We reviewed 123 second-generation uncemented total hip replacements performed on 115 patients by a single surgeon between 1993 and 1994. The acetabular component used in all cases was a fully porous-coated threaded hemispheric titanium shell (T-Tap ST) with a calcium ion stearate-free, isostatically compression-moulded polyethylene liner. The titanium femoral component used was a Taperloc with a reduced distal stem. No patient was lost to follow-up. Complete clinical and radiological follow-up was obtained for all 123 hips at a mean of 14 years (12 to 16). One femoral component was revised after a fracture, and three acetabular components for aseptic loosening. No additional femoral or acetabular components were judged loose by radiological criteria. Mild proximal femoral osteolysis was identified in two hips and minor acetabular osteolysis was present in four. The mean rate of penetration of the femoral head was 0.036 mm/year (0.000 to 0.227). These findings suggest that refinements in component design may be associated with excellent long-term fixation in cementless primary total hip replacement.
This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83). Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem. The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4). This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers.
New brands of joint prosthesis are released for general implantation with limited evidence of their long-term performance in patients. The CUSUM continuous monitoring method is a statistical testing procedure which could be used to provide prospective evaluation of brands as soon as implantation in patients begins and give early warning of poor performance. We describe the CUSUM and illustrate the potential value of this monitoring tool by applying it retrospectively to the 3M Capital Hip experience. The results show that if the clinical data and methodology had been available, the CUSUM would have given an alert to the underperformance of this prosthesis almost four years before the issue of a Hazard Notice by the Medical Devices Agency. This indicates that the CUSUM can be a valuable tool in monitoring joint prostheses, subject to timely and complete collection of data. Regional or national joint registries provide an opportunity for future centralised, continuous monitoring of all hip and knee prostheses using these techniques.
Periprosthetic joint infection (PJI) is a devastating
complication which can follow a total joint arthroplasty (TJA).
Although rare, this ongoing threat undermines the success of TJA,
a historically reputable procedure. It has haunted the orthopedic
community for decades and several ongoing studies have provided
insights and new approaches to effectively battle this multilayered
problem.
We identified 1305 femoral impaction bone grafting revisions using the Exeter stem performed between 1989 and 2002 in 30 hospitals throughout Sweden. There were 1188 patients with a mean age of 71 years (29 to 94) followed up for between five and 18 years. The participating departments reported 70 further revisions in total, of which 57 could also be identified on the Swedish National Arthroplasty Registry. Kaplan-Meier survivorship for all causes of failure was 94.0% (95% confidence interval (CI) 92 to 96) for women and 94.7% (95% CI, 92 to 96) for men at 15 years. Survivorship at 15 years for aseptic loosening was 99.1% (95% CI 98.4 to 99.5), for infection 98.6% (95% CI 97.6 to 99.2), for subsidence 99.0% (95% CI 98.2 to 99.4) and for fracture 98.7% (95% CI 97.9 to 99.2) Statistically significant predictors of failure were the year in which revision was conducted (p <
0.001). The number of previous revisions was slightly above the level of signifance (p = 0.056). Age, gender, the length of the stem and previous septic loosening were not predictors of failure (p = 0.213, p = 0.399, p = 0.337, p = 0.687, respectively). The difference in survivorship between high- and low-volume departments was only 3% at ten years. We conclude that impaction bone grafting with the Exeter stem has an excellent long-term survivorship following revision arthroplasty. The technique of impaction grafting appears to be reliable, can be learned rapidly and produces a predictably low incidence of aseptic loosening.
We evaluated the concentrations of chromium and cobalt ions in blood after metal-on-metal surface replacement arthroplasty using a wrought-forged, high carbon content chromium-cobalt alloy implant in 64 patients. At one year, mean whole blood ion levels were 1.61 μg/L (0.4 to 5.5) for chromium and 0.67 μg/L (0.23 to 2.09) for cobalt. The pre-operative ion levels, component size, female gender and the inclination of the acetabular component were inversely proportional to the values of chromium and/or cobalt ions at one year postoperatively. Other factors, such as age and level of activity, did not correlate with the levels of metal ions. We found that the levels of the ions in the serum were 1.39 and 1.37 times higher for chromium and cobalt respectively than those in the whole blood. The levels of metal ions obtained may be specific to the hip resurfacing implant and reflect its manufacturing process.
Using a computer-based quality assurance program, we analysed peri-operative data on 160 patients undergoing one-stage bilateral hip or knee arthroplasties under regional anaesthesia with routine anaesthetic monitoring and only using peripheral intravenous access for peri-operative safety. We monitored defined intra-operative adverse events such as hypotension, myocardial ischaemia, arrhythmias, hypovolaemia, hypertension and early post-operative complications. We also determined post-operative hip and knee function, and patient satisfaction with different aspects of the anaesthetic management. Those patients undergoing one-stage bilateral arthroplasties were matched according to a cross-stratification which used three variables (American Society of Anesthesiologists’ physical status scoring system, age and joint replaced) to patients undergoing unilateral hip or knee arthroplasties. Serious intra-operative adverse events were, with the exception of intra-operative hypotension, very infrequent in patients undergoing bilateral (nine adverse events) as well as unilateral arthroplasties (five adverse events). Early post-operative complications were also infrequent in both groups. However, the risks of receiving a heterologous blood transfusion (odds ratio 2.5; 95% confidence interval (CI) 1.3 to 5.0, estimated by exact conditional logistic regression) or vasoactive drugs (odds ratio 3.9; 95% CI 2.0 to 7.8) were significantly greater for patients undergoing bilateral operations. Patient satisfaction with anaesthesia was high; all patients who underwent the one-stage bilateral operation would choose the same anaesthetic technique again.
We present a comparison of patient-reported outcomes
(PROMs) in relation to patient age, in patients who had received
a total (TKR) or unicompartmental knee replacement (UKR). The outcome
was evaluated using the Oxford knee score (OKS), EuroQol (EQ-5D)
and satisfaction scores. Patients aged 65 to 84 years demonstrated
better pre-operative function scores than those aged <
65 years
(OKS, p = 0.03; EQ-5D, p = 0.048) and those aged
≥ 85 years (OKS, p = 0.03). Post-operative scores were comparable
across age groups, but a linear trend for greater post-operative
improvement in OKS and EQ-5D was seen with decreasing age (p <
0.033). The overall mean satisfaction score at six months was 84.9,
but those aged <
55 years exhibited a lower mean level of satisfaction (78.3)
compared with all other age groups (all p <
0.031). The cumulative
overall two-year revision rate was 1.3%. This study demonstrates
that good early outcomes, as measured by the OKS and EQ-5D, can
be anticipated following knee replacement regardless of the patient’s
age, although younger patients gain greater improvement. However,
the lower satisfaction in those aged
<
55 years is a concern, and suggests that outcome is not fully
encapsulated by the OKS and EQ-5D evaluation, and raises the question
whether the OKS alone is an appropriate measure of pain and function
in younger, more active individuals. Cite this article:
Progressive retroversion of a cemented stem is
predictive of early loosening and failure. We assessed the relationship
between direct post-operative stem anteversion, measured with CT,
and the resulting rotational stability, measured with repeated radiostereometric
analysis over ten years. The study comprised 60 cemented total hip
replacements using one of two types of matt collared stem with a
rounded cross-section. The patients were divided into three groups
depending on their measured post-operative anteversion (<
10°,
10° to 25°, >
25°). There was a strong correlation between direct
post-operative anteversion and later posterior rotation. At one
year the <
10° group showed significantly more progressive retroversion
together with distal migration, and this persisted to the ten-year
follow-up. In the <
10° group four of ten stems (40%) had been
revised at ten years, and an additional two stems (20%) were radiologically
loose. In the ‘normal’ (10° to 25°) anteversion group there was
one revised (3%) and one loose stem (3%) of a total of 30 stems,
and in the >
25° group one stem (5%) was revised and another loose (5%)
out of 20 stems. This poor outcome is partly dependent on the design
of this prosthesis, but the results strongly suggest that the initial
rotational position of cemented stems during surgery affects the
subsequent progressive retroversion, subsidence and eventual loosening.
The degree of retroversion may be sensitive to prosthetic design
and stem size, but <
10° of anteversion appears deleterious to
the long-term outcome for cemented hip prosthetic stems. Cite this article:
Periosteum is important for bone homoeostasis
through the release of bone morphogenetic proteins (BMPs) and their
effect on osteoprogenitor cells. Smoking has an adverse effect on
fracture healing and bone regeneration. The aim of this study was
to evaluate the effect of smoking on the expression of the BMPs
of human periosteum. Real-time polymerase chain reaction was performed
for BMP-2,-4,-6,-7 gene expression in periosteal samples obtained from
45 fractured bones (19 smokers, 26 non-smokers) and 60 non-fractured
bones (21 smokers, 39 non-smokers). A hierarchical model of BMP
gene expression (BMP-2 >
BMP-6 >
BMP-4 >
BMP-7) was demonstrated
in all samples. When smokers and non-smokers were compared, a remarkable
reduction in the gene expression of BMP-2, -4 and -6 was noticed
in smokers. The comparison of fracture and non-fracture groups demonstrated
a higher gene expression of BMP-2, -4 and -7 in the non-fracture
samples. Within the subgroups (fracture and non-fracture), BMP gene
expression in smokers was either lower but without statistical significance
in the majority of BMPs, or similar to that in non-smokers with
regard to BMP-4 in fracture and BMP-7 in non-fracture samples. In
smokers, BMP gene expression of human periosteum was reduced, demonstrating
the effect of smoking at the molecular level by reduction of mRNA
transcription of periosteal BMPs. Among the BMPs studied, BMP-2
gene expression was significantly
Radiostereometric analysis (RSA) can detect early
micromovement in unstable implant designs which are likely subsequently
to have a high failure rate. In 2010, the Articular Surface Replacement
(ASR) was withdrawn because of a high failure rate. In 19 ASR femoral
components, the mean micromovement over the first two years after implantation
was 0.107 mm ( We conclude that the ASR femoral component achieves initial stability
and that early migration is not the mode of failure for this resurfacing
arthroplasty.
The first 325 Exeter Universal stems (309 patients) implanted at the originating centre were inserted between March 1988 and February 1990 by a group of surgeons with differing experience. In this report we describe the clinical and radiological results at a mean of 15.7 years (14.7 to 17.3) after operation with no loss to follow-up. There were 97 patients (108 hips) with replacements still in situ and 31 (31 hips) who had undergone a further procedure. With an endpoint of revision for aseptic loosening, the survivorship at 17 years was 100% and 90.4% for the femoral and acetabular component, respectively. The mean Merle D’Aubigné and Postel scores at review were 5.4 (
Antibiotic prophylaxis is routinely administered during joint replacement surgery and may predispose patients to
Infection remains a significant and common complication after joint replacement and there is debate about which contributing factors are important. Few studies have investigated the effect of the operating time on infection. We collected data prospectively from 5277 hip and knee replacements which included the type of procedure, the operating time, the use of drains, the operating theatre, surgeon, age and gender. In a subgroup of 3449 knee replacements further analysis was carried out using the tourniquet time in place of the operating time. These variables were assessed by the use of generalised linear modelling against superficial, deep or joint-space post-operative infection as defined by the Australian Surgical-Site Infection criteria. The overall infection rate was 0.98%. In the replacement data set both male gender (z = 3.097, p = 0.00195) and prolonged operating time (z = 4.325, p <
0.001) were predictive of infection. In the knee subgroup male gender (z = 2.250, p = 0.02447), a longer tourniquet time (z = 2.867, p = 0.00414) and total knee replacement ( These findings support the view that a prolonged operating time and male gender are associated with an increased incidence of infection. Steps to minimise intra-operative delay should be instigated, and care should be exercised when introducing measures which prolong the duration of joint replacement.
The inherent challenges of total hip replacement
(THR) in children include the choice of implant for the often atypical
anatomical morphology, its fixation to an immature growing skeleton
and the bearing surface employed to achieve a successful long-term
result. We report the medium-term results of 52 consecutive uncemented
THRs undertaken in
35 paediatric patients with juvenile idiopathic arthritis. The mean
age at the time of surgery was 14.4 years (10 to 16). The median
follow-up was 10.5 years (6 to 15). During the study period 13 THRs
underwent revision surgery. With revision as an endpoint, subgroup
analysis revealed 100% survival of the 23 ceramic-on-ceramic THRs
and 55% (16 of 29) of the metal- or ceramic-on-polyethylene. This
resulted in 94% (95% CI 77.8 to 98.4) survivorship of the femoral
component and 62% (95% CI 41.0 to 78.0) of the acetabular component.
Revision of the acetabular component for wear and osteolysis were
the most common reasons for failure accounting for 11 of the 13
revisions. The success seen in patients with a ceramic-on-ceramic articulation
seems to indicate that this implant strategy has the potential to
make a major difference to the long-term outcome in this difficult
group of patients.
We report a systematic review and meta-analysis of published randomised controlled trials evaluating the efficacy of tranexamic acid (TXA) in reducing blood loss and transfusion in total hip replacement (THR). The data were evaluated using the generic evaluation tool designed by the Cochrane Bone, Joint and Muscle Trauma Group. We identified 11 clinical trials which were suitable for detailed extraction of data. There were no trials that used TXA in revision THR. A total of seven studies (comprising 350 patients) were eligible for the blood loss outcome data. The use of TXA reduced intra-operative blood loss by a mean of 104 ml (95% confidence interval (CI) −164 to −44, p = 0.0006, heterogeneity I2 0%), postoperative blood loss by a mean of 172 ml (95% CI −263 to −81, p = 0.0002, heterogeneity I2 63%) and total blood loss by a mean of 289 ml (95% CI −440 to −138, p <
0.0002, heterogeneity I2 54%). TXA led to a significant reduction in the proportion of patients requiring allogeneic blood transfusion (risk difference −0.20, 95% CI −0.29 to −0.11, p <
0.00001, I2 15%). There were no significant differences in deep-vein thrombosis, pulmonary embolism, infection rates or other complications among the study groups.
Systemic mastocytosis is a rare condition that often involves the bone marrow. We report the case of a patient with systemic mastocytosis who underwent total hip replacement. Technical difficulties encountered during the procedure included a narrow medullary canal and abnormally hard bone, later confirmed by laboratory measurements. Follow-up at five years showed a good clinical and radiological outcome.
The August 2012 Knee Roundup360 looks at: meniscal defects and a polyurethane scaffold; which is best between a single or double bundle; OA of the knee; how to resolve anterior knee pain; whether yoga can be bad for your menisci; metal ions in the serum; whether ACI is any good; the ACL; whether hyaluronic acid delays collagen degradation; and hyaluronan and patellar tendinopathy.
Our aim in this pilot study was to evaluate the fixation of, the bone remodelling around, and the clinical outcome after surgery of a new, uncemented, fully hydroxyapatite-coated, collared and tapered femoral component, designed specifically for elderly patients with a fracture of the femoral neck. We enrolled 50 patients, of at least 70 years of age, with an acute displaced fracture of the femoral neck in this prospective single-series study. They received a total hip replacement using the new component and were followed up regularly for two years. Fixation was evaluated by radiostereometric analysis and bone remodelling by dual-energy x-ray absorptiometry. Hip function and the health-related quality of life were assessed using the Harris hip score and the EuroQol-5D. Up to six weeks post-operatively there was a mean subsidence of 0.2 mm (−2.1 to +0.5) and a retroversion of a mean of 1.2° (−8.2° to +1.5°). No component migrated after three months. The patients had a continuous loss of peri-prosthetic bone which amounted to a mean of 16% (−49% to +10%) at two years. The mean Harris hip score was 82 (51 to 100) after two years. The two-year results from this pilot study indicate that this new, uncemented femoral component can be used for elderly patients with osteoporotic fractures of the femoral neck.