Total hip replacement (THR) is a very common
procedure undertaken in up to 285 000 Americans each year. Patient
satisfaction with THR is very high, with improvements in general
health, quality of life, and function while at the same time very
cost effective. Although the majority of patients have a high degree
of satisfaction with their THR, 27% experience some discomfort,
and up to 6% experience severe chronic pain. Although it can be
difficult to diagnose the cause of the pain in these patients, this
clinical issue should be approached systematically and thoroughly.
A detailed history and clinical examination can often provide the
correct diagnosis and guide the appropriate selection of investigations, which
will then serve to confirm the clinical diagnosis made. Cite this article:
The orientation of the acetabular component can
influence both the short- and long-term outcomes of total hip replacement
(THR). We performed a prospective, randomised, controlled trial
of two groups, comprising of 40 patients each, in order to compare
freehand introduction of the component with introduction using the transverse
acetabular ligament (TAL) as a reference for anteversion. Anteversion
and inclination were measured on pelvic radiographs. With respect to anteversion, in the freehand group 22.5% of the
components were outside the safe zone With respect to inclination, in the freehand group 37.5% of the
components were outside the safe zone The transverse acetabular ligament may be used to obtain the
appropriate anteversion when introducing the acetabular component
during THR, but not acetabular component inclination. Cite this article:
The hip joint is commonly involved in multiple epiphyseal dysplasia and patients may require total hip replacement before the age of 30 years. We retrospectively reviewed nine patients (16 hips) from four families. The diagnosis of multiple epiphyseal dysplasia was based on a family history, genetic counselling, clinical features and radiological findings. The mean age at surgery was 32 years (17 to 63), with a mean follow-up of 15.9 years (5.5 to 24). Of the 16 hips, ten required revision at a mean of 12.5 years (5 to 15) consisting of complete revision of the acetabular component in three hips and isolated exchange of the liner in seven. No femoral component has loosened or required revision during the period of follow-up. With revision for any reason, the 15-year survival was only 11.4% (95% confidence interval 1.4 to 21.4). However, when considering revision of the acetabular shell in isolation the survival at ten years was 93.7% (95% confidence interval 87.7 to 99.7), reducing to 76.7% (95% confidence interval 87.7 to 98.7) at 15 and 20 years, respectively.
Conventional uncemented femoral implants provide
dependable long-term fixation in patients with a wide range of functional
requirements. Yet challenges associated with proximal–distal femoral
dimensional mismatch, preservation of bone stock, and minimally
invasive approaches have led to exploration into alternative implant designs.
Short stem designs focusing on a stable metaphyseal fit have emerged
to address these issues in total hip replacement (THR). Uncemented
metaphyseal-engaging short stem implants are stable and are associated
with proximal bone remodeling closer to the metaphysis when compared
with conventional stems and they also have comparable clinical performances.
Short stem metaphyseal-engaging implants can meet the goals of a
successful THR, including tolerating a high level of patient function,
as well as durable fixation. Cite this article:
In a time of limited resources, the debate continues
over which types of hip prosthesis are clinically superior and more
cost-effective. Orthopaedic surgeons increasingly need robust economic
evidence to understand the full value of the operation, and to aid
decision making on the ‘package’ of procedures that are available
and to justify their practice beyond traditional clinical preference. In this paper we explore the current economic debate about the
merits of cemented and cementless total hip replacement, an issue
that continues to divide the orthopaedic community. Cite this article:
The use of plate-and-cable constructs to treat periprosthetic fractures around a well-fixed femoral component in total hip replacements has been reported to have high rates of failure. Our aim was to evaluate the results of a surgical treatment algorithm to use these lateral constructs reliably in Vancouver type-B1 and type-C fractures. The joint was dislocated and the stability of the femoral component was meticulously evaluated in 45 type-B1 fractures. This led to the identification of nine (20%) unstable components. The fracture was considered to be suitable for single plate-and-cable fixation by a direct reduction technique if the integrity of the medial cortex could be restored. Union was achieved in 29 of 30 fractures (97%) at a mean of 6.4 months (3 to 30) in 29 type-B1 and five type-C fractures. Three patients developed an infection and one construct failed. Using this algorithm plate-and-cable constructs can be used safely, but indirect reduction with minimal soft-tissue damage could lead to shorter times to union and lower rates of complications.
We describe 129 consecutive revision total hip replacements using a Charnley-Kerboull femoral component of standard length with impaction allografting. The mean follow-up was 8.2 years (2 to 16). Additionally, extramedullary reinforcement was performed using struts of cortical allograft in 49 hips and cerclage wires in 30. There was one intra-operative fracture of the femur but none later. Two femoral components subsided by 5 mm and 8 mm respectively, and were considered to be radiological failures. No further revision of a femoral component was required. The rate of survival of the femoral component at nine years, using radiological failure as the endpoint, was 98%. Our study showed that impaction grafting in association with a Charnley-Kerboull femoral component has a low rate of subsidence. Reconstruction of deficiencies of distal bone with struts of cortical allograft appeared to be an efficient way of preventing postoperative femoral fracture for up to 16 years.
Down’s syndrome is associated with a number of
musculoskeletal abnormalities, some of which predispose patients
to early symptomatic arthritis of the hip. The purpose of the present
study was to review the general and hip-specific factors potentially
compromising total hip replacement (THR) in patients with Down’s
syndrome, as well as to summarise both the surgical techniques that
may anticipate the potential adverse impact of these factors and
the clinical results reported to date. A search of the literature
was performed, and the findings further informed by the authors’
clinical experience, as well as that of the hip replacement in Down
Syndrome study group. The general factors identified include a high
incidence of ligamentous laxity, as well as associated muscle hypotonia
and gait abnormalities. Hip-specific factors include: a high incidence
of hip dysplasia, as well as a number of other acetabular, femoral
and combined femoroacetabular anatomical variations. Four studies
encompassing 42 hips, which reported the clinical outcomes of THR
in patients with Down’s syndrome, were identified. All patients
were successfully treated with standard acetabular and femoral components.
The use of supplementary acetabular screw fixation to enhance component
stability was frequently reported. The use of constrained liners
to treat intra-operative instability occurred in eight hips. Survival
rates of between 81% and 100% at a mean follow-up of 105 months
(6 to 292) are encouraging. Overall, while THR in patients with
Down’s syndrome does present some unique challenges, the overall
clinical results are good, providing these patients with reliable
pain relief and good function. Cite this article:
With greater numbers of younger patients undergoing
total hip replacement (THR), the effect of patient age on the diameter
of the femoral canal may become more relevant. This study aimed
to investigate the relationship between the diameter of the diaphysis
of the femoral canal with increasing age in a large number of patients
who underwent THR. A total of 1685 patients scheduled for THR had
their femoral dimensions recorded from calibrated radiographs. There
were 736 males and 949 females with mean ages of 67.1 years (34
to 92) and 70.2 years (29 to 92), respectively. The mean diameter
of the femoral canal was 13.3 mm (8.0 to 23.0) for males and 12.7
mm (6.0 to 26.0) for females. There was a poor correlation between
age and the diameter of the canal in males (r = 0.071, p = 0.05)
but a stronger correlation in females (r = 0.31, p <
0.001). The diameter of the femoral canal diameter of a female patient
undergoing THR could be predicted to increase by 3.2 mm between
the ages of 40 and 80 years, in contrast a male would be expected
to experience only a 0.6 mm increase during the same period. This
increase in the diameter of the canal with age might affect the
long-term survival of the femoral component in female patients. Cite this article:
We present the extended follow-up (≥ 20 years)
of a series of fully hydroxyapatite-coated femoral components used in
72 primary total hip replacements (THRs). Earlier results of this
cohort have been previously published. All procedures were performed
between 1986 and 1991. The series involved 45 women and 15 men with
12 bilateral procedures. Their mean age at the time of surgery was
60 years (46 to 80) and the mean duration of follow-up was 22.5
years (20 to 25). At final follow-up, the mean Merle d’Aubigné and
Postel hip scores were 5.5 (4.5 to 6), 3.8 (3.5 to 5) and 3.3 (3.0
to 5.0) for pain, mobility and function, respectively. Of the patients
92% were very satisfied at the time of final follow-up. There were seven revisions: six of the acetabular component for
aseptic loosening and one of both the stem and the acetabular component
for loosening due to deep infection. The survival of this prosthesis
at 22.5 years with revision for any reason as the endpoint was 91.7%
(95% confidence interval (CI) 84 to 99). Survival with aseptic loosening
of the stem as the endpoint was 100% (95% CI 90 to 100). This prosthesis provides pain relief in the long term. Survival
of this component is comparable to the best results for primary
THR with any means of fixation. Cite this article:
It has been suggested that the wear of ultra-high molecular weight polyethylene (UHMWPE) in total hip replacement is substantially reduced when the femoral head is ceramic rather than metal. However, studies of alumina and zirconia ceramic femoral heads on the penetration of an UHMWPE liner The purpose of this study was to examine the surface characteristics of 30 alumina and 24 zirconia ceramic femoral heads and to identify any phase transformation in the zirconia heads. We also studied the penetration rate of alumina and zirconia heads into contemporary UHMWPE liners. The alumina heads had been implanted for a mean of 11.3 years (8.1 to 16.2) and zirconia heads for a mean of 9.8 years (7.5 to 15). The mean surface roughness values of the explanted alumina heads (Ra 40.12 nm and Rpm 578.34 nm) were similar to those for the explanted zirconia heads (Ra 36.21 nm and Rpm 607.34 nm). The mean value of the monoclinic phase of two control zirconia heads was 1% (0.8% to 1.5%) and 1.2% (0.9% to 1.3%), respectively. The mean value of the monoclinic phase of 24 explanted zirconia heads was 7.3% (1% to 26%). In the alumina group, the mean linear penetration rate of the UMWPE liner was 0.10 mm/yr (0.09 to 0.12) in hips with low Ra and Rpm values (13.22 nm and 85.91 nm, respectively). The mean linear penetration rate of the UHMWPE liner was 0.13 mm/yr (0.07 to 0.23) in hips with high Ra and Rpm values (198.72 nm and 1329 nm, respectively). This difference was significant (p = 0.041). In the zirconia head group, the mean linear penetration rate of the UHMWPE liner was 0.09 mm/yr (0.07 to 0.14) in hips with low Ra and Rpm values (12.78 nm and 92.99 nm, respectively). The mean linear penetration rate of the UHMWPE liner was 0.12 mm/yr (0.08 to 0.22) in hips with high Ra and Rpm values (199.21 nm and 1381 nm, respectively). This difference was significant (p = 0.039). The explanted zirconia heads which had a minimal phase transformation had similar surface roughness and a similar penetration rate of UHMWPE liner as the explanted alumina head.
The purpose of this paper is to discuss the risk
factors, prevention strategies, classification, and treatment of
intra-operative femur fractures sustained during primary and revision
total hip arthroplasty.
In this study we present our experience with
four generations of uncemented total knee arthroplasty (TKA) from Smith
&
Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing
on the failure rates correlating with each design change. Beginning
in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix
were implanted by the senior author. The rate of revision for loosening
was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the
Tricon 2 with a HA coated tibial component, and 1.3% for the Profix
TKA. No loosening of the femoral component was seen with the Tricon
M, Tricon LS or Tricon 2, with no loosening seen of the tibial component
with the Profix TKA. Regarding revision for wear, the incidence
was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the
Tricon 2, and 0% for the Profix. These results demonstrate that
improvements in the design of uncemented components, including increased
polyethylene thickness, improved polyethylene quality, and the introduction
of hydroxyapatite coating, has improved the outcomes of uncemented
TKA over time.
We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component. All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant. This pattern of subsidence is regarded as being beneficial for uncemented fixation.
The ideal bearing surface for young patients
undergoing total hip replacement (THR) remains controversial. We report
the five-year results of a randomised controlled trial comparing
the clinical and radiological outcomes of 102 THRs in 91 patients
who were <
65 years of age. These patients were randomised to
receive a cobalt–chrome on ultra-high-molecular-weight polyethylene,
cobalt–chrome on highly cross-linked polyethylene, or a ceramic-on-ceramic
bearing. In all, 97 hip replacements in 87 patients were available
for review at five years. Two hips had been revised, one for infection
and one for peri-prosthetic fracture. At the final follow-up there were no significant differences
between the groups for the mean Western Ontario and McMaster Universities
osteoarthritis index (pain, p = 0.543; function, p = 0.10; stiffness,
p = 0.99), Short Form-12 (physical component, p = 0.878; mental
component, p = 0.818) or Harris hip scores (p = 0.22). Radiological
outcomes revealed no significant wear in the ceramic group. Comparison of standard and highly cross-linked polyethylene,
however, revealed an almost threefold difference in the mean annual
linear wear rates (0.151 mm/year
Orientation of the native acetabular plane as defined by the transverse acetabular ligament (TAL) and the posterior labrum was measured intra-operatively using computer-assisted navigation in 39 hips. In order to assess the influence of alignment on impingement, the range of movement was calculated for that defined by the TAL and the posterior labrum and compared with a standard acetabular component position (abduction 45°/anteversion 15°). With respect to the registration of the plane defined by the TAL and the posterior labrum, there was moderate interobserver agreement (r = 0.64, p <
0.001) and intra-observer reproducibility (r = 0.73, p <
0.001). The mean acetabular component orientation achieved was abduction of 41° (32° to 51°) and anteversion of 18° (−1° to 36°). With respect to the Lewinnek safe zone (abduction 40° ±10°, anteversion 15° ±10°), 35 of the 39 acetabular components were within this zone. However, there was no improvement in the range of movement (p = 0.94) and no significant difference in impingement (p = 0.085). Alignment of the acetabular component with the TAL and the posterior labrum might reduce the variability of acetabular component placement in total hip replacement. However, there is only a moderate interobserver agreement and intra-observer reliability in the alignment of the acetabular component using the TAL and the posterior labrum. No reduction in impingement was found when the acetabular component was aligned with the TAL and the posterior labrum, compared with a standard acetabular component position.
The purpose of this study was to define immediate post-operative ‘quality’ in total hip replacements and to study prospectively the occurrence of failure based on these definitions of quality. The evaluation and assessment of failure were based on ten radiological and clinical criteria. The cumulative summation (CUSUM) test was used to study 200 procedures over a one-year period. Technical criteria defined failure in 17 cases (8.5%), those related to the femoral component in nine (4.5%), the acetabular component in 32 (16%) and those relating to discharge from hospital in five (2.5%). Overall, the procedure was considered to have failed in 57 of the 200 total hip replacements (28.5%). The use of a new design of acetabular component was associated with more failures. For the CUSUM test, the level of adequate performance was set at a rate of failure of 20% and the level of inadequate performance set at a failure rate of 40%; no alarm was raised by the test, indicating that there was no evidence of inadequate performance. The use of a continuous monitoring statistical method is useful to ensure that the quality of total hip replacement is maintained, especially as newer implants are introduced.
Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed.
We evaluated all cases involving the combined use of a subtrochanteric derotational femoral shortening osteotomy with a cemented Exeter stem performed at our institution. With severe developmental dysplasia of the hip an osteotomy is often necessary to achieve shortening and derotation of the proximal femur. Reduction can be maintained with a 3.5 mm compression plate while the implant is cemented into place. Such a plate was used to stabilise the osteotomy in all cases. Intramedullary autograft helps to prevent cement interposition at the osteotomy site and promotes healing. There were 15 female patients (18 hips) with a mean age of 51 years (33 to 75) who had a Crowe IV dysplasia of the hip and were followed up for a mean of 114 months (52 to 168). None was lost to follow-up. All clinical scores were collected prospectively. The Charnley modification of the Merle D’Aubigné-Postel scores for pain, function and range of movement showed a statistically significant improvement from a mean of 2.4 (1 to 4), 2.3 (1 to 4), 3.4 (1 to 6) to 5.2 (3 to 6), 4.4 (3 to 6), 5.2 (4 to 6), respectively. Three acetabular revisions were required for aseptic loosening; one required femoral revision for access. One osteotomy failed to unite at 14 months and was revised successfully. No other case required a femoral revision. No postoperative sciatic nerve palsy was observed. Cemented Exeter femoral components perform well in the treatment of Crowe IV dysplasia with this procedure.
We prospectively evaluated the long-term outcome of 158 consecutive patients who underwent revision total hip replacement using uncemented computer-assisted design-computer-assisted manufacture femoral components. There were 97 men and 61 women. Their mean age was 63.1 years (34.6 to 85.9). The mean follow-up was 10.8 years (10 to 12). The mean Oxford, Harris and Western Ontario and McMaster hip scores improved from 41.1, 44.2 and 52.4 pre-operatively to 18.2, 89.3 and 12.3, respectively (p <
0.0001, for each). Six patients required further surgery. The overall survival of the femoral component was 97% (95% confidence interval 94.5 to 99.7). These results are comparable to those of previously published reports for revision total hip replacement using either cemented or uncemented components.