleeve was placed over the used femoral trunnion, to which a ceramic
head was added. There were 100 alumina and 65 Delta bearing couples
inserted. The mean Harris hip score improved significantly from 71.3 (9.0
to 100.0) pre-operatively to 91.0 (41.0 to 100.0) at a mean follow
up of 4.8 years (2.1 to 12.5) (p <
0.001). No patients reported
squeaking of the hip. There were two fractures of the ceramic head, both in alumina
bearings. No liners were seen to fracture. No fractures were observed
in components made of Delta ceramic. At 8.3 years post-operatively
the survival with any cause of failure as the endpoint was 96.6%
(95% confidence interval (CI) 85.7 to 99.3) for the acetabular component and
94.0% (95% CI 82.1 to 98.4) for the femoral component. The technique of revising the acetabular component in the presence
of a well-fixed femoral component with a ceramic head placed on
a titanium sleeve over the used trunnion is a useful adjunct in
revision hip practice. The use of Delta ceramic is recommended. Cite this article:
The Kaplan-Meier estimation is widely used in orthopedics to
calculate the probability of revision surgery. Using data from a
long-term follow-up study, we aimed to assess the amount of bias
introduced by the Kaplan-Meier estimator in a competing risk setting. We describe both the Kaplan-Meier estimator and the competing
risk model, and explain why the competing risk model is a more appropriate
approach to estimate the probability of revision surgery when patients
die in a hip revision surgery cohort. In our study, a total of 62 acetabular
revisions were performed. After a mean of 25 years, no patients
were lost to follow-up, 13 patients had undergone revision surgery
and 33 patients died of causes unrelated to their hip.Objectives
Methods
Periprosthetic femoral fracture (PFF) is a potentially
devastating complication after total hip arthroplasty, with historically
high rates of complication and failure because of the technical
challenges of surgery, as well as the prevalence of advanced age
and comorbidity in the patients at risk. This study describes the short-term outcome after revision arthroplasty
using a modular, titanium, tapered, conical stem for PFF in a series
of 38 fractures in 37 patients. The mean age of the cohort was 77 years (47 to 96). A total of
27 patients had an American Society of Anesthesiologists grade of
at least 3. At a mean follow-up of 35 months (4 to 66) the mean
Oxford Hip Score (OHS) was 35 (15 to 48) and comorbidity was significantly
associated with a poorer OHS. All fractures united and no stem needed
to be revised. Three hips in three patients required further surgery
for infection, recurrent PFF and recurrent dislocation and three
other patients required closed manipulation for a single dislocation.
One stem subsided more than 5 mm but then stabilised and required
no further intervention. In this series, a modular, tapered, conical stem provided a versatile
reconstruction solution with a low rate of complications. Cite this article:
The conventional method for reconstructing acetabular
bone loss at revision surgery includes using structural bone allograft.
The disadvantages of this technique promoted the advent of metallic
but biocompatible porous implants to fill bone defects enhancing
initial and long-term stability of the acetabular component. This
paper presents the indications, surgical technique and the outcome
of using porous metal acetabular augments for reconstructing acetabular
defects. Cite this article:
We hypothesised that the use of tantalum (Ta)
acetabular components in revision total hip arthroplasty (THA) was protective
against subsequent failure due to infection. We identified 966 patients
(421 men, 545 women and 990 hips) who had undergone revision THA
between 2000 and 2013. The mean follow up was 40.2 months (3 months
to 13.1 years). The mean age of the men and women was 62.3 years
(31 to 90) and 65.1 years (25 to 92), respectively. Titanium (Ti) acetabular components were used in 536 hips while
Ta components were used in 454 hips. In total, 73 (7.3%) hips experienced
subsequent acetabular failure. The incidence of failure was lower
in the Ta group at 4.4% (20/454) compared with 9.9% (53/536) in
the Ti group (p <
0.001, odds ratio 2.38; 95% CI 1.37 to 4.27).
Among the 144 hips (64 Ta, 80 Ti) for which revision had been performed
because of infection, failure due to a subsequent infection was
lower in the Ta group at 3.1% (2/64) compared with 17.5% (14/80)
for the Ti group (p = 0.006). Thus, the use of Ta acetabular components during revision THA
was associated with a lower incidence of failure from all causes
and Ta components were associated with a lower incidence of subsequent
infection when used in patients with periprosthetic joint infection. Cite this article: